Instek GPT-12001, GPT-12002, GPT-12003, GPT-12004, GPT-15001 Users Manual

...
Electrical Safety Analyzer
GPT-10000 Series
USER MANUAL
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Table of Contents
Table of Contents
SAFETY INSTRUCTIONS .................................................. 5
GETTING STARTED .......................................................... 9
GPT-10000 Series Overview ................. 10
Front Panel Overview ........................... 15
Rear Panel Overview ............................ 19
Set Up .................................................. 22
OPERATION .................................................................. 29
Menu Tree ............................................ 31
Test Lead Connection ........................... 36
Manual Tests ....................................... 39
Special MANU Test Mode (000) ........... 90
Sweep Function.................................... 92
Automatic Tests ................................... 95
System Settings ................................. 119
EXTERNAL CONTROL ................................................... 165
External Control Overview .................. 166
REMOTE CONTROL ...................................................... 172
Interface Configuration ...................... 173
Command Syntax ............................... 178
Command List ................................... 181
Error Messages .................................. 238
FAQ .............................................................................. 239
APPENDIX .................................................................... 242
Fuse Replacement .............................. 242
Tester Errors ...................................... 243
3
GPT-10000 Series User Manual
Factory Default Parameters ................ 246
GPT-10000 Specifications .................. 248
GPT-15001/15002/15003 Dimensions 256
GPT-15004 Dimensions ...................... 257
GPT-12001/12002/12003 Dimensions 258
GPT-12004 Dimensions ...................... 259
Declaration of Conformity .................. 260
Declaration of Conformity .................. 261
INDEX .......................................................................... 262
4
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the instrument or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Frame or Chassis Terminal Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
5
GPT-10000 Series User Manual
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not disassemble the GPT-10000 unless you
are qualified.
Position Guideline
WARNING
The rear position of the GPT-10000 should be
placed in an area with easy accessible for power disconnection, that is, unplugging the power cord with ease.
Keep away from the device under test which
connects with the GPT-10000 when test is underway. In addition, while test is ongoing, never touch the device under test, the GPT­10000 as well as other relevant units.
Any inappropriate manner that is unspecified
by the manufacturer may result in irreversible harms or impaired protection by the GPT-10000.
Safety Guidelines
6
SAFETY INSTRUCTIONS
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The GPT-10000 does not fall under category II, III or IV.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Power Supply
WARNING
AC Input voltage range:
AC 100V – 240V ± 10%
Frequency: 50Hz/60Hz To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.
Cleaning the GPT-10000
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 70% (no condensation) Altitude: < 2000m Temperature: 0˚C~40˚C
7
GPT-10000 Series User Manual
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The GPT-10000 falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -10°C to 70°C Relative Humidity: 85% (no condensation)
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
8
GETTING STARTED
GPT-10000 Series Overview ............................................ 10
Series lineup ..............................................................................................10
Lineup Overview......................................................................................11
Main Features ...........................................................................................11
Accessories ................................................................................................12
Package Contents .....................................................................................14
Front Panel Overview ..................................................... 15
GPT-12001/12002/12003/12004/15001/15002/15003/15004 ...........15
Rear Panel Overview ...................................................... 19
GPT-12001/12002/12003/12004/15001/15002/15003/15004 ...........19
Set Up ........................................................................... 22
Tilting the Stand .......................................................................................22
Line Voltage Connection and Power Up ..............................................23
Installing the Optional GPIB Card .........................................................24
Workplace Precautions ............................................................................25
Operating Precautions .............................................................................26
Basic Safety Checks ..................................................................................28
GETTING STARTED
This chapter describes the safety analyzer in a nutshell, including its main features and front / rear panel introduction. After going through the overview, please read the safety considerations in the Set Up chapter.
9
GPT-10000 Series User Manual
GPT-10000 Series Overview
Series lineup
The GPT-10000 Series Safety Analyzers are AC/DC withstanding voltage, insulation resistance, ground bond and continuity safety analyzers.
By and large, GPT-10000 Series has 2 major categories, one is GPT­12XXX models, and the other is GPT-15XXX models. The subordinate models of 2 categories share the same test functions but with different
specifications. We use the term “X” for the 2nd digit of model names
described below to stand for both 2 categories in common.
The GPT-1X001 is AC withstanding voltage and continuity tester, the GPT-1X002 is AC/DC withstanding voltage and continuity tester and the GPT-1X003 is AC/DC withstanding voltage, insulation resistance and continuity tester. The GPT-1X004 includes all the test functions of the other models, plus the ground bond testing. See the following Lineup Overview for more details.
The GPT-10000 Series can store up to 100 manual tests, as well as run up to 10 manual tests sequentially as an automatic test, allowing the safety analyzers to accommodate any number of safety standards, including IEC, EN, UL, CSA, GB, JIS and others.
Note: Throughout this user manual, the terms ACW, DCW, IR, GB and CONT refer to AC Withstanding, DC Withstanding, Insulation Resistance, Ground Bond and Continuity testing, respectively.
10
GETTING STARTED
Model name
ACW
DCW
IR
GB
CONT
GPT-12001
  GPT-12002
 GPT-12003
GPT-12004
GPT-15001
*
 GPT-15002
*
GPT-15003
*
 
GPT-15004
*
  
Short Current > 200mA
Performance
ACW: 0.05kV~5kVAC DCW: 0.05kV~6kVDC IR: 50V~1200V (50V steps)* GB: 3A~32A CONT: 100mA (fixed)
Lineup Overview
Main Features
11
GPT-10000 Series User Manual
Features
Ramp up time control Ramp down time control Safety discharge 100 test conditions (MANU mode) 100 automatic tests (AUTO mode) Over temperature, voltage and current
protection
Pass, Fail, Test, High Voltage and Ready
indicators
PWM output (90% efficiency, increased
reliability)
Interlock (configurable) Rear panel output
Interface
Remote control start/stop interface terminal RS232/USB interface for programming Optional GPIB interface for programming Optional LAN interface for programming Signal I/O port for pass/fail/test monitoring
and start/stop control/interlock
Standard Accessories
Part number
Description GHT-115x1
Test lead
Region dependent
Power cord
GTL-215x1
GB test lead
(GPT-12004/GPT-15004 only)
GHT-119
Remote terminal cable
N/A
Interlock key
Accessories
12
GETTING STARTED
Optional Accessories
Part number
Description GHT-205
High Voltage Test Probe
GHT-113
High Voltage Test Pistol
GTL-232
RS232C cable
GTL-248
GPIB cable
GTL-246
USB cable (A to B type)
GRA-440
Rack Adapter Panel (19”, 4U) (GPT-12000/15000 only)
Options
Part number
Description
Opt.01 GPIB Interface
GPIB module
Opt.02 LAN Interface
LAN module
13
GPT-10000 Series User Manual
Opening the box
Contents (single unit)
GPT-10000 unit Quick Start Guide User manual CD CTC (Calibration
Traceable Certificate)
Power cord x1 (region
dependent)
GHT-115 test leads x1 GTL-215 GB test leads
x1 (GPT-12004/GPT­15004 only)
GHT-119 Remote
terminal cable
Interlock key
Note
Keep the packaging, including the box, polystyrene foam and plastic envelopes should the need arise to return the unit to GW Instek.
Package Contents
Check the contents before using the GPT-10000 series.
14
GETTING STARTED
1
2
3
4 5
6
7
8
9
0
A
B
C
D
E
F
G
Item
Description
1
STOP Button
2
START Button
3
POWER Switch
4
Test Function Keys (Green Zone)
5
Display
6
Mode Keys (AUTO, MANUAL, SYSTEM in Red Zone)
7
Soft Keys (Blue Zone)
8
USB A-Type Host Port
9
PASS/FAIL Indicators
0
REMOTE Terminal
A
Scroll Wheel
B
HIGH VOLTAGE Indicator
C
HIGH VOLTAGE Output Terminal
D
SENSE L & RETURN Terminal
E
SENSE H & Output Terminal
F
SOURCE L (GPT-12004/GPT-15004 only)
G
SOURCE H (GPT-12004/GPT-15004 only)
Front Panel Overview
GPT-12001/12002/12003/12004/15001/15002/15003/15004
15
GPT-10000 Series User Manual
STOP button
The STOP button is used to stop/cancel tests. The STOP button will also put the safety analyzer in the READY status to begin testing.
