Inovytec Ventway Sparrow, VWSP-100, VWSP-900 User Manual

VWSP-100 Civil Model
VWSP-900 Military Model
User Manual
Ventway Sparrow
Emergency and transport
ventilator
Document Number VWSP-500-03 October 2018
Ventway Sparrow User Manual
Page 2 of 110
Important
This User Manual is subject to periodic review, update and revision.
Do not use a defective product. Do not repair this product or any of its parts. If this device does not perform properly, contact an Inovytec representative.
The user of this product has sole responsibility for any malfunction that results from improper use, faulty maintenance, improper repair, unauthorized service, damage, or alteration by anyone other than Inovytec Medical Solutions Ltd.
The safety, reliability, and performance of this device can be assured only under the following conditions:
The device has been used according to the accompanying operating instructions.
All fittings, extensions, readjustments, changes, or repairs have been carried out by Inovytec Medical Solutions Ltd.'s authorized represen­tatives.
No part of this publication may be reproduced, stored in a retrieval system or transmitted in any form by any means, electronic, mechanical, photo reproductive, recording or otherwise without the express prior written permission of Inovytec Medical Solutions Ltd.
Inovytec Medical Solutions Ltd. reserves the right to change or improve its products and accompanying technical literature without specific notice of changes or improvements.
This product is protected by patents listed on the Inovytec website.
Ventway Sparrow User Manual
Page 3 of 110
Contact Information:
Inovytec Medical Solutions Ltd. 3 Hanagar St., POB 7282, Hod-Hasharon 4501306, Israel Tel: +972 9 779 41 35 Fax: +972 9 779 41 38
E-mail: Info@Inovytec.com Web Site: http://www.Inovytec.com
Disclaimer
Information provided by Inovytec Medical Solutions Ltd. is believed to be accurate and reliable. However, Inovytec assumes no responsibility for the use of such information, nor for any infringements of patents or other rights of third parties, that may result from its use.
FDA Tracking Requirements
U.S. Federal Law (21 CFR 821) requires the tracking of ventilators. Under this law, owners of this ventilator must notify Inovytec Medical Solutions Ltd. if this product is received; lost, stolen, or destroyed; donated or resold; or otherwise distributed to a different organization. If any such event occurs, contact Inovytec in writing with the following information:
Originator's organization – Company name, address, contact name, and
contact phone number
Model number, and serial number of the ventilator Disposition of the ventilator (for example, received, lost, stolen,
destroyed, distributed to another organization), new location and/or organization (if known and different from originator’s organization) – company name, address, contact name, and contact phone number
Date when the change took effect
Please address this information to Inovytec Medical Solutions Ltd. at the address given above.
PLEASE READ THIS USER MANUAL BEFORE OPERATING THE SYSTEM.
Ventway Sparrow User Manual
Page 4 of 110
Contents
1. About This User Manual .......................................................................................................... 9
1.1. TYPES OF WARNINGS, CAUTIONS AND NOTES ....................................................................... 9
1.2. GLOSSARY AND ABBREVIATIONS ................................................................................................... 10
2. Overview of System .............................................................................................................. 11
2.1. DESCRIPTION OF DEVICE ............................................................................................................. 11
3. Conditions for Use ................................................................................................................ 12
3.1. INTENDED USE ......................................................................................................................... 12
3.2. INDICATIONS FOR USE ................................................................................................................ 12
3.3. CONTRAINDICATIONS ................................................................................................................. 13
3.4. LIMITATIONS OF USE ................................................................................................................. 13
4. Safety ................................................................................................................................... 14
4.1. ELECTRICAL SAFETY ................................................................................................................... 14
4.2. EMC COMPLIANCE ................................................................................................................... 14
4.3. SAFETY INSTRUCTIONS ............................................................................................................... 14
5. System Components ............................................................................................................. 18
5.1. UNPACKING THE DEVICE ............................................................................................................. 18
5.2. VENTILATOR FRONT PANEL ....................................................................................................... 19
5.3. VENTILATOR REAR PANEL......................................................................................................... 19
5.4. VENTILATOR USE CONFIGURATION ............................................................................................. 20
5.5. PATIENT CIRCUIT ....................................................................................................................... 21
5.6. PNEUMATIC SECTION THEORY OF OPERATION ................................................................................ 22
6. Connecting the Ventilator ..................................................................................................... 25
6.1. FRONT PANEL CONNECTIONS ....................................................................................................... 25
6.2. REAR PANEL CONNECTIONS......................................................................................................... 25
7. Power On and Display Startup Screen ................................................................................... 26
7.1. QUICK START ........................................................................................................................... 26
8. Navigating the GUI screens ................................................................................................... 28
8.1. SELECTING SCREEN OPTIONS ....................................................................................................... 28
8.2. EDITING FIELDS ........................................................................................................................ 28
8.3. NAVIGATING BETWEEN SCREENS .................................................................................................. 28
8.4. CONFIRMING OR CANCELLING A MESSAGE ...................................................................................... 29
8.5. SYMBOLS USED IN THE SYSTEM ..................................................................................................... 29
8.6. SYSTEM INDICATORS .................................................................................................................. 30
8.6.1. Battery Status .............................................................................................................. 30
8.6.2. indicators ..................................................................................................................... 31
8.6.3. Alert ............................................................................................................................. 31
9. Getting Started with Ventway Sparrow ................................................................................. 32
9.1. DISCONNECT PATIENT ................................................................................................................ 32
9.2. PATIENT WEIGHT ...................................................................................................................... 33
9.3. VENTILATION MODE .................................................................................................................. 34
9.4. VENTILATION PARAMETERS ......................................................................................................... 35
9.5. NUMERICAL REPRESENTATION OF BREATH PARAMETERS .................................................................... 36
9.6. PRESSURE AND FLOW GRAPHS ...................................................................................................... 37
10. Main Menu ......................................................................................................................... 38
Ventway Sparrow User Manual
Page 5 of 110
10.1. MAIN MENU/NEW PATIENT (DISCONNECT PATIENT SCREEN)........................................................... 39
10.2. MAIN MENU/VENT PARAMS .................................................................................................... 39
10.2.1. MAIN MENU/Vent Params/Trigger Sensitivity .......................................................... 40
10.2.2. MAIN MENU/Vent Params/Vent Mode ..................................................................... 40
10.2.3. MAIN MENU/Vent Params/O2 Enrichment ................................................................ 40
10.2.4. MAIN MENU/Vent Params/Patient Weight ............................................................... 40
10.3. MAIN MENU/ALERT SETTINGS ................................................................................................ 41
10.4. MAIN MENU/ADVANCED SETTINGS ............................................................................................ 42
10.4.1. MAIN MENU/Adv Settings/Load Default ................................................................... 42
10.4.2. MAIN MENU/ADV SETTINGS/Brightness ................................................................... 42
10.4.3. MAIN MENU/Adv Settings/Alert volume ................................................................... 43
10.4.4. MAIN MENU/Adv Settings/Language ........................................................................ 43
10.4.5. MAIN MENU/Adv Settings/Tech mode ...................................................................... 43
10.4.5.1. MAIN MENU/ADV SETTINGS/TECH MODE/Set Time ......................................... 44
10.4.5.2. MAIN MENU/ADV SETTINGS/TECH MODE/Calibration ..................................... 44
10.4.5.3. MAIN MENU/ADV SETTINGS/TECH MODE/Work Hours ................................... 45
10.4.5.4. MAIN MENU/ADV SETTINGS/TECH MODE/Self Test ......................................... 46
10.4.5.4.1. MAIN MENU/ADV SETTINGS/TECH MODE/Self Test /Flow and Pressure .. 47
10.4.5.4.2. MAIN MENU/ADV SETTINGS/TECH MODE/Self Test /CVT ........................ 48
10.4.5.4.3. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /VVT ...................... 49
10.4.5.4.4. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /Blower Test ......... 50
10.4.5.5. MAIN MENU/ADV SETTINGS/TECH MODE/Logbook ......................................... 51
10.4.5.6. MAIN MENU/ADV SETTINGS/TECH MODE/Export Log ...................................... 51
10.4.5.7. MAIN MENU/ADV SETTINGS/TECH MODE/SW Version .................................... 51
10.4.5.8. MAIN MENU/ADV SETTINGS/TECH MODE/SW Update..................................... 52
10.4.5.9. MAIN MENU/ADV SETTINGS/TECH MODE/System Mode ................................. 52
10.4.5.10. MAIN MENU/ADV SETTINGS/TECH MODE/Altitude ........................................ 52
11. Warnings and Alerts ............................................................................................................ 53
11.1. SUMMARY OF ALERT TYPES ....................................................................................................... 53
11.2. SUMMARY OF ALERT LEVELS ...................................................................................................... 54
11.3. WARNING EXAMPLE ................................................................................................................ 55
11.4. ALERT EXAMPLE ..................................................................................................................... 55
12. Default Parameters ............................................................................................................. 56
12.1. START VOLUME VENTILATION .................................................................................................... 56
12.1.1. Oxygen supply ............................................................................................................ 56
12.1.1.1. Connecting the Air Inlet to the Demand Valve .................................................. 57
12.1.2. Recommended devices for monitoring of oxygen ...................................................... 58
12.2. ALERT DEFAULT PARAMETERS: 5 KG TO 70+ KG ........................................................................... 59
12.3. ALERT DEFAULT PARAMETERS: ALL PATIENT WEIGHTS ..................................................................... 59
13. Labels and Symbols ............................................................................................................. 60
13.1. LABELS ................................................................................................................................. 60
13.2. SYMBOLS .............................................................................................................................. 61
14. Cleaning and Disinfecting .................................................................................................... 62
15. Ventilation Methods ........................................................................................................... 63
15.1. SIMV-VC (PS)FLOW CHART..................................................................................................... 63
Ventway Sparrow User Manual
Page 6 of 110
