Integra DOSE IT, 171000 Operating Instructions Manual

DOSE IT Operating instructions
171250_V08
Declaration of conformity
INTEGRA Biosciences AG –
7205 Zizers, Switzerland
declares on its own responsibility that the device
Zizers, November 4, 2016
Description Model
DOSE IT P910 171 000
EU Directives (DoW: Date of Withdrawal) Before DoW DoW After DoW
Low Voltage Equipment 2006/95/EC 20.04.2016 2014/35/EU
Electromagnetic Compatibility 2004/108/EC 20.04.2016 2014/30/EU
Restriction of Hazardous Substances 2011/65/EU
Waste Electrical and Electronic Equipment 2012/19/EU
EU Regulations
Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH)
1907/2006
Ecodesign - Power supplies 278/2009
Standards for EU
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
EN 61010-1: 2010
Electrical equipment for measurement, control and laboratory use ­EMC requirements.
EN 61326-1: 2013
Standards for Canada and USA
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
CAN/CSA-C22.2
No. 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
UL 61010-1
Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Part 15 of the
FCC Rules
Class A
Elmar Morscher CEO
Thomas Neher Quality Manager
DOSE IT – Operating instructions V08
www.integra-biosciences.com 3
Table of contents
Chapter 1 Introduction
1.1 Intended use ..................................................................................5
1.2 Symbols used.................................................................................5
1.3 Safety notes ...................................................................................6
Chapter 2 Description of the device
2.1 Scope of delivery............................................................................7
2.2 Overview of the DOSE IT...............................................................7
Chapter 3 Installation
3.1 Power supply connection ...............................................................9
3.2 Retort rod and filling arm (optional)................................................9
Chapter 4 Operation
4.1 Selection and loading of the tubing ..............................................10
4.2 Parameter settings.......................................................................13
4.3 Adjusting the dispensing program................................................15
4.4 Calibrating the pump ....................................................................18
4.5 Performing the dispensing process..............................................19
4.6 Working in Pump mode................................................................19
4.7 Using the double pumphead configuration (optional)...................20
4.8 Process documentation (optional)................................................21
4.9 Using the foot and benchtop switch .............................................22
4.10 Remote control by a PC .............................................................22
Chapter 5 Maintenance
5.1 Cleaning and servicing.................................................................23
5.2 Decontamination ..........................................................................23
5.3 Leak Test .....................................................................................23
5.4 Disposal .......................................................................................24
Chapter 6 Technical data
6.1 Environmental conditions.............................................................25
6.2 Specifications ...............................................................................25
6.3 Current consumption depending on input voltage........................25
6.4 Chemical compatibility .................................................................26
DOSE IT – Operating instructions V08
4 INTEGRA Biosciences AG
Chapter 7 Accessories and consumables
7.1 Consumables ...............................................................................27
7.2 Spare part ....................................................................................28
7.3 Accessories..................................................................................29
Imprint .................................................................................................30
DOSE IT – Operating instructions V08 Introduction
www.integra-biosciences.com 5
1 Introduction
1.1 Intended use
This is a general-purpose laboratory instrument. Any use of this instrument in a medical or IVD setting is the sole responsibility of the user.
The programmable peristaltic pump DOSE IT is designed for pumping and dispensing liquids in a volume range of 0.1 ml to 9999 ml using silicone tubings.
If the DOSE IT is used in a manner not specified by INTEGRA, the protection provided by the DOSE IT may be impaired.
1.2 Symbols used
This operating instruction manual makes specific reference to residual dangers using the symbols shown below.
1.2.1 Hazard warnings in these operating instructions
W
ARNING
This safety symbol advises of hazards that could result in injury or death. It also indicates hazards for machinery, materials, and the environment. It is essential that you follow the relevant precautions.
C
AUTION
This symbol cautions about potential material damage or data loss in a microprocessor controller. Follow the instructions.
N
OTE
This symbol identifies important notes regarding the correct operation of the device and labor-saving features.
1.2.2 Hazard warnings on the device
R
OTATING COG WHEELS
Risk to squash your fingers.
DOSE IT – Operating instructions V08 Introduction
6 INTEGRA Biosciences AG
1.3 Safety notes
The DOSE IT conforms to the state of technology and the recognized safety rules, and is operationally safe. Operate the DOSE IT when in perfect conditions only and observing this user manual.
