Instek GCT-9040 User's Manual

Page 1
Ground Bond Tester
GCT-9040
USER MANUAL
GW INSTEK PART NO. 82CT-90400E01
ISO-9001 CERTIFIED MANUFACTURER
Page 2
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Page 3
Table of Contents
Table of Contents
SAFETY INSTRUCTIONS ................................................... 5
GETTING STARTED ........................................................... 9
GCT-9040 Overview ............................. 10
Appearance .......................................... 14
Set Up .................................................. 19
OPERATION .................................................................... 25
Menu Tree ............................................ 26
Test Lead Connection ........................... 30
GB Manual Testing .............................. 36
Linked ACW/DCW/IR Tests .................. 57
Common Utility Settings ...................... 66
EXTERNAL CONTROL ..................................................... 74
External Control Overview .................... 75
REMOTE CONTROL ........................................................ 81
Interface Configuration ........................ 82
Command Syntax ................................. 86
Command List for GCT-9040 ................ 89
Error Messages .................................. 124
FAQ ............................................................................... 125
APPENDIX ..................................................................... 127
Fuse Replacement .............................. 127
Error Messages .................................. 128
GCT-9040 Specifications .................... 130
GCT-9040 Dimensions ....................... 131
Declaration of Conformity .................. 133
3
Page 4
GCT-9040 User Manual
INDEX ............................................................................ 134
4
Page 5
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the instrument or to other properties.
Attention Refer to the Manual
Protective Conductor Terminal
Frame or Chassis Terminal Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
5
Page 6
GCT-9040 User Manual
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
CAUTION
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not disassemble the GCT-9040 unless you
are qualified.
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The GCT-9040 does not fall under category II, III or IV.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Power Supply
WARNING
AC Input voltage range:
100/120/220/230VAC ±10%
Frequency: 50Hz/60Hz To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.
Safety Guidelines
6
Page 7
SAFETY INSTRUCTIONS
Cleaning the GCT-9040
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 70% (no condensation) Altitude: < 2000m Temperature: 0˚C~40˚C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The GCT-9040 falls under degree
2.
Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -10°C to 70°C Relative Humidity: 85% (no condensation)
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
7
Page 8
GCT-9040 User Manual
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the tester in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
8
Page 9
GETTING STARTED
POWER
START STOP
REMOTE
MANU EDIT/SAVE UTILITY
ESC PAGE
PASS FAIL READY TEST
CAUTION
GB Rx
SENSE H
SOURCE H
SENSE L
SOURCE L
40A MAX.
GCT-9040
Ground Bond Tester
GCT-9040 Overview ......................................................... 10
Introduction to the GCT-9040 ................................................................ 10
Main Features .......................................................................................... 11
Accessories ............................................................................................... 11
Package Contents .................................................................................... 13
Appearance ..................................................................... 14
Front Panel ............................................................................................... 14
Rear Panel ................................................................................................. 17
Set Up ............................................................................. 19
Line Voltage Connection and Power Up ............................................. 19
Installing the Optional GPIB Card ........................................................ 21
Workplace Precautions ........................................................................... 22
Operating Precautions ............................................................................ 23
GETTING STARTED
This chapter describes the safety tester in a nutshell, including its main features and front / rear panel introduction. After going through the overview, please read the safety considerations in the Set Up chapter.
9
Page 10
GCT-9040 User Manual
GCT-9040 Overview
Introduction to the GCT-9040
GCT-9040 is a ground bond tester that can operate as a standalone unit or in conjunction with a GPT-9000 Series safety tester to perform additional ACW/DCW/IR tests before, after or simultaneously(ACW/DCW only) with GB tests.
When the GCT-9040 operates with a GPT-9000 Series safety tester, it uses a dedicated LINK port and remote commands to communicate in a master-slave fashion. The GCT-9040 operates as the master and the GPT-9000 Series unit operates as a slave.
The GCT-9040 can store up to 100 manual tests allowing the safety testers to accommodate any number of safety standards, including IEC, EN, UL, CSA, GB, JIS and others.
Note: Throughout this user manual, the terms ACW, DCW, IR and GB refer to AC Withstanding, DC Withstanding, Insulation Resistance and Ground Bond testing, respectively. GPT-9000 refers to any of the GPT-98XX or GPT-99XX models, GPT-9000A refers to any of the GPT-99XXA models. For information specific to the GPT­9000, please see the GW Instek website, http://www.gwinstek.com.
10
Page 11
Main Features
Performance
GB: 3A~40A ac
Features
100 test conditions (MANU mode) Over temperature, voltage and current
protection
Pass, Fail, Test, Caution and Ready indicators PWM output (90% efficiency, increased
reliability)
Interlock (configurable).
Interface
Remote control start/stop interface terminal USB interface for programming Optional GPIB interface for programming Signal I/O port for pass/fail/test monitoring
and start/stop control/interlock
LINK port to control slave units for ACW, DCW
and IR tests.
Standard Accessories
Part number
Description Region dependent
Power cord
GTL-215 x1
GB Test leads
N/A
Remote terminal male plug
N/A
Interlock key
GTL-247
Type A-A USB cable
GTL-132
Link cable
GETTING STARTED
Accessories
11
Page 12
GCT-9040 User Manual
Optional Accessories
Part number
Description GTL-248
GPIB cable
GTL-251
GPIB-USB-HS (high speed)
GRA-417
Rack Adapter Panel (19”, 4U)
Options
Part number
Description
Opt.01 GPIB Interface
GPIB card
12
Page 13
GETTING STARTED
Opening the box
Contents (single unit)
GCT-9040 unit Quick Start guide User manual CD CTC (Calibration
Traceable Certificate)
Type A-A USB cable
Power cord x1 (region
dependent)
GTL-215 test leads x1 GTL-132 LINK cable Remote terminal male
plug
Interlock key
Note
Keep the packaging, including the box, polystyrene foam and plastic envelopes should the need arise to return the unit to GW Instek.
Package Contents
Check the contents before using the GCT-9040.
13
Page 14
GCT-9040 User Manual
PASS/FAIL indicators
Directional keysFunction keys
CAUTION indicator
Sense H and L terminals
REMOTE terminal
STOP button
START button
POWER
button
Display
Scroll wheel
Configuration keys
READY indicator
TEST indicator
POWER
START STOP
REMOTE
MANU EDIT/SAVE UTILITY
ESC PAGE
PASS FAIL READY TEST
CAUTION
GB Rx
SENSE H
SOURCE H
SENSE L
SOURCE L
40A MAX.
GCT-9040
Ground Bond Tester
Source H and L terminals
Display
240 X 64 dot matrix display (LCD)
Function keys
The function keys correspond to the soft-keys directly above on the main display.
Pass/Fail indicators
PASS
FAIL
The PASS and FAIL indicators light up upon a PASS or FAIL test result at the end of a manual test or automatic test.
ESC key
ESC
The ESC key is used to exit out of a menu or cancel a setting.
PAGE key
PAGE
The PAGE key is used in the EDIT status to toggle the HI/LO setting between mΩ & V.
Directional arrow keys
The directional arrow keys are used to navigate menus and parameter settings.
Appearance
Front Panel
14
Page 15
GETTING STARTED
READY indicator
READY
The READY indicator is lit when the tester is ready to begin testing. The STOP button is used to put the tester into READY status.
TEST indicator
TEST
The TEST indicator is lit when a test is on. The START button is used to put the tester into TEST status.
CAUTION indicator
CAUTION
40A MAX.
The CAUTION indicator will light up when the output terminals are active. Only after the test has finished or stopped will the indicator turn off.
SENSE and SOURCE terminals
GB Rx
SENSE H
SOURCE H
SENSE L
SOURCE L
The SOURCE H, SOURCE L, SENSE H and SENSE L terminals are used for GB testing.
Scroll wheel
The scroll wheel is used to edit parameter values.
UTILITY key
UTILITY
Used to enter the Common Utility menu.
EDIT/SAVE key
EDIT/SAVE
Used to start editing MANU tests as well as save settings and parameters.
MANU key
MANU
The MANU key is used to select manual tests.
15
Page 16
GCT-9040 User Manual
REMOTE terminal
REMOTE
The REMOTE terminal is used to connect to a remote controller.
STOP button
STOP
The STOP button is used to stop/cancel tests. The STOP button will also put the tester in the READY status to begin testing.
START button
START
The START button is used to start tests.
The START button can be used to start tests when the tester is in the READY status. Pressing the START button will put the tester in the TEST status.
POWER switch
POWER
Turns the power on. The tester will always start up with the last test setting from when the instrument was last powered down.
16
Page 17
Rear Panel
SIGNAL I/O USB A port LINK port Fan
Fuse selectorLine voltage
Optional GPIB port
GND
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING CONDUCTOR MUST BE
ONLY WITH SPECIFIED TYPE AND RATED FUSE. NO OPERATOR SERVICEABLE COMPONENTS INSIDE. DO NOT REMOVE COVERS. REFER SERVICING TO
FOR CONTINUED FIRE PROTECTION. REPLACE
CONNECTED TO GROUND.
QUALIFIED PERSONNEL.
WARNING
AC
LINE VOLTAGE
100V
230V
220V
120V
SELECTION
RANGE (50/60Hz)
250V
250V
T 7A
FUSE
T 4A
POWER MAX.
700VA
GND
SIGNAL I / O LINK
SER. NO. LB
ENSURE THE POWER IS REMOVED FROM THE INSTRUMENT BEFORE REPLACING THE FUSE
GPIB
90-110V 108-132V 198-242V 207-250V
SIGNAL I/O port
SIGNAL I / O
The SIGNAL I/O port is used to monitor the tester status (PASS, FAIL, TEST) and input (START/ STOP signals). It is also used with the Interlock key.
USB A port
Used for remote control and firmware updates.
LINK port
LINK
With the GCT-9040 acting as the master unit, the LINK port is used to control a GPT-9XXX slave unit to perform additional ACW, DCW or IR tests.
Fan/Fan Vents
Exhaust fan. Allow enough room for the fan to vent. Do not block the fan openings.
