♦ Complete, miniaturized patch clamp amplifier system
with small digital headstages using revolutionary
CLAMP microchips from Intan Technologies
♦ Voltage clamp operation with low current noise floor:
< 3 pA
♦ Current clamp operation with low voltage noise floor:
< 20 µV
♦ Fast transient capacitance compensation: 0-20 pF
range; bridge balance in current clamp mode
♦ Automated measurement of pipette / seal resistance
and cell membrane parameters
♦ Clamp operation controlled by software and/or external
analog voltage command signals
♦ Small headstage size: 6.1 cm × 2.8 cm × 2.0 cm (not
including pipette holder)
♦ Standard threaded Teflon pipette electrode connector
♦ Standard dovetail for mounting to micromanipulator
♦ All-digital interface cable is immune to noise pickup
♦ Standard USB interface to host computer
♦ Free, open-source, multi-platform GUI software
www.intantech.com ● info@intantech.com1
over 5 kHz bandwidth with highest sensitivity
rms
over 10 kHz bandwidth
rms
Description
The Intan Technologies CLAMP patch clamp amplifier
system allows users to perform signal channel or multichannel patch clamp electrophysiology or electrochemistry
experiments using the revolutionary new Intan CLAMP
voltage/current clamp chips.
The Intan CLAMP system incorporates all analog circuitry
and many digital control blocks in close proximity to the
electrode, resulting in a miniaturized “headstage” that is
actually a complete patch clamp amplifier. Thanks to the
integration of nearly all patch clamp circuit elements onto a
single silicon chip, the Intan-powered patch clamp amplifier
is smaller than traditional analog headstages that perform
only a small fraction of the total amplification task.
The Intan digital CLAMP headstages are interfaced to the
Intan CLAMP Controller which coordinates the operation of
the voltage/current clamps as well as sampling auxiliary
digital and analog signals in synchrony with the measured
electrophysiological signals. The interface cables are purely
digital with independently isolated grounds, making them
immune to noise pickup or signal degradation.
The Intan CLAMP Controller connects to a host PC via a
standard USB bus, and operates using free, open-source
software.
Intan CLAMP System
System Architecture and Theory of Operation
Traditional Analog Patch Clamp Amplifier
The diagram above shows the major functional components of a traditional patch clamp amplifier system. A traditional analog
headstage module contains sensitive analog electronics that act as a pre-amplifier with voltage clamp and current clamp capability.
In voltage clamp mode, the headstage measures the electrode current and converts this to a proportional small voltage. In current
clamp mode, the headstage buffers the electrode voltage but provides no additional amplification. Voltage and current clamp
control is provided by analog voltages that convey the desired clamping levels.
The headstage is connected to a remote computer-controlled amplifier via a shielded interface cable that carries small analog
voltages (typically in the millivolt range) and is thus susceptible to noise pickup. The amplifier module / digitizer consists of one or
more rack-mounted boxes that contain additional analog amplifiers, analog-to-digital converters (ADCs), digital-to-analog
converters (DACs), and a digital controller that coordinates control of these devices. The amplifier module and digitizer are
interfaced to a host PC that sequences particular experiments and records the measured data.
Intan Digital Patch Clamp Amplifier
The CLAMP chips from Intan Technologies combine all analog circuitry and many digital control blocks on a single chip, permitting
the construction of Intan-powered digital headstages (see diagram above). A small circuit board containing an Intan CLAMP
chip and a small number of support components form a complete patch clamp amplifier with a purely digital interface. The Intan digital “headstage” is actually a complete, miniaturized patch clamp amplifier. Thanks to the integration of nearly all patch
clamp circuit elements onto a single silicon chip (see “actual size” inset above), the Intan-powered patch clamp amplifier is smaller
than many traditional analog headstages that perform only a small fraction of the total amplification task.
The digital interface cable uses a standard serial communication protocol (Serial Peripheral Interface, or SPI) and is no longer
susceptible to noise pickup. Sensitive analog signals are digitized at the source, not several meters away. The purely digital
interface makes it easy to electrically isolate each headstage and eliminate ground loops. The need for a large computer-controlled
amplifier module is completely eliminated, replaced by an Intan CLAMP controller that can easily control multiple digital
headstages simultaneously.