START button
The START button is used to start tests. The START button can be used to start tests when the tester is in the READY status. Pressing the START button will put the tester in the TEST status.
POWER switch
Turns the power on. The safety analyzer will always start up with the last test setting from when the instrument was last powered down.
Test Function Keys
The keys indicate the 5 testing functions including ACW, DCW, IR, GB and CONT. Pressing one of the keys enters the function settings.
Display
7” Color TFT LCD display in 800 X 480 resolution
AUTO button
Press to enter the AUTO test mode.
MANUAL button
Press to enter the MANUAL test mode.
SYSTEM button
Press to enter the SYSTEM mode.
Soft Keys
The Soft keys correspond to the menu keys directly above on the main display.
USB Host Port
It can connect with USB flash drive for data and log import/export and for firmware update. Also, it is able to connect with barcode scanner for convenient tests.
16
GETTING STARTED
Pass/Fail indicators
The PASS and FAIL indicators light up upon a PASS or FAIL test result at the end of a manual test or automatic test.
REMOTE terminal
The REMOTE terminal is used to connect to a remote controller.
Scroll wheel
The scroll wheel is used to edit parameter values.
HIGH VOLTAGE indicator
The HIGH VOLTAGE indicator will light up red when an output terminal is active. Only after the test has finished or stopped will the indicator turn off.
HIGH VOLTAGE output terminal
The HIGH VOLTAGE terminal output is used for outputting the testing voltage in ACW, DCW and IR tests. The terminal is recessed for safety. This terminal is used in conjunction with the RETURN terminal.
WARNING
USE EXTREME CAUTION. Do not touch the HIGH VOLTAGE terminal during testing.
RETURN terminal
The RETURN terminal is used for ACW, DCW, IR and CONT tests.
17
GPT-10000 Series User Manual
OUTPUT and RETURN terminals
All models
The OUTPUT terminal (red) and RETURN terminal (black) are used for CONT (Continuity) test.
SENSE H/L and SOURCE H/L terminals
GPT-12004, GPT-15004 only
The SENSE H, SENSE L, SOURCE H and SOURCE L, terminals are used for GB (Ground Bond) test.
18
GETTING STARTED
1
3
7 8 9 A
B
C
D
2
4
5
6
E
Item
Description
1
HIGH VOLTAGE Output Terminal
2
HIGH VOLTAGE Indicator
3
SENSE H & Output Terminal
4
SENSE L & RETURN Terminal
5
SOURCE H (GPT-12004/GPT-15004 only)
6
SOURCE L (GPT-12004/GPT-15004 only)
7
Fan 8 RS-232 Port
9
USB B-Type Interface Port
A
Signal I/O Port
B
GPIB Port/Ethernet LAN Port (Optional)
C
GND
D
AC Mains Input (Power Cord Socket)
E
HIGH VOLTAGE pilot lamp
Rear Panel Overview
GPT-12001/12002/12003/12004/15001/15002/15003/15004
19
GPT-10000 Series User Manual
SIGNAL I/O port
The SIGNAL I/O port is used to monitor the tester status (PASS, FAIL, TEST) and input (START/ STOP signals). It is also used with the Interlock key.
USB B-Type port
The USB B-Type port is used for remote control.
RS232 interface port
The RS-232 port is used for remote control.
Fan/Fan Vents
Exhaust fan. Allow enough room for the fan to vent. Do not block the fan openings.
GND
Connect the GND (ground) terminal to the earth ground.
AC Mains Input
AC Mains Input for Power Cord Socket: 100 – 240 VAC ±10%.
The fuse holder contains the AC mains fuse. For fuse replacement details, see page 45.
Optional LAN port
Optional LAN port for remote control.
Optional GPIB port
Optional GPIB interface for remote control.
20
GETTING STARTED
HIGH VOLTAGE output terminal
The HIGH VOLTAGE terminal output is used for outputting the testing voltage in ACW, DCW and IR tests. The terminal is recessed for safety and used in conjunction with the RETURN terminal.
WARNING
USE EXTREME CAUTION. Do not touch the HIGH VOLTAGE terminal during testing.
HIGH VOLTAGE pilot lamp
The HIGH VOLTAGE pilot lamp will light up red when an output terminal is active. Only after the test has finished or stopped will the lamp turn off.
RETURN terminal
The RETURN terminal is used for ACW, DCW, IR and CONT tests.
OUTPUT and RETURN terminals
All models
The OUTPUT terminal (red) and RETURN terminal (black) are used for CONT (Continuity) test.
SENSE H/L and SOURCE H/L terminals
GPT-12004, GPT-15004 only
The SENSE H, SENSE L, SOURCE H and SOURCE L, terminals are used for GB (Ground Bond) test.
21
GPT-10000 Series User Manual
Horizontal position
Place the unit on a flat surface horizontally.
Tilt stand position
Gently pull the 2 stands out from the bottom and the unit will be placed in the tilt stand position.
Set Up
Tilting the Stand
22
GETTING STARTED
Background
The GPT-10000 accepts line voltages of 100 - 240V at 50Hz or 60Hz.
Steps
1. Connect the power cord to
the AC Mains Input socket on the rear panel.
2. If the power cord does not
have an earth ground, ensure the ground terminal is connected to an earth ground.
Warning
Ensure the power cord is connected to an earth ground. Failure could be harmful to the operator and instrument.
3. Press the Power button.
4. When the unit is powered up, the display will
show the last time parameters in either MANU or AUTO test mode as shown below.
Line Voltage Connection and Power Up
23
GPT-10000 Series User Manual
Background
The optional GPIB is a user-installable option. Follow the instructions below to install the GPIB card.
WARNING
Before installing optional GPIB/LAN card ensure the GPT-10000 is turned off and disconnected from power.
Steps
1. Remove screws from the rear panel cover plate.
2. Insert the GPIB/LAN card into the opening of
rear panel. Push the card gently until it is fully inserted followed by fastening the screws.
GPIB Card
LAN Card
Installing the Optional GPIB Card
24
GETTING STARTED
Background
The GPT-10000 is a high voltage instrument that outputs dangerous voltages. The following section describes precautions and procedures that must be followed to ensure a safe work environment.
WARNING
The GPT-10000 generates voltages in excess of 5kVAC or 6kVDC. Follow all safety precautions, warnings and directions given in the following section when using the instrument.
1. Only technically qualified personnel should be
allowed to operate the safety analyzer.
2. The operating workplace must be fully isolated,
especially when the instrument is in operation. The instrument should be clearly labeled with appropriate warning signage.
3. The operator should not wear any conductive
materials, jewelry, badges, or other items, such wrist watches.
4. The operator should wear insulation gloves for
high voltage protection.
5. Ensure the earth ground of the line voltage is
properly grounded.
6. Ensure any devices that are adversely affected
by magnetic fields are not placed near the tester.
Workplace Precautions
25
GPT-10000 Series User Manual
Background
The GPT-10000 is a high voltage instrument that outputs dangerous voltages. The following section describes precautions and procedures that must be followed to ensure that the tester is operated in a safe manner.
WARNING
The GPT-10000 generates voltages of up to 5kVAC or 6kVDC. Follow all safety precautions, warnings and directions given in the following section when using the instrument.
1. Never touch the safety analyzer, lead wires,
terminals, probes and other connected equipment when the tester is testing.
2. Do not turn the safety analyzer on and off
quickly or repeatedly. When turning the power off, please allow a few moments before turning the power back on. This will allow the protection circuits to properly initialize.
Do not turn the power off when a test is running, unless in an emergency.