Residual dangers can emanate from the device if it is used or operated incorrectly by untrained personnel. All persons entrusted with operation of the DOSE IT must have read and understood this user manual and, in particular, the safety notes, or must have been instructed by their superior such that they are able to operate the device without danger.
Please observe the danger references on the device.
Do not carry out any conversion work or alterations on the device.
Irrespective of the safety instructions set out here, applicable provisions and regulations must be observed in addition; for example, GLP, GMP, FDA, of the employer's liability insurance associations, of the health authorities and of the trading standards authorities.
Please visit our website www.integra-biosciences.com
on a regular basis for up to date
information regarding REACH classified chemicals contained in our products.
DOSE IT – Operating instructions V08 Description of the device
www.integra-biosciences.com 7
2 Description of the device
2.1 Scope of delivery
• DOSE IT P910 peristaltic pump
• Silicone tubing, autoclavable, 4 mm inner diameter (ID)
• 2x Aspiration/dispensing tubes 10 cm, stainless steel, 4 mm ID, one end dented
• Aspiration tube 35 cm, stainless steel, 4 mm ID, one end dented
• Tube collar
• AC adapter
• Operating Instructions
2.2 Overview of the DOSE IT
Front view
1Housing
2 Operating panel with display and keypad
3 Pumphead
Operating panel
4 Back-key
5 Function keys
6 Arrow keys
4
5
6
DOSE IT – Operating instructions V08 Description of the device
8 INTEGRA Biosciences AG
Rear view
7 Screw threads for fixation of the retort rod
8 Handhold
9 Connections and mains switch (see below)
78 9
Connections and mains switch
10 RS-232 interface for connection with serial printer
11 RS-232 interface for service connection
12 AUX connection for controlling a second DOSE IT
13 Mains switch
14 Connection for foot switch, benchtop switch,
MEDIAJET or external relay contact (potential free, normally open, 3.5 mm jack plug) for remote control
15 Mains connection socket
10
11
12
13
14
15
DOSE IT – Operating instructions V08 Installation
www.integra-biosciences.com 9
3 Installation
The DOSE IT is a ready-to-use peristaltic pump that requires only a minimal installation procedure. It shall be installed on a planar surface, in a dry and dust-free environment.
N
OTE
Before proceeding with the installation compare the contents of the package with the list “2.1
Scope of delivery” on page 7. Should something be missing or should you find a faulty component, please contact your local INTEGRA Biosciences representative.
3.1 Power supply connection
Connect the AC adapter cable to the DOSE IT mains connection socket and plug it to the electricity mains.
W
ARNING
The supply voltage must meet the requirements of the AC adapter:
100 - 240 VAC, 50 - 60 Hz.
3.2 Retort rod and filling arm (optional)
Fix the retort rod with the two screws on the rear side, using
a Phillips screwdriver (PH2), see "
Screw threads for fixation
of the retort rod" on page 8.
Mount the filling arm onto the rod and tighten it with the side
lever in order to prevent it from moving during dispensing.
DOSE IT – Operating instructions V08 Operation
10 INTEGRA Biosciences AG
4 Operation
The DOSE IT operates in different modes: DISPENSE mode for distribution of a defined volume into containers and PUMP mode for continuous pumping of liquids. In addition, customized applications can be defined (software version 2.0 and higher).
4.1 Selection and loading of the tubing
4.1.1 Tubing selection
The pumphead of the DOSE IT is compatible with differently sized tubing allowing the user to dispense a wide range of dose volumes. The main criteria for the choice of tubing are the speed and precision requirements of a specific application. The larger the inner diameter (ID) of the tubing, the faster but less precise is the dispensing.
As a reference for selecting the appropriate inner diameter of the tubing, typical volume and speed values for different tubing sizes are listed in Table 1.
N
OTE
The indicated volume values relate to the specified reproducibility and not to the actual limits of the system. The values have been determined experimentally and may vary slightly depending on the experimental set up.
Compatibility with the pumphead and specified reproducibility can be guaranteed for original INTEGRA Biosciences’ tubing only.
N
OTE
8 mm ID tubing is only recommended for the “PUMP” program mode (for accuracy reasons, drips between the aliquots and draw backs in the tubing may be a problem).