GND
GND
Connect the GND (ground) terminal to the earth ground.
GETTING STARTED
17
Page 18
GCT-9040 User Manual
Line voltage input
220
230
100
120
Line voltage input: 100/120/220/230VAC ±10%
Line voltage fuse
220
230
100
120
Line voltage selector and fuse:
100V/120V
T7A 250V
220V/230V
T4A 250V
Optional GPIB port
GPIB
Optional GPIB interface for remote control.
18
Page 19
GETTING STARTED
Background
Before powering up the GCT-9040 ensure the correct voltage has been selected on the rear panel. The GCT-9040 supports line voltages of 100V/120V/220V and 230V.
Steps
1. Check the line voltage and the fuse
in the fuse holder.
Page 127
The desired line voltage should line up with the arrow on the fuse holder.
2
2
0
1
0
0
120
230
2. Connect the power cord to
the AC voltage input.
2
2
0
1
0
0
120
230
3. If the power cord does not
have an earth ground, ensure the ground terminal is connected to an earth ground.
GND
Warning
Ensure the power cord is connected to an earth ground. Failure could be harmful to the operator and instrument.
4. Press the Power button.
POWER
Set Up
Line Voltage Connection and Power Up
19
Page 20
GCT-9040 User Manual
5. When the unit is powering up, all the LED
indicators will light. Check to make sure all 5 LED indicators are working.
6. Check to make sure the System Self Test passes
without errors.
S y s e m t H a r w a rd
e C
h
C h e c i g . .n .
F i r w a rm
e C h
SYS T E S M E L F T E S T
k e c i g . .n .k e c i g . .n .k
After the System Self Test completes, the tester will go into VIEW status and be ready to operate.
20 0
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
VIEW status
MMA UN N A_M A NU = * * * -
V
WARNING
See the Appendix on page 128 for details if a self­test error is detected.
20
Page 21
GETTING STARTED
Background
The optional GPIB is a user-installable option. Follow the instructions below to install the GPIB card.
WARNING
Before installing the optional GPIB card ensure the GCT-9040 is turned off and disconnected from power.
Steps
1. Remove the screws from the rear panel cover
plate.
2. Insert the GPIB card into the two slots on either
side of the opening. Push the card gently until it is fully inserted.
Installing the Optional GPIB Card
21
Page 22
GCT-9040 User Manual
Background
The GCT-9040 is a high current instrument that is capable of delivering up to 40A. The following section describes precautions and procedures that must be followed to ensure a safe work environment.
WARNING
The GCT-9040 generates high current in excess of 40A. Follow all safety precautions, warnings and directions given in the following section when using the instrument.
1. Only technically qualified personnel should be
allowed to operate the tester.
2. The operating workplace must be fully isolated,
especially when the instrument is in operation. The instrument should be clearly labeled with appropriate warning signage.
3. The operator should not wear any conductive
materials, jewelry, badges, or other items, such wrist watches.
4. The operator should wear insulation gloves for
electrical protection.
5. Ensure the earth ground of the line voltage is
properly grounded.
6. When high current flows through the test leads
a strong magnetic field is produced. Ensure any devices that are adversely affected by magnetic fields are not placed near the tester.
Workplace Precautions
22
Page 23
GETTING STARTED
Background
The GCT-9040 is a high current instrument. The following section describes precautions and procedures that must be followed to ensure that the tester is operated in a safe manner.
WARNING
The GCT-9040 generates outputs up to 40A ac. Follow all safety precautions, warnings and directions given in the following section when using the instrument.
1. Never touch the tester, lead wires, terminals,
probes and other connected equipment when the tester is testing.
2. After testing, any probes, terminals or ports
may become extremely hot due to the high test current. Do not touch these points to avoid burns.
3. Do not turn the tester on and off quickly or
repeatedly. When turning the power off, please allow a few moments before turning the power back on. This will allow the protection circuits to properly initialize.
Do not turn the power off when a test is running, unless in an emergency.
4. Only use those test leads supplied with the
instrument. Leads with inappropriate gauges can be dangerous to both the operator and the instrument. Never use the Sense leads on the SOURCE terminals.
Operating Precautions
23
Page 24
GCT-9040 User Manual
5. Ensure the earth ground of the line voltage is
properly grounded.
6. Only connect the test leads to the SOURCE
H/SENSE H terminals before the start of a test. Keep the test leads disconnected at all other times.
7. Always press the STOP button when pausing
testing.
8. Do not leave the tester unattended. Always
turn the power off when leaving the testing area.
9. When remotely controlling the tester, ensure
adequate safety measures are in place to prevent:
Inadvertent output of the test current.
Accidental contact with the instrument during
testing. Ensure that the instrument and DUT are fully isolated when the instrument is remotely controlled.
24
Page 25
OPERATION
Menu Tree ....................................................................... 26
Menu Tree Overview .............................................................................. 27
Test Lead Connection ...................................................... 30
GB Connection ......................................................................................... 30
Grounding Mode Note ........................................................................... 33
GB Manual Testing .......................................................... 36
Choose/Recall a Manual Test Number ................................................ 37
Edit Manual Test Settings ...................................................................... 38
Setting the GB Test Current ................................................................... 39
Setting the Test Frequency ..................................................................... 39
Setting the Upper and Lower Limits .................................................... 40
Setting a Reference Value ....................................................................... 42
Setting the Test Time (Timer) ................................................................ 43
Creating a MANU Test File Name ........................................................ 44
Saving and Exiting EDIT Status ............................................................ 45
Running a MANU Test ........................................................................... 47
PASS / FAIL MANU Test ...................................................................... 51
Zeroing of the Test Leads ....................................................................... 54
Linked ACW/DCW/IR Tests .............................................. 57
LINK Connection .................................................................................... 58
LINK Test Configuration ....................................................................... 60
Run LINK Test ......................................................................................... 63
OPERATION
25
Page 26
GCT-9040 User Manual
status
SET
T
status
OTS
P
status
YAER
D
status
EIV
W
status
IDE
T
Press
START
Save the MANU
test
Press STOP
Press EDIT/ SAVE
Press STOP
Common Utility
Settings
1
Press
ESC
Press EDIT/
SAVE
Toggle between ref/HI/LO V and
ref/HI/LO Ω
Press PAGE
1 Press EDIT/SAVE to save settings, or ESC to cancel and return to the previous screen.
Press
UTILITY
status
REZ
O
Press ZERO
soft-key
Press
START
Short test leads
Press STOP
Press
MANU
Press STOP
Press
START
status
SAP
S
status
IAF
L
Press STOP
X2
Menu Tree
This section describes the overall structure of the operation statuses and modes for the GCT-9040 ground bond safety tester. The tester has one testing mode (MANU mode) and 5 main operation statuses (VIEW, EDIT, READY, TEST and STOP).
26
Page 27
OPERATION
VIEW status
VIEW status is used to view the parameters of the selected manual test. The VIEW status is also used to select the MANU test number. VIEW status is the default state of the unit.
20 0
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
VIEW status
MMA UN N A_M A NU = * * * -
V
EDIT status
EDIT status is used to edit the manual test parameters. Pressing the EDIT/SAVE key will save any changes. Pressing the ESC key will cancel any changes.
VOL = .0 0 00
I N KL G B G - W C O NF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00MMA UN N A_ T2M A NU = * * * - 0 0
EDIT status
V
F R E Q = 0 H z 6 V IH = .0 0 08
G B V = 0 . 8 0 0 V
READY status
When the tester is in READY status, it is ready to begin testing. Pressing the START button will begin testing and put the tester into TEST status. Pressing the MANU key will return the tester to VIEW status.
200
AER D Y
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
READY status
MMA UN N A_M A NU = * * * -
V
E ROZ
Menu Tree Overview
27
Page 28
GCT-9040 User Manual
TEST status
TEST status is active when a MANU test is running. Pressing STOP will cancel the test.
20 0
E ROZ I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
SET
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
T
TEST status
MMA UN N A_M A NU = * * * -
0 051
V
STOP status
STOP status is shown when a manual test did not finish running and has been stopped by the operator. Pressing STOP will return the tester to READY status.
20 0
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
OTS
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
P
STOP status
MMA UN N A_M A NU = * * * -
0 051
V
ZERO status
ZERO status is shown when performing a zeroing test for the test leads. Performing a zeroing test will automatically set the REF# value for the selected test.
E ROZ
20 0
I N KL G B G - W CONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
REZ
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
. 0 V ER F =# 0 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
O
MMA UN N A_M A NU = * * * -
0 051
V
ZERO status
MANU test number
When in the VIEW mode, you can select a MANU test number with the scroll wheel.
20 0
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MANU test number
MMA UN N A_M A NU = * * * -
V
28
Page 29
OPERATION
Common Utility Settings
This utility controls the LCD, buzzer, interface, control and time settings. These settings are system wide.
L C D C o t r an L C D B r g h ti
n se
L C D B U Z Z N T EI R C T R L
s
s t : 8
MOC MO U N T I L I T Y
B: R I G TH
C T R L T I ME
29
Page 30
GCT-9040 User Manual
Background
GB tests use the SENSE H/L and SOURCE H/L terminals with the GTL-215 test leads.
Single GB Connection
Source H
DUT
FLOATING
GPT-9000
Sense L
Sense H
Source L
Ground bond tests are usually connected between the ground pin (conductor terminal) of
the DUT’s power cord or power socket and a
conductive point on the DUT chassis.
ACW/DCW/IR and GB Connection
GPT-9XXX (ACW/DCW)
GCT-9XXX (GB tests)
Source H
Source L
Sense H
Sense L
Return terminal
DUT
FLOATING
High Voltage terminal
ACW, DCW and IR withstanding tests are usually performed by testing the withstanding potential (ACW/DCW) or resistance (IR) from the neutral or live pin and the ground pin (conductor terminal) of the DUT.
Test Lead Connection
This section describes how to connect the GCT-9040 to a DUT for ground bond testing. The DUT should be floating with respect to ground and should be powered off.
