7.6 Electrical data for outputs and inputs............................................................................ 48
7.7 RTD and T/C accuracy table ........................................................................................... 49
8 Appendix51
8.1 Installation and control drawings .................................................................................. 51
9 Notes56
4
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
IPAQ C330/R330
1.1 Intended use
The IPAQ 330 signal conditioner is a universal two-wire transmitter intended to be used in
industrial environments and designed for:
• Temperature measurements with resistance thermometers
• Temperature measurements with thermocouples
• Voltage measurements in a range up to 1000 mV
• Resistance measurement up to 10 kΩ
• Measurements with potentiometers
C330 / C330X is intended for installation in a DIN B-head or larger
according to EN 50446.
R330 / R330X is intended for installation on a 35 mm DIN-rail
according to EN 60715/DIN 50022.
The transmitters are configured from a PC by using the ConSoft program and a transmitter
configuration kit (USB connection), or by a smartphone with built-in NFC support.
NFC applies only to later versions of IPAQ 330, see table on page 10
SAFETY INSTRUCTIONS
1
DANGER!
You may only use transmitters labelled with the "Ex" symbol in potentially explosive areas or
connect them to a sensor located in those areas. Additionally always note the zone(s) for which
the devices have an approval. Otherwise the transmitters might cause an explosion that can
result in fatal injuries.
DANGER!
Responsibility for the correct use of the devices with special regard to suitability, intended use
and the field of application lies solely with the operator. To avoid any kind of incorrect use, also
note the information in the chapter "Device description".
DANGER!
The transmitters do not contain any serviceable parts inside. Any substitution of components
may impair the intrinsic safety of the versions with an Ex approval. Always send defective devices
to the manufacturer or the local distributor for repair or exchange. If this is the case, attach a
clear description of the malfunction for warranty claims.
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose. To avoid any kind of incorrect use, also note the information in the chapter
"Device description".
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
This device is a Group 1, Class A device as specified within CISPR11. It is intended for use in
industrial environment. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted as well as radiated disturbances.
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
5
1
SAFETY INSTRUCTIONS
1.2 Certifications
1.2.1 EU directive compliance
CE marking
The device fulfils all applicable statutory requirements of the following EU directives:
• EMC Directive 2014/30/EU, harmonized standards EN 61326-1 and EN 61326-2-3
• Devices for use in hazardous areas: ATEX Directive 2014/34/EU, harmonized standards
EN 60079-0 and EN 60079-11
• CE Directive 93/68/EC
• RoHS Directive 2011/65/EU, harmonized standard EN 50581.
• Radio Equipment Directive (RED) 2014/53/EU.
The manufacturer certifies successful testing of the product by applying the CE marking.
IPAQ C330/R330
1.2.2 Ex approvals
C330X (intrinsically safe), partno. 70C330X010
ATEXKIWA 16ATEX0038 XII 1 G Ex ia IIC T6...T4 Ga
IECExIECEx KIWA 16.0016XEx ia IIC T6...T4 Ga
USAFM17US0283XIS CL I Div 1 GP A-D, T4...T6
CanadaFM17CA0144X
C330X (intrinsically safe), partno. 70C330X012
ATEXKIWA 16ATEX0038 XII 1 G Ex ia IIC T6...T4 Ga
IECExIECEx KIWA 16.0016XEx ia IIC T6...T4 Ga
R330X (intrinsically safe), partno. 70R330X010
ATEXKIWA 16ATEX0040 XII 1 G Ex ia IIC T6...T4 Ga
IECExIECEx KIWA 16.0018XEx ia IIC T6...T4 Ga
USAFM17US0283XIS CL I Div 1 GP A-D, T4...T6
CanadaFM17CA0144X
R330X (intrinsically safe), partno. 70R330X012
ATEXKIWA 16ATEX0040 XII 1 G Ex ia IIC T6...T4 Ga
IECExIECEx KIWA 16.0018XEx ia IIC T6...T4 Ga
Cl I Zn 0 AEx/Ex ia IIC T4...T6 Ga
Cl I Zn 0 AEx/Ex ia IIC T4...T6 Ga
See also “Specific Conditions of Use" in the Ex certificates in the download area of the
manufacturer's website
6
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
IPAQ C330/R330
1.3 Safety instructions from the manufacturer
1.3.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
SAFETY INSTRUCTIONS
1
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.3.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
7
1
SAFETY INSTRUCTIONS
1.3.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation or operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.3.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of icons as shown below.