Table 1: Typical volume and speed values
Tubing inner diameter (ID) 1 mm 2 mm 3 mm 4 mm 6 mm
Volume (ml) at CV* < 1 % > 0.5 > 1 > 3 > 7 > 15
Speed range (ml / min) 0.6 – 52 2.1 – 203 4.8 – 475 8.4 – 837 16 – 1634
* Coefficient of variation.
DOSE IT – Operating instructions V08 Operation
www.integra-biosciences.com 11
4.1.2 Loading of the tubing in the pumphead
W
ARNING
Always switch off the DOSE IT when loading the tubing and when manipulating the open pumphead.
C
AUTION
Do not adjust the tubing clamps according to the scale on the pumphead sides because the tubing will be unnecessarily squeezed and might be damaged. Also the pumping precision will be diminished. Use instead the adjustments described in Table 2.
1 Open the pumphead
2 Adjust the tubing clamps
The jagged, V-shaped tubing clamps on both sides of the pumphead need to be adjusted and guarantee that the tubing does not slip when the rotor turns. The tubing clamps are set by adjusting the finger screws in such a way as to match the diameter of the tubing.
DOSE IT – Operating instructions V08 Operation
12 INTEGRA Biosciences AG
Table 2: Adjustments of the clamps for differently sized tubing
4.1.3 Connecting the dispensing and aspiration tubes
Clamp Tubing ID 1.6-marking of clamps
1 mm between 3.2 and 4.8 scale values
2 mm matches the 4.8 value on the scale
3 mm matches the 6.4 value on the scale
4 – 8 mm Completely open the clamps as shown beside.
Turn the finger screws right till the stop position.
3 Insert the tubing
Place the tubing over the rollers and center it so that it comes out of the pump head over the clamps on both sides. Close the pumphead and make sure that the tub­ing exits the pumphead unhindered and bending down­wards. The tubing must not be bent upwards or towards the front or back.
The aspiration/dispensing tubes delivered with the DOSE IT facilitate the dosing process and are fitted on the end of the tubing. The aspiration tube is dented on one end in order to guarantee a free flowing of the liquid entering the tubing. Make sure not to insert the dented end of the aspiration tube into the silicone tubing.
The tube collar can either be screwed on the aspiration tube to weigh it down or to the dispensing tube as a spacer for use with a clamp.
Dented end
Tube collar
Aspiration/ dispensing
tube
Silicone tubing
DOSE IT – Operating instructions V08 Operation
www.integra-biosciences.com 13
4.2 Parameter settings
4.2.1 System parameters
4.2.2 Program parameters
Some parameters described are not displayed in every program mode.
System-Parameter Description
Language Set the desired language. Time/Date Set time and date. Display contrast Change the contrast of the display. Access code Provides user access restrictions to parameters. Info General information on the unit:
• Version of the software
• Version of the electronics
• Serial number.
Parameter Description Range Default
Name Allows to name the program with up to 16
alphanumeric symbols. Press the arrow keys simultaneously to delete the program name.
--
Type Defines the program mode: “P
UMP” for continu-
ous pumping of a liquid, “D
ISPENSE” for the
dosing of liquids, or “C
USTOMIZE” to create a
customized application.
Dispense Pump Customize
Dispense
Tubing-ID (mm)
Defines the Inner Diameter (ID) of the tubing. 1.0 – 8.0 mm 4.0 m m
Flow rate (ml/min)
Allows to set the pump speed. The flow rate depends on the tubing inner diameter.
Time (h min s)
Defines the duration of the pumping. Entering 0 h 0 min 0 s puts the unit in continuous opera­tion, which is represented by the infinite symbol (.
1'' – 9 h 59' 59''1h
Volume Defines the volume of a dose in the dispensing
mode.
0.01 – 9999 ml 10 ml
Repetitions Defines the number of doses in the dispensing
mode. Entering 0 results in an unlimited num­ber of doses, which is represented by the infinite symbol (.
1 – 999
20
DOSE IT – Operating instructions V08 Operation
14 INTEGRA Biosciences AG
Customize mode
The ‘Customize’ mode allows to program a customized application consisting of up to 20 single steps:
Pause Defines the duration of the interval between
doses in the dispensing mode. If switching to manual, the dispensing of a dose has to be activated manually.
