Inspect USA CO2-9914SD Operation Manual

one probe, SD card real time data recorder
CO2, %RH, Temp., Dew point, Wet bulb
CO2 METER + HUMIDITY/Temp.
Model : CO2-9914SD
Your purchase of this CO2 METER ,+ HUMIDITY/Temp. with SD CARD DATA RECORDER marks a step forward for you into the field of precision measurement. Although this Meter is a complex and delicate instrument, its durable structure
this manual within easy reach.
OPERATION MANUAL
TABLE OF CONTENTS
1. FEATURES.............................................................................................................
1
2. SPECIFICATIONS.......................................................................................................
2
2-1 General Specifications...............................................................................................
2
2-2 Electrical Specifications........................................................................
4
3. FRONT PANEL DESCRIPTION...........................................................................
6
4
MEASURING PROCEDURE............................................................................................
8
5. OTHER FUNCTION...................................................................................................
10
5-1 Data Hold........................................................................................................................
10
5-2 Record ( Max./ Min. reading )....................................................................
10
5-3 LCD Backlight ON/OFF.....................................................................................
10
6. DATALOGGER...............................................................................................................
11
6-1 Preparation before execute datalogger function...............................................................................
11
6-2 Auto Datalogger ( Set sampling time 1 second )...........................................................................
12
6-3 Manual Datalogger ( Set sampling time = 0 second ).........................................................................
13
6-4 Check time information.....................................................................................................................
13
6-5 Check sampling time information....................................................................................................
14
6-6 SD Card Data structure......................................................................................................
14
7. Saving data from the SD card to the computer.......................................................
15
8. ADVANCED SETTING....................................................................................................
17
9. POWER SUPPLY from DC ADAPTER..........................................................................
22
10. BATTERY REPLACEMENT............................................................................................
22
11. SYSTEM RESET................................................................................................................
23
12. RS232 PC SERIAL INTERFACE....................................................................................
23
13. PATENT....................................................................................................................
25
1. FEATURES
* Real time recorder, save the data into the SD memory
card and can be down load to the Excel, extra software
is no need. User can make the further data or graphic analysis by themselves. under the Excel software.
* At the same time, the SD memory card can record `1
probe 4 Function data ( CO2/Temp. , %RH/Temp. /Dew /Wet ) two group analog data with the time information into the one Excel file at the same time.
* Manual datalogger is available, during execute the
manual datalogger function, it can set the different location no. ( position 1 to position 99 ).
* Air quality measurement application, multi-function :
CO2 (Carbon dioxide ), Humidity/Dew/Wet , temperature measurement. * CO2 range : 0 to 10,000 ppm x 1 ppm. * Humidity range: 10 to 95 %RH. x 0.1 % RH * Dew point Temp. and Wet bulb Temp. measurement. * Temp. range : 0 to 50.0 , /. * CO2 sensor : NDIR, long term reliability. * Humidity sensor : Precision capacitance sensor * Alarm setting with the beeper sound output. * Sampling time for data recorder is 2 seconds to 3600 seconds. * Complete set with 1 probes :
CO2/Temp.,Humidity/Temp probe , main meter and the hard
carrying case. * Separate probe, easy for remote measurement. * Meter can cooperate with 2 GB to 16 GB SD card, SD
card is optional. * RS232/USB computer interface. * Patented.
1
2. SPECIFICATIONS
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 52 mm x 38 mm (Dot Matrix)
LCD with green backlight ( ON/OFF ).
Measurement CO2 (Carbon dioxide )
Humidity Dew point Temp., Wet bulb Temp. Temperature
Sensor
CO2
NDIR
* Nondispersive infrared sensor
structure
Humidity
Precision capacitance sensor
Temp.
Precision thermistor
Datalogger
Auto
2 sec to 3600 sec.
Sampling Time
@ Sampling time can set to 1 second,
Setting range
but memory data may loss.
Manual Push the data logger button
once will save data one time.
@ Set the sampling time to
0 second.
@ Manual mode, can also select the
1 to 99 position ( Location ) no.