3. Only use those test leads supplied with the
instrument. Leads with inappropriate gauges can be dangerous to both the operator and the instrument. For GB testing, never use the Sense leads on the SOURCE terminals.
4. Do not short the HIGH VOLTAGE terminal
with ground. Doing so could charge the chassis to dangerously high voltages.
Operating Precautions
26
GETTING STARTED
5. Ensure the earth ground of the line voltage is
properly grounded.
6. Only connect the test leads to the HIGH
VOLTAGE/SOURCE H/SENSE H terminals before the start of a test. Keep the test leads disconnected at all other times.
7. Always press the STOP button when pausing
testing.
8. Do not leave the safety analyzer unattended.
Always turn the power off when leaving the testing area.
9. When remotely controlling the safety analyzer,
ensure adequate safety measures are in place to prevent:
Inadvertent output of the test voltage.
Accidental contact with the instrument during
testing. Ensure that the instrument and DUT are fully isolated when the instrument is remotely controlled.
10. Ensure an adequate discharge time for the
DUT.
When DCW or IR tests are performed, the DUT, test leads and probes become highly charged. The GPT-10000 has discharge circuitry to discharge the DUT after each test. The time required for a DUT to discharge depends on the DUT and test voltage.
Never disconnect the safety analyzer before a discharge is completed.
27
GPT-10000 Series User Manual
Background
The GPT-10000 is a high voltage device and as such, daily safety checks should be made to ensure safe operation.
1. Ensure all test leads are not broken and are free
from defects such as cracks or splitting.
2. Ensure the safety analyzer is always connected
to an earth ground.
3. Test the safety analyzer operation with a low
voltage/current output: Ensure the safety analyzer generates a FAIL judgment when the HIGH VOLTAGE and RETURN terminals are shorted (using the lowest voltage/current as the testing parameters).
WARNING
Do not use high voltages/currents when the HIGH VOLTAGE and RETURN terminals are shorted. It may result in damage to the instrument.
Basic Safety Checks
28
OPERATION
Menu Tree...................................................................... 31
Menu Tree Overview ...............................................................................32
Test Lead Connection ..................................................... 36
ACW, DCW, IR Connection ....................................................................36
GB Connection ..........................................................................................37
CONT Connection ....................................................................................38
Manual Tests ................................................................. 39
Setting the Test Function .........................................................................40
Choose/Recall a Manual Test Number .................................................41
Creating a MANU Test File Name .........................................................42
Setting the Upper and Lower Limits .....................................................43
Setting the Test Time ...............................................................................45
Setting the Ramp Up Time ......................................................................47
Setting the Ramp Down Time ................................................................49
Setting the Test Voltage or Test Current ...............................................51
Setting the Test Frequency ......................................................................53
Setting a Reference Value ........................................................................54
Setting an Initial Voltage .........................................................................55
Setting the Wait Time ..............................................................................57
Setting the ARC Function ........................................................................59
Setting MAX HOLD .................................................................................62
Setting PASS HOLD .................................................................................63
Setting IR Mode ........................................................................................64
Setting GND OFFSET ..............................................................................66
Setting GB Contact ...................................................................................68
Zero Check for the Test Leads ................................................................70
Setting the Grounding Mode ..................................................................73
Setting Contact Check ..............................................................................78
Running a MANU Test ............................................................................80
OPERATION
29
GPT-10000 Series User Manual
PASS / FAIL MANU Test ....................................................................... 85
Special MANU Test Mode (000) ....................................... 90
Sweep Function............................................................... 92
Automatic Tests .............................................................. 95
Choose/Recall an AUTO Test ................................................................ 96
Creating an AUTO Test File Name ........................................................ 97
Adding a Step to the AUTO Test ........................................................... 98
Continuous AUTO Tests ....................................................................... 100
AUTO Test Page Editing ....................................................................... 102
Running an Automatic Test .................................................................. 107
Automatic Test Results .......................................................................... 113
30
OPERATION
Menu Tree
This section describes the overall structure of the operation statuses and modes for the GPT-10000 safety analyzers. The testers have two main testing modes (MANU, AUTO), one system mode (SYSTEM) and 5 main operation statuses (READY, TEST, PASS, FAIL and STOP).
31
GPT-10000 Series User Manual
MANU Mode
MANU mode is used to create and/or execute a single test. Only under MANU mode can parameters be edited for each manual test.
MANU mode
AUTO Mode
AUTO mode indicates that the tester is automatic, which consists of a sequential AUTO test of up to 10 MANU steps. Also, several groups of AUTO tests can be further interconnected for an advanced AUTO test.
AUTO mode
Menu Tree Overview
32
OPERATION
SYSTEM Mode
System mode covers the Display Set, Buzzer, Interface, Control, System Time, Data Initialize, Information, Statistics, USB Disk and CONTACT CHK settings. These settings are system-wide and applied to both MANU and AUTO tests.
SYSTEM mode
READY Status (Yellow Color)
When the tester is in READY status of MANU or AUTO test, it is ready to begin testing. Pressing the START button will begin testing and put the tester into TEST status. Pressing the AUTO key will change from MANU – READY status to AUTO – READY status and vice versa.
READY status in MANU test
READY status in AUTO test
33
GPT-10000 Series User Manual
TEST Status (Orange Color)
TEST status is active when a MANU test or AUTO test is running. Pressing STOP will cancel the MANU test or the remaining steps in an AUTO test instantly. The TEST status in AUTO test is identical with that of MANU test.
TEST status in MANU test
PASS Status (Green Color)
When a MANU test result is within the range of HI and LOW sets, the PASS status is shown on display. For AUTO test, the PASS status only shows when all the affiliated test steps are passed.
PASS status in MANU test
PASS status in AUTO test
34
OPERATION
FAIL Status (RED Color)
When a MANU test result is beyond the range of HI and LOW sets, the FAIL status is shown on display. For AUTO test, the FAIL status is shown when any of the test steps fails, even only one of them.
FAIL status in MANU test
FAIL status in AUTO test
STOP Status (Red Color)
STOP status is shown when an AUTO test did not finish running and has been stopped by user. Pressing STOP will return the tester to READY status. STOP status is not shown in MANU test as it returns to READY status directly after user pressed STOP button in MANU test.
STOP status in AUTO test
35
GPT-10000 Series User Manual
Background
ACW, DCW and IR tests use the HIGH VOLTAGE terminal and RETURN terminal with the GHT-115 test leads.
ACW, DCW, IR Connection
High Voltage terminal
DUT
GPT-10000
Return terminal
Steps
1. Turn the power off on the safety analyzer.
2. Connect the high voltage test lead (red) to the
HIGH VOLTAGE terminal and screw firmly into place.
3. Connect the return test lead (white) into the
RETURN terminal and screw the protector bar into place, as shown below.
HIGH
VOLTAGE
Terminal
RETURN
Terminal
Test Lead Connection
This section describes how to connect the GPT-10000 to a DUT for withstanding, insulation resistance, ground bond as well as continuity testing.
ACW, DCW, IR Connection
36
GB Connection
Background
GB tests use the SENSE H/L and SOURCE H/L terminals with the GTL-215 test leads.
GB Connection
Source H
DUT
GPT-10000
Sense L
Sense H
Source L
Steps
1. Turn the power off on the safety analyzer.
2. Connect the Sense H lead to the SENSE H
terminal.
3. Connect the Sense L lead to the SENSE L
terminal.
4. Connect the Source H lead to the SOURCE H
terminal.
5. Connect the Source L lead to the SOURCE L
terminal.
Sense H lead Sense L lead
Source L leadSource H lead
OPERATION
37
GPT-10000 Series User Manual
Background
CONT tests use the OUTPUT and RETURN terminals with the GTL-115 test leads.
CONT Connection
DUT
GPT-10000
Return terminal
Output
Steps
1. Turn the power off on the safety analyzer.
2. Connect the OUTPUT test lead (red) to the
OUTPUT terminal.
