InMode BodyTite Operator's Manual

BodyTite
InMode RF
Version: DO607095B
Operator Manual: BodyTite™ System DO607095B
Copyright © InMode MD Ltd. Date: October 2017
All rights reserved. Contents of this publication may not be reproduced in any form without
the written permission of InMode MD Ltd.
🏭 InMode MD Ltd.
Tabor House, Industrial Park South POB 44, Yokneam 2069201, Israel For information call: +972-4-9096317
Contents
Section 1: Introduction ........................................................................................... 1
1.1 Before You Start ........................................................................................................... 1
1.2 System Overview .......................................................................................................... 1
1.3 Conventions used in the Manual ................................................................................. 1
1.4 Explanation of the Symbols Used on the System......................................................... 1
Section 2: Safety ..................................................................................................... 3
2.1 The Patient ................................................................................................................... 3
2.2 Treating Attendant ....................................................................................................... 4
2.3 Cautions ....................................................................................................................... 4
2.4 Electrical and Mechanical Safety ................................................................................. 4
2.5 Fire Hazards .................................................................................................................. 5
2.6 Safety Features of the System ..................................................................................... 5
2.7 Safe Use of the Active Accessories............................................................................... 6
2.8 Warnings ...................................................................................................................... 6
2.9 Device and Handpiece Labels ....................................................................................... 8
2.10 Equipment Classification .............................................................................................. 9
Section 3: System Installation ............................................................................... 10
3.1 Electrical Requirements ............................................................................................. 10
3.2 Environmental Requirements .................................................................................... 10
3.3 Equipment List ........................................................................................................... 10
3.4 Unpacking................................................................................................................... 11
3.5 Installation ................................................................................................................. 12
3.6 Moving the System: ................................................................................................... 12
3.7 System Disposal ......................................................................................................... 13
Section 4: Device Description ................................................................................ 14
4.1 Rear Panel .................................................................................................................. 14
4.2 Front Panel and Operator Control Panel ................................................................... 15
4.3 Software Screens........................................................................................................ 15
4.3.1 Sound Indicator ......................................................................................................... 18
4.3.2 External Cut-Off Temperature Control ...................................................................... 18
4.3.3 Internal Cut-Off Temperature Control ...................................................................... 18
4.3.4 Treatment Time Control ............................................................................................ 18
4.4 Handpieces ................................................................................................................. 18
BodyTite Operator Manual
Section 1: Introduction
Section 5: System Operation ................................................................................. 20
5.1 Device Start-Up .......................................................................................................... 20
5.2 System Shutdown ...................................................................................................... 20
Section 6: Treatment Information ......................................................................... 21
6.1 Training Requirements ............................................................................................... 21
6.2 Indications for Use ..................................................................................................... 21
6.3 Contraindications ....................................................................................................... 21
6.4 Possible Adverse Effects............................................................................................. 22
6.4.1 Gamma sterilized - Instruction Prior to Use .............................................................. 22
6.5 Pre-Treatment Recommendations ............................................................................ 23
6.6 Treatment Recommendations ................................................................................... 23
6.7 Post-Treatment Recommendations ........................................................................... 23
System Maintenance ............................................................................................ 24
7.1 Cleaning the Device .................................................................................................... 24
Troubleshooting ................................................................................................... 25
8.1 Description of Faults .................................................................................................. 25
System Specifications............................................................................................ 27
9.1 BodyTite Output Power Curves .................................................................................. 28
9.2 EMC Safety for the BodyTite Device .......................................................................... 29
BodyTite Operator Manual
Section 1: Introduction
1 of 30
Section 1: Introduction
This chapter provides a system overview and conventions used in this manual.
1.1 Before You Start
The manual and the equipment it describes are for use only by qualified medical
professionals trained in the particular technique to be performed.
Federal (USA) law restricts sale of this device to, or on the order of a physician. Read this manual to become familiar with all safety requirements and operating
procedures before attempting to operate the System.
1.2 System Overview
The BodyTite device is based on the InMode RF™ platform. BodyTite employs radiofrequency (RF) energy for use in dermatological and general surgical procedures for electrocoagulation and hemostasis. The System operates when the Handpiece is connected.
The System provides individual adjustment of treatment parameters to achieve
maximum efficiency and safety for the specific treatment.
The System provides enhanced safety while minimizing possible side effects by
monitoring RF parameters and tissue temperature.
