Instek MDO-2000EG, MDO-2000EX User Manual

Mixed-Domain Oscilloscope
MDO-2000EG & MDO-2000EX
USER MANUAL
ISO-9001 CERTIFIED MANUFACTURER
May 2017
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan
TABLE OF CONTENTS
Table of Contents
SAFETY INSTRUCTIONS .................................................. 4
GETTING STARTED .......................................................... 9
MDO-2000EG/2000EX Series Overview 10
Appearance .......................................... 14
Set Up .................................................. 27
Built-in Help ........................................ 37
MEASUREMENT ............................................................ 38
Basic Measurement ............................. 39
Automatic Measurement ...................... 46
Cursor Measurement ........................... 60
Math Operation ................................... 67
ADVANCED CONFIGURATION ....................................... 77
Acquisition .......................................... 80
Segmented Memory Acquisition .......... 85
Display................................................. 97
Horizontal View ................................. 102
Vertical View (Channel) ...................... 110
Bus Key Configuration........................ 118
Trigger ............................................... 140
Search ................................................ 172
System Settings and Miscellaneous
Settings ............................................. 181
ARBITRARY WAVE GENERATOR .................................... 186
Arbitrary Wave Generator Operation .. 187
SPECTRUM ANALYZER .................................................. 220
Spectrum Analyzer operation ............. 221
1
MDO-2000EG & 2000EX User Manual
DMM ............................................................................ 239
DMM function ................................... 240
POWER SUPPLY ............................................................ 249
Power supply function ........................ 250
APPLICATIONS ............................................................. 253
Introduction ....................................... 254
Go-NoGo application ......................... 256
DVM application ................................ 261
Data Log application .......................... 263
Digital Filter application .................... 265
Mask application................................ 267
SAVE/RECALL ............................................................... 277
File Format/Utility ............................. 278
Create/Edit Labels .............................. 284
Save ................................................... 287
Recall ................................................. 295
Reference Waveforms ......................... 301
FILE UTILITIES ............................................................. 303
File Navigation .................................. 304
Create Folder...................................... 306
Rename File ....................................... 307
Delete File or Folder .......................... 308
Copy File to USB ................................ 309
HARDCOPY KEY ........................................................... 310
REMOTE CONTROL CONFIG ........................................ 314
Interface Configuration ...................... 315
MAINTENANCE ............................................................ 325
2
TABLE OF CONTENTS
FAQ .............................................................................. 331
APPENDIX .................................................................... 333
Updating the Firmware ...................... 334
MDO-2000EG/2000EX Specifications . 336
Probe Specifications .......................... 342
Dimensions ....................................... 344
Declaration of Conformity .................. 345
INDEX .......................................................................... 346
3
MDO-2000EG & 2000EX User Manual
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the equipment or to other properties.
DANGER High Voltage
(Note)
Attention required Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the Product name.
4
SAFETY INSTRUCTIONS
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
WARNING
CAUTION
Make sure the BNC input voltage does not
exceed 300Vrms.
Never connect a hazardous live voltage to the
ground side of the BNC connectors. It might lead to fire and electric shock.
Do not place any heavy object on the MDO-
2000EG/2000EX.
Avoid severe impact or rough handling that
leads to damaging the MDO-2000EG/2000EX.
Do not discharge static electricity to the MDO-
2000EG/2000EX.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not perform measurement at a power source
or building installation (Note below).
Do not disassemble the MDO-2000EG/2000EX
unless you are qualified.
Note
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The MDO-2000EG/2000EX falls under category I.
Measurement category IV is for measurements performed at the
source of low-voltage installation.
Measurement category III is for measurements performed in the
building installation.
Measurement category II is for measurements performed on
circuits directly connected to the low voltage installation.
Measurement category I is for measurement performed on
circuits not directly connected to Mains.
Safety Guidelines
5
MDO-2000EG & 2000EX User Manual
Power Supply
WARNING
AC Input voltage: 100 - 240V AC, 50 - 60Hz,
auto selection. Power consumption: 40W for MDO-2000EG; 50W for MDO-2000EX.
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid electrical shock.
Cleaning the MDO­2000EG/2000EX
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh materials
such as benzene, toluene, xylene and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: ≤80%, 40°C or below; 45%,
41°C ~ 50°C
Altitude: < 2000m Temperature: 0°C to 50°C
Note
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The MDO-2000EG/2000EX falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
6
SAFETY INSTRUCTIONS
Storage environment
Location: Indoor Temperature: -10°C to 60°C Humidity: Up to 93% RH (non-condensing) /
≤40ºC, up to 65% RH (non-condensing) / 41ºC ~ 60 ºC
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
7
MDO-2000EG & 2000EX User Manual
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the oscilloscope in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
8
GETTING STARTED
MDO-2000EG/2000EX Series Overview .................... 10
Integrated instruments and series lineup ..................................... 10
Main Features ................................................................................... 11
Accessories ........................................................................................ 12
Appearance .............................................................. 14
MDO-2000EG/2000EX 4-channel models Front Panel ......... 14
MDO-2000EG Rear Panel ............................................................. 22
MDO-2000EX Rear Panel ............................................................. 22
Display ............................................................................................... 24
Set Up ..................................................................... 27
Tilt Stand ........................................................................................... 27
Power Up .......................................................................................... 28
First Time Use .................................................................................. 29
How to Use This Manual ............................................................... 32
Built-in Help ............................................................ 37
GETTING STARTED
This chapter describes the MDO-2000EG/2000EX in a nutshell, including its main features and front / rear panel. After going through the overview, follow the Set Up section to properly set up the device for first time use. The Set Up section also includes an introduction on how to use this manual effectively.
9
MDO-2000EG & 2000EX User Manual
Model name
AWG
(1)
DMM
(2)
SA
(3)
Power
(4)
MDO-2000EG
✗ ✓ ✗
MDO-2000EX
✓ ✓ ✓
Model name
Frequency bandwidth
Input channels
Max. Real-time Sampling Rate
MDO-2072EG / 2072EX
70MHz
2
1GSa/s
MDO-2074EG / 2074EX
70MHz
4
1GSa/s
MDO-2102EG / 2102EX
100MHz
2
1GSa/s
MDO-2104EG / 2104EX
100MHz
4
1GSa/s
MDO-2202EG / 2202EX
200MHz
2
1GSa/s
MDO-2204EG / 2204EX
200MHz
4
1GSa/s
MDO-2000EG/2000EX Series Overview
Integrated instruments and series lineup
The MDO-2000EG and MDO-2000EX have different hardware implemented instruments:
(1) Arbitrary Wave Generator (2) Digital Multi-Meter (3) Spectrum Analyzer (4) Power Supply
The MDO-2000EG/2000EX series consists of 12 models, divided into 2-channel and 4-channel versions. Note that throughout the user manual, the term “MDO-2000EG/2000EX” refers to all models of the series, unless stated otherwise.
10
GETTING STARTED
Features
8 inch, 800 x 480, WVGA TFT display. Available from 70MHz to 200MHz. Real-time sampling rate of 1GSa/s (2 channel
models), 1GSa/s max. (4 channel models).
Deep memory: 10M points record length. Waveform capture rate of 120,000 waveforms
per second.
Vertical sensitivity: 1mV/div~10V/div. Segmented Memory: Optimizes the acquisition
memory to selectively capture only the important signal details. Up to 29,000 successive waveform segments can be captured with a time-tag resolution of 4ns.
Waveform Search: Allows the scope to search
for a number of different signal events.
Arbitrary Wave generator: Full-function dual
channel arbitrary waveform generator.
Spectrum Analyzer: A handy tool to perform
signal analysis in the frequency domain.
DMM: Up to 5000 counts and DCV, DCA, ACV,
ACA, R, Diode, Continuity and temperature measurement were supported (MDO-2000EX only)
Power Supply: Dual channel output, 1V to 5V
continuously adjustable (0.1V step) (MDO­2000EX only)
Powerful embedded applications such as: Data
Logging, Digital Voltmeter, Go-No Go, Mask, Digital filter, etc..
On-screen Help. 32 MB internal flash disk.
Main Features
11
Interface
USB host port: front panel, for storage devices. USB device port: rear panel, for remote control
or printing.
Ethernet port as standard. Probe compensation output with selectable
output frequency (1kHz ~ 200kHz).
Calibration output.
Accessories
Standard Accessories
Part number
Description N/A region dependent
Power cord
GTP-70B-4, for MDO-2072EG/2072EX MDO-2074EG/2074EX
Passive probe; 70 MHz
GTP-100B-4, for MDO-2102EG/2102EX MDO-2104EG/2104EX
Passive probe; 100 MHz
GTP-200B-4, for MDO-2202EG/2202EX MDO-2204EG/2204EX
Passive probe; 200 MHz
GTL-105A
Test lead for power supply (for MDO-2000EX only).