3. Connect the RETURN test lead (black) into the
RETURN terminal and screw the protector bar into place, as shown below.
OUTPUT
Terminal
RETURN
Terminal
CONT Connection
38
OPERATION
Manual Tests
This section describes how to create, edit and run a single ACW, DCW, IR, GB and CONT manual tests. Each Manual setting described in this chapter only applies to the selected manual test – no other manual tests are affected.
Each manual test can be stored/recalled to/from one of 100 memory locations. Each stored manual test can be used as a test step when creating an AUTO test (page 92).
Setting the Test Function → from page 40. Choose/Recall a Manual Test Number → from page 41. Creating a MANU Test File Name → from page 42. Setting the Upper and Lower Limits → from page 43. Setting the Test Time → from page 45. Setting the Ramp Up Time → from page 47. Setting the Ramp Down Time → from page 49. Setting the Test Voltage or Test Current → from page 51. Setting the Test Frequency from page 53. Setting a Reference Value → from page 54. Setting an Initial Voltage → from page 55. Setting the Wait Time → from page 57. Setting the ARC Function → from page 59. Setting MAX HOLD → from page 62. Setting PASS HOLD → from page 63. Setting IR Mode → from page 64. Setting GND OFFSET → from page 66. Setting GB Contact → from page 68. Zero Check for the Test Leads → from page 70. Setting the Grounding Mode → from page 73. Setting Contact Check → from page 78. Running a MANU Test from page 80. PASS / FAIL MANU Test → from page 85. Special MANU Test Mode (000) from page 90. Sweep Function → from page 92.
39
GPT-10000 Series User Manual
Background
There are five test functions, AC Withstand, DC Withstand, Insulation Resistance, Ground Bond and Continuity tests.
Steps
1. If the tester is in AUTO or SYSTEM
mode, press the MANUAL key to put the tester into MANU mode.
2. To choose the test function, press the ACW,
DCW, IR, GB or CONT key on the front panel.
3. The key of selected test function is lit, and the
test function selected is shown on the upper-left corner of the display.
Selected Test Function
Setting the Test Function
40
OPERATION
Background
ACW, DCW, IR, GB and CONT tests can only be created and edited in the MANU mode. MANU number 001 to 100 can be saved and thus be loaded when editing/creating a MANU test or AUTO test. MANU number 000 is a special mode. See page 90 for details on the special mode.
Steps
1. If the tester is in AUTO or SYSTEM
mode, simply press the MANUAL key to switch to MANU mode.
2. Use the scroll wheel to choose the
MANU number.
MANU #
001~100 (MANU# 000 is a special mode)
MANU number cursor
Note
Manual number can only be selected
or recalled when the “READY” status shows on the screen. If the “FAIL”
status appears, it is required to press STOP key first before selecting or recalling procedure.
Choose/Recall a Manual Test Number
41
GPT-10000 Series User Manual
Background
Each manual test can have a user-defined name (default: MANU_NAME) up to 10 characters long. See the available list of characters below.
Character List
0 1 2 3 4 5 6 7 8 9
W X Y ZM VO P RQ T USNLA B C D E F G H I J K
w x y zm vo p rq t usnla b c d e f g h i j k
_
Steps
1. Press the UP / DOWN arrow soft-
keys to bring the cursor to the MANU_NAME (default name) field. The characters table will appear in the right hand accordingly.
MANU NAME cursor Characters Table
2. Use the scroll wheel to scroll
through the available characters.
3. Press the LEFT / RIGHT arrow soft-
keys to move the cursor to the next character.
4. The MANU test file name is set when the
cursor is moved to another setting.
Creating a MANU Test File Name
42
OPERATION
Background
There is both a LOW and HI judgment setting. When the measured value is below the LOW SET setting, the test will be judged as FAIL. When the value exceeds the HI SET setting the test will be judged as FAIL. Any measurement between the LOW SET and HI SET setting is judged as PASS. The LOW SET limit cannot be made greater than the HI SET limit.
Steps
1. Press the UP / DOWN arrow soft-
keys to bring the cursor to the HI SET setting.
HI SET cursor
2. Use the scroll wheel to set the HI
SET limit.
ACW (HI)
DCW (HI)
IR (HI) GB (HI)
CONT (HI)
001uA~42.00mA (GPT-12XXX) 001uA~110.0mA (GPT-15XXX) 001uA~11.00mA (GPT-12XXX) 001uA~21.00mA (GPT-15XXX)
000.2M~50.00G, OFF
000.1m~650.0m
00.01~80.00
Setting the Upper and Lower Limits
43
GPT-10000 Series User Manual
3. Press the UP / DOWN arrow soft-
keys to bring the cursor to the LOW SET setting.
LOW SET cursor
4. Use the scroll wheel to set the
LOW SET limit.
ACW (LOW)
DCW (LOW)
IR (LOW) GB (LOW) CONT (LOW)
000uA~41.99mA (GPT-12XXX) 000uA~109.9mA (GPT-15XXX) 000uA~10.99mA (GPT-12XXX) 001uA~20.99mA (GPT-15XXX)
000.1M~49.99G
000.0m~649.9m
00.00~79.99
Note
*Please note that the resolution of the measured value depends on the resolution of HI SET setting.
Note
The LOW SET setting is limited by the HI SET setting. The LOW SET limit cannot be greater than the HI SET limit.
When setting the current, be aware that a maximum of 200VA can be set for ACW and 50W for DCW, respectively in terms of GPT-12XXX series.
As for GPT-15XXX series, a maximum of 500VA can be set for ACW and 100W for DCW, respectively.
44
OPERATION
Background
This setting is used to set the test time for a test. The test time determines how long the test voltage or current is applied to the DUT. This test time does not include RAMP UP time or RAMP DOWN time (note: GB and CONT do not have RAMP UP or RAMP DOWN). The test time can be set from 0.3 seconds to 999.9 seconds for ACW, DCW, IR, GB and CONT, with a resolution of 0.1 seconds for all modes. Also, the test time can be turned off when using the ACW or DCW test functions.
Each test has a RAMP UP and RAMP DOWN time (except GB and CONT), respectively. Refer to page 47 & 49 for more details.
Output Voltage Timing Chart (Resistive load)
Test V
time
TEST TIME
RAMP UP
RAMP DOWN
Initial V
(Adjustable)
Steps
1. Press the UP / DOWN arrow soft-
keys to bring the cursor to the TEST TIME setting.
TEST TIME cursor
Setting the Test Time
45
GPT-10000 Series User Manual
2. Use the scroll wheel to set the
TEST TIMER value.
ACW DCW IR GB
CONT
OFF, 000.3s~999.9s OFF, 000.3s~999.9s
000.3s~999.9s
000.3s~999.9s
000.3s~999.9s
Note
With the ACW test function, when the test current is beyond 30mA, the Ramp Up Time + Test Time cannot exceed 240 seconds. At this current level, the tester also needs to pause after a test for a time equal to or greater than the output time.
Turn Off Test Time
When in either ACW or DCW test, the TEST TIME can be turned off, which means the test without test time will last infinitely until FAIL judgment occurs.
Identical with the regular setting for TEST TIME, turn off the timer by using the scroll wheel to set OFF for TEST TIME value.
TEST TIME OFF
46
OPERATION
Background
The Ramp Up time is the total time taken for the tester to reach the test voltage level. The Ramp Up time starts with a start voltage of 50 volts. The Ramp Up time can be set from 000.1 to 999.9 seconds. The Ramp Up time is only applicable for ACW, DCW and IR tests.
Output Voltage Timing Chart (Resistive load)
Test V
time
TEST TIME
RAMP UP
RAMP DOWN
Initial V
(Adjustable)
Steps
1. Press the UP / DOWN arrow soft-
keys to bring the cursor to the RAMP TIME setting.
RAMP TIME cursor
2. Use the scroll wheel to set the
RAMP TIME value.