To preserve legacy “external command” and “signal monitor” functions associated with traditional patch clamp systems, the Intan
CLAMP controller incorporates ADCs and DACs to allow for real-time control of voltage and current clamp via analog voltages (in
addition to software clamp control), and real-time monitoring of measured and/or clamp voltages/currents via analog signals.
www.intantech.com ● info@intantech.com2
Intan CLAMP System
The voltage clamp operates over a wide range with steps as
The current clamp produces currents up to ±127 nA and
steps as small as 5 pA. Intracellular voltages may be
. The plot below shows the intracellular
Intan CLAMP Headstages
An Intan CLAMP headstage is shown above with a pipette holder and glass pipette attached (not included). Intan headstages
mate with thread-mounted pipette holders that are compatible with Axon amplifiers. Pipette holders sold by Molecular Devices
(the 1-HL-U), Warner Instruments (the Q series), and other companies may be used with Intan headstages. Glass pipettes must
be custom made with a pipette puller shortly before recording.
Each headstage has standard 0.7” (17.8 mm) wide dovetail connectors on the side and the bottom. These mate with
micromanipulators from companies such as Sutter Instruments to permit precision positioning and movement of the headstage.
The rear side of the headstage has a receptacle for a 1-mm diameter grounding pin. This should be used to tie the headstage to
a grounding point near the experiment. Intan headstages are electrically isolated from the controller and from earth ground; this
can prevent noise pickup caused by unintentional ground loops. If a Faraday cage is used, it should be connected to the black
I/O GND binding post on the rear panel of the controller. Some experimentation may be needed to find the optimal grounding
scheme for a particular patch clamp rig.
The blue interface cable delivers ±3.3V power to the headstage and provides a digital serial interface between the headstage and
the controller. These SPI (serial peripheral interface) cables use low-voltage differential signaling (LVDS) with twisted-pair wires
for noise-free operation. Each Intan CLAMP headstage comes with a 6-foot (1.8-meter) interface cable. Additional SPI interface
cables may be purchased from Intan and daisy-chained to span longer distances.
Each headstage contains a small circuit board with an Intan CLAMP1 chip that performs all voltage and current clamping and
voltage and current sensing (see previous page). The detailed operation of this microchip is described in the Intan CLAMP Voltage/Current Amplifier Chip Datasheet, available from the Downloads page of the Intan website.
The figures below show actual data obtained using an Intan CLAMP system.
small as 2.5 mV. A low noise floor (below 3 pA
sensitive range) allows tiny currents to be resolved. The plot
below shows a cell membrane test:
in the most
rms
observed across a wide ±300 mV range with a noise floor
below 20 µV
potential of a neuron during current injection:
rms
www.intantech.com ● info@intantech.com3
Intan CLAMP System
Controller Hardware
Front Panel
The front panel of the Intan CLAMP Controller provides connection ports for Intan CLAMP headstages as well as auxiliary digital
and analog inputs. From left to right:
•Intan CLAMP headstage ports: These ports, labeled A and B, provide connection points for up to two headstages via
12-wire digital SPI (serial peripheral interface) cables. Each headstage port is electrically isolated from the controller
and from earth ground. Indicator lights provide information on the status of each port: green and yellow LEDs show that
proper voltage supplies are being provided for each headstage. Red LEDs are activated when the software recognizes
a headstage plugged into a port.
•Digital inputs: Two BNC sockets are provided for recording digital signals in synchrony with the headstage signals.
The digital inputs accept TTL-level signals. Any voltage between 0V and +0.8V is read as a digital “low”. Any voltage
between +2.0V and +5.5V is read as a digital “high”. Voltages delivered to these sockets should not exceed the range
of 0V to +5.5V. These signals may be used to record discrete events associated with an experiment.
•Analog inputs: Two BNC sockets are provided for recording general-purpose analog signals, or for controlling the
headstages using externally provided analog waveforms. Signals are digitized with 16-bit ADCs over a range of -10.24V
to +10.24V. Voltages delivered to these sockets should not exceed this range.
•Status indicators: Status indicator A is illuminated when the CLAMP Controller software has successfully connected
to the unit. Status indicator B is controlled by DIGITAL IN 1; status indicator C is controlled by DIGITAL IN 2. These
LEDs can be used to monitor the status of digital signals that are recorded in synchrony with the patch clamp amplifiers.
•Power indicator: This red LED is illuminated when the Intan CLAMP Controller is powered.
Rear Panel
The rear panel of the Intan CLAMP Controller provides auxiliary output lines as well as other ports and switches. From left to right:
•Analog outputs: Two BNC sockets are provided for monitoring measured signals or clamp signals from CLAMP
headstages. The headstages communicate with the controller using purely digital signals, but 16-bit DACs are used to
reconstruct analog signals with desired scaling factors. The control software allows users to route selected signals to
any analog output ports. These ports have a -10.24V to +10.24V voltage range.
