Integra 4014, 4015, 4021, 4022, 4023 Operating Instructions Manual

...
VIAFLO Electronic Pipettes Operating instructions
900-00002-00_V05
Declaration of conformity | Konformitätserklärung | Déclaration de conformité | Declaración de conformidad | Dichiarazione di conformità
INTEGRA Biosciences – 2 Wentworth Drive, Hudson, New Hampshire 03051
declares on its own responsibility that the product | erklärt in alleiniger Verantwortung, dass das Produkt | déclare sous sa responsabilité exclusive, que le produit | declara bajo su propia responsabilidad que el producto | dichiara sotto la propria responsabilità che il prodotto
VIAFLO Electronic Pipettes
in accordance with EC directives | gemäss der EU-Richtlinien | est conforme au terme directives CE | de acuerdo con las directivas CE | in conformità alle direttive CE
2006/95/EC Low voltage equipment
2004/108/EC Electromagnetic compatibility
2011/65/EC Restriction of Hazardous Substances
2002/96/EC Waste Electrical and Electronic Equipment
is in compliance with the following normative documents: | mit den folgenden normativen Dokumenten übereinstimmt: | aux documents normatifs ci-après: | cumple las documentos normativos: | soddisfa le normative seguenti:
EN 61010-1 Safety requirements for electrical equipment for measurement,
EN 61326-1 Electrical equipment for measurement, control and laboratory
Standards for Canada and USA
CAN/CSA-C22.2 No. 61010-1
UL Std. No. 61010-1 Safety requirements for electrical equipment for measurement,
FCC, Part 15, Class A Emission
Models: 4011, 4012, 4013, 4014, 4015, 4021, 4022, 4023, 4024, 4031, 4032, 4033, 4034, 4041, 4042, 4121, 4122, 4123, 4124, 4131, 4132, 4143, 4144, 4163, 4164
control and laboratory use - General requirements.
use - EMC requirements.
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
control and laboratory use - General requirements.
Hudson, New Hampshire, USA - April 5, 2013
Gary Nelson President
George Kalmakis VP Operations
VIAFLO Electronic Pipettes – Operating instructions V05
Table of Contents
Chapter 1 Introduction
1.1 Symbols used...................................................................................7
1.2 Intended use.....................................................................................7
1.3 Safety notes .....................................................................................8
Chapter 2 Description of the device
2.1 Scope of delivery..............................................................................9
2.2 Pipette configurations.......................................................................9
2.3 Overview of the VIAFLO Electronic Pipettes..................................10
2.3.1 VIAFLO pipette parts ..........................................................10
2.3.2 Back view............................................................................11
2.3.3 Display ................................................................................11
2.3.4 Touch wheel........................................................................12
2.3.5 Left and right arrow buttons ................................................12
2.3.6 Purge button .......................................................................12
2.3.7 Run key...............................................................................12
2.3.8 Tip ejector ...........................................................................13
2.3.9 Reset button........................................................................13
Chapter 3 Installation
3.1 Operating environment ...................................................................14
3.2 Charging the battery.......................................................................14
3.2.1 Charging the battery on a stand..........................................14
3.2.2 Charging the battery with UPS............................................15
3.3 Toolbox - adapt your VIAFLO Electronic Pipettes..........................15
3.3.1 Preferences.........................................................................16
3.3.2 Calibration & Services.........................................................17
3.3.3 Communications .................................................................18
3.3.4 Owner Information ..............................................................18
3.3.5 Help Language....................................................................19
3.3.6 Write protect........................................................................19
3.3.7 Password key......................................................................19
www.integra-biosciences.com 3
VIAFLO Electronic Pipettes – Operating instructions V05
Chapter 4 Operation
4.1 Turn on/off the device.....................................................................20
4.2 Attaching and removing GripTips ...................................................20
4.3 Start pipetting .................................................................................21
4.3.1 Pipetting..............................................................................21
4.3.2 Blowout modes ...................................................................21
4.3.3 Recommendations for pipetting ..........................................22
4.4 Pipetting options and settings ........................................................23
4.4.1 Edit option...........................................................................23
4.4.2 Volume selection.................................................................23
4.4.3 Speed selection ..................................................................24
4.4.4 Pace....................................................................................25
4.4.5 Count, Mix Cycle and Rows................................................25
4.4.6 Help.....................................................................................25
4.5 VOYAGER tip spacing ...................................................................26