ACW DCW IR
000.1s~999.9s
000.1s~999.9s
000.1s~999.9s
Setting the Ramp Up Time
47
GPT-10000 Series User Manual
Ramp Time Duration Indicator
After pressing START to begin a test with set RAMP TIME, a section at the lower right corner of display shows the counting duration of RAMP TIME, which will run to the set value followed by the test time. See the screenshot shown below.
RAMP TIME duration indicator
48
OPERATION
Background
The Ramp Down time is the time taken for the DUT to discharge the test voltage level. The Ramp Down time can be set from 000.0 to 999.9 seconds. The Ramp DOWN time is only applicable for ACW, DCW and IR tests.
Output Voltage Timing Chart (Resistive load)
Test V
time
TEST TIME
RAMP UP
RAMP DOWN
Initial V
(Adjustable)
Steps
1. Press the PAGE soft-key to move to
the 2/3 page where RAMP DOWN setting appears for ACW and DCW.
As for IR test, the RAMP DOWN setting shows in the 2/2 page.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the RAMP DOWN setting.
RAMP DOWN cursor
Setting the Ramp Down Time
49
GPT-10000 Series User Manual
3. Use the scroll wheel to set the
RAMP DOWN value.
ACW DCW IR
000.0s~999.9s
000.0s~999.9s
000.0s~999.9s
Ramp Down Duration Indicator
After the set TEST TIME is fully completed, a section at the lower right corner of display shows the counting duration of RAMP DOWN, which will run to the set value by user. See the screenshot shown below.
RAMP DOWN duration indicator
50
OPERATION
Background
The test voltage can be set from 0.050kV to
5.1kV for ACW, 0.050kV to 6.1kV for DCW and
0.050 to 1.2kV for IR (50V steps*). For GB tests the test current can be set from 3A to 33A. As for CONT test, the test current is fixedly set at the default value of 100mA.
Steps
1. Press the UP / DOWN arrow soft-
keys to bring the cursor to the voltage or ampere setting depending on selected test function.
Test Voltage / Current cursor region
2. Use the scroll wheel to set the test
voltage or ampere level.
ACW DCW IR GB CONT
0.050kV ~ 5.1kV 1
0.050kV ~ 6.1kV 2
0.05kV ~ 1.2kV (50V steps)
3.00A ~ 33.00A 100mA 3
Setting the Test Voltage or Test Current
51
GPT-10000 Series User Manual
1
At least 0.3 seconds is needed to reach a set
voltage of 50V/10mA.
2
At least 0.3 seconds is needed to reach a set
voltage of 50V/2mA.
3
Test current for CONT is fixed at 100mA
Note
When setting the current, be aware that a maximum of 200VA can be set for ACW and 50W for DCW, respectively in terms of GPT-12XXX series.
As for GPT-15XXX series, a maximum of 500VA can be set for ACW and 100W for DCW, respectively.
The ground bond voltage (GBV) is calculated as the HI SET limit x Test Current.
52
OPERATION
Background
A test frequency of 60Hz or 50Hz can be set, regardless of the input line voltage. The test frequency setting only applies to ACW and GB tests.
Note
The test frequency can only be set for ACW or GB tests.
Steps
1. Press the PAGE soft-key to move
to the 2/3 page where FREQ setting appears for ACW.
As for GB test, the FREQ setting shows in the 1/2 page directly.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the FREQ setting.
FREQ setting cursor
3. Use the scroll wheel to set the test
frequency.
ACW, GB
50Hz, 60Hz
Setting the Test Frequency
53
GPT-10000 Series User Manual
Background
The REF VALUE acts as an offset. The REF VALUE is subtracted from the measured current (ACW, DCW) or measured resistance (IR, GB, CONT).
Steps
1. Press the PAGE soft-key to move to
the 3/3 page where REF VALUE setting appears for ACW and DCW.
As for IR and GB, the REF VALUE setting shows in the 2/2 page.
The REF VALUE setting appears in the 1/1 page directly for CONT.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the REF VALUE setting.
REF VALUE cursor
3. Use the scroll wheel to set the REF
value.
Setting a Reference Value
54
OPERATION
ACW
DCW
IR
GB
CONT
000uA~ 41.99mA (GPT-12XXX) *HI SET + REF value ≤ 42.00mA
000uA~ 109.9mA (GPT-15XXX) *HI SET+REF value 110.0mA
000uA~ 10.99mA (GPT-12XXX) *HI SET + REF value ≤ 11.00mA
000uA~ 20.99mA (GPT-15XXX) *HI SET+REF value 21.00mA
000.0M~50.00G
000.0m~650.0m *ISET x (HI SET + REF value) is no greater than 7.2V
00.00~80.00 *ISET(100mA) x (HI SET + REF
value) is no greater than 8V
Note
For IR test, a reference value of tester can be automatically created via the GND OFFSET function. See page 66 for details.
For GB and CONT tests, a reference value of test lead can be automatically created via the ZERO CHECK function. See page 70 for details.
Background
In essence, the test voltage for both ACW and DCW will gradually and linearly rise up, from zero, to the target set voltage in accord with the set RAMP TIME ahead of the TEST TIME.
Nevertheless, under certain circumstances, user may have preferences on the percentage of starting test voltage. Therefore, the INIT VOLTAGE provides another alternative for different applications on user side.
Setting an Initial Voltage
55
GPT-10000 Series User Manual
It is easy to set a preferred percentage of the test voltage in the INIT VOLTAGE setting and the starting test voltage will commence from the value corresponding to the set percentage relative to the target test voltage.
Note
INIT VOLTAGE setting is only applicable to both ACW and DCW tests.
Steps
1. Press the PAGE soft-key to move to
the 3/3 page where the INIT VOLTAGE setting appears for ACW and DCW.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the INIT VOLTAGE setting.
INIT VOLTAGE cursor
3. Use the scroll wheel to set the
percentage of INIT VOLTAGE.
INIT VOLTAGE
000% ~ 099%
56
OPERATION
Background
The Wait Time refers to the pending time before FAIL judgment appears. By default, FAIL judgment appears when Test Time has reached 0.3 second at the earliest manner. However, when user sets 1.0 second for Wait Time on the tester with 0.5 second of Ramp Up time and 1.0 second of Test Time, the FAIL judgment will be shown when Test Time has reached 0.5 second. In short, Wait Time is the pending duration which dominates the priority over both Ramp UP time and Test Time in terms of timing of FAIL judgment.
The WAIT TIME is only applicable for ACW, DCW and IR tests.
Steps
1. Press the PAGE soft-key to move to
the 2/3 page where WAIT TIME setting appears for ACW and DCW.
As for IR test, the WAIT TIME setting shows in the 1/2 page.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the WAIT TIME setting.
Setting the Wait Time
57
GPT-10000 Series User Manual
WAIT TIME cursor
3. Use the scroll wheel to set the
WAIT TIME value.
ACW DCW IR
000.0s~999.9s
000.0s~999.9s
000.0s~999.9s
Wait Time Indicator
While the WAIT TIME is set, the indicator of WAIT TIME will be shown on the display in the set duration during a test progress for clear identification for user.
WAIT TIME indicator
58
OPERATION
Background
ARC detection, otherwise known as flashover detection, detects fast voltage or current transients that are not normally detected. Arcing is usually an indicator of poor withstanding insulation, electrode gaps or other insulating problems that cause temporary spikes in current or voltage during ACW and DCW testing.
There are three ARC detection settings: OFF, ON & CONT, ON & STOP. The ON & CONT setting will detect arcs over the ARC current level and continue the test, the ON & STOP setting will stop the test when an arc is detected.
ARC mode settings only apply to both ACW and DCW tests.
Steps
1. Press the UP / DOWN arrow soft-
keys to bring the cursor to the ARC FUNC setting.
ARC FUNC cursor
2. Use the scroll wheel to set the ARC
modes setting.
Setting the ARC Function
59
GPT-10000 Series User Manual
ARC MODES:
OFF ON & CONT ON & STOP
3. If the ARC MODE was set to either
ON & CONT, or ON & STOP, the ARC current level can be edited. Press the DOWN arrow soft-key to bring the cursor to the ARC SET setting field.
