Instek AEL-5019-425-112.5, AEL-5004-350-37.5, AEL-5002-425-18.75, AEL-5023-350-112.5, AEL-5006-350-56 User Manual

...
AC/DC High Power Electronic Load
AEL-5000 Series
USER MANUAL
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Table of Contents
Table of Contents
SAFETY INSTRUCTIONS ................................................... 4
GETTING STARTED ........................................................... 8
The most complete measurement function ................ 9
AEL-5000 Series Introduction .................................. 11
Accessories.............................................................. 16
Operating Mode Description ................................... 18
Operating Area ........................................................ 22
Appearance .............................................................. 27
FUNCTION DESCRIPTION .............................................. 39
Function keys description ........................................ 40
Store or Recall functions ......................................... 56
Sequence Functions ................................................. 57
Wave Function description ....................................... 61
Test Function description ........................................ 71
Entry key description ............................................. 102
CONNECTION .............................................................. 104
Rear Panel ............................................................. 105
Connecting the I-monitor to an oscilloscope ......... 109
Master/Slave Description ...................................... 110
2 operating modes for Master/Slave ...................... 112
REMOTE operating ................................................ 117
INSTALLATION ............................................................. 120
Check line voltage .................................................. 121
Grounding requirements ....................................... 122
Power up ............................................................... 123
Connection to the load Input Terminal .................. 124
Interface Card ........................................................ 125
1
AEL-5000 User Manual
RS232 interface option ........................................... 126
GPIB interface option ............................................ 127
USB interface option .............................................. 127
LAN interface option ............................................. 127
I/O connection ...................................................... 128
Load wire inductance ............................................. 129
Parallel and three-phase control ............................ 133
REMOTE CONTROL ...................................................... 134
Interface Configuration .......................................... 135
Communication Interface programming command
list ......................................................................... 137
Command Syntax ................................................... 148
Command List ....................................................... 150
PRESET Commands ............................................... 152
Limit Commands ................................................... 165
STATE commands .................................................. 167
System Commands ................................................ 173
Measure Commands .............................................. 176
APPLICATION ............................................................... 178
Local sense connections ........................................ 179
Remote sense connections .................................... 181
Constant Current mode and LIN mode application 183
Constant Resistance mode application .................. 185
Constant Voltage mode application ....................... 187
Constant Power mode application ......................... 189
Battery discharge test application .......................... 191
Current protection component test ........................ 194
AC rectified load simulation................................... 197
Parallel operation .................................................. 198
Inrush Current ....................................................... 200
Power Supply OCP testing ..................................... 203
Power Supply OPP testing ..................................... 205
SHORT testing ....................................................... 207
2
Table of Contents
BW Setting............................................................. 209
Special waveform applications ............................... 210
APPENDIX .................................................................... 211
Replacing the Fuse................................................. 212
AEL-5000 Default Settings ..................................... 214
AEL-5000 Dimensions ............................................ 218
AEL-5000 series Specifications .............................. 222
Declaration of Conformity ...................................... 244
GPIB programming Example .................................. 245
AEL-5000 Series USB Instruction ........................... 249
AEL-5000 series Auto, Sequence funct ion provide EDIT, ENTER, EXIT, TEST and STORE 5 keys
operation ............................................................... 251
AEL-5000 Series LAN Instruction ........................... 255
3
AEL-5000 User Manual
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the instrument or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Earth (ground) Terminal
Frame or Chassis Terminal
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
4
SAFETY INSTRUCTIONS
General Guideline
CAUTION
Do not place any heavy object on the
instrument. Note: Only 2 units can be stacked vertically.
Avoid severe impact or rough handling that
leads to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only crimped wires, not bare wires, for the
terminals.
Do not block the cooling fan opening. Do not disassemble the instrument unless you
are qualified.
The equipment is not for measurements
performed for CAT II, III and IV.
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in
the building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
0 is for measurements performed on circuits not directly
connected to Mains.