4.5.1 Set the number of tip positions and the tip spacing ............26
4.5.2 Tip spacing operation..........................................................27
4.5.3 Homing the tips...................................................................27
4.6 Troubleshooting/FAQ .....................................................................28
4.6.1 General ...............................................................................28
4.6.2 Electronic ............................................................................29
Chapter 5 Pipetting modes
5.1 Overview pipetting modes ..............................................................31
5.2 Detailed description of pipetting modes .........................................32
5.2.1 Pipet mode..........................................................................32
5.2.2 Repeat dispense mode .......................................................33
5.2.3 Sample dilute mode ............................................................34
5.2.4 Pipet/mix mode ...................................................................34
5.2.5 Manual pipet mode .............................................................35
5.2.6 Reverse pipet mode............................................................36
5.2.7 Variable dispense mode .....................................................37
5.2.8 Variable aspirate mode .......................................................38
5.2.9 Sample dilute/mix mode......................................................39
5.2.10 Serial dilution mode ..........................................................40
5.3 Custom step-based programming mode ........................................41
5.3.1 Create a custom program ...................................................41
5.3.2 Program tip adjustments in a custom program ...................42
5.3.3 Modify existing programs ....................................................42
5.3.4 Example of custom mode ...................................................43
4 INTEGRA Biosciences AG
VIAFLO Electronic Pipettes – Operating instructions V05
Chapter 6 Maintenance
6.1 Cleaning .........................................................................................44
6.1.1 Assembly of single channel pipettes...................................44
6.1.2 Assembly of multichannel fixed spacing pipettes................46
6.2 Sterilization.....................................................................................48
6.2.1 Autoclaving the disassembled components ........................ 49
6.3 Servicing.........................................................................................49
6.3.1 Shipping to INTEGRA Biosciences.....................................49
6.3.2 Changing O-rings of tip fittings............................................49
6.3.3 Lubrication ..........................................................................50
6.4 Calibration ......................................................................................51
6.4.1 Materials .............................................................................51
6.4.2 Definitions ...........................................................................51
6.4.3 Test Conditions and Environment.......................................52
6.4.4 Pre-wetting of tips ...............................................................52
6.4.5 Leak test .............................................................................53
6.4.6 Obtaining the Actual Volume ..............................................54
6.4.7 Calculation of accuracy and precision ................................55
6.4.8 Adjusting VIAFLO Electronic Pipettes.................................56
6.5 Equipment disposal ........................................................................58
Chapter 7 Technical Data
7.1 Environmental conditions ...............................................................59
7.2 Specification of the device..............................................................59
7.3 Pipetting speed...............................................................................59
7.4 Pipette Specifications .....................................................................60
7.5 Z Correction Factors.......................................................................63
Chapter 8 Accessories
8.1 Accessories ....................................................................................64
8.2 Consumables .................................................................................64
www.integra-biosciences.com 5
VIAFLO Pipettes – Operating instructions V05
Imprint
© 2013 INTEGRA Biosciences AG
All rights to this documentation are reserved. In particular the rights of reproduction, processing, translation and the form of presentation lie with INTEGRA Biosciences AG. Neither the complete documentation nor parts thereof may be reproduced in any way, or stored and processed using electronic media or distributed in any other way without the written consent of INTEGRA Biosciences AG.
This operating instruction manual has part number 900-00002-00, the version is V05. It applies to firmware version 2.44 or higher until a newer revision is released.
VIAFLO, VOYAGER, VIALINK, and GripTip are trademarks of INTEGRA Biosciences, Hudson, NH.
Manufacturer INTEGRA Biosciences AG
CH-7205 Zizers, Switzerland T +41 81 286 95 30 F +41 81 286 95 33 info@integra-biosciences.com www.integra-biosciences.com
INTEGRA Biosciences Corp.
Hudson, NH 03051, USA T +1 603 578 5800 F +1 603 577 5529
Customer service
Please contact your local INTEGRA Biosciences representative. To find out name and address of your local representative go to www.integra-biosciences.com
Further information and operating instructions in other languages are available on www.integra-biosciences.com
6 INTEGRA Biosciences AG
.
or on request info@integra-biosciences.com.
VIAFLO Pipettes – Operating instructions V05 Introduction