0.1– 99.9 s, manual
1.0 s
Direction Defines the rotational direction of the rotor and
hence the flow direction. / C W m e a n s c l o c k ­wise, / CCW means counter-clockwise.
/ C W / C C W
/ C W
Dispense direction
Defines the rotational direction for the step type “D
ISPENSE”. For the step type “ASPIRATION” the
rotor rotates oppositely.
/ C W / C C W
/ C W
No. of heads
Defines the number of pump heads. 1, 2 1
Step 1–20 Defines single steps of a customized applica-
tion.
1–20 -
Step Description Parameter
Dispense Defines the dosing of liquids analog to the program mode
“Dispense”. The rotational direction is defined by the parameter “Dispense direction’.
Volume Repetitions Flow rate Pause
Dispense Ramp
Defines dispensing with accelerating or decelerating speed, needed e.g. for density gradients. Just define the start and end flow to ramp the dispensing speed up or down. If the start flow is identical to the speed of the previ­ous step, and the pause of the previous step is set to man­ual, the liquid can be dispensed without pause.
Volume Repetitions Pause Start flow End flow
Aspirate Defines the aspiration of liquids analog to the program
mode “Dispense” with opposite rotational direction.
Volume Repetitions Flow rate Pause
Aspirate Ramp
Defines the aspiration of liquids with accelerating or decel­erating speed. Just define the start and end flow to ramp the aspirating speed up or down. If the speed of the follow­ing step is identical to the end flow, and the pause is set to manual, the DOSE IT continues without pause.
Volume Repetitions Pause Start flow End flow
Parameter Description Range Default
DOSE IT – Operating instructions V08 Operation
www.integra-biosciences.com 15
4.3 Adjusting the dispensing program
4.3.1 Choosing the language
N
OTE
Because it is easiest to operate the DOSE IT by following the instructions on the screen, please select the language you are most familiar with.
Pump Defines continuous pumping of a liquid. Direction
Flow rate Time
Wait Defines the duration of the interval between two steps.
Manual mode means that the next step has to be activated manually.
1'' – 9 h 59' 59'' manual
Cycles Defines the number of cycles of a sequence. A cycle
always initiates a return back to a defined step, e.g. step 1. You can define several cycles within your customized sequence.
Number of cycles 1 –999, Goto step ¥
- No operation, used as default setup of step. -
Step Description Parameter
1 Switch on the DOSE IT by pressing the mains switch at
the back. The A
PPLICATIONS window appears. The first four
programs in the application list are already pre-defined with default values.
2 Press SYSTEM to access the SYSTEM window and set the
language of your choice by accessing the LANGUAGE menu. In the S
YSTEM window you can also set the display
contrast, the date and the time, and an access code, if required.
DOSE IT – Operating instructions V08 Operation
16 INTEGRA Biosciences AG
4.3.2 Access code
The DOSE IT system and parameter settings can be protected by a code, if activated (Access code required ON). Before any parameter changes can be performed, an access code must be entered. A standard user can only run defined programs.
Select the A
CCESS CODE window and enter the default access code “3473” by typing the
appropriate key until the correct number appears on the screen. Press the back-key.
4.3.3 Setting the dispensing parameters
N
OTE
The DOSE IT has a memory capacity for storing 20 programs, all of which can be configured according to your specific applications.
1 Select the line “Access code required” and press
C
HANGE. Use the arrow keys to select “Access code
required: ON” and press S
AVE.
2 For changing the Access code select this line and press
S
AVE.
3 Change the default access code to your personal code,
if required. Enter the numbers by the arrow keys and select the digit with the function keys (, ). Keep this code in a save place. Press S
AVE.
1 Press the back-key to exit the SYSTEM window and to
return to the A
PPLICATION window. Press PARAMETER to
configure the program that is backlit.
2 Use the arrow keys to select the listed parameters and
press C
HANGE to set different values.
All parameters and their values are described in “4.2.2
Program parameters” on page 13.
DOSE IT – Operating instructions V08 Operation
www.integra-biosciences.com 17
4.3.4 Defining a customized program
The following example of a customized program can be used to fill 10 density gradient tubes. The careful and continuous dispensing without pauses prevents stirring up of layers.
Parameter Value Description of settings
Name Density
gradient
Allows to name your customized program.
