Inspiration Tecotherm Neo Quick Setup Manual

Tecotherm Neo
Quick Set-up Guide
Prior to operation: Refer to Tecotherm Neo Instructions for Use.
You will need:
Tecotherm Neo
TECOmed Coolant Fluid
Connecting Hoses
Aqua Pad/ Wrap Mattress
Interlayer
Pre-Fill Mattress
A minimum volume of 250 ml TECOmed fluid must be filled into mattress
Connect the male mattress connectors to the female connectors of the
fill-up bottle
Turn bottle until cap is directed downwards and keep it in a position 0.5m
higher than the mattress
Fill-Up the Tecotherm Neo with TECOmed fluid
Ensure mattress hoses are disconnected from the lower pair of device ports (6)
Undo fill-up bottle and fill with TECOmed fluid up to the 450 ml mark
Connect fill-up set to refill port connector counterparts at the device interface (7)
Green marked connector always fits in to the right hand side port.
Turn on the Tecotherm Neo, select and start treatment mode
“Constant mattress temperature”
Turn bottle until cap is directed downwards
Wait until the fluid level inside fill-up bottle stops falling
Now quit treatment mode
Attach the prefilled mattress to hoses and connect to lower device ports (6)
Connect the green marked male connector of the fill-up bottle to the right refill
port connector. This time do not connect the left hand male connector
Select “Constant mattress temperature” and fill the Tecotherm with another
100 ml of TECOmed fluid
Disconnect fill up set
Note: Move mattress around to ensure all air bubbles have been removed.
Optional:
Rectal Probe
Skin Probe
Treatment Modes
In the main menu select the treatment mode required by scrolling to one of,
Complete Treatment Mode, Servo Mode or Constant Mattress Temperature Mode
Once selected, scroll down and select the parameters to be changed
as appropriate
Once all parameters are selected press start
Note: If using either servo mode you will be asked to insert a rectal probe into the patient and connect to the Tecotherm Neo, if doing so it is recommended to do so by approx 6 cm without forcing the probe so as not to cause internal damage to t he patient.
Caution: In servo modes the Tecotherm Neo first pre-conditions the mattress temperature, once rectal probes are inserted press ‘START’ to commence servo mode otherwise the Tecotherm Neo stays in constant mattress mode.
FEMALE CONNECTORS
NON- RETURN VALVE
GREEN CONNECTOR
(7) FILL UP CONNECTION*
PORTS
(T5) ALARM SILENCER*
(1) POWER ON/OFF*
(6) MATTRESS
CONNECTION PORTS*
*Numbers relate to those shown in the Instructions for Use.