1 Introduction

These operating instructions contain all the information required for installation, operation and maintenance of the VIAFLO Electronic Pipettes. This chapter informs about the symbols used in these operating instructions, the intended use of the VIAFLO Electronic Pipettes and the general safety instructions.

1.1 Symbols used

The operating instructions specifically advise of residual risks with the following symbols:
ARNING
W
This safety symbol warns against hazards that could result in injury. Italso indicates hazards for machinery, materials and the environment. It is essential that you follow the corresponding precautions.
CAUTION
This symbol cautions against potential material damage or the loss of data in a microprocessor controller. Follow the instructions.
OTE
N
This symbol identifies important notes regardi ng the correct operation of the device and labour-saving features.

1.2 Intended use

VIAFLO Electronic Pipettes are microprocessor controlled and stepper motor driven pipettes. They are used for aspirating and dispensing liquids in the volume range of 0.5 ­5000 µl using GripTip pipette tips.
www.integra-biosciences.com 7
VIAFLO Pipettes – Operating instructions V05 Introduction

1.3 Safety notes

VIAFLO Electronic Pipettes comply to the recognized safety regulations and is safe to operate. VIAFLO Electronic Pipettes can only be operated when in perfect condition and while observing these operating instructions.
The device may be associated with residual risks if it is used or operated improperly by untrained personnel. Any person operating the VIAFLO Electronic Pipettes must have read and understood these operating instructions, and particularly, the safety notes, or must have been instructed by supervisors so that safe operation of the device is guaranteed.
Use only an INTEGRA battery.
Regardless of the listed safety notes, additional applicable regulations and guidelines of trade associations, health authorities, trade supervisory offices, etc. must be observed.
Do not open or modify the VIAFLO Electronic Pipettes in any way. Repairs may only be performed by INTEGRA Biosciences AG or by an authorised after-sales service member.
Parts may be replaced with original INTEGRA Biosciences parts only.
ARNING
W
Do not use the VIAFLO Electronic Pipettes near flammable material or in explosive areas. Also, do not pipette highly flammable liquids such as acetone or ether.
When handling dangerous substances, comply with the material safety data sheet (MSDS) and with all safety guidelines such as the use of protective clothing and safety goggles.
CAUTION
Do not immerse the VIAFLO Electronic Pipettes in liquid. The fluid can damage internal parts. Avoid pipetting of liquids whose vapors could attack the materials PA (polyamide), POM (polyoxymethylene), FPM (fluor -rubber), NBR (nitrile-rubber), CR (chloroprene), silicone. Corrosive vapors could also damage metallic parts inside the device.
OTE
N
Prolonged exposure of the VIAFLO Electronic Pipettes to UV-light can cause discolouration and/or yellowing of the control unit. However, this will not affect the performance of the device in any way.
8 INTEGRA Biosciences AG
VIAFLO Pipettes – Operating instructions V05 Description of the device

2 Description of the device

2.1 Scope of delivery

• VIAFLO pipette
• Rechargeable battery (located inside the pipette, Li-ion, 3.7 V, 1050 mAh)
• Bag of spare O-rings (300 and 1250 µl versions only)
• Accuracy and precision certificate
AUTION
C
Verify the scope of delivery when unpacking the device and check for potential transportation damage. Do not operate a device that is damaged, instead contact your local dealer.

2.2 Pipette configurations

VIAFLO single and multichannel pipettes couple an easy-to-use touch wheel controller with intuitive programming. They are microprocessor controlled and stepper motor driven. In addition, the VIAFLO VOYAGER adjustable tip spacing pipettes provides you with the ability to vary tip spacing.
www.integra-biosciences.com 9
VIAFLO Pipettes – Operating instructions V05 Description of the device
1
2
5
6
7
8
3
4
9
11
12
10

