Integra VIAFLO 96, VIAFLO 384 Operating Instructions Manual

VIAFLO 96/384 Operating instructions
125950_V11
Declaration of conformity
INTEGRA Biosciences AG – 7205 Zizers, Switzerland
declares on its own responsibility that the devices
VIAFLO 96 6000, 6001
VIAFLO 384 6030, 6031
comply with:
EU Directives (DoW: Date of Withdrawal) Before DoW DoW After DoW Low Voltage Equipment 2006/95/EC 20.04.2016 2014/35/EU Electromagnetic Compatibility 2004/108/EC 20.04.2016 2014/30/EU Restriction of Hazardous Substances 2011/65/EU Waste Electrical and Electronic Equipment 2012/19/EU
EU Regulations
Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH)
Standards for EU
Safety requirements for electrical equipment for measurement,
EN 61010-1: 2010
control and laboratory use - General requirements. Particular requirements for automatic and semi-automatic
EN 61010-2-81: 2015
laboratory equipment for analysis and other purposes. Electrical equipment for measurement, control and laboratory use
EN 61326-1: 2013
- EMC requirements.
1907/2006
Standards for Canada and USA
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements
Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes
a
.
a
.
a
.
Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interfer­ence received, including interference that may cause undesired operation.
a
NRTL certificate number (TÜV Süd): U8 17 05 42035 007
Zizers, June 19, 2017
Elmar Morscher CEO
Thomas Neher Quality Manager
CAN/CSA-C22.2
No. 61010-1
UL 61010-1
UL 61010-2-81
Part 15 of the
FCC Rules
Class A
VIAFLO 96/384 – Operating instructions V11
Table of Contents
Chapter 1 Introduction
1.1 Symbols used...................................................................................7
1.2 Intended use.....................................................................................7
1.3 Safety notes .....................................................................................8
Chapter 2 Description of the device
2.1 Scope of delivery..............................................................................9
2.2 Overview of VIAFLO 96/384.............................................................9
2.2.1 VIAFLO 96/384 device..........................................................9
2.2.2 VIAFLO 96/384 control unit.................................................10
2.2.3 Display ................................................................................10
2.2.4 Touch wheel........................................................................11
2.2.5 Left and right arrow buttons ................................................11
2.2.6 PURGE button ....................................................................11
2.2.7 RUN key..............................................................................11
Chapter 3 Installation
3.1 Operating environment...................................................................12
3.2 Setting up and moving of the instrument........................................12
3.3 Getting started................................................................................13
3.4 Toolbox - adapt your VIAFLO 96/384.............................................14
3.4.1 Change head ......................................................................14
3.4.2 Storage of pipetting heads..................................................15
3.4.3 Position Settings .................................................................16
3.4.4 Preferences.........................................................................16
3.4.5 Calibration & Services.........................................................18
3.4.6 Communications .................................................................19
3.4.7 Device Information..............................................................19
3.4.8 Language............................................................................19
3.4.9 Write protect........................................................................20
www.integra-biosciences.com 3
VIAFLO 96/384 – Operating instructions V11
Chapter 4 Operation
4.1 Turn on/off the device.....................................................................21
4.2 Attaching and removing GripTips ...................................................21
4.2.1 Loading tips from a complete box.......................................21
4.2.2 Partial tip loading ................................................................22
4.2.3 Ejecting used GripTips........................................................22
4.3 Start pipetting .................................................................................23
4.3.1 Pipetting..............................................................................23
4.3.2 Set Z-Position .....................................................................23
4.3.3 Blowout modes ...................................................................23
4.3.4 Recommendations for pipetting ..........................................24
4.4 Pipetting options and settings ........................................................25
4.4.1 Edit option...........................................................................25
4.4.2 Volume selection.................................................................25