15.2. CPAP SPONTANEOUS BREATHING MODE ................................................................................... 66
15.3. BACKUP VENTILATION MODE .................................................................................................... 69
15.3.1. Backup Ventilation in CPAP ........................................................................................ 69
15.3.2. Backup Ventilation before starting patient ventilation .............................................. 70
15.4. SIMV VC (CPR) .................................................................................................................... 72
15.5. CPAP (CPR) ......................................................................................................................... 73
16. Service and Maintenance .................................................................................................... 74
16.1. DEVICE CALIBRATION AND SOFTWARE UPGRADES .......................................................................... 74
17. Troubleshooting .................................................................................................................. 75
18. Specifications ...................................................................................................................... 76
18.1. DIMENSIONS AND WEIGHT ....................................................................................................... 76
18.2. ENVIRONMENTAL SPECIFICATIONS .............................................................................................. 77
18.3. HARSH ENVIRONMENTAL CONDITIONS ............................................................................. 77
18.3.1. Operation in extreme high / low temperatures ......................................................... 77
18.3.2. Operation in High or Low Altitude ............................................................................. 78
18.3.3. Airborne particulates ................................................................................................. 78
18.4. POWER SUPPLY ...................................................................................................................... 79
18.5. VENTILATION PERFORMANCE ..................................................................................................... 80
18.6. STANDARDS AND SAFETY REQUIREMENTS ..................................................................................... 81
19. Cleaning and Routine Maintenance ..................................................................................... 82
20. Batteries ............................................................................................................................. 83
20.1. BATTERY MAINTENANCE .......................................................................................................... 83
21. Parts and Accessories .......................................................................................................... 84
22. Regulatory .......................................................................................................................... 85
23. Warranty ............................................................................................................................ 86
24. Appendix – Test Alerts ........................................................................................................ 87
24.1. BACKUP VENTILATION .............................................................................................................. 87
24.2. START BACKUP VENTILATION WITH UNKNOWN WEIGHT .................................................................... 88
24.3. CPAP VENTILATION WITH BACKUP VENTILATION ............................................................................ 89
24.4. PATIENT DISCONNECT .............................................................................................................. 90
24.5. HIGH PEEP ........................................................................................................................... 91
24.6. VALVE BLOCKED ...................................................................................................................... 92
24.7. PRESSURE ALERT ..................................................................................................................... 93
24.8. MINUTE VOLUME (MV) ALERT .................................................................................................. 94
24.9. LEAK ALERT ........................................................................................................................... 95
24.10. TIDAL VOLUME ALERT ............................................................................................................ 96
24.11. I:E ALERT ............................................................................................................................ 97
24.12. APNEA ALERT ....................................................................................................................... 98
24.13. POWER ALERT ...................................................................................................................... 99
24.14. LOW BATTERY ALERT ........................................................................................................... 100
24.15. BATTERY TYPE ALERT ............................................................................................................ 101
24.16. VOLTAGE ALERT .................................................................................................................. 102
24.17. TEMPERATURE ALERT ........................................................................................................... 103
24.18. TUBE DISCONNECT ALERT ..................................................................................................... 104
24.19. SERVICE REQUIRED ALERT ..................................................................................................... 105
24.20. FILTER REPLACEMENT REQUIRED ............................................................................................ 106
Ventway Sparrow User Manual
Page 7 of 110
24.21. ALTITUDE CHANGE ALERT ..................................................................................................... 107
24.22. SHUTDOWN ALERT .............................................................................................................. 108
24.23. VENTILATION DURING STANDBY (DUE TO PATIENT INSPIRATORY EFFORT) .......................................... 109
25. Appendix – Menu Hierarchy .............................................................................................. 110
Ventway Sparrow User Manual
Page 8 of 110
Obtaining Help
If you have a ventilator problem that you cannot solve, and the ventilator was purchased directly from Inovytec, you may contact Inovytec at
sales@Inovytec.com.
If you have a ventilator problem that you cannot solve, and the ventilator was purchased from an authorized Inovytec distributor, please contact your distributor directly to report the problem.
Note: If this ventilator has not been purchased directly from Inovytec, please ensure that it has been purchased from an authorized distributor of Inovytec. To obtain a list of authorized distributors, contact Inovytec at
sales@Inovytec.com.
Ventway Sparrow User Manual
Page 9 of 110
1. ABOUT T HIS USER MANUAL
This User Manual provides the information necessary to operate and maintain the Ventway Sparrow ventilator.
PLEASE READ THIS USER MANUAL BEFORE OPERATING THE SYSTEM. If any part of this User Manual is not clear, contact Customer Support for assistance.
1.1. TYPES OF WARNINGS, CAUTIONS AND NOTES
Three types of special message appear in this User Manual:
Warning: A warning indicates precautions to avoid the possibility of personal injury or death.
Caution: A caution indicates a condition that may lead to damage to equipment, or a lower quality of treat­ment.
Note: A note provides other important information.
PLEASE RETAIN THIS USER MANUAL FOR FUTURE REFERENCE.
Ventway Sparrow User Manual
Page 10 of 110
1.2. GLOSSARY AND ABBREVIATIONS
Apnea Temporary cessation of breathing
BPM Breaths Per Minute
CPAP Continuous Positive Airway Pressure
lpm Liters per minute
Mandatory Breath
Ventilator initiated breath
MV Minute Volume
NIV Non-Invasive Ventilation
Peak Flow Maximum volumetric flow
PEEP Peak End Expiratory Pressure
PIP Peak Inspiratory Pressure
Pressure support
Preset pressure delivered to the patient, on top of the PEEP, during triggered breath
PSV Pressure Support Ventilation
SIMV VC PS
Synchronized Intermittent Mechanical Ventilation with Volume Control and Pressure Support
Te Expiratory Time
Ti Inspiratory Time
Tidal Volume
Normal volume of air displaced between normal inhalation and exhalation when extra effort is not applied
Triggered breath
Patient initiated breath
TVe Expired Tidal volume
TV
i
Inspired Tidal volume
Ventway Sparrow User Manual
Page 11 of 110
2. OVER VIE W OF SYSTEM
2.1. DESCRIPTION OF DEVICE
The Ventway is an emergency portable ventilator, used for transport, EMS and military applications.
The situations for which it is intended are characterized by attendance of first responders with limited triage capabilities, requiring a simple yet highly effective ventilator, that is self-sufficient and lightweight.
The ventilator is suitable for noninvasive ventilation for a full non-vented ventilation face mask or invasive ventilation via an endotracheal tube, tra­cheostomy and laryngeal mask airway (LMA).
Note: The power supply needs to be firmly and permanently secured in any EMS environment in which the ventilator is used.
Ventway Sparrow User Manual
Page 12 of 110
3. CONDITIONS FOR USE
3.1. INTENDED USE
The Ventway Sparrow lung ventilator is intended for emergency use and transportation of adult and pediatric patients weighing at least 5 kg (11 lb). It may be used for invasive or noninvasive ventilation presets. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician.
3.2. INDICATIONS FOR USE
The Ventway Sparrow ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg, who require the following types of ventilatory support: SIMV - VC (PS), CPAP.
Ventway Sparrow User Manual
Page 13 of 110
3.3. CONTRAINDICATIONS
See specific patient instructions regarding performing ventilation or any
use of life support equipment.
Acute Pneumothorax
3.4. LIMITATIONS OF USE
Clinical situations potentially affecting accuracy or performance:
Controlling the flow in the presence of difficult airways, such as severe
lung blockage and asymmetric air entrance to the lung
Low compliance of the airways Asynchronization between patient and ventilator Barotrauma Behavior of the ventilator in case of barotrauma, monitoring and alerting
in these cases.
Note: The use of humidification is not recommended, due to potential blockage of control and measurement tubes.
Ventway Sparrow User Manual
Page 14 of 110
4. SAFETY
4.1. ELECTRICAL SAFETY
The device complies with requirements of IEC/EN 60601-1 for general requirements for safety of medical electrical equipment:
Class I Equipment BF type applied part Mode of operation: Continuous measurement Degree of mobility: Portable
4.2. EMC COMPLIANCE
The unit has Class B compliance.
4.3. SAFETY INSTRUCT IONS
Warnings
Basic safety precautions always should be taken, including all those listed below. DO NOT USE BEFORE READING THIS USER MANUAL.
DO NOT use this device for any purpose other than specified in this manual without written consent and approval from Inovytec Medical Solutions Ltd. In case of VENTILATOR failure, the lack of immediate access to appropriate alternative means of ventilation can result in PATIENT death. US Federal Law restricts the sale of this instrument only by, or on the order of, a physician. The exhaled volume of the patient can differ from the measured exhaled volume due to leaks around the mask. The device shall not be used in a hyperbaric chamber.
The device shall not be used with nitric oxide and explosive or highly flammable gas mixtures. The device shall not be used with helium or mixtures with helium.
The device should not be used on unattended patients.
The device accuracy can be affected by the gas added by use of a nebulizer.
Ventway Sparrow User Manual
Page 15 of 110
Cautions
If the device packaging is not intact, do not use the device.
If the device does not turn on, or is not working correctly, discontinue use. Refer servicing or replacement to qualified service personnel. Do not disassemble any part of the system components. This system is not user-serviceable. Do not use the equipment if it is not working properly or if it has suffered any damage, for example, by dropping the equipment or splashing water on it. If the LCD screen is cracked or damaged, check whether the screen can be used, and if not, do not use the device. If the power button is damaged or stuck, disconnect the patient from the device and remove the battery. If the rotator switch does not allow changing parameters, the device cannot be used. The Patient Circuit is single use only. If it is not removed from a new container, it may have already been used and should not be used. The Patient Circuit can be used for the same patient up to five days.
Do not use a Patient Circuit that is not the original Patient Circuit of the device. Confirm that the expiration date, found on the Patient Circuit packaging bag, has not been reached. The device should be used under medical supervision.
When using external oxygen enrichment, please note the following:
When using demand valve, the oxygen enrichment level will
reach a minimum of 95%.
When using reservoir bag, the oxygen enrichment level may
vary depending on oxygen flow rate. It is the user’s responsibility to retain information about the patient (by USB connection). Storage capacity may be sufficient for at least one year of ventilation.
Ventway Sparrow User Manual
Page 16 of 110
Cautions
Repairs should be undertaken only by personnel trained or auth­orized by Inovytec Medical Solutions Ltd. Do not modify this equip­ment without authorization from Inovytec Medical Solutions Ltd. The device may not operate correctly if used or stored outside the relevant temperature or humidity ranges, as described in the per­formance specifications. Strictly follow the warning instructions in this manual.