GTL-110
Test lead for AGW.
GTL-207
Test lead for DMM (for MDO­2000EX only).
Standard Apps
Name
Description
Go-NoGo
Go-NoGo testing app.
DataLog
Waveform or image data logging app.
DVM
Digital Voltmeter app.
MDO-2000EG & 2000EX User Manual
12
GETTING STARTED
Digital Filter
High ,low or band pass digital filter for analog inputs.
Mask
Creates shape templates for signal comparison.
Remote Disk
Allows the scope to mount a network share drive.
Demo mode
Demonstration mode that is used with the GDB-03 demo board.
Optional Accessories
Part number
Description GDB-03
Demo board.
Drivers, others
USB driver
LabVIEW driver
13
MDO-2000EG & 2000EX User Manual
Side menu keys
LCD
Variable knob and Select key
Autoset, Run/Stop, Single and Default settings
Trigger controls
Function keys
USB Host port, Probe Compensation terminals
Power button
Hardcopy key
Math, Reference and Bus keys
Bottom menu keys
Horizontal controls
Vertical controls
Option key
Menu key
DMM ports
CH1~CH4
LCD
Variable knob and Select key
Autoset, Run/Stop, Single and Default settings
CH1~CH2, EXT TRIG
Trigger controls
Function keys
USB Host port, Probe Compensation terminals
Hardcopy key
Math, Reference and Bus keys
Bottom menu keys
Vertical controls
Option key
Menu key
Side menu keys
DMM ports
Power button
Horizontal controls
EXT TRIG
Appearance
MDO-2000EG/2000EX 4-channel models Front Panel
MDO-2000EG/2000EX 2-channel models Front Panel
14
GETTING STARTED
LCD Display
8” WVGA TFT color LCD. 800 x 480 resolution, wide angle view display.
Menu Off Key
Use the Menu Off key to hide the onscreen menu system.
Option Key
The Option key is used to access installed options.
Menu Keys
The side menu and bottom menu keys are used to make selections from the soft-menus on the LCD user interface.
To choose menu items, use the 7 Bottom menu keys located on the bottom of the display panel.
To select a variable or option from a menu, use the side menu keys on the side of the panel. See page 32 for details.
VARIABLE
POSITION
HORIZONTAL
TRIGGER
LEVEL
PUSH TO
ZERO
PUSH TO
ZERO
SCALE
Measure
Display
Cursor
Help
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Select
Search
Set/Clear
Zoom
Menu
50 %
Force-Trig
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALESCALE
PUSH TO
ZERO
PUSH TO
ZERO
M
R
B
Digital Storage Oscilloscope
GDS-2202E
200 MHz 1 GS/s
Visual Persistence Oscilloscope
Side menu keys
Bottom menu keys
Hardcopy Key
The Hardcopy key is a quick-save or quick-print key, depending on its configuration. For more information see pages 312 (save) or 311 (print).
15
MDO-2000EG & 2000EX User Manual
Variable Knob and Select Key
Select
VARIABLE
The Variable knob is used to increase/decrease values or to move between parameters.
The Select key, when lit up, is used to make selections.
Function Keys
The Function keys are used to enter and configure different functions on the MDO-2000EG/2000EX.
Measure
Measure
Configures and runs automatic measurements.
Cursor
Cursor
Configures and runs cursor measurements.
APP
APP
Configures and runs GW Instek applications.
Acquire
Acquire
Configures the acquisition mode, including Segmented Memory acquisition.
Display
Display
Configures the display settings.
Help
Help
Shows the Help menu.
Save/Recall
Save/Recall
Used to save and recall waveforms, images, panel settings.
16
GETTING STARTED
Utility
Utility
Configures the Hardcopy key, display time, language, probe compensation and calibration. It also accesses the file utilities menu.
Autoset
Autoset
Press the Autoset key to automatically set the trigger, horizontal scale and vertical scale.
Run/Stop Key
Run/Stop
Press to Freeze (Stop) or continue (Run) signal acquisition (page 42). The run stop key is also used to run or stop Segmented Memory acquisition (page 88).
Single
Single
Sets the acquisition mode to single triggering mode.
Default Setup
Default
Resets the oscilloscope to the default settings.
Horizontal Controls
The horizontal controls are used to change the position of the cursor, set the time base settings, zoom into the waveforms/traces and search for events.
Horizontal Position
POSITION
PUSH TO
ZERO
The Position knob is used to position the waveforms/traces horizontally on the display screen. Pressing the knob will reset the position to zero.
SCALE
SCALE
The Scale knob is used to change the horizontal scale (TIME/DIV).
17
MDO-2000EG & 2000EX User Manual
Zoom
Zoom
Press Zoom in combination with the horizontal Position knob.
Play/Pause
The Play/Pause key allows you to view each search event in succession – to effectively “play” through each search event. It is also used to play through a waveform/trace in zoom mode.
Search
Search
The Search key accesses the search function menu to set the search type, source and threshold.
Search Arrows
Use the arrow keys to navigate the search events.
Set/Clear
Set/Clear
Use the Set/Clear key to set or clear points of interest when using the search function.
Trigger Controls
The trigger controls are used to control the trigger level and options.
Level Knob
LEVEL
Used to set the trigger level. Pressing the knob will reset the level to zero.
Trigger Menu Key
Menu
Used to bring up the trigger menu.
50% Key
50 %
Sets the trigger level to the half way point (50%).
Force - Trig
Force-Trig
Press to force an immediate trigger of the waveform.
18
GETTING STARTED
Vertical POSITION
POSITION
PUSH TO
ZERO
Sets the vertical position of the waveform/trace. Push the knob to reset the vertical position to zero.
Channel Menu Key
CH1
Press the CH1~4 key to set and configure the corresponding channel.
(Vertical)SCALE Knob
SCALE
Sets the vertical scale of the channel (TIME/DIV).
External Trigger Input
EXT TRIG
Accepts external trigger signals (page 140). Only on 2 channel models.
Input impedance: 1MΩ Voltage input: ±15V (peak), EXT trigger capacitance: 16pF.
Math Key
MATH
M
Use the Math key to set and configure math functions.
Reference Key
REF
R
Press the Reference key to set or remove reference waveforms.
BUS Key
BUS
B
The Bus key is used for serial bus (UART, I2C, SPI, CAN, LIN) configuration.
19
MDO-2000EG & 2000EX User Manual
Channel Inputs
CH1
Accepts input signals. Input impedance: 1MΩ. Capacitance: 16pF CAT I
USB Host Port
2V
Type A, 1.1/2.0 compatible. Used for data transfer.
Ground Terminal
2V
Accepts the DUT ground lead for common ground.
Probe Compensation Output
2V
The probe compensation output is used for probe compensation. It also has an adjustable output frequency.
By default this port outputs a 2Vpp, square wave signal at 1kHz for probe compensation.
Please see page 184 for details.
Power Switch
Used to turn the power on/off.
: ON
: OFF
DMM Ports
(For MDO­2000EX only)
10A MAX
FUSED
V
W
COMA mA
CAT II 600V CAT III 300VFUSED
600mA MAX
20
GETTING STARTED
mA
Accepts up to 600mA. Fuse protection: 1A.
A
Accepts up to 10A. Fuse protection: 10A
COM
Com port
V
W
Voltage, Resistance and Diode port.
Max voltage: 600V
21
MDO-2000EG & 2000EX User Manual
Calibration output
Key lock Power input socket
Fan
Go/No Go output
LAN portUSB Device port
AWG channels output
Calibration output
Key lock Power input socket
Fan
Go/No Go output
LAN portUSB Device port
AWG channels output
Power supply output
Calibration Output
Outputs the signal for vertical scale accuracy calibration (page 327).
MDO-2000EG Rear Panel
MDO-2000EX Rear Panel
22
GETTING STARTED
USB Device Port
The USB Device port is used for remote control.
LAN (Ethernet) Port
The LAN port is used for remote control over a network or when combined with the Remote Disk app, allows the scope to be mounted to a share disk.
Power Input Socket
Power cord socket accepts AC mains, 100 ~ 240V, 50/60Hz.
For power up sequence, see page
28.
Security Slot
Kensington security slot compatible.
Go-No Go Output
Outputs Go-No Go test results (page 256) as a 500us pulse signal.
AWG Output
GEN 1
Output the GEN1 or GEN2 signal from the Arbitrary Wave Generator function. (see page 188).
Power Supply outputs
5V/1A Dual power supply outputs, (For MDO-2000EX only, see page
249).
