Integra PIPETBOY acu 2, 155015, PIPETGIRL, 155016, 155018 Operating Instructions Manual

...
PIPETBOY acu 2 Operating instructions
466079_V13
Declaration of conformity
INTEGRA Biosciences AG –
7205 Zizers, Switzerland
declares on its own responsibility that the devices
EU Regulations
Standards for EU
Standards for Canada and USA
Zizers, November 4, 2016
Description Models
PIPETBOY acu 2 155000, 155015, 155016, 155017, 155 018, 155 019
PIPETGIRL 155021
EU Directives (DoW: Date of Withdrawal) Before DoW DoW After DoW
Low Voltage Equipment 2006/95/EC 20.04.2016 2014/35/EU
Electromagnetic Compatibility 2004/108/EC 20.04.2016 2014/30/EU
Restriction of Hazardous Substances 2011/65/EU
Waste Electrical and Electronic Equipment 2012/19/EU
Battery Directive 2006/66/EC
Registration, Evaluation, Authorisation and Restriction of Chemicals (REACH)
1907/2006
Capacity Labelling of Portable Secondary Batteries 1103/2010
Ecodesign - Power supplies 278/2009
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
EN 61010-1: 2010
Electrical equipment for measurement, control and laboratory use ­EMC requirements.
EN 61326-1: 2013
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
CAN/CSA-C22.2
No. 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use - General requirements.
UL 61010-1
Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interfer­ence received, including interference that may cause undesired operation.
Part 15 of the
FCC Rules
Class A
Elmar Morscher CEO
Thomas Neher Quality Manager
PIPETBOY acu 2 - Operating instructions V13
www.integra-biosciences.com 3
Table of Contents
1 Introduction ................................................................................................................ 4
2 Description of the device............................................................................................ 5
3 Installation .................................................................................................................. 6
4 Operation ................................................................................................................... 7
5 Maintenance............................................................................................................. 10
6 Technical Data ......................................................................................................... 11
7 Accessories.............................................................................................................. 12
Imprint
© 2018 INTEGRA Biosciences AG
This operating instruction manual applies to serial number 1 350 000 or higher of PIPETBOY acu 2 / PIPETGIRL.
Manufacturer
Customer service
Please contact your local INTEGRA Biosciences representative, see www.integra-biosciences.com
or contact info@integra-biosciences.com.
INTEGRA Biosciences AG
CH-7205 Zizers, Switzerland T +41 81 286 95 30 F +41 81 286 95 33
INTEGRA Biosciences Corp.
Hudson, NH 03051, USA T +1 603 578 5800
F +1 603 577 5529 info@integra-biosciences.com www.integra-biosciences.com
PIPETBOY acu 2 - Operating instructions V13 Introduction
4 INTEGRA Biosciences AG
1 Introduction
1.1 Intended use
PIPETBOY acu 2 is a pipette controller designed for aspirating and dispensing aqueous solutions with plastic or glass pipettes of 1 to 100 ml volumes. It is intended for measurement, control and laboratory use. Any use of this instrument in a medical or IVD setting is under the sole responsibility of the user.
PIPETGIRL is a special model of PIPETBOY acu 2 and provides the same functions as PIPETBOY acu 2.
1.2 Safety notes
1) Do not use or charge PIPETBOY acu 2 in an atmosphere with danger of explosion.
Also, do not pipette highly flammable liquids such as acetone or ether.
2) When handling dangerous substances, comply with the material safety data sheet
(MSDS) and with all safety guidelines such as the use of protective clothing and safety goggles. Never point a pipette in anyone’s direction.
3) Avoid pipetting of liquids whose vapours could attack the materials PA (polyamide),
POM (polyoxymethylene), FPM (fluor-rubber), NBR (nitrile-rubber), CR (chloroprene), silicone. Corrosive vapours could also damage metallic parts inside the device.
4) Use an original INTEGRA Biosciences mains adapter only and protect it from moisture,
otherwise PIPETBOY acu 2 might be damaged.
5) Prolonged exposure of PIPETBOY acu 2 to UV-light can cause discolouration and/or
yellowing of the plastic housing. However, this will not affect the performance of the device in any way.
6) Old Li-ion batteries may cause a safety risk. We recommend to replace the battery after
3 years of use. Also replace the battery if the charging intervals are unusually short or if the charging takes much longer than usual (4 hours or more). These are indicators that the battery has reached the end of its life-cycle.
7) Li-ion technology bears the risk of thermal runaway and cell rupture if the battery was
damaged. Do not expose the battery to heat (> 60°C) and avoid mechanical stress. Batteries which were subject to deep discharges may develop internal short circuits, leading to an increased self-discharge rate and heating during battery charging. This may also result in thermal runaway and cell rupture.
8) To extend the battery life-cycle, it is recommended to charge the battery every 2
months if the pipette controller is not used regularly. If the pipette controller is not used for more than 6 months, remove the battery from the instrument.
Regardless of the listed safety notes, additionally applicable regulations and guidelines of trade associations, health authorities, trade supervisory offices, etc. must be observed.
Please visit our website www.integra-biosciences.com
on a regular basis for up to date
information regarding REACH classified chemicals contained in our products.