This instrument is fragile. To prevent damage, please handle with care, including while packing and unpacking. Ensure that the system is only used by a trained person familiar with all system operating procedures. EMS personnel should complete a training program before operating the Ventway Sparrow. User is prohibited from changing, adding, removing or disassembling any system parts. Warranty shall not apply to any defects, failure or damage caused by improper use and/or improper or inadequate maintenance and care. The unit is classified as Class IIb, continuously operated, ordinary equipment with applied part and with signal input/output parts. The device is not intended for use in the presence of flammable sub­stances. To avoid damage to the screen, do not expose the instrument to direct sunlight for prolonged periods. The system is approved for IP45 in operation mode with oxygen enrichment. To prevent damage to the instrument or patient cable, avoid liquid spillage while cleaning. It is strongly recommended that all Ventway Sparrow parts be replaced with parts purchased from Inovytec Medical Solutions Ltd. or an authorized distributor. Use of other parts may damage the unit and void the warranty. The ventilator is suitable for noninvasive ventilation for full non­vented ventilation face mask or invasive ventilation via an endotracheal tube, tracheostomy and laryngeal mask airway (LMA).
Ventway Sparrow User Manual
Page 17 of 110
Cautions
During NIV (Non-Invasive Ventilation) the user should use a capnograph in order to monitor the CO2 level of the patient.
Covering the ventilator is prohibited. Ensure that no Latex or natural rubber parts are in patient pathways.
When adding medication to the gas flowing into the patient by using an MDI or nebulizer, please position between mask/ETT and exha­lation valve.
Do not obstruct the gas intake ports.
Discarded used or unused patient circuit is classified as clinical waste. As such, the user is responsible for complying with all local and national regulations regarding discarding of clinical waste.
Notes
Dispose of this device and used sensors in accordance with local regulations. Use the equipment only for the purpose described in these instructions for use. The contents of this manual are subject to change without prior notice. The user or any technical personnel who are not formally authorized by Inovytec Medical Solutions Ltd. should not open the device under any circumstances. Opening the device could damage the unit and will void the warranty provided by Inovytec Medical Solutions Ltd.
Ventway Sparrow User Manual
Page 18 of 110
5. SYST EM COMPONENTS
5.1. UNPACKING THE DEVICE
Package contents
1 Ventway Sparrow ventilator
2 User Manual and device documentation
3 Battery pack - Inside ventilator battery
compartment
4 Power supply
Ventway Sparrow User Manual
Page 19 of 110
5.2. VENTILATOR – FRONT PANEL
5.3. VENTILATOR – REAR PANEL
Warning: Do not block item (1), the air/oxygen inlet.
Rear Pane:
(1) Air/Lo
w pressure oxygen inlet,
(2) P
ower supply connector,
(3) USB
connector, (4) Power On/Off button, (5) Battery pack lock, (6) Battery pack, (7) Filter compartment screws
Front Panel:
(1) control knob,
(2) display,
(3) control and sensing tubes port,
(4) patient port
Ventway Sparrow User Manual
Page 20 of 110
5.4. VENTILATOR – USE CONFIGURATION
During transport, the ventilator is recommended to be placed in a horizontal position.
Ventway Sparrow User Manual
Page 21 of 110
5.5. PATIENT CIRCUIT
Patient circuits specifications Adult Pediatric
Nominal overall length 180 / 240 cm 180 / 240 cm
Internal volume 690 / 920 CC 318 /424 CC
Inspiratory Resistance to flow 2.7 cmH2O @ 60
lpm
5.4 cmH2O @ 30 lpm
Dead space 17 cc 16.5 cc
Circuit compliance 0.001 L/cmH2O 0.008 L/cmH2O
Corrugated tube diameter 22 mm 15 mm
Delivered tidal volumes Tv > 200 cc 50 ml < Tv < 200 cc
Exhalation valve color Transparent Blue
An authorized HME and anti-bacterial filter must be used at all times, in order to protect both patient and ventilator from infection. Assemble the filter between the patient and transducer; if not possible, then between the ventilator outlet and the patient circuit 22 mm connector.
Warning: All parts of the patient circuit are single-use. Discard after use. Handle carefully to avoid cross-contamination. Warning: Do not use antistatic or electrically conductive hoses or tubing in the ventilator breathing system.
Diagram of patient circuit
Ventway Sparrow User Manual
Page 22 of 110
5.6. PNEUMATIC SE CTION – THEORY OF OP ERA TION
Page 22 of 110
Ventway Sparrow User Manual
Page 23 of 110
Explanation of Diagram
1
Solenoid valves exhaust – scavenging of gas residue from the solenoids , passed to inlet to prevent oxygen buildup inside the
enclosure.
2
Manifold assembly – encapsulates all electromechanical control elements, membranes, valves and air tubes in one compact
element. This design eliminates the use of air tubes and pneumatic connectors.
3
Inspiratory control valve – controls the main inspiratory valve.
4
Exhalation control valve – controls the patient exhalation valve membrane, located in the patient circuit transducer.
5
Exhalation control tube – transferring pressure from the exhalation control valve to the exhalation valve membrane. It is a
part of the patient circuit.
6
Inspiratory tube – a 15 mm or 22 mm tube, connecting the ventilator and the patient. Used for delivering the gas mixture to
the patient’s airways.
7
PEEP valve – Spring activated membrane (same membrane used for exhalation control). Adjustable by rotating a knob.
8
Anti asphyxia valve – designed to allow the patient to inhale through an emergency intake port separated from the regular
inlet, in case of failure of the ventilator.
9
Blower – A radial blower, referred to sometimes as “turbine”. Intended to generate pressure and flow of gas mixture either
from the ambient air or from a mixture of air and oxygen.
10
Bleeding conduit and valve – orifice used for passing small amounts of flow, compensating for any leaks that may occur in
the patient circuit, and remove any CO2 from the patient circuit.
11
Main inspiratory valve – normally open valve. Main valve designed to allow passage of gas from the blower to the patient.
12
Flow and pressure sensor – a part of the patient circuit. A constant orifice sensor used for acquiring the flow and pressure in
the patient’s airways and passing them to pressure sensors located in the ventilator’s electronic unit.
Page 23 of 110
Ventway Sparrow User Manual
Page 24 of 110
Inspiratory (1) vs. Expiratory cycle
During the inspiratory cycle, the exhalation valve membrane is closed, allowing the gas from the ventilator to be diverted to the patient. Once the inspiration cycle is complete, the membrane opens, allowing the gas from the patient to be exhaled through the exhalation valve.
The expiratory gas passes through the membrane which is constantly supported by a spring, creating the preset PEEP value. During expiratory flow, a continuous low flow is delivered into the patient circuit, in order to prevent any re­breathing and to compensate for small leaks that may exist in the various circuit components, facemask or LMA (Laryngeal Mask Airway).
Page 24 of 110
6. CONNECT ING THE VENTILATOR
6.1. FRONT PANEL CONNECTIONS
6.2. REAR PANEL CONNECTIONS
1.
Air inlet/Oxygen supply connector – Air inlet and oxygen connector to
demand valve (see Section 12.1.1.1. Connecting the Air Inlet to the Demand Valve).
2. DC Power connector – Connect power line (AC/DC or DC/DC power supply)
Caution:
Insert t
he three conductor plug into a
properly wired,
three
-
wire,
grounded, hospital grade receptacle. If a three-wire receptacle is not available, a qualified electrician must install one in accordance with the governing electrical code. Do not under any circumstances remove the grounding connector from the power plug. Do not use extension cords or adapters of any type. The power cord and plug must be intact and undamaged.
3. USB connector – This connector is intended for technicians who
require access to the logbook and device software.
4.
Power On/Off button
5. Battery safety lock
6. Battery pack
7. Filter compartment screws
1 Pressure and flow measurement port 2 Pressure measurement port 3 Expiratory valve port – control tube 4 Ventilator gas output
Page 25 of 110
Ventway Sparrow User Manual
Page 26 of 110
7. POWE R ON A ND DISPL AY STARTUP SCR EEN
To start the system, press and hold the Power On button on the back panel for 3 seconds.
Back panel indicating the Power On button
7.1. QUICK START
In an emergency situation, Quick Start allows you to start the ventilator immediately, using preset default options.
To perform a Quick Start, press and hold the Power On button for ten seconds. The following screen will display:
Quick Start initial screen
Ventway Sparrow User Manual
Page 27 of 110
After you confirm by selecting the check mark and pressing the control know, specify the Patient Type:
Patient Type screen
Now the ventilator will operate in Backup mode:
Ventilator running after Quick Start
For information on Backup Mode, see Section 15.3. Backup Ventilation Mode.
Ventway Sparrow User Manual
Page 28 of 110
8. NAVIGAT ING THE GUI SCREENS
8.1. SELECTING SCREEN OPTIONS
To navigate between the screen options, turn the control knob on the left side of the device. When the desired option has been marked by positioning the marker on its location, press the knob to select the option.
8.2. EDITING FIELDS
While turning the control knob, fields that can be modified are highlighted. To edit a field, press the control knob when posi­tioned on the field. The field will change color. Rotate the control knob to view different values for the field, and press the knob to select a value.
8.3. NAVIGATING BETWEEN SCREENS
Press the Back button ( ) to return to the previous screen.
Press the Next Screen button ( ) to return to the previous screen.
Control knob
Patient Weight highlighted
Ventway Sparrow User Manual
Page 29 of 110
8.4. CONFIRMING OR CANCELLING A MESSAGE
When a message is displayed by the system that allows a confirm or cancel response, press Cancel ( ) to reject the action, or press Confirm ( ) to accept the action.
8.5. SYMBOLS USED IN THE SYSTEM
Back
Next screen
Language
Graphs (Flow / Pressure)
Confirm
Cancel
Manual breath
Main Menu
Ventway Sparrow User Manual
Page 30 of 110
8.6. SYSTEM INDICAT ORS
The top left corner of the display shows battery status and system indicators.
8.6.1. BATTERY STATUS
No battery, external power source used
Non-rechargeable battery, no external power source
Non-rechargeable battery, external power source connected
Rechargeable
battery,
external
power
source
connected
(charg
-
ing)
Rechargeable
battery,
external
power
source
connected
(fully
charged)
Rechargeable battery, no external power source connected
Rechargeable battery, not charging
due to battery
tem
perature
protection or faulty battery
Battery charge level is indicated by the following signs :
Battery full
Medium energy level
Low energy level
Battery in critical energy level
Ventway Sparrow User Manual
Page 31 of 110
8.6.2. INDICATORS
_T_
Patient trigger detected
_Z_
Pressure sensors are currently being zeroed
Note: While zeroing sensors, the ventilator will not detect patient triggers.
The bottom left corner of the display shows the active alerts indicator.
8.6.3. ALERT
Caution:
It is
the user’s
responsibility
to
ensure
that the
alert
s are working
properly. See Section 24. Appendix – Test Alerts for instructions on testing the alerts.