Do NOT position the equipment so that it is difficult to disconnect the appliance inlet or the power plug.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Safety Guidelines
5
AEL-5000 User Manual
Power Supply
WARNING
AC Input voltage range: 100Vac~240Vac ± 10% Frequency: 47-63Hz Power for every model
Model
Power
AEL-5002-350-18.75, AEL-5003-350-28 AEL-5004-350-37.5 AEL-5002-425-18.75 AEL-5003-425-28 AEL-5004-425-37.5
AEL-5003-480-18.75 AEL-5004-480-28
150VA
AEL-5006-350-56 AEL-5008-350-75 AEL-5006-425-56 AEL-5008-425-75
270VA
AEL-5012-350-112.5 AEL-5012-425-112.5
390VA
AEL-5015-350-112.5 AEL-5015-425-112.5
510VA
AEL-5019-350-112.5 AEL-5019-425-112.5
630VA
AEL-5023-350-112.5 AEL-5023-425-112.5
750VA
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.
To avoid electric shock, the power cord
protective grounding conductor must be connected to ground. No operator serviceable components inside. Do not remove covers. Refer servicing to qualified personnel.
6
SAFETY INSTRUCTIONS
Cleaning
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Temperature: 0°C to 40°C Humidity: 0 to 85% RH Altitude: <2000m Overvoltage category II
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The instrument falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -20°C to 70°C Humidity: <90% RH
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
7
AEL-5000 User Manual
AEL-5000 LOAD Operating mode
AEL-5000 Ser ies Op erating Mo des
daoLCDdaoLCA
Sine Wave
Voltage Source
Constant Current
Mode
Linear Constant
Current Mode
Constant Resistance
Mode
Linear Constant
Current Mode
Constant Resistance
Mode
Constant Power
Mode
Constant Voltage
Mode
Rectifier Load
Non-linear Mode
Rectifier Load
Non-linear Mode
Constant Current
Mode
Linear Constant
Current Mode
Constant Resistance
Mode
Constant Power
Mode
Constant Voltage
Mode
Rectifier Load
Non-linear Mode
+
Constant Resistance
Rectifier Load
Non-linear Mode
+
Constant Resistance
PWM, Step, SquareWave
Voltage Source
GETTING STARTED
AEL-5000 Series is suitable for the step, square and sine wave of the AC Power device test. Especially for the uninterruptible power supply UPS, Inverter, fuses, circuit breakers, power regulator AVR, Battery, AC/ DC power supply/ components ... and so on, absolutely is the best test solution in the market.
8
GETTING STARTED
AEL-5000 test mode
Fuse / Br eaker Trip /
Non -trip Time Test
AC Power Sour ce Sh ort
Circ uit / OPP / OCP Test
DC P ower Sourc e Sh ort
Cir cuit / OPP / OC P Test
AEL-5000 Load Variou s Test Mo des
UPS
Eff icie ncy Test
PV Inverter Eff iciency Test
UPS Trans fer T ime Test
UPS Back-up Time Test
daoLCDdaoLCA
Battery Discharge Time Test
Fuse / Br eaker Trip /
Non -trip Time Test
Fuse / Br eaker Trip /
Non -trip Time Test
AC Power Sour ce Sh ort
Circ uit / OPP / OCP Test
Turbo Mode
Cu rrent / Pow er 2 Times
Fuse / Br eaker Trip /
Non -trip Time Test
Bat tery D ischa rge
Capa city Test
AC Pow er Sou rce S hort
Circ uit / OPP / OCP Test
Turbo Mode
Cu rrent / Pow er 2 Time s
The most complete measurement function
AEL-5000 Series AC/ DC electronic load has built-in 16-bit precision measurement circuit, providing accurate measurement values, measuring items include voltage rms (Vrms), current rms (Arms), watts (Watt), volt ampere (VA), crest factor (CF), power factor (PF), voltage total harmonic distortion (VTHD), voltage harmonics (VH), current total harmonic distortion (ITHD), current Harmonics (IH), peak current (Ipeak), maximum ampere (Amax), minimum ampere (Amin), maximum voltage (Vmax), and minimum voltage (Vmin).
