Waste Electrical and Electronic Equipment 2012/19/EU
Battery Directive2006/66/EC
EU Regulations
Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH)
Capacity Labelling of Portable Secondary Batteries 1103/2010
Ecodesign - Power supplies278/2009
Standards for EU
Safety requirements for electrical equipment for measurement,
control and laboratory use - General requirements.
Electrical equipment for measurement, control and laboratory use EMC requirements.
7205 Zizers, Switzerland
1907/2006
EN 61010-1: 2010
EN 61326-1: 2013
Standards for Canada and USA
Safety requirements for electrical equipment for measurement,
control and laboratory use - General requirements.
Safety requirements for electrical equipment for measurement,
control and laboratory use - General requirements.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including
interference that may cause undesired operation.
CH-7205 Zizers, Suisse
T +41 81 286 95 30
F +41 81 286 95 33
info@integra-biosciences.com
www.integra-biosciences.com
Customer service
Please contact your local INTEGRA Biosciences representative, see
www.integra-biosciences.com
or contact info@integra-biosciences.com.
INTEGRA BIOSCIENCES CORP.
Hudson, NH 03051, USA
T +1 603 578 5800
F +1 603 577 5529
English
www.integra-biosciences.com3
Page 4
VACUSIP - Operating instructions V04Introduction
1Introduction
These operating instructions contain all the information required for installation, operation
and maintenance of the VACUSIP aspiration system.
English
1.1Intended use
This is a general-purpose laboratory instrument. Any use of this instrument in a medical
or IVD setting is under the sole responsibility of the user.
VACUSIP is a vacuum-based system for aspiration of non explosive aqueous solutions,
such as biological buffers and media.
If the VA CUS IP is used i n a man ner not specified by INTEGRA Biosciences, the protection
provided by the VACUSIP may be impaired.
1.2Safety notes
WARNING
Read these operating instructions carefully before use and pay particular
attention to sections containi ng thi s symb ol .
1.2.1 General
1) Do not carry out any conversions and alterations on the device.
2) Defective parts may only be replaced with original INTEGRA Biosciences spare parts
according to the INTEGRA Biosciences operating or service instructions.
3) Do not use or charge VACUSIP in an atmosphere with danger of explosion. Also, do
not aspirate highly flammable liquids such as acetone or ether.
4) When handling dangerous substances, comply with the material safety data sheet
(MSDS) and with all safety guidelines such as the use of protective clothing and safety
goggles.
5) Prolonged exposure of VACUSIP to UV-light can cause discoloration and/or yellowing
of the plastic housing. However, this will not affect the performance of the device in any
way.
rubber products, O-rings, filter) should be stored in a light protected place at ambient
temperature to prevent premature material aging caused by prolonged exposure to UVlight.
Regardless of the listed safety notes, additionally applicable regulations and guidelines of
trade associations, health authorities, trade supervisory offices, etc. must be observed.
Please visit our website www.integra-biosciences.com
information regarding REACH classified chemicals contained in our products.
on a regular basis for up to date
4INTEGRA Biosciences AG
Page 5
VACUSIP - Operating instructions V04Introduction
1.2.2 Battery (model 159 000)
1) Charge the VACUSIP device using only the charger supplied by INTEGRA
Biosciences.
2) Old Li-ion batteries may cause a safety risk. We recommend to replace the battery
after 3 years of use. Also replace the battery if the charging intervals are unusually
short or if the charging takes much longer than usual (6 hours or more). These are
indicators that the battery has reached the end of its life-cycle.
3) Li-ion technology bears the risk of thermal runaway and cell rupture if the battery was
damaged. Do not expose the battery to heat (> 60 °C) and avoid mechanical stress.
Batteries which were subject to deep discharges may develop internal short circuits,
leading to an increased self-discharge rate and heating during battery charging. This
may also result in thermal runaway and cell rupture.
