Instek GPD-3303D, GPD-3303S Instruction Manual

Distributed By TestEquity - www.testequity.com - 800-732-3457
DC Power Supply
GPD-3303 Series
USER MANUAL
GW INSTEK PART NO.
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.
TABLE OF CONTENTS
Distributed By TestEquity - www.testequity.com - 800-732-3457
Table of Contents
GPD-3303 Series User Manual
Error Messages ....................................27H38
Command List......................................28H39
Command Details ................................ 29H40
SAFETY INSTRUCTIONS .................................................. 5
OVERVIEW....................................................................... 9
Introduction........................................... 9
Series Lineup / Main Features ............. 11
Principle of Operation.......................... 12
Front Panel Overview........................... 14
Rear Panel Overview ............................ 17
CV/CC Crossover Characteristics ......... 18
SETUP ........................................................................... 19
Power Up ............................................. 19
Load Cable Connection ........................ 20
Output On/Off..................................... 21
Beep On/Off ........................................ 21
Front Panel Lock ..................................22
OPERATION .................................................................. 23
CH1/CH2 Independent Mode .............. 23
CH3 Independent Mode....................... 25
CH1/CH2 Tracking Series Mode........... 27
CH1/CH2 Tracking Parallel Mode ........19H32
FAQ................................................................................30H45
APPENDIX......................................................................31H46
Fuse Replacement................................ 32H46
Specifications ......................................33H47
Declaration of Conformity.................... 34H49
INDEX ............................................................................35H50
SAVE/RECALL SETUP ..................................................... 20H34
Save Setup ...........................................21H34
Recall Setup......................................... 22H35
REMOTE CONTROL....................................................... 23H36
Remote Control Setup.......................... 24H36
Remote Connection Step...................... 25H37
Command Syntax ................................. 26H38
3
4
SAFETY INSTRUCTIONS
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow when operating the GPD­3303 series and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition for the GPD­3303 series.

Safety Symbols

These safety symbols may appear in this manual or on the GPD-3303 series.
WARNING
CAUTION
Warning: Identifies conditions or practices that could result in injury or loss of life.
Caution: Identifies conditions or practices that could result in damage to the GPD-3303 series or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal

Safety Guidelines

General Guidelines
CAUTION
Power Supply
WARNING
Do not place any heavy object on the GPD-3303
series.
Avoid severe impacts or rough handling that
leads to damaging the GPD-3303 series.
Do not discharge static electricity to the GPD-
3303 series.
Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at circuits directly
connected to Mains (see note below).
Do not disassemble the GPD-3303 series unless
you are qualified as service personnel.
(Measurement categories) EN 61010-1:2001 specifies the measurement categories and their requirements as follows. The GPD-3303 series falls under category I.
Measurement category IV is for measurement performed at the source of low-voltage installation.
Measurement category III is for measurement performed in the building installation.
Measurement category II is for measurement performed on the circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on circuits not directly connected to Mains.
AC Input voltage: 100V/120V/220V/230V
±10%, 50/60Hz
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid electrical shock.
Earth (ground) Terminal
Fuse
Fuse type: 100V/120V: T6.3A/250V,
220V/230V: T3.15A/250V
WARNING
Make sure the correct type of fuse is installed
before power up.
5
6
SAFETY INSTRUCTIONS
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
Cleaning the GPD-3303 series
Operation Environment
To ensure fire protection, replace the fuse only
with the specified type and rating.
Disconnect the power cord before fuse
replacement.
Make sure the cause of fuse blowout is fixed
before fuse replacement.
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals or cleaners containing
harsh products such as benzene, toluene, xylene, and acetone.
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (note below)
Relative Humidity: < 80%
Altitude: < 2000m
Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2001 specifies the pollution degrees and their requirements as follows. The GPD-3303 series falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non- conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor
Relative Humidity: < 70%
Temperature: 10°C to 70°C

Power cord for the United Kingdom

When using the GPD-3303 series in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the following code:
Green/ Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with the letter E or by the earth symbol Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.
or coloured Green or
2
should be protected by a 3A
7
8
OVERVIEW
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
OVERVIEW
This chapter describes the GPD-3303 series in a nutshell, including its main features and front / rear panel introduction. After going through the overview, follow the Setup chapter (page19) to properly pow

