AeroNOx Nitric Oxide Titration & Monitoring System delivers nitric oxide gas and
measures nitric oxide and nitrogen dioxide gas concentrations in parts per million (ppm).
“Caution: U. S. Federal and Canadian law restricts this device to sale by or on the order
of a physician or other licensed medical practitioner. Outside Canada and the U. S.,
check with local laws for applicable restrictions.”
“Persons using this device should be trained and experienced in the use of this device to
assure effective administration of nitric oxide, and to avoid injury to the patient or to
others resulting from inhalation of excess nitric oxide, nitrogen dioxide, or other reaction
products.”
“Persons using this device who may be particularly sensitive to nitric oxide or nitrogen
dioxide, or who may be exposed to these gases for prolonged periods as a result of the
use of this device, should be aware that the AeroNOx does not scavenge the exhaust
gas, and that this gas is vented from the underside of the AeroNOx, or through the side
vent in the event that the bottom port is occluded. Ambient concentrations of nitric oxide
or nitrogen dioxide expected to result from the use of this device are less than 50 ppb.”
WARNING
The use of oxygen increases the danger of fire. Auxiliary equipment producing sparks should
not be placed near the AeroNOx. Small quantities such as ether or alcohol left near the
AeroNOx can cause fire.
Classification
According to the standard EN60601-1 of the International Electrotechnical Commission,
Medical electrical equipment, Part 1: General requirements for safety, the AeroNOx is
classified as follows:
Class I / Internally Powered, according to the type of protection against electric
shock.
Ordinary, according to the degree of protection against harmful ingress of water.
Continuous operation for the mode of operation.
Document No. 715-7000Revision E Page 5 of 63
SECTION II - UNPACKING AND SET-UP INSTRUCTIONS
Verify that the shipping carton (s) contain one of the following groups of equipment. Complete
descriptions of these groupings can be found in this section.
1. AeroNOx
Nitric Oxide Titration & Monitoring System
2. AeroNOx
Nitric Oxide Titration & Monitoring System
Transport, P/N 731-9148
Bedside Option
Document No. 715-7000Revision E Page 6 of 63
The AeroNOx Transport P/N 731-9148
Nitric Oxide Titration & Monitoring System
Component Part Number Quantity
AeroNOx
SS Hose with Quick Connect (3 ft.) 731-9371 1
AeroNOx Operating Manual
AeroNOx Technical Manual
12 V Universal Power Supply 293-0040 1
Medical Grade Power Cord 711-0179 1
SS NO Transport Delivery Regulators, CGA 626 731-9142 2
AeroNOx “NO WORRIES” Connector Sample Pak
AeroNOx Calibration Circuit
Single Calibration Screwdriver 416-0010 1
731-9138 1
715-7000 1
715-7001 1
731-9373 1
415-0000 1
The AeroNOx must be used with calibration gas and
accessories specified by the manufacturer.
Symbols
The following symbols appear in the AeroNOx documentation and labels. These
internationally recognized symbols are defined by the International Electrotechnical
Commission, IEC 417A and IEC 878 or by the Institute of Electrical and Electronics
Engineers, 315(a).
Direct current
Positive to center pin, Negative to outer ring of connection
Recycle or dispose of properly, contains sealed lead batteries
Attention, consult accompanying documents
Document No. 715-7000Revision E Page 7 of 63
The AeroNOx Transport P/N 731-9148
Nitric Oxide Titration & Monitoring System
Prepare the AeroNOx Transport Nitric Oxide
Titration & Monitoring System
for initial operation as follows:
1. Unpack the AeroNOx and inspect for damage.
2. Install battery as per SECTION III, “Battery”.
3. Unpack 12 V universal power supply (P/N 293-0040) and Medical Grade Power Cord
(P/N 711-0179). Plug in AeroNOx and charge for ~ 24 hrs.
4. Calibrate AeroNOx. (See SECTION V.)
5. Attach CGA 626 regulators to the transport tanks as per “Purge procedures to follow when changing tanks and/or regulators” in this section.
6. Connect selected regulator outlet to the AeroNOx gas inlet using the stainless steel hose
with quick connect adapters.
7. Turn on nitric oxide gas supply.
8. Attach the delivery line from the AeroNOx
delivery line (P/N 415-0001) found in the AeroNOx
Bagger (See Diagram # 2) or the AeroNOx
“NO WORRIES” connector sample
pak to the female luer connector marked “NITRIC OXIDE DELIVERY OUTLET” on the
AeroNOx. Tee the other end of the delivery line into the patient breathing circuit at least
30 - 40 cm upstream of the sampling site (see Diagram # 1).
9. Attach the “NO WORRIES” sample line (P/N 415-0004) to the male luer connector
marked “ANALYSIS SAMPLE INLET” on the AeroNOx.
10. Select the appropriate sample gas connector from your “NO WORRIES” connector
sample pak and attach it to the free end of the sample line. *NB if attaching the sample
line to the bagger, skip this step.
11. Connect the sample gas connector to the patient breathing circuit. The connector should
be placed in the inspiratory limb, just prior to the patient wye connection (refer to Diagram
# 1).
OR
Connect the sample line to the sampling port on the bagger (see Diagram # 2).
12. “Flush” your system according to “Suggested flush procedure for the delivery portion of
the AeroNOx
system”, or according to the instructions for the AeroNOx Bagger.
13. You are now ready to deliver NO gas and sample mixed inspired gases.
CAUTION!
Never turn on the NO delivery gas without first turning on the
ventilator or bagger flow. Failure to do this will result in undiluted
delivery gas entering the sampling chamber and exposing the
sensors to levels of NO and/or NO
which may damage the
2
sensors.
CAUTION!
NO2 gas may have collected in the AeroNOx delivery circuitry. Prior
to delivering gas to the patient, “flush” your delivery system with
fresh NO delivery gas. This should be repeated every time you begin
NO gas administration. Analyze the gas for high NO
levels prior to
2
patient connection.
Document No. 715-7000Revision E Page 8 of 63
Suggested “flush” procedure for the delivery portion of the AeroNOx system:
1. Set up the AeroNOx for delivery as outlined in “AeroNOx Transport Nitric Oxide
Titration & Monitoring System for initial operation”.
2. Hook up the ventilator to a test lung (patient should be manually ventilated during this
procedure and the ventilator parameters should be set to the patient’s pre-NO gas
settings).
3. Inactivate the “Safety Shut Off” system by simultaneously depressing the NO and NO2
alarm silence buttons for 5 seconds or until the display reads “SAFETY OFF”. For more
information on the “Safety Shut Off” system, see “Safety Shut Off” in SECTION III.
4. Turn on the NO gas flow to 2 lpm by turning the knob on the precision metering valve in a
counter-clockwise direction. This knob is located in the bottom right hand quadrant of the
front panel of the AeroNOx below the LCD flow display. Allow the gas to flow into the
ventilator circuit.
5. As soon as the NO2 value has dropped to a stable level, allow the gas to run for an
additional 3 seconds and then adjust the NO flow to the desired level (see SECTION VII Calculations for NO Delivery to determine flow required for desired NO concentration).
6. Re-activate the “Safety Shut Off” system by simultaneously depressing the NO and NO2
alarm silence buttons for at least 5 seconds or until the display reads “SAFETY ON”.
7. Once the desired NO concentration has been reached, ensure that NO2 level is
acceptable for patient administration. Below you will find a table which will give you the
approximate NO2 values you can expect for a given NO concentration and ventilator flow
at an FiO2 of 1.0.
Minute Volume / Ventilator Flow
[NO] 5 10 15 20
5 ppm
10 ppm
20 ppm
40 ppm
80 ppm
0.5 0.1 0.1 0.1
0.5 0.2 0.2 0.1
0.5 0.3 0.2 0.2
0.8 0.6 0.4 0.4
2.0 1.5 1.2 1.1
NO2 in PPM
The table above is for reference only.
Factors such as the accuracy of the ventilator flow, percent error in delivery apparatus,
and human error, may all affect the actual delivered patient dose. The calculated NO flow
should be compared with the analyzed NO dose to confirm accurate NO dosing. Should
the actual NO flow differ from the calculated NO flow by more than 10% beyond
published specifications, the cause must be determined and corrected immediately. If the
cause can not be immediately determined, ensure patient safety and contact International
Biomedical at 1-800-433-5615 for further assistance.
8. Connect the patient to the ventilator and monitor according to study and/or institution
protocol.
CAUTION!
The flush procedure must be performed each time NO therapy is
started. This includes initial therapy starts, tank changes, and restarting therapy after trials off NO.
Document No. 715-7000Revision E Page 9 of 63
Purge Procedures to follow when changing tanks and/or regulators.
Important!
Purge Procedure Required before Use
Please follow purge instructions below to ensure gas purity. Failure to follow these
instructions may introduce potentially harmful contaminants into the patient’s
breathing gas or may affect the monitoring analyzer’s accuracy by introduction of
contaminants into the calibration gas.
Warning: Perform cylinder connection and purge procedures in well-ventilated areas
to prevent inadvertent exposure to nitric oxide or nitrogen dioxide gas. Follow your
facility’s safety procedures for handling medical gas cylinders.
Purge Procedures for use with Medical Gas Regulators:
1. Connect cylinder to a matching CGA nitric oxide or nitrogen dioxide
regulator only.
2. Open, then immediately close the cylinder valve.
3. Release (bleed) all of the gas from the regulator.
4. Repeat steps 2 and 3 four more times for a total of five purge cycles.
5. Leave the regulator installed until it is time to change to a new cylinder.
6. Repeat the purge procedure any time a regulator is reattached.
Any time a regulator is installed on a tank or cylinder of compressed gas, certain precautions
must be followed. This is to prevent contamination of the gas in the tank and in the system by
air that is trapped in the dead space of the regulator and fittings. To eliminate the possibility of
the oxygen in this air reacting with the nitric oxide to form nitrogen dioxide in the system, the
regulator and fittings must be purged before use. The valve on the tank must not be opened
and left open until the regulator is purged. The stainless steel line must also be purged prior to
connection to the AeroNOx.
CAUTION!
Should there be a sudden need to change therapy tanks, a second tank
should always be purged and ready for immediate use. Perform the
purge procedure immediately upon installation of a new regulator.
Although the dead space volume in the regulator and hose assembly is physically small, if it had
been exposed to room air for a period of time it will contain sufficient oxygen to convert a
significant amount of nitric oxide to nitrogen dioxide.
Document No. 715-7000Revision E Page 10 of 63
The following purge procedure must be followed any time a regulator is installed on a tank of
gas. It should be done immediately after connecting a regulator to a tank to avoid contaminating
a tank of gas with the air trapped in the dead space of the regulator in the event someone turns
on and leaves on the valve on the tank.