In addition to these measurement functions, it also provides time
measurement, such as UPS back up time, fuses and circuit breakers’
trip or blow time and Off-line UPS transfer time. Note*1ms= milli - siemens = 1/kΩ Note*2The operating temperature range is 0 ~ 40° Caccuracy of
this specification is 25 ° C ± 5° C
9
AEL-5000 User Manual
AEL-5000 Series Introduction................................... 11
Model Line Up ................................................................................ 11
Main Features .................................................................................. 12
Protection features .......................................................................... 14
Accessories .............................................................. 16
AEL-5002-xxx-18.75/AEL-5003-xxx-28/
AEL-5004-xxx-37.5 ........................................................................ 16
AEL-5006-xxx-56/AEL-5008-xxx-78/AEL-5012-xxx-
112.5/AEL-5015-xxx-112.5/AEL-5019-xxx-112.5
/AEL-5023-xxx/112.5 ................................................................... 16
Operating Mode Description ................................... 18
AC load mode .................................................................................. 18
DC load mode ................................................................................. 20
Operating Area ........................................................ 22
Appearance .............................................................. 27
Front Panel ....................................................................................... 27
LCD Display .................................................................................... 29
10
GETTING STARTED
Model
Voltage (Volt)
Current
Power
AEL-5002-350-18.75
50~350Vrms/500Vdc
18.75 Arms/ 56.25Apeak
1875 W
AEL-5003-350-28
50~350Vrms/500Vdc
28 Arms / 84Apeak
2800W
AEL-5004-350-37.5
50~350Vrms/500Vdc
37.5 Arms / 112.5Apeak
3750 W
AEL-5006-350-56
50~350Vrms/500Vdc
56.0Arms/168Aprak
5600W
AEL-5008-350-75
50~350Vrms/500Vdc
75.0Arms/225Aprak
7500W
AEL-5012-350-112.5
50~350Vrms/500Vdc
112.5.0Arms/337.5Aprak
11250W
AEL-5015-350-112.5
50~350Vrms/500Vdc
112.5.0Arms/337.5Aprak
15000W
AEL-5019-350-112.5
50~350Vrms/500Vdc
112.5.0Arms/337.5Aprak
18750W
AEL-5023-350-112.5
50~350Vrms/500Vdc
112.5.0Arms/337.5Aprak
22500W
AEL-5002-425-18.75
50~425Vrms/600Vdc
18.75 Arms/ 56.25Apeak
1875 W
AEL-5003-425-28
50~425Vrms/600Vdc
28 Arms / 84Apeak
2800W
AEL-5004-425-37.5
50~425Vrms/600Vdc
37.5 Arms / 112.5Apeak
3750 W
AEL-5006-425-56
50~425Vrms/600Vdc
56.0Arms/168Aprak
5600W
AEL-5008-425-75
50~425Vrms/600Vdc
75.0Arms/225Aprak
7500W
AEL-5012-425-112.5
50~425Vrms/600Vdc
112.5.0Arms/337.5Aprak
11250W
AEL-5015-425-112.5
50~425Vrms/600Vdc
112.5.0Arms/337.5Aprak
15000W
AEL-5019-425-112.5
50~425Vrms/600Vdc
112.5.0Arms/337.5Aprak
18750W
AEL-5023-425-112.5
50~425Vrms/600Vdc
112.5.0Arms/337.5Aprak
22500W
AEL-5003-480-18.75
50~480Vrms/700Vdc
18.75 Arms / 56.25Apeak
2800W
AEL-5004-480-28
50~480Vrms/700Vdc
28 Arms / 84Apeak
3750 W
Model
Voltage (Volt)
Current
Power
AEL-5002-350-18.75
50~350Vrms/500Vdc
37.5Arms/56.25Apeak