Introduction

Overview
Independent / Trac ki ng S er ie s / Tracking Parallel
GPD-3303, the regulated DC power supply series, are light weight, adjustable, multifunctional work stations. They have three independent outputs: two with adjustable voltage level and one with fixed level selectable from 2.5V, 3.3V and 5V. The GPD-3303 series can be used for logic circuits where various output voltage or current are needed, and for tracking mode definition systems where plus and minus voltages with insignificant error are required.
The three output modes of GPD-3303 series, independent, tracking series, and tracking parallel, can be selected through pressing the TRACKING key on the front panel. In the independent mode, the output voltage and current of each channel are controlled separately. The isolation degree, from output terminal to chassis or from output terminal to output terminal, is 300V. In the tracking modes, both the CH1 and CH2 outputs are automatically connected in series or parallel; no need to connect output leads. In the series mode, the output voltage is doubled; in the parallel mode, the output current is doubled.
er up and set operation environment.
Constant Voltage/ Constant Current
matic
Auto tracking mode
Dynamic load
Except for CH3, each output channel is completely transistorized and well-regulated, and works in constant voltage (CV) or constant current (CC) mode. Even at the maximum output current, a fully rated, continuously adjustable output voltage is provided. For a big load, the power supply can be used as a CV source; while for a small load, a CC source. When in the CV mode (independent or tracking mode), output current (overload or short circuit) can be controlled via the front panel. When in the CC mode (independent mode only), the maximum (ceiling) output voltage can be controlled via the front panel. The power supply will automatically cross over from CV to CC operation when the output current reaches the target value. The power supply will automatically cross over from CC to CV when the output voltage reaches the target value. For more details about CV/CC mode operation, see page18.
The front panel display (CH1, CH2) shows the output voltage or current. When operating in the tracking mode, the power supply will automatically connect to the auto- tracking mode.
When used in audio production lines, the power supply will provide a continuous or dynamic load connector. When the connectors are connected to the position “ON”, a stable DC current power will be provided for audio power amplifiers.
9
10
OVERVIEW
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

Series Lineup / Main Features

Series Lineup

Model V Meter A Meter USB Tracking Error
GPD-3303D 3 digit 3 digit Yes
GPD-3303S 5 digit 4 digit Yes

Main Features

Performance
Operation
Low noise: Cooling fan controlled by Heatsink
temperature
Compact size, light weight
Constant Voltage / Constant Current operation
Tracking Series / Tracking parallel operation
Output On/Off control
3 outputs: 30V/3A x 2, 2.5V/3.3V/5V/3A x 1
Digital panel control
4 sets of panel setup save/recall
0.5% + 50mV of Master
0.5% + 10mV of Master

Principle of Operation

Overview
Block diagram
The power supply consists of the following.
AC input circuit
Transformer
Bias power supply including rectifier, filter,
pre-regulator and reference voltage source
Main regulator circuit including the main
rectifier and filter, series regulator, current comparator, voltage comparator, reference voltage amplifier, remote device and relay control circuit
The block diagram below shows the circuit arrangement. The single phase input power is connected to the transformer through the input circuit. Details of each part are described in the next page.
Coarse and fine Voltage/Current control
Software calibration
Buzzer output
Key lock function
Protection
Interface
Overload protection
Reverse polarity protection
USB for remote control
11
12
OVERVIEW
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
Rectifier
Main Rectifier
Current Limiter
Overload
The auxiliary rectifiers D1011~ D1014 provide bias voltage filtered by the capacitors C102 and C103, for the pre-regulators U101 and U102. They provide a regulated voltage for other modules.
The main rectifier is a full wave bridge rectifier. It provides the power after the rectifier is filtered by the capacitor C101, and then regulated via a series­wound regulator, which is finally delivered to the output terminal.
U104 acts as a current limiter. When the current is over predetermined rating, U104 is activated and decreases the current. U208 provides a reference voltage. U206 is the inverter amplifier. U103 is a comparator amplifier which compares reference voltage and feedback voltage, and then delivers to Q103, Q104, which then calibrates the output voltage.
When the unit is overloaded, Q107 activates to control the current magnitude of Q104, to limit the output current. The relay control circuit controls the power dissipation in the series-wound regulated circuit.

