MB/TS25 PC211 UPGRADE Battery Charger
Operators Manual
Press the ENTR key rapidly twice
The display will briefly display CAL.
To re-calibrate current press ASET.
Enter the number of amps displayed on the measuring equipment, then press
ENTR.
The charger will adjust the charge amps so that the measuring equipment will
then display the same as the charger.
To recalibrate voltage press VSET
Enter the number of volts as displayed on the measuring equipment and press
ENTR.
The charger display will then adjust the display to show the same reading as
the measuring equipment.
Once the process is completed press STOP to end the charge.
Charge circuit calibration is complete.
If necessary the same process is performed for channel B.
5.3.2 Capacity Test Re-Calibration.
Connect the measuring equipment in accordance with the calibration equipment
connection diagram to Channel A.
Set the battery charger to test mode.
Set capacity test amps to 25A using the ASET key.
Set the termination voltage to 20V using the VSET key.
Set the capacity test duration to 010 using the TSET key.
Press START to begin the capacity test process.
Allow the capacity test current to stabilise.
Press the ENTR key rapidly twice
The display will briefly display CAL.
To re-calibrate current press ASET.
Enter the number of amps displayed on the measuring equipment, then press
ENTR.
The charger will adjust the charge amps so that the measuring equipment will
then display the same as the charger.
To recalibrate voltage press VSET
Enter the number of volts as displayed on the measuring equipment and press
ENTR.
The charger display will then adjust the display to show the same reading as
the measuring equipment.
Once this process is completed press STOP to end the capacity test.
Capacity test circuit calibration is complete.
Doc: DWG1006-34-R3 TS25 PC211 Upgrade Operators manual.odtPage 27 of 44Copyright Material of Intelligent Charging Limited © 2015
Printed On : 02/09/15