Tubing-ID 3.0 mm Define the inner diameter of the tubing inserted. Dispense
Direction
/ C W Define the rotational direction of the rotor, e. g. clock-
wise. No. of heads 1 Define the number of pump heads installed. Step 1: Dispense Volume: 2.5 ml
Repetitions: 1
Pause: manual (here without pause)
Flow rate: 3.8 ml/min (i.e. speed: 4.0 rpm). Step 2: Dispense
Ramp
Volume: 12.5 ml
Repetitions: 1
Pause: manual (here without pause)
Start flow: 3.8 ml/min (i.e. speed: 4.0 rpm)
End flow: 22.8 ml/min (i.e. speed: 24.0 rpm). Step 3: Wait Manual, i.e. the next cycle starting with step 1 has to be
activated manually by pressing N
EXT STEP.
Step 4: Cycles Number of cycles: 10
Goto step: 1.
The figure depicts the customized program described above.
DOSE IT – Operating instructions V08 Operation
18 INTEGRA Biosciences AG
4.4 Calibrating the pump
NOTE
The calibration procedure tunes the pump to deliver a precise output. The output of the pump is determined by the tubing's inner diameter; therefore a calibration is necessary for tubing with different inner diameter. Also calibrate the pump after changing the fluid to be pumped. It is recommended to calibrate the pump at regular intervals to maintain its precision.
I
MPORTANT NOTE
A calibration value is stored for each of the 20 programs that can be saved. Therefore a calibration for each of the 20 programs is necessary, even when the tubing inner diameter remains the same. Within each program, only the last calibration value is saved, hence calibrate whenever there is a change in the viscosity of the pumped liquid, of the flow rate or of the tubing inner diameter.
Each step within a customized program has to be calibrated separately.
1 Select a program and press START to access the
program window. Put the tubing end with the aspiration tube in the liquid that has to be dispensed and press P
RIME to fill the tubing with liquid without bubbles.
2 Press the CALIBRATE button to enter the calibration
dialogue window and DOSING to obtain a calibration sample. To reduce the measurement error it is possible to obtain up to ten such dosing samples (by repeatedly pressing the D
OSING button).
3 After one or more calibration samples have been
dispensed and measured, press the C
ONTINUE button
and enter the measured volume (CHANGE button). To end the calibration procedure press CONTINUE.
DOSE IT – Operating instructions V08 Operation
www.integra-biosciences.com 19
4.5 Performing the dispensing process
4.6 Working in P
UMP mode
1 In the APPLICATION window select a program and press
S
TART to access the program window. Press PRIME to fill
the tubing with liquid until no more bubbles are in the tube and calibrate if necessary.
To begin dispensing, press S
TART.
2 While dispensing, there are two possibilities to interrupt
the process:
• P
AUSE – the process is interrupted after completing
the dispensing of the current dose.
• S
TOP – the process is interrupted immediately.
After interruption, the process can either be continued, aborted or parameters can be changed (the latter is not possible in the customize mode). If parameters have been changed, the user is asked to either save or discard the changes before exiting the program.
The completion of the program is announced by a beep and a message appears on the screen.
To continuously pump a liquid instead of dispensing it, change the T
YPE-parameter from DISPENSE to PUMP. To
set the other parameters follow the instructions de­scribed above and in “4.2.2 Program parameters” on page 13.
The calibration procedure in P
UMP mode works as described above. Note that the
calibration volume is indicated in the calibration window and that it varies depending on the tubing size and on the selected flow rate.
To use a pumping program, follow the instructions on the screen. Note that you can change the flow rate without interrupting the pumping by using the arrow keys (not possible within customized mode).
DOSE IT – Operating instructions V08 Operation
20 INTEGRA Biosciences AG
4.7 Using the double pumphead configuration (optional)
The DOSE IT can be used with a double pumphead and the corresponding 8 mm ID tubing set (171 088) when it is necessary to quickly dispense large volumes. The double pumphead configuration also reduces flow pulsation.
W
ARNING
Always switch off the DOSE IT when assembling the double pumphead configuration.
Be aware of the sharp edges of the pumpheads when assembling them and when opening and closing the first pumphead for tubing insertion.
When placing the tubing set (171 088) into the double pumphead assembly, gently stretch the tubing so that inside the pumphead the tubing is taut (avoid loading the tubing in a loose, slackened way).