1. Install and tighten the regulator onto the tank.
2. Insert the quick-connect on one end of the stainless steel hose into the quick-connect
fitting on the regulator. Leave the other end of the stainless steel hose free.
3. Crack open and immediately close the tank valve.
4. Gently press the center tip of the quick-connect on the free end of the stainless steel hose
against a clean hard surface and allow the pressure to bleed to zero. A small amount of
high concentration delivery gas will be released at 50 psig. Keep the hose away from
patients and clinicians in order to avoid exposure to the gas.
5. Repeat Step 3 and Step 4 above five times (five complete purge cycles).
6. Connect equipment and open the tank valve as needed.
Any time a regulator is installed onto a cylinder and before each patient use, the
AeroNOx system should be purged. - NO EXCEPTIONS!!
AeroNOx Bagger: Instructions for use with the AeroNOx
(See Diagram # 2.)
CAUTION!
Oxygen and Nitric Oxide mix in the reservoir bag and stagnant flow will cause NO2 to be
formed. Turn off NO flow when not actively bagging patient. Flush the bagger with
oxygen for a least 20 seconds to completely remove NO and NO2 before and after patient
use.
Concentrations of nitrogen dioxide in the reservoir bag may exceed 1 ppm. Large tidal
volumes may expose patients to the nitrogen dioxide in the reservoir bag.
Flush the bagger with oxygen for at least 20 seconds to completely remove NO and NO
after any
interruption in active patient bagging.
2
The AeroNOx may not be able to detect rapid, short acting fluctuations in the
concentration of the delivered gas, including nitrogen dioxide. For this reason, the
AeroNOx Bagger is designed for short-term use only.
If a patient requires nitric oxide concentrations other than 20 ppm, a separate system
capable of providing the required concentration must be available. DO NOT USE THE AERONOX BAGGER TO DELIVER CONCENTRATIONS IN EXCESS OF 20 PPM!
The generation of nitrogen dioxide increases rapidly above this concentration.
Do not alter the length of the Bagger Gas Supply Tubing as this may cause generation of
excessive levels of nitrogen dioxide.
Document No. 715-7000Revision E Page 11 of 63
Do not substitute AeroNOx Bagger components! The AeroNOx Bagger has been
designed and tested for patient safety with the components included in the current
configuration.
The AeroNOx Bagger is designed for use only with the AeroNOx Delivery System. Do
not attempt to use the AeroNOx with any other manual resuscitator. Do not attempt to
use the AeroNOx Bagger with any other delivery system.
A backup system must always be available in the event that the primary system should
fail. For instructions for use of a Back-Up System in the event of a failure of the
AeroNOx system, please see SECTION III, Back-Up system.
The AeroNOx Bagger is intended to be connected directly to the patient’s endotracheal
tube. Do not insert any additional tubing between the AeroNOx Bagger and the
endotracheal tube.
The AeroNOx Bagger is for single patient use only. Do not reprocess.
Persons using this device should be trained and experienced in the use of this device to
assure affective administration of nitric oxide, and to avoid injury to the patient or to
others resulting from inhalation of excess nitric oxide, nitrogen dioxide, or other reaction
products.
Persons using this device who may be particularly sensitive to nitric oxide, or who may be
exposed to these gases for prolonged periods as a result of the use of this device, should
be aware that the gases exiting the AeroNOx Bagger are not scavenged.
1. Connect the large end of the bagger gas supply tubing to an oxygen source. Ensure that
the small end of this tubing is connected to the NO / O2 gas inlet port on the bagger.
2. Connect AeroNOx Bagger gas delivery line to the AeroNOx delivery gas port.
3. Connect male end of the sample line (P/N 415-0004) to the sampling port on the bagger
gas supply tubing and the female end to the AeroNOx
sampling port.
4. Set O
seconds to flush out any residual NO or NO
flow on the O2 flow meter to exactly 10 LPM. Let the oxygen run for at least 20
2
left in the AeroNOx Bagger.
2
5. Turn on NO delivery gas source to 0.25 LPM. If you have correctly set your O
exactly 10 LPM, the delivered NO concentration will be ~20 ppm and delivered NO
flow at
2
2
values should not exceed 0.2 ppm. Do not exceed 20 ppm during manual ventilation!
6. Connect test lung and set overflow valve to desired positive end expiratory pressure
(PEEP).
Document No. 715-7000Revision E Page 12 of 63
7. Continue to run the AeroNOx Bagger with the test lung until the analyzed NO and NO
values stabilize. This will prevent high doses of NO
, which can form in the bagger and
2
delivery line, from being delivered to your patient. The NO2 sensor in the AeroNOx
Bagger may not react quickly enough to detect NO2 boluses. If you do not have time
to use a test lung, in an emergency situation for example, you must still flush the
bagger with fresh gas for at least 20 seconds prior to patient connection!
8. Analyze NO and NO
concentration prior to patient administration.
2
9. Flush bagger with O
NO
formation.
2
after each patient use to flush out residual NO and prevent
2
AeroNOx Bagger Specifications:
O2 Gas Flow
NO Gas Flow
Delivered [NO]
NO2 Generated
Reservoir Volume
Bagger Dimensions
Tidal Volume
Maximum 0.2 ppm @ 20 ppm Delivered [NO]
12” from Reservoir Bag to Patient Connection, ½” I. D.
Maximum 0.5 L (500 cc)
10 LPM
0.25 LPM
~ 20 ppm
0.5 L
Maximum 100 bpm at Pressures 18/5 (pip/peep)
Breath Rate
Maximum 50 bpm at Pressures 40/5
(Pressures Listed in cmH2O)
I:E Ratio
Variable
2
Diagram # 1
NO Titration with a Transport Ventilator
Document No. 715-7000Revision E Page 13 of 63
Diagram # 2 NO Titration with AeroNOx Bagger
Document No. 715-7000Revision E Page 14 of 63
The AeroNOx Bedside
Nitric Oxide Titration & Monitoring System
Component Part NumberQuantity
AeroNOx
AeroNOx Operating Manual
AeroNOx Technical Manual
AeroNOx Cart including Stand, Platform, Basket, & Power Bar
AeroNOx Mounting Block
731-9138 1
715-7000 1
715-7001 1
1
1
12 V Universal Power Supply 293-0040 1
SS Hose with Quick Connect Fittings 731-9371 1
Medical Grade Power Cord 711-0179 1
CGA 626 NO Delivery Regulators 731-9142 2
AeroNOx “NO WORRIES” Connector Sample Pak
AeroNOx Calibration Circuit
731-9373 1
415-0000 1
Single Calibration Screwdriver 416-0010 1
Recommended Bedside Gases: (Not Included)
Concentration: 800 ppm NO in balance N2 of pharmaceutical
grade gas
2
Size: 38” (1 meter) height by 7” (18 cm) diameter
The AeroNOx must be used with calibration gas and accessories
specified by the manufacturer.
Document No. 715-7000Revision E Page 15 of 63
The AeroNOx Bedside Set-Up
Prepare the AeroNOx Bedside Nitric Oxide Titration & Monitoring System
for initial operation as follows:
1. Unpack the AeroNOx
and inspect for damage.
2. Install battery as per SECTION III, “Battery”.
3. Secure the therapeutic tanks to the AeroNOx cart.
4. Unpack 12 V universal power supply (P/N 293-0040) and Medical Grade Power Cord
(P/N 711-0179). Plug in AeroNOx and charge for ~ 24 hrs.
5. Calibrate AeroNOx
Gas specified by the manufacturer and the instructions found in
using the Calibration Kit (P/N 731-9145) NO and NO2 Calibration
SECTION V of this manual.
6. Attach CGA 626 regulators (P/N 731-7142) to delivery gas cylinders as per “Purge procedures to follow when changing tanks and/or regulators” in this section.
7. Connect selected regulator outlet to the AeroNOx gas inlet using the stainless steel hose
with quick connect adapters.
8. Turn on nitric oxide gas supply.
9. Attach the AeroNOx delivery line (P/N 415-0001) found in the AeroNOx “NO
WORRIES” connector sample pak to the female luer connector marked “NITRIC OXIDE
DELIVERY OUTLET” on the AeroNOx. Tee the other end of the delivery line into the
patient breathing circuit at least 30 - 40 cm upstream of the sampling site. Attach the “NO
WORRIES” sample line (P/N 415-0004) to the male luer connector marked “ANALYSIS
SAMPLE INLET” on the AeroNOx.
10. Select the appropriate sample gas connector from your “NO WORRIES” connector
sample pak and attach it to the free end of the sample line.
11. Connect the sample gas connector to the patient breathing circuit. The connector should
be placed in the inspiratory limb, just prior to the patient wye connection (refer to Diagram
# 3).
12. “Flush” your system according to “Suggested “flush” procedure for the delivery portion of
the AeroNOx system”.
13. You are now ready to deliver NO gas and sample mixed inspired gases.
CAUTION!
Never turn on the NO delivery gas without first turning on the
ventilator or bagger flow. Failure to do this will result in undiluted
delivery gas entering the sampling chamber and exposing the sensors
to levels of NO and/or NO
which may damage the sensors.
2
CAUTION!
gas may have collected in the AeroNOx delivery circuitry. Prior
NO
2
to delivering gas to the patient, “flush” your delivery system with
fresh NO delivery gas. This should be repeated every time you begin
NO gas administration. Analyze the gas for high NO
levels prior to
2
patient connection.
Document No. 715-7000Revision E Page 16 of 63
Suggested “flush” procedure for the delivery portion of the AeroNOx system:
1. Set up the AeroNOx for delivery as outlined in “AeroNOx Bedside Set-Up”.
2. Hook up the ventilator to a test lung (patient should be manually ventilated during this
procedure and the ventilator parameters should be set to the patient’s pre-NO gas
settings).
3. Inactivate the “Safety Shut Off” System by simultaneously depressing the NO and NO2
alarm silence buttons for 5 seconds or until the display reads “SAFETY OFF”. For more
information on the “Safety Shut Off” system, see “Safety Shut Off” in SECTION III.
4. Turn on the NO gas flow to 2 lpm by turning the knob on the precision metering valve in a
counter-clockwise direction. This knob is located in the bottom right hand quadrant of the
front panel of the AeroNOx below the LCD flow display. Allow the gas to flow into the
ventilator circuit.
5. As soon as the NO2 value has dropped to a stable level, allow the gas to run for an
additional 3 seconds and then adjust the NO flow to the desired level (see SECTION VII Calculations for NO Delivery to determine flow required for desired NO concentration).
6. Re-activate the “Safety Shut Off” system by simultaneously depressing the NO and NO2
alarm silence buttons for at least 5 seconds or until the display reads “SAFETY ON”.