23
MDO-2000EG & 2000EX User Manual
Memory bar
Reference
waveform
Analog
Waveform
Channel status Horizontal status
Trigger configuration
Waveform
frequency
Date and time
Trigger position
Trigger status
Acquisition mode
Trigger level
Channel Indicators
Memory length
and sample rate
Analog Waveforms
Shows the analog input signal waveforms.
Channel 1: Yellow
Channel 2: Blue
Channel 3: Pink
Channel 4: Green
Bus decoding
Shows serial bus data decoding. The values are displayed in hex or binary.
Reference waveform
Reference waveform(s) can be displayed for reference, comparison or other operations.
Channel Indicators
The channel indicators for each activated channels are located at the zero volt level of each signals. Any active channel is shown with a solid color.
Example:
Analog channel indicator
Reference waveform indicator
Math indicator
Display
Below is a general description of the main display. As the display changes while activating the different functions of the MDO­2000EG/2000EX, please refer to each function sub-chapters of this user manual for more details.
24
GETTING STARTED
Trigger Position
Shows the position of the trigger.
Horizontal Status
Shows the horizontal scale and position.
Date and Time
Current date and time (page 183).
Trigger Level
Shows the trigger level on the graticule.
Memory Bar
The ratio and the position of the displayed waveform compared to the internal memory (page
102).
Trigger Status
Trig’d
Triggered.
PrTrig
Pre-trigger.
Trig?
Not triggered, display not updated.
Stop
Trigger stopped. Also appears in Run/Stop (page 42).
Roll
Roll mode.
Auto
Auto trigger mode.
For trigger details, see page 140.
Acquisition Mode
Normal mode
Peak detect mode
Average mode
For acquisition details, see page 80.
Signal Frequency
Shows the trigger source frequency.
Indicates the frequency is less than 2Hz (lower frequency limit).
Trigger Configuration
Trigger source, slope, voltage and coupling.
Horizontal Status
Horizontal scale, horizontal position.
25
MDO-2000EG & 2000EX User Manual
For trigger details, see page 140.
Channel Status
Channel 1, DC coupling, 2V/Div.
For channel details, see page 110.
26
Set Up
Tilt
To tilt, pull the legs forward, as shown below.
Stand
To stand the scope upright, push the legs back under the casing as shown below.
Tilt Stand
GETTING STARTED
27
Power Up
Requirements
The MDO-2000EG/2000EX accepts line voltages of 100 ~ 240V at 50 or 60Hz.
Step
1. Connect the power cord to
the rear panel socket.
2. Press the POWER key. The
display becomes active in ~ 30 seconds.
: ON
: OFF
Note
The MDO-2000EG/2000EX recovers the state right before the power is turned OFF. The default settings can be recovered by pressing the Default key on the front panel. For details, see page 296.
MDO-2000EG & 2000EX User Manual
28
First Time Use
Background
This section describes how to connect, adjust the scale and compensate the probe. Before operating the MDO-2000EG/2000EX in a new environment, run these steps to make sure the instrument performs at its full potential.
1. Power On
Follow the procedures on the previous page.
2. Firmware
Update to the latest firmware.
Page 334
3. Set the Date and Time
Set the date and time.
Page 183.
4. Reset System
Reset the system by recalling the factory settings. Press the Default key on the front panel. For details, see page 296.
Default
5. Connect the probe
Connect the probe that you will use for measurements to the Channel 1 input and to the probe compensation output. This output provides by default a 2V peak to peak, 1kHz square wave for signal compensation.
Set the probe attenuation to x10 if the probe has adjustable attenuation.
GETTING STARTED
29
MDO-2000EG & 2000EX User Manual
x1
x10
X 10
X 1
CH1
2V
6. Capture Signal (Autoset)
Press the Autoset key. A square waveform appears on the center of the screen. For Autoset details, see page 40.
Autoset
7. Select Vector Waveform
Press the Display key, and set the display to Vector on the bottom menu.
Display
30
GETTING STARTED
8. Compensate the probe
Turn the adjustment point on the probe to make the square waveform as flat as possible.
9. Start operations
Continue with the other operations.
Measurement: page 38
Configuration: page 77
Using the Spectrum Analyzer: page 220.
Using the Power Supply: page 249.
Using the DMM: page
239.
Applications: page 220 Save/Recall: page 277
File Utilities: page 303
Hardcopy key: page 310
Remote Control: page 314
Maintenance: page 325
Under
Compensation
Normal
Over
Compensation
31
MDO-2000EG & 2000EX User Manual
Background
This section describes the conventions used in this manual to operate the MDO-2000EG/2000EX.
Throughout the manual any reference to pressing a menu key refers to the keys directly below or beside any menu icons or parameters.
When the user manual says to “toggle” a value or
parameter, press the corresponding menu item. Pressing the item will toggle the value or parameter.
Active parameters are highlighted for each menu item. For example in the example below, Coupling is currently set to DC.
If a menu item can be toggled from one value or parameter to another, the available options will be visible, with the current option highlighted. In the example below the slope can be toggled from a rising slope to a falling slope or either slope.
Menu item
Parameter
Menu item
Active
parameter
Optional
parameters
Menu item
Selecting a Menu Item, Parameter or Variable
When the user manual says to “select” a value
from one of the side menu parameters, first press the corresponding menu key and use the Variable knob to either scroll through a parameter list or to increase or decrease a variable.
How to Use This Manual
32
GETTING STARTED
Example 1
1
2
3
3. Press a bottom menu key to access
the side menu.
4. Press a side menu key to either set
a parameter or to access a sub menu.
5. If accessing a sub menu or setting
a variable parameter, use the Variable knob to scroll through menu items or variables. Use the Select key to confirm and exit. The Select key is lit-up when such selection can be made.
Select
VARIABLE
6. Press the same bottom menu key
again to reduce the side menu.
Example 2
For some variables, a circular arrow icon indicates that the variable for that menu key can be edited with the Variable knob.
33
MDO-2000EG & 2000EX User Manual
1
7. Press the desired menu key to select it. The
circular arrow will become highlighted.
8. Use the Variable knob to edit the value.
Toggling a Menu Parameter
1
9. Press the bottom menu key
to toggle the parameter.
Reduce Side Menu
1
34
GETTING STARTED
10. To reduce the side menu, press the
corresponding bottom menu that brought up the side menu.
For example: Press the Source soft-key to reduce the Source menu.
Reduce Lower Menu
11. Press the relevant function
key again to reduce the bottom menu. For example: press the Trigger Menu key to reduce the trigger menu.
POSITION
SCALE
TRIGGER
LEVEL
ZERO
PUSH TO
ZERO
PUSH TO ZERO
SCALE
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Zoom
Menu
50 %
Force-Trig
1
Remove All Menus
35
MDO-2000EG & 2000EX User Manual
12. Press the Menu Off key to
reduce the side menu, press again to reduce the bottom menu.
VARIABLE
POSITION
HORIZONTAL
POSITION
POSITION
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALE
TRIGGER
LEVEL
PUSH TO
ZERO
SCALESCALESCALE
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
SCALE
CH1 CH2 CH3 CH4
2V
Measure
Display
Cursor
Help
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Select
Search
Set/Clear
Zoom
CH1 CH2 CH3 CH4
M
R
B
Menu
50 %
Force-Trig
1
Remove On­Screen Messages
13. The Menu Off key can also be
used to remove any on screen messages.
VARIABLE
POSITION
HORIZONTAL
POSITION
POSITION
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALE
TRIGGER
LEVEL
PUSH TO
ZERO
SCALESCALESCALE
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
SCALE
CH1 CH2 CH3 CH4
2V
Measure
Display
Cursor
Help
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Select
Search
Set/Clear
Zoom
CH1 CH2 CH3 CH4
M
R
B
Menu
50 %
Force-Trig
1
36
GETTING STARTED
The Help key accesses a context sensitive help menu. The help menu contains information on how to use the front panel keys.
Panel Operation
1. Press the Help key. The
display changes to Help mode.
Help
2. Use the Variable knob to scroll up and down
through the Help contents. Press Select to view the help on the selected item.
Example: Help on the Display key
Home Key
Press the Home key to return to the main help screen.
Go Back
Press the Back key to go to the previous menu page.
Exit
Press the Help key again or press the Exit key to exit the Help mode.