GB Connection
30
Page 31
OPERATION
Steps for GB Connection (GCT-9040)
1. Turn the power off on the safety tester.
2. Using the GTL-215 GB test lead, a. Connect the Sense H lead to the SENSE
H terminal.
b. Connect the Sense L lead to the SENSE
L terminal.
c. Connect the Source H lead to the
SOURCE H terminal.
d. Connect the Source L lead to the
SOURCE L terminal.
Sense H lead Sense L lead
Source L leadSource H lead
SENSE H
GB Rx
SENSE L
SOURCE LSOURCE H
e. Connect the H clip of the GB test lead
to a conductive point on the DUT chassis.
f. Connect L clip of the GB test lead to the
ground (conductor terminal) pin of the DUT power cord or power socket.
DUT
Ground pin from power cord
or power socket
Conductive point
on the chassis
L clip
H clip
G
N
L
31
Page 32
GCT-9040 User Manual
Steps for ACW/DCW/IR Connection (GPT-9XXX)
1. Turn the power off on the GPT-9XXX safety
tester.
2. Using the GHT-114 test leads, a. Connect the high voltage test lead(red)
to the HIGH VOLTAGE terminal and screw firmly into place.
b. Connect the return test lead(white) into
the RETURN terminal and screw the protector bar into place, as shown below.
HIGH VOLTAGE
Terminal
RETURN Terminal
c. Connect the high potential clip to
neutral or live pin of the power cord or socket of the DUT that you want to test.
d. Connect the return clip to the ground
pin (conductor terminal) of the power cord or socket.
32
Page 33
OPERATION
DUT
Ground pin from power cord
or power socket
Live or neutral pin from
power cord or socket.
L clip
H clip
G
N
L
Background
The GCT-9040 operates with the SOURCE L terminal floating with respect to the earth ground. For the GCT-9040 and GPT-9XXX Series, this is referred to as having the GROUND MODE is set to OFF. On the GCT­9040 this mode is fixed.
When performing additional ACW or DCW tests, a slave GPT-9XXX Series unit can operate with the GROUND MODE set to ON or OFF.
Grounding Mode Note
33
Page 34
Overview
GROUND MODE = OFF
This mode is for DUTs that are floating and not directly connected to an earth ground. When GROUND MODE is set to OFF any stray capacitance/resistance that leaks to the earth ground from the DUT side of the testing circuit are not measured. GB and IR tests must operate with GROUND MODE = OFF.
GROUND MODE = ON
When GROUND MODE is set to ON, the GPT­9XXX grounds the return terminal to the ground. This mode measures the potential of the HIGH VOLTAGE terminal with respect to earth ground. This means that any stray capacitance/resistance that leaks to earth ground will also be measured.
For further details about the grounding mode for the ACW, DCW and IR tests, please see the GPT-9XXX user manual.
Ground Mode Icons
The ground mode icons represent the ground mode settings.
E
F R EQ = 0 H z 6
IDE
0 0 m AER F =# 0 .0MMA UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
GROUND
MODE = OFF
m A
E
F R E Q = 0 H z 6
IDE
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
GROUND
MODE = ON
m A
GCT-9040 User Manual
34
Page 35
OPERATION
Warning
The GCT-9040 is a floating device. The DUT ground mode setup depends on the GPT-9XXX ground mode setting.
GPT-9XXX (ACW/DCW)
GCT-9040 (GB tests)
Source H
Source L
Sense H
Sense L
Return terminal
DUT
High Voltage terminal
GROUND MODE =
ON or OFF
GROUND MODE = OFF
GROUNDED DUT
1
2
GPT-9XXX (ACW/DCW)
GCT-9040 (GB tests)
Source H
Source L
Sense H
Sense L
Return terminal
DUT
High Voltage terminal
GROUND MODE =
ON or OFF
GROUND MODE = OFF
DUT FLOATING
1
2
35
Page 36
GCT-9040 User Manual
GB Manual Testing
This section describes how to create, edit and run a single GB safety test. Each Manual setting described in this chapter only applies to the selected manual test – no other manual tests are affected.
Each manual test can be created/stored/recalled to/from one of 100 memory locations.
Choose/Recall a Manual Test number → from page 37. Edit Manual Test Settings → from page 38. Setting the GB Test Current→ from page 39. Setting the Test Frequency → from page 39. Setting the Upper and Lower Limits → from page 40. Setting a Reference Value → from page 42. Setting the Test Time (Timer) → from page 43. Creating a MANU Test File Name→ from page 44. Saving and Exiting EDIT Status→ from page 45. Running a MANU Test → from page 47. PASS / FAIL MANU Test → from page 51. Zeroing of the Test Leads → from page 54
Before operating the GCT-9040 please read the safety precautions as outlined in the Set Up chapter on page 19.
36
Page 37
OPERATION
Background
Up to 100 different manual tests can be saved/recalled. MANU tests are selected in the VIEW status.
Steps
1. Ensure that the tester is in VIEW status.
See the menu tree on page 26 for details on how to enter the VIEW status.
20 0
I N KL G B G - W CONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
VIEW status
MMA UN N A_M A NU = * * * -
V
2. Use the scroll wheel to choose the
MANU number.
MANU #
001~100
20 0
I N KL G B G - W CONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MANU test number
MMA UN N A_M A NU = * * * -
V
Note
The MANU number can only be chosen in VIEW status. If in the EDIT status, switch to the VIEW status by pressing the EDIT/SAVE or ESC key.
Choose/Recall a Manual Test Number
37
Page 38
GCT-9040 User Manual
Background
To edit any of the manual test settings, the tester must be in EDIT status.
Any settings or parameters that are edited only apply to the currently selected MANU number.
Steps
1. Press the EDIT/SAVE key when in
VIEW status to enter the EDIT status. This will enter the EDIT status for the currently chosen test number.
EDIT/SAVE
EIV
W
VOL = .0 0 00
I N KL G B G - W C ONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00MM A UN N A_ T2M AN U = * * * - 0 0
V
F R EQ = 0 H z 6 V IH = .0 0 08
G B V = 0 . 8 0 0 V
Currently selected
MANU test number
2. The Status changes from VIEW to EDIT.
Note
Pressing the EDIT/SAVE key again will save the settings for the current test and return back to VIEW status.
Edit Manual Test Settings
38
Page 39
OPERATION
Background
The GB test current can be set from 3A to 40A ac.
Steps
1. Ensure the tester is in EDIT status.
Page 38
2. Press the UP / DOWN arrow keys
to bring the cursor to the current setting.
VOL = .0 0 00
I N KL G B G - W C ONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00MM A UN N A_ T2M A NU = * * * - 0 0
V
F R E Q = 0 H z 6 V IH = .0 0 08
G B V = 0 . 8 0 0 V
cursor
3. Use the scroll wheel to set the
current level.
GB
3.00A ~ 40.00A ac
Note
The ground bond voltage (GBV) is calculated as the HI limit voltage + REF voltage (or (HI limit Ω + REFΩ) X test current).
Background
A test frequency of 60Hz or 50Hz can be set, regardless of the input line voltage.
Steps
1. Ensure the tester is in EDIT status.
Page 38
Setting the GB Test Current
Setting the Test Frequency
39
Page 40
GCT-9040 User Manual
2. Press the UP / DOWN arrow keys
to bring the cursor to the FREQ setting.
VOL = .0 0 00
I N KL G B G - W CONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00MM A UN N A_ T2M A NU = * * * - 0 0
V
F R E Q = 0 H z 6 V IH = .0 0 08
G B V = 0 . 8 0 0 V
cursor
3. Use the scroll wheel to set the test
frequency.
Frequency
50Hz, 60Hz
Background
There is both a LO and HI judgment setting. When the measured value is below the LO SET setting, the test will be judged as FAIL. When the value exceeds the HI SET setting the test will be judged as FAIL. Any measurement between the LO SET and HI SET setting is judged as PASS. The LO SET limit cannot be made greater than the HI SET limit.
Steps
1. Ensure the tester is in EDIT status.
Page 38
2. Press the PAGE to choose the HI
and LO value units (voltage or resistance).
PAGE
Setting the Upper and Lower Limits
40
Page 41
OPERATION
Note
Toggling the HI/LO units will also toggle the REF# units as well (REF# m, REF# V). The REF# units and the HI/LO units are the same. If the units are changed, the REF# value will need to be set again. See page 42 for setting the REF# value.
3. Press the HI/LO soft-key or use
the UP / DOWN arrow keys to bring the cursor to the HI limit setting.
I / LH O
OR
4. Use the scroll wheel to set the HI
limit.
HI
000.1m~ 650.0m
0.001V ~ 7.200V
5. Repeat steps 2 and 3 for the LO limit setting.
LO
000.0m~ 649.9m
0.000V ~ 7.199V
Example: V units
VOL = .0 0 00
I N KL G B G - W CONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00MM A UN N A_ T2M A NU = * * * - 0 0
V
F R E Q = 0 H z 6 V IH = .0 0 08
G B V = 0 . 8 0 0 V
HI limit LO limit
Example: units
m
OL = 00 . 00
I N KL G B G - W CONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
. 0
m
ER F =# 2 00MM A UN NA_ T2M A NU = * * * - 0 0
m
F R E Q = 0 H z 6
m
IH = 05 . 00
G B V = 5 . 0 0 0 V
HI limit LO limit
Note
The LO SET setting is limited by the HI SET setting. The LO SET limit cannot be greater than the HI SET limit.
41
Page 42
GCT-9040 User Manual
Background
The REF# acts as an offset. The REF# value is subtracted from the measured current (ACW, DCW) or measured resistance (IR, GB).
Steps
1. Ensure the tester is in EDIT status.
Page 38
2. Press the PAGE soft-key to choose
the REF# value unit (voltage or resistance).
PAGE
Note
Toggling the REF# units will also toggle the HI/LO units (HI/LO m, HI/LO V). The REF# units and the HI/LO units are the same. If the units are changed, the HI/LO settings will need to be set again. See page 40 for setting the HI/LO values.
3. Press the UP / DOWN arrow keys
to bring the cursor to the REF# setting.
4. Use the scroll wheel to set the
REF# value.