IPAQ C330/R330
8
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
IPAQ C330/R330
1.3.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This warning refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death.
SAFETY INSTRUCTIONS
1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.4 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
9
2
DEVICE DESCRIPTION
2.1 General description
The IPAQ 330 signal conditioner is a smart, digital 2-wire compatible transmitter for
temperature measurements and other measurement applications in an industrial environment.
For further information about the possible measurements refer to
Configuration of the transmitter is possible with:
• PC configuration software ConSoft with PC configuration kit ICON-X
• NFC interface in a portable device such as a smartphone. NFC applies only to later versions
of IPAQ 330, see table below . The smartphone app - INOR Connect, makes it possible to read,
write, save and share settings directly in the field.
The PC configuration software ConSoft is used for configuration, display and documentation.
The current ConSoft version is available for downloading on our website.
You can find configuration instructions in the ConSoft reference manual.
ConSoft is compatible with Windows XP/Vista/7/8/8.1/10.
Intended use
IPAQ C330/R330
on page 5.
Different versions of the IPAQ 330
IPAQPart NumberNFC
C330 / C330X70C3300010 / 70C330X010No
70C3300012 / 70C330X012Yes
R330 / R330X70R3300010 / 70R330X010No
70R3300012 / 70R330X012Yes
IPAQ C330 - In-head transmitter
The IPAQ C330 is a smart, digital 2-wire universal in-head transmitter for temperature
measurements and other measurement applications in an industrial environment. For further
information about the possible measurements refer to
The in-head transmitter is optionally available in an intrinsically safe version for installation in
potentially explosive areas. These devices wear the "Ex" symbol and have an approval for
mounting into classified hazardous area, Zone 0, 1 and 2 / DIV 1 and 2
All in-head versions are intended for installation in a DIN B-head or larger according to EN
50446 / DIN 43729. As an alternative you can also mount the in-head version on a 35 mm rail
according to EN 60715 / DIN 50022 with the help of the rail installation kit ( refer to
kit for in-head transmitters
on page 15).
Intended use
on page 5 .
Rail mounting
IPAQ R330 - Rail-mount transmitter
The IPAQ R330 rail-mount transmitter has the same features as the in-head version.
10
The rail-mount transmitter is optionally available in an intrinsically safe version for installation
into potentially explosive areas. All devices with an Ex approval wear the "Ex" symbol. The railmount transmitter is intended for installation on a 35 mm rail according to EN 60715 / DIN 50022
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
IPAQ C330/R330
2.2 Nameplate
Look at the device nameplate to ensure that the device is delivered according to your order.
The transmitter can be identified by the information on the nameplates. For the different
versions of IPAQ 330 refer to
Nameplate for in-head transmitter - Partno. 70C330X010
C330X
www.inor.com
Figure 2-1: Example for round nameplate
1 Product name
2 CE marking (EC conformity)
3 Ex-relevant electrical data and Ex approvals
4 Part number, serial number (yyww = year and week of manufacturing) and batch number
5 Printable field, sensor configuration
6 Connections
7 Manufacturer and address
0539 II 1G Ex ia IIC T6...T4 Ga
KIWA 16ATEX0038 X
IECEx KIWA 16.0016X
DEVICE DESCRIPTION
NFC configuration and logging
12
234
70C330X010
Nyyww.xxxxxx
PO zzzzzzzzz
1
T/C
RTD
PO Box 9125
SE-200 39 Malmö
+
2
on page 24
SWEDEN
Ui = Vmax: 30 V
Pi = Pmax: 0.9 W
Ii = Imax: 100 mA
Per Control Dwg
No. 4005488501
IS CL I DIV 1 GP ABCD T4...T6
FM17US0283X
FM17CA0144X
Cl I Zn 0 AEx/Ex ia IIC T4...T6 Ga