7. Once the desired NO concentration has been reached, ensure that NO2 level is
acceptable for patient administration. Following, you will find a table which will give you
the approximate NO2 values for a given NO concentration and ventilator flow at an FiO2 of
1.0.
Minute Volume / Ventilator Flow
[NO] 5 10 15 20
5 ppm
10 ppm
20 ppm
40 ppm
80 ppm
0.5 0.1 0.1 0.1
0.5 0.2 0.2 0.1
0.5 0.3 0.2 0.2
0.8 0.6 0.4 0.4
2.0 1.5 1.2 1.1
NO2 in PPM
The table above is for reference only.
Factors such as the accuracy of the ventilator flow, percent error in delivery apparatus,
human error, and accuracy of calibration of the AeroNOx
may all affect the actual
delivered patient dose. The calculated NO flow should be compared with the analyzed
NO dose to confirm accurate NO dosing. Should the actual NO flow differ from the
calculated NO flow by more than 10% beyond published specifications, the cause must
be determined and corrected immediately. If the cause can not be immediately
determined, ensure patient safety and contact International Biomedical at 1-800-4335615 for further assistance.
8. Connect the patient to the ventilator and monitor according to study and/or institution
protocol.
CAUTION!
The flush procedure must be performed each time NO therapy is
started. This includes initial therapy starts, tank changes, and restarting therapy after trials off NO.
Document No. 715-7000Revision E Page 17 of 63
Diagram # 3 AeroNOx Bedside Set-Up
The Sensormedics 3100A or 3100B
To set up the AeroNOx for use with the 3100A & 3100B, tee in the AeroNOx Delivery Line
(P/N 415-0001) immediately post humidifier. To connect the “NO WORRIES” sample line, you
will require the 3100A / 3100B sample port adapter (P/N 415-0009). The port can be placed in
one of three positions. All three positions are labeled as “temperature probe ports” in the
3100A / 3100B literature. The “set back” temperature port is pictured in Figure 1 and Figure 2.
The proximal temperature port is pictured in Figure 3. The third port (not shown) is the second
proximal temperature probe port located next to the “dump valve” on the inspired limb of the
flexible circuit.
Document No. 715-7000Revision E Page 18 of 63
Sampling Ports for the Sensormedics 3100A / 3100B:
Figure 1 The “Set Back” Temperature Probe Port
Figure 2 The “Set Back” Temperature Probe Port
Document No. 715-7000Revision E Page 19 of 63
Figure 3 The Proximal Temperature Probe Port
Document No. 715-7000Revision E Page 20 of 63
AeroNOx
Portable Cylinder Cart and Platform
Initial Set-Up: See package insert for complete instructions.
Tools Required: Allen key (supplied)
1. Remove screws (Item 11) from base pole.
2. Lift support wheel (Item 9) so hole locations in base pole and support wheel pole (Item 7
& 9) are aligned.
3. Insert screw (Item 11) and tighten securely into base pole.
4. Attach basket assembly to tank bracket via the four button head 10-32 5/16 screws
(Item 22) provided.
The AeroNOx Cart is a strong six-wheeled cart with two locking wheels. It can hold two
2040 L Aluminum cylinders which are ~ 38” (1 meter) in height and ~ 7” (18 cm) in
diameter.
Beneath the basket holder is a medical grade 120 V power bar with a 3 prong plug and
four power outlets.
The AeroNOx platform has a handle to allow movement of the cart, and a hook designed
to mount to the bed’s foot or headboard which allows the cart, cylinders, and AeroNOx
system to be transported with the patient within the hospital. (See Diagram # 5.)
The AeroNOx unit is attached to the cart platform by a metal block which fits into a ½”
hole in the black platform. The AeroNOx
aluminum sliding bracket, allowing secure bedside use and quick disconnect capabilities.
(See Diagram # 4.)
Document No. 715-7000Revision E Page 21 of 63
Description
unit is secured onto the mounting block by an
Diagram # 4
Room to Room Transport with the AeroNOx & Cart
The AeroNOx bedside system and tanks can be moved short distances with the patient by
hooking the black platform onto the bed’s foot or head board, and raising the bed slightly, so as
to partially un-weight the wheels.
Diagram # 5
AeroNOx & Cart
(tanks not shown for clarity)
ready to attach to the bed
in the lowered position.
AeroNOx and Cart partially
lifted by raising the bed
when the cart is attached to
the bed rail.
CAUTION!
A fully loaded AeroNOx
system with tanks weighs over 150 lbs (~ 68 kgs).
Do not use the bed to lift the system completely off of the floor due to risk of
serious injury to the patient, clinicians, or damage to the bed.
Stability & Safety:
The cart and monitor platform are designed to be stable when a NO cylinder is strapped to the
cart. The unit occupies minimal floor area at the bedside, and can be placed behind or beside
the mechanical ventilator.
CAUTION!
Support the cart when changing out tanks at the bedside.
Document No. 715-7000Revision E Page 22 of 63
SECTION III - DEVICE DESCRIPTION
General
The AeroNOx
(NO), nitrogen dioxide (NO2) and oxygen (O2) analyzer. The AeroNOx is a stand-alone,
lightweight, portable unit designed to continuously monitor NO, NO2, and O2 concentrations in a
breathing circuit.
The AeroNOx system is specifically designed for the delivery and monitoring of gaseous nitric
oxide (NO) in parts per million (ppm) concentrations.
The construction materials are compatible with NO and will not contaminate the gas stream.
systemincludes an integrated nitric oxide gas delivery system and nitric oxide
Recommended Use:
The AeroNOx Portable Nitric Oxide Titration & Monitoring System is designed for use with
continuous flow systems as either a:
Primary nitric oxide delivery system
Transport system
Backup delivery system in case of emergency
The AeroNOx has been tested for use with the AeroNOx Bagger, the Bird VIP (in
continuous flow modes only), the Sechrist 100IVB, the SensorMedics 3100A &
3100B, the Infant Star 100 Neonatal Ventilator, the MVP-10 Ventilator, the CV2i
Ventilator, and the Percussionaire Bronchotron Ventilator.
The AeroNOx is not MRI compatible.
Document No. 715-7000Revision E Page 23 of 63
Diagram # 6 The INOstat Flow Inflating Bagger:
Use as a back-up system (independent of the AeroNOx system).
Document No. 715-7000Revision E Page 24 of 63
Back-Up System:
An alternate means of delivering nitric oxide to your patient must be available should your
primary system fail. For this reason International Biomedical has an available back-up system,
the INOstat Kit (P/N 731-9147), for use in the event that your AeroNOx system should fail.
This system consists of a flow inflating bagger, and a specially designed regulator with a preset
NO delivery flow of 0.25 LPM.
Check INOstat Kit or other back up system for the AeroNOx for
functionality on initial set up and periodically during use.
To ensure that the back up system to the AeroNOx is always ready for emergency use, the
functionality of the system should be confirmed on initial set up and periodically during use.
To check the functionality for INOstat Kit baggers, follow the instructions provided in the
package inserts. These instructions read:
Set Up:
1. Connect the large end of the bagger gas supply tubing to an oxygen source. Ensure that the small end of
this tubing is connected to the NO/O
2. Connect the back up delivery regulator (P/N 731-9143) to the INOstat NO gas delivery line.
3. Ensure that the sampling port on the bagger gas supply tubing is capped.
4. Place the system in a place that is readily accessible in the event of an emergency.
CAUTION! Confirm that the system is functional and delivering 20 ppm on initial set up by connecting the
sampling system of the AeroNOx
periodically.
to the port that is normally kept capped on the bagger. Recheck functionality
CAUTION!
Oxygen and Nitric Oxide mix in the reservoir bag and stagnant flow will cause NO2 to be
formed. Turn off NO flow when not actively bagging patient. Flush the bagger with
oxygen for at least 20 seconds to completely remove NO and NO2 before and after
patient use.
Concentrations of nitrogen dioxide in the reservoir bag may exceed 1 ppm. Large tidal
volumes may expose patients to the nitrogen dioxide in the reservoir bag.
Flush the bagger with oxygen for at least 20 seconds to completely remove NO and NO
after any
interruption in active patient bagging.
Rapid, short acting fluctuations in the concentration of the delivered gas, including
nitrogen dioxide are possible. For this reason, the INOstat Bagger is designed for
short-term use only.
gas inlet port on the bagger.
2
2
Document No. 715-7000Revision E Page 25 of 63
If a patient requires nitric oxide concentrations other than 20 ppm, a separate system
capable of providing the required concentration must be available. DO NOT USE THE INOSTAT BAGGER TO DELIVER CONCENTRATIONS IN EXCESS OF 20 PPM! The
generation of nitrogen dioxide increases rapidly above this concentration.
Do not alter the length of the Bagger Gas Supply Tubing as this may cause generation of
excessive levels of nitrogen dioxide.
Do not substitute INOstat Bagger components! The INOstat Bagger has been
designed and tested for patient safety with the components included in the current
configuration.
A backup system must always be available in the event that the primary system should
fail.
The INOstat Bagger is intended to be connected directly to the patient’s endotracheal
tube. Do not insert any additional tubing between the INOstat Bagger and the
endotracheal tube.
The INOstat Bagger is for single patient use only. Do not reprocess.
The INOstat Kit is for use only in the event of primary nitric oxide delivery device
(AeroNOx) failure and is not intended for use as a primary system.
The Back-Up delivery regulator in the INOstat Kit has a preset flow and is for use only
with the INOstat Bagger. Do not attempt to use this regulator as a delivery regulator for
any other application.
Persons using this device should be trained and experienced in the use of this device to
assure affective administration of nitric oxide, and to avoid injury to the patient or to
others resulting from inhalation of excess nitric oxide, nitrogen dioxide, or other reaction
products.
Persons using this device who may be particularly sensitive to nitric oxide, or who may be
exposed to these gases for prolonged periods as a result of the use of this device, should
be aware that the gases exiting the AeroNOx
Bagger are not scavenged.
When used according to the instructions in this manual, the INOstat Kit will deliver a fixed
concentration of 20 ppm NO ( 4 ppm). For this reason, the INOstat Kit should be used only
during periods when the AeroNOx
has failed or cannot be used for other unanticipated
reasons, unless the patient is known to have no adverse effects at 20 ppm.
Document No. 715-7000Revision E Page 26 of 63
INOstat Kit: Instructions for Use
Your back-up system should be set up and ready for use at all times during inhaled nitric oxide
therapy. Clinicians should familiarize themselves with the use of the INOstat Kit. The
adjustable overflow valve in the flow inflating manual resuscitator (Bagger) will help to control
the amount of PEEP delivered. It is adjusted by turning the valve to adjust the size of the
opening to atmosphere.
Set-Up:
1. Connect the large end of the bagger gas supply tubing to an oxygen source. Ensure that
the small end of this tubing is connected to the NO/O2 gas inlet port on the bagger.