4) To extend the battery life-cycle, it is recommended to charge the battery every 2
months if the VACUSIP is not used regularly. If the VACUSIP is not used for more than
6 months, unplug the battery, see 3.3.5
5) If despite all safety measures the VACUSIP Li-ion battery catches fire and toxic
(typically dense, white) fumes emanate from the VACUSIP device, immediately
evacuate the working area. The VACUSIP housing is made of special flame retardant
material that has proven its capability to contain a battery fire. During a battery fire the
housing surface may become very hot. Do not touch the housing by hand and wait at
least 20 minutes after fumes have disappeared before approaching the device. Do no
longer use the VACUSIP device.
.
English
www.integra-biosciences.com5
Page 6
VACUSIP - Operating instructions V04Description of the device
1
3
2
4
5
6
7
8
13
9
10
11
2Description of the device
Scope of delivery
• VACUSIP base unit without battery (model 159 010) or with battery (model 159 000)
• Mains adapter
• INTEGRA glass bottle 500 ml with standard GL 45 lid
English
• GL 45 lid with tube fittings
• Tubing and hydrophobic filter with an additional spare filter
• VACUSIP hand operator
• 8-channel stainless steel tips
• 1-channel stainless steel tip
• 1-channel adapter for disposable tips
1 On/off switch
2 Indicator LED
3 Base unit
4 Stainless steel rod
5 VACUSIP hand operator
6 Hand operator mount
7 Lid with tube fittings
8 500 ml bottle for liquid collection
Rear view
12
9 Hydrophobic filter
10 Silicone filter adapter
11 Air outlet
12 Opening to insert stainless steel rod
6INTEGRA Biosciences AG
13 Mains connection socket
Page 7
VACUSIP - Operating instructions V04Installation
3Installation
3.1Power supply
Insert the mains adapter cable into the VACUSIP mains connection socket (13) and plug
it into the power source.
ARNING
W
Use an original INTEGRA Biosciences mains adapter only (see “6.1
Specifications” on page 12 for voltage requirements)
3.2Charging of VACUSIP battery model
A full charge takes 5 hours. Before the first use, VACUSIP should be charged for at least
5 hours.
When the LED (2
prolong battery life. VACUSIP has an integrated overcharge protection.
VACUSIP can be used while it is being charged.
3.3Set-up
3.3.1 Installing the VACUSIP hand operator mount
Find the enclosed stainless steel rod (4) and push it into the opening (12).
Attach the hand operator mount (6
) starts to blink green, the battery needs to be recharged immediately to
) to the top of the stainless steel rod (4).
ARNING
W
Do not carry VACUSIP by holding the rod or the hand operator mount. Always
grab the instrument at the back to carry it around.
English
3.3.2 Assembling the hand operator
Insert the inner part in the silicone
cover, as shown on the picture
beside.
3.3.3 Connecting the filter
Attach the shorter silicone tubing to the filter (9) and then plug the filter into the silicone
filter adapter with the blue side facing the instrument.
VACUSIP should never be used without a filter.
www.integra-biosciences.com7
Page 8
VACUSIP - Operating instructions V04Installation
3.3.4 Connecting the silicone tubing
Place the bottle in the instrument and screw on the green lid with tube fittings (7).
Attach the silicone tubing coming from the filter to the fitting on the lid named “FILTER”.
Attach the longer silicone tubing to the fitting in the lid named “LIQ” (liquid) and the other
English
end to the VACUSIP hand operator (5
3.3.5 Replacing the battery
Unscrew the 5 screws on the bottom of the device with a
Phillips screwdriver nr. 1 and remove the bottom lid.
Unplug the old battery and replace it with an original
INTEGRA Biosciences rechargeable battery, see “7
Accessories and consumables” on page 14. Make sure that
it is inserted with the correct polarity (+/-).
Place back the battery in the battery compartment with the
wire facing to the inside of the instrument.
Close the bottom lid and fix it with the 5 screws.
).
8INTEGRA Biosciences AG
Page 9
VACUSIP - Operating instructions V04Operation
4Operation
4.1Vacuum activation and notes on use
Switch on VACUSIP to start the vacuum pump (the LED lights up green).
Attach a suitable adapter to the hand operator.