REF# () REF# (V)
000.0m~ 650.0m
0.000V ~ 7.200V
Example: V units
VOL = .0 0 00
I N KL G B G - W C ONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 50MM A UN N A_ T2M AN U = * * * - 0 0
V
F R EQ = 0 H z 6 V IH = .0 0 08
G B V = 1 . 3 0 0 V
REF#
Setting a Reference Value
42
Page 43
OPERATION
Example: units
REF#
m
OL = 00 . 00
I N KL G B G - W C ONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
. 0
m
ER F =# 2 00MM A UN N A_ T2M AN U = * * * - 0 0
m
F R EQ = 0 H z 6
m
IH = 05 . 00
G B V = 5 . 2 0 0 V
Note
Limitations: ((REF# (Ω) + HI Set (Ω) ) x I Set) < 7.2V. ((REF# (V) + HI Set (V)) < 7.2V.
Note
A reference offset can be automatically created using the zeroing function. See page 54 for details.
Background
The TIMER setting is used to set the test time for the current test. The test time determines how long the test current is applied to the DUT. The test time can be set from 0.5 to 999.9 seconds, with a resolution of 0.1 seconds.
Test I
time
Start I
Initial time
TEST TIME
(Approximately 100ms)
Note
Each test has an initial test time of approximately 100ms. This time cannot be edited.
Steps
1. Ensure the tester is in EDIT status.
Page 38
Setting the Test Time (Timer)
43
Page 44
GCT-9040 User Manual
2. Press the TIMER soft-key or use
the UP/DOWN arrow keys to bring the cursor to the TIMER setting.
I M ET R
OR
VOL = .0 0 00
I N KL G B G - W CONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00MM A UN N A_ T2M A NU = * * * - 0 0
V
F R E Q = 0 H z 6 V IH = .0 0 08
G B V = 0 . 8 0 0 V
cursor
3. Use the scroll wheel to set the
TIMER value.
GB
000.5s~999.9s
Background
Each MANU test can have a user-defined test file name (default: MANU_NAME) up to 10 characters long. See the character list below for the allowed characters.
Character List:
0 1 2 3 4 5 6 7 8 9
W X Y ZM VO P RQ T USNLA B C D E F G H I J K
w x y zm vo p rq t usnla b c d e f g h i j k
+ - * / _ = : ? ( ) < > [ ]
Steps
1. Ensure the tester is in EDIT status.
Page 38
Creating a MANU Test File Name
44
Page 45
OPERATION
2. Use the UP/DOWN arrow keys to
bring the cursor to the MANU test file name at the top of the screen. The test file name is initially set as MANU_NAME.
cursor
VOL = .0 0 00
I N KL G B G - W CONF T I / LH O I M ET R
E
1
0 00
A
IDE
MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00MM A UN N A_ T2M A NU = * * * - 0 0
V
F R E Q = 0 H z 6 V IH = .0 0 08
G B V = 0 . 8 0 0 V
3. Use the scroll wheel to scroll
through the available characters.
4. Press the Left/Right arrow keys to
go the next character.
5. The MANU test file name is set when the
current test setting is saved or when the cursor is moved to another setting.
Background
After all test parameters have been set, the test can be saved.
Steps
1. When in EDIT status, press the
EDIT/SAVE key to save the current test. This will enter the VIEW status for the chosen test number.
EDIT/SAVE
Saving and Exiting EDIT Status
45
Page 46
GCT-9040 User Manual
20 0
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MM A UN N A_M AN U = * * * -
V
IDE
T
2. The Status changes from EDIT to VIEW.
Note
Pressing the EDIT/SAVE key again will return the tester back to EDIT status for the current test.
46
Page 47
OPERATION
Background
A test can be run when the tester is in READY status.
Note
The tester cannot start to run a test under the following conditions:
A protection setting has been tripped; when a
protection setting has been tripped the corresponding error message is displayed on the screen. See page 129 for a comprehensive list of the all the setting errors.
The INTERLOCK function is ON and the
Interlock key is not inserted in the signal I/O port (page 70).
The STOP signal has been received remotely.
If Double Action is ON, ensure the START button is pressed immediately after the STOP button (<0.5s).
Note
When a test is running the current output cannot be changed.
Steps
1. Ensure the tester is in VIEW status
for the current test. Save the current test if necessary.
Page 45
20 0
I N KL G B G - W CONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
VIEW status
MM A UN N A_M AN U = * * * -
V
Running a MANU Test
47
Page 48
GCT-9040 User Manual
2. Press the STOP button to put the
tester into the READY status.
STOP
20 0
AER D Y
I N KL G B G - W CONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
READY status
MM A UN N A_M AN U = * * * -
V
E ROZ
3. The READY indicator will be lit
blue when in the READY status.
READY
4. Press the START button when the
tester is in the READY status. The manual test starts automatically and the tester goes into the TEST status.
START
5. The TEST indicator will be lit
orange when in the TEST status.
TEST
20 0
E ROZ I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
SET
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
T
TEST status
MM A UN N A_M AN U = * * * -
0 051
V
6. The test will start by showing the remaining
test time. The test will continue until the test is finished or the test is stopped.
48
Page 49
OPERATION
20 0
E R OZ I N KL G B G - W C O NF T
E
F R EQ = 0 H z 6
1
0 00
A
V OL = .0 0 00
SET
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
T
TEST status
MM A UN N A_MA N U = * * * -
0 051
V
Remaining test time
Test Current
Measured /V
Measured /V
Stop the Test
1. To stop the test at any time when it
is running, press the STOP button. The test will stop immediately. When the STOP button is pressed, a judgment is not made on the test.
All panel keys except the STOP button are locked when the tester is in STOP status.
STOP
20 0
I N KL G B G - W CONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
OTS
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
P
STOP status
MM A UN N A_M AN U = * * * -
0 051
V
2. To put the tester back into READY
status, press the STOP button again.
STOP
Exit TEST Status
To exit testing, press the MANU key when the tester is in the READY status. The tester will revert to the VIEW status for the current test.
MANU
49
Page 50
GCT-9040 User Manual
20 0
I N KL G B G - W C ONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MM A UN N A_M A N U = * * * -
V
Note
Do not touch any terminals, test leads or any other connections when the test is on.
50
Page 51
OPERATION
Background
If the test is allowed to run to completion (the test is not stopped or a protection setting is not tripped) then the tester will judge the test as either PASS or FAIL.
Note
The test will be judged PASS when:
The HI and LO limits have not been tripped
during the test time.
The test will be judged FAIL when:
Either the HI or LO limit has been tripped
during the test time.
A protection setting has been tripped during
the test time. See page 129 for a list of error messages.
PASS Judgment
1. When the test is judged as PASS,
PASS will be displayed, the buzzer will sound and the PASS indicator will be lit green.
PASS
E ROZ I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
SAP
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
S
MM A UN N A_M A N U = * * * - 20 0
0 051
V
2. The PASS judgment will be held on the display
until the STOP or START button is pressed.
Pressing the STOP button will return the tester to the READY status.
STOP
PASS / FAIL MANU Test
51
Page 52
GCT-9040 User Manual
Pressing the START button will restart the test.
START
Note
The buzzer will only sound if the Pass Sound is set to ON. See page 67 for buzzer details.
The START button is disabled when the buzzer is beeping.
PASS Timing Diagrams
The timing diagrams below show the GB timing for the START status, TEST status and PASS judgment. All the test lines are active low.
GB PASS Timing
Output I
time
START TEST PASS
0.5s
Initial time
Total Test Time
(Approximately 100ms)
Signal IO
Time Cut = ON
Signal IO
Time Cut =
OFF
FAIL Judgment
1. When the test is judged as FAIL,
FAIL will be displayed, the buzzer will sound and the FAIL indicator will be lit red.
As soon as a test is judged FAIL, power is cut from the terminals.
FAIL
52
Page 53
OPERATION
20 0
E ROZ I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 07
IAF
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
L
MM A UN N A_M A N U = * * * -
0 051
V
2. The FAIL judgment will be held on
the display until the STOP button is pressed. Pressing the STOP button twice will return the tester to the READY status.
STOP
X2
3. The READY indicator will be lit
blue in the READY status.
READY
20 0
AER D Y
I N KL G B G - W CONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 07
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
READY status
MM A UN N A_M AN U = * * * -
V
E ROZ
Note
The buzzer will only sound if Fail Sound is set to ON. See page 67 for details.
FAIL Timing Diagrams
The timing diagrams below show the GB timing for the START status, TEST status and FAIL judgment. All test lines are active low.
GB FAIL Timing
Output I
time
START TEST FAIL
0.5s
Initial time
Total Test Time
(Approximately 100ms)
Signal IO
Time Cut = ON
Signal IO
Time Cut =
OFF
53
Page 54
GCT-9040 User Manual
Background
The Zeroing function is used to determine the resistance of the test leads for GB tests. When a zero check is performed, the reference is automatically set to the measured resistance of the test leads.
Steps
1. Ensure the tester is in VIEW status
for the current GB test. Save the current test if necessary.
Page 45
20 0
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MM A UN N A_M A N U = * * * -
V
2. Short the positive and negative alligator clips as
shown below.
3. Press the STOP button to put the
tester into the READY status.
STOP
Zeroing of the Test Leads
54
Page 55
OPERATION
4. The ZERO function can be
activated by pressing the corresponding soft-key in the READY status. The ZERO soft-key will be highlighted.
E R OZ
5. Press the START button to perform
the zero check. The tester will go into the ZERO status.
START
6. The TEST indicator will be lit
orange when in the ZERO status.
TEST
E ROZ
20 0
I N KL G B G - W C ONF T
E
F R EQ = 0 H z 6
1
0 00
A
V OL = .0 0 00
REZ
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
. 0 V ER F =# 0 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
O
MM A UN N A_MA N U = * * * -
0 051
V
Measured /V
ZERO status
7. When the zero check has finished, the tester
will return back to the VIEW status. The resistance of the test leads will be automatically set as the Reference value.