Front Panel Overview

Display
Vol tMeter
GPD-3303S (5 digits)
Displays CH1 or CH2 output voltage.
GPD-3303D (3 digits)
13
AmpMeter
GPD-3303S (4 digits)
14
Displays CH1 or CH2 output current.
GPD-3303D (3 digits)
OVERVIEW
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
Control Panel
Memory Keys
CH1/CH2/Beep
ys
Ke
Parallel/Se
ries
Keys
Lock Key
Output Key
Saves or recalls panel settings. Four settings, 1 ~ 4, are available. For save/recall details, see page34.
Selects the output channel for level adjustment. For level setting details, see page23. Pressing and holding CH2 key enab
les beep sound. For
details, see page21.
Activates Tracking Parallel operation or Tracking Series operation, For details, see page27.
Locks or unlocks the front panel settings. For details, see page
42H22.
Turns the output on or off.
Te rm in al s
Default Terminals
GND Terminal
CH1 Output
CH1 CV/CC Indicator
CH2 Output
European Terminals
Accepts a grounding wire.
Outputs CH1 voltage and current.
Indicates CH1 Constant Voltage or Constant Current state.
Outputs CH2 voltage and current.
Vol t a g e K nobs
Adjusts the output voltage level for CH1 or CH2. Pressing the knob switches coarse and fine level setting.
Current Knobs
Adjusts the output current level for CH1 or CH2. Pressing the knob switches coarse and fine level setting.
Power Switch
Turns On
or Off the main power.
For power up sequence, see page
43H19.
15
CH2 CV/CC/PAR Indicator
CH3 Output
CH3 Overload Indicator
CH3 Voltage Selector
16
Indicates CH2 Constant Voltage, Constant Current, or Tracking Parallel operation mode.
Outputs CH3 voltage and current.
Indicates when CH3 output current is overloaded.
Selects CH3 output voltage: 2.5V,
3.3V, or 5V.
OVERVIEW
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

Rear Panel Overview

USB Connector AC SelectorCooling Fan
USB Connector
Power Cord / Fuse Socket
Accepts a USB slave connector for command-based remote control (page36).

CV/CC Crossover Characteristics

Background
CV mode
CC mode
Diagram
The GPD-3303 series automatically switches between constant voltage mode (CV) and constant current mode (CC), according to load condition.
When the current level is smaller than the output
setting, the GPD-3303 series operates in Constant Voltage mode. The indicator on the front panel
turns green (C.V.) The Voltage level is kept at the setting and the Current level fluctuates according to the load condition until it reaches the output current setting.
When the current level reaches the output setting, the GPD-3303 series starts operating in Constant Current mode. The indicator on the front panel turns red (C.C.) The Current level is kept at the setting but the Voltage level becomes lower than the setting, in order to suppress the output power level from overload. When the current level becomes lower than the setting, the GPD-3303 series goes back to the Constant Voltage mode.
Vout
Power Co Fuse Socket
rd /
The power cord socket accepts the AC mains: 115V/230V, 50/60Hz. For power up details, see page19.
The
fuse holder contains the AC main fuse. For fuse replacement details, see page46.
AC Sele
ctor
Selects AC voltage: 100V/ 120V/ 220V/ 230V.
17
18
Vmax
Constant
Voltage
Constant
Current
Imax
Iout
SETUP
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
SETUP
This chapter describes how to properly power up and configure the GPD-3303 series before operation.

Power Up

Select AC voltage
Connect AC power cord
Before powering up the power supply, select the AC input voltage from the rear panel.
Connect the AC power cord to the rear panel socket.

Load Cable Connection

GTL-104
GTL-105
GTL-203, 204
1. Turn the terminal counterclockwise and loosen the screw.
2. Insert the cable terminal.
3. Turn the terminal clockwise and tighten the screw.
Insert the plug into the socket.
Insert the plug into the terminal.
Power On
Power Off
Press the Power switch to turn on the power. The display shows the initialization screen with the model name, followed by the last recalled settings.
Press the Power switch again to turn off the power.
Wire type
Wire size (AWG) Maximum current (A)
20 2.5
16 6
19
12 16
20
When using load cables other than the attached, make sure they have enough current capacity for minimizing cable loss and load line impedance. Voltage drop across a wire should not excess 0.5V. The following list is the wire current rating at 450A/cm
18 4
14 10
2
.
SETUP
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

Output On/Off

Panel operation
Automatic output off
Pressing the Output key turns on all CH 1/2/3 outputs.
The key LED also turns on. Pressing the Output key again turns the output and the key LED off.
Any of the following actions during output on automatically turns it off. They might involve sudden and harmful change in the output level.
Change the operation mode between
independent / tracking series / tracking parallel
Recalling other setups from the memory
Storing the setup into the memory

Beep On/Off

Panel operation
By default, the beep sound is enabled. To turn off the beep, press the CH2 key for 2 seconds.
A beep comes out and the beep setting will be turned off. To enable the beep, press the CH2 key again for 2 seconds.
Panel operation
Press the LOCK key to lock the front panel key operation. The key LED
turns on. To unlock, press the LOCK key for 2 seconds. The key LED also turns off.
Note
The OUTPUT key is not affected by the lock operation.