To aspirate press the soft buttons on both sides of the
VACUSIP hand operator.
Once the vacuum is reached, the pump stops until the vacuum falls below threshold.
VACUSIP switches off automatically after the pump has not been running for 10 minutes
or 40 minutes after initial switch-on. This feature can be temporarily deactivated by
pressing the power button for at least 4 seconds when switching on the instrument. The
power-off function is automatically restored after turning off VACUSIP.
Check the level of liquid in the bottle frequently. The liquid level should never reach the lid.
Prevent liquid or foam overflow from the collection bottle into the filter. In case the filter
gets accidentally wetted or soiled, exchange it immediately or otherwise the retention of
further liquid cannot be guaranteed. Note that the hydrophobic filter holds back aqueous
solutions, whereas retention of apolar liquids is limited.
ARNING
W
When working with dangerous substances, you mu st comply with the material
safety data sheet (MSDS) and the additionally applicable regulations of trade
associations, health authorities, supervisory offices, etc. In addition, when
working with hazardous volatile substances or biohazardous agents, either
place the VACUSIP inside a biosafety cabinet or connect the instrument's air
outlet (11) to a safety ventilation by means of a tube (ID=4 mm) and an
aspiration needle (Part. No. 155 502).
4.2Code of indicator LED
English
LED codeMeaning
Constantly greenUnit is powered on and no errors occurred.
Blinking green Battery level low.
Blinking red once
per second
Blinking red 5 times
per second
Constantly redBattery error. Press the power button to continue using the
Blinking red and green Error in vacuum sensor calibration. Please contact your supplier.
www.integra-biosciences.com9
Leakage detection. Pump turns off if end vacuum has not been
reached after 5 minutes.
Vacuum generation is too fast (~100 mbar/s). Indicates a
clogged part.
mains power supply.
Page 10
VACUSIP – Operating instructions V04Operation
English
(model 159000) is discharged. Check indicator LED light.
VACUSIP automatically switches off if it is not used or 40
minutes after initial switch-on.
tubing, filter and hand operator for leaks. If the maximum
vacuum is no reached for 5 min, the pump is turned off.
Empty bottle and exchange filter.
Unit automatically switched off.Switch on the unit (LED lights up green).
4.3Troubleshooting
ProblemProbable causeRemedy
Device is not running.No electrical power.Check if mains adapter cable is plugged in or if battery
10INTEGRA Biosciences AG
Aspiration speed is low. Leak in the vacuum system.Close the bottle lid tightly and attach tubing firmly. Check
Liquid overflowed and blocked the
Tubing is blocked.Clean or exchange the tubing.
Hand operator is clogged.Take apart the hand operator to clean it. Use an ultrasonic
filter.
bath to clean it of debris.
The LED blinks red rapidly.
Check if filter is blocked and exchange it, if necessary.
Replace the hand operator.
Check if the filter adapter is blocked.
Clogging of a part close to the
pump.
VACUSIP turns off after
a few seconds or
minutes.
The pump automatically stops if the end vacuum is not
reached after 5 min. Check for leaks in the system and push
the power button again.
Leak in the vacuum system.The LED blinks red slowly.
Page 11
VACUSIP - Operating instructions V04Maintenance
5Maintenance
5.1Cleaning and servicing
WARNING
Always switch off the VACUSIP and disconnect it from the mains when carrying
out maintenance work.
Clean the VACUSIP housing with a lint-free cloth lightly soaked with mild soap solution in
distilled water or with a 70 % dilution of Isopropyl or Ethanol. Never use bleach (sodium
hypochlorite) or other solvents.
It is recommended to change the hydrophobic filter regularly, at least once a year.
Exchange the filter immediately if liquid has entered.
When working with solvents and other hazardous materials always rinse the hand
operator and tubing with water followed by 70% Ethanol.
5.2Autoclaving and chemical sterilization
5.2.1 Autoclaving
All parts contacting the liquid can be autoclaved: 500 ml glass bottle, lid with tube fittings,
silicone tubing and hand operator. The hand operator should be taken apart for optimal
decontamination. To disassemble it, hold the silicone cover and pull out the inner part.