Help
Built-in Help
37
MDO-2000EG & 2000EX User Manual
Basic Measurement ................................................. 39
Channel Activation .......................................................................... 39
Autoset ............................................................................................... 40
Run/Stop ........................................................................................... 42
Horizontal Position/Scale .............................................................. 43
Vertical Position/Scale .................................................................... 45
Automatic Measurement .......................................... 46
Measurement Items ......................................................................... 46
Add Measurement ............................................................................ 50
Remove Measurement ..................................................................... 52
Gated mode ....................................................................................... 53
Display All mode .............................................................................. 54
High Low Function ......................................................................... 55
Statistics ............................................................................................. 56
Reference Levels .............................................................................. 58
Cursor Measurement ............................................... 60
Use Horizontal Cursors .................................................................. 60
Use Vertical Cursors ........................................................................ 63
Math Operation ....................................................... 67
Basic Math Overview & Operators .............................................. 67
Addition/Subtraction/Multiplication/Division ......................... 67
FFT Overview & Window Functions .......................................... 69
FFT Operation ................................................................................. 70
Advanced Math Overview .............................................................. 72
Advanced Math Operation ............................................................. 73
MEASUREMENT
38
MEASUREMENT
Activate Channel
To activate an input channel, press a channel key.
When activated, the channel key will light up. The corresponding channel menu will also appear.
CH1 CH1
Each channel is associated with the color shown beside each channel’s vertical SCALE dial: CH1: yellow, CH2: blue, CH3: pink and CH4: green.
When a channel is activated, it is shown above the bottom menu system.
CH1 CH3CH2 CH4
Basic Measurement
This section describes the basic operations required in capturing, viewing and measuring the input signal. For more detailed or more specific operations, see the following chapters.
Advanced Configuration → from page 77 Arbitrary Wave Generator → from page 186 Spectrum Analyzer→ from page 220 Digital Multi-meter (MDO-2000EX only)→ from page 239 Power Supply (MDO-2000EX only)→ from page 249 Applications→ from page 253
Before operating the oscilloscope, please see the Getting Started chapter, page 9.
Channel Activation
39
De-activate Channel
To de-activate a channel, press the corresponding channel key again. If the channel menu is not open, press the channel key twice (the first press shows the Channel menu).
CH1 CH1
Default Setup
To activate the default state, press Default (this will reset the system and recall the factory defaults, see page 296).
Default
Autoset
Background
The Autoset function automatically configures the panel settings to position the input signal(s) to the best viewing condition. The MDO-2000EG/2000EX automatically configures the following parameters:
Horizontal scale Vertical scale Trigger source channel
There are two operating modes for the Autoset function: Fit Screen Mode and AC Priority Mode.
Fit Screen Mode will fit the waveform to the best scale, including any DC components (offset). AC priority mode will scale the waveform to the screen by removing any DC component.
Panel Operation
1. Connect the input signal to the
MDO-2000EG/2000EX and press the Autoset key.
Autoset
2. The waveform appears in the center of the
display.
MDO-2000EG & 2000EX User Manual
40
MEASUREMENT
Before
After
3. To undo Autoset, press Undo
Autoset from the bottom menu.
Change modes
4. Choose between Fit Screen Mode
and AC Priority Mode from the bottom menu.
5. Press the Autoset key again to use
Autoset in the new mode.
Autoset
Fit Screen Mode
AC Priority
Limitation
Autoset does not work in the following situations:
Input signal frequency is less than 20Hz Input signal amplitude is less than 10mV
Note
The Autoset key does NOT automatically activate the channels to which input signals are connected.
41
Run/Stop
Background
By default, the waveform on the display is constantly updated (Run mode). Freezing the waveform by stopping signal acquisition (Stop mode) allows flexible observation and analysis. To enter Stop mode, two methods are available: pressing the Run/Stop key or using the Single Trigger mode.
Stop mode icon
When in Stop mode, the Stop icon appears at the top of the display.
Triggered icon
Freeze Waveform using the Run/Stop Key
Press the Run/Stop key once. The Run/Stop key turns red. The waveform and signal acquisition freezes.
Stop:
Run/Stop Run/Stop
To unfreeze, press the Run/Stop key again. The Run/Stop key turns green again.
Run:
Run/Stop Run/Stop
Freeze Waveform by Single Trigger Mode
Press the Single key to go into the Single Trigger mode. The Single key turns bright white.
In the Single Trigger mode, the scope will be put into the pre­trigger mode until the scope encounters the next trigger point. After the scope has triggered, it will remain in Stop mode, until the Single key is pressed again or the Run/Stop key is pressed.
SingleSingle
Waveform Operation
The waveform can be moved or scaled in both Run and Stop mode, but in different manners. For details, see page 102 (Horizontal position/scale) and page 110 (Vertical position/scale).
MDO-2000EG & 2000EX User Manual
42
MEASUREMENT
Set Horizontal Position
The horizontal position knob moves the waveform left and right.
POSITION
PUSH TO
ZERO
Set Horizontal Position to 0
Pressing the horizontal position knob will reset the horizontal position to 0.
POSITION
PUSH TO
ZERO
Alternatively, pressing the Acquire key and then pressing Reset H Position to 0s from the bottom menu will also reset the horizontal position.
Acquire
As the waveform moves, the display bar on the top of the display indicates the portion of the waveform currently shown on the display and the position of the horizontal marker on the waveform.
Position Indicator
The horizontal position is shown at the bottom of the display grid to the right of the H icon.
Horizontal Position/Scale
For more detailed configuration, see page 102.
43
MDO-2000EG & 2000EX User Manual
Select Horizontal Scale
To select the timebase, turn the horizontal SCALE knob; left (slow) or right (fast).
SCALE
Range
1ns/div ~ 100s/div, 1-2-5 increments
The scale is displayed to the left of the H icon at the bottom of the screen.
Display bar
The display bar indicates how much of the waveform is displayed on the screen at any given time. Changes to timebase will be reflected on the display bar.
Fast
Medium
Slow
Stop mode
In the Stop mode, the waveform size changes according to the scale.
Note
The Sample rate changes according to the timebase and record length. See page 84.
44
MEASUREMENT
Set Vertical Position
To move the waveform up or down, turn the vertical position knob for each channel.
POSITION
PUSH TO
ZERO
Push the vertical position knob to reset the position to 0.
As the waveform moves, the vertical position of the channel indicator appears on the display.
POSITION
PUSH TO
ZERO
Run/Stop mode
The waveform can be moved vertically in both Run and Stop mode.
Select Vertical Scale
To change the vertical scale, turn the vertical SCALE knob; left (down) or right (up).
SCALE
Range
1mV/div ~ 10V/div
1-2-5 increments
The vertical scale indicator for each channel on the bottom of the display changes accordingly.
Vertical Position/Scale
For more detailed configuration, see page 110.
45
MDO-2000EG & 2000EX User Manual
V/I Measurements
Time Meas.
Delay Meas.
Overview
FPREShoot
RPREShoot
FOVShoot
ROVShoot
Cycle Area
Area
Cycle RMS
RMS
Cycle Mean
Mean
Low
High
Amplitude
Min
Max
Pk-Pk
Frequency
Dutycycle
-Width
+Width
FallTime
RiseTime
Period
-Edges
+Edges
-Pulses
+Pulses
% Flicker
Flicker Idx
A
B
(A-B)
(A+B)
A1
A1
A1+A2
A2
Voltage/Current Measurement
Pk-Pk
(peak to peak)
Difference between positive and negative peak. (=max − min)
Max
Positive peak.
Min
Negative peak.
FRR
FRF
FFR
Phase
LFF
LFR
LRF
LRR
FFF
Automatic Measurement
The automatic measurement function measures and updates major items for Voltage/Current, Time, and Delay type measurements.
Measurement Items
46
MEASUREMENT
Amplitude
Difference between the global high value and the global low value, measured over the entire waveform or gated region. (=high − low)
High
Global high voltage. See page 55 for details.
Low
Global low voltage. See page 55 for details.
Mean
The arithmetic mean value is calculated for all data samples as specified by the Gating option.
Cycle Mean
The arithmetic mean value is calculated for all data samples within the first cycle found in the gated region.
RMS
The root mean square of all data samples specified by the Gating option.
Cycle RMS
The root mean square value is calculated for all data samples within the first cycle found in the gated region.
Area
Measures the positive area of the waveform and subtracts it from the negative area. The ground level determines the division between positive and negative areas.
Cycle Area
The Summation based on all data samples within the first cycle found in the gated region.
ROVShoot
Rise overshoot
47
MDO-2000EG & 2000EX User Manual
FOVShoot
Fall overshoot
RPREShoot
Rise preshoot
FPREShoot
Fall preshoot
Time Measurement
Frequency
Frequency of the waveform.
Period
Waveform cycle time. (=1/Freq)
RiseTime
The time required for the leading edge of the first pulse to rise from the low reference value to the high reference value.
FallTime
The time required for the falling edge of the first pulse to fall from the high reference value to the low reference value.
+Width
Positive pulse width.