Front Panel Lock

List of beep
The following operations beep when the beep setting is on.
Power on
INDEP – SER – PARA
mode switching
Setup save/recall
Voltage/current knob
fine/coarse switching
Output on/off
Panel lock/unlock
CH1/CH2 output
level knob switching
Voltage/current level
reaching minimum (zero) level
21
22
OPERATION
Distributed By TestEquity - www.testequity.com - 800-732-3457
OPERATION

CH1/CH2 Independent Mode

Background / Connection
CH1 and CH2 outputs work independent of each other.
GPD-3303 Series User Manual
Note: this diagram shows non-European terminals.
3. Set the CH1 output voltage and current. Press the CH1 key (LED turns on) and then use the Voltage and Current knob. By default, the Voltage and Current knob work in the coarse mode. To activate the fine mode, press the knob to turn the FINE LED on.
Coarse: 0.1V or 0.1A @ rotation click
Fine: the smallest digit @ rotation click
(For CH1)
(Fine control)
Output rating
Panel operation
0 ~ 30V/0~3A for each channel
1. Make sure the PARA/ INDEP and SER/INDEP keys are turned off (the key LEDs are off).
2. Connect the load to the front panel terminals, CH1 +/, CH2 +/−.
23
24
4. Repeat the above settings for the CH2.
5. To turn on the output, press the output key. The key LED turns on and the CH1 / CH2 indicator shows the output mode, CV or CC.
(CH1)
(CH2)
OPERATION
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

CH3 Independent Mode

Background / Connection
Output rating
The CH3 rating is 2.5V/3.3V/5V, 3A fixed. It works independently from CH1 and CH2, regardless of their modes.
2.5V/3.3V/5V, 3A fixed
CV CC
3. To turn on the output, press the output key. The key LED turns on.
When the output Current level exceeds 3A, the overload indicator turns red and CH3 operation mode switches from Constant Voltage to Constant Current.
Note: “overload” in this case does not mean an abnormal operation.
No Tracking Series/Parallel
Panel operation
CH3 does not have tracking series/parallel mode. Also, CH3 output is not affected by CH1 and CH2 modes.
1. Connect the load to the front panel CH3 +/ terminal. (the diagram shows non­European terminals)
2. Select the output voltage,
2.5V/3.3V/5V using the CH3 voltage selector key.
25
26
OPERATION
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

CH1/CH2 Tracking Series Mode

Background

Tracking series without common terminal

Connection
Tracking series operation doubles the Voltage capacity of the GPD-3303 series by internally connecting CH1 (Master) and CH2 (Slave) in serial and combining the output to a single channel. CH1 (Master) controls the combined Voltage output level.
The following describes two type of configurations depending on the common ground usage.
2. Connect the load to the front panel terminals, CH1+ & CH2− (Single supply).
Note: this diagram shows non-European terminals.
3. Press the CH2 key (LED turns on) and then use the Current knob to set the CH2 output current to the maximum level (3.0A). By default, the Voltage and Current knob work in the
(Fine control)
coarse mode. To activate the fine mode, press the knob to turn the FINE LED on.
Coarse: 0.1V or 0.1A @ rotation click
Fine: the smallest digit @ rotation click
Output rating
0 ~ 60V/0 ~ 3A
1. Press the SER/INDEP key to activate the tracking series mode. The key LED turns on.
27
4. Press the CH1 key (LED turns on) and then use the Voltage and Current knob to set the output voltage and current level.
5. To turn on the output, press the output key. The key LED turns on.
6. Refer to the CH1 (Master) meter and indicator for the output setting level and CV/CC status.
28
OPERATION
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
Voltage level
Double the reading on the CH1 Voltage meter. In the above case, the actual output is 20.0 x 2 = 40.0V.
Current level
CH1 meter reading shows the output Current. In the above case, 2.000A. (CH2 Current control must be in the Maximum position=3.0A).

Tracking series with common terminal

Connection
1. Press the SER/INDEP key to activate the tracking series mode. The key LED turns on.
2. Connect the load to the front panel terminals, CH1+ & CH2. Use the CH1 () terminal as the common line connection.
Note: this diagram shows non-European terminals.
3. Press the CH1 key (LED turns on) and use the Voltage knob to set the master & slave
(master & slave)
output voltage (the same level for both channels). By default, the Voltage and Current knob work in the
(Fine control)
coarse mode. To activate the fine mode, press the knob to turn the FINE LED on.
Coarse: 0.1V or 0.1A @ rotation click
Fine: the smallest digit @ rotation click
Output rating
0~30V/0~3A for CH1 ~ COM
0~–30V/0~3A for CH2 ~ COM
4. Use the Current knob to set the master output current.
29
30
5. To turn on the output (and LED), press the output key.
OPERATION
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
Master (CH1)
Master (CH1)
6. For the master (CH1) output level and CV/CC status, refer to the CH1 meter and indicator.
CH1 meter reading shows the
voltage level
current level
7. Press the CH2 key (LED turns on) and use the Current knob to set the slave output current.
8. For the slave (CH2) output level and CV/CC status, refer to the CH1/CH2 meter and CH2 indicator.
output voltage. In the above case,
20.0V.
CH1 meter reading shows the output current. In the above case,
2.000A.