W
ARNING
Visually inspect the tubing set Y-pieces for the absence of cracks in the glass. Cracked Y-pieces may burst during operation and shall not be used.
In the affected programs, change the parameter “No. of heads” before using the double pumphead configuration.
1 Open the pumphead to gain access to the two fixation
screws and unscrew the pumphead from the housing using an Allen key # 3.
2 Fix the extension pumphead (171 090) to the housing
using the screws obtained after removing the original pumphead.
Fixation screws
3 Assemble the two pumpheads by locking the original
pumphead onto the extension pumphead. When turning the original pumphead in place, ensure that the lock mechanism engages with an audible click.
Important: The rollers of the two pumpheads must be out of phase in order to avoid excessive strain on the motor (torque limits) and to obtain a reduction in flow pulsation.
DOSE IT – Operating instructions V08 Operation
www.integra-biosciences.com 21
4.8 Process documentation (optional)
To document the dispensing and pumping performed with a DOSE IT, relevant data can be recorded either on paper or in an electronic text file. The following data – exemplified by a dispensing program – are recorded:
• At the start of the process
======================================== DOSE-IT (SN:00000000)
***Start dispense Date/time: 02.11.2006/18:01
***Program settings Name: MY PRG1 Tubing-ID: 4.0mm Volume: 10.0ml Repetitions: 20 Pause: 1.0s Flow rate: 500.0ml/min Direction: CW No. of heads: 1 ========================================
• After calibration
======================================== DOSE-IT (SN:00000000) ***Calibration done Date/time: 02.11.2006/18:05
Tubing-ID: 4.0mm ========================================
• At the end of the process
======================================== DOSE-IT (SN:00000000)
***End dispense Date/time: 02.11.2006/18:02
Repetitions: 20
***Program settings Name: MY PRG1 Tubing-ID: 4.0mm Volume: 12.0ml Pause: 1.0s Flow rate: 500.0ml/min Direction: CW No. of heads: 1 ========================================
Note: This parameter
has been changed
during the process.
DOSE IT – Operating instructions V08 Operation
22 INTEGRA Biosciences AG
Printing on paper occurs automatically when a label printer is connected to the RS - 232 (2) port of the DOSE IT. INTEGRA Biosciences recommends the EPSON TM-U220 matrix printer and your local INTEGRA dealer can assist you to connect this printer.
To generate an electronic text file, connect a PC to the RS - 232 (2) port of the DOSE IT via a crossed female-female RS - 232 cable and record the process data using the HyperTerminal software installed on your PC.
RS - 232 (2) interface settings:
Transmission speed:......... 9600 bps
Databits:............................ 8 bits
Parity:................................ No
Stopbits: ............................ 1
Handshake:....................... Xon / Xoff
4.9 Using the foot and benchtop switch
The optional foot switch or benchtop switch can be used for hands-free operation. There are no specific setting needed, just connect one of the switches.
Depending on program mode and step, the switch can be used instead of the S
TART,
PAUSE or STOP key. However, the keys will still work, even if a switch is connected.
In D
ISPENSE mode, set the PAUSE parameter to manual. The first pedal/button pulse starts
the set number of doses, the next pulse will pause the process, and so on until the process is completed.
In P
UMP mode, the pump starts with the first pedal/button pulse (=START). With the next
pulse it stops (=STOP) and so on.
4.10 Remote control by a PC
Remote operation can be performed by simulating the foot switch, e. g. with a PC and an input/output interface box as described above for the foot switch.
DOSE IT – Operating instructions V08 Maintenance
www.integra-biosciences.com 23
5 Maintenance
5.1 Cleaning and servicing
WARNING
Always switch off the DOSE IT and disconnect it from the mains when carrying out maintenance work.
If the DOSE IT gets soiled, it can be cleaned with a cloth moistened with soapy water or with a 70 % solution of Ethanol.
Regular servicing of the DOSE IT by a qualified partner of INTEGRA Biosciences is recommended.
5.2 Decontamination
The aspiration/dispensing tubes, tube collars and silicone tubing contacting the liquid can be autoclaved at 121°C, 1 bar overpressure for at least 20 minutes. Silicone may become brittle after extensive autoclaving. Replace the tubings if they are damaged.