2.3 Overview of the VIAFLO Electronic Pipettes

2.3.1 VIAFLO pipette parts
1 Display
2 Back button, to navigate backward
3 Touch wheel, spin to scroll and move the cursor
4 OK button, to make a selection
5 Left and right arrow buttons, for selections
6 Purge button, to empty tips
7 Run key, to start operations
8 Tip ejector
9 Finger hook, facilitates easy operation
10 Volume indicator label, color matches GripTip rack
insert.
11 Ejector sleeve
12 Tri-lobe tip interface
10 INTEGRA Biosciences AG
VIAFLO Pipettes – Operating instructions V05 Description of the device
13
14
15
16
RUN
Current mode
Active step
OK
2.3.2 Back view
13 Reset button
14 Power connector
15 Charging stand interface
16 Battery
2.3.3 Display
The Display shows all pipetting options.
Battery life indicator
Instruction Vol umes
Options
www.integra-biosciences.com 11
VIAFLO Pipettes – Operating instructions V05 Description of the device
2.3.4 Touch wheel
The Touch wheel is fully operational with only one hand. Rotational finger movements translate into up and down cursor movement on the display. The Touch wheel is fully functional with the use of latex gloves.
Move finger on the Touch wheel to choose (and highlight) an option on the display. Press OK (4 make the selection.
When a setting dial is displayed, spin the Touch wheel to change the value and press OK.
2.3.5 Left and right arrow buttons
At times, you will see and on a display screen. These buttons are used to select options.
Press to select the option indicated with the left arrow (HELP, in the example beside). Press to select the option indicated with the right arrow (RUN, in the example).
For Voyager models, these buttons are also used for tip spacing adjustments.
2.3.6 Purge button
During pipetting, you can interrupt the current pipetting protocol and purge all remaining liquid currently in the GripTips. To do so, press Purge (6
).
) to
The pipette will display a prompt:
To proceed, press and release the Run key (7 of the dispense, the first step in the current program will be displayed.
2.3.7 Run key
Press and release the Run key to initiate aspiration, dispense, mix, purge, and special pipetting operations. This button is centrally located for left- or right-handed actuation.
During dispense, you can press and hold the Run key to perform a two-step blowout, see “4.3.2 Blowout modes” on page 21.
12 INTEGRA Biosciences AG
). Upon completion
VIAFLO Pipettes – Operating instructions V05 Description of the device
2.3.8 Tip ejector
The tip ejector easily ejects tips from the tip cones.
The serial number can be found beneath the tip ejector. Press and hold the tip ejector in the down position to see the 7-digit serial number.
2.3.9 Reset button
The reset button (13 RAM in the pipette. The programs stored in memory are maintained. Once pressed, the VIAFLO startup screen will be displayed.
Press any key to continue and allow the instrument to initialize and home the pipette. The routine ends by displaying the Main Menu.
) is located on the back of the VIAFLO pipettes. It is used to reset the
www.integra-biosciences.com 13
VIAFLO Pipettes – Operating instructions V05 Installation

3 Installation

3.1 Operating environment

The VIAFLO Electronic Pipettes have been designed for use in a laboratory. They shall be operated in a dry and dust-free location with a temperature of 5–40 °C and a maximal (non-condensing) relative humidity of 80 %.

3.2 Charging the battery

All VIAFLO hand-held pipettes share the same rechargeable, long-life, lithium-ion battery that has a capacity of 1050 mAmp/hours. The battery should be charged for at least one hour before pipetting. A full charge takes 2.5 hours and will provide approximately 3500 pipetting cycles.
A battery indicator is provided on the pipette display (2.3.3
• Blinking red icon: Low battery. A recharging message will appear.
• Blinking grey icon: Pipette is recharging.
• Blinking green icon: Pipette is recharging and battery is 90% charged.
• Solid green icon: Battery is fully charged and the pipette is still connected to the mains adapter.
• Three static grey bars: Battery is fully charged.
AUTION
C
Use only the approved INTEGRA battery, power supply or charging stand. Use of an incompatible power transformer can damage the pipette.
Battery charging can take place in two ways, using a pipette charging stand or the Universal Voltage Power Supply (UPS), that allows for pipetting while charging the battery, see (“8.1
3.2.1 Charging the battery on a stand
Use one of the different charging stands - single position with two- (#4210) or four prong connector (#4211), or four position with two prong connector (#4215) - to charge the battery.
Accessories” on page 64).
Place the pipette on the charging stand by fitting the power receptacle (15, on the top back of the pipette) over the prong connector on the top of the stand.
Plug the cable of the adequate power supply to the socket at the base of the stand (see “8.1 on page 64).
) indicating different states:
Accessories”
AUTION
C
Always use the correct power supply for the charging stand.
14 INTEGRA Biosciences AG
VIAFLO Pipettes – Operating instructions V05 Installation
The pipette will turn fully on when placed on the stand. When the dimming time is achieved the startup screen is displayed on the pipette. When the turnoff time is reached, the pipette turns off. For disconnecting simply lift the pipette up from the stand.
3.2.2 Charging the battery with UPS
With the optional Universal Voltage Power Supply (UPS, #4200), you can use the pipette while charging through the line cord.
Insert the UPS connector into the receptacle on the top back of the pipette (14
It will take approximately 30 seconds of charging to store enough voltage to begin running the motor.
The pipette will turn fully on when the line cord is connected. The current screen remains displayed. If the pipette is idle while charging, the display may dim (see “4.1 Turn on/off the device” on page 20), but it will continue to show the battery life indicator. When the display shut off time is reached, the instrument will shut off.
). Plug the UPS into a wall outlet.