CH1/CH2 Tracking Parallel Mode

Background / Connection
Output rating
Tracking parallel operation doubles the current capacity of the GPD-3303 series by internally connecting CH1 and CH2 in parallel and combining the output to a single channel. CH1 controls the combined output.
0 ~ 30V/0 ~ 6A
1. Press the PARA/INDEP key to activate the tracking parallel mode. The key LED turns on.
Slave (CH2)
voltage level
Slave (CH2)
current level
The CH2 meter reading shows the output voltage. In the above case, 20.0V.
The CH2 meter reading shows the output current. In the above case, 3.000A.
31
32
2. Connect the load to the CH1 +/ terminals.
Note: this diagram shows non-European terminals.
OPERATION
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
3. To turn on the output, press the output key. The key LED turns on.
4. The CH2 indicator turns red, indicating tracking parallel (PARA) mode.
5. Press the CH1 key (LED turns on) and then use the Voltage and Current knob to set the output voltage and current. The CH2 output control is disabled. By default, the Voltage and Current knob work in the coarse mode. To activate the fine mode, press the knob to turn the FINE LED on.
6. For the output level and CV/CC status, refer to the CH1 meter and indicator.
(Fine control)
SAVE/RECALL SETUP

Save Setup

Background
Contents
The front panel settings can be stored into one of the four internal memories.
The following list shows the setup contents.
Independent / tracking series / tracking parallel
mode
CH1/CH2 knob selection
Fine/coarse editing mode
Output voltage/current level
The following settings are always saved as “off”.
Output on/off
Front panel lock/unlock
Voltage level
Current level
The CH1 meter reading shows the output voltage. In the above case, 20.0V.
Double the amount of CH1 current meter reading. In the above case, 2.0A x 2 = 4.0A.
33
Panel operation
Note
34
Press one of the 1~4 Memory keys for 2 seconds, for example memory 1. The panel settings are saved in memory 1 and the key LED turns on. When the panel settings are modified, the LED turns off.
When a setting is stored, the output automatically turns off.
SAVE/RECALL SETUP
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

Recall Setup

Background
Contents
Panel operation
The front panel settings can be recalled from one of the four internal memories.
The following list shows the setup contents.
Independent / tracking series / tracking parallel
mode
CH1/CH2 knob selection
Fine/coarse editing mode
Output voltage/current level
The following settings are always recalled as “off”.
Output on/off
Front panel lock/unlock
Press one of the 1~4 Memory keys, for example memory 1. The panel settings saved in memory 1 are recalled. The key LED turns on. When the panel settings are modified, the LED turns off.
REMOTE CONTROL

Remote Control Setup

Background
Interface
COM setting
The GPD-3303D and GPD-3303S are capable of being remotely controlled via the USB connection.
USB slave port, rear panel
Set up the COM port inside the PC according to the following list.
Baud rate: 9600
Parity bit: None
Data bit: 8
Stop bit: 1
Data flow control: None
Note
When a setting is recalled, the output automatically turns off.
35
Functionality check
36
Run this query command via the terminal application such as MTTTY (Multi-threaded TTY).
*IDN?
This should return the identification information: Manufacturer, model name, serial number, firmware version.
GW INSTEK, GPD-3303x, SN: xxxxxxxx, Vx.xx
REMOTE CONTROL
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

Remote Connection Step

Entering the remote control mode
Leaving the remote control mode
1. Connect the USB cable to the slave port.
2. The connection will be automatically established, and the front panel shows “USB…YES” message.
3. The power supply also automatically enters the lock state (the Lock key will become activated).
1. Disconnect the USB cable from the rear.
2. The display shows “USB…NO” message.
Command format
Output channel
Parameter Type Description Example
<Boolean>
<NR1>
<NR2>
Note
1 (CH1) or 2 (CH2)
boolean logic 0 (off), 1 (on)
integers 0, 1, 2, 3
decimal numbers
Commands are not case-sensitive.
1: command header
2: output channel
3: separator
4: parameter
0.1, 3.14, 8.5

Error Messages

Command Syntax

The following error messages might appear when the GPS-3303D or 3303S cannot accept the command.
Message contents Descriptions
Program mnemonic too long
The command length must be 15 characters or less.
3. Unlock the power supply by keep pressing the Lock key until it turns off.
4. The power supply goes back to the local operation mode.
Invalid character
37
Missing parameter
Data out of range
Command not allowed
Undefined header
38
Invalid characters, such as symbols, are entered. Example: VOUT#
The parameter is missing from the command. Example: VSET: (should have a number)
The entered value exceeds the specification. Example: VSET:33 (should be 32V)
The entered command is not allowed in the circumstance. Example: trying to set CH2 output while in the tracking mode.
The entered command does not exist, or the syntax is wrong.
REMOTE CONTROL
Distributed By TestEquity - www.testequity.com - 800-732-3457

Command List

Detailed descriptions of each command start from the next page.
The “HELP” command shows all the below commands and their
meanings, except for the HELP command itself.