Figure 2-2: Example for bottom nameplate - Partno. 70C330X010
1 Electrical data for output
2 Ex-relevant electrical data
3 Ex approvals - North America
4 Electronic/electric device waste marking
5 Control drawing number
Nameplate for in-head transmitter - Partno. 70C330X012
12
234
C330X
www.inor.com
0539 II 1G Ex ia IIC T6...T4 Ga
KIWA 16ATEX0038 X
IECEx KIWA 16.0016X
70C330abcd
Nyyww.xxxxxx
PO zzzzzzzzz
Figure 2-3: Example for round nameplate
1 Product name
2 CE marking (EC conformity)
3 Ex-relevant electrical data and Ex approvals
4 Part number, serial number (yyww = year and week of manufacturing) and batch number
5 Printable field, sensor configuration
6 Connections
7 Manufacturer and address
1
T/C
RTD
PO Box 9125
SE-200 39 Malmö
+
SWEDEN
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
11
2
DEVICE DESCRIPTION
Ui = Vmax: 30 V
Pi = Pmax: 0.9 W
Ii = Imax: 100 mA
Figure 2-4: Example for bottom nameplate - Partno. 70C330X012
1 Electrical data for output
2 Additional certifications
Nameplate for rail-mount transmitter - Partno. 70R330X010
IPAQ C330/R330
R330X
8
70R330X010
Nyyww.xxxxxx
PO ZZZZZZZZZ
PO Box 9125
SE-200 39 Malmö
SWEDEN
KIWA 16ATEX0040 X
IECEx KIWA 16.0018X
0539 II 1 G Ex ia IIC T6...T4 Ga
FM17US0283X
Max. ambient 85 °C
Per Control Dwg
No. 4005499601
www.inor.com
Figure 2-5: Example for nameplate
1 Product name
2 Part number, serial number (yyww = year and week of manufacturing) and batch number
3 Manufacturer and address
4 Printable field, sensor configuration
5 Ex-relevant electrical data and Ex approvals
6 Electrical data for output and input
7 Control drawing number
8 CE marking (EC conformity) and electronic/electric device waste marking
FM17CA0144X
IS CL I DIV 1 GP ABCD T4...T6
Cl I Zn 0 AEx/Ex ia IIC T4...T6 Ga
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
IPAQ C330/R330
Nameplate for rail-mount transmitter - Partno. 70R330X012
DEVICE DESCRIPTION
2
R330X
Max. ambient 85 °C
8
www.inor.com
Figure 2-6: Example for nameplate
1 Product name
2 Part number, serial number (yyww = year and week of manufacturing) and batch number
3 Manufacturer and address
4 Printable field, sensor configuration
5 Ex-relevant electrical data and Ex approvals
6 Additional certifications.
7 Electrical data for output and input
8 Max. ambient temperature
9 CE marking (EC conformity) and electronic/electric device waste marking
2.3 Scope of delivery
70R330abcd
Nyyww.xxxxxx
PO zzzzzzzzz
KIWA 16ATEX0040 X
IECEx KIWA 16.0018X
0539 II 1 G Ex ia IIC T6...T4 Ga
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
Do a check of the packing list to make sure that you have all the elements given in the order.
Look at the device nameplate to ensure that the device is delivered according to your order.
The scope of delivery always consists of the transmitter and its documentation.
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
13
3
INSTALLATION
3.1 In-head transmitter
DANGER!
IPAQ C330X must be installed in an enclosure having an Ingress Protection suitable for the
actual use but at least IP20.
The transmitter is intended for installation in DIN B connection head or larger. For detailed
information refer to
Dimensions
IPAQ C330/R330
on page 45.
Figure 3-1: Connection head installation kit
1 M4 screw
2 Spring
3 Lock washer
4 Wires from the measuring insert.
5 MI cable
The connection head installation kit does not belong to the standard scope of delivery of the
transmitter, you have to order it separately.
DANGER!
The transmitter is optionally available in an intrinsically safe version (zone 0, 1 and 2) for
installation in potentially explosive atmospheres. The intrinsically safe version must be supplied
by an intrinsically safe power supply unit or Zener barrier placed outside of the potentially
explosive zone.
WARNING!
The transmitter has been developed for an operating temperature of -40...+85
°
C / -40...+185°F.