2. Connect the back-up delivery regulator (P/N 731-9143) to the INOstat NO gas delivery
line.
3. Ensure that the sampling port on the bagger gas supply tubing is capped.
4. Place the system in a place that is readily accessible in the event of an emergency.
Using the Bagger:
1. Start the oxygen flow at exactly 10 LPM.
2. Open up the valve on the back-up delivery gas cylinder. NO flow is preset at 0.25 LPM.
3. Allow 20 seconds for the combined gas supply to flow freely from the bagger. This will
flush out any residual NO2 that may have formed in the bagger or delivery system since
the bagger was last used.
4. Begin manually ventilating your patient. If you have properly set the O2 flow at exactly 10
LPM, the delivered NO dose will be ~ 20 ppm.
5. Once you are finished using the bagger, turn off the NO gas flow and run O2 through the
bagger (for at least 60 seconds) to flush out any residual NO or NO2.
INOstat Bagger Specifications:
O2 Gas Flow
NO Gas Flow
Delivered [NO]
NO2 Generated
Reservoir Volume
Bagger Dimensions
Tidal Volume
Maximum 0.2 ppm @ 20 ppm Delivered [NO]
12” from Reservoir Bag to Patient Connection, ½” I. D.
Maximum 0.5 L (500 cc)
10 LPM
0.25 LPM
~ 20 ppm
0.5 L
Maximum 100 bpm at Pressures 18/5 (pip/peep)
Breath Rate
Maximum 50 bpm at Pressures 40/5
(Pressures Listed in cmH2O)
I:E Ratio
Variable
Document No. 715-7000Revision E Page 27 of 63
Titration Delivery System
Nitric oxide gas is regulated to 50 psig by a CGA 626 nitric oxide regulator. It is then delivered
via stainless steel high pressure hose to the AeroNOx unit. (See Diagram # 7.) A precision
metering valve (controlled by the operator on the front panel) regulates the NO gas flow to the
patient circuit.
CAUTION!
Powering the unit off will shut off the delivery gas flow. Gas flow is
terminated when the unit is powered off, whether intentionally, or
during a complete loss of power.
The NO gas flow is measured by a mass flow meter before it is delivered via the “NO
WORRIES” delivery line (P/N 415-0001) into the breathing circuit. The mass flow meter’s range
is from 0 - 2 LPM in 10 mL/min (0.01 LPM) increments.
CAUTION!
DO NOT exceed 2.00 LPM as displayed on the AeroNOx flow meter
digital display. Displayed values in excess of 2.00 LPM are not
accurate. If the flow on the AeroNOx flow meter is set at a value
greater than 2.00 LPM, the delivered flow will be higher than the
displayed flow value.
The suggested titration point is in the inspiratory circuit at least 30 - 40 cm upstream of the
sampling site. The high concentration source gas (~800 ppm NO, balance N2) is diluted in the
breathing circuit’s air / oxygen mixture to the desired dosage.
This distance is required to ensure complete mixing of the nitric oxide with the carrier gas.
Distances greater than 30 - 40 cm may increase the amount of NO2 generated.
The initial nitric oxide flow rate required to get a desired [NO] can be estimated from the
following dilution formula. For more information, see
Troubleshooting.
NO Flow (LPM) =
Vent Flow (LPM) [NO] Desired
Source [NO] in Tank
SECTION VII, Calculations and
Document No. 715-7000Revision E Page 28 of 63
NO / NO2 and O2 Sampling
The AeroNOx system continuously draws a mixed gas sample @ ~ 150 mL/min from the
breathing circuit at a point in the inspiratory circuit nearest the patient connection. (See Diagram # 7.) The sample is drawn through a Nafion gas drying element, which removes
water vapor only, using a patented electrically charged membrane. The dry sample gas is then
drawn through a 5 micron hydrophobic filter into the device and into the sensor housing where
gas analysis of NO (ppm), NO2 (ppm) and O2 (%) takes place.
CAUTION!
Do not use the AeroNOx without the “NO WORRIES” sample line (P/N
415-0004) which includes both the Nafion
filter for removing condensing humidity from the sample gas. Failure to
do so may impair the function of the AeroNOx as well as decrease the
life of the internal components. WARRANTY MAY NOT BE HONORED
FOR DAMAGE CAUSED IN THIS MANNER!
Diagram # 7 AeroNOx: Gas Supply, NO Delivery, and Gas Sampling
tubing and the hydrophobic
Document No. 715-7000Revision E Page 29 of 63
Sensors
The AeroNOx is supplied with one nitric oxide sensor and one nitrogen dioxide sensor in the
delivery & analysis module. AeroNOx sensors are designed to provide trouble-free use for
about twelve (12) months. Lifetime of the AeroNOx sensor will depend entirely on the
concentration of NO or NO2 being analyzed. The normal lifetime is about 1000 hours of
continuous use when exposed to 100 ppm NO and 10 ppm NO2 respectively.
The AeroNOx sensors must have a bias voltage going to them at all times. The battery must
be charged in order to maintain this “bias voltage”. The AeroNOx maintains this voltage even
when the power button is in the “off” position. This means that the battery is being drained even
when the unit is powered off. For this reason, it is recommended that the AeroNOx be plugged
in whenever possible to avoid the possibility of draining the battery to the point where it can no
longer maintain the “bias voltage” necessary to keep the sensors in a ready state.
Unbiased sensors will need up to 48 hours to fully bias. This applies to the installation of new
sensors as well as to sensors which have lost their bias due to a loss of battery power.
Unbiased sensors will read a value greater than zero when exposed to room air. This value will
continue to drop as the sensor biases. Once the value has dropped to zero, the AeroNOx must
be re-calibrated prior to use.
CAUTION!
The NO2 sensor may easily be damaged by inadvertent high levels of
NO2. Use nitrogen or air to flush the system after high levels of NO
(> 100 ppm) or NO2 (> 20 ppm) have been introduced to the sensor.
CAUTION!
Condensing humidity will impair the function of the AeroNOx
sensors. Use only “NO WORRIES” sampling line (P/N 415-0004)
which utilizes Nafion tubing to prevent condensing humidity from
reaching the sensors.
In the unlikely event that the sensor(s) gets wet, turn the unit on and
sample dry room air to allow the AeroNOx
module to dry out.
Once dry, the AeroNOx
delivery & analysis
sensors will function
normally.
Analysis
Electrochemical sensors in the AeroNOx measure NO and NO2 concentrations in parts per
million (ppm). NO values are displayed to the nearest ppm while NO2 values are displayed to
the nearest tenth (0.1) parts per million. The electrochemical sensors operate by allowing NO
and NO2 molecules to diffuse across a membrane located on the top of the sensor. Once inside
the cell, these molecules react in an electrolyte solution augmenting an electrical current
between an anode and a cathode. The augmented current is proportional to the NO or the NO2
concentration. The AeroNOx
also samples oxygen % for added safety.
Document No. 715-7000Revision E Page 30 of 63
Alarms
The AeroNOx unit has adjustable alarm limits for NO & NO2. Audible and visual alarms are
activated at the value that is set by the user, with the exception of the NO2 high alarm setting
which can manually be adjusted to 99 ppm, but has a factory set default alarm of 11 ppm should
the user set the alarm at 11 or higher. By carefully choosing alarm limits, any deviations in NO
or NO2 concentrations will be known as soon as possible.
CAUTION!
There are no alarms for the oxygen analysis portion of the AeroNOx.
An auxiliary oxygen analyzer, such as the MiniOX 3000 by MSA, with
audible and visual alarms must be used in conjunction with the
AeroNOx.
Safety Shut Off
For patient safety, the AeroNOx has incorporated a “safety” shut-off system for the delivery of
NO gas. The system is designed to avoid inadvertent high doses of NO being delivered to the
patient by terminating the flow of NO gas to the delivery line. The system is activated when the
NO dose analyzed is 5 ppm above that which has been set as the high NO alarm (or 100 ppm
whichever is the lesser value) and 1 ppm above that set for the high NO2 alarm (or 12 ppm
whichever is the lesser value). The system is designed to restore the flow of gas to the delivery
line once the analyzed NO and/or NO2 values drop back to below the 5 ppm above and 1 ppm
above alarm values. The default setting for this alarm is “activated”, that is, no action is required
by the user to enable the safety system. To verify that the system is activated, simultaneously
depress the NO and NO2 alarm silence buttons. The display will read “SAFETY ON”. Do not
depress the alarm silence buttons for more than 4 seconds as this will disable the
“safety” shut-off system!
Should the “safety” shut-off system be triggered, (that is, an alarm condition has occurred) there
will be an audible as well as a visual alarm, and the display will alternately read “NITRIC OFF”
and the sensor data.
Your patient will not receive any NO gas from the AeroNOx
condition!
It is recommended that you manually ventilate your patient with a system such as the INOstat
Kit, capable of delivering a safe and known concentration of NO to your patient, while you
troubleshoot the system to determine why the “safety” shut-off system was triggered. For
information on the INOstat Kit back-up delivery system, see SECTION III, Back-Up System.
To override the system, simultaneously depress the alarm silence buttons for 5 seconds. The
display will read “SAFETY ON” until 5 seconds have elapsed at which time it will read “SAFETY
OFF”. At this point, the “safety” shut-off system has been inactivated. During the time that this
system remains inactivated, the display will alternately display the “SAFETY OFF” message and
the sensor data. To restore the “safety” shut-off to the activated mode, you may either
simultaneously depress the alarm silence buttons for 5 seconds, or you may power the machine
off and then back on again.
system during this
Document No. 715-7000Revision E Page 31 of 63
Diagram # 8 AeroNOx Front Panel
AeroNOx Front Panel
CONTROL FUNCTION
1. Display of NO & NO2 in
ppm
2. “Zero” NO & NO2
3. Power
4. LITE
5. HIGH / LOW Adjustable
Alarm Settings
6. Alarm Silence
7. Cal. NO & Cal. NO2
Calibration Pots
8. Oxygen %
9. O2 Cal. Pot.
10. NO Flow Meter
11. NO Flow Control
12. Check Sample Line
Back-lit, digital liquid crystal display of NO and NO2
ppm, calibration status and battery condition. (See
“Battery” SECTION III.) A “plug” character in the
display indicates that the wall power is connected
and that the unit is charging.
Automatically zeros the corresponding NO or NO2
sensor. Displays “DONE!” on the LCD when
zeroed.
Turns on / off power, except that which is
necessary to maintain the sensor in a state of
readiness.
Illuminates a back light for the LCD when
depressed.
Adjustable thumb-wheel alarm settings for high
alarms (NO & NO2) and low alarms (NO).
Silences audible alarms for one minute.