4.4.3 Speed selection ..................................................................26
4.4.4 Custom................................................................................27
4.4.5 Pace....................................................................................27
4.4.6 Count, Mix Cycle and Rows................................................27
4.4.7 Pipetting between 96 and 384 well plates...........................28
4.4.8 Pipetting with the three position stage................................29
4.5 Troubleshooting/FAQ .....................................................................30
Chapter 5 Pipetting modes
5.1 Overview pipetting modes ..............................................................31
5.2 Detailed description of pipetting modes .........................................32
5.2.1 Pipet mode..........................................................................32
5.2.2 Repeat dispense mode.......................................................33
5.2.3 Sample dilute mode ............................................................34
5.2.4 Pipet/mix mode ...................................................................34
5.2.5 Manual pipet mode .............................................................35
5.2.6 Reverse pipet mode............................................................36
5.2.7 Variable dispense mode .....................................................37
5.2.8 Variable aspirate mode.......................................................38
5.2.9 Sample dilute/mix mode......................................................39
5.2.10 Serial dilution mode ..........................................................40
5.3 Custom step-based programming mode ........................................41
5.3.1 Manual custom program .....................................................42
5.3.2 Modify existing programs....................................................44
5.3.3 Automatic custom program.................................................44
5.4 VIALINK..........................................................................................46
4 INTEGRA Biosciences AG
VIAFLO 96/384 – Operating instructions V11
Chapter 6 Maintenance
6.1 Cleaning and Decontamination ......................................................47
6.1.1 Base unit.............................................................................47
6.1.2 Pipetting heads ...................................................................47
6.2 Prolonged period of non-use..........................................................47
6.3 Leak test.........................................................................................48
6.4 Servicing.........................................................................................49
6.4.1 Blocked pipetting unit..........................................................49
6.4.2 Changing O-rings of tip fittings............................................50
6.5 Calibration ......................................................................................51
6.6 Equipment disposal ........................................................................51
Chapter 7 Technical Data
7.1 Environmental conditions ...............................................................52
7.2 Specification of the device..............................................................52
7.3 Intellectual Property........................................................................53
7.4 Pipetting specifications...................................................................54
Chapter 8 Accessories and consumables
8.1 Accessories....................................................................................55
8.2 Consumables .................................................................................56
www.integra-biosciences.com 5
VIAFLO 96/384 – Operating instructions V11
Imprint
© 2017 INTEGRA Biosciences AG All rights to this documentation are reserved. In particular the rights of reproduction,
processing, translation and the form of presentation lie with INTEGRA Biosciences AG. Neither the complete documentation nor parts thereof may be reproduced in any way, or stored and processed using electronic media or distributed in any other way without the written consent of INTEGRA Biosciences AG.
This operating instruction manual has part number 125950, the version is V11. It applies to software version as of (see Toolbox - Device information):
Control unit (FW) 3.10 or higher Instrument base (HW) 3.20 or higher
until a newer revision is released.
Manufacturer INTEGRA Biosciences AG
CH-7205 Zizers, Switzerland T +41 81 286 95 30 F +41 81 286 95 33
INTEGRA Biosciences Corp.
Hudson, NH 03051, USA T +1 603 578 5800 F +1 603 577 5529
info@integra-biosciences.com www.integra-biosciences.com
Customer service
Please contact your local INTEGRA Biosciences AG representative. To find out name and address of your local representative go to www.integra-biosciences.com
Further information and operating instructions in other languages are available on www.integra-biosciences.com
6 INTEGRA Biosciences AG
.
or on request info@integra-biosciences.com.
VIAFLO 96/384 – Operating instructions V11 Introduction