Normal operation
Alert is silenced.
Note: While the alert is silenced (maximum 120 seconds) this indicator is displayed, and the num­ber of active alerts is displayed in parentheses next to the icon.
Note: The icon serves as an activation button. Pressing it will open a list of up to five active alerts ordered by priority.
Ventway Sparrow User Manual
Page 32 of 110
9. GETT ING ST ARTED WITH VENTWAY SPARROW
The initial set of screens allow you to start the ventilator and specify settings such as patient weight, ventilation type, and other basic parameters.
Note: A system alert sound is activated during system startup, indicating a valid Circuit Verification test.
Note: An emergency quick start can be performed if necessary. See Section 7.1. Quick Start.
The first screen showing the Ventway Sparrow logo allows the user to identify whether the display is working properly.
9.1. DISCONNECT PATIENT
Initial screen
Normal procedure after turning on the ventilator is to verify that the patient is disconnected , then press OK on this initial screen. The system will activate the blower to verify the correct operation of the device and various compon­ents (e.g. patient circuit, sensors and electrical components).
The self test screen will display for about 8 seconds.
Self Test screen
Ventway Sparrow User Manual
Page 33 of 110
If the self test fails, the user is advised to check the following:
No oxygen demand valve connected to the ventilator No oxygen reservoir Dirty/obstructed inlet filter Pressure and control tube connected properly (kinks, disconnec-
tion or rupture of the tubes)
No HME or filter are connected to the exhalation valve Patient is not connected
9.2. PATIENT WEIGHT
Rotate the scroll knob to select the weight of the patient:
Pediatric and adult weight selection screens
Note: If the Main Menu is selected instead of OK, the START option will be unavailable, since the weight of the patient is unknown and ventilation parameters are not set. For details on the Main Menu option, see Section 10. Main Menu.
Ventway Sparrow User Manual
Page 34 of 110
9.3. VENTILATION MODE
Select the initial choice of ventilation mode. This can be changed later. See Section 15. Ventilation Methods for details about each ventilation mode.
Ventilation Mode screen
After ventilation mode is selected, select Invasive or Non-Invasive Mode:
You will be prompted to connect the patient:
Select the preference regarding Invasive or Non-Invasive ventilation mode.
Ventway Sparrow User Manual
Page 35 of 110
Alerts
Menu
Manual
Breath
Graph
9.4. VENTILATION PARAMETERS
The Ventilation Parameters screen displays the relevant patient information , updated on a real-time basis:
Ventilation Parameters screen
Select BPM to change the breaths per minute delivered. Select VTe to change the desired VTe
Select ALERTS to view the current active (silenced) alerts. Select GRAPH to display the Pressure or Flow graphs – see Section 9.6.
Pressure and flow graphs.
Select MANUAL BREATH to give one breath to the patient using the
current parameters.
Select MENU to display the Menu screen – see Section 10. Main Menu.
Note:
When Non Invasive ventilation mode is applied, only the set
value of VT is displayed in parenthesis. VTE and VTI are displayed as part of the graphs screens of pressure and flow. In Non Invasive ven­tilation the ventilator is programmed to compensate for the loss of volume due to leaks from the facemask or LMA. Actual delivered tidal volume can only be interpolated, and is not displayed.
Example of Non-Invasive Mode display
Ventway Sparrow User Manual
Page 36 of 110
9.5. NUMERICAL REPRESENTATION OF BREATH PARAMETERS
The measured and computed values that are displayed in the system are calculated using the filtering and smoothing techniques described below.
Patient pressure is displayed to the user by a bar (graphic refreshment frequency is 10 Hz). The sample rate of the pressure value is 500 Hz. The Peak Inspiratory Pressure is displayed as a blue background, remaining throughout the current breath, while the current pressure is displayed in green.
Altimeter input is averaged per second and linearly interpolated. Oxygen percentage correction factor for density correction is taken from a
table. Oxygen value is determined by the user.
Flow correction factor is taken from the last performed volume calibration. Real time flow values are calculated using linear interpolation and corrected
by the oxygen percentage, altitude (atmospheric pressure) and calibration correction factors.
Inspiratory and expiratory tidal volumes are calculated by integrating the positive and negative flows over time. The negative (expiratory) tidal volume and minute volume are displayed to the user. The positive tidal volume is used to determine timing of the current breath termination and for blower speed corrections.
Please note that due to possible leaks when using non-invasive ventilation, VTe and VTi may vary substantially. Nevertheless, in non invasive mode, a correction to leaks is performed by the ventilator.
BPM is calculated as one minute divided by the latest breath length. Minute volume is calculated as the latest BPM multiplied by the latest
expiratory tidal volume. I:E is displayed to the user and is calculated per each breath as the inspiration
time divided by the expiration time. The latest PEEP is displayed to the user, and is updated per each breath, as
the pressure measured at the end of expiration. PIP is displayed to the user by a bar graph. It is the highest level of pressure
applied to the lungs during inhalation.
Ventway Sparrow User Manual
Page 37 of 110
9.6. PRESSURE AND FLOW GRAPHS
This display shows the graphs of the pressure (in units of cmH2O) over time, and flow (in units of LPM) vs. time.
Rotation of the knob will switch between flow and pressure graphs.
Flow Graph screen
Pressure Graph screen
During sensor zeroing, diagonal lines are drawn on the graph to indicate no input from the sensors.
Flow graph screen during sensor zeroing
VT
E
VT
E
Ventway Sparrow User Manual
Page 38 of 110
10. MAI N MENU
The Main Menu screen allows the user to view and set various system values.
Menu screen
STOP VENT/START VENT
Stop or start the respiratory function of the device.
NEW PATIENT
View the Disconnect Patient screen and reset all selected parameters. See Section 10.1. Main Menu/New Patient (Disconnect Patient Screen).
VENT. PARAMS
View patient weight, ventilation mode, and other parameters.
See Section 10.2. Main Menu/Vent Params.
ALERT SETTINGS
Set the thresholds at which the alerts will be displayed. See Section 10.3. MAIN MENU/Alert settings.
ADV SETTINGS
Set advanced settings, such as technician mode, brightness, language. See Section 10.4. Main Menu/Advanced Settings.
Ventway Sparrow User Manual
Page 39 of 110
10.1. MAIN MENU/NEW PATIENT (DISCONNECT PATIENT SCREEN)
This screen allows the user to restart existing ventilation parameters, e.g. when ventilating a new patient.
Disconnect Patient screen
10.2. MAIN MENU/VENT PARAMS
These screens allow the user to set the patient weight, ventilation mode, and other ventilation parameters.
Vent. Param. screen (page 1) Vent. Param. screen (page 2)
Ventway Sparrow User Manual
Page 40 of 110
10.2.1. MAIN MENU/VENT PARAMS/TRIGGER SENSITIVITY
Set the trigger sensitivity for the pressure measurements.
Trigger Sensitivity screen
Caution: When closed suction catheterization is performed, patient trigger sensitivity must be turned to "off". Caution: During transport ventilation, when the patient is often moved and subject to abrupt bumps, it is recommended to decrease pressure triggers sensitivity, in order to avoid auto triggers (e.g. -5 instead of -2).
.
10.2.2. MAIN MENU/VENT PARAMS/VENT MODE
The ventilation mode was set initially when the device was turned on. See Section 9.3. Ventilation Mode to change the ventilation mode.
10.2.3. MAIN MENU/VENT PARAMS/O2 ENRICHMENT
10.2.4. MAIN MENU/VENT PARAMS/PATIENT WEIGHT
Ventway Sparrow User Manual
Page 41 of 110
The patient weight was set initially when the device was turned on. See Section 9.2. Patient Weight to change the patient weight.
10.3. MAIN MENU/ALERT SETTINGS
The Alert Settings screens allow the user to set the threshold values for each type of ventilation parameters alert.
Alert Param. screen (page 1) Alert Param. screen (page 2)
See also Section 11. Warnings and Alerts for more information about warning and alert messages.
Caution: Boot Fail Alert
The Boot Fail screen appears if the system experiences an error when starting. Replace ventilator and contact Inovytec support.
Boot Fail screen
Ventway Sparrow User Manual
Page 42 of 110
10.4. MAIN MENU/ADVANCED SETTINGS
The Advanced Settings menu screen is shown below.
Advanced Settings menu
These options are explained in the following sections.
10.4.1. MAIN MENU/ADV SETTINGS/LOAD DEFAULT
Set all parameters to their default values by the set weight.
The Load Default screen
10.4.2. MAIN MENU/ADV SETTINGS/BRIGHTNESS
Change the brightness of the display.
Ventway Sparrow User Manual
Page 43 of 110
10.4.3. MAIN MENU/ADV SETTINGS/ALERT VOLUME
Change the volume of all sound cues of the system.
10.4.4. MAIN MENU/ADV SETTINGS/LANGUAGE
Change the system display language.
10.4.5. MAIN MENU/ADV SETTINGS/TECH MODE
Tech mode allows the setting of the time, total hours of operation, and provides a system self-test function.
Tech Mode screens
Ventway Sparrow User Manual
Page 44 of 110
10.4.5.1. MAIN MENU/ADV SETTINGS/TECH MODE/SET TIME
Set the system date and time.
The Set Time screen
10.4.5.2. MAIN MENU/ADV SETTINGS/TECH MODE/CALIBRATION
Under Tech mode, the user may perform volume calibration by choosing the calibration screen. The Calibration screen requires a device key to be entered:
Calibration screen: Enter key
Then the Patient Type screen displays:
Choose the patient type, before calibrating
Ventway Sparrow User Manual
Page 45 of 110
The use of fixed volume 500 cc syringe provides an easy and accurate solution for volume calibration. The user must move the syringe handle, completing a cycle in one to two seconds, resulting in an average flow of 15 to 30 lpm. The calculated tidal volume is displayed for each stroke, and once the tidal vol­umes (VTI and VTE) are between 490 to 510 cc, the user may finish the cali­bration process by pressing the scroll knob.
Volume Calibration screen
The ventilator compensates for each stroke for tidal volume inaccuracies. The tidal volume per stroke becomes closer to 500 cc, once it is automatically mul­tiplied by a correction factor. At the beginning of each stroke, the ventilator generates a sound, indicating to the user to start moving the syringe handle.