If the surface of the DOSE IT has been in contact with biohazardous material, it must be decontaminated in accordance to good laboratory practice. Do not spray directly on the instrument but use a lint-free cloth, lightly soaked with a disinfectant and wipe dry directly after decontamination. Never use acetone or other solvents! Follow the instructions provided by the disinfectant manufacturer.
5.3 Leak Test
It is recommended to perform a leak test about every 3 months or when dosing errors occur. A leak test will reveal worn or defect tubing or pumphead.
Procedure:
1) Load a silicone tubing in the pumphead.
2) Position the dispensing end of the tubing below the pumphead height into a container.
3) Put the aspiration tube in liquid and press P
RIME to fill the tubing completely without
any bubbles.
4) Lift the aspirating end of the tubing as high as possible above the pumphead. The
system is tight, if the liquid level do not decrease and no liquid flow out of the lower dispensing end.
The moving parts of the pumphead should be checked from time to time for freedom of movement.
Occasionally lightly grease the lever (1) and the rollers (see image) with Teflon lubrication oil.
The rotor shaft runs on sealed bearings which do not require lubrication.
1
DOSE IT – Operating instructions V08 Maintenance
24 INTEGRA Biosciences AG
If the system is leaking, you have to check whether the tubing or the pumphead is worn:
Repeat the leak test with a new tubing (only use original INTEGRA tubings, see “7.1 Consumables” on page 27).
• If system is tight, the tubing was worn and the problem is solved.
• If the system is still leaking, the pumphead is worn (e.g. lever loose, mechanical play)
and need to be replaced, see “7.2 Spare part” on page 28
.
N
OTE
Work around until a new pumphead is available: Put a rubber band (5 to 8 mm wide) around the pumphead to help holding it fully closed during oper­ation. Repeat the leak test to confirm that the work around works.
5.4 Disposal
The DOSE IT is labeled with the “crossed-out bin” symbol to indicate that this equipment must not be disposed of with unsorted municipal waste. Instead, it is your responsibility to correctly dispose of your waste equipment by hand­ing it over to an authorized facility for separate collection and recycling. It is also your responsibility to decontaminate the equipment in case of biological, chemical, and/or radiological contamination so as to protect from
health hazards the persons involved in the disposal and recycling of equipment.
For more information about where you can drop off your waste equipment for recycling, please contact your local dealer from whom you originally purchased the product or your local council.
By doing so, you will help conserve natural resources and you will ensure that your waste equipment is recycled in a manner that protects human health and the environment. Thank you!
DOSE IT – Operating instructions V08 Technical data
www.integra-biosciences.com 25
6 Technical data
6.1 Environmental conditions
6.2 Specifications
6.3 Current consumption depending on input voltage
Operation
Temperature range 5–35 °C Humidity range Max. rel. humidity 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % rel. humidity at 40 °C.
Altitude range < 2000 m
Dose volume 0.01 ml – 9999 ml
Flow rate 0.6 ml / min – 5 l / min
Tubing inner diameter 1 – 8 mm
Tubing wall thickness 1.5 mm
Materials Housing: Polyurethane
Base plate, rear panel: stainless steel Membrane keypad: Polyester Tubing: Silicone Pump head body rear: Polypropylene Pump head body front: IXEF Pump head tube clamps: Nylon
Pump head roller: Nylatron
Dimensions (H x W x D) 203 x 210 x 191 mm
Weight 3.5 kg
Interfaces 2 x RS-232
Electricity supply Input: 100 – 240 VAC, 50 / 60 Hz
Output: 46.6 – 49.4 VDC, 70 W
Current consumption
Input voltage Standby
Operation with 1 or 2 pump heads
Pause mode
100 VAC 45 mA 260 mA 280 mA 230 VAC 15 mA 120 mA 130 mA
DOSE IT – Operating instructions V08 Technical data
26 INTEGRA Biosciences AG
6.4 Chemical compatibility
A particular advantage of peristaltic pumps is that the fluid remains enclosed inside the tubing. Even chemical aggressive liquids can be transferred without the risk of damaging parts of the pump as long they are compatible with the silicone tubing. The table below rates the compatibility of silicone to a few of the chemicals commonly used in laboratories. INTEGRA Biosciences AG assumes no liability for the information contained in the table.
Compatibility ratings for silicone tubing: A = good, little to minor effects (0-5 % volume swell). B = fair, moderate effects, not recommended for continuous use (5-10 % volume swell). C = critical, not recommended (10 % or greater volume swell).