ARC SET cursor
4. Use the scroll wheel to edit the ARC
SET level.
ACW
DCW
1.000mA~80.00mA (GPT-12XXX)
1.000mA~200.0mA (GPT-15XXX)
1.000mA~20.00mA (GPT-12XXX)
1.000mA~40.00mA (GPT-15XXX)
5. If the ARC MODE was set to either
ON & CONT, or ON & STOP, the ARC speed, which indicates the threshold for width of detected ARC, can be edited. Press the PAGE soft-key to move to the 2/3 page where ARC SPEED setting appears for ACW and DCW.
60
OPERATION
6. Press the UP / DOWN arrow soft-
keys to bring the cursor to the ARC SPEED setting field.
ARC SPEED cursor
7. Use the scroll wheel to select the
ARC SPEED modes.
ARC SPEED
FAST
Threshold for the narrowest width of detected arc, which is the most sensitive manner.
NORMAL
Threshold for the general width of detected arc.
SLOW
Threshold for the widest width of detected arc, which is the manner of high tolerance.
61
GPT-10000 Series User Manual
Background
The MAX HOLD setting will hold the maximum current measured in the ACW and DCW tests or the maximum resistance measured in the IR and GB tests.
For instance, when running an IR test with 120 seconds of test time and MAX HOLD enabled, the highest resistance measured in the 30 seconds of the test time will be retained on display until the next largest value. If there is no further maximum resistance occurred, the value measured in 30 seconds will be remained till the end of the test of 120 seconds.
Steps
1. Press the PAGE soft-key to move to
the 2/3 page where MAX HOLD setting appears for ACW and DCW.
As for IR and GB, the MAX HOLD setting shows in the 2/2 page.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the MAX HOLD setting.
MAX HOLD cursor
Setting MAX HOLD
62
OPERATION
3. Use the scroll wheel to set MAX
HOLD.
MAX HOLD
OFF, ON
Background
The PASS HOLD setting refers to the holding duration after PASS judgment is shown on the display. When the PASS HOLD setting is set, a PASS judgment is held until the set duration is fully reached.
Note
The PASS HOLD setting only applies to MANU tests. This setting is ignored when running AUTO test.
Steps
1. Press the PAGE soft-key to move to
the 3/3 page where PASS HOLD setting appears for ACW.
And it is the 2/3 page where PASS HOLD setting appears for DCW.
As for IR and GB, the PASS HOLD setting shows in the 2/2 page.
The PASS HOLD setting appears in the 1/1 page directly for CONT.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the PASS HOLD setting.
Setting PASS HOLD
63
GPT-10000 Series User Manual
PASS HOLD cursor
3. Use the scroll wheel to set PASS
HOLD duration.
PASS HOLD
000.0s ~ 999.9s, ON
Note
The STOP key can be pressed at any time in
the set duration of PASS HOLD to promptly halt the set PASS HOLD duration. In short, user can stop, if necessary, the duration of PASS HOLD any time.
When ON is selected, the duration of PASS
HOLD will remain indefinitely until the STOP key is further pressed.
Background
The IR MODE setting, which contains three options, STOP ON FAIL, STOP ON PASS, TIMER, only applies to IR test.
When IR MODE is set to STOP ON FAIL, the tester will show the FAIL judgment, if available, in the 0.3 second of test time at the earliest manner, regardless of the set test time.
When set to PASS ON FAIL, the tester will show the PASS judgment, if available, in the 0.3
Setting IR Mode
64
OPERATION
second of test time at the earliest manner, regardless of the set test time.
The TIMER mode will run a test in a full course completely in accordance with the set test time, whether the final judgment is PASS or FAIL.
Note
If the DUT is under the situation of abnormal measurement, e.g., short circuit, the FAIL judgment of SHORT warning, though TIMER is set, will appear in the early manner regardless of the set test time.
Steps
1. Press the PAGE soft-key to move to
the 2/2 page where IR MODE setting appears for IR test.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the IR MODE setting.
IR MODE cursor
3. Use the scroll wheel to set the IR
MODE.
IR MODE
STOP ON FAIL
STOP ON PASS
TIMER
65
GPT-10000 Series User Manual
Background
The GND OFFSET is used to determine the offset resistance of the tester. When a GND OFFSET is performed, the reference is automatically set to the measured resistance.
Note
GND OFFSET setting is only applicable to IR test.
Steps
1. Press the PAGE soft-key to move to
the 2/2 page where GND OFFSET setting appears for IR testing.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the GND OFFSET setting. When selecting ON, the ZERO CHECK indicator will be shown on the display.
GND OFFSET cursor ZERO CHECK indicator
Setting GND OFFSET
66
OPERATION
3. Press the START button to perform
the GND OFFSET. The resistance of the tester, after the GND OFFSET has finished, will be added into the REF VALUE field as the display shown below.
Resistance of the tester
67
GPT-10000 Series User Manual
Background
Basically, GB test has no ramp up time and thus starts from the set test time by user directly. However, due to some cases where a buffer time before test time is in fact required for GB test, e.g., in conveyor where DUTs are tested for GB by batches and certain buffer duration needed for test leads or jigs connecting with DUTs, the GB CONTACT setting practically allows user to apply to customized scenarios when necessary occurs.
Note
GB CONTACT setting is only applicable to GB test.
Steps
1. Press the UP / DOWN arrow soft-
keys to bring the cursor to the GB CONTACT setting.
GB CONTACT cursor
2. Use the scroll wheel to set the
value of GB CONTACT
GB CONTACT
000.0 s ~ 999.9 s
Setting GB Contact
68
OPERATION
GB CONTACT Duration Indicator
After every parameter including GB CONTACT is well set, press START to begin the GB test. A section at the lower right corner of display shows the counting duration of GB CONTACT, which will run to the set value followed by the test time. See the screenshot shown below.
GB CONTACT duration indicator
69
GPT-10000 Series User Manual
Background
The Zeroing function is used to determine the resistance of the test leads for GB and CONT tests. When a ZERO CHECK is performed, the reference is automatically set to the measured resistance of the test leads.
Note
ZERO CHECK setting is only applicable to both GB and CONT tests.
Steps
1. Short the positive and negative alligator clips as
shown below.
2. Press the PAGE soft-key to move to
the 2/2 page where ZERO CHECK setting appears for GB testing.
As for CONT, ZERO CHECK setting shows in the 1/1 page directly.
3. Press the UP / DOWN arrow soft-
keys to bring the cursor to the ZERO CHECK setting. When selecting ON, the ZERO CHECK indicator will be shown on the display.
Zero Check for the Test Leads
70
OPERATION
ZERO CHECK cursor ZERO CHECK indicator
4. Press the START button to perform
the zero check. The resistance of the test leads, after the ZERO CHECK has finished, will be added into the REF VALUE field as the display shown below.
Resistance of the test leads
Note
Remember to replace the test leads to the proper position on the DUT before testing.
71
GPT-10000 Series User Manual
FAIL – GBI LOW
If SOURCE H/L terminals are open or poorly connected, the FAIL – GBI LOW status will appear on the screen. Please re-check the connection of SOURCE H/L terminals again.
FAIL – GBI LOW status
REF VALUE = 0
Press STOP button to exit and the resistance of test leads were not properly added into the REF VALUE, which shows 000.0 mΩ as shown below. Re-check the connection of SOURCE H/L terminals and press START button again to proceed to the ZERO CHECK procedure.
REF VALUE = 0
72
OPERATION
Background
When GROUND MODE is set to ON, the GPT­10000 grounds the return terminal to the ground. This mode is best for DUTs that are grounded to an earth ground by their chassis, fixtures or operation environment. This mode measures the potential of the HIGH VOLTAGE terminal with respect to earth ground. This means that additional noise which leaks to earth ground will also be measured. This is the safest testing mode, though potentially not as accurate.
