Inspect EB612, UV888 User Manual

1
EN
Portable UV Monitor
Model: EB612 / UV888
USER MANUAL
CONTENTS
Contents................................................................ 1
Overview................................................................ 2
Front View............................................................ 2
Back View............................................................ 2
LCD Display.......................................................... 3
Getting Started........................................................ 3
Strap..................................................................... 3
Batteries................................................................ 3
Enter Settings....................................................... 4
Display ................................................................... 4
UVI Display........................................................... 4
UV Exposure Timer................................................. 5
UV Exposure Countdown Timer............................ 5
Manual Countdown Timer...................................... 5
Settings.................................................................. 6
Key Lock................................................................ 6
Reset .............................................................................. 6
Precautions............................................................ 6
Specications........................................................ 7
About Oregon Scientic....................................... 7
EU - Declaration Of Conformity............................ 7
FCC Statement....................................................... 7
Declaration Of Conformity.................................... 8
EB612_IM_EU_EN_R3 2006.3.30, 6:16 PM1
2
EN
INTRODUCTION
Thank you for selecting the Oregon ScienticTM Portable UV Monitor.
NOTE Please keep this manual handy as you use your new product. It contains practical step-by-step instructions, as well as technical specications and warnings you should know.
OVERVIEW
FRONT VIEW
1. Sensor
2. UV: measure UV level
3. MODE / SET: toggle display modes, enter settings mode
4. Strap loophole
5. START / STOP / UP: start, stop and pause timers; adjust settings
6. CLEAR / DOWN: stop timers; adjust settings
BACK VIEW
1. RESET
2. Battery compartment
EB612_IM_EU_EN_R3 2006.3.30, 6:16 PM2
3
EN
LCD DISPLAY
1. UV Index icon
2. UV measurement icon
3. Exposure alarm
4. Low battery icon
5. Exposure timer icon
6. Clock / timer
7. Temperature display
8. Key lock icon
9. Skin type / Sun Protection Factor icons
GETTING STARTED
STRAP
Wear on the wrist or attach using the strap as shown.
BATTERIES
The unit uses a CR2032 battery.
Insert the battery before rst use, matching the polarity (+ and -):
appears when the battery is low.
The unit powers on automatically after the battery is inserted.
  
EB612_IM_EU_EN_R3 2006.3.30, 6:16 PM3
4
EN
SPF setting options:
The SPF range is between 1-70. Enter an SPF setting that corresponds to the SPF protection of the sunscreen product you are using.
DISPLAY
To toggle display modes: Press MODE / SET to toggle display mode between:
Clock Mode: time, UV level and temperature
Timer Mode: timer, UV level and temperature
UV mode: UV index, UV level and UV exposure time / timer.
UVI DISPLAY
UV INDEX LEVEL UVI DISPLAY
0-2 Low
3-5 Medium
6-7 High
8-10 Very high
11-25 Extremely high
25+ Over range
SKIN TYPE
1
2
3
4
TANS / BURNS
Never tans; always burns
Sometimes tans; usually burns
Usually tans; sometimes burns
Always tans; rarely burns
SKIN COLOUR IN UNEXPOSED AREA
Pale or milky white; alabaster
Very light brown, sometimes freckles
Light tan, brown or olive; distinctly pigmented
Brown, dark brown or black
EYE COLOUR
Blue
Blue /Green
Gray/ Brown
Brown
and “HH” on LCD display
ENTER SETTINGS
Upon rst use, battery change, or reset the monitor automatically enters setting mode.
To enter settings:
1. Press UP and DOWN to select the desired setting, press and hold to quickly scroll through the setting values.
2. Press MODE / SET to conrm.
3. The settings order is: 12/24 hr format, hour, minute,
oC/o
F, skin type, SPF.
Use the table below to help you determine your skin type. First use the TANS / BURNS columns. If unsure, use the colour of your skin in an area that does not get exposed to sunlight to help guide you. Alternatively, eye colour can also be used as a guideline.
Skin type setting options:
EB612_IM_EU_EN_R3 2006.3.30, 6:16 PM4
5
EN
To measure UVI:
1. Press UV. The sensor will measure continuously for 4 seconds. Rotate the unit by 360
o
while the sensor is
reading to provide a more accurate measurement.
2. Unless there is an exposure timer activated skin type and SPF will be displayed consecutively during this time.
3. UV index, UV level and exposure timers will be automatically be updated.
NOTE
If no timer is set the display will clear after 30mins.
UV EXPOSURE TIMER
After taking a UV measurement in UV mode the monitor calculates a recommended maximum exposure time based on the skin type and SPF set by the user.
NOTE
Please set your skin type before using the exposure
timer. If you change th e skin type or SPF, the exposure time will be adjusted automatically.
The monitor automatically measures the UV level every 10mins when a timer is set. If the UV index has increased it will then adjust the timer to take account of the new levels.
IMPORTAN T The calc ul at ed expo su re time is a suggestion only.
UV EXPOSURE COUNTDOWN TIMER
To start a UV countdown:
1. After UV exposure time has been calculated press START / STOP to start the timer.
2. To pause the timer press START / STOP.
3. Press START / STOP again to resume or press CLEAR to cancel the timer.
4. When the timer has elapsed an alarm will sound.
To silence the alarm:
Press any button.
NOTE
Press UV during a countdown to manually force a UV measurement. The timer will automatically adjust the exposure time.
After pausing and restarting the timer, press UV to ensure the timer remains accurate.