10.4.5.3. MAIN MENU/ADV SETTINGS/TECH MODE/WORK HOURS
Work Hours screen
Ventway Sparrow User Manual
Page 46 of 110
10.4.5.4. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST
Self Test screen
The Self Test screen allows four tests:
FLOW AND PRESS – see Section:
10.4.5.4.1 MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /FLOW AND PRESSURE
CVT (circuit verification test) see Section:
10.4.5.4.2 MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /CVT
VVT (ventilator verification test) see Section:
10.4.5.4.3. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /VVT
BLOWER TEST – see Section:
10.4.5.4.4. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /BLOWER TEST
Ventway Sparrow User Manual
Page 47 of 110
10.4.5.4.1. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /FLOW AND PRESSURE
While the test is in progress, a progress bar displays:
Self test progress bar
When the test is complete, a test result screen displays:
Self test results screen
In this example, the test results indicate that the ventilation circuit is ready for use, and the ventilator pneumatic and electrical systems are in order.
Ventway Sparrow User Manual
Page 48 of 110
10.4.5.4.2. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /CVT
Follow the instructions on the screen for the Circuit Verification Test:
CVT screen
Use a cap to seal off the patient transducer valve:
Press the navigation knob on the ventilator front panel to begin the test. A pop-up appears, indicating the test has begun.
After several seconds, another pop-up will direct the user to remove the cap blocking the transducer.
After the ventilator does further testing, the test results will display.
Press the control knob to complete the C.V.T.
CVT Results screen
The following information is displayed:
PRESS. CHECK: Pressure is over 60 [cmH2O]
PRESS. TIME: Time to reach pressure of 40 [cmH2O]. FLOW CHECK: Flow is over 120 [lpm]. FLOW TIME: Acceleration time of blower to 100 [lpm].
Ventway Sparrow User Manual
Page 49 of 110
10.4.5.4.3. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST /VVT
Follow the instructions on the screen for the Ventilator Verification Test:
VVT screen
In the VVT process, the ventilator is checked for:
System alert – audio
Patient pressure measurement and pressure performance
Blower pressure measurement
Transducer tubes leak
Motor speed measurement
Two Solenoid valves
Solenoid safety release mechanism
Flow performance and flow zeroing accuracy
Watchdog safety device
The user specifies two pressure values, while the patient circuit is con­nected to a manometer. After the user dials in the first pressure value, the blower increases the pressure and the user dials in the second pressure value. The values will be used to calibrate the pressure sensor:
VVT pressure calibration screen
Ventway Sparrow User Manual
Page 50 of 110
10.4.5.4.4. MAIN MENU/ADV SETTINGS/TECH MODE/SELF TEST
/BLOWER TEST
Follow the instructions on the screen for the Blower Test.
Ventway Sparrow User Manual
Page 51 of 110
10.4.5.5. MAIN MENU/ADV SETTINGS/TECH MODE/LOGBOOK
The Logbook screen shows alerts, indications and user interaction with the ventilator:
Example of an alert
Note: The logbook is displayed in technician mode. Download is possible only in technician mode.
Note: Press BACK to exit log or SCROLL to continue viewing the log.
10.4.5.6. MAIN MENU/ADV SETTINGS/TECH MODE/EXPORT LOG
10.4.5.7. MAIN MENU/ADV SETTINGS/TECH MODE/SW VERSION
Ventway Sparrow User Manual
Page 52 of 110
10.4.5.8. MAIN MENU/ADV SETTINGS/TECH MODE/SW UPDATE
10.4.5.9. MAIN MENU/ADV SETTINGS/TECH MODE/SYSTEM MODE
10.4.5.10. MAIN MENU/ADV SETTINGS/TECH MODE/ALTITUDE
The altitude above sea level is shown on the bottom of the second Tech Mode submenu.
Example of altitude display: 72 meters
Ventway Sparrow User Manual
Page 53 of 110
11. WARNIN GS AND ALERTS
This section provides information about alerts, and examples of typical system messages.
11.1. SUMMARY OF ALERT TYPES
The following table summarizes the various types of alerts.
Note: If no patient weight was set, the user will not be able to change any default alert parameters. Note: Minimal alarm sound level is 80 dB at a distance of one meter from the ventilator.
Alert type
Value
Range/Display
User
Adjustable
Respiratory rate
High: 12
-
60 bpm
Low: 1-20 bpm
Inspiratory pressure
Limit: 10
-
65 cmH
2
O
Alert: 10-55 cmH2O
Apnea
5-
120 seconds
Leak
0-
100%
Low Tidal Volume Delivered
Off or 15%
-
85%
Volume Limit Reached
70-
1800 mL
Inverse I:E Ratio
ON/OFF
Alert Volume
0 (off) 1
-34
Pressure limit
10-
60 cmH
2
O
Pressure
alert limit
10-
55 cmH
2
O
AC Powe
r Disconnect
Screen icon and an audio cue
Additional
Low Battery
Estimated battery energy
level + Alert
Empty Battery
Screen Icon + Alert
Battery Disconnect
Screen Icon
(swapping)
Check Sensor tubes
Alert
Tube Disconnect
Alert
High PEEP
Alert
Service Notice
Alert
System Boot Failed
Alert
Ventway Sparrow User Manual
Page 54 of 110
11.2. SUMMARY OF ALERT LEVELS
Alerts are defined with three levels of importance:
High: Five beeps every two seconds Medium: Three beeps every five seconds Low: One beep one time only. Not shown in list of active alerts.
Alert category
Alert
High level Blower malfunction
Tube disconnect Patient disconnect Apnea
System recovered from a crash
Battery empty Sensor disconnect Low respiratory rate High minute volume Low minute volume High inspiratory pressure Low inspiratory pressure Leak Inverse I:E ratio Battery low High temperature Expiratory valve blocked High PEEP
Low PEEP
Alert category
Alert
Medium level
High respiratory rate Volume limit reached Low tidal volume Low pressure
Low level Replace filter
Service needed Altitude out of range External power
connected/disconnected (icon instead of popup message)
Battery charge protection active
Ventway Sparrow User Manual
Page 55 of 110
11.3. WARNING EXAMPLE
Here is an example of a warning message:
Warning message
11.4. ALERT EXAMPLE
A typical alert appears as follows:
Example of an alert
Note: Press cancel sign (big X) to acknowledge the message and cancel the alert.
A High PEEP alert is an exception to the normal action taken with an alert message, since the user may choose to accept the high PEEP value, if it was intentionally changed. This action saves valuable time adjusting the PEEP level.
High PEEP alert
Ventway Sparrow User Manual
Page 56 of 110
12. DEF AULT PAR AMETERS
12.1. START VOLUME VENTILATION
Patient
Weight
Rate
Tidal
Volume
Max.
Pressure
(Kg)
(bpm)
(ml)
(cm
H2O)
5 35 30 21
10 30 100
23
15 25 150
25
20 20 200
30
30 18 300
33
40 16 400
35
50 14 500
40
60 12 600
40
70+
12
700
40
12.1.1. OXYGEN SUPPLY
When a high-pressure oxygen source that is connected to a demand valve is not available, the Ventway Sparrow ventilator can accept oxygen from a low­pressure oxygen source such as a reservoir bag connected to a flow meter.
To do this, use an optional low-pressure oxygen enrichment system attached to the ventilator air inlet port through an optional Ventway adapter. Adjust the flow of oxygen to reach the desired value of FiO2. Select the "O2 ENRICHMENT" option in the VENT. PARAMS menu. The FiO2 value must be measured with a calibrated external oxygen analyzer.
If a high-pressure oxygen source is available, an Inovytec-approved demand valve can be used to connect to the air inlet port, delivering between 30% to 95% FiO2 to the patient. Measure FiO2 with a calibrated external oxygen
analyzer.
The oxygen supply pressure shall be according to manufacturer specifications (usually 40-60 psi).
Ventway Sparrow User Manual
Page 57 of 110
12.1.1.1. CONNECTING THE AIR INLET TO THE DEMAND VALVE
The oxygen demand valve is connected to the device as shown below. The oxygen demand valve connector is specially threaded to fit securely on top of the air/oxygen inlet.
Connecting the oxygen demand valve to the air inlet connector
The demand valve/mixer may be adjusted so to mix atmospheric air and oxygen in ratios between 30% to 95%. The user may change the oxygen percentage of the gas mixture by rotating a knob at the base of the demand valve.
The high pressure connection to the demand valve should be performed using a standard female DISS connector.
Ventway Sparrow User Manual
Page 58 of 110
12.1.2. RECOMMENDED DEVICES FOR MONITORING OF OXYGEN
Company Name Product Model Precision Medical PM5900 Oxygen Monitor
Maxtec MaxO2 ME® ENVITEC MySign®O
Caution: Use of the low-pressure oxygen system at concentrations
above 60% is NOT recommended, as higher values combined with varying minute volume due to spontaneous breathing of the patient may cause inadvertent PEEP.
Caution: Please read the manufacturer's instructions before using the oxygen monitoring device.
Caution: Oxygen enrichment delay time until reaching 95% oxygen at the patient port: 45 [S].
Caution: Measurement of oxygen should be performed using an external oxygen sensor. The location of the sensor should be at the outlet of the ventilator between the ventilator and the patient circuit.
Ventway Sparrow User Manual
Page 59 of 110
12.2. ALERT DEFAULT PARAMETERS: 5 KG TO 70+ KG
Alert Parameter
Weight
5kg 10kg 15kg 20kg 30kg 40kg 50kg 60kg 70+
kg
Pressure
Alert (cm H2O)
21 23 25 30 33 35 40 40 40
Pressure
Limit
(cm H2O)
25 28 30 35 40 42 47 50 50
High
Rate
(BPM)
45 40 35 30 30 30 30 25 25
Low
Rate
(BPM)
12 10 10 10 8 6 6 6 6
High
Min.
Vol. (LPM)
2.5 4.2 5.3 5.6 7.6 9 9.8 10.1
12
Low
Min
Vol. (LPM)
1 1 1 2 2 3 3 4 4
12.3. ALERT DEFAULT PARAMETERS: ALL PATIENT WEIGHTS
PEEP > 2.5 cm H2O from set value Low pressure 5 cm H2O
Leak (%) 100% (off) Inverse I:E Ratio ON Low Tidal Volume Alert Range 85% of set volume
Ventway Sparrow User Manual
Page 60 of 110
13. LAB ELS AND SYMBOLS
13.1. LABELS
A number of internationally recognized symbols are found on the labels. These relate to safety requirements and standards and are described below.