1 Introduction

VIAFLO 96 and VIAFLO 384 are separate instruments. The abbreviation VIAFLO 96/384 is used if information is relevant to both instruments. These operating instructions contain all the information required for installation, operation and maintenance of VIAFLO 96/384. This chapter informs about the symbols used in these operating instructions, the intended use of VIAFLO 96/384 and the general safety instructions.

1.1 Symbols used

The operating instructions specifically advise of residual risks with the following symbols:
ARNING
W
This safety symbol warns against hazards that could result in injury. Italso indicates hazards for machinery, materials and the environment. It is essential that you follow the corresponding precautions.
AUTION
C
This symbol cautions against potential material damage or the loss of data in a microprocessor controller. Follow the instructions.
NOTE
This symbol identifies important notes regardi ng the correct operation of the device and labor-saving features.
The device is marked with the following symbols:
IOHAZARD
B
The instrument can be potentially biohazardous due to the use of biozazardous substances by the operator.
UNPLUG
The power cable must be unplugged before the housing of the instrument is opened by a service technician.

1.2 Intended use

This is general purpose laboratory instrument for use in research only. Any use of this instrument in a medical or IVD setting is the responsibility of the user.
VIAFLO 96/384 is an electronic 96/384-channel hand held pipette designed for aspirating and dispensing aqueous solutions in the volume range of 0.5 µl to 1250 µl using GripTip pipette tips. VIAFLO 96/384 is used like a hand held pipette. The movement and positioning of the 96-/384-channel pipette is supported by a servo assisted steering mechanism which allows fast, precise and stress free multichannel pipetting. VIAFLO 96 can only be used with 96-channel pipetting heads, while VIAFLO 384 is compatible to both 96- and 384-channel pipetting heads.
www.integra-biosciences.com 7
VIAFLO 96/384 – Operating instructions V11 Introduction
If VIAFLO 96/384 is used in a manner not specified by INTEGRA Biosciences, the protection provided by the VIAFLO 96/384 may be impaired.

1.3 Safety notes

VIAFLO 96/384 complies to the recognized safety regulations and is safe to operate. VIAFLO 96/384 can only be operated when in perfect condition and while observing these operating instructions.
The device may be associated with residual risks if it is used or operated improperly by untrained personnel. Any person operating the VIAFLO 96/384 must have read and understood these operating instructions, and particularly, the safety notes, or must have been instructed by supervisors so that safe operation of the device is guaranteed.
ARNING
W
Do not use the VIAFLO 96/384 near flammable material or in explosive areas. Also, do not pipette highly flammable liquids such as acetone or ether.
When handling dangerous substances, comply with the material safety data sheet (MSDS) and with all safety guidelines such as the use of protective clothing and safety goggles.
AUTION
C
Do not immerse the pipetting head in liquid. The fluid can damage intern al parts. Avoid pipetting of liquids whose vapors could attack the materials PA (polyamide), POM (polyoxymethylene), FPM (fluor-rubber), NBR (nitrile­rubber), CR (chloroprene), silicone. Corrosive vapors could also damage metallic parts inside the device.
Do not open or modify VIAFLO 96/384 in any way . Th e sheet m eta l must not be removed. Repairs may only be performed by INTEGRA Biosciences AG or by an authorized after-sales service member.
Parts may be replaced with original INTEGRA Biosciences parts only. Never insert the mains adapter of any INTEGRA electronic hand pipette into
the power connector of the control unit (5
NOTE
Prolonged exposure of the VIAFLO 96/384 to UV-light can cause discoloration and/or yellowing of the control unit. However, this will not affect the performance of the device in any way.
Regardless of the listed safety notes, additional applicable regulations and guidelines of trade associations, health authorities, trade supervisory offices, etc. must be observed.
Please visit our website www.integra-biosciences.com information regarding REACH classified chemicals contained in our products.
).
on a regular basis for up to date
8 INTEGRA Biosciences AG
VIAFLO 96/384 – Operating instructions V11 Description of the device
4
5
8 A
8 B
10
6
1
2
9
3
7

2 Description of the device

2.1 Scope of delivery

• VIAFLO 96/384 device
• Power cable
• USB cable type A to B
• Operating instructions
AUTION
C
Verify the scope of delivery when unpacking the device and check for potential transportation damage. Do not operate a device that is damaged, instead contact your local dealer.