3.3 Toolbox - adapt your VIAFLO Electronic Pipettes

The Toolbox provides options to adapt the device to appropriate applications, setting personal preferences, calibration, computer connectivity and storing owner information.
Toolbox mode Description
Preferences Customizes the system parameters.
Calibration & Service Sets calibration and service history options.
Communications Enables communication between your VIAFLO Electronic
Pipettes and a PC.
Owner Information View your pipettor’s serial number and set a personal ID.
Help Language Sets language specific help.
Write Protect Protects programs or menu options from modification.
Password Key Displays the encrypted password.
www.integra-biosciences.com 15
VIAFLO Pipettes – Operating instructions V05 Installation
3.3.1 Preferences
Preferences customizes your system parameters. Select a preference and press OK to access.
Preference Description Range
Sound Simple tones indicate completion of operations and
errors. Select an option and press OK to change the status of the beep tone between On and Off:
Step complete: at the end of a program step
Program complete: at program completion
Purge key: when Purge is pressed.
Error message: when an error message appears or when illegal data entry is attempted.
Spinner: When spinning the Touch w h e el.
Last dispense: before the last dispense in Repeat Dispense and Variable Dispense.
Display Customizes your display. Press OK to select an option,
use the Touch wheel to display the desired value.
Start up display: Select the startup display at one of the following: VIAFLO Logo, None or Custom (up to two, uploaded with VIALINK). Press to Save your selection.
Display dim time: The display will dim after a set num­ber of minutes. A shorter dim time helps preserve bat­tery life. Press OK to save your selection.
Turn off time: By default, the pipette will turn off after 5 minutes of idle time. A shorter turn off time helps pre­serve battery life. You can change this setting. Press OK to save your selection.
Main Menu Select a function to be hidden from the main menu (Off)
and press OK, e.g. Tip Spacing, Pipet, Repeat Dispense, Dilute, Pipet/Mix. On/Off
Spinner Adjust your Touch wheel spin sensitivity. Low, Medium,
Pipetting Select an option and press OK.
Purge key speed: Choose the desired purge speed and press OK to save your selection.
Delayed blowin: Choose a timed delay between the blowout and the blowin (delay before the piston homes) at the end of a dispense, if no two step blowout is performed, see “4.3.2 Blowout modes” on page 21
After changing desired settings, press to Save.
On/Off
VIAFLO Logo, None, Custom 1 or 2
Never, 1-20 min
5-20 min, 1 hour
High
1-10
0/0.5/1/1.5 s
.
16 INTEGRA Biosciences AG
VIAFLO Pipettes – Operating instructions V05 Installation
3.3.2 Calibration & Services
These options enable you to set calibration features and review service history .
Calibration & Services
Calibration Allows for re-calibration of the VIAFLO Electronic
Calibration Reminder
Time or Cycles
Service His­tory
Calibrate Tip Spacing
After changing desired settings, press to Save.
Description Range
Pipettes to restore accuracy. The calibration factors for Pipet and Repeat type are displayed. To edit the calibration volumes, press .
Target Volume: This is the volume you are interested in using for the calibration.
Actual Volume: This is the measured volume obtained when dispensing the target volume.
Factory Reset: Resets the correction factor back to the original factory setting. Press to apply the factory setting.
Current Factor: Displays the factor currently in use.
Sets a calibration reminder based on a specified time frame or number of pipetting cycles. When the calibration reminder is displayed, press any key to confirm. However, the reminder will reappear every time the pipettor is turned on until you change the reminder time or use the reset option.
Timer: Press OK to turn the reminder timer On or Off.
Remind in: Use the Touch wheel to set a reminder interval for calibration (time in months or in thousands of cycles).
Reset: Resets the timer to the defined calibration inter­val. With this option highlighted, press OK and Save to enable.
Displays notes of any service that took place on the VIAFLO Electronic Pipettes listed newest entry first.
Starts calibration of tip spacing of VOYAGER pipettes. -
-
On/Off 1 -12 months or 10k - 240k cycles
-
www.integra-biosciences.com 17
VIAFLO Pipettes – Operating instructions V05 Installation
3.3.3 Communications
The VIAFLO Electronic Pipettes can be programmed from a PC via a Single Pipette Programming Stand (#4211) or via wireless Bluetooth connection.
VIALINK is a pipette management software for the PC and is used to manage VIAFLO Electronic Pipettes. It can be downloaded from the INTEGRA website in the product section and is free of charge for all VIAFLO Electronic Pipettes customers. A detailed description of the software, along with the operating instructions, can be found on the website as well.
Communi­cations
Serial Turn serial communication mode “On” to begin bi-
Bluetooth Each pipette needs its own Blue Tooth chip. Blue Tooth
Press OK to toggle between On and Off.
3.3.4 Owner Information
Owner Information
VIAFLO The first line is highlighted. Press to enter personal
In addition, information about your VIAFLO Electronic Pipettes, such as owner identification, serial number (SN), pipette size, indication of single or multi channel, and version number, are displayed automatically.
Description Range
On/Off directional communication. To exit the communications mode follow screen prompt or reset the VIAFLO pipette
). Serial communication further allows (in addition to
(13 the functions of Bluetooth):
• Firmware updates
• Bi-directional communication from PC to pipette.
On/Off communication has an open field connectivity of about 100 meters. It can be used for:
• Create and manage custom pipetting programs
• Image downloads
• Service history tracking
Description Range
-
information for your VIAFLO Electronic Pipettes.
•Use the Touch wheel to highlight a character and
press OK. You can press to Delete the last charac­ter entered. After entering the desired text, press to Save.
18 INTEGRA Biosciences AG
VIAFLO Pipettes – Operating instructions V05 Installation
3.3.5 Help Language
Help Language
Help Language
3.3.6 Write protect
Select this option to protect programs and menu options from inadvertent modification. The pipetting programs can still be used.
Write protect
Keep the password in a safe place.
Description Range
You can choose the language in which all help screens are displayed (English, French, German, Spanish, simplified Chinese, Japanese). Scroll to the desired language and press to Save.
Description Range
Select an option and press OK to switch protection On or Off:
Standard programs
Custom programs
Calibration
Too lbo x
VOYAGER
Password protect: Protect the access to the write pro-
tect menu by selecting “On”.
Edit password, if password protect is “On”. To enter a
password use the Touch wheel to highlight a charac- ter and press OK. Press to save the password. The password must be entered before you can access the write protect menu.
-
On/Off
3.3.7 Password key
This menu displays the encrypted password, if set. Should you lose your password, contact INTEGRA Biosciences to retrieve your password using these numbers.
www.integra-biosciences.com 19
VIAFLO Pipettes – Operating instructions V05 Operation