1.3 Conventions used in the Manual
The following conventions in the form of notes and warnings are used in this manual:
WARNING! This information is extremely important!
ATTENTION! Consult Accompanying Document.
Provides general information that is important to keep in mind.
1.4 Explanation of the Symbols Used on the System
BodyTite Operator Manual
Section 1: Introduction
2 of 30
Symbol
Description
CSA marking (212603 CSA master contract number)
Do not discard in trash. Electronic equipment should be disposed of in an appropriate manner
Fuse
Type BF Equipment
HF Isolated Patient Circuit
This equipment intentionally supplies non-ionizing RF energy
Follow operating instructions
Table 1-1: Device Symbols
BodyTite Operator Manual
Section 2: Safety
3 of 30
Section 2: Safety
This chapter describes safety issues regarding the use and maintenance of the BodyTite System, with a special emphasis on electrical safety.
The System is designed for safe and reliable treatment when used in accordance with proper operation and maintenance procedures. Only trained, qualified practitioners can use the System. The operator and all other personnel operating or maintaining the System should be familiar with the safety information provided in this Section.
The primary consideration should be to maximize safety for both treating attendant and patient.
Read this chapter to be familiar with all safety requirements and operating
procedures prior to System operation.
The RF energy can cause injury if used improperly. High voltage is present inside the System. Always be aware of the possible dangers and take proper safe-guards as
described in the manual.
2.1 The Patient
Well-trained staff is a key for assuring patient safety. A patient history should be
completed prior to scheduling. Patients should be fully informed of the treatment details, the likely results and any risks associated with the treatment.
Metal jewelry and metal accessories that are within the activation range of the
Handpiece should be removed to avoid accidental RF conduction. The metal item(s) will be removed prior to use of the equipment.
Patients will not be in contact with any metal or other alternate pathway to ground
whilst the System is in use.
BodyTite Operator Manual
Section 2: Safety
4 of 30
2.2 Treating Attendant
Only authorized individuals with appropriate training and knowledge should operate,
assist in the operation of, or provide maintenance to the BodyTite System.
Personnel should not operate the System until they have been fully educated in its
use. Make sure that all treatment personnel are familiar with the System controls and know how to shut down the System instantly.
There are no user-serviceable parts in the system, and all service and repair must be
performed only by the factory or authorized field service technicians.
2.3 Cautions
The following cautions should be heeded for safe System use:
Do not touch the System’s inner parts. Service is supplied by company-authorized personnel only. To avoid damage, do not allow the Handpiece to come in contact with hard
materials.
2.4 Electrical and Mechanical Safety
Keep all covers and panels of the System closed. Removing the covers creates a
safety hazard.
Keep hands away from the applicator during the System startup. Perform maintenance procedures when the System is shut down and disconnected
from the power.
The System is grounded through the grounding conductor in the power cable. This
protective grounding is essential for safe operation.
Move the System slowly and carefully. The System weighs approximately 20kg (44lb.)
and may cause injury if proper care is not used when moving it.
Provide as much distance as possible between the system, RF Handpiece and other
electronic equipment as the activated RF generator may cause interference between them.
BodyTite Operator Manual
Section 2: Safety
5 of 30
2.5 Fire Hazards
Do not use the System in the presence of explosive or flammable materials. Materials conducting RF energy may cause temperature rise of the absorbing
material. Do not use the System in the presence of explosive or flammable materials conductive to RF.
Keep drapes and towels moist to prevent them from igniting and burning. Use non-
flammable prepping solutions.
Do not use flammable substances when preparing the skin for treatment. Be
especially careful with the use of oxygen.
If alcohol is used for cleaning and disinfecting, allow it to dry thoroughly before the
System is used.
2.6 Safety Features of the System
The System incorporates the following safety features. All personnel operating the System should be familiar with these features.
System has unique password to avoid device operation by non-authorized
personnel.
The RF energy cannot be activated unless the applicator and footswitch have been
connected to the System.
An audible tone indicates energy activation. During activation, the System performs a self-test of the hardware. Hardware is tested every 10ms to ensure proper operation of electrical circuit. Tissue impedance monitoring prevents accidental energy emission to the patient. Skin surface is monitored during the treatment. RF energy delivery is terminated
when skin temperature accidentally reaches the Cut-off level.
System starts at a low setting. Internal and skin surface temperature and impedance are constantly monitored.