Turning these potentiometers adjusts the digital
display to calibrate the analyzer against a known
reference NO and NO
gas source.
2
Oxygen sensor with LED display
High point 100% calibration
Mass flow meter with 3.5 digit LED display, range
0 - 2 LPM in 10 mL/min (0.01 LPM) increments
Precision metering valve
Illuminates when sample line becomes occluded
Document No. 715-7000Revision E Page 32 of 63
Diagram # 9
1. Quick Connect Gas Inlet Female high pressure gas inlet, receives male
connector on SS hose from delivery regulator
2. Sample Gas Inlet Male luer connector. Hydrophobic filter from
sample line (P/N 415-0004) attaches here
3. Delivery Gas Outlet Female luer connector. Delivery line (P/N
415-0001) attaches here
CAUTION!
Luer fittings are common to both the AeroNOx and to most
intravenous systems. Careful attention must be paid when
connecting and disconnecting luer connections to avoid inadvertent
administration of NO gas into a patient I. V. system.
Quick Connect Gas Fittings
The AeroNOx was designed with quick connect gas fittings for quick and easy connection from
the AeroNOx
empty to a full cylinder, or when switching from a bedside to a transport cylinder and vice versa.
To connect the fittings, insert the male quick connect on the SS hose into the female quick
connect located on the AeroNOx unit and on the delivery regulators and push them together
until an audible “click” is heard and the two connections stay together.
Document No. 715-7000Revision E Page 33 of 63
to the delivery gas regulators. This can be utilized either when changing from an
CAUTION!
Ensure the regulator is purged prior to connection to the patient
circuit. See “Purge procedures to follow when changing tanks
and/or regulators” in SECTION II.
To disconnect, pull the sleeve on the female connector toward the male connector. The quick
connects will release from each other with very little effort.
THE NIPPLE SEAL SHOWN IN THE PHOTO BELOW MAY LEAK
IF NOT SECURELY FITTED TO THE NIPPLE.
TO ENSURE AN ADEQUATE SEAL BETWEEN THE TANK VALVE AND THE
REGULATOR, HAND TIGHTEN THE NIPPLE SEAL UNTIL IT IS FULLY
SEATED. DO NOT OVERTIGHTEN.
Luer Lock Connections
The sample gas inlet and the delivery gas outlet are equipped with luer connectors. These
connectors are very common in the hospital environment because of the quick way that air tight
connections can be made. These are the same connections found on I. V. tubing.
To attach the sample line (P/N 415-0004) to the sample gas inlet, simply fit the female luer on
the hydrophobic filter on the sample line to the male luer located below and to the left of the
handle on the AeroNOx and turn the hydrophobic filter in a clockwise direction until a firm
connection is established. To remove, turn the hydrophobic filter in a counterclockwise
direction, then gently pull the sample line free.
To attach the delivery line to the AeroNOx, fit the male end of the delivery line onto the female
luer located below and to the right of the handle on the AeroNOx
direction until a firm connection is established. To remove, turn the male fitting in a
counterclockwise direction.
CAUTION!
Do not use the AeroNOx without the “NO WORRIES” sample line (P/N
415-0004) which includes both the Nafion
tubing and the hydrophobic
filter for removing condensing humidity from the sample gas. Failure to
do so may impair the function of the AeroNOx as well as decrease the
life of the internal components. WARRANTY MAY NOT BE HONORED
FOR DAMAGE CAUSED IN THIS MANNER!
and turn in a clockwise
Document No. 715-7000Revision E Page 34 of 63
Battery
The AeroNOx is supplied with a rechargeable 6 volt sealed lead acid battery. The unit is
shipped with the battery disconnected to prevent the battery from fully discharging during
shipping. To re-connect the battery, remove the battery compartment panel on the back of the
AeroNOx. Connect the cable on the battery to the cable in the compartment. The expected life
of the battery is approximately one (1) year. Battery life varies considerably depending on the
amount of clinical use while the unit is unplugged.
The amount of charge that the battery holds is indicated by the letter “B” on the left hand side of
the display. This character will appear as soon as the unit is disconnected from the power
supply. As the charge that the battery holds increases, the letter “B” fills in until it resembles a
solid “B”. There are six levels of charge indicated by how much of the “B” is filled in. When the
power supply is connected to the AeroNOx, a character that resembles a power plug appears
in place of the “B”. This character indicates that the unit is plugged in and is charging.
If “LOW BATTERY” Alarm Occurs:
An audible alarm as well as a visual “LOW BATTERY” message will appear alternately
with the sensor data when the battery no longer carries enough charge to reliably operate
the electronics. This message will remain on the screen for approximately 4 minutes,
depending on the condition of the battery. If the unit remains disconnected from the power
supply, the screen will begin to alternately display the messages “CONNECT CHARGER”,
“DATA INVALID”, and “NITRIC OFF” until the display fades out for lack of power. GAS FLOW TO YOUR PATIENT WILL BE DISCONTINUED AT THIS TIME! Gas flow to your
patient is always terminated once the unit is powered off, intentionally or non-intentionally.
Should you completely lose battery function to the AeroNOx:
Plug in the 12 V universal power supply and allow the internal charging circuit to recharge
the battery.
After the battery is recharged ~ 24 hours, recalibrate the sensors using the procedure in
SECTION V of this manual.
Battery Maintenance & Storage:
The battery discharges even when the monitor is turned off. For this reason, the
AeroNOx
battery discharge.
Should the AeroNOx
plugged in, the battery should be disconnected to prevent the battery from fully
discharging.
A fully discharged battery requires ~ 30 minutes to become functional and ~ 24 hrs to fully
recharge.
should be kept plugged in whenever possible in order to prevent complete
need to be stored for periods greater than one week without being
Document No. 715-7000Revision E Page 35 of 63
Once the battery has discharged below a certain voltage, bias to the sensors is lost. An
unbiased sensor can take up to 48 hours to become functional again. Unbiased sensors will
read a value greater than zero when exposed to room air. This value will continue to drop as the
sensor biases. Once the value has dropped to zero, the AeroNOx must be re-calibrated prior
to use.
CAUTION!
Use only P/N 888-0115 as the battery source for the AeroNOx.
CAUTION!
Disconnect the battery when the unit is not in use for extended
periods of time or if the system must remain unplugged for extended
periods of time.
CAUTION!
Attempting to run the AeroNOx for longer than five
hours on battery only could result in an interruption of
Nitric Oxide Therapy.
Document No. 715-7000Revision E Page 36 of 63
Universal Power Supply
The universal power supply accepts a 100 - 240 V A/C ~ 50 - 60 Hz input, and supplies a +12 V
D/C @ 3.3 A max. output.
For the AeroNOx, the universal power supply is both a voltage source for the internal charging
circuit and an A/C power supply.
It will simultaneously power the analyzer and charge the internal battery. The internal charging
circuit is self-regulating and will sense the condition of the battery, charging and not charging as
required.
It is recommended that the power supply be used whenever possible, to ensure the
battery is kept fully charged.
The analyzer is powered by the internal battery at all times. The plugging in or
unplugging of the external power supply from either end will have no effect on the
analyzer operation or battery life.
CAUTION!
Use only the universal power supply (P/N 293-0040) as a power supply to
the AeroNOx.
Document No. 715-7000Revision E Page 37 of 63
SECTION IV - SPECIFICATIONS
NOTE: Standard conditions referred to throughout the specifications consist of the
following:
Temperature (transport & storage) - 20 C to + 40 C
Operating temperature + 5 C to + 40 C
Altitude up to 3800 meters or 12,000 ft.
Relative Humidity 0 - 93% non-condensing
Functional Performance
Measurement range
0 to 100 ppm nitric oxide
0 to 12 ppm nitrogen dioxide
5 to 100 % oxygen
The AeroNOx NO and NO2 sensors are not pressure dependent and are thus unaffected by
altitude changes during transport.
Oxygen Sensor
The oxygen sensor in the AeroNOx
is affected by altitude (partial pressure) changes. The
above graph shows a predictable linear slope for the first 4000 feet and then a nominal change
between 4000 and 12,000 feet. As altitude increases, the oxygen percentage display will
decrease to reflect the change in partial pressure at the corresponding change in altitude. The
following formula can be used to calculate the predicted change in oxygen percentage at a
specific altitude.
NOTE
: Remember to first determine the effect of the NO delivery on the oxygen
percentage as shown in SECTION VII, Calculations and Troubleshooting.
Document No. 715-7000Revision E Page 39 of 63
Step 1 - Calculate predicted change in oxygen percentage by multiplying the anticipated altitude
reading by - 4.71 10-3. Use this factor between sea level and 4000 feet. If the flying altitude
will be greater than 4000 feet above sea level, then use - 2.56 10-3 factor. For example:
If the reading on the AeroNOx is 60% at an altitude of 2000 feet above sea level, then to
calculate what the oxygen will be reading at a flying altitude of 8,000 feet above sea level you
would calculate as follows:
Calculated difference in Oxygen reading = (-2.56 10-3) New altitude
= (-2.56 10-3) 8,000
= -20.5%
New Oxygen reading = Current reading - Calculated difference in Oxygen reading
= 60% - 20.5%
= 39.5%
Flow Meter Display
The flow meter display will be affected nominally with a change in the altitude. However, the
delivered and displayed NO will not change. The average change with flow meter displayed for
0.56 LPM at 2000 feet above sea level read 0.53 (SD=0.02) at 12,000 feet above sea level.
Alarms
Manually adjusted thumb wheel switches with 1 ppm resolution
NO: Low adjustable from 0 - 99, high adjustable from 0 - 99. Audible and visual alarm
when actual value is at set alarm value
NO2: High audible and visual alarm when actual value is at set alarm value. Factory
default setting for high NO2 alarm is 11 ppm when thumb wheels are adjusted to 11 ppm
or higher (thumb wheels adjustable to 99 ppm).
Depressing the alarm silence button will deactivate the audible alarm, but not the visual
alarm, for one minute. If the alarm condition corrects itself, the audible and visual alarm
will deactivate. If the actual NO/NO2 value reenters the alarm condition, the audible and
visual alarms will be activated, even if one minute has not elapsed since the alarm silence
button was pushed.
Sample Gas Flow
150 cc/min 15%
sampled gas is vented to the room through a vent located on the right-hand side of the
AeroNOx
Sample Inlet Pressure
Cyclic pressure up to 80 cmH2O (Ti = 3.0 sec. and 6.6 cycles per minute)
Expected Life of Life-Limited Components under conditions of typical clinical usage:
NO Sensor ...................................min. 2000 hours
Range
Display Resolution
Accuracy (Including Linearity)
~ 11 lb. (~ 4.8 kg.)
0 - 2 LPM
10 cc
5% F. S.