20 0
AER D Y
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 08
G B R = m MT I E = 0 0 1 . 0 SR
6 1 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
Reference value
MM A UN N A_MA N U = * * * -
V
E ROZ
Note
Remember to replace the test leads to the proper position on the DUT before testing.
55
Page 56
GCT-9040 User Manual
I<SET
If SOURCE H/L terminals are open or poorly connected, then an I<SET error will appear on the screen. Stop the test and re-check the connection again and try again.
20 0
I N KL G B G - W C ONF T
E
F R EQ = 0 H z 6 V OL = .0 0 00
G B R = 9 9 9 . 9 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
MM A UN N A_MA N U = * * * -
IAF
L< I S E T
I<SET error message
FAIL status
R = 0
Stop the test and perform the zero check again.
20 0
I N KL G B G - W C O NF T
E
F R E Q= 0 H z 6 V OL = .0 0 00
G B R = 0 0 0 . 0 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
MM A UN N A_MA N U = * * * -
IAF
L= R 0
R = 0 error message
FAIL status
56
Page 57
OPERATION
Linked ACW/DCW/IR Tests
This section describes how to perform additional ACW, DCW or IR tests with a slave GPT-9XXX unit using the GCT-9040 as the master unit. ACW and DCW tests can be performed before, after or simultaneously with a GB test. IR tests can only be performed before or after a GB test.
The LINK port on the GCT-9040 is used to control the linked slave unit. The slave unit can be used in MANU or AUTO mode, depending on the setup. In addition, remote commands can also be issued to slave units via the GCT-9040.
LINK Connection → from page 58 LINK Test Configuration→ from page 60 Run LINK Test→ from page 63
Before operating the GCT-9040 please read the safety precautions as outlined in the Set Up chapter on page 19 as well as the safety information in the user manuals’ of the GPT-9XXX units that function as the slave units when using the link connection.
57
Page 58
GCT-9040 User Manual
Background
Any GPT-98XX or GPT-99XX unit can be used in conjunction with the GCT-9040 to perform additional ACW, DCW or IR tests.
The LINK port on the GCT-9040 and the RS-232 port on the slave unit is used to connect both devices. The GPT-9XXX interface should be configured to the RS-232, with a baud of
115200.
Steps
1. Ensure the power is off on both the GCT-9040
and the GPT-9XXX slave unit.
2. Connect the LINK cable (GTL-132) to the LINK
port on the GCT-9040 and to the RS-232 port on the GPT-9XXX unit.
3. Connect the DUT as shown on page 30.
Note: Stack the units vertically to use the cable.
LINK port
RS-232 port
GCT-9040
master unit
GPT-9XXX
slave unit
LINK cable
LINK Connection
58
Page 59
OPERATION
Note
No more than 2 units should be stacked vertically.
59
Page 60
GCT-9040 User Manual
Note
First ensure the GPT-9XXX interface s set to RS­232 and that the baud rate is set to 115200. See the Remote Control chapter of the corresponding user manual for configuration details. Not setting the interface correctly will result in failure of the LINK mode connection.
Steps
1. Turn on both units after the link cable has been
connected*.
2. The GCT-9040 will be in VIEW mode upon
startup.
Note
*The GCT-9040 will perform a search to see if a slave unit is connected with the LINK cable upon startup. If the slave GPT-9XXX is found then the LINK connection will automatically be activated at startup. Note, however, that the GPT-9XXX unit will need to have been fully powered on before the GCT-9040 is powered on for the search to succeed.
Load GB Test
3. Use the scroll wheel on the GCT-
9040 to select a MANU test number. This will be the GB test used.
20 0
I N KL G B G - W C ONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MM A UN N A_M A N U = * * * -
V
LINK Test Configuration
60
Page 61
OPERATION
4. The GPT-9XXX unit will automatically go into
the RMT-VIEW status upon startup when connected with the link cable. The last used MANU or AUTO mode test will be loaded on startup.
20 0MA N U = * * * -
V I E W
H z RF E =Q 6 0
A CW D CW I R
E
0
500
k V
MT I E = 0 0 1 . 0 SR
. 0 m AER F =# 0 00
2 0 . 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_
m A
TMR
(GPT-9XXX screen)
Load ACW, DCW, IR Test
5. If you need to select a different
ACW, DCW or IR test, follow the steps below on the GPT-9XXX:
a. Press the STOP button on
the GPT-9XXX to exit from remote mode and return the unit to READY status.
STOP
b. Press the MANU/AUTO
key to select a MANU test.
OR
Hold the MANU/AUTO key to select an AUTO test.
MANU/AUTO
c. Use the scroll wheel to
select the desired MANU or AUTO test.
6. Press the LINK soft-key on the
GCT-9040 if the LINK connection is not already activated.
I N KL
61
Page 62
GCT-9040 User Manual
The LINK soft-key will be highlighted when the LINK connection is active.
20 0
I NKL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MM A UN N A_M A N U = * * * -
V
The GPT-9XXX will enter the RMT-VIEW mode if it wasn’t already.
7. On the GCT-9040 press the F1 soft-
key to toggle the order of the testing mode:
G + W
G-W
W-G
G+W
GB then ACW/DCW/IR test.
ACW/DCW/IR then GB test.
Simultaneously test GB with ACW or DCW. IR is not supported in this mode.
8. Press the F2 soft-key to set the test
FAIL behavior:
C ONF T
FCONT
FSTOP
If a test is judged FAIL, the tester will continue to the next test.
If a test is judged FAIL, the test will stop.
62
Page 63
Run LINK Test
Background
The operation for running a linked test is the same as running a single GB test.
Note
The tester cannot start to run a test under the following conditions:
Any protection modes have been tripped.
The INTERLOCK function is ON and the
Interlock key is not inserted in the signal I/O port (page 80).
The STOP signal has been received remotely.
If Double Action is ON, ensure the START button is pressed immediately after the STOP button (<0.5s).
Warning
Do not touch any terminals, test leads or the DUT when a test is running.
Start Testing
1. Press the STOP button on the GCT-
9040 to put the tester into READY status if it is not already (from the VIEW status).
STOP
2. Press the START button on the
GCT-9040 to start the LINK tests.
TEST will be displayed on the screen of the GCT-9040 or GPT­9XXX when a particular test is being run:
START
OPERATION
63
Page 64
GCT-9040 User Manual
GCT-9040 (GB) Test Example:
20 0
E ROZ I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
SET
G B R = 0 0 5 . 1 m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
T
MM A UN N A_M A N U = * * * -
0 051
V
GPT-9XXX (ACW/DCW/IR) Test Example:
20 0
T E S T
H z RF E =Q 6 0
A CW D CW I R
E
0
500
k V
MT I E = 0 0 1 . 0 SR
. 0 m AER F =# 0 00
2 0 . 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_M A N U = * * * -
000
0
m A
TMR
How the tests are performed depends on the testing mode:
G-W
The ground bond test will run first followed by the ACW/DCW/IR manual or AUTO test(s).
W-G
ACW/DCW/IR manual or AUTO test(s) run first followed by the ground bond test.
G+W
Simultaneously run the ground bond and the first ACW or DCW test. Any remaining tests on the slave unit will then run. IR is not supported in this mode.
Stop a Running Test
3. To stop the LINK test at any time
when it is running, press the STOP button. The current test will stop immediately. When the STOP button is pressed, a judgment is not made on the current test and any remaining tests are aborted.
STOP
64
Page 65
OPERATION
4. To put the tester back into READY
status, press the STOP button again.
STOP
Exit Testing
To exit testing, press the MANU key on the GCT-9040 when the tester is in the READY status. The tester will revert to the VIEW status for the current test.
MANU
20 0
I NKL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
EIV
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
W
MM A UN N A_M A N U = * * * -
V
Note
All panel keys except the STOP and START buttons are locked when the tester has been stopped.
All the results up until when the test was stopped are shown on-screen; GB results on the GCT-9040 screen, ACW/DCW/IR results on the GPT-9XXX slave unit.
Please see page 51 to see the PASS/FAIL details for GB tests. Please see the GPT-9XXX user manual for PASS/FAIL details for ACW, DCW and IT MANU/AUTO tests.
65
Page 66
GCT-9040 User Manual
Description
The LCD settings include contrast and brightness controls.
Steps
1. Ensure the tester is in VIEW status.
Save the current test if necessary.
Page 26
2. Press the UTILITY key.
UTILITY
3. Press the LCD soft-key to bring up
the LCD Common Utility menu.
L C D
T I ME
L C D C o t r an L C D B r g h ti
n se
L C D B U Z Z N T EI R C T R L
s : B R I G H T
s t : 5
MOC MO U N T I L I T Y
4. Use the UP/DOWN arrow keys to
choose a menu item: LCD Contrast, LCD Brightness.
Common Utility Settings
The Common Utility settings are system-wide settings that apply to all the MANU tests.
The Common Utility menu includes the following settings:
LCD Settings → from page 66. Buzzer Settings → from page 67. Interface Settings → from page 68. Control Settings → from page 70. Signal Time Settings → from page 72.
LCD Settings
66
Page 67
5. Use the scroll wheel to select a
parameter for the chosen menu item.
LCD Contrast LCD Brightness
1(low) ~ 8(high) BRIGHT, DARK
6. Press EDIT/SAVE to save the
settings and exit to VIEW status.
EDIT/SAVE
Note
The ESC key can be pressed at any time to cancel and exit back to VIEW status.
Buzzer Settings
Description
The Buzzer settings allow you to set whether the buzzer will sound for PASS/FAIL judgments. The buzzer time can also be set for the PASS/FAIL judgments. The buzzer settings are system-wide.
Steps
1. Ensure the tester is in VIEW status.
Save the current test if necessary.
Page 26
2. Press the UTILITY key.
UTILITY
3. Press the BUZZ soft-key to bring
up the Buzzer Common Utility menu.
B U Z Z
T I MEL C D B U Z Z N T EI R C T R L
MOC MO U N T I L I T Y
P a s s S u n do
: NO
F a i l S u n do
: FO F
T I M E : 0 0 . 50
S
OPERATION
67
Page 68
GCT-9040 User Manual
4. Use the UP/DOWN arrow keys to
choose a menu item: Pass Sound or Fail Sound.
