Universal PC-Programmable 2-wire transmitter
IPAQ C330 is a universal, isolated, temperature transmitter with additional voltage and resistance input. Its robust design and high quality gives excellent performance and accuracy also under harsh conditions.
IPAQ C330 supports communication via NFC® (Near-field communication) and Bluetooth® which makes it possible to configure and monitor the transmitter remotely.
• High accuracy and long term stability
• 50-point Customized Linearization and Callendar-Van Dusen
• Accepts RTD, T/C, mV and Ω
• Sensor error and system (sensor/transmitter) error correction for highest total accuracy
• Low temperature drift
• Configuration via USB or NFC without external power
• Runtime counter - hour counter for elapsed operational time
• Rugged design tested for 10 g vibrations
• High security - Password protection and date of changes logged
Specifications:
Input RTD 2-, 3-, 4-wire connection Pt100 (α =0.00385) -200 to +850 °C / -328 to +1562 °F PtX 10 X 1000 (α =0.00385) Upper range depending on X-value Pt100 (α =0.003916) -200 to +850 °C / -328 to +1562 °F Ni1001), Ni1202) -60 to +250 °C / -76 to +482 °F Ni10001) -50 to +180 °C / -58 to +356 °F Cu103) -50 to +200 °C / -58 to +392 °F
Input Resistance / potentiometer 0 to 10000 / 100 to 10000 Input Thermocouples Types B, C, D, E, J, K, N, R, S, T Input mV -10 to +1000 mV Sensor failure Upscale (≥21.0 mA) or downscale (≤3.6 mA) action Adjustments - Zero Any value within range limits Adjustments - Minimum spans
Pt100, Pt1000, Ni100, Ni1000 10 °C / 18 °F Potentiometer 10 T/C, mV 2 mV
Output 4-20 / 20-4 mA, temperature linear Operating temperature -40 to +85 °C / -40 to +185 °F Galvanic isolation 1500 VAC, 1 min Power supply IPAQ C330 8.0...36.0 VDC
IPAQ C330X 8.0...30.0 VDC Intrinsic safety IPAQ C330X ATEX: IPAQ C330X IECEx: Ex ia IIC T6…T4 Ga4) IPAQ C330X cFMus: IS CL I Div 1 GP A-D, T6...T4 Cl I Zn 0 AEx/Ex ia IIC T6...T4 Ga4)
Typical accuracy ±0.08°C or ±0.08% of span Connection head DIN B or larger
1)
DIN 43760,
II 1 G Ex ia IIC T6…T4 Ga4)
2)
Edison No.7,
3)
Edison No.15
4)
For Tambient, see the manual
Input connections
See data sheet for more alternatives
RTD
2-wire connection
RTD
3-wire connection
RTD
4-wire connection
Potentiometer
0%
100%
100%0%
T/C
mV
3-wire connection
Resistance
3-wire connection
Thermocouple
Voltage
mV
Output connections
R
I
OUT
LOAD
Output load diagram
R
()
LOAD
C330X
Supply voltage U (V DC)
R
=(U-8)/0.022
LOAD
C330
Dimensions
33 / 1.3
mm/inches
7 / 0.28
44.5 / 1.75
19.5 / 0.77
Ordering information
IPAQ C330 70C3300012 IPAQ C330X (ATEX / IECEx) 70C330X012 IPAQ C330X (cFMus) 70C330F012 PC configuration kit (USB-conn.) 70CFGUSX01 ICON-BT Bluetooth kit 70CFGBT001 Head mounting kit 70ADA00017 Rail mounting kit 70ADA00015