Precision Metering Valve ~ 12 turns to fully open
Gas Supply Recommend NO @ 800 ppm N2
Gas Fittings CGA 626, Quick Connect & Luer
Power Supply
Battery Life
Internal 6 V sealed lead acid battery
charged by Universal power supply
Fully charged will last at least 5 hours
Battery life ~ 1 year
Safety
EMC: Compliance with EN60601-1-2
Flammable anesthetics: Not for use in the presence of flammable anesthetics
Regulatory Approvals: CSA, ETL, CE
Specifications are subject to change without notice.
Document No. 715-7000Revision E Page 41 of 63
SECTION V - CALIBRATION
Calibration of the AeroNOx NO / NO2 / O2 Analyzer
The AeroNOx must be used with calibration gas and accessories specified
by the manufacturer.
2. Use only the calibration circuit supplied, P/N 415-0000. Failure to use the supplied
calibration circuit may result in damage to the AeroNOx or inaccurate calibration results.
OXYGEN SENSOR CALIBRATION
3. Connect the nipple end of the calibration circuit to oxygen tubing.
4. Run wall O
5. Connect the luer connector of the calibration circuit to the sample inlet of the AeroNOx.
6. Allow the O2 reading to stabilize for 2 - 4 minutes. Calibrate oxygen sensor @ 100% by
adjusting the calibration potentiometer using the calibration screwdriver (P/N 416-0010)
provided. ** When using the oxygen analyzer to analyze oxygen concentrations
below 30%, the oxygen analyzer should be re-calibrated to 21% oxygen to improve
the accuracy in the lower range.
The NO and NO2 cells can be zeroed in oxygen or in room air. You may choose to leave the
oxygen running during the zero procedure.
7. NO Zero Procedure: Zero the NO sensor by depressing and holding the NO zero button
down. The words ZERO NO appear on the screen. The sensor is zeroed when the word
DONE! appears.
8. NO2 Zero Procedure: Zero the NO2 sensor by depressing and holding the NO2 zero
button down. The words ZERO NO2 appear on the screen. The sensor is zeroed when
the word DONE! appears.
9. Disconnect calibration circuit from the AeroNOx.
10. Disconnect the O2 and attach NO2 calibration gas to the calibration circuit.
11. Turn on the calibration gas flow by opening the valve located on the calibration gas
cylinder. The calibration regulators (P/N 731-9141) are factory set to deliver 0.5 LPM. If
you are not using these regulators, be sure to adjust the calibration gas flow to 0.5 LPM.
12. Connect calibration circuit to the AeroNOx
13. Allow the NO2 reading to stabilize for 2 - 4 minutes, or until the reading has not changed
for at least 30 seconds. Adjust the NO2 potentiometer on the front panel with the
calibration screwdriver (clockwise to increase, counterclockwise to decrease) until the
display matches the chemist’s reported “ACTUAL” NO
14. Disconnect the NO
LPM (or allow room air to be drawn in to the analysis sample inlet) until both sensors read
zero.
through the cal. circuit @ 0.5 LPM.
2
ZERO BOTH NO & NO2 SENSORS
CALIBRATE NO2 SENSOR
as in Step 5.
PURGE NO / NO2 SENSORS
calibration gas, and run oxygen through the calibration circuit @ 0.5
2
reading on the gas cylinder label.
2
Document No. 715-7000Revision E Page 43 of 63
CALIBRATE NO SENSOR
15. Disconnect calibration circuit from the AeroNOx.
16. Disconnect the O2 and attach NO calibration gas to the calibration circuit.
17. Turn on the calibration gas flow by opening the valve located on the calibration gas
cylinder. The calibration regulators (P/N 731-9141) are factory set to deliver 0.5 LPM. If
you are not using these regulators, be sure to adjust the calibration gas flow to 0.5 LPM.
18. Connect calibration circuit to the AeroNOx as in Step 5.
19. Allow the NO reading to stabilize for 2 - 4 minutes, or until the reading has not changed
for at least 30 seconds. Adjust the NO potentiometer on the front panel with the
calibration screwdriver (clockwise to increase, counterclockwise to decrease) until the
display matches the chemist’s reported “ACTUAL” NO reading on the gas cylinder label.
20. Disconnect the calibration circuit and allow room air to be sampled into the AeroNOx
until the NO and NO2 displays return to zero and the Oxygen sensor reads 21%. This is
to flush out any calibration gas remaining in the AeroNOx sampling circuit after
calibration. This step will help ensure the longevity of the AeroNOx sensors.
READY TO GO
The AeroNOx is calibrated and ready for use.
Important Reminders
1. The AeroNOx has been designed and tested for use with calibration gases
specified by the manufacturer. International Biomedical can not guarantee the
accuracy of the AeroNOx when calibration gases other than specified gases are
used.
2. Use only the calibration circuit provided to avoid damage to the AeroNOx or
inaccurate calibration results.
3. Zero NO & NO2 sensors daily & calibrate weekly when in use.
4. The AeroNOx holds its calibration even when the unit has been powered off. This
allows for the unit to be calibrated and then stored until it is needed. There is no
need to recalibrate the AeroNOx unless a period of one week or greater has
elapsed since the last calibration.
Helpful Hint:
It is a good practice to calibrate the AeroNOx
weekly even when not in
use. This ensures that the AeroNOx is always ready for use.
Document No. 715-7000Revision E Page 44 of 63
SECTION VI - NITRIC OXIDE DELIVERY
The AeroNOx Nitric Oxide Titration & Monitoring System was designed to deliver a continuous
flow of NO gas into the patient’s bulk gas flow.
Inspiratory Limb NO Titration in Constant Flow Ventilators or Other Respiratory Gas
Administration Systems
This technique is modeled after the system tested by Betit and colleagues published in
Respiratory CareJuly 1995 Vol 40 No. 7 p 706-715. With constant flow ventilators, stability of
the [NO] is assured because both the NO gas and the gas flow from the ventilator are fixed
(constant).
Precautions for Inspiratory Limb NO Titration
1. Carefully select “tight” NO & NO2 alarms on the NO / NO2 analyzer.
2. Where applicable, ensure that the minute ventilation supplied by the ventilator or by other
respiratory gas administration systems is consistent. Carefully select “tight” minute
volume alarms. This is critical.
3. Titrate the desired NO dose before you attach the system to the patient, using a test lung
with both the ventilator & NO delivery system running for approximately 1 minute prior to
patient connection.
4. Ensure that the dose measured by the AeroNOx is the actual desired nitric oxide
concentration and that NO2 values are acceptable. For acceptable limits, see the “NO2 in
ppm” table in SECTION VII.
Ventilators Safe for Use with the AeroNOx
In general, any ventilator which functions as a true continuous flow ventilator may be used with
the AeroNOx. Some ventilators have modes labeled “continuous flow” but may not function as
true continuous flow ventilators. Please consult the manufacturer of the ventilator you intend to
use with the AeroNOx to determine if, and in what modes, your ventilator will operate as a
continuous flow ventilator.
CAUTION!
The AeroNOx
only! Check with your ventilator manufacturer to determine the flow
characteristics of the ventilator you wish to use.
The following ventilators have been tested for use with the AeroNOx:
1. Sechrist 100IVB in all modes
2. Bird VIP in conjunction with the Infant Flow Sensor in the following time cycled modes:
a) Assist Control
b) SIMV
c) CPAP
3. Infant Star 100 Neonatal Ventilator
4. SensorMedics 3100A & 3100B
5. BioMed Devices MVP-10 Ventilator
6. BioMed Devices CV2i Ventilator
7. Percussionaire Bronchotron Ventilator
is designed to be used with continuous flow ventilators
Document No. 715-7000Revision E Page 45 of 63
(
)
Nitric Oxide Concentration Profiles
The following graph represents the concentrations of nitric oxide representative of the delivered
concentration to the patient throughout the respiratory cycle when using the INOstat Bagger.
The graph below represents the concentration of nitric oxide representative of the delivered
concentration to the patient during mechanical ventilation with each ventilator tested for use with
the AeroNOx.
Nitric Oxide Concentration
Ventilator Pressure
Concentration
NO
Nitric Oxide
Document No. 715-7000Revision E Page 46 of 63
NO = 20 1 ppm
Ventilator Pressure
PIP/PEEP = 24/4 cmH2O
Time
SECTION VII - CALCULATIONS & TROUBLESHOOTING
1. CALCULATIONS FOR NITRIC OXIDE DELIVERY
Where to start?
How to estimate initial NO flow
What is the diluted FiO
Estimating your FiO
?
2
after dilution with NO gas
2
How are we doing?
Oxygen Index Calculation
2. CYLINDER DURATION NOMOGRAMS
How long will this tank last?
3. AERONOX TROUBLESHOOTING GUIDE
When all else fails, read the manual...
CALCULATIONS FOR NITRIC OXIDE DELIVERY
Where to start?
It is imperative that you determine what NO flow to set, and what the NO flow reading
should be during NO administration for the desired dose.
To determine the initial Nitric Oxide (NO) flow rate, given your NO source gas
concentration, [NO], ventilator minute ventilation & desired NO dose, use the formula
below.
To determine NO flow for use with the AeroNOx Bagger or other respiratory gas
administration systems, substitute O
Initial NO Flow (LPM) =
Example:
Minute volume, or ventilator flow = 10 LPM.
Nitric oxide source tank = 800 ppm
You want to deliver 25 ppm to your patient.
Initial NO Flow (LPM) =
10 LPM 25 ppm Desired [NO]
800 ppm Tank [NO]
flow for ventilator flow in the equation below.
2
Vent Flow (LPM) Desired [NO] ppm
Source Tank [NO] ppm
250
=
800
= 0.31 LPM
Document No. 715-7000Revision E Page 47 of 63
Set 800 ppm NO source gas flow @ 0.31 LPM to get ~ 25 ppm NO diluted in 10 LPM
fresh gas flow.
The chart below was developed using the previous calculations. You may use it as a starting
point for setting your NO flow rate. Please note that these are reference points only. The
actual dose delivered must be measured by the analysis portion of the AeroNOx.
Minute Volume / Ventilator / Bagger Flow
[NO] 5 10 15 20
5 ppm
10 ppm
20 ppm
40 ppm
80 ppm
0.03 0.06 0.09 0.13
0.06 0.13 0.19 0.25
0.13 0.25 0.38 0.50
0.25 0.50 0.75 1.00
0.50 1.00 1.50 2.00
NO Flow in LPM
The following table represents the average NO2 value measured at an FiO2 of 1.0 during testing
with the ventilators listed in this manual. Use it as a reference for what NO
values you can
2
expect when delivering the NO concentrations listed in the column on the left.