5. Use the scroll wheel to select a
parameter for the chosen menu item.
Pass Sound Fail Sound
ON (000.2s~999.9s), OFF ON (000.2s~999.9s), OFF
6. Press EDIT/SAVE to save the
settings and exit to the VIEW status.
EDIT/SAVE
Note
For linked tests, the Pass Sound and Fail Sound settings also apply to the overall PASS/FAIL of the
all tests
, not just each individual test.
Note
The ESC key can be pressed at any time to cancel and exit back to VIEW status.
Description
The interface settings choose the remote interface configuration. USB and GPIB (optional) can be selected.
Steps
1. Ensure the tester is in VIEW status.
Save the current test if necessary.
Page 26
2. Press the UTILITY key.
UTILITY
3. Press the INTER soft-key to bring
up the Interface Common Utility menu.
N T EI R
Interface Settings
68
Page 69
OPERATION
T I ME
I n t
L C D B U Z Z N T EI R C T R L
MOC MO U N T I L I T Y
N Y (LU S B Of a c :e
e r
1 1 5 2 0 )0
Note
By default the interface is fixed to USB with a baud of 115200 when the GPIB interface is not installed.
4. Use the scroll wheel to select USB,
or GPIB (if installed).
5. For GPIB, use the UP/DOWN
arrow keys to choose Address.
6. Use the scroll wheel to select the
GPIB address.
GPIB address
0~30
7. Press EDIT/SAVE to save the
settings and exit to VIEW status.
EDIT/SAVE
Note
Ensure the GPIB address matches the host machine.
Note
The ESC key can be pressed at any time to cancel and exit back to VIEW status.
69
Page 70
GCT-9040 User Manual
Description
The Control settings are accessed in the COMMON UTILITY menu. The Control settings include: Start Control, Double Action, Key Lock and Interlock.
Start Control is used to determine how a test is started. Tests can be started via the front panel (START/STOP buttons), from a remote controller or via the SIGNAL I/O port.
The Double Action function is a safety feature used to prevent accidentally starting a test. Normally to start a test, the START button is pressed when the tester is in the READY status. To start a test when Double Action is ON, the STOP button must first be pressed, followed by the START button within 500ms.
Key Lock disables the front panel keys from changing the test number, mode or testing parameters. Only the Utility menu and any keys required for testing are not disabled.
The Interlock function is a safety feature. The interlock function prevents a test from running, unless the interlock pins on the signal I/O port connector are shorted. The included interlock key can be used for this purpose. See page 80 for details.
Steps
1. Ensure the tester is in VIEW status.
Save the current test if necessary.
Page 45
2. Press the UTILITY key.
UTILITY
Control Settings
70
Page 71
OPERATION
3. Press the CTRL soft-key to bring
up the Control Common Utility menu.
T R LC
T I ME
K e y L o k : Oc
F F
S t a r t t r lC
D o u b l e A c t
i no
L C D B U Z Z N T EI R C T R L
:
: F R ON T P A N
MOC MO U N T I L I T Y
E L
FO F
I N T E R L C K :O
O FF
4. Use the UP/DOWN arrow keys to
choose a menu item: Start Ctrl, Double Action, Key Lock or INTERLOCK.
5. Use the scroll wheel to select the
setting for the chosen menu item.
Start Ctrl
Double Action Key Lock INTERLOCK
FRONT PANEL, REMOTE CONNECT, SIGNAL IO ON, OFF ON, OFF ON, OFF
6. Press EDIT/SAVE to save the
settings and exit to VIEW status.
EDIT/SAVE
Note
The Double Action setting is ignored when the GCT-9040 is being controlled remotely using the USB or GPIB interfaces.
71
Page 72
GCT-9040 User Manual
Note
If a test is started with INTERLOCK ON, but the interlock signal I/O pins are not shorted (either with the included interlock key or manually), the INTERLOCK OPEN message will be displayed, preventing the test from starting.
AER D Y
I N KL G B G - W C O NF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
V
E ROZ
I N T E R L OC K O P E N
Interlock open message
Description
The Signal I/O Time Cut setting determines how the I/O fail signal is output for a FAIL or PASS judgment.
When set to OFF, the fail/pass signal stays low on a judgment.
time
FAIL or PASS
Pass/Fail judgment
When turned ON, the fail/pass signal is pulsed (active low) for the duration of the signal I/O time cut time + buzzer fail or pass time (see page 67).
time
FAIL
Judgment
I/O cut time + buzzer
pass/fail time
Steps
1. Ensure the tester is in VIEW status.
Save the current test if necessary.
Page 26
Signal I/O Time Cut Settings
72
Page 73
OPERATION
2. Press the UTILITY key.
UTILITY
3. Press the TIME soft-key to bring
up the Time Common Utility menu.
T I M E
T I ME
S I G
L C D B U Z Z N T EI R C T R L
MOC MO U N T I L I T Y
L I O
N A
C T :UT I ME S0 0 0 . 1
4. Use the scroll wheel to set the time
or to turn the option OFF.
SIGNAL IO TIME CUT
OFF, 000.1 ~ 999.9
5. Press EDIT/SAVE to save the
settings and exit to VIEW status.
EDIT/SAVE
Note
The ESC key can be pressed at any time to cancel and exit back to VIEW status.
73
Page 74
GCT-9040 User Manual
External Control Overview ............................................... 75
Remote Terminal Overview .................................................................... 75
Remote Controller Operation ................................................................. 76
SIGNAL I/O Overview ........................................................................... 77
Using the SIGNAL I/O to Start/Stop Tests ......................................... 79
Using the Interlock Key ........................................................................... 80
EXTERNAL CONTROL
The External Control chapter covers the REMOTE terminal and the SIGNAL I/O port.
74
Page 75
EXTERNAL CONTROL
Overview
The REMOTE terminal connector is a standard 5-pin DIN terminal suitable for a remote controller.
WARNING
Keep any cables that are connected to the REMOTE terminal away from the Source and Sense terminals.
Pin Assignment and Connection
REMOTE
1
4
2
5
3
RMT_START
RMT_STOP
Pin
Pin name
Description
1 RMT_STOP
Remote Stop signal
2 COM
Common line
3 Not used
4
RMT_START
Remote Start signal
5 Not used
Signal Properties
High level input voltage
2.4V~3.3V
Low level input voltage
0~0.8V
Input period
Minimum of 1ms
External Control Overview
The External Control section describes the front panel REMOTE terminal connection and the rear panel SIGNAL I/O port.
Remote Terminal Overview
75
Page 76
GCT-9040 User Manual
Description
The GCT-9040 accepts external remote controllers with a START and STOP button. To use the REMOTE terminal, the GCT-9040 must first be configured to accept a remote controller.
Operating a remote controller is the same as operating the START and STOP buttons on the front panel.
Steps
1. Insert the lead of the remote
controller into the REMOTE terminal.
REMOTE
2. Configure the Start Ctrl option to
REMOTE CONNECT in the Common Utility menu.
Page 70
3. The tester will now only be able to
start a test using a remote controller.
NOTE
Even if the GCT-9040 is configured to use the REMOTE CONNECT option, the STOP button on the front panel can still be used to stop a test.
4. To return the operation control to
the front panel, configure the Start Ctrl option to FRONT PANEL.
Page 70
Remote Controller Operation
76
Page 77
EXTERNAL CONTROL
Overview
The SIGNAL I/O port can be used to remotely start/stop tests and monitor the test status of the instrument. The SIGNAL I/O port is also used for the interlock function (page 70).
The SIGNAL I/O port uses a DB-9 pin female connector.
Pin Assignment
6 97 8
1 2 3 4 5
Pin name
Pin
Description
INTERLOCK1
1
When INTERLOCK is ON, a test is only allowed to start when both INTERLOCK pins are shorted.
INTERLOCK2
2
INPUT_COM
3
Common input line
INPUT_START
4
Start signal input
INPUT_STOP
5
Stop signal input
OUTPUT_TEST
6
Indicates that a test is in progress
OUTPUT_FAIL
7
Indicates that a test has failed
OUTPUT_PASS
8
Indicates that a test has passed
OUTPUT_COM
9
Common output line
Interlock connection
PIN 2 INTERLOCK2
PIN 1 INTERLOCK1
Input Connection
PIN 5 INPUT_STOP
PIN 3 INPUT_COM
PIN 4 INPUT_START
SIGNAL I/O Overview
77
Page 78
GCT-9040 User Manual
Output Connection
PIN 6 OUTPUT_TEST
PIN 9 OUTPUT_COM
PIN 8 OUTPUT_PASS
PIN 7 OUTPUT_FAIL
Signal Properties
Input Signals
High level input voltage
5V ~ 32V
Low level input voltage
0V ~ 1V
Low level input current
Maximum of -5mA
Input period
Minimum of 1ms
Output Signals
Output Type
Relay form A
Output Rated Voltage
30VDC
Maximum output current
0.5A
78
Page 79
EXTERNAL CONTROL
Background
To use the SIGNAL I/O port, the Start Ctrl settings have to be set to SIGNAL I/O in the Common Utility menu.
Panel operation
1. Set the Start Ctrl option to SIGNAL
I/O.
Page 70
2. Connect the Input/Output signals
to the SIGNAL I/O port.
3. To start the testing, short the
INPUT_STOP and INPUT_COM line for a minimum of 1ms to put the tester into READY status.
4. To start the testing, short the
INPUT_START and INPUT_COM lines for a minimum of 1ms.
5. To stop the testing, temporarily
short the INPUT_STOP and INPUT_COM line again.
NOTE
Even if the GCT-9040 is configured to use the SIGNAL I/O interface, the STOP button on the front panel can still be used to stop a test.
Using the SIGNAL I/O to Start/Stop Tests
79
Page 80
GCT-9040 User Manual
Background
When the INTERLOCK function is set to ON, tests are only allowed to start when both Interlock pins on the signal I/O port are shorted. Using the Interlock key will short the INTERLOCK1 and INTERLOCK2 pins on the signal I/O port. See page 77 for the Signal I/O pin assignment.