2.2 Overview of VIAFLO 96/384

2.2.1 VIAFLO 96/384 device
1 Base unit, to move left and right (X axis) 2 Carrying handle of base unit 3 Pipetting unit, to move up and down (Z axis) 4 Tip load button 5 Control unit, see 2.2.2 6 Knob of Side cover, covers the pipetting head 7 Instrument deck 8 Plate holders on position A and B 9 Plate slider, with front (384), zero (96) and back
(384) position to index 96 or 384 well plates (Y axis)
10 Main switch
www.integra-biosciences.com 9
VIAFLO 96/384 – Operating instructions V11 Description of the device
11
12
15
13
14
19
Current mode
Active step
OK
2.2.2 VIAFLO 96/384 control unit
11 Display 12 Back button, to navigate backward 13 Touch wheel, spin to scroll and move the cursor 14 OK button, to make a selection 15 Left and right arrow buttons, for selections 16 PURGE button, to empty tips 17 RUN key, to start operations 18 Tip ejector 19 Finger hook, facilitates easy operation
2.2.3 Display
The Display shows all pipetting options.
10 INTEGRA Biosciences AG
Instruction Volumes
Options
Arrow buttons to set Z height
VIAFLO 96/384 – Operating instructions V11 Description of the device
2.2.4 Touch wheel
The Touch wheel is fully operational with only one hand. Rotational finger movements translate into up and down cursor movement on the display. The Touch wheel is fully functional with the use of latex gloves.
Move finger on the Touch wheel to choose (and highlight) an option on the display. Press OK (14 make the selection.
When a setting dial is displayed, spin the Touch wheel to change the value and press OK.
2.2.5 Left and right arrow buttons
At times, you will see and on a display screen. These buttons are used to select options.
Press to select the option indicated with the left arrow (HELP, in the example beside). Press  to select the option indicated with the right arrow (RUN, in the example).
2.2.6 PURGE button
During pipetting, you can interrupt the current pipetting protocol and purge all remaining liquid currently in the GripTips. To do so, press PURGE (16
).
) to
The pipette will display a prompt. To proceed, press and release the RUN key (17
of the dispense, the first step in the current program will be displayed.
2.2.7 RUN key
Press and release the RUN key (17 special pipetting operations.
During dispense, you can press and hold the RUN key to perform a two-step blowout, see
Blowout modes” on page 23.
“4.3.3
www.integra-biosciences.com 11
) to initiate aspiration, dispense, mix, purge, and
). Upon completion
VIAFLO 96/384 – Operating instructions V11 Installation
2
7

3 Installation

3.1 Operating environment

VIAFLO 96/384 has been designed for use in a laboratory. It shall be operated in a dry and dust-free location with a temperature of 5– 40 °C and a maximal (non-condensing) relative humidity of 80 %.

3.2 Setting up and moving of the instrument

The VIAFLO 96/384 must be set up on a cleaned, dry and horizontal surface.
.
Hold the Carrying handle (2 Instrument deck (7
) to lift VIAFLO 96/384.
) and the
WARNING
The device must not be carried on the control unit because it will be damaged.
Leave at least 5 cm of free space on the back for air circulation and to allow easy connection of the mains cable.
ARNING
W
It must always be possible to manually disconnect the plug of the VIAFLO 96/ 384 from electricity supply. The corresponding socket shall be within easy reach of the operator and be clearly labeled as the disconnecting device of VIAFLO 96/384.
Only use a 3 core mains cable with protective earth to connect VIAFLO 96/384 with the power source. The socket is located on the reverse side of the instrument deck.
Put the Plate holders (8
) on the Instrument deck (7) on Position A and/or B. They can both be placed on either side. Move the Plate holder from left to right until the two bolts click into the two holes.
Alternatively, put the three position stage on the Instrument deck (7
) until the bolts click
into the four holes, see “8.1 Accessories” on page 55).
12 INTEGRA Biosciences AG
VIAFLO 96/384 – Operating instructions V11 Installation
WARNING
VIAFLO 96/384 must be secured with the “Park Head” function before carrying.
Before VIAFLO 96/384 can be relocated, the pipetting and base unit must be fixed. Select “Park Head” from “Calibration & Service” of the Toolbox menu, see “3.4.5 Calibration & Services” on page 18. Make sure GripTips have been ejected, clear the Instrument deck (7
) and press the RUN key (17) to start the park routine. The unit moves to the park position and is anchored on the base with a bolt. Switch off VIAFLO 96/384 and disconnect it from the electricity mains.

3.3 Getting started

Turn on VIAFLO 96/384 (see "Turn on/off the device" on page 21). To adapt the device to the appropriate applications, select the following functions of the Toolbox menu in this order:
Change head: Select the toolbox menu “Change head” and insert the appropriate
pipetting head, see "
Load 96/384 GripTips, see "Attaching and removing GripTips" on page 21
Position settings - Head alignment: Define the center of the wells of a 96-or 384-well
plate, see "
Position settings - Tip align: VIAFLO 96/384 base unit can be moved left and right
(X direction). Tip align locks onto the wells of a plate and helps to guide the tips into the microplate wells.
Z-Height: VIAFLO 96/384 pipetting unit can be moved up and down (Z direction). A
minimum height can be defined in every pipetting mode, e. g. to set optimal tip immer­sion depth, see “4.3.2
Preferences: Define system parameters, see "Preferences" on page 16.
Position Settings" on page 16.
Change head" on page 14.
Set Z-Position” on page 23.
www.integra-biosciences.com 13
VIAFLO 96/384 – Operating instructions V11 Installation

