www.insize.com
MN-ISU-300D-E
ISU-300D
ULTRASONIC THICKNESS GAUGE
OPERATION MANUAL

Attention
In order to obtain good measurement precision, you need to clear away
the rusty, sundry, grease, etc. on the working surface.
Please use the couplant on the working surface, measure repeatedly
around the target area and take average value.
Please clean the couplant on the transducer and working surface after
measure.
Please don’t pull the transducer cable when use the instrument.
Specification Of Transducers
Code
ISU-T07
ISU-T06
ISU-T12
ISU-T13
Frequency
5.0MHz 10.8mm
7.5MHz
2.0MHz
5.0MHz
Diameter( )Ød
8.5mm
16.3mm
13mm
Measuring range
T-E: 0.8-300mm
E-E: 4-25mm
0.7-50mm
3-350mm
3-100mm
Code
ISU-T07
ISU-T06
ISU-T12
ISU-T13
Minimum size of
measurement
(diameter × wall thickness)
T-E: 20×1.2mm
E-E: 20×4mm
pipe for
15×1.2mm
30×4mm
25×3mm
Applicable
temperature
<60°C
<60°C
<60°C
<350°C
Application
general use
for thin material
for casting steel
for high
temperature
1 10

Glycerin
Gold
Ice
Inconel
Iron
Iron (cast)
Lead
Magnesium
Mercury
Molybdenum
Monel
Neoprene
Nickel
Nylon, 6.6
Oil (SAE 30)
Platinum
Plexiglass
Polythylene
Polystyrene
Polyurethane
Quartz
Rubber, Buty
Silver
Steel, Mild
Steel, Stainless
Teflon
Tin
Titanium
Tungsten
Uranium
Water
Zinc
19000.075
0.130
0.160
0.220
0.230
0.180
0.085
0.230
0.057
0.250
0.210
0.063
0.220
0.100
0.067
0.130
0.110
0.070
0.093
0.0700
0.230
0.070
0.140
0.233
0.228
0.060
0.130
0.240
0.200
0.130
0.584
0.170 4200
3200
4000
5700
5900
4600
2200
5800
1400
6300
5400
1600
5600
2600
1700
3300
1700
1900
2400
1900
5800
1800
3600
5920
5800
1400
3300
6100
5200
3400
1480
Technical Specification
Measurement range
Resolution
Velocity range
Measurement rate
Average mode
Limited setting
Measuring Units
Memory
Data output
Display
Battery
Operating temperature
Measuring temperature
Dimensions
Weight
T-E: 0.8mm~300mm
E-E: 4mm~25mm
0.01mm(0.001〞), 0.1mm (0.01〞)
1000m/s~9999m/s
4 /s and 10/s in fast mode
2 to 9 times average measurement
With Low-high indication and alarm
mm / inch
Memory of 5000 readings with location number
USB to PC
128×64 LCD with back light
2 x AAA Batteries
-20℃~+50℃
-20℃~+350℃(according to the probes)
116mm
0.22kg
Standard delivery
Mainunit
Transducer ISU-T07
Battery(AAA)
Couplant (for ISU-T06, ISU-T07, ISU-T12)
USB cable and software disc
Optional Accessory
Transducer
Couplant (for ISU-T13)
ISU-T06, ISU-T12, ISU-T13
ISU-HT5-COUPLANT
L)×64mm(W)×27mm H( ( )
1pc
1pc
2pcs
1bottle
1pc of each
9 2

Codel
ISU-T07
ISU-T06
ISU-T12
ISU-T13
Measuring Range Temperature
T-E: 0.8-300mm/0.03-12.00"
E-E: 4.0-25mm/0.16-1"
0.7-50mm/0.03-2.00"
3.0-350mm/0.12-13.00"
3.0-100mm/0.12-4.00"
Overview the display unit
1
2
1. LCD Screen
5. 4.00mm Block
2. Key Pad 3. Battery Pack
Key Functions
4
Frequency
5.0MHZ
7.5MHZ
2.0MHZ
5.0MHZ
5
<60℃/140°F
<60℃/140°F
<60℃/140°F
<350℃/662°F
3
4. Transducer Port
Press key into the menu, the screen will display:
1. Power off 2. Gain adjustment
5. Default4. Contrast
3. Language
6. Information
Press or key into the selected item, press key to confirm.
1. Power off
Auto shut down after 1 Min. 3 Min. 5 Min. or Never can be
selectable.
2. Gain adjustment
Press key into the “Gain adjustment”, the screen will display:
A. High B. Medium
C. Low
D. Automatic
Press or key to select desired item.
Press to confirm.
3. Language
To change the language: English, Portuguese, Czech.
4. Contrast
To change the contrast of the screen.
5. Default
When the “ Default ”is selected , the gauge will recover the default
parameter.
6. Information
The screen displays the version number and transducer number.
Sound velocity
Material
Air
Aluminum
Alumina Oxide
Beryllium
Boron Carbide
Brass
Cadmium
Copper
Glass(crown)
Inch/µS
0.013
0.250
0.390
0.510 12900
0.430
0.170
0.110
0.180
0.210
Sound Velocity
M/s
330
6300
9900
11000
4300
2800
4700
5300
3
8