3750W
AEL-5003-350-28
50~350Vrms/500Vdc
56Arms/84Apeak
5600W
AEL-5004-350-37.5
50~350Vrms/500Vdc
75.0Arms/112.5Apeak
7500W
AEL-5006-350-56
50~350Vrms/500Vdc
112.0Arms/168Aprak
11200W
AEL-5008-350-75
50~350Vrms/500Vdc
150.0Arms/225Aprak
15000W
AEL-5012-350-112.5
50~350Vrms/500Vdc
225.0Arms/337.5Aprak
22500W
AEL-5015-350-112.5
50~350Vrms/500Vdc
225.0Arms/337.5Aprak
30000W
AEL-5019-350-112.5
50~350Vrms/500Vdc
225.0Arms/337.5Aprak
37500W
AEL-5000 Series Introduction
Model Line Up
When Turbo is off
When Turbo is on
11
AEL-5000 User Manual
AEL-5023-350-112.5
50~350Vrms/500Vdc
225.0Arms/337.5Aprak
45000W
AEL-5002-425-18.75
50~425Vrms/600Vdc
37.5Arms/56.25Apeak
3750W
AEL-5003-425-28
50~425Vrms/600Vdc
56Arms/84Apeak
5600W
AEL-5004-425-37.5
50~425Vrms/600Vdc
75.0Arms/112.5Apeak
7500W
AEL-5006-425-56
50~425Vrms/600Vdc
112.0Arms/168Aprak
11200W
AEL-5008-425-75
50~425Vrms/600Vdc
150.0Arms/225Aprak
15000W
AEL-5012-425-112.5
50~425Vrms/600Vdc
225.0Arms/337.5Aprak
22500W
AEL-5015-425-112.5
50~425Vrms/600Vdc
225.0Arms/337.5Aprak
30000W
AEL-5019-425-112.5
50~425Vrms/600Vdc
225.0Arms/337.5Aprak
37500W
AEL-5023-425-112.5
50~425Vrms/600Vdc
225.0Arms/337.5Aprak
45000W
AEL-5003-480-18.75
50~480Vrms/700Vdc
37.5Arms/56.25Apeak
5600W
AEL-5004-480-28
50~480Vrms/700Vdc
56Arms/84Apeak
7500W
Performance
Four meters can be displayed V/A/W Meter,
display the Voltage (Vrms, Vpeak, Vmax., Vmin), Current (Irms, I Peak, Imax. Imin.) Watt, Voltampere (VA), Frequency, Crest Factor, Power Factor, Total Harmonic Distortion of Voltage (VTHD), Voltage Harmonic (VH), Total Harmonic Distortion of Current (ITHD), Current Harmonic (IH) Remote Control via a choice of Computer interfaces.
Support on-load boot; at first set Load ON to
support on-load boot, inverter or uninterruptible power supply is turned on directly with the set load current, used to verify whether the starter is stable when the Inverter is connected.
Supports the loading and unloading angle
control; the loading and unloading angle control, the full range of 0-359 degrees can be set to verify whether the Inverter output voltage transient response is stable when the actual electrical plugging and unplugging, and whether Overshoot/Undershoot is within the
Main Features
12
GETTING STARTED
allowable range.
Support positive half-cycle or negative half-
cycle loading; used to verify whether the inverter output voltage remains stable when the actual appliance has only positive half­cycle or negative half-cycle load current.
Supports SCR/TRIAC current phase
modulation waveforms, 90 degree Trailing edge and Leading Edge.
Supports the Inrush Current of the power
supply at startup and the Surge Current test when the load is suddenly plugged in (Hot Plug-in).
Features
AC / DC load with CC, Linear CC, CR, CV, CP
and Rectifier Load mode
Frequency Range : DC, 40~440Hz Crest factor adjustable range : 1.4~5.0 Power factor (PF) adjustable range: 0~1 lead or
(~1~0)lag
Built-in test modes include UPS Efficiency, PV
Inverter Efficiency, UPS Back-up time, Battery Discharge time, UPS transfer time, Fuse/ Breaker Trip / Non-Trip, short circuit simulation, OCP, OPP, etc.