When GROUND MODE is set to OFF, the return terminal is floating with respect to the earth ground. This mode is for DUTs that are floating and not directly connected to an earth ground. This is more accurate than when GROUND MODE is set to ON as less noise will be measured. For this reason, this testing mode is able to measure with better stability.
ACW/DCW, GROUND MODE ON, DUT grounded
Setting the Grounding Mode
73
GPT-10000 Series User Manual
ACW/DCW, GROUND MODE OFF, DUT floating
Note
If the current value, which results from comparison between I1 and I2 current, is above 3mA, once user unexpectedly touches the DUT, the GFCI, Ground Fault Circuit Interrupter, function activates and output will be stopped immediately so that protection mechanism will be well triggered at once.
IR, GROUND MODE ON, DUT grounded
74
OPERATION
IR, GROUND MODE OFF, DUT floating
GB, GROUND MODE ON, DUT grounded
GB, GROUND MODE OFF, DUT floating
75
GPT-10000 Series User Manual
Cont., GROUND MODE ON, DUT grounded
Note
In terms of Continuity test, it is compulsory to ground the DUT and thus GROUND MODE is ON.
Warning
When GROUND MODE is set to OFF, the DUT, fixtures or connected instrumentation cannot be grounded. This will short circuit the internal circuitry during a test.
For ACW and DCW tests, if it is not known whether the DUT test setup is grounded or not, always set GROUND MODE to ON.
Only set GROUND MODE to OFF when the DUT is floating electrically.
Steps
1. Press the PAGE soft-key to move to
the 2/3 page where GROUND MODE setting appears for ACW and DCW.
As for IR and GB, the GROUND MODE setting shows in the 1/2 page.
2. Press the UP / DOWN arrow soft-
keys to bring the cursor to the GROUND MODE setting.
76
OPERATION
GROUND MODE cursor
3. Use the scroll wheel to set the
GROUND MODE.
GROUND MODE
OFF, ON
4. The GROUND MODE icon on the display
changes accordingly.
GROUND MODE ON
GROUND MODE OFF
Note
Under the IR test mode, when GROUND MODE is ON but test time is se t < 0.5s, the error message TEST TIMR<0.5swill be shown, by which user is not able to start the IR test mode unless the test time is reset to > 0.5s. Refer to page 45 for how to set the test time manually.
GROUND MODE ON Error Message Prompts
0.3 s Test Time
77
GPT-10000 Series User Manual
Background
The CONTACT CHK function is used to determine if open circuit or short circuit occurs between the test leads and DUT under the ACW, DCW and IR tests. Before activating this function, it is first required to define a reference value along with relevant thresholds, for which refer to page 161.
Note
CONTACT CHK setting is only applicable to ACW, DCW and IR test modes.
Steps
1. After well setting up the test leads connection
with DUT, refer to page 161 for how to define a reference value and relevant thresholds firstly.
2. Press the PAGE soft-key to move to
the 3/3 page where CONTACT CHK setting appears for ACW, DCW and IR tests.
3. Press the UP / DOWN arrow soft-
keys to bring the cursor to the CONTACT CHK and turn it ON.
CONTACT CHK ON
Setting Contact Check
78
OPERATION
4. After pressing the START button,
the GPT-10000 unit will perform the CONTACT CHK before running a MANU test. If the measured current is lower than the reference value by user-defined
percentage, the “OPEN” status
appears on the screen. While the measured current is higher than the reference value by user-defined
percentage, the “SHORT” status
appears instead.
OPEN Status
OPEN Status detected
SHORT Status
SHORT Status detected
79
GPT-10000 Series User Manual
Background
A test can be run when the tester is in READY status.
Note
The tester cannot start to run a test under the following conditions:
A protection setting has been tripped; when a
protection setting has been tripped the corresponding error message is displayed on the screen. See page 238 for a comprehensive list of the all the setting errors.
The INTERLOCK function is ON and the
Interlock key is not inserted in the signal I/O port (page 130).
The STOP signal has been received remotely.
If Double Action is ON, ensure the START
button is pressed immediately after the STOP button (<0.5s).
Note
When a test is running the voltage output cannot be changed, unless the test is under the special manual mode. See page 90 for details.
Steps
1. Ensure the tester is in READY
status for the test to come.
Page 33
Running a MANU Test
80
OPERATION
READY status
2. Press the START button when the
tester is in the READY status. The manual test starts accordingly and the tester goes into the TEST status.
TEST status
3. The test will start by showing the ongoing ramp
up time followed by the ongoing test time and the ongoing ramp down time. The test will continue until the test is finished or stopped.
81
GPT-10000 Series User Manual
RAMP UP TIME
Ongoing RAMP UP TIME
TEST TIME
Ongoing TEST TIME
RAMP DOWN TIME
Ongoing RAMP DOWN TIME
Note
RAMP DOWN time only appears when user has activated it. See page 49 for details.
82
OPERATION
ACW Example
Measured Current
Test Voltage
DCW Example
Measured Current
Test Voltage
IR Example
Measured Resistance
Test Voltage
83
GPT-10000 Series User Manual
GB Example
Measured Resistance
Test Current
CONT Example
Measured Resistance
Test Current
Stop the Test
1. To stop the test at any time when it
is running, press the STOP button. The test will stop immediately. When the STOP button is pressed, a judgment is not made and the tester will restore to READY status.
Note
Do not touch any terminals, test leads or any other connections when the test is on.
84
OPERATION
Background
If the test is allowed to run to completion (the test is not stopped or a protection setting is not tripped) then the tester will judge the test as either PASS or FAIL.
Note
The test will be judged PASS when:
The HI SET and LO SET limits have not been
tripped during the test time.
The test will be judged FAIL when:
Either the HI SET or LO SET limit has been
tripped during the test time.
A protection setting has been tripped during
the test time. See page 238 for a list of error messages.
PASS Judgment
1. When the test is judged as PASS,
PASS will be displayed on screen, the buzzer will sound and the PASS indicator will be lit green.
PASS Judgment
PASS / FAIL MANU Test
85
GPT-10000 Series User Manual
2. The tester will immediately restore back to the
READY status after PASS judgment. However, if the PASS HOLD is activated, PASS judgment will persist until the set duration of PASS HOLD is fully met. Refer to page 63 for details.
In addition, pressing the STOP button during the set duration of PASS HOLD can return to READY status immediately.
Note
The START button is disabled when the buzzer is beeping.
PASS Timing Diagrams
The timing diagrams below show the ACW, DCW, IR, GB and CONT timing for the START status, TEST status and PASS judgment.
ACW PASS Timing
RAMP DOWN & DISCHARGE
Output V
time
START TEST PASS
RAMP UPTEST TIME
DCW PASS Timing
Output V
time
START TEST PASS
RAMP UPTEST TIME
RAMP DOWN & DISCHARGE
86
OPERATION
IR PASS Timing
Output V
time
START TEST PASS
RAMP UPTEST TIME
RAMP DOWN & DISCHARGE
GB PASS Timing
START
PASS
Output I
TEST
Contact
RAMP Test
time
Time0.1s 0.1s
UP Time
CONT PASS Timing
Output I
time
START TEST PASS
Test Time
FAIL Judgment
1. When the test is judged as FAIL,
FAIL will be displayed on screen, the buzzer will sound and the FAIL indicator will be lit red.
As soon as a test is judged FAIL, power is cut from the terminals.
87
GPT-10000 Series User Manual
FAIL Judgment
2. The FAIL judgment will be held on
the display until the STOP button is pressed. Pressing the STOP button will return the tester back to the READY status.
FAIL Timing Diagrams
The timing diagrams below show the ACW, DCW, IR, GB and CONT timing for the START status, TEST status and FAIL judgment.