3.4 Toolbox - adapt your VIAFLO 96/384

The Toolbox provides options to adapt the device to appropriate applications, setting personal preferences, calibration, computer connectivity and storing owner information.
Toolbox mode Description
Change Head Allows to change the pipetting head. Position Settings Sets the tip alignment, head alignment and Z-height. Preferences Customizes the system parameters. Calibration & Service Sets calibration and service history options and parks the head. Communications Enables communication between your VIAFLO 96/384 and a
PC.
Device Information View your pipette’s serial number, software version and set a
personal ID. Language Sets display language. Write Protect Protects programs or menu options from modification.
The Help information describes the modes and some settings, press  to select the Help option.
3.4.1 Change head
N
OTE
VIAFLO 96 can only be used with 96-channel pipetting heads, while VIAFLO 384 is compatible with both 96- and 384-channel pipetting heads.
The following pipetting heads are available:
Pipetting head Channels Volume range
12.5 µl 96 0.5–12.5 µl 50 µl 96 2–50 µl
125 µl 96 5–125 µl 300 µl 96 10–300 µl
1250 µl 96 50–1250 µl
12.5 µl 384 0.5–12.5 µl 50 µl 384 2–50 µl
125 µl 384 5–125 µl
14 INTEGRA Biosciences AG
VIAFLO 96/384 – Operating instructions V11 Installation
a
a
b
Remove the pipetting head from the case. Make sure the pipetting head is in extended position (see
image to the left). If necessary, pull apart the Piston plate (a) on the top as far as possible.
To install the appropriate pipetting head select the Toolbox option “Change Head”.
Loosen the knob and remove the Side cover (6 Remove the existing head, if installed. Push the appropriate pipetting head into the slide block
and make sure the Piston plate (a) slides onto the brazen rails.
Mount the Side cover, screw the knob tightly and press OK to continue.
OTE
N
Pipetting heads with a serial number >50274 feature an ejection lever (b) to facilitate the removal of the head. Pull up the lever to loosen the pipetting head, then remove it from the instrument.
Check whether the Pipet factor on the display corresponds to the Pipet factor labeled on the side of the head itself or indicated in the latest calibration certificate. Press OK  if the factors coincide. If they don't, click Edit.
).
With the factor “Pipet” selected click Edit to change the Pipet factor accordingly. Using the Touch Wheel, select the digit to be modified, then press OK and use the Touch Wheel to select a number. Press OK to confirm. Once finished, press Save to apply the correct Pipet factor to the instrument.
Reset sets the factor to 1.0000. Changing the factor “Repeat” is only necessary under special circumstances. Refer to the calibration document (PROTO_VIAFLO96_384_ calibration), available on request.
3.4.2 Storage of pipetting heads
Once a pipetting head is removed from VIAFLO 96/384 it must be stored in the appropriate casing to protect it from dust and mechanical damage.
AUTION
C
Pipetting heads with 384-channels must always be stored in extended position.
www.integra-biosciences.com 15
VIAFLO 96/384 – Operating instructions V11 Installation
3.4.3 Position Settings
Position settings contain options that help to find correct pipetting positions. These settings are globally stored and are valid for all pipetting heads. Perform these alignments with GripTips attached, see “4.2 Attaching and removing GripTips” on page 21.
Position settings
Tip Align Optimally aligns the tips to target the center of the wells
Head Align­ment
Z Height Activate to enable overriding of the defined Z-height.
Description Range
on a 96-or 384-well plate.
•Select Position A or B and press OK to activate the tip alignment.
• Highlight the Strength option and press OK. Set align- ment support strength 1 (low) to 3 (high). Press OK.
Column Detent activates column positions of a plate on Position A or B, e. g. for serial dilutions. Press OK.
Press to save.
Use a 96 well plate for a 96-channel and a 384-well plate for a 384-channel pipetting head to define the center of the wells. This setting is required only once per pipetting head. Move the unit to Position A or B until the selected position, e.g. A, is displayed.
• Move the unit down and align the tips to the center of the wells. Select Set to save this position.
• Repeat at position B.
Keep pushing down the control unit to overcome the Z­height, see “4.3.2
Set Z-Position” on page 23.