Turbo mode, which can withstand up to twice
the current (225A) and power (45KW) electronic load in a short time, the most suitable for Fuse / Breaker and AC power short circuit, OCP, OPP test.
Up to three parallel up to 22500W and three-
phase or Y load synchronization control.
Maximum power of single-phase can up to
180KW, 3-phase total power up to 540KW 3­phase or Y connection parallel connection can be controlled by external voltage for CC, Linear CC, CR, CP, CV mode (Option)
Measure the fuse and circuit breaker trip or
13
AEL-5000 User Manual
blow time
Measure the UPS OFF- line transfer time
(Transfer time)
Perform short circuit simulation(can set the
short circuit time), OCP, OPP test
Over voltage warning, over current, over
power, over temperature protection.
150 set Store/Recall memory.
Interface
Optional interface: GPIB, RS232, USB, LAN.
Overvoltage protection
The Electronic Load input will turn OFF if the overvoltage circuit is tripped. The message OVP will be displayed on the LCD. When the OVP fault has been removed the load can be set to sink power again. While the unit will attempt to protect itself given an OVP state it is strongly advised to guard against any potential OVP fault state by using external protection and the correctly rated electronic load.
The Overvoltage protection circuit is set at a predetermined voltage and cannot be adjusted. The OVP level is 105% of the AEL-5000 Series nominal voltage rating.
Caution
Never apply an AC voltage to the input of the AEL-5000 Series Load. Do not apply a DC voltage that is higher than AEL-5000 Series Load rating. If this advice is ignored it is likely that damage will be caused to the electronic load module. This damage will not be covered by the warranty.
Over current protection (OCP)
The OCP protection will engage if the current being taken by the load reaches 105% of the load
module’s maximum current. The message OCP
Protection features
The protection features of the AEL-5000 series electronic load modules are as follows:
14
GETTING STARTED
will be displayed on the front panel and the unit will switch to its LOAD OFF state. Once the source of the over current has been removed the load can be switched on again.
Over power protection (OPP)
The AEL-5000 Series Electronic Load monitors the power dissipation level. The input to the load is automatically switched to LOAD OFF if the power dissipation is greater than 105% of the rated power input. If an over power condition occurs the display will show OPP
Over temperature protection
The load internal temperature at the heat sink is monitored. If the temperature reaches approximately 100°C the OTP message will be displayed and the unit will automatically switch to the LOAD OFF state. If an OTP error occurs please check the ambient temperature is between 0 to 40°C. Also ensure that the front and rear air vents of the mainframe are not obstructed. The air flow is taken from the front of the mainframe and exhausted from the rear. Therefore a suitable gap needs to be left at the rear of the mainframe. A minimum of 15cm is recommended. After a suitable cooling period the load can be switched.
15
AEL-5000 User Manual
Standard Accessories
Description
PCs
AEL-5000 series operation manual
It can be downloaded from GW instek website.
1 Terminal PTV1-12; PIN TRML 6
6
SLS10B RED; PLUG CONN 20A RED
1
SLS10B BLK; PLUG CONN 20A BLK
1 RNB 22-6S RING TRML,#4
2 HD-DSUB
15pin MALE to MALE 150cm
1
Standard Accessories
Description
PCs
AEL-5000 series operation manual
It can be downloaded from GW instek website.