Press key to enter/confirm.
3. Calibration
Press key into menu.
Press or key into “Calibration”.
Measure the test piece with known thickness.
If measured value is different than the known value of the test piece,
adjust the measured value by pressing or key and then press
key. The gauge will return to measure mode.
6
Memory
Press key into the menu.
Press or key into “ Memory”, the screen will display:
1. Memory unit 2. Memory read
3. Delete all memory 4. Date transfer
Press or key into the selected item, press key to confirm.
1. Memory unit
The gauge has a memory capacity of 5000 measurements. The
memory location was composed by alphabet A-Z + 0000-4900.
You can select an Alphabet + an initial number freely for beginning
to store the value and the next number will be followed automatically.
Press key into the menu.
Press or key to select Alphabet A-Z which you want.
Press key to confirm.
After taking every measurement, press key to store the value with
a location number.
2. Memory read
Press key into the menu - “Memory” - “Memory Read” (Under the
measurement).
Press or key to select desired Alphabet. Then the desired group
of value can be readable by pressing or key.
3. Delete memory
Delete all memory.
4. Date transfer
Press key into the menu.
Press or key into “Date Transfer”, the screen will display:
The data can be transferred to PC using the data view and can be
stored as DOC, TXT, Or Excel.
For more detailed information, refer to the “Installation Manual”
enclosed in the CD.
7
Function
Display screen
⑦
⑧
Key
①
Function
--On/Off Key
--Esc. Menu
--Menu Key
--Confirm Key
--Up Arrow Key
--Backlight Key
--Down Arrow Key
--Calibration Key
--Left Arrow Key
(Under the measurement)
--Right Arrow Key
(Under the measurement)
--Read data
--Storage Key
②
③
④
⑤
⑥
7 4

Measurement value
①
②
③
④
⑤
⑥
⑦
⑧
Battery life
Measurement model
Resolution
Current transducer model
Current memory location
Current velocity
Gain setting
Basic gauge operations
Switch on
1
Press key to turn on the gauge.
Probe zero
2
The gauge does an automatic zeroing of the transducer thus eliminating
the need for an on-block zero. And then the gauge came into the
measurement mode directly.
Please make sure the transducer is not coupled to the test piece
when the gauge is first turned on and that there is no coolant on the
end of the transducer. The transducer should also be at the room
temperature, clean without any noticeable wear.
3
Backlight
Press key to turn on / off the backlight.
Measurement setting
4
Press key into the measurment mode setting.
Press or key to select desired measurment mode.
Press key to confirm selection.
Press key to esc menu and into the measurement.
ISU-250C Ultrasonic Thickness Gauge offers two measurment
modes, they are T-E mode and E-E mode.
1. T-E Mode
Standard - It is available for the normal measurement.
Minimum - The gauge will catch and display the minimum measured
thickness during one measurement process. It is available for measuring
the thickness of curve surface or pipe.
Difference - The gauge will display a thickness value as an absolute
number of what has been inputted. For example, input value = 5.00mm
and the real thickness is 5.03mm, the display will show diff 0.03mm. If
the real value is 4.97mm,the gauge will display -0.03mm.
Average - The gauge will display the average thickness of 2-9
measurements.
Limitation - The gauge will alarm you of low or high thresholds via
audible sound.
Scan - The gauge will alarm for each fast measurement. And will
display the all measured thickness upon the complete measurement
finished. It is available for measuring the thickness of test piece with
high temperature surface.
2. E-E Mode
When the E-E mode is selected, only probe ISU-T07 is available. This
function allows you to make measurement between two consecutive
back wall echoes. Therefore, a good usage of the E-E option is for
measuring through coatings (Max. 2mm) to measure only the true
metal thickness.
5
Setting
1. Velocity rate
Velocity setting
Press key into“ VEL. SETTING” state.
There are 9 velocities for materials pre-stored in gauge. You can
select one by pressing or key.
Press key to confirm.
Velocity input
When the soun d v elo cit y of 9 materials is not sa tis fied with the
requirements of the users, there is a sound velocity table which give
the sound velocity of various materials in the appendix.
Press key into “Velocity input ” state.
Press or key to select the place where you want to store
and put the new velocity.
Press key to confirm.
Velocity measurement
Measuring the sample which thickness is known.
Press key into“ Velocity measurement ”state.
Press or key to up and down the value of velocity to determine
the thickness as the same as the value of sample that is measured.
Press key to confirm.
Velocity store
After you set the velocity, the instrument can store the velocity value.
2. Resolution
Press key into“ Resolution ”.
Press or to select resolution and unit.
A. 0.1mm B. 0.01mm
C. 0.01in
D. 0.001in
5 6