VWSP-100 Civil Model
VWSP-900 Military Model
Ventway Sparrow User Manual
Page 61 of 110
13.2. SYMBOLS
The following table explains the meaning of each symbol on the label.
Symbol Meaning
Consult instructions for use
Manufacturer
European approval mark
Authorized representative in the European Community
Serial Number
Catalogue number
Batch code
Direct current
Do not dispose of, contact for recycling
FCC Symbol
Caution: law prohibits dispensing without prescription
Caution, consult accompanying documents
Type BF Applied Part
Ventway Sparrow User Manual
Page 62 of 110
14. CLE ANING AND DISINFECTING
Caution: The system is approved for IP45 in operation mode with oxygen enrichment. To avoid damage to the instrument or patient cable, be careful of liquid spillage while cleaning. Caution: Do not expose the instrument, patient cable or sensors to sprays, or any other type of solvents. Caution: Be sure to turn the power off and disconnect the AC power cord from the power source before performing cleaning procedures.
This instrument requires routine cleaning, which includes removal of any soil or dirt from the external surfaces. A soft cloth dampened lightly with water may be used.
Spray the entire surface of the device with 70% alcohol.
Leave the alcohol on the device for an exposure time of 2-3 minutes.
Wipe the surface of the device with Pharma-Wipes 70% alcohol.
Inspect the device for residual debris. Should any debris remain, repeat the entire cleaning procedure until all debris is removed.
Ventway Sparrow User Manual
Page 63 of 110
15. VENTILATION ME THO DS
15.1. SIMV-VC (PS)FLOW CHART
In SIMV-VC, the patient is supplied with the set tidal volume VT during the mandatory breaths. The mandatory breaths are synchronized with the patient’s own breathing attempts. Mandatory breaths are prevented from being applied during spontaneous expiration, by providing that a patient­triggered mandatory breath can only be triggered within a trigger window.
If the expiration phase (and with it the spontaneous breathing time) is shortened on account of synchronization, the next expiration phase will be extended. This adaptation prevents a change in the number of mandatory breaths. If no independent breathing attempt is detected during the trigger window, the machine-triggered mandatory breaths are applied. Thus, the minute volume MV remains constant over time. If the breathing attempts of the patient are insufficient to trigger the mandatory breath, the machine-triggered mandatory breaths are applied. The patient can breathe spontaneously at PEEP level during the expiration phase. During spon­taneous breathing at PEEP level, the patient can be pressure-supported using PS.
Please note:
1. The number of breaths is set by the patient weight (more weight = less
breaths/minute). BPM=mandatory breaths.
2. It is volume controlled (although pressure is constantly being mon-
itored).
3. Either the machine or patient can trigger the breath.
4. Trigger definition: If the patient reaches -2 cm H2O in less than 300
msec. then a trigger breath is initiated.
5. Trigger activation window: 300 msec after termination of breath until
initiation of new breath If no independent breathing attempt is detected during the trigger window, the machine-triggered mandatory breath is applied.
Ventway Sparrow User Manual
Page 64 of 110
6. If no spontaneous breathing attempt is detected during the inspiratory trigger window, the machine-triggered mandatory breath is applied.
The mandatory tidal volume (VT) results from the pressure difference between PEEP and Pinsp, the lung mechanics and the breathing effort of the patient. Pinsp is the inspiratory pressure (pressure in the lung throughout the inspiratory cycle). If the lung mechanics changes (pressure, water in the lungs etc.) the ventilation will not change the volume, and the only change may be the PIP (within the safety/alert pressure values).
Ventway Sparrow User Manual
Page 65 of 110
Note: Through the entire ventilation cycle, we need to test that:
The patient is still connected (no de-
crease in PEEP pressure)
Expiratory flow is undisturbed (pressure
has decreased – nothing is blocking the ex. valve)
Value of Ti/Te is OK (within the norm)
.
Ventway Sparrow User Manual
Page 66 of 110
15.2. CPAP – SPONTANEOUS BREATHING MODE
During the spontaneous ventilation modes, the patient carries out the majority of the breathing effort. The pressure support ventilation level PS can be adjusted. The default PSV level is 18 cm H2O.
As long as the patient’s airways are below the PEEP + PSV level, the ven­tilator will perform ventilation support. This setting directly affects the flow and thus the supplied tidal volume (VTI).
Ventway Sparrow User Manual
Page 67 of 110
Ventway Sparrow User Manual
Page 68 of 110
CPAP Backup ventilation
Note: BPM is based on patient
weight and other parameters.
Ventway Sparrow User Manual
Page 69 of 110
15.3. BACKUP VENTILATION MODE
Note: Backup will not occur when the user inserts weight, selects a ventilation mode and does not press OK. In this situation, a trig­ger is received from the patient and the respiration is performed according to the ventilation mode.
Backup mode is an alternative ventilation method taking place when the ventilator patient circuit is connected to the patient in these situations:
1. Patient is ventilated in CPAP mode from more than 10 seconds for pediatric (5KG) and up to more than 25 seconds for an adult (70+ KG).
2. Patient is connected to the ventilator, and for some reason, the user did not start ventilation after connecting the patient. The backup starts after three breaths.
3. Patient is connected to the ventilator, and user has paused venti­lation. In this case when trigger is sensed by the ventilator, the ven­tilation will automatically resume.
The result of using this backup mode is that the patient will be ventilated in case of losing the ability to initiate spontaneous breaths in CPAP mode, or if an emergency or user error has occurred in the preliminary stages of ventilation setup, or when pausing the ventilator for some reason without resuming ventilation.
15.3.1. BACKUP VENTILATION IN CPAP
If during CPAP ventilation, the patient stops initiating spontaneous breaths from more than 10 seconds for pediatric (5KG) and up to more than 25 seconds for an adult (70+ KG) --> Apnea alert appears.
Then backup ventilation is initiated:
-Screen displays an alert: "Backup Ventilation Started" and "OK" button for user approval appears. Ventilation will start even if user has not pressed OK. Alert will appear on screen while audio prompt is activated.
BPM = patient weight based
Volume, Pressure, MV and all other alerts which are patient weight
based remain unchanged.
Ventway Sparrow User Manual
Page 70 of 110
Ventilation method is changed to SIMV - VC - PS with the last patient
weight parameters dialed in.
BPM value is patient weight based. Resuming to CPAP ventilation will be possible either by patient trigger
(will alert on screen: "CPAP ventilation resumed" - "OK"), or user switching back to CPAP or SIMV - VC - PS manually, through the "MENU" option on the main screen.
All alerts and user inputs will be logged in logbook. An alert is issued "Backup ventilation active"--"OK".
15.3.2. BACKUP VENTILATION BEFORE STARTING PATIENT VENTILATION
If patient is connected to the ventilator, the following conditions may also initiate backup ventilation:
a) Ventilation has not started yet, but ventilator is turned on (either patient weight, ventilation method or start command was not chosen by the user). b) Patient trigger sensed.
If conditions a) and b) above were met, then backup ventilation starts.
"Backup ventilation mode" alert is activated - "OK" displays.
o If the user approves, the Patient Type window appears,
and then the backup starts.
o If the user does not approve, the action is canceled and
the ventilation does not start.
The "Backup ventilation mode" icon appears on top of the screen
after last alert was either accepted or discarded.
If Patient weight was entered, the device ventilates according to the given weight parameters.
If Patient weight was not entered:
1. Initial pressure support value is set to PS=10 cm H2O.
2.
Tidal volume is calculated during ventilation.
3. 3 breaths are given to the patient at identical parameters (PS=10 cmH2O) at 10 BPM rate.
Ventway Sparrow User Manual
Page 71 of 110
4. After 3 breaths, the patient weight is evaluated by using TV average of the last 3 breaths:
TV[cc] Patient weight [kg]
50-70 5
71-140 10 141-210 15 211-280 20 281-420 30 421-560 40 561-700 50 701-840 60
Over 841 70+
5. Evaluating patient weight will establish the other ventilation par­ameters as per a more complete table given in the URS/SRS, including maximum TV, BPM etc.
If Patient weight and ventilation mode were entered, ventilation will start in selected mode and will issue an alert “Ventilation started – “OK”. If patient trigger is sensed, then spontaneous breath is initiated with the same ventilation parameters.
• An alert appears "Backup ventilation active -- OK".
• User may change the ventilation parameters by entering MENU and selecting new ventilation parameters.
If the ventilator did not complete the initial verification, an alert will be activated 15 seconds after starting backup ventilation: "Initial ven­tilator verification needed --OK". This alert will appear only once.
3. Backup Ventilation in Pause mode In case the user chose to pause the ventilation for any reason, and patient trigger (pressure -2.5 cm H
2
O or -5 lpm or greater) was sensed, then ventilation is resumed with the exact same parameters as set by the user. The alert is issued: "Ventilation is resumed -- OK".
Ventway Sparrow User Manual
Page 72 of 110
15.4. SIMV VC (CPR)
In SIMV-VC (CPR), as in the normal SIMV-VC (PS), the patient is supplied with the set tidal volume VT during the mandatory breaths.
The only difference in this setup, is that patient triggering is switched to OFF. This is very useful during cardiopulmonary resuscitation combining chest compressions.
Chest compression is known to activate undesirable patient triggers causing the ventilator to continue delivering spontaneous breaths at a high rate (uncontrolled auto triggering).
By turning off the triggers, the ventilator will not alert due to high minute volume, breath rate or inadequate I:E ratio.
Ventway Sparrow User Manual
Page 73 of 110
15.5. CPAP (CPR)
This setup delivers constant pressure into the patient airways as in the regular CPAP ventilation method, but without any pressure support during triggers (no response to spontaneous breaths).
Constant pressure values can be set between 5 to 10 cm H2O, without activation of triggers. PEEP is set by the external PEEP control knob on the patient transducer.
This setup saves the time of entering the ventilation parameters menu, trigger sensitivity, and decreasing the trigger sensitivity.
CPR mode may be used when prolonged cardiopulmonary resuscitation (CPR) is performed. The following parameters are displayed in CPR mode :
BPM – Number of compressions sensed by the patient pressure sensor. The number may vary due to the strength, frequency of com­pressions, and leaks from the patient airways.
Rate – Audio prompt frequency ranging from 80 to 120 may be activated in order to indicate rhythm of compressions to the caregiver.
PEEP – Adjust by rotating the rotator knob of the exhalation valve.
Pressure bar graph – Pulmonary pressure is displayed on the bar graph.