To avoid destruction or damage of the device, always assure that the operating temperature or
ambient temperature does not exceed the permissible range. The thermowell also transfer the
process temperature to the transmitter housing. If the process temperature is close to or
exceeds the maximum temperature of the transmitter, then the temperature in the transmitter
housing can rise above the maximum permissible temperature. One way to decrease the head
transfer via thermowell is to install the transmitter further away from the heat source. Inversely
similar measurements can be done if the temperature gets below specified minimum
temperature.
14
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
IPAQ C330/R330
3.2 Rail mounting kit for in-head transmitters
The rail mounting kit allows to install the in-head transmitter on a rail according to EN 60715 /
DIN 50022.
The kit does not belong to the standard scope of delivery. You have to order it separately.
The screws in the kit is not to be used with this transmitter.
Rail mounting kit for in-head transmitters
INSTALLATION
3
Figure 3-2: Rail mounting kit for in-head transmitters
1 Rail (not included in the kit)
2 Screws (not needed)
3 Clip
4 Transmitter
Installation procedure: Step 1
1 Place the transmitter on the clip.
2 Press down the transmitter until it is attached to the clip.
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
15
3
INSTALLATION
Installation procedure: Step 2
1 Hook one end of the clip into the rail.
2 Push the other end down until it snaps onto the rail.
3 Release by pushing the hook, and at the same time lift the clip out of the rail.
3.3 Rail-mount transmitter
DANGER!
IPAQ R330X must be installed in an enclosure having an ingress protection suitable for the
actual use but at least IP20.
IPAQ C330/R330
These transmitters are intended for installation on a 35 mm rail according to EN 60715 /
DIN 50022.
Figure 3-3: Rail installation
1 Fix the upper part of the transmitter onto the rail.
2 Press the lower part of the transmitter against the rail.
3 To remove the transmitter, bend the locking device using a small screwdriver. Carefully pull
the transmitter in the forward direction.
CAUTION!
The manufacturer has developed the R330 for an operating temperature range of
-40...+85
°
C / -40...+185°F.
To avoid destruction or damage of the device, always note the following items:
•
Assure that the operating temperature or the ambient temperature does not exceed the
permissible range.
16
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
IPAQ C330/R330
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
The transmitter is protected against polarity reversal. No damage will occur to the device if the
polarity of the supply voltage is switched. The output will then indicate 0 mA.
DANGER!
In potentially explosive areas, the intrinsically safe versions must be used.
The C330X / R330X (intrinsically safe) can be installed in potentially hazardous area zone 0, 1 and
2.
The C330X / R330X (intrinsically safe)can be installed in potentially hazardous area zone 1 and 2
and the input may be connected to temperature sensors installed in zone 0.
The intrinsically safe versions must be supplied by an intrinsically safe power supply unit or
Zener barrier placed outside of the potentially explosive area.
ELECTRICAL CONNECTIONS
4
DANGER!
Never connect or operate a non-Ex version of a transmitter in potentially explosive areas,
otherwise it might cause an explosion that can result in fatal injuries! Before you connect and
operate a transmitter version with an Ex approval, always note the following items to avoid an
explosion which may result in fatal injuries
•
Connect the Ex version only to Ex approved sensors or sensors that meet the requirements
for "simple apparatus" in EN 60079-11.
•
Observe the corresponding regulations, the declaration of conformity, the Ex type test
certificate of the device and the relevant instructions of this document.
•
The intrinsically safe versions must be supplied by an intrinsically safe power supply unit or a
Zener barrier placed outside of the potentially explosive area.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate (applies only to the intrinsically
safe versions).
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
17
4
ELECTRICAL CONNECTIONS
DANGER!
For the transmitters used in hazardous areas following special conditions have to be considered:
The communication interface (USB connection) may only be connected to the associated ICON
interface if the temperature transmitter is outside of the hazardous area. If certified ICON-X
interface is used, a connected sensor may be located in the hazardous area. If non-Ex ICON
interface is used, a connected sensor shall not be located in the hazardous area.
For the permitted ambient temperature range, refer to Temperature data for areas with
potentially explosive atmospheres on page 46
The transmitter shall be mounted into a suitable enclosure that provides a degree of protection
of at least IP20.
The NFC Interface for communication must not be used in hazardous areas.
IPAQ C330/R330
18
www.inor.com01/2019 - 4007317401 – MA IPAQ 330 R01 en
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.