•Audio line out jack: This standard 3.5-mm stereo phone jack allows users to connect an audio amplifier to the controller
and listen to the signals routed to the two analog output ports. ANALOG OUT 1 is connected to the left channel;
ANALOG OUT 2 is connected to the right channel. This port cannot drive speakers directly; an audio amplifier should
be used, and the volume should be adjusted carefully to ensure that excessive levels are not delivered to speakers.
• High-speed port: This connector is reserved for future use.
• I/O expansion port: This connector is used to add an Intan I/O Expander. This board is described in the next section.
It provides six additional analog inputs and outputs and 14 additional digital inputs and outputs for more complex
experiments. Signals on this port are digital and serially encoded, and are not easily accessed without the I/O Expander.
www.intantech.com ● info@intantech.com4
Intan CLAMP System
•CONFIG switches: Configuration switches 1-3 are reserved for future use. Switch 4 (CONFIG4) is used to select the
voltage level of the digital output ports (see next item). With CONFIG4 in the down position, 3.3V digital signals are
generated. With CONFIG4 in the up position, 5.0V digital signals are generated.
•Digital outputs: Two BNC sockets produce either 3.3V or 5.0V digital signals (see previous item) that can be
synchronized with software-generated voltage clamp or current clamp waveforms.
• USB port: A USB 2.0 port provides two connection to a host computer running the control software.
• Sample clock out: This port generates a digital square wave proportional to the headstage sampling rate when the
headstages are active. While the software acquires data with a sampling rate of 50 kS/s, the headstages are actually
sampled at 200 kS/s to reduce noise, so this port produces a 200 kHz square wave during operation. The voltage level
of this signal is set by the CONFIG4 switch.
•Mark out: This port generates digital pulses marking the onset and offset of voltage clamp or current clamp waveforms.
The voltage level of this signal is set by the CONFIG4 switch.
•I/O GND: This binding post is connected to the controller system ground used by all analog and digital inputs and
outputs. This is the preferred ground to use for Faraday cage and other shielding connections.
•Chassis GND: This binding post is connected to the controller chassis and to the grounding conductor of the AC power
socket.
•Power switch and fuse holder: The unit uses two standard 1A 250V 5x20mm slow blow fuses that can be replaced
by opening the fuse holder to the right of the power switch. The power cord must be removed to access the fuses.
•AC power socket: The controller is powered by 90-260V AC power, and is compatible with international voltage levels.
A US-style power cord is supplied with the controller. International customers must use an adapter to accommodate
non-US power sockets. The center grounding conductor must be connected to earth ground to avoid electric shock
hazards.
Mounting
The Intan CLAMP controller can be rack mounted on a standard 19” instrument rack using provided hardware, or it can be used
on a bench top by folding out the feet on the bottom of the case:
www.intantech.com ● info@intantech.com5
Intan CLAMP System
Intan I/O Expander
Intan Technologies offers an optional I/O Expander (sold separately) that provides an additional six analog inputs and outputs and
an additional 14 digital inputs and outputs. This unit is shown below:
Front Panel
The front panel of the Intan I/O Expander provides auxiliary digital and analog inputs, and analog outputs. From left to right:
•Analog outputs: Two analog outputs for monitoring measured signals or clamp signals from CLAMP headstages. (Four
more analog outputs are provided on the rear panel.) These ports have a -10.24V to +10.24V voltage range.
•Digital inputs: Six BNC sockets are provided for recording digital signals in synchrony with the headstage signals.
(Eight more digital inputs are provided on the rear panel.)
•Analog inputs: Six BNC sockets are provided for recording analog signals, or for controlling the headstages using
externally provided analog waveforms. Signals are digitized with 16-bit ADCs over a range of -10.24V to +10.24V.
•Power indicator: This red LED is illuminated when the Intan I/O Expander is powered. The I/O Expander receives low-
voltage DC power over an interface cable from the controller.
Rear Panel
The rear panel of the Intan I/O Expander provides auxiliary input and outputs lines. From left to right:
• Interface port: This connector is used to interface with the main controller unit.
• Analog outputs: Four analog outputs for monitoring measured signals or clamp signals from CLAMP headstages. (Two
more analog outputs are provided on the front panel.) These ports have a -10.24V to +10.24V voltage range.
•Digital outputs: Six BNC sockets produce either 3.3V or 5.0V digital signals that can be synchronized with software-
generated voltage clamp or current clamp waveforms. The CONFIG4 switch on the main Intan controller selects the
voltage level used by these ports.
•Digital inputs 9-16: Eight additional digital inputs are provided on screw terminal blocks. System ground connections
are also provided on the ends of the terminal block.