Manual breath – The activation of manual breath will result in a single
pressure support breath, of default value (PEEP + 18 cmH2O). The caregiver may change the PSV value by entering the Ventilation Parameters menu.
Caution: If an unexpected shutoff event occurs, the ventilator will alert for the unexpected shutoff when it is next activated. If the incident occurred during patient ventilation, and the event took place no more than two minutes before, the ventilator will continue with the last ventilation parameters automatically.
Display in CPR mode
Ventway Sparrow User Manual
Page 74 of 110
16. SER VICE AND MAINTENANCE
Note: The user or any technical personnel who are not formally authorized by Inovytec Medical Solutions Ltd. must not open the Ventway Sparrow device under any circumstances. Opening the Ventway Sparrow device may damage the unit and will void the warranty provided by Inovytec Medical Solutions Ltd.
The system requires maintenance on a routine basis of 15,000 operation hours or three years (whichever is first). Service should only be provided by an authorized Inovytec Medical Solutions Ltd. representative.
16.1. DEVICE CALIBRATION AND SOFTWARE UPGRADES
Calibration and upgrade of the software are performed by the factory, or in technician mode using a password provided by Inovytec.
Ventway Sparrow User Manual
Page 75 of 110
17. TRO UBLESHOOTING
The following table lists some typical conditions that may occur when using the system.
Condition Possible Cause Recommended Action
Battery disconnected
Not inserted properly
Firmly insert battery in its place
DC connector connected and no sign of charging
Power supply not connected to power source or DC connector
Check connection to ven­tilator and power source
Patient circuit verification failed
Accessories or patient connected, control tube connector disconnected
Check patient circuit con­nections
Ventway Sparrow User Manual
Page 76 of 110
18. SPE CIF ICATIONS
18.1. DIMENSIONS AND WEIGHT
Width 165 mm
Length 167 mm
Height 60 mm
Weight Civilian version: 1 kg with batteries
Military version: 1.3 kg with batteries
Ventway Sparrow User Manual
Page 77 of 110
18.2. ENVIRONMENTAL SPECIFICATIONS
Operating
Temperature
-18 °C to 50 °C (0 °F to 122
°F)
Storage
Temperature
-20 °C to 70 °C (-4 °F to
158
°F)
Relative
Humidity
10% to 90%
Water
and
Dust
Resistance
IP45 (with
oxygen
enrichment
)
IP20 (when using power supply)
Note:
First digit
(“4
” In
IP45
, or “2” in
IP20)
relates to object size protection (<1 mm).
Second digit (“5” in IP45 or “0” in IP20) indi­cates protection from water jets from any direction.
Atmospheric
Pressure
700
hPa
to
1060
hPa
Altitude
-
370
to
4500
meters
(-
1,200 to 14,764 feet)
Total
External
Sound
Level
54 dBa
at
0.6 meter
distance
Max acoustic energy level
55 [dBa]
18.3. HARSH ENVIRONMENTAL CONDITIONS
The Ventway Sparrow is designed to operate in extreme environments, such as ground and transport. The Ventilator continuously monitors the ambient pressure, temperature and other parameters, and compensates for these changes.
Although the unit may activate a low priority alert that indicates ex­ceeding regular operation conditions, please remember that these alerts are advisory, and the unit will continue to operate normally.
18.3.1. OPERATION IN EXTREME HIGH / LOW TEMPERATURES
During high temperature operation, the ventilator may prompt an alert regarding its high temperature charge limit. The battery protection circuit will stop charging below 0°C and over 50°C, as measured inside the battery pack.
Ventway Sparrow User Manual
Page 78 of 110
Normal ventilator operation may create heat in the battery compart­ment. When operating in a hot environment you should remove the ven­tilator padded case, increasing heat dissipation into the environment.
During low temperature operation, it is recommended to use the venti­lator padded case. This will insulate the unit from the outside temper­ature and retain the internal heat created by the blower motor and other electrical heat-generating components.
18.3.2. OPERATION IN HIGH OR LOW ALTITUDE
The Ventway Sparrow is designed to operate at an altitude of -370 to 4,500 meters (-1,200 to 15,000 feet). An internal altimeter measures the ambient pressure, allowing the ventilator to compensate the flow calculations for the changes in density and pressure.
If the operation altitude is outside of the specified values, a low priority alert will be activated. In this case the user should monitor the Peak Inspiratory Pressure (PIP) and adjust the tidal volume so the same PIP is achieved. It is also advised to monitor the breathing sounds and chest excursion to assure the unit maintains adequate ventilation.
18.3.3. AIRBORNE PARTICULATES
The inlet filter of the Ventway Sparrow provides protection of gas flow paths through the inlet and emergency intake ports. However, in areas where fine dust is present, it is recommended to use a disposable bac­terial/viral filter.
This may prolong the lifetime of the internal filter. Visually inspect the filter for dust and dirt buildup for extended operation in dusty environments.
If the filter becomes dirty, it must be replaced. Circuit Verification Test (CVT) must be performed without the filter attached to the ventilator, since it may cause a pressure drop that will result in not passing the test.
Ventway Sparrow User Manual
Page 79 of 110
18.4. POWER SUPPLY
External
AC-DC
Adapter
Input
100
to 264 VAC, 50-60 Hz, max
1.6 A
Output 16-24 VDC , 120 W
Internal
Battery
8 x CR123
cells
for 12VDC
configuration
(2 x 4S1P
)
4 x 18650
Li-
Ion for 14.8VDC
rechargeable
configuration (4S1P)
Recharge
Time
6 hours
Operating
Time
(internal
battery)
4 hours
Note: T
he time required for the ventilator to warm from the minimum
storage temperature between
uses until it is ready for intended use is 20
minutes in STP conditions.
Note: T
he time required for the ventilator to cool from the maximum storage
temperature between uses until it is ready for intended use is 30 minutes.
Ventway Sparrow User Manual
Page 80 of 110
18.5. VENTILATION PERFORMANCE
Note:
Measurement accuracy of calibration e
quipment may
affect the respiratory parameters display by the following parameters:
Volume calculation : +/-2% or 20 [cc] Inspiratory and expiratory times : +/- 0.02 [s]
Pressure :
+/-
0.75% or +/
- 0.1 [mbar]
Note:
If temperature exceeds the range of
-18°
C to 50
°
C, a
temperature alert may be activated. This may result in breath parameters inaccuracies that exceed the tolerances specified
below.
Note:
If humidity exceed
s 95% and condensation occurs, it
may lead to water droplet buildup in the sensing
tubes
.
Note:
If t
he ventilator is exposed to altitudes outside of the
defined values, the ventilator may be unable to accurately compensate for volume and pressure deviations.
Respiratory
Rate
1 – 12 ±1, 12 – 60 ± 2
bpm
Tidal
Volume
50-1800
mL
Acc
uracy
of
Respiratory Volume Measurement
±15%
+ 4 ml
Inspiratory
Pressure
Limit
5 to 65 cm H2O
Inspiratory
Time
0.3
to 3
±10%
sec
Peak
Flow
(PIF)
130 (-
5% to +10%)
% L/min
Spontaneous to 180 ±10% L/min
Oxygen
Mix
(FiO
2
) 21%
to
95%
±5%
O2
PEEP
3 to 13 cm H2O ±1, (external
ly adjusted)
Trigger
Sensitivity
-2 to 10 cm
H2O Pressure,
Off
PS 5 to 45 cm
H2O
Controlled
Pressure
5 to 65 cm
H2O
FiO
2
low
pressure
enrichment
21%
to
100%
(low
and
high
pressure
oxygen
source)
Ventway Sparrow User Manual
Page 81 of 110
18.6. STANDARDS AND SAFETY REQUIREMENTS
The Ventway Sparrow meets the requirements of the following inter­national standards:
IEC 60601-1 Medical electrical equipment — Part 1: General
requirements for basic safety and essential per­formance
IEC 60601-1-2 Medical electrical equipment — Part 1-2: General
requirements for basic safety and essential per­formance — Collateral Standard: Electromagnetic disturbances — Requirements and tests
IEC 60601-1-12 Medical electrical equipment - Part 1-12: General
requirements for basic safety and essential per­formance - Collateral Standard: Requirements for medical electrical equipment and medical electri­cal systems intended for use in the emergency medical services environment
ISO 80601-2-12 Medical electrical equipment -- Part 2-12:
Particular requirements for basic safety and essential performance of critical care ventilators
ISO 10651-3 Lung ventilators for medical use -- Part 3: Par-
ticular requirements for emergency and transport ventilators
EN 794-3 Lung ventilators. Particular requirements for
emergency and transport ventilators
RTCA DO-160G Environmental Conditions and Test Procedures
for Airborne Equipment Rotary wing – Helicopters
EN 1789 Medical vehicles and their equipment. Road
ambulances
Ventway Sparrow User Manual
Page 82 of 110
19. CLE ANING AND R OUT INE MAINTEN ANC E
Part Procedure Comments Ventilator 1. Spray entire
surface with 70% Alcohol
2. Do not wipe device for 1 min.
3. Wipe device with pharma wipes 70% alcohol
Do not allow liquid to penetrate into the ven­tilator.
Inspect for residual de­bris.
Air Inlet Filter Replace every 300
hours (or 1 month) of operation, or as necessary.
Do not attempt to clean or reuse the air inlet filter.
Ventway Sparrow User Manual
Page 83 of 110
20. BATTERIES
The Sparrow can be configured with two battery options: an internal, re­chargeable, battery pack, and a rechargeable battery pack. The battery should be charged or replaced when not in use. For long storage periods, the battery should not be inserted into the battery compartment of the ventilator.
20.1. BATTERY MAINTENANCE
Rechargeable batteries
The internal battery pack is charged from the external device connector, using the supplied 18-24 volt regulated power supply.
Caution:
If
the batteries
do
not
charge,
replace
the char
ger or the
batteries.
Caution:
Battery failure may occur if maintenance is not performed on
time and according to instructions.
Note:
If
charging
at
temperatures
over
35°C (
95°F)
, charging
time
may
be
extended, and battery thermal protection may be a
ctivated preventing
the battery charge.
Shelf life: 4 years. First time charging: 6 hours. Permissible battery pack charging temperature range: 0°C to 50°C (32°F to 122°F) Permissible temperature range for long-term storage: 10°C to 35°C (50°F to 95°F)
Charge Interval: 6 months in standard temperature and pressure Shelf life: 4 years or 1000 cycles at 25°C (77°F). A cycle is defined as one charge and one discharge.