4 Operation

4.1 Turn on/off the device

Turn on: Press and release the Run key (7
AUTION
C
Remove hands from the Tri-lobe tip interface (12) at switch on and during homing. Also do not touch the touch wheel, because it is calibrated during the start up process.
The pipette flashes the startup screen and performs a full motor homing routine, ensuring the motor is in the run position. “Home” is the base point for the pipette. Homing is the process whereby the pipette motor moves the piston(s) to a sensor position. This position ensures that no liquid remains in the tips. For the VOYAGER, homing also includes the tip spacing motor. After homing, the tips move to the last position they occupied.
After homing the menu last used is displayed. Press the Back button (2 as necessary to return to the Main menu.
Turn off: To turn off the pipette, press and hold the Back button (2
OTE
N
The VIAFLO Electronic Pipettes will dim and tu rn off automatically after a preset duration of inactivity. This duration is 5 minutes, by default, and configurable with the Toolbox (see “3.3.1 Preferences” on page 16).
) to turn on the pipette.
) as many times
) for 3 seconds.

4.2 Attaching and removing GripTips

CAUTION
To ensure optimal performance of your VIAFLO Electronic Pipettes always use GripTips which have been designated for the use with the VIAFLO Electronic Pipettes, see “8.2
The unique tri-lobe design of the VIAFLO Electronic Pipettes reduces attachment and ejection forces, ensures a perfect fitting to prevent the tips from falling off and to provide a perfect sealing. On a multichannel pipette all tips sit on the same height.
Tri-lobes minimize contact surface and temporarily deforms the tip shaft
Shoulder provides a positive stop to prevent over tightening and all tips seal at the same height
O-ring provides forgiving and robust seal surface for the pipet tip
20 INTEGRA Biosciences AG
Consumables” on page 64.