•Digital outputs 9-16: Eight additional digital outputs are provided on screw terminal blocks. System ground connections
are also provided on the ends of the terminal block. The CONFIG4 switch on the main Intan controller selects the
voltage level used by these ports.
www.intantech.com ● info@intantech.com6
Intan CLAMP System
Software Installation
Before attaching the Intan CLAMP Controller to a computer,
first install the drivers provided on the Intan website. (If you
have already installed drivers for other Intan USB-linked
products, you do not need to reinstall these drivers.) After
installing the drivers, connect the CLAMP Controller to the
computer using the supplied USB cable, plug in the
Controller, and turn on the power switch on the rear panel.
Windows users may download the compiled CLAMP UI
(User Interface) software from the Intan website. The
executable, DLL files, and
the same directory. To run the software, double click the
executable.
Plug in at least one CLAMP headstage to one of the ports
on the front panel and start the software. The software takes
a few seconds to calibrate the attached headstages and
then opens two windows: a Control Window (see right) and
a Data Display Window (see below).
If any errors show up when the software is run the first time,
these can be corrected by installing the
Redistributable Packages for Visual Studio 2013and the
Visual C++ Redistributable Packages for Visual Studio 2015
from Microsoft.
main.bit file should be kept in
Visual C++
the Record button is enabled. Clicking this button starts
data acquisition and saves the data to a file. Auxiliary data
from all ANALOG IN, DIGITAL IN, and DIGITAL OUT ports
are also saved unless the Save Auxiliary I/O item in the
Options menu is unchecked.
Control Window
Run and Run Once: These buttons start data acquisition
and display the measured and clamp data in real time
without saving data to disk. The Stop button halts data
acquisition. All data are acquired at 50 kSamples/s.
Select Base Filename: This button opens a dialog that
prompts the user to select a path and filename for saved
data files. A time and date stamp will be added to all saved
data files. Once a valid base filename has been selected,
Port Tabs: Clicking on these tabs (i.e., Port A, Port B)
selects headstages attached to the ports on the front panel
of the Intan CLAMP Controller. Each headstage may be
configured independently with its own holding levels, clamp
mode, capacitance compensation settings, etc. A softwaregenerated voltage clamp or current clamp waveform may
only be run on one headstage at a time; all other headstages
run in holding mode, but measurements are made on all
headstages simultaneously. Any external commands
www.intantech.com ● info@intantech.com7
Intan CLAMP System
coming from ANALOG IN channels will be applied to all
enabled headstages (see External Command Tab below).
Capacitance Compensation: This control enables and
adjusts the fast transient capacitance compensation circuit
in each headstage which is used to cancel pipette
capacitance. Capacitance compensation is typically
performed when the electrode is in the bath but prior to cell
contact. It may be refined after a gigaseal has been
obtained. See the Capacitance Compensation section
below for more details.
Status: This indicator shows the current clamp status of the
selected headstage.
Voltage Clamp Controls
Clicking on the Voltage Clamp tab displays the following
voltage clamp controls:
Pipette Offset: When a pipette electrode is placed in an
electrolytic solution, a small DC potential is generated by the
interface of dissimilar materials. This control allows the user
to adjust the pipette offset to null out this built-in potential.
Pipette offset should be nulled after the electrode is placed
in solution but prior to contact with a cell. If the Auto button
is clicked, the pipette offset will be adjusted to minimize
measured current. The lock can be clicked to prevent this
value from being changed accidentally.
The pipette offset may be adjusted manually by clicking on
the blue voltage indicator and turning the mouse scroll
wheel, moving the cursor keys up or down, or typing a
number.
Holding Voltage: This control sets a baseline clamp voltage
that is superimposed on all other voltage clamp commands.
After breaking into a cell, this is typically set close to -70 mV
to match the intracellular potential and prevent large
currents from flowing into or out of the cell.
Voltage Clamp Command: These buttons allow the user to
select between different types of voltage clamp waveforms.
Holding Only can be used to monitor spontaneous activity
in a cell. Seal Test produces a repeating voltage step, and
is typically used to monitor the pipette resistance as a
gigaseal is being formed or to estimate cell parameters (see
below). Resistance is similar to Seal Test, but adds a
running plot of resistance vs. time so that resistance trends
may be monitored visually. Multi-Step generates multiple
voltage clamp steps and superimposes the plots. Arbitrary
generates a waveform specified by a user file (see Arbitrary
Waveform Specification below for details).
Current Measurement Range: This control allows the user
to select the range of the current that is sensed by the
headstage: ±5 nA, ±10 nA, ±100 nA, or ±1000 nA. Using
the smallest acceptable range minimizes the noise floor.