BodyTite Operator Manual
Section 2: Safety
6 of 30
2.7 Safe Use of the Active Accessories
Examine the connection of the Handpiece through the connector to the System
before using. Improper connection may result in arcs and sparks, accessory malfunction, or unintended treatment effects.
Do not wrap the Handpiece cords around metal objects. It may induce current that
could lead to electrical shock, fire or injury to the patient or personnel.
Ensure that return electrode is in full contact with the skin. Bad coupling of the
return electrode with the skin results in a specific warning sound, a message on the screen, and disabling of RF.
Handpieces are for single use only. Do not try to reuse the handpiece. Re-use of the
handpiece may create loss of integrity of handpiece components that may affect the performance and make the device non-functional.
Do not connect a wet accessory to the System. Do not immerse the applicator under water at any time. The Handpiece is gamma-sterilized for a single use only and CANNOT be
autoclaved or re-sterilized by any other technology.
2.8 Warnings
This equipment is for use only by trained, licensed physicians. Only Handpieces manufactured or approved by InMode MD Ltd. should be
used with the BodyTite System.
Connect the power cord to a properly polarized and grounded power source
with the frequency and voltage characteristics that match those listed on the back of the unit.
Connect the System power cord to a properly grounded receptacle. Do not use
power plug adapters.
Always turn off and unplug the device before cleaning. The patient should not come into contact with metal parts which are earthed
or which have an appreciable capacitance to earth. The use of antistatic sheeting is recommended for this purpose.
BodyTite Operator Manual
Section 2: Safety
7 of 30
Use the lowest output setting necessary to achieve the desired treatment
effect. The higher RF energy is applied, the greater the possibility of unintended thermal damage.
When BodyTite and physiological monitoring EQUIPMENT are used
simultaneously on the same patient; any monitoring electrodes should be placed as far as possible from the handpiece electrodes. Needle monitoring electrodes are not recommended.
Failure of the equipment could result in an unintended increase of output
power.
The cables of the Handpiece should be positioned in such a way that contact
with the patient or other leads is avoided.
Fire / Explosion Hazard - The following substances will contribute to increased
fire and explosion hazards in the operating room:
Flammable substances (such as alcohol based skin prepping agents and
tinctures).
Naturally-occurring flammable gases which may accumulate in body cavities
such as the bowel.
Oxygen enriched atmospheres. Oxidizing agents (such as nitrous oxide [N2O] atmospheres). Endogenous gases. The RF energy and heating associated with the System can provide an
ignition source. Observe fire precautions at all times. When using BodyTite in the same room with any of these substances or gases, prevent their accumulation or pooling within the area where BodyTite procedures are performed.
The operation of BodyTite may adversely influence the operation of other
electronic EQUIPMENT.
To avoid the RISK of electric shock, this equipment must only be connected to a
SUPPLY MAINS with protective earth.
Do not use BodyTite on patients with pacemakers or internal defibrillators.
BodyTite Operator Manual
Section 2: Safety
8 of 30
2.9 Device and Handpiece Labels
As required by national and international regulatory agencies, appropriate Warning and information labels have been attached in specific locations on the instrument as identified below.
The following device labels are located on BodyTite device console and the handpieces:
Figure 2-1:System Certification and Identification Label
Figure 2-2: 2 Footswitch label for all applications
BodyTite Operator Manual
Section 2: Safety
9 of 30
Figure 2-3: BodyTite Handpiece Labels
2.10 Equipment Classification
The following is a list of the different equipment used and their classifications. Electric shock protection: Class I, Defibrillation-proof Type BF. Protection against ingress of liquids: Ordinary equipment. Not suitable for use in presence of flammable substance. Power receptacle must include protective earth, and must be checked before
connecting the system.
System is classified as a IIb device defined by the Medical Device Directive
(93/42/EEC).
BodyTite Operator Manual
Section 3: System Installation
10 of 30
Section 3: System Installation
3.1 Electrical Requirements
The System will require a separate line supply of single phase (100Vac; 15A) or
(115Vac; 15A) or (230Vac; 15A) or (240Vac; 15A) 50-60Hz.
Power receptacles must be within 15 feet of the System site. The System should not share a power line with other equipment. Power receptacle must include protective earth, and must be checked before
connecting the system.
For continued protection against fire, replace the fuse on-ly with one of the
same type and rating.
Proper grounding is essential for safe operation.
