Intelligent Weighing Technology PG Series, Intell-Lab PG Series User Manual

1. Environmental Conditions
a. Place the balance on a stable platform, away from vibration, air
currents, direct sunlight and strong electromagnetic waves.
b. Ensure the balance is level by using the adjustable feet.
Operating temperature: 41 °F to 95 °F / 5 °C to 35 °C Temperature fluctuation ≤5 °C / hour.
Humidity 50% to 85% RH.
2. Operation
void the warranty. Plug in the power cord and wait 5 minutes before turning on the balance.
b. Turn on the balance using the ON/OFF key. Following the
8.8.8.8.8. display, the balance will display maximum capacity, “S--CAL”, then 0.00 g in weighing mode. Please allow at least 15 minutes for the balance to warm up before using. (Longer if the temperature is below 60 °F).
3. Calibration
a. Single range calibration (recommended). Press “CAL” for 3
seconds. The display will show “CA-dA”, then display the weight to be applied (in grams). Apply the weight, the display will show CAL, followed by the weight value. The balance may now be used. If the weight reading is not correct (due to environmental conditions), repeat the process.
4. Weighing
a. When the black dot at the bottom left of the display has
disappeared, the balance is stable and ready to begin weighing. Please press the “TARE” key before weighing.
b. If the display shows “--------“, the balance has been overloaded.
Remove the weight and begin again with a smaller load.
5. Tare
a. When the balance is stable, press the “TARE” key to always
start weighing at zero.
6. Backlight
a. Press the “BL” backlight button to turn the backlight on and off.
7. Zero-Tracking and Auto-Tare
a. Turn the balance off. Press and hold the “CAL” key and at the
same time, turn the balance on. The display will show “-Zero-”. Press the “TARE” key. The display will show “Zero*d”. Press the “TARE” key to change the value from 0 to 5, “Zero0d” meaning no zero-tracking and the sensitivity is the highest. “Zero5d” is the factory setting. Press backlight to return to weighing.
b. Press the “CAL” key. The display will show “-tArE-“. Press the
“TARE” key to adjust the “*” from 0 to 9. “tArE0d” means there is no auto-tare. “tArE9d” is the factory setting.
c. “Press the “BL” backlight key to confirm settings. The display
will then show “0.0”. Your settings will remain the same when the balance is turned on again.
8. Other Functions
a. Units. Press “MODE” for 3 seconds. “Unit” will flash. Then
press the “TARE” key and the current unit will flash. Press the “MODE” key to toggle between units. Press the “TARE” key to confirm. The balance will return to weighing in the unit chosen.
b. Counting. Press the “MODE” key for 3 seconds until the
display flashes. Press the “MODE” key again to choose “Count”. Then press the “TARE” key to confirm. The display will show “pcs”. Press the “MODE” key again to choose from “10pcs” to “500pcs”. The larger the number, the higher the count accuracy. Put the sample parts on the balance pan and press the “TARE” key to confirm the sample size. The display will show “----“. If the sample size is too small the display will show “no-Cou”.
c. Percentage. Press the “MODE” key for 3 seconds until the
display flashes. Press the “MODE” key again to choose “PER”. Then press the “TARE” key to confirm. The display will show “100%”. Place the sample to be weighed on the pan and press the “TARE” key to confirm the sample as 100%. Remove the sample and place the goods on the pan. The display will show “###%”. If the amount on the pan is too small a weight the display will show “no-PER”.
d. Print. Press the “MODE” key until “Prt” is displayed. Press the
“TARE” key to confirm. Press the “MODE” key to change between “hAnd”, “Auto” and “Contin”. Press the “TARE” key to accept.
i. “hAnd” – Print on demand
ii. “AUto” – Print when the balance is stable
iii. “Contin” – Print continuously
RS 232 port parameter ASCII code transmission data format: Every 10BIT as a Data frame. 1 start bit, 8 data bits, 1 stop bit, no parity Baud rate: 9600 bps
Protect the unit against moisture. Please disconnect the unit from the main power supply before cleaning. Do not
use any liquids or aerosol sprays to clean this unit. Use only a clean dampened cloth.
Intelligent Weighing Technology, Inc.
www.intelligentwt.com
A Higher Level of Precision…
A Higher Level of Performance
Intell-LabToploading Balance PG Series
User Manual
© 2010 Intelligent Weighing Technology, Inc. All rights reserved worldwide. The information contained herein is the property of Intelligent Weighing Technology, Inc. and is supplied without liability for errors or
omissions. No part may be reproduced or used except as authorized by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodies.
Loading...