Cell Parameters: If this box is checked, the passive cell
parameters R
resistance), and C
(series resistance), RM (membrane
S
(membrane capacitance) will be
M
estimated by fitting an exponential curve to the measured
voltage clamp data (see Estimating Membrane Parameters
below for details). Unchecking this box will “freeze” these
values.
Zap: This button must be enabled by clicking the lock icon.
Clicking the Zap button produces a brief, high voltage pulse
(typically 1 V) that can be used to break into a cell once a
gigaseal has been established (although most researchers
use pressure pulses to break the membrane). The
parameters of the zap pulse can be adjusted by clicking on
the blue numerical display.
Plot total Vclamp: Checking this box adds a second plot on
the Clamp Voltage graph in the lower half of the Data Display
Window. The primary plot shows the voltage clamp
waveform generated by the Voltage Clamp Command
controls above. The second plot shows the total voltage
clamp waveform applied by the headstage, which is a linear
superposition of the software-defined voltage clamp
command, the pipette offset, and any external command
delivered to an ANALOG IN port, if this option is enabled
(see External Command Tab below).
Plot Vcell: Checking this box adds a plot to the Clamp
Voltage graph in the lower half of the Data Display Window
that shows the cell voltage V
, which is defined as the
cell
clamp voltage minus the voltage drop across the series
resistance. This voltage drop is the series resistance (R
S
multiplied by the instantaneous measured electrode current.
The Intan CLAMP System does not allow for series
resistance compensation in voltage clamp mode, but
observing V
can allow users to manually compensate for
cell
some aspects of the voltage drop across the series
resistance and perhaps adjust the voltage clamp waveforms
to yield the desired voltage at the cell.
Current Clamp Controls
Clicking on the Current Clamp tab displays the following
current clamp controls:
Pipette Offset: See “Voltage Clamp Controls” for a
description of this control.
Holding Current: This control sets a baseline clamp current
that is superimposed on all other current clamp commands.
This can be used to bias the cell to a desired intracellular
potential or suppress spontaneous activity. The Zero Current button may be clicked to set the holding current to
zero quickly.
Current Clamp Command: These buttons allow the user to
select between different types of current clamp waveforms.
Holding Only can be used to monitor spontaneous activity
in a cell. Tuning produces a repeating current step, and can
www.intantech.com ● info@intantech.com8
)
Intan CLAMP System
be used to tune the bridge balance (see below). Multi-Step
generates multiple current clamp steps and superimposes
the plots. Arbitrary generates a waveform specified by a
user file (see Arbitrary Waveform Specification below for
details).
Current Clamp Range: This control allows the user to
select the range and resolution (i.e., step size) of the current
clamp. The current-output DAC that generates the
headstage clamp current has a range of ±127 steps. The
step size can be 5 pA, 50 pA, 0.5 nA, or 1 nA.
Bridge Balance: Series resistance compensation may be
performed in current clamp mode using this control.
Checking this box enables bridge balance, where the
voltage drop across the series resistance of the pipette is
subtracted from the voltage measurements. The default
value of series resistance is the last measurement of R
made in voltage clamp mode (see Cell Parameters above),
but this value can be adjusted here to minimize step artifacts
observed during Tuning current clamp pulses (see the
Bridge Balance section below for more details).
Buzz: This button must be enabled by clicking the lock icon.
Clicking this button modulates the capacitance
compensation circuitry to produce small AC current pulses
that may assist in forming a membrane seal around sharp
intracellular electrodes. The parameters of the buzz pulse
can be adjusted by clicking on the blue parameter display.
S
recording. To control headstages using only the external
commands, select Holding Only in the software clamp
commands and run the system.
Multiple headstages may be controlled simultaneously using
external command signals. Voltage waveforms may be
generated using legacy hardware (e.g., NIDAQ boards) and
connected to ANALOG IN ports to generate clamp control.
Plot total Iclamp: Checking this box adds a second plot on
the Clamp Current graph in the Data Display Window. The
primary plot shows the current clamp waveform generated
by the Current Clamp Command controls above. The
second plot shows the total current clamp waveform applied
by the headstage, which is a linear superposition of the
software-defined current clamp command and any external
command delivered to an ANALOG IN port, if this option is
enabled (see below).
Other Controls
Display Resistance: Checking this box enables real-time
display of the measured pipette resistance. This can be
used when establishing a gigaseal. Average “best fit” lines
used to calculate resistance are superimposed on the
measured waveform in the Data Display Window when this
feature is enabled.