Panel operation
1. Insert the Interlock
key into the SIGNAL I/O port on the rear panel.
2. Set the INTERLOCK option to ON
in the Common Utility.
Page 70
Note
With INTERLOCK set to ON, the tester can now only start a test when the Interlock key is connected. Do not remove the interlock after starting a test. It must be still be connected after a test has started or is running.
Set INTERLOCK to OFF to disable this feature.
Using the Interlock Key
80
Page 81
REMOTE CONTROL
Interface Configuration ................................................... 82
Command Syntax ............................................................ 86
Command List ................................................................. 89
Error Messages .............................................................. 124
REMOTE CONTROL
This chapter describes basic configuration of IEEE488.2 based remote control. The remote interface supports USB and GPIB.
81
Page 82
GCT-9040 User Manual
USB Configuration
PC side connector
Type A, host
GCT-9040 side connector
Rear panel Type A
USB Class
CDC (communications device class) (VCP, Virtual Com Port)
Panel operation
1. Connect the USB cable to the rear
panel USB A port.
2. Set the interface to USB from the
Common Utility menu.
Page 68
Note
When USB is used for remote control, an RS232 port is simulated. Check the Windows Device Manager for the baud rate and other RS232 settings.
Note the baud rate is fixed to 115200 baud when using the USB interface.
Interface Configuration
USB Remote Interface
82
Page 83
REMOTE CONTROL
GPIB Configuration
Address
0-30
Panel operation
1. Connect the GPIB cable to the
rear panel GPIB port.
GPIB
2. Set the interface to GPIB and set
the GPIB address from the Common Utility menu.
Page 68
Functionality check
Invoke a terminal application such as Hyper Terminal.
To check the COM port number and other settings, see the Device Manager in the PC. For Windows 7; Control panel → Hardware and Sound Device Manager.
Run this query command via the terminal after the instrument has been configured for USB remote control (page 82).
*idn?
This should return the Model number, Serial number, and Firmware version in the following format:
GCT-9040, XXXXXXXXXXXX, V1.00
GPIB Remote Interface
The GPIB card is an optional extra. Please see page 21 for installation details.
USB Remote Control Function Check
83
Page 84
GCT-9040 User Manual
Model number : GCT-9040 Serial number :12 character serial number
Firmware version : V1.00
Display
When the panel is being remotely controlled via the USB or GPIB interfaces, RMT will be displayed on the screen.
20 0
I N KL G B G - W C ONF T
E
F R E Q = 0 H z 6
1
0 00
A
V OL = .0 0 00
G B R = m MT I E = 0 0 1 . 0 SR
0 0 V ER F =# . 00
V IH = .0 0 08
0 0 . 1 S0A M P =R
MM A UN N A_M A N U = * * * -
V
V I E W
TMR
Background
When the instrument is remotely controlled, all panel keys except the STOP button are disabled.
Steps
1. When RMT is on the display, press
the STOP button. The panel goes to the READY status.
STOP
2. From the READY status the tester
can go into one of two states: TEST or VIEW.
To put the tester into VIEW status,
press the MANU key.
To put the tester in TEST status,
press the START button. This will start the manual test. For more details on running a manual test, see page 47.
MANU
START
Return to Panel Control
84
Page 85
REMOTE CONTROL
Note
To put the tester back to RMT, simply issue another remote control command.
85
Page 86
GCT-9040 User Manual
Compatible Standard
IEEE488.2
Partial compatibility
SCPI, 1999
Partial compatibility
Command Structure
SCPI commands follow a tree-like structure, organized into nodes. Each level of the command tree is a node. Each keyword in an SCPI command represents each node in the command tree. Each keyword (node) of an SCPI command is separated by a colon (:).
For example, the diagram below shows an SCPI sub-structure and a command example.
MANU
ACW
CLOSet
VOLTage
CHISet
MANU:ACW:VOLTage
Command types
There are a number of different instrument commands and queries. A command sends instructions or data to the unit and a query receives data or status information from the unit.
Command types
Setting
A single or compound command with/without a parameter
Example
MANU:STEP 1
Command Syntax
86
Page 87
REMOTE CONTROL
Query
A query is a simple or compound command followed by a question mark (?). A parameter (data) is returned.
Example
MANU1:ACW:VOLTage?
Command Forms
Commands and queries have two different forms, long and short. The command syntax is written with the short form of the command in capitals and the remainder (long form) in lower case.
The commands can be written in capitals or lower-case, just so long as the short or long forms are complete. An incomplete command will not be recognized.
Below are examples of correctly written commands.
Long form
SYSTem:BUZZer:KEYSound SYSTEM:BUZZER:KEYSOUND system:buzzer:keysound
Short form
SYST:BUZZ:KEYS syst:buzz:keys
Command Format
1 32
MANU:STEP 100
1. Command header
2. Space
3. Parameter
Parameters
Type
Description
Example
<Boolean>
Boolean logic
0, 1
<NR1>
integers
0, 1, 2, 3
<NR2>
decimal numbers
0.1, 3.14, 8.5
87
Page 88
GCT-9040 User Manual
<NR3>
floating point
4.5e-1, 8.25e+1
<NRf>
any of NR1, 2, 3
1, 1.5, 4.5e-1
<string>
ASCII text string
TEST_NAME
Message Terminator
CR, LF
Carriage Return, Line feed code
88
Page 89
REMOTE CONTROL
System Commands
SYSTem:LCD:CONTrast ........................................... 91
SYSTem:LCD:BRIGhtness ......................................... 91
SYSTem:BUZZer:PSOUND ....................................... 92
SYSTem:BUZZer:FSOUND ....................................... 92
SYSTem:BUZZer:PTIMe ............................................ 92
SYSTem:BUZZer:FTIMe ............................................ 93
SYSTem:ERRor ........................................................... 93
SYSTem1:ERRor.......................................................... 94
SYSTem:GPIB:VERSion ............................................. 95
Function Commands
FUNCtion:TEST .......................................................... 96
FUNCtion1:TEST ........................................................ 97
MEASure<x>............................................................... 97
MAIN1:FUNCtion ...................................................... 98
Manual Commands
MANU:STEP ............................................................. 101
MANU1:STEP ........................................................... 101
MANU:NAME .......................................................... 101
MANU1:RTIMe ........................................................ 102
MANU1:EDIT:MODE .............................................. 102
MANU1:ACW:VOLTage ......................................... 103
MANU1:ACW:CHISet ............................................. 103
MANU1:ACW:CLOSet ............................................ 104
MANU1:ACW:TTIMe .............................................. 105
MANU1:ACW:FREQuency ..................................... 105
MANU1:ACW:REF .................................................. 106
MANU1:ACW:ARCCurrent ................................... 106
MANU1:DCW:VOLTage ......................................... 107
MANU1:DCW:CHISet ............................................. 107
MANU1:DCW:CLOSet ............................................ 108
MANU1:DCW:TTIMe .............................................. 109
MANU1:DCW:REF .................................................. 109
MANU1:DCW:ARCCurrent ................................... 110
MANU1:IR:VOLTage .............................................. 110
Command List for GCT-9040
89
Page 90
GCT-9040 User Manual
MANU1:IR:RHISet ................................................... 111
MANU1:IR:RLOSet .................................................. 111
MANU1:IR:TTIMe ................................................... 112
MANU1:IR:REF ........................................................ 112
MANU:GB:CUTMode ............................................. 113
MANU:GB:CURRent ............................................... 113
MANU:GB:RHISet ................................................... 114
MANU:GB:RLOSet .................................................. 114
MANU:GB:VHISet ................................................... 115
MANU:GB:VLOSet .................................................. 115
MANU:GB:TTIMe .................................................... 116
MANU:GB:FREQuency ........................................... 116
MANU:GB:REF ......................................................... 117
MANU:GB:ZEROCHECK ....................................... 117
MANU1:UTILity:ARCMode ................................... 118
MANU1:UTILity:PASShold .................................... 118
MANU1:UTILity:FAILmode .................................. 118
MANU1:UTILity:MAXHold ................................... 119
MANU1:UTILity:GROUNDMODE ....................... 119
MANU<x>:EDIT:SHOW ......................................... 120
TESTok:RETurn ........................................................ 120
Auto Commands
TESTok:RETurn ........................................................ 120
Common Commands
*CLS ............................................................................ 121
*IDN ........................................................................... 121
*IDN1 ......................................................................... 121
Remote Commands
*RMTOFF ................................................................... 123
Special Functions
InterLock Key Open ................................................. 123
90
Page 91
REMOTE CONTROL
SYSTem:LCD:CONTrast ........................................................................ 91
SYSTem:LCD:BRIGhtness ...................................................................... 91
SYSTem:BUZZer:PSOUND .................................................................... 92
SYSTem:BUZZer:FSOUND .................................................................... 92
SYSTem:BUZZer:PTIMe ......................................................................... 92
SYSTem:BUZZer:FTIMe ......................................................................... 93
SYSTem:ERRor ........................................................................................ 93
SYSTem1:ERRor ...................................................................................... 94
SYSTem:GPIB:VERSion .......................................................................... 95
SYSTem:LCD:CONTrast
Set
Query
Description
Sets the contrast of the LCD display from 1 (low) to 8 (bright).
Syntax
Query Syntax
SYSTem:LCD:CONTrast <NR1>
SYSTem:LCD:CONTrast?
Parameter/ Return parameter
<NR1>
1~8
Example
SYST:LCD:CONT 5
Sets the display contrast to 5.
SYSTem:LCD:BRIGhtness
Set
Query
Description
Sets the brightness of the LCD display from 1(dark) to 2(bright).
Syntax
Query Syntax
SYSTem:LCD:BRIGhtness <NR1>
SYSTem:LCD:BRIGhtness?
Parameter/ Return parameter
<NR1>
1 (dark), 2 (bright)
System Commands
91
Page 92
GCT-9040 User Manual
Example
SYST:LCD:BRIG 2
Sets the display brightness to bright.
SYSTem:BUZZer:PSOUND
Set
Query
Description
Turns the buzzer sound on or off for a PASS judgment.
