INSURGICAL ™ MAXLiFe™ Battery Charger
Catalog # 3004022
INSTRUCTIONS FOR USE
INTRODUCTION
The information and procedures described in this manual are intended to assist medical
professionals in the safe and effective use, care, cleaning, and long-term maintenance of
INSURGICAL Battery Charger.
INTENDED USE
The INSURGICAL Battery Charger is intended for use with the INSURGICAL MAXLiFe battery Cat.
# 3002022 only.
WARNINGS
Only trained and experienced medical professionals should use this equipment. Failure
to comply with the INSURGICAL Battery Charger instructions for use (IFU) may result in
patient and/or medical staff injury.
Do not use if damage is apparent.
Use of eye protection is required while operating equipment.
DO NOT modify.
DO NOT use this equipment in the presence of flammable anesthetic and air, or with
oxygen or nitrous oxide.
DO NOT operate the Battery Charger with a damaged power cord or plug.
DO NOT modify the ground of the Battery Charger power cord.
Install the power cord of the Battery Charger directly into wall electrical outlet.
DO NOT disassemble or service the Battery Charger. Return to INSURGICAL for service
or repair. Failure to comply may result in electric shock or fire.
Always disconnect the power cord from the Battery Charger before performing cleaning
to reduce the risk of electric shock.
CAUTIONS:
DO NOT sterilize the Battery Charger.
DO NOT connect multiple Battery Charger units in a series.
ENVIRONMENTAL SPECIFICATIONS
02/2013 110-2008-001rB
Battery Charge Indicator Lights
Battery Charger - Cat # 3004022
FEATURES
Power Connection – connects and disconnects Battery Charger from facility power.
Charging Bay – accepts and charges INSURGICAL batteries (Cat. # 3002022).
INSURGICAL Battery Charger accommodates up to four (4) batteries.
Charging Bay and battery are mechanically protected, preventing incorrect introduction
of battery to charging bay.
Charging Bay Indicator Lights – provide battery status information when illuminated.
Overcharge Protection – terminates battery charge at 15.77 VDC.
Short Circuit Protection – prevents charger from commencing when battery has an
electrical short.
02/2013 110-2008-001rB
* Flashing yellow indicator light signals user that the battery has been charged more than 200 of
its rated 300+ charging cycles. Yellow indicator light will flash 3 times then resume normal
charging.