Width
Negative pulse width.
Duty Cycle
Ratio of signal pulse compared with whole cycle. =100x (Pulse Width/Cycle)
+Pulses
Measures the number of positive pulses.
-Pulses
Measures the number of negative pulses.
+Edges
Measures the number of positive edges.
48
MEASUREMENT
-Edges
Measures the number of negative edges.
% Flicker
A
B
(A-B)
(A+B)
Ratio in percentage of the peak-to-peak value to the sum of peak values.
Flicker Idx
A1
A1
A1+A2
A2
Ratio of the area above the average to the total area during one cycle.
Delay Measurement
FRR
Time between:
Source 1 first rising edge and Source 2 first rising edge.
FRF
Time between: Source 1 first rising edge and Source 2 first falling edge.
FFR
Time between: Source 1 first falling edge and Source 2 first rising edge.
FFF
Time between: Source 1 first falling edge and Source 2 first falling edge.
LRR
Time between: Source 1 first rising edge and Source 2 last rising edge.
LFR
Time between: Source 1 first rising edge and Source 2 last falling edge.
LRF
Time between: Source 1 first falling edge and Source 2 last rising edge.
LFF
Time between: Source 1 first falling edge and Source 2 last falling edge.
49
MDO-2000EG & 2000EX User Manual
Phase
The phase difference of two signals, calculated in degrees.
360x
t2
t1
Note
The in-built help system can be used to see detailed automatic measurement definitions.
Add Measurement Item
1. Press the Measure key.
Measure
2. Press Add Measurement from the
bottom menu.
3. Choose either a V/I, Time or Delay
measurement from the side menu and choose the type of measurement you wish to add.
V/I (Voltage/ Current)
Pk-Pk, Max, Min, Amplitude, High, Low, Mean, Cycle Mean, RMS, Cycle RMS, Area, Cycle Area, ROVShoot, FOVShoot, RPREShoot, FPREShoot
Time
Frequency, Period, RiseTime, FallTime, +Width, –Width, Duty Cycle, +Pulses, -Pulses, +Edges, ­Edges, %Flicker, FlickerIndex
Add Measurement
The Add Measurement function allows you to add up to eight automatic measurement items on the bottom of the screen from any channel source.
50
MEASUREMENT
Delay
FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF, Phase
4. All of the chosen automatic measurements will
be displayed in a window on the bottom of the screen. The channel number and channel color indicate the measurement source. For the analog inputs: yellow = CH1, blue = CH2, pink = CH3, green = CH4.
Choose a Source
The channel source for measurement items can be set either before or when selecting a measurement item.
5. To set the source, press either the
Source1 or Source2 key from the side menu and choose the source.
Range
Source1: CH1~CH4, Math
Source2: CH1~CH4, Math
(Note)
Source 2 is only applicable to Delay measurements.
Only selected Time Frequency measurements can be applied to Frequency, Period, +Width, ­Width and Dutycycle.
51
MDO-2000EG & 2000EX User Manual
Remove Measurement Item
1. Press the Measure key.
Measure
2. Press Remove Measurement from the
bottom menu.
3. Press Select Measurement and select
the item that you want to remove from the measurement list.
Remove All Items
Press Remove All to remove all the measurement items.
Remove Measurement
Individual measurements can be removed at any time using the Remove Measurement function.
52
MEASUREMENT
Set Gating Mode
1. Press the Measure key.
Measure
2. Press Gating from the bottom
menu.
3. Choose one of the gating modes
from the side menu: Off (full record), Screen, Between Cursors.
Cursors On Screen
If Between Cursors is selected, the cursor positions can be edited by using the cursor menu.
Page 60
Gated mode
Some automatic measurements can be limited to a “gated” area
between cursors. Gating is useful for measuring a magnified waveform or when using a fast time base. The Gated mode has three possible configurations: Off (Full Record), Screen and Between Cursors.
53
MDO-2000EG & 2000EX User Manual
View Measurement Results
1. Press the Measure key.
Measure
2. Press Display All from the bottom
menu.
3. Press Source from the side menu
and choose a measurement source.
Range
CH1~CH4, Math
4. The results of Voltage and Time type
measurements appear on the display.
Remove Measurements
To remove the measurement results, press OFF.
Delay Measurements
Delay type measurements are not available in this mode as only one channel is used as the source. Use the individual measurement mode (page 50) instead.
Display All mode
Display All mode shows and updates all items from Voltage and Time type measurements.
54
MEASUREMENT
Background
The High-Low function is used to select the method for determining the value of the High-Low measurement values.
Auto
Automatically chooses the best high-low setting for each waveform when measuring.
Histogram
Uses histograms to determine the high-low values. This mode ignores any pre-shoot or overshoot values. This mode is particularly useful for pulse-type waveforms
low
high
Min-max
Sets the high-low values as the minimum or maximum measured values.
low
high
Set High-Low
1. Press the Measure key.
Measure
2. Press High-Low from the bottom
menu.
High Low Function
55
3. Select the type of High-Low settings from the
side menu.
High-Low Settings:
Histogram, Min-Max, Auto
Restore Default High-Low Settings
To return to the default High-Low settings, press Set to Defaults.
Statistics
Background
The Statistics function can be used to view a number of statistics for the selected automatic measurements. The following information is displayed with the Statistics function:
Value
Currently measured value
Mean
The mean value is calculated from a number of automatic measurement results. The number of samples used to determine the mean can be user-defined.
Min
The minimum value observed from a series of measured results for the selected automatic measurement items.
MDO-2000EG & 2000EX User Manual
56
MEASUREMENT
Max
The maximum value observed from a series of measured results for the selected automatic measurement items.
Standard Deviation
The variance of the currently measured value from the mean. The standard deviation equals the squared root of the variance value. Measuring the standard deviation can, for example, determine the severity of jitter in a signal. The number of samples used to determine the standard deviation can be user-defined.
Panel Operation
1. Press the Measure key.
Measure
2. Select at least one automatic
measurement.
Page 50
3. Press Statistics from the bottom
menu.
4. Set the number of samples to be
used in the mean and standard deviation calculations.
Samples:
2~1000
5. Press Statistics and turn Statistics
on.
6. The statistics for each automatic measurement
will appear at the bottom of the display in a table.
57
MDO-2000EG & 2000EX User Manual
Reset Statistics
To reset the standard deviation calculations, press Reset Statistics.
Background
The reference level settings determine the measurement threshold levels for some measurements like the Rise Time measurement.
High Ref: Sets the high reference level.
Mid Ref: Sets the middle reference for the first and second waveforms.
Low Ref: Sets the low reference level.
Panel Operation
1. Press the Measure key.
Measure
2. Press Reference Levels from the
bottom menu.
Reference Levels
58
MEASUREMENT
3. Set the reference levels from the side menu.
Ensure the reference levels do not cross over.
High Ref
0.0% ~ 100%
Mid Ref
0.0% ~ 100%
0.0% ~ 100%
Low Ref
0.0% ~ 100%
Default Settings
4. Press Set to Defaults to set the
reference levels back to the default settings.
59
MDO-2000EG & 2000EX User Manual
Panel Operation
1. Press the Cursor key once.
Cursor
2. Press H Cursor from the bottom
menu if it is not already selected.
3. When the H Cursor is selected,
repeatedly pressing the H Cursor key or the Select key will toggle which cursor is selected.
OR
Select
Range
Description
Left cursor ( ) movable, right cursor position fixed
Right cursor ( ) movable, left cursor position fixed
Left and right cursor ( + ) movable together
4. The cursor position
information appears on the top left hand side of the screen
Cursor
Hor. position, Voltage/Current
Cursor
Hor. position, Voltage/Current
Delta (difference between cursors)
dV/dt or dI/dt
Cursor Measurement
Horizontal or vertical cursors are used to show the position and values of waveform measurements and math operation results. These results cover voltage, time, frequency and other math operations. When the cursors (horizontal, vertical or both) are activated, they will be shown on the main display unless turned off.
Use Horizontal Cursors
60
MEASUREMENT
5. Use the Variable knob to move the
movable cursor(s) left or right.
VARIABLE
(Note)
The selected cursor(s) will move along the active waveform. To move along another waveform, select its corresponding channel and press the cursor key again to re-enter the cursor menu.
Select Units
6. To change the units of the
horizontal position, press H Unit.
Units
S, Hz, % (ratio), ˚(phase)
Phase or Ratio Reference
7. To set the 0% and 100% ratio or the
0˚ and 360˚ phase references for the current cursor positions, press Set
Cursor Positions As 100%.
Example
Horizontal
cursors
61
MDO-2000EG & 2000EX User Manual
FFT
FFT cursors can use different units. For FFT details, see page
69.