Minute Volume / Ventilator Flow
[NO] 5 10 15 20
5 ppm
10 ppm
20 ppm
40 ppm
80 ppm
0.5 0.1 0.1 0.1
0.5 0.2 0.2 0.1
0.5 0.3 0.2 0.2
0.8 0.6 0.4 0.4
2.0 1.5 1.2 1.1
NO2 in ppm
The tables above are for reference only. Factors such as the accuracy of the ventilator flow,
percent error in delivery apparatus, and human error, may all affect the actual delivered patient
dose. The calculated NO flow should be compared with the analyzed NO dose to confirm
accurate NO dosing. Should the actual NO flow differ from the calculated NO flow by more than
10% beyond published specifications, the cause must be determined and corrected immediately.
If the cause can not be immediately determined, ensure patient safety and contact International
Biomedical at 1-800-433-5615 for further assistance.
CAUTION!
NO flow must be continually evaluated during Nitric Oxide
administration to ensure accurate dosing!
Document No. 715-7000Revision E Page 48 of 63
What is the diluted FiO2?
Estimating your FiO2 after dilution with NO gas, given your NO flow setting, and total ventilator
gas flow
Estimated FiO2 =
Initial FiO2 - (NO Flow / Total Flow O2 + Total NO Flow) 100
Example:
Initial FiO2 = 1.0 (100%)
NO flow = 0.31 LPM (800 ppm gas, bal Nitrogen)
Ventilator flow = 10 LPM
Total flow = 10.31 LPM
In this example, OI improved markedly (23 to 8.8) after inhaled nitric oxide administration.
Document No. 715-7000Revision E Page 50 of 63
Calculation of Cylinder Duration
How long will this tank last?
The volume of gas in your cylinder(s) is a function of the filling pressure and the capacity
(size) of the cylinder. For compressed gases, this relationship is linear and can be
expressed as a tank factor which is equal to cylinder volume / pressure expressed in
LPM.
GAS SUPPLY
E. g. Large NO Cylinders
Capacity (L): 2040
Maximum Pressure (full): 2000 psig
Tank Factor =
The amount of therapeutic gas remaining in a cylinder can be estimated in minutes or hours,
provided that three things are known:
1. Tank Factor
2. Tank Pressure
3. Flow Rate
Cylinder Duration (Minutes) =
The two following nomograms can be used to estimate cylinder duration.
CYLINDER DURATION
Below you will find an example of how to determine cylinder duration based
on the following:
Rounded down to the nearest quarter hour
Based on set flow rate and tank pressure
Time listed is time to run the cylinder dry (cylinders should be
changed out at 250 psig)
E. g. #1 - Cylinder Duration for a Cylinder with 2040 L @ 2000 psig
Calibration gas regulators have a preset flow of 0.5 LPM. Since sensor stabilization usually
takes ~ 2 - 4 minutes, one sensor calibration should take ~ 2 - 4 minutes and use ~ 2 L of
calibration gas.
Document No. 715-7000Revision E Page 52 of 63
AERONOX TROUBLESHOOTING GUIDE
FRONT PANEL - OPERATIONAL MESSAGES
Alarm
Silence NO
Alarm
Silence NO2
ZERO NO
ZERO NO2
ZERO NO &
ZERO NO
PRESSED
TOGETHER
Visual LED
BUTTON
PRESSED
2
Silences audible alarm for 1
minute
Silences audible alarm for 1
minute
Sets calibration baseline for
selected sensor
Sets calibration baseline for
selected sensor
Displays the total time in
hours that both sensors have
been in service
Alerts user that the sample
line has become obstructed
PURPOSE MESSAGE
ALARM SILENCE
Visual alarm persists until rectified
ALARM SILENCE
Visual alarm persists until rectified
CAL NO, after 5 seconds, DONE!
CAL NO2, after 5 seconds, DONE!
SENSOR TIME
, after 5 seconds, time
in hours that both sensors have been
in service
SAMPLE OBSTRUCTION
Document No. 715-7000Revision E Page 53 of 63
ERROR / WARNING MESSAGES
MESSAGE DESCRIPTION ACTION
“LOW
BATTERY”
alternating
with sensor
data
“NO ZERO
FAILURE”
“NO2 ZERO
FAILURE”
“ROM
FAILURE”
“CHECK
SAMPLE
LINE”
Battery voltage < minimum
operating voltage
Zero baseline calibration has been
corrupted.
Zero baseline calibration has been
corrupted.
Data stored in read only memory
has been corrupted or damaged.
Sample line has become occluded,
pump may have shut down.
Try to recharge the battery as outlined
in SECTION III, Battery.
If the battery still does not function,
either the charger or battery must be
inspected by a biomedical engineer.
Power unit off and then back on.
Calibrate as per SECTION V.
Power unit off and then back on.
Calibrate as per SECTION V.
Return to International Biomedical for
repair.
Check sample line for occlusion.
Remove occlusion. Power unit off for
3 - 5 seconds, then power back on.
This will re-set pump flow.
To re-enable the “safety” shut-off,
“SAFETY
OFF”
alternately
displayed with
sensor data
“Safety” shut-off system has been
disabled.
a) simultaneously depress the NO and
NO2 alarm silence buttons for 5
seconds until the display reads
“SAFETY ON” or
b) power the unit off, then back on
again.
NO gas flow to your patient has
Manually ventilate your patient with a
bagger capable of delivering a safe
and known concentration of NO gas to
the patient. Determine the cause of
the alarm. Do not resume NO
administration with the AeroNOx until
it is safe to do so.
“NITRIC OFF”
alternately
displayed with
sensor data
been discontinued because NO
analysis indicated NO delivery
levels in excess of 5 ppm above
the set “Hi” alarm limit or 100 ppm
whichever is the less value or NO2
levels 1 ppm above the set “Hi”
alarm limit or 12 ppm whichever is
the less value.
“DATA
INVALID”
alternately
displayed with
“CONNECT
The battery voltage has dropped
below a point where data can be
accurately displayed.
Connect the AeroNOx
supply and plug into a wall outlet.
CHARGER”
to the power
Document No. 715-7000Revision E Page 54 of 63
COMMON PROBLEMS ENCOUNTERED
SYMPTOM CAUSE ACTION
1. Possible system leaks
Sensors are slow
to calibrate or will
not calibrate
NO / NO2 values
drift during
calibration, will not
hold cal. values
NO / NO2 values
drift during clinical
use
Screen remains
blank when unit is
powered up.
Pump will not run
when unit is
powered on.
“AERONOX” is
displayed even
when the unit is
powered off
CAUTION!
Disconnect the battery when the unit is not in use for extended periods of time or if the
system must remain unplugged for extended periods of time.
2. Sensors may need
replacing. Please refer to
the section on Sensors for
sensor life.
3. Calibration gas may have
expired.
1. Sensors may need
replacing. Please refer to
the section on Sensors for
sensor life.
2. Sensors may need
additional time to be
biased
1. Sensors may need
replacing. Please refer to
the section on Sensors for
sensor life.
2. Sensors may need
additional time to be
biased.
3. Incomplete gas mixing in
your patient circuit may be
occurring.
1. Internal battery has
become discharged.
2. Internal battery requires
replacement.
1. Battery is not fully
charged.
2. Internal battery requires
replacement.
1. Ensure O
oxygen during O
during NO or NO
If not, check for external leaks.
Send to Biomedical engineering for
internal system leak check.
2. Have Biomedical engineering
change sensors.
3. Check “Analysis date” on cal. gas
label.
1. Have Biomedical Engineering
replace sensor(s).
2. Allow 24 - 48 hours for sensors to
bias. Calibrate as per SECTION V.
1. Have Biomedical Engineering
replace sensor(s).
2. Allow 24 - 48 hours for sensors to
bias. Calibrate as per SECTION V.
3. Refer to SECTION VI, Nitric Oxide
Delivery Techniques.
1. Plug “Universal Power Supply” into
the AeroNOx
for at least 30 min.
2. Have Biomedical Engineering
replace the battery.
1. Plug “Universal Power Supply” into
the AeroNOx
for at least 30 min.
2. Have Biomedical Engineering
replace the battery.
analyzer reads 100%
2
sampling and 0%
2
cal. gas sampling.
2
and allow to charge
and allow to charge
Document No. 715-7000Revision E Page 55 of 63
SYMPTOM CAUSE ACTION
NO screen
displays “1600”
and/or NO
screen
2
displays “21”
NO2 screen
displays 21 and
will not change
Either or both
screens display a
(-) hash mark
NO / NO2 values
drop suddenly
after water has
entered the
AeroNOx
sampling block
NO sensor will not
return to zero
1. Bias voltage to the
sensors has been lost.
1. NO
sensor has been
2
exposed to NO
values in
2
excess of 21 ppm.
1. Sensors have drifted
below zero.
1. Water is preventing the
sampling gas from
reaching the sensor
membrane.
1. Sensor has been exposed
to high levels of NO (>100
ppm).
1. Recharge or replace battery and
allow 24 - 48 hours for the sensors
to re-bias.
1. Allow room air or oxygen to be
aspirated into the AeroNOx
sampling block until NO
returns to zero. Calibrate the NO
value
2
2
channel. Exposure to high levels of
NO
will decrease the sensor’s life.
2
1. Zero sensors as outlined in the
calibration procedure (SECTION V).
1. Run room air or oxygen through the
AeroNOx
to dry out the sampling
chamber. Calibrate the analyzer as
in SECTION V prior to putting the
AeroNOx
back into clinical use.
Use sample line P/N 415-0004 to
prevent water from entering the
AeroNOx
.
1. Run room air or oxygen through the
analysis port until the value returns
to zero. The length of time this will
take will depend on how much NO
the sensor was exposed to and for
how long. Exposing the sensors to
high levels of NO will decrease the
sensor’s life.
Document No. 715-7000Revision E Page 56 of 63
SECTION VIII - CLEANING & MAINTENANCE
CLEANING
1. Clean the AeroNOx
cable assembly surfaces by wiping them with a cloth moistened with a mild detergent or
bactericidal agent. Avoid using alcohol on the liquid crystal display.
Caution!
DO NOT sterilize the AeroNOx
sterilization techniques.
DO NOT clean any parts with acetone or other strong solvents.
DO NOT immerse or allow water or any other liquid to enter the
AeroNOx
REPAIRS AND REPLACEMENT
All routine maintenance, repairs, and replacement of standard parts should be conducted
according to procedures outlined in the Technical manual. For all clinical or technical issues not
addressed in the manual, please contact International Biomedical directly at:
CUSTOMER SERVICE
Should you need clinical or technical information, please feel free to contact International
Biomedical. To help us to help you, please have the following information available:
A complete description of the problem / observation
Model #
Serial #
Your Institution address
Your fax #
The above information is necessary to determine warranty status as well as to gather the
information necessary for us to fill out a Return Merchandise Authorization (RMA) should you
need to send anything back for repair. Any equipment sent to International Biomedical without an RMA will not be accepted for delivery. Do not ship any products to International
Biomedical without first obtaining an RMA.
system with mild detergent. Clean the AeroNOx analyzer and
analyzer by steam autoclave or gas
analyzer or housing.