ACW FAIL Timing
DISCHARGE
Output V
time
START TEST FAIL
RAMP UPTEST TIME
88
OPERATION
DCW FAIL Timing
DISCHARGE
Output V
time
START TEST FAIL
RAMP UPTEST TIME
IR FAIL Timing
DISCHARGE
Output V
time
START TEST FAIL
RAMP UPTEST TIME
GB FAIL Timing
START
FAIL
Output I
TEST
Contact
RAMP Test
time
Time0.1s 0.1s
UP Time
CONT FAIL Timing
Output I
time
START TEST FAIL
Test Time
89
GPT-10000 Series User Manual
Special Test Mode Overview
When MANU number 000 is selected, the special test mode is activated. Under the special test mode, the voltage can be changed during a test in real time (ACW, DCW only). The test function can also be changed when in READY status, unlike under normal operation.
Separate settings can be saved under the special test mode for each of the testing functions: ACW, DCW, IR, GB and CONT. This means different test setups for ACW, DCW, IR, GB and CONT can be saved within the MANU number 000 concurrently.
Steps
1. Choose MANU number 000 to
enter the special test mode.
Page 41
2. The settings of a previous test can
be loaded by pressing the corresponding soft-keys on the front panel.
For example, if you are currently in DCW mode, pressing the ACW key will load the ACW settings that were previously stored in the special manual mode.
Special MANU Test Mode (000)
90
OPERATION
3. Set all the necessary parameters for
a test and save.
Note: A different test setup can be saved for each test function (ACW, DCW, IR, GB and CONT). Below is an example of ACW function in special manual mode.
Pages 42 ~ 73
Special MANU Number 000
Running the Test
1. In special test mode (000), tests are
started and stopped in the same way as for the normal manual test mode. See page 78 for details.
2. If required, the scroll wheel can be
used to set the voltage level in real­time as the test is running under either ACW or DCW mode.
ACW DCW
0.050kV ~ 5kV
0.050kV ~ 6kV
Results
Test judgments are the same as those for the normal manual tests. Please see the PASS/FAIL MANU Test section for details.
Page 85
91
GPT-10000 Series User Manual
Sweep Function Overview
The GPT-10000 Series has access to the sweep mode function, which creates a graph of one of the ACW, DCW, IR, GB or CONT tests in either Manual test or the special MANU mode. The graph will plot the output voltage, current or resistance versus time. After the test has been completed, the test current, voltage or resistance at any point in time can be fetched and viewed in the graph.
Below is an example of the resultant sweep plot of a DCW test where a DC voltage is ramped up to a user-defined level until the HI SET current level has been tripped or the test time runs out.
Test V Meas. I
time
Start time
RAMP TIME
Legend: Voltage: Current
TEST time
Voltage
Current
The test items that are plotted on the sweep graph depend on the type of test that is performed.
TEST
Graph Test Items
ACW
Measured voltage, measured current (V, I)
DCW
Measured voltage, measured current (V, I)
IR
Measured voltage, measured resistance (V, R)
GB
Measured current, measured resistance (I, R)
CONT
Measured current, measured resistance (I, R)
Sweep Function
92
OPERATION
Steps of View Sweep Graph
1. When a test has finished, press
the corresponding button, e.g., DCW button for DCW test, to view the result of the sweep in an intuitive graph.
Graph Test Items:
TEST
GREEN
BLUE
ACW
Measured voltage
Measured current
DCW
Measured voltage
Measured current
IR
Measured voltage
Measured resistance
GB
Measured current
Measured resistance
CONT
Measured current
Measured resistance
DCW Sweep Graph Example
Cursor
Measured Current
(READ I in Blue)
Measured Voltage
(READ V in Green)
The values of point by cursor
93
GPT-10000 Series User Manual
2. Use the scroll wheel to move the
cursor on the time axis (red highlight in x-axis). The measured values on the green and blue lines at that particular point in time are shown within the table below (orange highlight). Also, the test function along with the test number is clearly shown within the table. The HI in y-axis (yellow highlight) along with the dotted line in red indicates the HI SET value and the point of tripped time.
Turn Pages
3. The resultant graph will be over 1 page
when test time is beyond 650 steps (the interval of each step is 0.1s). In this case, press PAGE soft-key to switch among each page for full graphs.
Fast-forward
4. Press the MOVE soft-key before
moving the cursor to fast-forward steps by 10 times (x 10), which is practical when steps are many. Press the MOVE soft-key again to return back to the normal “x 1” speed.
Exit the Results Graph
To exit the sweep graph, press the corresponding button again to return back to Manual test.
94
OPERATION
Automatic Tests
This section describes how to create, edit and run automatic tests. Automatic tests allow you to link up to 10 different MANU tests and run them sequentially within a single AUTO test. Each stored MANU test is used as a test step when creating an AUTO test. In addition, up to 5 groups of AUTO test can be interconnected together to present an ever-advanced AUTO tests.
Choose/Recall an AUTO Test→ from page 96 Creating an AUTO Test File Name → from page 97 Adding a Step to the AUTO Test → from page 98 Continuous AUTO Tests → from page 100 AUTO Test Page Editing → from page 102 Running an Automatic Test → from page 107 Automatic Test Results → from page 113
Before operating the GPT-10000 please read the safety precautions as outlined in the Set Up chapter on page 22.
95
GPT-10000 Series User Manual
Background
The tester must first be put into AUTO mode to create or run automatic tests.
Up to 100 automatic tests can be saved or recalled.
Steps
1. If the tester is in MANU or
SYSTEM mode, press the AUTO key on the front panel. This will put the tester into Auto mode.
2. After entering the AUTO mode,
first use the scroll wheel to choose the AUTO number.
AUTO #
001~100
AUTO number cursor READY status
Note
The AUTO number can only be chosen in READY status. If the status is either PASS or FAIL, press the STOP button to restore back to the READY status.
Choose/Recall an AUTO Test
96
OPERATION
Background
Each automatic test can have a user-defined test file name (Default: AUTO_NAME) up to 10 characters long. See the character list below for the allowed characters.
Character List
0 1 2 3 4 5 6 7 8 9
W X Y ZM VO P RQ T USNLA B C D E F G H I J K
w x y zm vo p rq t usnla b c d e f g h i j k
_
Steps
1. Use the LEFT/RIGHT arrow soft-
keys to move the cursor to the AUTO_NAME (default name) field. The characters table will appear in the right hand accordingly.
AUTO name cursor Character Table
2. Use the scroll wheel to scroll
through the available characters.
3. Press the LEFT / RIGHT arrow soft-
keys to move the cursor to the next character.
Creating an AUTO Test File Name
97
GPT-10000 Series User Manual
4. The AUTO test file name is set when the
current AUTO test is saved or when the cursor is moved to another setting.
Background
Up to 10 MANU tests (steps) can be added to an automatic (AUTO) test. Each step is added in a sequential order.
Steps
1. Press the DOWN arrow key to
bring the cursor to the MANU STEP number.
MANU STEP number cursor
2. Use the scroll wheel to choose a
MANU STEP number to add to the automatic test.
MANU STEP number
001~100, CON
CON
It indicates that this group of AUTO test can be connected with the next group. Refer to page 100 for more details.
Adding a Step to the AUTO Test
98
OPERATION
3. Further press the DOWN arrow
key followed by using the scroll wheel to choose another MANU STEP number to add to the automatic test.
MANU STEP number cursor (2nd)
4. Repeat the previous steps for any other MANU
tests that you wish to add to the automatic test.
99
GPT-10000 Series User Manual
Background
As mentioned previously, up to 10 MANU steps can be grouped to form an AUTO test and user can designate each step from MANU step number 1 to 100 for an AUTO test. However, it is available to interconnect different AUTO tests together to present a series of AUTO tests.
Steps
1. Follow the steps of “Adding a Step to the
AUTO Test” in page 98 first. See the example below where 5 MANU steps have been added into the AUTO-001 group.
2. Press the DOWN arrow key to
bring the cursor to the next MANU STEP field followed by using the scroll wheel to choose CON from the MANU STEP options.
CON is chosen from MENU STEP
Continuous AUTO Tests
100
Loading...