Analog/Digital I/O Tabs
External Command Tab: Controls in this tab allow the user
to control the voltage clamp or current clamp in real time via
external command signals from any of the ANALOG IN
ports. Voltage clamp and current clamp scaling factors may
be selected for each headstage. The external command is
superimposed on any software-defined clamp command,
and only takes effect when the system is running or
Signal Out Tab: Controls in this tab allow the user to route
the measured voltage or current signal to any of the
ANALOG OUT ports for monitoring on an oscilloscope or
speaker. (The ANALOG OUT 1 and 2 signals are connected
to the left and right channels of the AUDIO LINE OUT jack
on the rear panel.) Voltage and current scaling factors can
be selected here as well. DACs in the Intan CLAMP
Controller reconstruct an analog waveform from the digital
www.intantech.com ● info@intantech.com9
Intan CLAMP System
headstage data in real time. Note: Any low-pass Bessel
filtering or bridge balancing is not applied to these
waveforms as these operations are performed in software.
Clamp Out Tab: Controls in this tab allow the user to
produce an analog waveform corresponding to the clamp
command signal on any of the ANALOG OUT ports. Voltage
and current scale factors can be selected here for voltage
clamp or current clamp commands. DACs in the Intan
CLAMP Controller reconstruct an analog waveform from the
digital clamp commands sent to the headstage.
Marker Out Tab: When a voltage clamp or current clamp
waveform is active, a digital signal on the MARK OUT port
on the rear panel of the Intan CLAMP Controllers goes high
when the clamp waveforms departs from the holding level.
This signal may also be routed to any of the DIGITAL OUT
ports using controls in this tab. Note: If a signal is routed to
a DIGITAL OUT port, then it will be saved in the auxiliary I/O
file.
Data Display Window
Low-Pass Filter: This control selects the cutoff frequency
th
for a 4
voltages and currents.
-order Bessel low-pass filter applied to measured
Save Displayed: This button is enabled by selecting a base
filename in the Control Window. When the button is clicked,
the data visible in the Data Display Window are saved to
disk.
Clear: The button clears the plots in this window.
Capacitance
Compensation
Each CLAMP headstage supports pipette capacitance
cancellation in the range of 0-20 pF. In practice fast
transient capacitance compensation is usually tuned when
the pipette is placed into bath, before contact with a cell is
made. As shown in the plot below, a voltage clamp step is
applied to the electrode and the capacitance compensation
is adjusted to eliminate the transient peaks in the measured
current. The pipette resistance is easily measured during
this procedure as well by checking the Display Resistance
box. Note that excessive capacitance compensation can
cause increased noise levels and instability.
Autoscale: This box may be unchecked to enable manual
control of the plotting axes. Axes may be adjusted by
clicking and dragging along the axis (see figure below),
using the cursor keys, the page up/down keys, or other keys
described in the Keyboard Shortcuts dialog accessible
through the Help menu in the Control Window.
Overlay multi-steps: When multi-step clamp commands
are selected, the results from each voltage clamp or current
clamp step are superimposed on top of one another unless
this box is unchecked.
Estimating Membrane
Parameters
In whole-cell patch clamp electrophysiology, the electrode is
electrically connected to the inside of a cell as shown in the
diagram below. The tiny opening in the cell membrane
creates a large access resistance R
pipette resistance R
can have a value in the tens of megohms. To first order the
cell’s membrane behaves as a capacitance C
the tens of picofarads) with some parallel resistance R
(typically in the hundreds of megohms) in series with a
reversal potential V
to -90 mV). These membrane parameters, as well as the
series resistance R
voltage clamp measurement and checking the Cell Parameters box.
. The resulting series resistance RS
P
(usually in the range of -30 mV
R
, can be estimated using a simple
S
which adds to the
A
(typically in
M
M
www.intantech.com ● info@intantech.com10
Intan CLAMP System
Note that the electrode parasitic capacitance CP is not
shown in the diagram below because it is assumed that this
element has already been cancelled by the fast transient
capacitance compensation circuitry prior to cell contact.
After breaking into a cell, the voltage clamp is typically set to
a value near the cell’s expected resting potential (e.g., -70
mV) to prevent the activation of voltage-gated ion channels.
A small voltage step is applied around this potential (e.g., a
step from -70 mV to -60 mV) and the resulting current is
measured. The plot below shows a typical clamp voltage
profile and measured current waveform. The current takes
the form of a decaying exponential with a time constant τ:
·exp(-t/τ) + I1. (These exponential current peaks are
i(t) = I
0
much wider and slower than the “fast transient” peaks
caused by uncompensated pipette capacitance)
The Intan CLAMP system does not support series
resistance compensation in voltage clamp mode, but the
Plot VCell control (see Voltage Clamp Controls above) can
be used to estimate the true cell potential by subtracting the
voltage drop across the series resistance in real time. The
user may then adjust the voltage clamp parameters to
produce the desired voltage in the cell.