Data error no. 0.1% of total saved data max. Memory Card SD memory card. 1 GB to 16 GB. Advanced
* SD memory card Format
setting
* Set clock time *
Set sampling time
@ main setting * Auto power OFF management
* Set beep Sound ON/OFF * Decimal point of SD card setting * Temp. unit setting * Alarm value setting * Altitude value setting
2
Data Hold Freeze the display reading. Memory Recall Maximum & Minimum value. Sampling Time Approx. 1 second. of Display Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
Power Supply * Alkaline DC 1.5 V battery
( UM3, AA ) x 6 PCs, or equivalent.
*
DC 9V adapter input. ( AC/DC power
adapter is optional ).
Power Current
CO2 Normal operation ( w/o SD card save measure- data and LCD Backlight is OFF) : ment Approx. DC 136.5 mA.
When SD card save the data and LCD
Backlight is OFF) :
Approx. DC 166 mA.
Humidity Normal operation ( w/o SD card save measure- data and LCD Backlight is OFF) : ment Approx. DC 10.5 mA.
When SD card save the data and LCD
Backlight is OFF) :
Approx. DC 40 mA.
*
If LCD backlight on, the power
consumption will increase approx. 12 mA.
Operating 0 to 50 . ( 32 to 122 ). Temperature Operating Less than 80% R.H. Humidity Weight 350 g/0.77 LB. Dimension Meter 177 x 68 x 45 mm
CO2/Humidity length 175 mm x 28 probe
3
Accessories
Instruction manual....................................................................
1 PC
Included
Hard carrying case, CA-08............................................
1 PC
CO2 / Humidity probe..................................................
1 PC
Optional
SD memory card ( 4 GB )
Accessories
AC to DC 9V adapter.
USB cable, USB-01. RS232 cable, UPCB-02. Data Acquisition software, SW-U801-WIN.
2-2 Electrical Specifications (23±5 ℃)
CO2 ( Carbon dioxide )
Range 0 to 10,000 ppm CO2 Resolution 1 ppm ( Carbon
Accuracy
±40 ppm
dioxide ) * ≦1,000 ppm.
±(50ppm +3% of reading)
23 ± 5 . * > 1,000 ppm ≦ 3,000 ppm.
±(50ppm +5% of reading) typically
* > 3,000 ppm ≦ 10,000 ppm.,
Repeatability ± 20 ppm * ≦ 3,000 ppm. Temperature Range 0 to 50 ,32 to 122 .
Resolution 0.1 degree
Accuracy
: ±0.8 : ±1.5 .
Humidity/Temperature
Range 5 % to 95 % R.H. Humidity Resolution 0.1 % R.H.
Accuracy 70% RH :
±(3% reading + 1% RH).
< 70% RH :
±3% RH.
Range 0 to 50 ,32 to 122 . Temperature Resolution 0.1 degree
Accuracy ±0.8 .
±1.5 .
4
Dew Point Temp. ( Humidity )
Range -25.3 to 48.9
Resolution 0.1
Range
-13.5 to 120.1 .
Resolution 0.1 .
Remark :
* Dew Point display value is calculated from the
Humidity/Temp. measurement automatically.
*
The Dew Point accuracy is sum accuracy value of Humidity
& Temperature measurement..
Wet bulb Temp. ( Humidity )
Range -21.6 to 50.0
Resolution 0.1
Range -6.9 to 122.0 .
Resolution 0.1 .
Remark :
* Wet bulb display value is calculated from the Humidity/Temp.
measurement automatically.
* The Welt bulb accuracy is sum accuracy value of Humidity
& Temperature measurement..
@ Above specification tests under the environment RF Field Strength less than 3
V/M & frequency less than 30 MHz only.
5
3. FRONT PANEL DESCRIPTION
Fig. 1
6
3-1 Display
3-2 Power Button ( Backlight Button ) 3-3 Hold Button ( ESC Button ) 3-4 REC Button
3-5 Alarm Button ( ▲ Button )
3-6 Function Button ( ▼ Button )
3-7 Time Button ( SET Button, Sampling check )
3-8 Logger Button ( Enter Button ) 3-9 Stand 3-10 Battery Compartment/Cover 3-11 Battery Cover Screw 3-12 Tripod Fix Nut 3-13 CO2/Humidity Probe Input Socket 3-16 DC 9V Power Adapter Input Socket 3-17 Reset Button
3-18 RS-232 Output Terminal
3-19 SD Card Socket 3-20 Humidity Sensing Head 3-21 CO2/Humidity Probe Plug 3-22 CO2 Sensing Head
7
Loading...