Cursor
Hor. position, dB/Voltage
Cursor
Hor. Position, dB/Voltage
Delta (difference between cursors)
dV/dt or d/dt
Example
Horizontal
cursors
XY Mode
XY mode cursors measure a number of X by Y measurements. See page 82.
Cursor
Time, rectangular, polar co­ordinates, product, ratio.
Cursor
Time, rectangular, polar co­ordinates, product, ratio.
Delta (difference between cursors)
62
MEASUREMENT
Example
Horizontal
cursors
Panel Operation/ Range
1. Press the Cursor key twice.
Cursor
x2
2. Press V Cursor from the bottom
menu if it is not already selected.
3. When the V Cursor is selected,
repeatedly pressing the V Cursor key or the Select key will toggle which vertical cursor is selected.
OR
Select
Range
Upper cursor movable, lower cursor position fixed
Lower cursor movable, upper cursor position fixed
Upper and lower cursor movable together
Use Vertical Cursors
63
MDO-2000EG & 2000EX User Manual
4. The cursor position
information appears on the top left hand side of the screen.
,
Time: cursor 1, cursor 2
,
Voltage/Current: cursor1, cursor2
Delta (difference between cursors)
dV/dt or dI/dt
5. Use the Variable knob to move the
cursor(s) up or down.
VARIABLE
Select Units
6. To change the units of the vertical
position, press V Unit.
Units
Base (source wave units), % (ratio)
Base or Ratio Reference
7. To set the 0% and 100% ratio
references for the current cursor position, press Set Cursor Positions As 100%.
Example
Horizontal
cursors
Vertical cursors
64
MEASUREMENT
FFT
FFT has different content. For FFT details, see page 69.
,
Frequency/Time: cursor1, cursor2
,
dB/V: cursor1, cursor2
Delta (difference between cursors)
d/dt
Example
Horizontal
cursors
Vertical
cursors
XY Mode
XY mode cursors measure a number of X by Y measurements. See page 82.
Cursor
Rectangular, polar co-ordinates, product, ratio.
Cursor
Rectangular, polar co-ordinates, product, ratio.
Delta (difference between cursors)
65
MDO-2000EG & 2000EX User Manual
Example
Horizontal
cursors
Vertical cursors
66
MEASUREMENT
Background
The Math function performs basic math functions (addition, subtraction, multiplication, division) on the input signals or the reference waveforms. The resultant waveform will be shown on the screen in real-time.
Addition (+)
Adds the amplitude of two signals.
Source
CH1~4, Ref1~4
Subtraction (–)
Extracts the amplitude difference between two signals.
Source
CH1~4, Ref1~4
Multiplication (×)
Multiplies the amplitude of two signals. Source
CH1~4, Ref1~4
Division (÷)
Divides the amplitude of two signals. Source
CH1~4, Ref1~4
Panel Operation
1. Press the Math key.
MATH
M
2. Press the Math key on the lower
bezel.
3. Select Source 1 from the side menu
Range
CH1~4, Ref1~4
Math Operation
Basic Math Overview & Operators
Addition/Subtraction/Multiplication/Division
67
MDO-2000EG & 2000EX User Manual
4. Press Operator to choose the math
operation.
Range
+, -, ×, ÷
5. Select Source 2 from the side menu.
Range
CH1~4, Ref1~4
6. The math measurement result appears on the
display. The vertical scale of the math waveform appears at the bottom of the screen.
From left: Math function, source1, operator, source2, Unit/div
Example
Math
Source 2
Source 1
Position and Unit
To move the math waveform vertically, press the Position key from the side menu and use the Variable knob to set the position.
Range
12.00 Div ~ +12.00 Div
68
MEASUREMENT
To change the unit/div settings, press Unit/div, then use the Variable knob to change the unit/div.
The units that are displayed depend on which operator has been selected, and whether the probe for the selected channel has been set to voltage or current.
Operator:
Unit/div:
Multiplication Division Addition/Subtraction
VV, AA or W V/V, A/A V or A
Turn Off Math
To turn off the Math result from the display, press the Math key again.
MATH
M
Background
The FFT function performs a Fast Fourier Transform on one of the input signals or the reference waveforms. The resultant spectrum will be shown on the screen in real-time. Four types of window function are available: Hanning, Hamming, Rectangular, and Blackman, as described below.
Hanning
Frequency resolution
Good
Amplitude resolution
Not good
Suitable for....
Frequency measurement on periodic waveforms
FFT Overview & Window Functions
69
MDO-2000EG & 2000EX User Manual
Hamming
Frequency resolution
Good
Amplitude resolution
Not good
Suitable for....
Frequency measurement on periodic waveforms
Rectangular
Frequency resolution
Very good
Amplitude resolution
Bad
Suitable for....
Single-shot phenomenon (this mode is the same as having no window at all)
Blackman
Frequency resolution
Bad
Amplitude resolution
Very good
Suitable for....
Amplitude measurement on periodic waveforms
(Note)
For more complete measures and functions in the frequency domain of a signal, please also refer to the Spectrum Analyzer option of the MDO-2000EG on page 220.
Panel Operation
1. Press the Math key.
MATH
M
2. Press FFT from the bottom menu.
3. Select the Source from the side
menu.
Range
CH1~4, Ref~4
FFT Operation
70
MEASUREMENT
4. Press the Vertical Units key from
the side menu to select the vertical units used.
Range
Linear RMS, dBV RMS
5. Press the Window key from the side
menu and select the window type.
Range
Hanning, Hamming, Rectangular, and Blackman.
6. The FFT result represents the frequency-
domain representation of a signal. Hence, the horizontal scale changes from time to frequency, and the vertical scale from voltage/current to dB/RMS.
Math
Source
Position and Scale
To move the FFT waveform vertically, press Vertical until the Div parameter is highlighted and then use the Variable knob.
Range
–12.00 Div ~ +12.00 Div
To select the vertical scale of the FFT waveform, press Vertical until the dB or voltage parameters are highlighted and then use the Variable knob.
71
MDO-2000EG & 2000EX User Manual
Range
2mV~1kV RMS (Linear RMS), 1~20 dB (dB VRMS)
Horizontal Position and Scale
To move the FFT waveform horizontally, press Horizontal until the Frequency parameter is highlighted and then use the Variable knob.
Range
0Hz ~ half of the sampling frequency
To select the horizontal scale of the FFT waveform, press Horizontal repeatedly until the Hz/div parameter is highlighted and then use the Variable knob.
Background
The advanced math function allows complex math expressions to be created based on the input sources, reference waveforms or even the automatic measurements available from the Measure menu (see page 46).
An overview of each of the major parameters that can be used in the advanced math function are shown below:
Expression
Displays the function expression as it is created.
Source
Selects the source signal.
Source
CH1~4, Ref1~4
Function
Adds a mathematical function to the expression.
Function
Intg, Diff, log, Ln, Exp, Sqrt, Abs, Rad, Deg, Sin, Cos, Tan, Asin, Acos, Atan
Advanced Math Overview
72
MEASUREMENT
Variable
Adds a user-specified variable to the expression.
Source
CH1~4, Ref1~4
Operator
Adds an operator or parenthesis to the function expression.
Operator
+, -, *, /, (, ), !(, <, >, <=, >=, ==, !=, ||, &&
Figure
Adds a value to the expression.
Figure
Integers, floating point, or floating point with exponent values.
Measurement
Adds automatic measurements to the expression. Not all automatic measurements are supported.
Measurement
Pk-Pk, Max, Min, Amp, High, Low, Mean, CycleMean, RMS, CycleRMS, Area, CycleArea, ROVShoot, FOVShoot, Freq, Period, Rise, Fall, PosWidth, NegWidth, Dutycycle, FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF, Phase, RPRFShoot, FPREShoot, +Pulses, -Pulses, +Edges, -Edges
Panel Operation
1. Press the Math key.
M
MATH
2. Press Advanced Math from the
bottom menu.
3. Press Edit Expression.
Advanced Math Operation
73
MDO-2000EG & 2000EX User Manual
4. The Edit f(x) screen appears. CH1 + CH2 is
shown in the expression box as an example at startup.
5. Press Clear to clear the expression
entry area.
6. Use the Variable knob and the Select
key to create an expression.
Use the Variable knob to highlight a source, function, variable, operator, figure or measurement in orange.
Press the Select key to make the selection.
If a particular parameter is grayed out, it indicates that the particular parameter is not available at that time.
Select
VARIABLE
Back Space
7. To delete the last parameter press
Back Space.
74
MEASUREMENT
8. When the expression is complete,
press OK Accept.