Recommendation regarding autoclaving: glass bottle for at least 20 min at 121 °C.
The bottle lid has to be always unscrewed and kept loosely on the bottle during
autoclaving.
Silicone may become brittle after extensive autoclaving. Replace the tubings if they are
damaged.
ARNING
W
The hand operator can be autoclaved sporadically for decontamination,
however it is not recommended to autoclave it regularly.
The filter and the instrument itself cannot be autoclaved.
5.2.2 Chemical sterilization
Two methods for chemical inactivation of biohazardous agents are commonly used:
• A suitable disinfectant is added to the full collection bottle and left to react for the
recommended time.
• The disinfectant is added to the empty bottle so that the aspirated liquid gets sterilized
as it accumulates in the bottle. Use disinfectants that are free of chlorine or other
corrosive agents, especially when corrosive vapors are generated and aspirated.
English
www.integra-biosciences.com11
Page 12
VACUSIP - Operating instructions V04Technical data
5.3Equipment disposal
The VACUSIP device must not be disposed of with unsorted municipal waste.
Do not dispose of the device in a fire.
VACUSIP (159000) contains a Li-ion battery. Do not modify the battery in any
English
way. Dispose of the VACUSIP device and the battery separately in accordance
with the laws and regulations in your area governing disposal of devices
containing Li-ion batteries.
6Technical data
6.1Specifications
Vacuum range–250 mbar +/- 20%
Flow rate (air)Pump: 1.5 l/min +/- 30% at ambient pressure
125 x 120 x 162 mm (base unit)
Height: unit with bottle and hand operator mount 345 mm
Unit complete: 970 g (with glass bottle, tubing and hand operator)
Device input: 8–10 VDC, 6 W
Temperature: 5 – 40 °C
Humidity: max. rel. humidity 80% for temperatures up to 31 °C
decreasing linearly to 50% rel. humidity at 40 °C.
6.2Chemical compatibility
The table below lists VACUSIP parts that come into contact with the aspirated liquid or its
aerosols and vapors, and rates the compatibility of these parts to a few of the chemicals
commonly used in laboratories. To determine the compatibility of a component to a
chemical not listed in the table, please consult one of the several tables available on the
internet. Note that the rating refers to soaking in the concentrated chemical; however,
more relevant here is the attenuated effect resulting from vapors and the diluted chemical.
It is recommended to test the compatibility of relevant components to a specific chemical
prior to extensive use.
INTEGRA Biosciences does not warrant that the information in the table is accurate or
complete and that any material is suitable for any purpose.
12INTEGRA Biosciences AG
Page 13
VACUSIP – Operating instructions V04Technical data
Phenol PhOH
Dimethyl sulfoxide
DMSO
Nitric acid (20 %)
Hydrochloric acid
HCl (20 %)
Sodium acetate
(3M, pH 5.2)
Sodium hydroxide
NaOH
Isopropyl alcohol IPA
Ethanol EtOH
Acetic acid AcOH
English
JAVEL (e.g. NaClO)
INOXBCAABACAAA
Outer partSilicone ABAAACACAC
Adapters POMCCAAAABCAA
for continuous use.
determined by test.
Compatibility ratings:
A = Good, no or minor effects.
B = Fair, moderate effects, not recommended
Chemical compatibility chart
www.integra-biosciences.com13
C = Critical, not recommended, suitability to be
PartsMaterials
Hand operatorInner partPPAAAAAAAAAA
LidsPPAAAAAAAAAA
GasketEPDM AAAAAAAAAA
HousingPVCACBAAAAACB
TubingSilicone ABAAACACAC
BottleSingle-usePPAAAAAAAAAA
FilterMembranePTFEAAAAAAAAAA
Filter adapterSiliconeABAAACACAC
VACUSIP baseHousingPPAAAAAAAAAA
Pump housingPOMCCAAAABCAA
Pump membraneNBRABBAAAACBC
Page 14
VACUSIP - Operating instructions V04Accessories and consumables