Round terminal PVL-1-4
2
Round terminal RNYBS8-4
2
Terminal PTV1-12
6
HD-DSUB
15pin MALE to MALE 150cm
1 Optional Accessories
Description
PCs
GPIB+RS232 interface
PEL-030
1
RS232 interface
PEL-023
1
GPIB interface
PEL-022
1
Accessories
AEL-5002-xxx-18.75/AEL-5003-xxx-28/AEL-5004-xxx-37.5
AEL-5006-xxx-56/AEL-5008-xxx-78/AEL-5012-xxx-112.5/AEL­5015-xxx-112.5/AEL-5019-xxx-112.5/AEL-5023-xxx/112.5
16
GETTING STARTED
USB interface + USB driver(The driver can be downloaded from GW instek website)
PEL-025
1
LAN interface + LAN driver (The driver can be downloaded from GW instek website)
PEL-024
1
GPIB cable
GTL-250 GPIB Cable,0.6m
1
GPIB cable
GTL-248 GPIB Cable,2m
1
USB cable
GTL-246 USB Cable,1.2m
1
AEL-5000, AEL-5006, AEL-5008, AEL-5012 and AEL-5015 handle
PEL-028
1
AEL-5002, AEL-5003 and AEL-5004 handle
PEL-029
1
AEL-5000, AEL-5000 GPIB and RS­232 interface
PEL-030
1
17
AEL-5000 User Manual
CC Mode
With the operating mode of Constant Current, the AEL-5000 Series electronic load will sink a current in accordance with the programmed value regardless of the input voltage
Current Setting
Linear C.C. Mode
During Linear C.C. mode, the load current input into AEL-5000 Series High Power Electronic Load depends on the current setting regardless of the input voltage, e.g., the current setting remains unchanged. Please refer to Fig.1-8. The load input current signal will follow input voltage signal that is useful for step wave-form and square wave­form device.
Current Setting
CR Mode
At Constant Resistance mode, the AEL-5000 Series Electronic Load will sink a current linearly proportional to the load input voltage in accordance with the programmed resistance setting
Operating Mode Description
AC load mode
18
GETTING STARTED
Resistance Setting
CP Mode
At Constant Power mode, the AEL-5000 Series Electronic Load will attempt to sink load power (load voltage * load current) in accordance with the programmed power.
Power Setting
CV Mode
At Constant Voltage mode, the AEL-5000 Series Electronic Load will attempt to sink enough current until the load input voltage reaches the programmed value.
Voltage Setting
19
DC load mode
CC Mode
With the operating mode of Constant Current, the AEL-5000 Series electronic load will sink a current in accordance with the programmed value regardless of the input voltage
Current Setting
CR Mode
At Constant Resistance mode, the AEL-5000 Series Electronic Load will sink a current linearly proportional to the load input voltage in accordance with the programmed resistance setting
Resistance Setting
CP Mode
At Constant Power mode, the AEL-5000 Series Electronic Load will attempt to sink load power (load voltage * load current) in accordance with the programmed power.
AEL-5000 User Manual
20
GETTING STARTED
Power Setting
CV Mode
At Constant Voltage mode, the AEL-5000 Series Electronic Load will attempt to sink enough current until the load input voltage reaches the programmed value.
Voltage Setting
21
AEL-5000 User Manual
AEL-5002-350-18.75
350V
10V
18.75A6.25A
1875W
Power Curve
Voltage
Current
100V
50V
12.5A 18.5A
20V
30V
5.35A
AEL-5003-350-28
350V
10V
28A8A
2800W
Power Curv
Voltage
Current
100V
50V
18.75A 27.75A
20V
30V
9.375A
AEL-5004-350-37.5
350V
10V
37.5A
12.5A
3750W
Power Curve
Voltage
Current
100V
50V
25A 37A
20V
30V
10.7A
AEL-5002-425-18.75
425V
10V
18.75A6.25A
1875W
Power Curve
Voltage
Current
100V
50V
12.5A 18.5A
20V
30V
4.4A
Operating Area
AEL-5000 Series AC/DC electronic load can be used to work with GPIB, RS232, USB or LAN interface and panel manual operation can be made available.
The electronic load operating environment temperature is 0 ° C ~ 40 ° C, full power operation for a period of time may produce OTP.