Example: CH1 + CH2
Advanced Math
Source1
Source2
Set the VAR1 & VAR2
9. Press VAR1 or VAR2 to set
VAR1/VAR2 if they were used in the expression created previously.
10. Press Mantissa.
Use the Left and Right arrow keys to select a digit and use the variable knob to set the value of the selected digit.
11. Press Exponent.
Use the Variable knob to set the exponent of the variable.
12. Press Go Back to finish editing
VAR1 or VAR2.
Vertical Position and Scale
13. Press Unit/div and use the Variable
knob to set the vertical scale of the math waveform.
75
MDO-2000EG & 2000EX User Manual
14. Press Position and use the Variable
knob to set the vertical position of the math waveform on the display.
Clear Advanced Math
To clear the advanced math result from the display, press the Math key again.
MATH
M
76
ADVANCED CONFIGURATION
Acquisition .............................................................. 80
Select Acquisition Mode ................................................................. 80
Show Waveform in XY Mode ....................................................... 82
Set the Record Length .................................................................... 84
Segmented Memory Acquisition .............................. 85
Segments Display ............................................................................. 87
Set the Number of Segments ......................................................... 87
Run Segmented Memory ................................................................ 88
Navigate Segmented Memory ........................................................ 90
Play Through Each Segment ......................................................... 91
Measurement on Segments ............................................................ 91
Segment Info .................................................................................... 96
Display .................................................................... 97
Display Waveform as Dots or Vectors ........................................ 97
Set the Level of Persistence ........................................................... 97
Set the Intensity Level ..................................................................... 98
Select Display Graticule .................................................................. 99
Freeze the Waveform (Run/Stop) .............................................. 100
Turn Off Menu .............................................................................. 101
Horizontal View ......................................................102
Move Waveform Position Horizontally ..................................... 102
Select Horizontal Scale .................................................................. 103
Select Waveform Update Mode .................................................. 104
Zoom Waveform Horizontally .................................................... 106
Play/Pause ....................................................................................... 108
Vertical View (Channel) ..........................................110
Move Waveform Position Vertically .......................................... 110
Select Vertical Scale ....................................................................... 111
Select Coupling Mode ................................................................... 111
Input Impedance ............................................................................ 112
Invert Waveform Vertically .......................................................... 112
ADVANCED
CONFIGURATION
77
MDO-2000EG & 2000EX User Manual
Limit Bandwidth............................................................................. 113
Expand by Ground/Center ......................................................... 114
Select Probe Type .......................................................................... 115
Select Probe Attenuation Level ................................................... 116
Set the Deskew ............................................................................... 116
Bus Key Configuration ........................................... 118
Bus Display ..................................................................................... 118
Serial Bus Overview ....................................................................... 120
UART Serial Bus Configuration .................................................. 122
I2C Serial Bus Interface ................................................................. 124
SPI Serial Bus Interface ................................................................ 125
CAN Serial Bus Interface ............................................................. 127
LIN Serial Bus Interface ............................................................... 129
Bus Encoding.................................................................................. 130
Threshold configuration ............................................................... 130
Serial Bus Event Tables ................................................................ 132
Event Tables Format ..................................................................... 135
Adding a Label to a Bus ................................................................ 136
Using Cursors with the Serial Bus ............................................... 138
Trigger ................................................................... 140
Trigger Type Overview ................................................................. 140
Trigger: types and sources ............................................................ 142
Trigger Parameter Overview ........................................................ 143
Setup Holdoff Level ...................................................................... 148
Setup Trigger Mode ....................................................................... 149
Using the Edge Trigger ................................................................. 149
Using Advanced Delay Trigger .................................................... 151
Using Pulse Width Trigger ........................................................... 152
Using Video Trigger ...................................................................... 154
Pulse Runt trigger ........................................................................... 155
Using Rise and Fall Trigger .......................................................... 157
Using the Timeout Trigger ........................................................... 158
Using the Bus Trigger.................................................................... 160
UART BUS Trigger Settings ........................................... 160
I2C Bus Trigger Settings .................................................. 161
SPI Bus Trigger Settings .................................................. 164
CAN Bus Trigger .............................................................. 166
LIN Bus Trigger ................................................................ 169
Search ................................................................... 172
Configuring Search Events ........................................................... 172
Copying Search Event To/From Trigger Events .................... 174
Search Event Navigation .............................................................. 174
Save Search Marks ......................................................................... 175
78
ADVANCED CONFIGURATION
Setting/Clearing Single Search Events ....................................... 176
FFT Peak ......................................................................................... 177
System Settings and Miscellaneous Settings ..........181
Select Menu Language................................................................... 181
View System Information ............................................................. 181
Erase Memory ................................................................................ 182
Set Date and Time ......................................................................... 183
Probe Compensation Frequency ................................................. 184
QR Code Reader Function........................................................... 184
79
MDO-2000EG & 2000EX User Manual
Background
The acquisition mode determines how the samples are used to reconstruct a waveform.
Sample
This is the default acquisition mode. Every sample from each acquisition is used.
Peak detect
Only the minimum and maximum value pairs for each acquisition interval (bucket) are used. This mode is useful for catching abnormal glitches in the signal.
Average
Multiple acquired data is averaged. This mode is useful for drawing a noise-free waveform. To select the average number, use the Variable knob.
Average number: 2, 4, 8, 16, 32, 64, 128, 256
Panel Operation
1. Press the Acquire key.
Acquire
2. To set the Acquisition mode, press
Mode on the bottom menu.
Acquisition
The Acquisition process samples the analog input signals and converts them into digital format for internal processing.
Select Acquisition Mode
80
ADVANCED CONFIGURATION
3. Select an acquisition mode from
the side menu.
4. If Average was chosen, set the
number of samples to be used for the average function.
Mode
Sample, Peak Detect, Average
Average sample
2, 4, 8, 16, 32, 64, 128, 256
Example
Sample
Peak Detect
Average (256 times)
81
MDO-2000EG & 2000EX User Manual
Background
The XY mode maps the input of channel 1 to the input of channel 2. In 4 channel models, the input of channel 3 can be mapped to the input of channel
4. This mode is useful for observing the phase
relationship between waveforms.
Connection
1. Connect the signals to
Channel 1 (X-axis) and Channel 2 (Y-axis) or Channel 3 (X2-axis) and Channel 4 (Y2-axis).
CH3CH2CH1 CH4
YX X2 Y2
2. Make sure a channel pair is
active (CH1&CH2 or CH3&CH4). Press the Channel key if necessary. A channel is active if the channel key is lit.
CH1 CH1
CH2 CH2
Panel Operation
3. Press the Acquire menu key.
Acquire
4. Press XY from the bottom menu.
5. Choose Triggered XY from the side
menu.
Show Waveform in XY Mode
82
ADVANCED CONFIGURATION
X-Y mode is split into two windows. The top window shows the signals over the full time range. The bottom window shows XY mode.
To move the XY waveform position, use the vertical position knob: Channel 1 knob moves the XY waveform horizontally and Channel 2 knob moves the XY waveform vertically. Similarly, the X2 and Y2 axis can be positioned using the channel 3 and channel 4 vertical position knobs.
POSITION
CH1
POSITION
CH2
The horizontal position knob and horizontal Scale knob can still be used under the XY mode.
Turn Off XY Mode
To turn off XY mode, choose OFF (YT) mode.
Cursors and XY Mode
Cursors can be used with XY mode. See the Cursor chapter for details.
Page 60
83
MDO-2000EG & 2000EX User Manual
Background
The number of samples that can be stored is set by the record length. Record length is important in an oscilloscope as it allows longer waveforms to be recorded.
The maximum record length for the MDO­2000EG/2000EX depends on operating mode. The table below describes the record lengths that are available for each mode.
Limitations
Record Length
Normal
Zoom
FFT
FFT in Zoom Window
1k
✓ ✗ ✓
10k
✓ ✓ ✓
100k
✓ ✓ ✓
1M
✓ ✓ ✓
10M
✓ ✓ ✗
Panel Operation
1. Press the Acquire key.
Acquire
2. Press the Record Length key on the
bottom menu and choose the record length.
Record length
1000, 10k, 100k, 1M, 10M points
Note
The sampling rate may also be changed when the record length is changed.
Set the Record Length
84
ADVANCED CONFIGURATION
Normal memory acquisition period Rearm time Next trigger
Segmented Memory Acquisition
The advanced segmented memory utility allows the scope memory to be divided into different segments. Each time the scope is triggered, it only acquires data for one segment of memory at a time. This allows you to optimize the scope memory to only perform signal acquisition during important signal events.