To determine the compatibility silicone to a chemical not listed in the table, please consult one of the several tables available on the internet, e. g. Watson Marlow with material Pumpsil (silicone).
It is the responsibility of the users to ensure the chemical compatibility. Before using a critical chemical, immerse a short piece of the tubing in a closed container of the chemical for 48 hours. Check the tubing for sings of swelling, softening, discoloration, embrittlement or any other attack.
Chemical Rating
Acetic acid CH3COOH B Acetone C3H6O C Acetonitrile C2H3N C Dimethyl sulfoxide DMSO A Ethanol C2H5OH B Hydrochloric acid 33 % HCl C JAVEL NaClO A Phenol C6H5OH C Sodium carbonate Na2CO3 A Sodium hydroxide 50 % NaOH B
DOSE IT – Operating instructions V08 Accessories and consumables
www.integra-biosciences.com 27
7 Accessories and consumables
There are a number of accessories and options that adapt the DOSE IT to particular application requirements and working environments.
7.1 Consumables
7.1.1 Silicone tubing
ID
1
(mm)
1. ID = inner diameter
Part No.
length 2.5 m, autoclavable, wall thickness 1.5 mm.
1 171 021
2 171 022
3 171 023
4 171 024
6 171 026
8 171 028
2
2. only for pumping applications
length 25 m, (bulk roll), autoclavable, wall thickness 1.5 mm.
1 171 031
2 171 032
3 171 033
4 171 034
6 171 036
8 171 038
2
DOSE IT – Operating instructions V08 Accessories and consumables
28 INTEGRA Biosciences AG
7.1.2 Aspiration/dispensing tubes
7.1.3 Tube collars
7.2 Spare part
ID (mm) Part No.
Aspiration/dispensing tube, length 10 cm, stainless steel, one end dented.
1 171 051
2 171 052
3 171 053
4 171 054
6 171 056
8 171 058
Aspiration/dispensing tube, length 35 cm, stainless steel, one end dented.
4 171 064
6 171 066
8 171 068
ID (mm) Part No.
Used as weight for the aspiration tube or as clamp spacer for the dispensing tube. See also “4.1.3
Connecting the dispensing
and aspiration tubes” on page 12.
1 – 3 171 071
4 – 6 171 074
Part No.
Pumphead 313D 103 520
DOSE IT – Operating instructions V08 Accessories and consumables
www.integra-biosciences.com 29
7.3 Accessories
Part No.
Foot switch. 143 200
Benchtop switch. 171 081
Retort rod with filling arm. 171 091
Extension pumphead for double pumphead assembly. See also “4.7 Using the double pumphead configuration (optional)” on page 20.
171 090
Silicone tubing set for double pumphead (ID 8 mm), total length 3 m, autoclavable, incl.
• 2 short silicone tubings (ID 8 mm)
• 2 Y-pieces made of glass
• 2 long silicone tubings (ID 10 mm)
171 088
Pipette-tubing connector, for connecting pipettes to 2-6 mm inner diameter silicone tubing, autoclavable, 5-pack
171 077
DOSE IT – Operating instructions V08
30 INTEGRA Biosciences AG
Imprint
© 2018 INTEGRA Biosciences AG
All rights to this documentation are reserved. In particular the rights of reproduction, processing, translation and the form of presentation lie with INTEGRA Biosciences AG. Neither the complete documentation nor parts thereof may be reproduced in any way, or stored and processed using electronic media or distributed in any other way without the written consent of INTEGRA Biosciences AG.
This operating instruction manual has the part number 171 250, version number V08. It applies to firmware version 2.02 or higher until a newer revision is released.
Manufacturer
Customer service
Please contact your local INTEGRA Biosciences representative. To find out name and address of your local representative go to www.integra-biosciences.com
.
Further information and operating instructions in other languages are available on www.integra-biosciences.com
or on request (info@integra-biosciences.com).
INTEGRA Biosciences AG
CH-7205 Zizers, Switzerland T +41 81 286 95 30 F +41 81 286 95 33
INTEGRA Biosciences Corp.
Hudson, NH 03051, USA T +1 603 578 5800 F +1 603 577 5529
info@integra-biosciences.com www.integra-biosciences.com
Loading...