Syntax
Query Syntax
SYSTem:BUZZer:PSOUND{ON|OFF}
SYSTem:BUZZer:PSOUND ?
Parameter/ Return parameter
ON
PASS Sound on.
OFF
PASS Sound off.
Example
SYST:BUZZ:PSOUND ON
Turns the buzzer sound on for PASS judgments.
SYSTem:BUZZer:FSOUND
Set
Query
Description
Turns the buzzer sound on or off for a FAIL judgment.
Syntax
Query Syntax
SYSTem:BUZZer:FSOUND{ON|OFF}
SYSTem:BUZZer:FSOUND ?
Parameter/ Return parameter
ON
FAIL Sound on.
OFF
FAIL Sound off.
Example
SYST:BUZZ:FSOUND ON
Turns the buzzer sound on for FAIL judgments.
SYSTem:BUZZer:PTIMe
Set
Query
Description
Sets the PASS sound duration in seconds.
Syntax
Query Syntax
SYSTem:BUZZer:PTIMe <NR2>
SYSTem:BUZZer:PTIMe?
Parameter/ Return parameter
<NR2>
0.2~999.9
92
Page 93
REMOTE CONTROL
Example
SYST:BUZZ:PTIM 1
Sets the buzzer to 1 second for a PASS judgment.
SYSTem:BUZZer:FTIMe
Set
Query
Description
Sets the FAIL Sound duration in seconds.
Syntax
Query Syntax
SYSTem:BUZZer:FTIMe <NR2>
SYSTem:BUZZer:FTIMe?
Parameter/ Return parameter
<NR2>
0.2~999.9
Example
SYST:BUZZ:FTIM 1
Sets the buzzer to 1 second for a FAIL judgment.
SYSTem:ERRor
Query
Description
Returns any errors in the output buffer for the GCT-9040. See the error code table below for details.
Query Syntax
SYSTem:ERRor ?
Return parameter
<string>
Returns an error string that includes an error code and an error description.
Error Code Table
Error code, Error description
0,No Error
20,Command Error
21,Value Error
22,String Error
23,Query Error
24,Mode Error
27,GBV > 7.2V
31,Current Setting Error
34,Resistance HI SET Error
35,Resistance LO SET Error
36,REF Setting Error
93
Page 94
GCT-9040 User Manual
37,Frequency Setting Error
40,TEST Time Setting Error
41, LINK Setting Error (LINK GPT-9800/9900 Error)
42, Voltage HISet Error (GCT-9040 CUT V HI SET Error)
43, Voltage LOSet Error (GCT-9040 CUT V LO SET Error)
45, Buffer Error (Command buffer overflow)
Example
SYST:ERR ?
>0,No Error
Returns “0,No Error” as the error message.
SYSTem1:ERRor
Query
Description
Returns any errors in the output buffer of the slave GPT-9XXX unit. See the error code table below for details.
Query Syntax
SYSTem1:ERRor ?
Return parameter
<string>
Returns an error string that includes an error code and an error description.
Error Code Table
Error code, Error description
0,No Error
20,Command Error
21,Value Error
22,String Error
23,Query Error
24,Mode Error
25,Time Error
26,DC Over 50W (GPT-98XX), DC Over 100W (GPT­ 99XX/99XXA
27,GBV > 5.4V
30,Voltage Setting Error
31,Current Setting Error
32,Current HI SET Error
33,Current LO SET Error
94
Page 95
REMOTE CONTROL
34,Resistance HI SET Error
35,Resistance LO SET Error
36,REF Setting Error
37,Frequency Setting Error
38,ARC Setting Error
39,RAMP Time Setting Error
40,TEST Time Setting Error
45,Buffer Error
60,Get Data = 0 (GPT-9900 only get SWEEP Data)
Example
SYST1:ERR ?
>0,No Error
Returns “0,No Error” as the error message.
SYSTem:GPIB:VERSion
Query
Description
Queries the GPIB version.
Query Syntax
SYSTem:GPIB:VERSion?
Return parameter
<string>
Returns: The GPIB version as a string “GPIB,V1.00” or “No GPIB connected” if there is not a GPIB device configured/connected.
Query Example
SYST:GPIB:VERS?
>GPIB,V1.00 Returns the GPIB version.
95
Page 96
GCT-9040 User Manual
FUNCtion:TEST ........................................................................................ 96
FUNCtion1:TEST ...................................................................................... 97
MEASure<x> ............................................................................................ 97
MAIN1:FUNCtion .................................................................................... 98
FUNCtion:TEST
Set
Query
Description
Turns the currently selected test (output) on or off for the GCT-9040. If a slave is connected using the LINK cable, its output will also turn on, according to the order of the test settings (G-W, W-G, G+W).
Setting the FUNCtion:TEST command to OFF at the end of a test will also temporarily turn the PASS/FAIL buzzer sound off.
Note: Setting the FUNCtion:TEST command to OFF will not turn the test off for a connected slave unit. Please use the FUNCtion1:TEST OFF command.
Syntax
Query Syntax
FUNCtion:TEST {ON|OFF}
FUNCtion:TEST?
Parameter ON
Turns the test on.
OFF
Turns the test off.
Return parameter TEST ON
Test is on.
TEST OFF
Test is off.
Example
FUNC:TEST ON
Turns the output on.
Function Commands
96
Page 97
REMOTE CONTROL
FUNCtion1:TEST
Set
Query
Description
Turns the currently selected test (output) on or off for a slave GPT-9XXX.
Setting the FUNCtion:TEST command to OFF at the end of a test will also temporarily turn the PASS/FAIL buzzer sound off.
Syntax
Query Syntax
FUNCtion1:TEST {ON|OFF}
FUNCtion1:TEST?
Parameter
ON
Turns the test on for the slave unit.
OFF
Turns the test off on for the slave unit.
Return parameter TEST ON
Test is on for the slave unit.
TEST OFF
Test is off for slave unit.
Example
FUNC1:TEST ON
Turns the output on for the slave unit.
MEASure<x>
Query
Description
Returns the test parameters & results of the tester.
MANU mode: Returns the test parameters & results of a MANU test for both the GCT-9040 and the slave, if connected.
AUTO mode: Returns the test parameters & results of the selected step (1-16) of the AUTO test of the slave, if connected.
Return parameters: function, judgment/status, test voltage, test current/resistance, test time (time of completed test) or ramp time (elapsed time of test that has not been completed.
Query Syntax
MEASure<x>?
Parameter (MANU mode)
No parameter needed for MANU mode.
97
Page 98
GCT-9040 User Manual
Parameter (AUTO mode)
<x>
<NR1>1~16. Step number.
Return parameter
<string>LN <string>
Returns a string with the test status of the GCT-9040 test, followed by another string of the slave test, if connected, in the following format: function, judgment or status, test voltage, test current or resistance, test time or ramp time
Function
ACW, DCW, IR, GB
Judgment /Status
PASS, FAIL VIEW
Test voltage
voltage+unit
Test current /Test resistance
current+unit resistance+unit
Test time /Ramp time
T=time+S R=time+S
Example (in MANU mode)
MEAS?
> GB ,PASS ,03.00A ,000.0mohm,T=001.0S
ACW, FAIL , 0.024kV ,0.013 mA ,R=000.1S
Returns the test result of the current manual test first (1st line) then the slave manual test(2nd line), if connected.
Example (in AUTO mode)
MEAS10?
>IR, FAIL ,0.225kV ,999M ohm,T=010.3S
Returns step 10 of the automatic result of the connected slave unit.
MAIN1:FUNCtion
Set
Query
Description
Changes the mode between AUTO and MANU on the slave unit when in LINK mode.
Syntax
Query Syntax
MAIN1:FUNCtion {MANU|AUTO}
MAIN1:FUNCtion?
98
Page 99
REMOTE CONTROL
Parameter/ Return parameter
MANU
Puts the tester mode to MANU.
AUTO
Puts the tester mode to AUTO.
Example
MAIN1:FUNC MANU
Sets the tester to MANU mode for the slave unit.
99
Page 100
GCT-9040 User Manual
MANU:STEP ........................................................................................... 101
MANU1:STEP ......................................................................................... 101
MANU:NAME ........................................................................................ 101
MANU1:RTIMe ...................................................................................... 102
MANU1:EDIT:MODE ............................................................................ 102
MANU1:ACW:VOLTage ...................................................................... 103
MANU1:ACW:CHISet ........................................................................... 103
MANU1:ACW:CLOSet .......................................................................... 104
MANU1:ACW:TTIMe ........................................................................... 105
MANU1:ACW:FREQuency .................................................................. 105
MANU1:ACW:REF ................................................................................ 106
MANU1:ACW:ARCCurrent ................................................................. 106
MANU1:DCW:VOLTage ...................................................................... 107
MANU1:DCW:CHISet ........................................................................... 107
MANU1:DCW:CLOSet .......................................................................... 108
MANU1:DCW:TTIMe ........................................................................... 109
MANU1:DCW:REF ................................................................................ 109
MANU1:DCW:ARCCurrent ................................................................. 110
MANU1:IR:VOLTage ............................................................................ 110
MANU1:IR:RHISet ................................................................................. 111
MANU1:IR:RLOSet ................................................................................ 111
MANU1:IR:TTIMe ................................................................................. 112
MANU1:IR:REF ...................................................................................... 112
MANU:GB:CUTMode ........................................................................... 113
MANU:GB:CURRent ............................................................................. 113
MANU:GB:RHISet ................................................................................. 114
MANU:GB:RLOSet ................................................................................ 114
MANU:GB:VHISet ................................................................................. 115
MANU:GB:VLOSet ................................................................................ 115
MANU:GB:TTIMe .................................................................................. 116
MANU:GB:FREQuency ......................................................................... 116
MANU:GB:REF ...................................................................................... 117
MANU:GB:ZEROCHECK ..................................................................... 117
MANU1:UTILity:ARCMode ................................................................ 118
MANU1:UTILity:PASShold .................................................................. 118
MANU1:UTILity:FAILmode ................................................................ 118
Manual Commands
100
Loading...