For example, for a signal with a number of pulses, normally the oscilloscope will acquire the signal until the acquisition memory of the scope is filled up and then it will re-arm the trigger and then capture again. This could result in a number of events not being captured or captured at a less-than-desired resolution (depending on the horizontal scale and sampling rate). However, the segmented memory function would effectively allow you to capture more of the signal than you would otherwise. The diagrams below illustrate this point.
Normal acquisition mode example:
85
MDO-2000EG & 2000EX User Manual
SegmentnSegment
n+1
Segment
n+2
Segment
n+3
Segment
n+4
Segment
n+5
Segmented memory acquisition example:
As shown above, the memory is divided into segments to increase the number of events that can be effectively captured with the same
acquisition memory. Also notice that the scope doesn’t need to
rearm the trigger between each segment, this makes the segmented memory function especially useful for high speed signals. The time between each segment is also recorded so that accurate signal timing can also be measured.
The segmented memory function also supports automatic measurements for each segment or statistics for all the captured segments.
86
ADVANCED CONFIGURATION
Run/Stop Indicator
Progress Indicator
Progress Indicator
Indicates the number of segments that have to been captured relative to the set number of segments.
Run/Stop Indicator
Stop: The segments have finished acquiring or have been stopped.
Run: The scope is ready to acquire segments.
Note
Before the Segment function can be used, set the trigger settings as appropriate for the signal you wish to use. The number of segments that can be used depends wholly on the record length. See page 84 to set the record length.
Record length
Number of segments
1000 pt.
1 ~ 29000
10k pt.
1 ~ 2900
100k pt.
1 ~ 290
1M pt.
1 ~ 20
Segments Display
Set the Number of Segments
87
MDO-2000EG & 2000EX User Manual
10M pt.
1 ~ 2
Panel Operation
1. Press the Acquire key.
Acquire
2. Press Segments on the bottom
menu.
3. Press Select Segments and set the
number of segments from the side menu.
Num of Seg
1~29000 (record length dependant)
Set to Maximum
Sets to the maximum number
Set to Minimum
Sets to 1 segment
Note
The Select Segments icon is only available when Segments = OFF or when Segments is in the STOP mode (see the section below).
Background
Before the Segmented Memory function can be used, set the trigger settings as appropriate for the signal you wish to use. See page 140 for configuring the trigger settings.
Run Segments
1. Toggle Segments On from the bottom menu.
Note
The first time Segmented memory is turned on the segments will automatically be run. Each segment will be automatically captured.
Run Segmented Memory
88
ADVANCED CONFIGURATION
2. The scope will automatically start acquiring
segments. The progress of the segmented memory capture is shown in the Progress Indicator.
3. The Run Indicator will be shown when in the
Run mode and the Segments icon will also indicate that the function is in run mode.
Run Indicator
Progress Indicator
Segment (Run )icon
4. When the scope has finished acquiring
segments, press Segments Run to toggle the mode to the Segments Stop mode.
Alternatively, the Run/Stop key can be pressed.
Run/Stop
5. The Stop Indicator will be shown when in the
Stop mode.
Stop Indicator
Progress Indicator
Segment (Stop )icon
89
MDO-2000EG & 2000EX User Manual
The scope is now ready to navigate or analyze the acquired segments.
Rerun Segmented Acquisition
6. To rerun the segments, press the Segments Stop
key to toggle the mode back to the Segments Run mode.
Alternatively, press the Run/Stop key again.
Run/Stop
7. Repeat steps 3 and 4 in the section above when
the segmented acquisition has completed.
Background
After the segmented memory acquisitions have been captured you can navigate through each segment one at a time.
Operation
1. Press Select Segments from the
bottom menu. This key will be available in the Stop mode.
2. To navigate to the segment of
interest, press Current Seg from the side menu and use the Variable knob to scroll to the segment of interest.
Alternatively, the Set to Minimum and Set to Maximum keys can be used to jump to the first and last segment respectively.
Navigate Segmented Memory
90
ADVANCED CONFIGURATION
3. The position in time of the selected
segment relative to the time of the first segment is shown in the Segments Time key.
Background
When all the segments have been acquired, the play/pause key can be used to play back through each segment.
Operation
1. Make sure the scope is in Segments Stop mode.
See page 88 for details.
2. Press the Play/Pause key to run
through the acquired segments in numerical order.
Press the Play/Pause key again
to pause the playback.
When the scope has played
through to the last segment, pressing the Play/Pause key again will play through each segment again in reverse order.
Background
The Segmented memory function can be used in conjunction with the automatic measurements configured in the Measurement menu (see page
46).
Play Through Each Segment
Measurement on Segments
91
MDO-2000EG & 2000EX User Manual
Modes
Segments Measure
This function will either perform statistics calculations on the segments or tabulate a list of the measurement results for all the segments.
Segments Info
Provides configuration information common for all the acquired memory segments.
Segments Measure
The Segments Measure function allows you to view automatic measurements for the segments in statistical bins or as a list displaying the result of each automatic measurement.
Statistics
This function will bin the measurement results of a single automatic measurement into a user-defined number of bins. This allows you to easily view statistics for a large number of segments. For example, the statistics function will display the number of results for each bin and the measurement range of each bin for the selected automatic measurement.
Measurement List
Puts all the measurement results for a segment in a list. All the currently selected automatic measurement results are listed. A maximum of 8 automatic measurements can be used with this function.
92
ADVANCED CONFIGURATION
Note
To use automatic measurements with the segmented memory, automatic measurements must first be selected from the Measure menu before the segmented memory function is run.
Setup
Press the Measure key and select any single source measurement from the Add Measurement menu.
See page 50 for details on how to add automatic measurements.
Measure
Operation
1. Press Analyze Segments from the
Segments menu.
Note: This key will only be available in the Stop mode.
2. Press Segments Measure.
3. Select either the
statistics or the measurement list from the side menu.
Statistics
List
4. The statics table or measurement list appears on
the display.
Note that the more segments that you have, the longer it will take to calculate the statics or list the measurement results.
93
MDO-2000EG & 2000EX User Manual
5. For statistic measurements,
press Plot Source to choose which automatic measurement to use for the statistics calculations. The statistics for only one automatic measurement can be viewed at a time.
6. For the measurement list, press
Source and select the source channel for measurement.
Range
CH1 ~ CH4
Statistics Results
This function will bin the measurement results of the selected automatic measurement into a user­defined number of bins.
Setup
7. To select the number of bins for the
statistics, press Divided by and select the number of bins with the Variable knob.
Range
1~20 bins
8. Press Select and use the Variable
knob to view the measurement results for each bin.
94
ADVANCED CONFIGURATION
Example: Statistics
Statistics of currently
selected bin
Cursor/selected bin
Bin count
Measurement List
Puts all the measurement results for a segment in a list.
Setup
9. Press Select and use the variable
knob to scroll through each segment.
Example: Measurement List
Select cursor
Measurement results
Measurement types
95
Segment Info
Operation
1. Press Analyze Segments from the
bottom menu.
Note: This key will only be available in the Stop mode.
2. Press Segments Info.
3. A table showing all general setting information
for the segmented memory acquisitions is shown on the display.
Info:
Sample rate, Record length, Horizontal, Vertical
MDO-2000EG & 2000EX User Manual
96
ADVANCED CONFIGURATION
Background
When the waveform is displayed on the screen, it can be displayed as dots or vectors.
Panel Operation
1. Press the Display menu key.
Display
2. Press Dot / Vector to toggle between
Dot and Vector mode.
Range
Dots
Only the sampled dots are displayed.
Vectors
Both the sampled dots and the connecting line are displayed.
Example:
Vectors
Dots
Background
The persistence function allows the MDO­2000EG/2000EX to mimic the trace of a traditional analog oscilloscope. A waveform trace can be configured to “persist” for a designated amount of time.
Panel Operation
1. Press the Display menu key.
Display
Display
The Display menu defines how the waveforms and parameters appear on the main LCD display.
Display Waveform as Dots or Vectors
Set the Level of Persistence
97
MDO-2000EG & 2000EX User Manual
2. To set the persistence time, press
the Persistence menu button on the bottom bezel.
3. Use the Variable knob to select a
persistence time.
Time
16ms, 30ms, 60ms, 120ms, 240ms,
0.5s, 1s, 2s,~4s, Infinite, Off
Clear
To clear persistence, press Clear Persistence.
Background
The intensity level of a signal can also be set to mimic the intensity of an analog oscilloscope by setting the digital intensity level.
Panel Operation
1. Press the Display menu key.
Display
2. Press Intensity from the bottom
menu.
Waveform Intensity
3. To set the waveform intensity, press Waveform
Intensity and edit the intensity.
Range
0~100%
Example
Waveform Intensity 50%
Waveform Intensity 100%
Set the Intensity Level
98
Loading...