22
GETTING STARTED
AEL-5003-425-28
425V
10V
28A
6.58A
2800W
Power Curv
Voltage
Current
100V
50V
18.75A 27.75A
20V
30V
9.375A
AEL-5004-425-37.5
425V
10V
37.5A
12.5A
3750W
Power Curve
Voltage
Current
100V
50V
25A 37A
20V
30V
8.8A
AEL-5006-350-56
350V
10V
56A
18.75A
5600W
Power Curve
Voltage
Current
100V
50V
37.5A 55A
20V
30V
16A
AEL-5008-350-75
350V
10V
75A
25A
7500W
Power Curve
Voltage
Current
100V
50V
50A 74A
20V
30V
21.4A
AEL-5012-350-112.5
350V
10V
112.5A32.1A
11250W
Power Curve
Voltage
Current
100V
50V
75A 111A
20V
30V
37.5A
23
AEL-5000 User Manual
AEL-5015-350-112.5
350V
112.5A
15000W
Power Curve
Voltage
Current
133.3V
50V
75A 111A
20V
30V
42.8A
AEL-5019-350-112.5
350V
112.5A
18750W
Power Curve
Voltage
Current
166.6V
50V
75A
111A
20V
30V
53.5A
AEL-5023-350-112.5
350V
112.5A
22500W
Power Curve
Voltage
Current
200V
50V
75A 111A
20V
30V
64.2A
AEL-5006-425-56
425V
10V
56A
18.75A
5600W
Power Curve
Voltage
Current
100V
50V
37.5A 55A
20V
30V
13.1A
AEL-5008-425-75
425V
10V
75A
25A
7500W
Power Curve
Voltage
Current
100V
50V
50A 74A
20V
30V
17.6A
24
GETTING STARTED
AEL-5012-425-112.5
425V
10V
112.5A26.4A
11250W
Power Curve
Voltage
Current
100V
50V
75A 111A
20V
30V
37.5A
AEL-5015-425-112.5
425V
112.5A
15000W
Power Curve
Voltage
Current
133.3V
50V
75A 111A
20V
30V
35.2A
AEL-5019-425-112.5
425V
112.5A
18750W
Power Curve
Voltage
Current
166.6V
50V
75A
111A
20V
30V
44.1A
AEL-5023-425-112.5
425V
112.5A
22500W
Power Curve
Voltage
Current
200V
50V
75A 111A
20V
30V
52.9A
AEL-5003-480-18.75
480V
18.75A5.8A
2800W
Power Curve
Voltage
Current
149V
50V
12.5A
20V
25
AEL-5000 User Manual
AEL-5004-480-28
480V
28A
3750W
Power Curve
Voltage
Current
133V
50V
18.6A
20V
7.8A
26
GETTING STARTED
1
LCD Multi­function display
Four meters can display the voltage value at the same time Voltage(Vrms, Vpeak, Vmax, Vmin), Current (Irms, Ipeak, Imax, Imin.) , Watt, Voltampere (VA), Frequency, Crest Factor, Power Factor, Total Harmonic Distortion of Voltage(VTHD), Voltage Harmonic (VH), Total Harmonic Distortion of Current (ITHD), Current Harmonic (IH)
2
Meter Switch button
V/AW keys can set the display Rms/Peak/Max/Min, Meter key can select PF/CF/FREQ, switchable display WATT/VA/VAR keys, THD keys choose to display THD.
3
Operate function keys
Mode, Preset ON/OFF, Load ON/OFF, Sense ON/OFF, Level A/B, Config, Limit, Recall, Store, SET, Local, System operate keys.
4
Waveform library keys
These keys can be quickly set CF /2/2.5/3/3.5, PF0.6/0.7/0.8/0.9/1.0, FREQ Auto/50Hz/60Hz/400Hz
Appearance
Front Panel
27
AEL-5000 User Manual
5
Test function keys
These keys can select Short/OPP/OCP/Non­L/NL-CR/Fuse/Batt(Battery Discharge)Trans(UPS transfer time) test functions.
6
Number keypad
7
Knob setting
8
Power switch
9
Cursor and button setting
28