ISET<X>:<NR2>

ISET<X>?

Sets the output current.
Returns the output current setting.
GPD-3303 Series User Manual

Command Details

ISET<X>:<NR2>
Description
Panel operation
Sets the output current.
See page23

VSET<X>:<NR2>

VSET<X>?
IOUT<X>?
VOUT<X>?
TRACK<NR1>
BEEP<BOOLEAN>
OUT<BOOLEAN>
STATUS?
*IDN?
RCL<NR1>
SAV<NR1>
HELP?
ERR?
Sets the output voltage.
Returns the output voltage setting.
Returns the actual output current.
Returns the actual output voltage.
Selects the operation mode.
Turn on or off the beep.
Turn on or off the output.
Returns the GPS-3303D or GPS-3303S status.
Returns the GPS-3303D or GPD-3303S identification.
Recalls a panel setting.
Saves the panel setting.
Shows the command list.
Returns the instrument error messages.
Response time
Example ISET1:2.234
Minimum 70ms
ISET1:2.23
Sets the CH1 output current to
2.234A (for GPD-3303S) Sets the CH1 output current to
2.23A (for GPD-3303D)
ISET<X>?
Description
Panel operation
Response time
Example ISET1?
Returns the output current setting.
See page23
Minimum 80ms
Returns the CH1 output current setting
VSET<X>:<NR2>
Description
Panel operation
Response time
Example VSET1:20.345
Sets the output voltage.
See page23
Minimum 70ms
Sets the CH1 voltage to 20.345V
VSET1:20.3
(for GPD-3303S) Sets the CH1 voltage to 20.3V (for GPD-3303D)
39
40
REMOTE CONTROL
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual

VSET<X>?

Description
Response time
Example VSET1?
Returns the output voltage setting.
Minimum 80ms

IOUT<X>?

Description
Response time
Example IOUT1?
Returns the actual output current.
Minimum 80ms

VOUT<X>?

Description
Panel operation
Response time
Example VOUT1?
Returns the actual output voltage.
See page23
Minimum 80ms

TRACK<NR1>

Description
Panel operation
NR1
Response time
Selects the operation mode: independent, tracking series, or tracking parallel.
See page27
0: Independent 1: Tracking series 2: Tracking parallel
Minimum 70ms
Returns the CH1 voltage setting
Returns the CH1 output current
Returns the CH1 output voltage
Description
Panel operation
Response time
Example BEEP1
Turns the beep on or off.
See page21
Minimum 70ms
Turns on the beep

OUT<Boolean>

BEEP<Boolean>

Description
Panel operation
Response time
Example OUT1
Turns on or off the output.
See page21
Minimum 70ms
Turns on the output

STATUS?

Description
Response time
Contents
Returns the GPD-3303D or GPD-3303S status.
Minimum 400ms
8 bits in the following format
Bit Item Description 0 CH1 0=CC mode, 1=CV mode 1 CH2 0=CC mode, 1=CV mode 2, 3 Tracking 01=Independent, 11=Tracking series,
10=Tracking parallel 4 Beep 0=Off, 1=On 5 N/A N/A 6 Output 0=Off, 1=On 7 N/A N/A
Example TRACK0
Selects the independent mode
41
42
REMOTE CONTROL
Distributed By TestEquity - www.testequity.com - 800-732-3457

*IDN?

Description
Response time
Contents
Returns the GPD-3303D or GPD-3303S identification.
Minimum 300ms
GW INSTEK,GPD-3303x,SN: xxxxxxxx, Vx.xx
(Manufacturer, model name, serial number, firmware version)

RCL<NR1>

Description
Panel operation
NR1
Response time
Example RCL1
Recalls a panel setting.
See page35
1 – 4: Memory 1 to 4
Minimum 70ms