Bridge Balance
When applying current clamp pulses through real electrodes
connected to cells, the measured membrane potential is
distorted by an artifact caused by the voltage drop across
the electrode. Referring to the diagram below, we would like
to apply a current signal I
voltage V
equal to the intracellular potential V
elec
However, the clamp current flows through a series
resistance R
access resistance R
is equal to I
comprised of the pipette resistance RP and the
S
. This leads to a voltage offset ΔV that
A
.
clamp·RS
and measure an electrode
clamp
.
cell
Enabling Cell Parameters causes an exponential curve to
be fit to the data and the membrane parameters R
can be estimated, along with the series resistance R
and CM
M
.
S
Traditional patch clamp instruments sometimes use positive
feedback circuits to reduce the effects of the series
resistance R
in voltage clamp mode. (The measured
S
current flowing through the electrode creates a voltage drop
across R
, causing the voltage in the cell to be slightly
S
different from the clamp voltage.)
The plot above shows an example of this phenomenon for a
simple clamp current that starts at zero and pulses to I
Luckily, the intracellular potential V
the measured electrode voltage V
can be recovered from
cell
by subtracting the
elec
clamp current (at every instant in time) multiplied by R
stim
S
which was previously measured using the Cell Parameters
www.intantech.com ● info@intantech.com11
.
,
Intan CLAMP System
tool. In the old days of purely analog instrumentation, this
subtraction was performed using an analog circuit in a
technique called bridge balance or series resistance compensation. With modern computing power it is much
easier to perform this subtraction after the electrode voltage
has been digitized. The Bridge Balance tool in the Current
Clamp tab performs this function. In practice, the exact
value of series resistance should be adjusted while applying
small current pulses to eliminate the voltage jump shown in
the figure above.
Note that the clamp current is exactly the same with or
without bridge balance enabled; the stimulation experienced
by the cell is not affected, only the interpretation of the
measured data.
Arbitrary Waveform
Specification
Users may define arbitrary voltage clamp or current clamp
waveforms by creating CSV (comma separated values) text
files listing all clamp levels and durations comprising the
waveform.
Arbitrary waveform files must end in the .arb prefix, and the
first line of the file must read
waveform or
with the second line of the text file, each line specifies a
voltage or current clamp level and a time duration that this
clamp level should be held. Clamp levels are specified in
integer steps. (Smooth changes in clamp levels must be
approximated by “stair steps”.)
The standard voltage clamp step size is 2.5 mV, but if 2x Voltage Clamp Mode is enabled in the Options menu, the
step size is 5.0 mV (see next section for details). The
voltage clamp level can range between -255 and +255
steps. The current clamp step size is set by the Current
Clamp Range control in the Current Clamp tab. The step
size can be 5 pA, 50 pA, 0.5 nA, or 1 nA. The current clamp
level can range between -127 and +127 steps.
The duration each clamp level is held is specified by a
positive integer in the range of 1 to 65535. The actual
duration is calculated by multiplying this number by
20 microseconds (i.e., 1 / 50 kHz), so a value of 25000
signifies 500 milliseconds, for example. The maximum
value of 65535 sets a duration of 1.31 seconds; if longer
holding times are needed then repeated commands must be
used.
The clamp level and duration are specified by two numbers
separated with a comma. For example, to define a voltage
clamp waveform that holds at zero for 20 ms, then steps to
+20 mV for 40 ms, and then steps to -30 mV for 10 ms, the
.arb file would be:
ARBI for a current clamp waveform. Starting
ARBV for a voltage clamp
ARBV
0,1000
8,2000
-12,500
The total length of an arbitrary waveform specification is
limited to approximately 16000 lines.
By default, the MARK OUT digital signal for an arbitrary
waveform will go high for any nonzero clamp level. More
sophisticated digital signals can be specified by appending
two additional binary numbers (i.e., 0 or 1) to each line. The
first number signals when the MARK OUT signal should go
high. The second number signals when the optional
DIGITAL OUT signal (controlled by the Marker Out tab)
should go high.