MANUAL COUNTDOWN TIMER
To set countdown timer:
1. In timer mode, press and hold MODE / SET.
2. Use UP and DOWN to select hour and minutes.
3. Press MODE / SET to conrm.
To start manual countdown timer:
1. Press MODE / SET to select timer mode.
2. Press START / STOP To start the timer
3. To pause the timer press START / STOP.
4. Press START / STOP again to resume or press CLEAR to cancel the timer.
5. When the timer has elapsed an alarm will sound.
To silence the alarm:
Press any button.
EB612_IM_EU_EN_R3 2006.3.30, 6:16 PM5
6
EN
SETTINGS
To change clock settings:
1. In clock mode, press and hold MODE / SET.
2. Use UP and DOWN to select the desired setting.
3. Press MODE / SET to conrm. The settings order is: 12/24 hr format, hour, minute,
oC/o
F.
To change skin type / SPF settings
1. In UV mode press and hold MODE / SET.
2. Use UP and DOWN to select the desired setting.
3. Press MODE / SET to conrm. The settings order is: skin type / SPF.
KEY LOCK
To activate / de-activate key lock: Press and hold START / STOP / UP, then press CLEAR / DOWN. indicates key lock is activated.
RESET
In case of abnormal operation, use a thin blunt object to press the RESET button on the back of the unit.
PRECAUTIONS
This produ ct is engi neere d to give you years of satisfactory service if you handle it carefully. Here are a few precautions:
Do not subject the unit to excessive force, shock, dust, temperature or humidity, which may result in malfunction, shorter electronic life span, damaged battery and distorted parts.
Placement of this product on wood surfaces with certain types of nishes, such as clear varnish, may result in damage to the nish. Consult the furniture manufacturer’s care instructions for direction as to the types of objects that may safely be placed on the wood surface. Oregon Scientic shall not be responsible for any damage to wood surfaces from contact with this product.
Do not immerse the unit in water. If you spill liquid over it, dry it immediately with a soft, lint-free cloth.
Do not clean the unit with abrasive or corrosive materials.
Do not tamper with the unit’s internal components. Doing so will invalidate the warranty on the unit and may cause unnecessary damage. The unit contains no user-serviceable parts.
Only use fresh batteries as specied in the user’s instructions. Do not mix new and old batteries.
Do not use rechargeable batteries with this prod­uct.
Remove the batteries if storing this product for a long period of time.
Due to printing limitations, the displays shown in this manual may differ from the actual display.
The contents of this manual may not be reproduced without the permission of the manufacturer.
Do not dispose this product as unsorted municipal waste. Collection of such waste separately for special treatment is necessary.
NOTE The technical specications for this product and the contents of the user manual are subject to change without notice
EB612_IM_EU_EN_R3 2006.3.30, 6:16 PM6
7
EN
SPECIFICATIONS
TYPE DESCRIPTION
Unit dimension 80 x 43 x 20mm (H x W x D) (3.2 x 1.7 x 0.9 inches) Unit weight 78g (2.75 ounces)
without battery UVI range 1-25 Clock time 12 or 24-hour format Temperature unit °C / °F Alert sound 2 minutes Timer 1 min up to 39 hrs 59 mins Operating 0°C to 60°C
temperature (32°F to 140°F) Water resistance IPX-2 / JIS2 Power 1 x CR2032 battery
ABOUT OREGON SCIENTIFIC
Visit our website (www.oregonscientic.com) to learn more about Oregon Scientic products such as digital cameras; MP3 players; children’s electronic learning products and games; projection clocks; health and tness gear; weather stations; and digital and conference phones. The website also includes contact information for our Customer Care department in case you need to reach us, as well as frequently asked questions and customer downloads. W
e hope you will nd all the information you need on our website, however if you’re in the US and would like to contact the Oregon Scientic Customer Care department directly, please visit:
www2.oregonscientic.com/service/default.asp OR Call 1-800-853-8883.
For
international inquiries, please visit: www2.oregonsc
ientic.com/about/international.asp
EU-DECLARATION OF CONFORMITY
Hereby, Oregon Scientic, declares that this Portable UV Monitor Model: EB612 / UV888 is in compliance with EMC directive 89/336/CE. A copy of the signed and dated Declaration of Conformity is available on request via our Oregon Scientic Customer Service.
FCC STATEMENT
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, inclu ding int erfer ence that may caus e und esi red operation.
W
ARNING Changes or modifications not expressly
approved by the party responsible for compliance could void the user’s authority to operate the equipment.
EB612_IM_EU_EN_R3 2006.3.30, 6:16 PM7
8
EN
We
Name: Oregon Scientic, Inc. Address: 19861 SW 95
th
Ave.,Tualatin, Oregon 97062 USA Telephone No.: 1-800-853-8883
declare that the product
Product No.: EB612 Product Name: Portable UV Monitor Manufacturer: IDT Technology Limited Address: Block C, 9/F, Kaiser Estate, Phase 1,41 Man Yue St., Hung Hom, Kowloon, Hong Kong
is in conformity with Part 15 of the FCC Rules. Operation is subject to the following two conditions: 1) This device may not cause harmful interference. 2) This device must accept any interference received, including interference that may cause undesired operation.
NOTE This equipme nt has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency ene rgy and, if not ins talled and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and
receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio / TV technician for help.
DECLARATION OF CONFORMITY
The following information is not to be used as contact for support or sales. Please call our customer service number (listed on our website at www.oregonscienti c.com), or on the warranty card for this product) for all inquiries instead.