SAV<NR1>

Description
Panel operation
Stores the panel setting.
See page34
Recalls the panel setting stored in memory 1
GPD-3303 Series User Manual
Contents
ISET<x>:<NR2> Sets the value of current.
VSET<x>:<NR2> Sets the value of voltage. x:1=CH1,2=CH2.
ISET<x>? Return the value of current.
VSET<x>? Return the value of voltage.
IOUT<x>? Returns actual output current
VOUT<x>? Returns actual output voltage.
TRACK<NR1> Sets the output of the power supply working on independent or tracking mode. NR1:0=INDEP,1=SER,2=PARA;
BEEP<Boolean> Sets the BEEP state on or off.
OUT<Boolean> Sets the output state on or off.
STATUS? Returns the power supply state.
bit0:(CH1)0=CC,1=CV;bit1:(CH2)0=CC,1=CV;bit23=(TRACK)01=INDEP,1 1=SER,10=PARA;bit4:(BEEP)0=OFF,1=ON;bit6:(OUT)0=OFF,1=ON;
*IDN? Returns instrument identification.
RCL<NR0> Recall the setting data from the memory which previous saved.
SAV<NR0> Saves the setting data to memory.
NR0:1=Memory1,2=Memory2,3=Memory3,4=Memory4;
ERR? Returns instrument error messages.
NR1
Response time
Example SAV1
1 – 4: Memory 1 to 4
Minimum 70ms
Stores the panel setting in memory 1

HELP?

Description
Response time
Shows the command list.
Minimum 1000ms
43

ERR?