Adding to the previous example, if we wanted MARK OUT
to go high during the +20 mV clamp level and the selected
DIGITAL OUT signal to go high during the -30 mV clamp
level, we would use the following file:
ARBV
0,1000,0,0
8,2000,1,0
-12,500,0,1
2x Voltage Clamp Mode for
FSCV
The Intan CLAMP chip in each headstage uses an on-chip
DAC to generate a clamp voltage. The voltage level of this
DAC is set by an 8-bit magnitude variable and 1-bit sign
variable. In the normal mode of operation, this DAC has
2.5 mV steps, giving a total voltage clamp range of
±637.5 mV, which is more than enough range for cellular
patch clamp experiments. By selecting the 2x Voltage Clamp Mode from the Options menu, the DAC step size is
doubled to 5.0 mV, producing a range of ±1.275 V.
This wide voltage clamp range permits the Intan CLAMP
Controller to be used in fast-scan cyclic voltammetry (FSCV)
or other electrochemistry applications. In FSCV
experiments, voltages typically in the range of -0.4 V to
+1.3 V are swept rapidly across graphite electrodes while
the current is measured. (The Arbitrary Waveform function
can be used generate FSCV voltage clamp signals.)
Background currents on the order of 1 µA are subtracted
from baseline readings, and the resulting residual current
variations – typically less than 10 nA in magnitude – indicate
the presence of various neurotransmitters or other
chemicals.
When the 2x Voltage Clamp Mode is enabled, all clamp
voltages generated by the headstages are 2x larger than
they appear in the software. The controls and display axes
do not reflect this 2x increase. However, the MATLAB code
used to read saved data files (see next section) does
www.intantech.com ● info@intantech.com12
Intan CLAMP System
recognize data saved with the 2x mode enabled and
automatically adjusts voltage clamp waveforms to reflect the
true voltages.
www.intantech.com ● info@intantech.com13
Intan CLAMP System
Importing Recorded Data into MATLAB
Intan Technologies provides an open-source m-file (read_Intan_CLP_file.m) for importing data recorded from the CLAMP
Controller software into MATLAB. Make sure you have the latest version of this m-file to ensure compatibility with the newest
version of the control software. Running this m-file brings up a file selector dialog with which the user locates and selects the
desired .clp data file. The m-file then loads and parses the data file and returns a data structure containing the clamp and measured
waveforms, time vectors, and settings.
The CLAMP Controller software creates a new subdirectory each time data is saved. The directory has a name consisting of the
base filename specified by the user followed by a date stamp and a time stamp:
this subdirectory, a separate file is saved for each headstage, with a letter designating the headstage port. An additional file
designated AUX contains waveforms from the ANALOG IN, DIGITAL IN, and DIGITAL OUT ports sampled in synchrony with the
headstages. (If this file is not needed it can be disabled by unchecking the Save Auxiliary I/O item in the Options menu.) A typical
set of saved data files might have the following names:
The DigitalIn and DigitalOut vectors are 16-bit unsigned integers encoding all DIGITAL IN and DIGITAL OUT values.
(The Intan CLAMP Controller has two DIGITAL IN lines and two DIGITAL OUT lines; the Intan I/O Expander can be added to gain
access to all 16 signals.) For example, if DIGITAL IN 1 and DIGITAL IN 7 are both high and all other digital inputs are low, then
the value of
The time vectors in the auxiliary I/O files and individual headstage data files are identical.
DigitalIn at that point in time will be 2
0
+ 26 = 1 + 64 = 65.
www.intantech.com ● info@intantech.com15
Intan CLAMP System
intan
TECHNOLOGIES, LLC
Related CLAMP
Documentation
The following supporting datasheets may be found at
http://www.intantech.com/downloads:
♦ Intan CLAMP Voltage/Current Amplifier Chip
Datasheet
Pricing Information
See www.intantech.com for current pricing. All price
information is subject to change without notice. Quantities
may be limited. All orders are subject to current pricing at
time of acceptance by Intan Technologies. Additional
charges may apply for international purchases and
shipping.
Contact Information
This datasheet is meant to acquaint scientists and engineers
with the general characteristics of the Intan CLAMP system
developed at Intan Technologies. We value feedback from
potential end users.
For more information, contact Intan Technologies at:
Information furnished by Intan Technologies is believed to be accurate and reliable. However, no responsibility is assumed by Intan
Technologies for its use, nor for any infringements of patents or other rights of third parties that may result from its use. Specifications
subject to change without notice. Intan Technologies assumes no liability for applications assistance or customer product design.
Customers are responsible for their products and applications using Intan Technologies components. To minimize the risks
associated with customer products and applications, customers should provide adequate design and operating safeguards.
Intan Technologies’ products are not authorized for use as critical components in life support devices or systems. A critical component
is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the
life support device or system, or to affect its safety or effectiveness.
www.intantech.com ● info@intantech.com16
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.