IPAQ-L
T
1
T
2
6
+ -
5
mV
RTD
(Pt100)
T/C
RTD
100%
0%
RTD
RTD
(Pt100)
100%
0%
SPECIALISTS IN INDUSTRIAL TEMPERATURE MEASUREMENT
Universal
Programmable
Input connections
RTD
4-wire connection
2-wire Transmitters
IPAQ-L/-LX are universal, isolated 2-wire transmitters for temperature
and other measurement applications. They combine competitive pricing,
functionality and simple conguration.
Useful error correction functions improve the accuracy.
• Fully universal, linearized and isolated
• Accepts RTD, T/C, mV and Ω
• Sensor error and system (sensor/transmitter) error correction for
highest total accuracy
• Full access to all features while in operation
• NAMUR compliant
• Consistent sensor break function
• Simplied loop check-up with calibration output
• Test output without breaking the loop (IPAQ-L)
• IPRO, easy-to-use Windows conguration software
Specications:
Input RTD 3-, 4-wire connection
Pt100 (α=0.00385) -200 to +1000 °C / -328 to +1832 °F
Pt1000 (α=0.00385) -200 to +200 °C / -328 to +392 °F
PtX 10 ≤ X ≤ 1000 (α=0.00385) Upper range depending on X-value
Pt100 (α=0.003902) -200 to +1000 °C / -328 to +1832 °F
Pt100 (α=0.003916) -200 to +1000 °C / -328 to +1832 °F
1)
Ni100
Ni1000
Cu10
2)
, Ni120
-60 to +250 °C / -76 to +482 °F
1)
-100 to +150 °C / -148 to +302 °F
3)
-200 to +260 °C / -328 to +500 °F
Input Potentiometer/resistance 3-, 4-wire connection. 0 to 2000 Ω
Input Thermocouples Types B, C, E, J, K, L, N, R, S, T, U
Input mV -10 to +500 mV
Sensor failure User denable output
Adjustments-Zero Any value within range limits
Adjustments-Minimum spans
Pt100, Pt1000, Ni100, Ni1000 10 °C / 18 °F
Potentiometer 10 Ω
T/C, mV 2 mV
Output 4-20 / 20-4 mA, temperature linear
Operating temperature -20 to +70 °C / -4 to +158 °F
Galvanic isolation 1500 VAC, 1 min
Power supply IPAQ-L 7.5 to 36 VDC
IPAQ-LX 8 to 30 VDC
Intrinsic safety (Mounting in safe area)
IPAQ-LX ATEX: II (1) G [EEx ia] IIC
IPAQ-LX FM: IS Class I-III, DIV 1, GP A-G
IPAQ-LX CSA:
Class III
Typical accuracy ±0.1 % of span
Mounting Rail acc. to DIN EN50022, 35 mm
1)
DIN 43760
2)
Edison No. 7
3)
Edison No. 15
Class I, Groups A-D; Class II, Groups E-G;
RTD
3-wire connection
Pt100
Diff temperature T1 > T2
Potentiometer
3-wire connection
Potentiometer
4-wire connection
Voltage
mV
Thermocouple
Output connections
1)
R
LOAD
I
OUT
Test
(IPAQ-L)
1)
Ri ≤ 10 Ω
Output load diagram
R
(Ω)
LOAD
Supply voltage U (V DC)
R
=(U-7.5)/0.022 (IPAQ-L)
LOAD
R
=(U-8)/0.022 (IPAQ-LX)
LOAD
Dimensions
mm/inches
Ordering information
IPAQ-L 70IPL00001
IPAQ-LX (ATEX) 70IPLX0001
IPAQ-LX (FM, CSA) 70IPLX1001
PC Conguration Kit 70CFG00092
Conguration 70CAL00001