/ (On/Off) 1-3
(Off) (A or B)
-
/ (On/Off)
3.4.4 Preferences
Preferences customizes your system parameters. Select an option and press OK to access. After changing desired settings, press to Save.
Preferences
Deck Bright­ness
16 INTEGRA Biosciences AG
Description Range
Sets brightness of deck illumination.
• Use the Touch wheel to change the brightness: off, 1 (weak) to 10 (bright). Press OK.
Off 1-10
VIAFLO 96/384 – Operating instructions V11 Installation
Handle Sen­sitivity
Sound Select an option and press OK to change the status of
Display Brightness
Main Menu Select a function to be hidden from the main menu (Off)
Touch Wheel
Pipetting Select an option and press OK.
Sets the sensitivity of the control unit to move the pipetting unit and base unit.
• Use the Touch wheel to set a sensitivity 1 (low) to 9 (high). Press OK.
the beep tone between On and Off:
Step Complete: at the end of a program step
Program Complete: at program completion
Purge Key: when PURGE is pressed.
Messages: when a message appears.
Error Message: when an error message appears or when illegal data entry is attempted.
Touch Wheel: When spinning the Touch wheel.
Last dispense: before the last dispense in Repeat Dispense and Variable Dispense.
Sets brightness of the display. Use the Touch wheel to change the brightness: 1 (dim) to 10 (bright). Press OK.
and press OK:
• Pipet, Repeat Dispense, Dilute, Pipet/Mix, etc.
• The Automatic mode is by default deactivated. To acti­vate it, change the status to “” (On). See section 5.3.3 for a description of the Automatic mode.
Adjust your Touch wheel spin sensitivity. Low, Medium,
Purge Key Speed: Choose the desired purge speed.
Blowin Delay: Choose a timed delay between the blowout and the blowin at the end of a dispense, if no two step blowout is performed, see “4.3.3 modes” on page 23.
Extend Volume (not available for 12.5 µl pipetting head): For pipetting below the volume range specified: 50 μl pipette: (1.0)–2–50 μl 125 μl pipette: (2.0)–5–125 μl 300 μl pipette: (5.0)–10–300–(310) μl 1250 μl pipette: (25)–50–1250 μl The volumes in brackets refer to extended volumes, e.g. extend the minimal pipetting volume of a 125 μl head from 5 μl to 2 μl.
NOTE
Extending the volume range is not recommended. Full functionality and specified accuracy/precision cannot be guaranteed.
Blowout
1-9
/ (On/Off)
1-10
/ (On/Off)
High
1-10
None/
0.5-5.0 s
/ (On/Off)
www.integra-biosciences.com 17
VIAFLO 96/384 – Operating instructions V11 Installation
3.4.5 Calibration & Services
These options enable you to set calibration features, review service history and move the pipetting head into the parking position.
Calibration & Services
Calibration Allows for re-calibration of VIAFLO 96/384 to restore
Calibration Reminder
Service His­tory
Park Head Fixes the pipetting and base unit for safe transportation.
After changing desired settings, press to Save.
Description Range
accuracy. The calibration factors for Pipet and Repeat type are displayed. To edit the calibration volumes, press .
Target Volume: This is the volume you are interested in using for the calibration.
Actual Volume: This is the measured volume obtained when dispensing the target volume.
Current Factor: Displays the factor currently in use. This factor should be the same as specified on the head or the latest calibration certificate.
Reset: Resets the correction factor back to the origi­nal factory setting. Press to apply the factory setting.
Sets a calibration reminder based on a number of pipetting cycles. When the calibration reminder is displayed, press any key to confirm. However, the reminder will reappear every time the pipette is turned on until you change the reminder time or use the reset option.
Reminder: Press OK to turn the reminder timer On or Off.
Cycles: Use the Touch wheel to set a reminder inter- val for calibration (time in thousands of cycles). Press to set the timer to the defined calibration interval.
Remind in/Total Cycles: Displays the residual amount of cycles before calibration is required.
Reset: Resets the timer to the defined calibration interval. Press to enable.
Press to save. Displays notes of any service that took place on the
VIAFLO 96/384 listed newest entry first.
• Make sure GripTips are ejected, clear the deck and press the RUN key to start the park routine. The pipette moves to park position and can then be turned off.
-
/ (On/Off) 1k - 240k cycles
-
-
18 INTEGRA Biosciences AG
Loading...
+ 39 hidden pages