In North America, Call Toll Free:
1-800-433-5615
Internationally, Call 1-512-873-0033
FAX: 512-873-9090
Document No. 715-7000Revision E Page 57 of 63
INSTRUCTIONS FOR DISPOSAL
To dispose of the AeroNOx:
1. Remove the battery and dispose of it in accordance with local regulations, or recycle it by
calling the toll free number listed on the battery.
2. The remainder of the AeroNOx can be disposed of as per your hospital’s policy for nonhazardous materials, or it may be returned to International Biomedical for disposal.
Document No. 715-7000Revision E Page 58 of 63
SECTION IX - WARRANTY
Subject to the exceptions* and upon the conditions* stated below, International Biomedical
warrants that the products sold under this sales order shall be free from defects in workmanship
and materials for one year after delivery of the products to the original Buyer by International
Biomedical, and if any such products should prove to be defective within such one year period
International Biomedical agrees, at its option, (i) to correct by repair or at International
Biomedical’s election, by replacement with equivalent product any such defective product,
provided that investigation and factory inspection discloses that such defect developed under
normal and proper use (ii) to refund the purchase price.
The exceptions* and conditions* mentioned above are as follows:
a) Exchange, and/or factory repaired components are warranted for ninety (90) days from
ship date, from factory.
b) Upgraded parts are warranted for 6 (six) months from ship date from factory.
c) Electrochemical sensors are warranted for 6 (six) months from ship date to the original
buyer.
d) Components or accessories manufactured by International Biomedical which by their
nature are not intended to and will not function for one year are warranted only to give
reasonable service: what constitutes reasonable shall be determined solely by
International Biomedical. A complete list of such components and accessories is
maintained at the factory.
e) International Biomedical makes no warranty with respect to components or accessories
not manufactured by it in the event of defect in any such component or accessory.
International Biomedical will give reasonable assistance to the Buyer in obtaining from the
respective manufacturer whatever adjustment is authorized by the manufacturer’s own
warranty.
f) Any International Biomedical product claimed to be defective must, if required by
International Biomedical, be returned to the factory, transportation charges prepaid, and
will be returned to Buyer with transportation charges collect unless the product is found to
be defective due to workmanship or materials, in which case International Biomedical will
pay all transportation charges, subject to receipt of original shipping invoices. The
customer will be responsible for duty, taxes, border charges, or claims resulting from but
not limited to the improper processing of customs documents. Any damage incurred
during transit from the Buyer to International Biomedical due to poor or insufficient
packaging will be the responsibility of the Buyer.
g) If the product is a disposable or the like, it is warranted only to conform to the quantity
and content and for the period stated on the label at the time of delivery.
h) International Biomedical may from time to time provide a special printed warranty with
respect to a certain product, and where applicable, such warranty shall be deemed
incorporated herein by reference.
Document No. 715-7000Revision E Page 59 of 63
i) International Biomedical shall be released from all obligations under all warranties, either
expressed or implied, if any product covered hereby is repaired or modified by persons
other than its own authorized service personnel unless such repair by others is made with
the written consent of International Biomedical.
IT IS EXPRESSLY AGREED THAT THE ABOVE WARRANTY SHALL BE IN LIEU OF ALL
WARRANTIES OF FITNESS AND OF THE WARRANTY OF MERCHANTABILITY AND THAT
INTERNATIONAL BIOMEDICAL SHALL HAVE NO LIABILITY FOR SPECIAL OR
CONSEQUENTIAL DAMAGES OF ANY KIND OR FORM ANY CAUSE WHATSOEVER
ARISING OUT OF THE MANUFACTURE, USE, INABILITY TO USE, SALE, HANDLING,
REPAIR, MAINTENANCE, OR REPLACEMENT OF ANY OF THE PRODUCTS SOLD UNDER
THIS SALES ORDER.
Representations and warranties made by any person, including dealers and representatives of
International Biomedical, which are inconsistent or in conflict with the terms of this warranty,
shall not be binding upon International Biomedical unless reduced to writing and approved by an
expressly authorized officer of International Biomedical.
International Biomedical
8508 Cross Park Drive
Austin, Texas 78754
USA
Telephone: 1-512-873-0033
FAX: 1-512-873-9090
Toll Free: 1-800-433-5615
E-Mail: sales@int-bio.com
Website: int-bio.com
Document No. 715-7000Revision E Page 60 of 63
SECTION X - COMPETENCY BASED PERFORMANCE CHECK-OFF TOOL
Introduction
In order to ensure patient safety, safe operation of clinical equipment is a necessity. Each
employer is encouraged to have documented proof of employee competency in the operation of
clinical equipment. For this purpose, International Biomedical has made available this
competency based performance check-off.
This clinical check-off is based on criterion-referenced performance. Employee performance is
measured against pre-established standards of behavior. These behaviors are called critical
elements. Critical elements are single, discrete, observable behaviors that are mandatory for
meeting the standards of acceptability. Since competency in equipment operations is required,
all critical elements must be performed as specified in order to pass. 100% accuracy is
recommended for completion of the clinical performance check-off.
Employees are strongly encouraged to review the Operating manual for the AeroNOx Nitric
Oxide Titration & Monitoring System in order to familiarize themselves with the equipment
functions, and to practice performing these functions independently as per the competency
based check-off prior to the actual check-off by a clinician. During the actual check-off, the
clinician is to merely observe the employee perform the required functions as per the check-off
without offering assistance. As each function is performed, and if the criterion for performing the
specified critical element is met, the clinician can tick off that item in the “criterion met” column.
It may be necessary to provide teaching sessions or supervised practice sessions to staff prior
to the clinical competency based check-off.
The following two competency based check-off tools are designed for:
1) Routine (Weekly) AeroNOx Calibration
2) AeroNOx bedside or portable NO titration & analysis with a portable ventilator.
Document No. 715-7000Revision E Page 61 of 63
AeroNOx
1) Routine (Weekly) AeroNOx
Calibration
COMPETENCY BASED PERFORMANCE CHECK-OFF #1 of 2
Employee Name:
Date of Check-Off:
Supervised by:
1. Turn Power on. Use only the calibration circuit supplied (P/N 415-0000).
2. Connect the barbed fitting to standard oxygen tubing.
3. Run wall O2 through the cal. circuit @ 0.5 LPM.
4. Connect the luer connector of the calibration circuit to the sample inlet of the AeroNOx.
5. Allow the O2 reading to stabilize for 2 - 4 minutes. Calibrate oxygen sensor @ 100% by using the
calibration screwdriver to adjust the calibration potentiometer.
ZERO BOTH NO & NO2 SENSORS
6. Zero the NO sensor by depressing and holding the NO zero button down. ZERO NO appears on the
screen. The sensor is zeroed when the word DONE! appears.
7. Zero the NO2 sensor by depressing and holding the NO2 zero button down. ZERO NO2 appears on
the screen. The sensor is zeroed when the word DONE! Appears.
CALIBRATE NO2 SENSOR
8. Disconnect calibration circuit from the AeroNOx.
9. Disconnect the O2 and attach NO2 calibration gas to the calibration circuit.
10. Connect the calibration circuit to the AeroNOx as in Step 4.
11. Run the NO2 cal. gas @ 0.5 LPM. Re-attach calibration circuit as in Step 5. Allow the NO2 reading
to stabilize for 2 - 4 minutes. Adjust the NO
screwdriver until the reading matches the chemist’s reported NO
calibration gas tank.
PURGE NO / NO2 SENSORS Check
12. Disconnect the NO2 calibration gas, and run oxygen through the calibration circuit @ 0.5 LPM until
both sensors read zero.
CALIBRATE NO SENSOR
13. Disconnect the calibration circuit from the AeroNOx.
14. Disconnect the O2 and attach NO calibration gas, run the NO cal gas @ 0.5 LPM. Re-attach the
calibration circuit as in Step 4. Allow the NO reading to stabilize for 2 - 4 minutes. Adjust the NO
potentiometer on the front panel with the calibration screwdriver until the reading matches the
chemist’s reported NO concentration on the NO calibration gas tank.
15. Disconnect the calibration circuit and allow room air to be sampled into the AeroNOx until the NO
and NO
gas remaining in the AeroNOx
longevity of the AeroNOx
The AeroNOx is calibrated and ready for use.
displays return to zero and the Oxygen sensor reads 21%. This will flush out any calibration
2
sampling circuit after calibration. This step will help ensure the
sensors.
potentiometer on the front panel with the calibration
2
READY TO GO
concentration on the NO2
2
NAME: , employee # has successfully
completed the AeroNOx Calibration competency.
Date: Signature:
Check
Done
Document No. 715-7000Revision E Page 62 of 63
2) AeroNOx Portable & Bedside Set-Up
COMPETENCY BASED PERFORMANCE CHECK-OFF #2 of 2
Employee Name:
Date of Check-Off:
Supervised by:
CRITICAL ELEMENTS Check
1. Completed Calibration check off
2. AeroNOx SET-UP: TRANSPORT or BEDSIDE
Attach regulators to 800 ppm NO tanks
3. SOURCE GAS:
Attach high pressure quick connect from regulator(s) to AeroNOx
4. DELIVERY:
Attach delivery tubing (P/N 415-0001) from AeroNOx
the sampling site.
5. SAMPLING:
Attach AeroNOx
connection.
6. Double check [NO] of the cylinder & record = ppm
7. Double check NO tank pressure = psig
8. Calculate initial NO flow to achieve the desired [NO] (see SECTION V).
9. Ventilator flow / gas flow = LPM
10. Tank [NO] =
11. Desired [NO] = ppm
12. Calculated Initial NO flow =
13. Calculate maximal FiO2 (see SECTION VII).
Maximum FiO
14. Perform flush procedure.
15. Set NO flow @ the initial flow rate you calculated (# 12).
16. After readings stabilize, record the following from the AeroNOx
Record [NO]
Record [NO
Record % O
17. Calculate the cylinder NO duration in hours & minutes (see SECTION VII).
This
will last
luer sample line from AeroNOx to inspiratory limb / bagger near patient
ppm
LPM
=
2
ppm
] ppm
2
ppm
2
L tank of NO gas @ psig
Hrs Min @ LPM.
to pt. circuit at least 30 - 40 cm upstream from
.
NAME:
, employee # has successfully
completed the AeroNOx Portable & Bedside Set-Up competency.
Date: Signature:
Document No. 715-7000Revision E Page 63 of 63
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.