Max discharge current 10[A] Storage temperature for shipping state (about 40% capacity of fully charged state):
1 month
-30°C to 60°C (-22°F
to
140°F)
3 month
-20°C to 45°C (-4°F to 113°F)
1 year
-20°C to 20°C (-4°F to 68°F)
Non-Rechargeable batteries
Shelf life: 5 years. Long term storage: -20°C to 35°C (-4°F to 95°F)
Ventway Sparrow User Manual
Page 84 of 110
21. PARTS AND ACCE SSO RIES
This Section outlines information for ordering and shipment of replace­ment parts for the Ventway Sparrow.
All equipment and accessories are available directly from Inovytec Medical Solutions Ltd. or from an authorized distributor. For a list of Inovytec Medical Solutions Ltd. distributors please visit our web site at
sales@Inovytec.com.
When ordering parts, specify the serial number of the unit and the part number of the item(s) ordered. A schedule of part numbers may be found in the Equipment and Accessory Inventory list below.
Forward orders to:
Inovytec Medical Solutions Ltd. 3 Hanagar St., POB 7282, Hod-Hasharon 4501306, Israel Tel: +972 9 7794135
E-mail: Info@Inovytec.com
Web Site: http://www.Inovytec.com
Caution: It is strongly recommended that all Ventway Sparrow
parts be replaced with parts purchased from Inovytec Medical Solutions Ltd. or an authorized distributor. Use of other parts may damage the unit and could void the unit warranty.
Note: Dispose of this device and of used sensors in accordance with local regulations.
Ventway Sparrow User Manual
Page 85 of 110
22. REGULATORY
Manufacturer:
Inovytec Medical Solutions Ltd. 3 Hanagar St., POB 7282, Hod-Hasharon 4501306, Israel Tel: +972 9 7794135
E-mail: Info@Inovytec.com
Web Site: http://www.Inovytec.com
Européen Agent Information:
OBELIS S.A.
Bd Général Wahis, 53 B-1030 Brussels Belgium Telephone +3227325954 Fax: +3227326003
E-mail: mail@obelis.net Website www.obelis.net
0086
Ventway Sparrow User Manual
Page 86 of 110
23. WARRANTY
Service Support
Repairs of the System under warranty must be made by authorized repair centers. If the device needs repair, contact Inovytec Medical Solutions Ltd. service department or your local distributor.
If shipping the device is required, pack the device and its accessories carefully to prevent shipping damage.
Duration
Inovytec Medical Solutions Ltd. will repair or replace, at its sole discretion, the product or any defective part, provided it is returned to Inovytec Medical Solutions Ltd. service within 30 days.
Ventway Sparrow User Manual
Page 87 of 110
24. APP ENDIX – TES T ALERTS
The following instructions explain how to test all the alerts that can be activated by the system, to ensure that they are working properly.
24.1. BACKUP VENTILATION
Step Procedure Action Expected Results
1
Start a new patient via the menu.
Set the lung simulator simulated weight to 50 kg.
Message requesting to disconnect the patient.
2
Generate inhalation trigger.
The software switches to the running ventilation screen with unknown weight, in SIMV ventilation mode.
The software indicates that it is in backup ventilation.
3
Inspect the parameters on the screen.
The blower starts, with PS = DefaultPS, and presents the ventilation parameters.
4
Wait for two breathes.
The following ventilation parameters are automatically set according to the simulated patient weight:
Alert Pressure, Minute Volume M.V., BPM, Tidal volume Vt, Ti, Te, PS, trigger sensitivity.
5
Wait for a few minutes.
Ventilation continues with parameters suitable for a patient weighing approx. 50 kg.
The software keeps indicating that it is in backup ventilation mode.
The software instructs the user to perform patient circuit test (via the menu).
6
Stop ventilation. Ventilation stops.
Ventway Sparrow User Manual
Page 88 of 110
24.2. START BACKUP VENTILATION WITH UNKNOWN WEIGHT
Step Procedure Action Expected Results
1
Start a new patient via the menu.
Set the lung simulator simulated weight to 70 kg.
Message requesting to disconnect the patient.
2
Generate inhalation trigger.
The software switches to the running ventilation screen with unknown weight, in SIMV ventilation mode.
The software indicates that it is in backup ventilation.
3
Inspect the parameters on the screen.
The blower starts, with PS = DefaultPS, and presents the ventilation parameters.
4
Wait for two breathes.
The following ventilation parameters are automatically set according to the simulated patient weight:
Alert Pressure, Minute Volume M.V., BPM, Tidal volume Vt, Ti, Te, PS, trigger sensitivity.
5
Wait for a few minutes.
Ventilation continues with parameters suitable for a patient weighing approx. 70 kg.
The software keeps indicating that it is in backup ventilation mode.
The software instructs the user to perform patient circuit test (via the menu).
6
Stop ventilation. Ventilation stops.
Ventway Sparrow User Manual
Page 89 of 110
24.3. CPAP VENTILATION WITH BACKUP VENTILATION
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg.
Select 60 kg and then CPAP ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in CPAP mode.
CPAP - Apnea
3
Wait for the apnea timer to run out.
The software switches to SIMV mode, backup mode is indicated.
4
Wait for the next exhalation phase and for pressure to reach PEEP value.
Wait until Te time has passed.
The software switches to inhalation phase:
Inspiratory control valve opens and exhalation valve closes.
5
Allow for a few breaths cycles.
The software switches between phases, using SIMV ventilation mode.
6
Wait for the next exhalation phase and for pressure to reach PEEP value.
After more than MinRestDuration after PEEP reached, generate inhalation trigger.
Trigger is detected & inhalation phase starts (Inspiratory control valve opens and exhalation valve closes).
The software switches back to CPAP normal mode.
Ventway Sparrow User Manual
Page 90 of 110
24.4. PATIENT DISCONNECT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then CPAP ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in CPAP mode.
3
Decrease PEEP to PEEP -2.
The software alerts: “Patient disconnection”.
4
Silence the alert. Alert indication on the screen, but alert is silenced.
5
Select to view active alerts via the menu.
Patient Disconnection alert is shown.
Ventway Sparrow User Manual
Page 91 of 110
24.5. HIGH PEEP
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in SIMV mode.
3
Increase PEEP above the limit.
The software indicates to the user that PEEP is above the PEEP limit.
4
Silence the alert. Alert indication on the screen, but alert is silenced.
5
Select to view active alerts via the menu.
Alert is shown.
Ventway Sparrow User Manual
Page 92 of 110
24.6. VALVE BLOCKED
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Slowly decrease the expiratory flow.
The software alerts: “valve blocked”.
4
Increase flow back to normal.
The alert disappears.
5
Silence the alert. Alert indication on the screen, but alert is silenced.
6
Select to view active alerts via the menu.
Alert is shown.
Ventway Sparrow User Manual
Page 93 of 110
24.7. PRESSURE ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Slowly decrease the inspiration pressure under the min inspiration alert settings.
The software alerts: “Low Pressure”.
4
Increase pressure back to normal.
The alert disappears.
5
Slowly increase the inspiration pressure above the max inspiration alert settings.
The software alerts: “High Pressure”.
Caution: The interval from the moment that the airway pressure equals the high-pressure alarm limit, to the moment that the pressure starts to decline, must not exceed 80 ms.
6
Silence the alert. Alert indication on the screen, but alert is silenced.
7
Select to view active alerts via the menu.
Alert is shown.
Ventway Sparrow User Manual
Page 94 of 110
24.8. MINUTE VOLUME (MV) ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Slowly decrease MV under the min MV alert settings.
The software alerts: “MV Alert”.
4
Increase pressure back to normal.
The alert disappears.
5
Slowly increase MV above the max MV alert settings.
The software alerts: “MV Alert”.
6
Silence the alert. Alert indication on the screen, but alert is silenced.
7
Select to view active alerts via the menu.
Alert is shown.
Ventway Sparrow User Manual
Page 95 of 110
24.9. LEAK ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in SIMV mode.
3
Slowly increase a leak in the ventilation.
After a short pause, the software alerts: “Leak Alert”.
4
Close the leak. The alert disappears.
Ventway Sparrow User Manual
Page 96 of 110
24.10. TIDAL VOLUME ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in SIMV mode.
3
Slowly decrease the Tidal volume (TV) to 70%.
When TV is below 80%, the software alerts: “Low Tidal Volume Alert”.
4
Silence the alert. Alert indication on the screen, but alert is silenced.
5
Increase TV above 80%.
The alert disappears.
6
Change settings to disable TV.
TV is disabled.
7
Slowly decrease the Tidal volume (TV) to 50%.
Alarm is not set.
Ventway Sparrow User Manual
Page 97 of 110
24.11. I:E ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts.
The software switches to running ventilation, in SIMV mode.
3
Change the I:E ratio to be below 1.
The software alerts: “I:E Alert”.
4
Silence the alert. Alert indication on the screen, but alert is silenced.
5
Change the settings to disable I:E Alert.
I:E alert is off.
6
Change the I:E ratio to be below 1.
The software doesn’t issue an alert.
Ventway Sparrow User Manual
Page 98 of 110
24.12. APNEA ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then CPAP ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in CPAP mode.
3
Wait without generating inhalation triggers.
After the set time for Apnea alert, the software alerts: “Apnea Alert” and switches to backup ventilation mode.
4
Silence the alert. Alert indication on the screen, but alert is silenced.
5
Select to view active alerts via the menu.
Alert is shown.
Ventway Sparrow User Manual
Page 99 of 110
24.13. POWER ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in SIMV mode.
3
Disconnect the ventilator from the power.
The software indicate Power is disconnected.
4
Inspect the battery level.
Battery level is indicated to the user.
5
Reconnect the ventilator to power.
The software indicates Power is connected.
Ventway Sparrow User Manual
Page 100 of 110
24.14. LOW BATTERY ALERT
Step Procedure Action Expected Results
1
Start up the Sparrow ventilator.
Complete the patient circuit test.
The software switches to the ventilation phase, weight selection screen.
2
Set the simulator to the patient weight of 60 kg. Select 60 kg and then SIMV ventilation mode.
Mode is set. The blower starts. The software switches to running ventilation, in SIMV mode.
3
Wait until the battery level reaches CriticallyLowBattery level.
The software indicate CriticallyLowBattery level.
4
Silence the alert. Alert indication on the screen, but alert is silenced.
Loading...