INSURGICAL MAXLiFe Instructions For Use Manual

INSTRUCTIONS FOR USE
INSURGICAL® MAXLiFe® Battery Catalog # 3002022
INTRODUCTION
INTENDED USE
WARNINGS
CLEANING
The information and procedures described in this manual are intended to assist medical professionals in the safe and effective use, care, cleaning, and long-term maintenance of INSURGICAL rechargeable batteries.
The INSURGICAL battery Cat. # 3002022 is intended for use with the INSURGICAL hand pieces only. The batteries must be charged with the INSURGICAL MAXLiFe™ Battery Charger, Cat. # 3004022 ONLY.
INSURGICAL hand-held devices are provided STERILE. Contents of tray are sterile unless opened or damaged. If tray or lid is opened, damaged, or if USE BY Date is exceeded, DO NOT USE. SINGLE-USE. DO NOT RESTERILIZE.
1) INSURGICAL batteries are for use by properly trained personnel who are completely familiar with the required surgical techniques and instructions for use of the equipment.
3) To assure the safety of the patient and operating room personnel, use only INSURGICAL rechargeable batteries with INSURGICAL hand-held devices.
4) Do not expose battery to fire or incineration.
5) Do not allow battery contacts to contact metal objects. Contact with metal objects may result in electrical shock, burn, or injury to user.
6) Do not disassemble or modify battery.
7) Do not incinerate or heat the battery.
8) Do not use damaged battery.
9) Avoid contact with leaking battery. In the event of eye contact with electrolyte, wash with copious amounts of fresh water and seek medical attention immediately.
10) Do not sterilize battery.
With a cloth, wipe down the battery with a neutral enzymatic or neutral detergent solution before and after each use. CAUTION: Do not use solvents to disinfect the batteries. Battery poles must not contact water or solvents: Danger of short circuiting, causing burn or injury to user!
ENVIRONMENTAL
SPECIFICATIONS
OPERATIONAL SPECIFICATIONS
/CERTIFICATIONS
CAUTION
EXPLANATION OF SYMBOLS
Storage Temperatur:
Operating Conditions:
Batteries can be stored separately in a dry cool place or left in charger that is powered on.
Charging Power: 19.54VDC Output Power: 13.2VDC Conforms to: UL1642, UL2054, EN 60601, UN38.3
1) Batteries contain Lithium Iron Phosphate (LiFePo4), which can be recycled or disposed of as municipal waste. Battery should be fully discharged prior to recycling or disposal. Dispose or recycle in accordance with your local government regulations.
2) Always use fully charged batteries to ensure maximum run time.
3) Battery charge cycle is 2.5 hours from full discharge to full charge. Typical charge time will vary depending on amount of charge present when placed on charger.
INSTRUCTIONS FOR USE
INSURGICAL® MAXLiFe® Battery Catalog # 3002022
Cat. #3002022
Battery
Contacts
Battery
Compartment
Battery Insertion Shield
FEATURES
Battery
Safety Key
BATTERY INSTALLATION/
REMOVAL
Battery
Compartment
1) Finger Hold – recessed area aids in the removal of the battery after use.
2) Battery Charge Indicator – on board charge indicator that shows battery charge level via 3 indicator lights.
Finger Hold
Battery Charge Indicator
3) Safety Key – a mechanical safety feature that ensures battery cannot be placed into tool or charger incorrectly.
4) Battery Contacts – Conducts charge from battery charger to battery and distributes charge to hand held devices.
1) Circulating person: Remove the hand piece from its outer package, and transfer to the sterile field using aseptic technique.
2) Scrubbed person: Ensure the mode switch is in the LOCK position.
3) Scrubbed person: Ensure battery insertion shield is affixed securely to the hand piece. The battery insertion shield helps guide the battery into the battery compartment and prevents contamination
Battery
of the hand piece by the non sterile battery.
4) Circulating person: Determine battery charge level by pressing the TEST button on the battery.
Cover
A light on the battery will illuminate accordingly: Green Fully Charged
5) Circulating person: Insert the battery through the battery insertion shield into the battery compartment, pressing down until the battery is fully seated. Remove the battery insertion shield.
Tabs
6) Scrubbed person: Affix the battery compartment cover by guiding the tabs into the slots on each side of the battery compartment. Press firmly until the battery compartment cover is fully seated.
7) To remove the battery, pull while pressing the tabs on the battery compartment cover. IMPORTANT: Do not attempt to remove and replace battery while in the sterile field. Battery is not sterile.
8) To clean battery, wipe down with a cloth with a neutral enzymatic or neutral detergent solution before and after each use. IMPORTANT: do not use solvents to disinfect the batteries. Battery poles must not contact water or solvents: Danger of short circuiting!
9) DO NOT STERILIZE BATTERY.
Battery Charge Indicator Lights: Green Fully Charged Blue 50 to 80 % Charge Yellow Less than 50% Charge
Blue 50 to 80 % Charge Yellow Less than 50% Charge
SPECIFICATIONS
PRODUCT DISPOSAL
WARRANTY
Weight 274 g Voltage 13.2V Charging Time 30 Min to 2.5 Hours Run Time Fully Charged 25 to 40 minutes
Battery Packs contain Lithium Iron Phosphate (LiFePo4), which can be recycled or disposed of as municipal waste. Battery should be fully discharged prior to recycling or disposal. Dispose or recycle in accordance with local governmental regulations.
INSURGICAL warrants its batteries to be free from defects in material and workmanship for a period of ninety (90) days from the invoice date. The warranty is limited to the replacement of the product without charge. This warranty is void in the event of any of the following: abuse, misuse, or use in other than normal surgery environment, disassembly, alteration or unauthorized repair, or in the event that the product has not been used in reasonable manner and in compliance with the written instructions provided by INSURGICAL.
Contact INSURGICAL for complaint reporting at: quality@insurgical.com
Caution: Federal Law (USA) restricts this device to sell by or on the order of a physician.
INSURGICAL, INC.
9111 Jollyville Road #230 Austin, Texas 78759 +1 (800) 936-9296 insurgical.com
110-2009-001revD