Description
Response time
Contents
44
Checks the error status of the instrument and returns the last error message.
Minimum 70ms
See page38 for the list of error messages.
FAQ
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
FAQ
Q1. I pressed the panel lock key but the output still turns on/off.
A1. The output key is not affected by the panel lock key operation, for ensuring safety.
Q2. The CH3 overload indicator turned on – is this an error?
A2. No, it simply means that the CH3 output current reached the maximum 3.0A and the operation mode turned from CV (constant voltage) to CC (constant current). You can continue using the power supply, although reducing the output load is recommended.
Q3. The specifications does not match the real accuracies.
A3. Make sure that the power supply is powered on for at least 30 minutes, within +20°C – +30°C.
Q4. The internal memory is not recording the panel setting correctly – the output should be on.
A4. The output is always stored or recalled as “off” to ensure safety.
APPENDIX
23BFuse Replacement
Steps
1. Take off the power cord and remove the fuse
socket using a minus driver.
2. Replace the fuse in the holder.
For more information, contact your local dealer or GWInstek at www.gwinstek.com.tw
/ marketing@goodwill.com.tw.
45
Rating
46
100V/120V:T6.3A/250V
220V/230V:T3.15A/250V
APPENDIX
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
24BSpecifications
The specifications apply when the GPD-3303 series are powered on for at least 30 minutes under +20°C – +30°C.
Output Ratings CH1/CH2
0 ~ 30V / 0 ~ 3A
Independent
CH1/CH2
0 ~ 60V / 0 ~ 3A Series CH1/CH2
0 ~ 30V / 0 ~ 6A Parallel
CH3 2.5V/3.3V/5.0V, 3A Vol tage Regulation
Line Load
0.01% + 3mV
0.01% + 3mV (rating current 3A)
0.02% + 5mV (rating current > 3A)
Recovery Time
Ripple & Noise
1mVrms (5Hz ~ 1MHz)
100μs (50% load change, minimum
load 0.5A) Temperature
300ppm/°C Coefficient
Current
Line
0.2% + 3mA
Regulation Load Ripple & Noise
CH3 Specification
Regulation
0.2% + 3mA
3mArms
Line 5mV
Load 15mV
Ripple & Noise Tr ac kin g
Tracking Error ≤ 0.5%+10mV of Master (GPD-3303S)
Operation Parallel
Regulation
2mVrms
0.5%+50mV of Master (GPD-3303D)
Line: 0.01% + 3mV
Load: 0.01% + 3mV
(rating current 3A)
Load: 0.02% + 5mV
(rating current > 3A)
Meter Resolution
Series Regulation
Line: 0.01% + 5mV
Load: 300mV GPD-3303D Voltage: 100mV
Current: 10mA
GPD-3303S Voltage: 1mV
Current: 1mA
A Meter GPD-3303D 3.2A full scale, 3 digits 0.5" LED
display
GPD-3303S 3.2A full scale, 4 digits 0.4" LED
display V Meter GPD-3303D 32V full scale, 3 digits 0.5" LED display GPD-3303S 32V full scale, 5 digits 0.4" LED display Program Accuracy
GPD-3303D ± (0.5% of reading + 2digits)
± (0.5% of reading + 2digits) GPD-3303S ± (0.03% of reading + 10mV)
± (0.3% of reading + 10mA) Readback Accuracy
GPD-3303D ± (0.5% of reading + 2digits)
± (0.5% of reading + 2digits) GPD-3303S ± (0.03% of reading + 10mV)
± (0.3% of reading + 10mA) Insulation Chassis and
20M or above (DC 500V)
Te rm in al
Chassis and
30M or above (DC 500V)
AC cord Operation Environment
Indoor use, Altitude: 2000m
Ambient temperature 0 ~ 40°C
Relative humidity 80%
Installation category: II, Pollution degree: 2 Storage Environment
Ambient temperature –10 ~ 70°C
Relative humidity 70% Power Source AC 100V/120V/220V/230V±10%, 50/60Hz Accessories User manual x1
Test lead GTL-104 x 2, GTL-105 x 1
(Europe) Test lead GTL-203 x 1, GTL-204 x 2 Dimensions 210 (W) x 130 (H) x 265 (D) mm Weight Approx. 7kg
48BOptions
USB cable GTL-246 USB 2.0, A-B type
47
48
APPENDIX
Distributed By TestEquity - www.testequity.com - 800-732-3457
GPD-3303 Series User Manual
25BDeclaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
(1) No.7-1, Jhongsing Rd., Tucheng City, Taipei County, Taiwan (2) No. 69, Lu San Road, Suzhou City (Xin Qu), Jiangsu Sheng, China declare, that the below mentioned product
Type of Product: Power Supply Model Number: GPD-3303D/GPD-3303S
are herewith confirmed to comply with the requirements set out in the Council Directive on the Approximation of the Law of Member States relating to Electromagnetic Compatibility (2004/108/EC, 89/336/EEC, 92/31/EEC, 93/68/EEC) and Low Voltage Directive (73/23/EEC, 93/68/EEC). For the evaluation regarding the Electromagnetic Compatibility and Low Voltage Directive, the following standards were applied: EMC
EN 61326-1: 2006 Electrical equipment for measurement, control and laboratory use –– EMC requirements
Conducted Emission Radiated Emission
EN 55011: 1998 + A1:1999 + A2:2002 Current Harmonics EN 61000-3-2: 2000 + A2:2005 Voltage Fluctuations EN 61000-3-3: 1995 + A1:2001
------------------------- Conducted Susceptibility
------------------------- Power Frequency Magnetic Field
------------------------- Voltage Dip/ Interruption
Safety
Low Voltage Equipment Directive 73/23/EEC, 93/68/EEC
Safety Requirements IEC/EN 61010-1: 2001
ClassA
Electrostatic Discharge EN 61000-4-2: 1995 + A1:1998 + A2:2001 Radiated Immunity EN 61000-4-3: 2002 + A1:2002 Electrical Fast Transients EN 61000-4-4: 2004 Surge Immunity EN 61000-4-5: 1995 + A1:2001
EN 61000-4-6: 1996 + A1:2001
EN 61000-4-8: 1993 + A1:2001
EN 61000-4-11: 2004
INDEX
automatic out off ..................21
banana plug...........................20
beep setting
contents..............................21
manual ...............................21
remote control...................42
caution symbol .......................5
cleaning the instrument.........7
COM setting, remote control
............................................36
command list ........................39
recalling the list.................43
common terminal, tracking
series...................................27
cooling fan.............................17
safety instruction ................6
CV/CC
CH1/CH2 indicator .........24
CH3 indicator....................26
operation theory .........10, 18
EN61010
declaration of conformity 49
measurement category ......6
pollution degree .................7
EN61326-1 .............................49
environment
operation..............................7
specification ......................48
storage..................................8
error messages, remote
control..........................38, 44
front panel
lock (manual).................... 22
overview ........................... 14
fuse
rating ................................. 46
replacing ........................... 46
safety instruction ............... 6
GPD series
block diagram................... 12
dynamic load.................... 10
lineup................................. 11
list of features ................... 11
operation theory............... 12
technology overview......... 9
ground symbol....................... 5
identification information... 43
load connection.................... 20
operation mode
independent...................... 23
remote control .................. 41
specifications .................... 47
tracking parallel ............... 32
tracking series................... 27
output current setting
manual............................... 24
remote control .................. 40
output on/off
FAQ ................................... 45
manual............................... 21
remote................................ 42
output voltage setting
manual............................... 24
remote control .................. 40
49
50
INDEX
Distributed By TestEquity - www.testequity.com - 800-732-3457
overload indicator ................26
power supply
safety instruction ................6
setup...................................19
socket overview ................17
specification.......................48
protective ground symbol.....5
rear panel overview .............17
recall settings
manual ...............................35
remote ................................43
remote control
command syntax ..............38
connection test ..................36
error messages ..................38
interface............................. 36
save settings
manual............................... 34
remote................................ 43
service operation
about disassembly ............. 6
contact ............................... 45
status, instrument ................ 42
tracking mode
operation theory................. 9
UK power cord....................... 8
USB interface ........................36
warning symbol ..................... 5
wire, load ..............................20
51