3.2 Environmental Requirements
Corrosive materials can damage electronic parts; therefore, the System should
operate in a non-corrosive atmosphere.
For optimal operation of the System, maintain room temperature between 20-27
o
C
(68-79oF) and relative humidity of less than 80%.
3.3 Equipment List
The System includes the following:
System platform Footswitch Operator manual Power cord Handpieces
BodyTite Operator Manual
Section 3: System Installation
11 of 30
3.4 Unpacking
In order to unpack the device:
1. Remove the paper strip and open the box.
2. Remove accessories and foams around the device.
3. Take the device out of the box using top and bottom handles.
BodyTite Operator Manual
Section 3: System Installation
12 of 30
3.5 Installation
The System is designed for installation in a clinic environment. To install the System, perform the following tasks:
Check the System and all its components for damage. Connect the Footswitch. Connect the power cord to the System inlet. Plug the System power cord into an appropriate electrical outlet.
3.6 Moving the System:
Turn the System off. Disconnect the Power Cord. Disconnect the Footswitch. Release the wheel brakes. Slowly push or pull the System using the handle. When moving to another facility, lift the System to the vehicle and lay it carefully on
its side.
Never lift, pull or push the System using the operating panel. Always use the handles when moving the System. Upon unpacking check the System for mechanical damage (e.g., cracks in the
cable insulation).
Never lift, pull or push the System using the operating panel. Always use the handles when moving the System. Upon unpacking check the System for mechanical damage (e.g., cracks in the
cable insulation).
BodyTite Operator Manual
Section 3: System Installation
13 of 30
3.7 System Disposal
To comply with European Commission Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) and other country and state regulations, please DO NOT dispose of this equipment in any location other than designated locations.
BodyTite Operator Manual
Section 4: Device Description
14 of 30
Section 4: Device Description
4.1 Rear Panel
Power cord inlet
100-240V~, 1A, 50-60Hz.
Fuse holder
Rating is T 2A, 250V. Replace fuses if it is needed only with fuses having exactly the same rating.
Software flash memory plug
Software plug is a flash memory with the machine software. The software plug should be screwed to the connectors.
Foot switch connector
Foot switch is connected to the inlet. Foot switch activates RF energy if the system is in Ready mode. Place the foot switch on the floor near the treatment area.
Table 4-1: Rear Panel Symbols
Figure 4-1: Handpiece Connector (Arrow) on Front Panel
!!
BodyTite Operator Manual
Section 4: Device Description
15 of 30
4.2 Front Panel and Operator Control Panel
The Operator Control Panel is located on the upper side of the System. The Operator Control Panel consists of an LCD screen with touch panel.
On the front panel there is a black On/Off switch on the left, and handpiece connector on the right.
Power On-Off switch
Power switch turns power electronics On and Off.
LCD Screen
LCD Screen shows information about system mode and
treatment parameters.
The panel allows changing treatment parameters and
system mode.
Power electronics is not activated if the Handpiece is not connected to its connector on the front panel.
4.3 Software Screens
The Splash Screen appears after the On-Off switch is turned on.
Figure 4-2: 2 Splash Screen
*The Software version number is displayed on the top of the screen and depends on the product version.
BodyTite Operator Manual
Section 4: Device Description
16 of 30
After entering the individual code in the Login Screen, the system allows access to the Treatment Screen.
Default Login code 1234 can be changed in the Utilities Screen. The Volume level can also be reduced in the Utilities Screen.
Figure 4-3: Login Screen
Prior to entering the Treatment Screen, a self-test of the system module is performed. After the end of the self-test, the Treatment Screen appears.
Figure 4-4: Treatment Screen
BodyTite Operator Manual
Section 4: Device Description
17 of 30
Item
Function
Selection Frame
The frame selects parameters that can be changed.
External Cut-Off
This indicator shows the skin Cut-Off temperature, which is
adjustable between 35-42oC, with increments of 1oC in the tissue. The selected value indicates measured temperature which is maintained during the treatment according to the setting.
Internal Cut-Off
This indicator shows the internal Cut-Off temperature, which
is adjustable between 50-70oC, with increments of 1oC in the tissue. The selected value indicates measured temperature which is maintained during the treatment according to the setting.
External Temp.
Internal Temp.
This indicator shows measured external temperature. This indicator shows measured internal temperature.
Treatment Time
This indicates the time that Cut-Off temperature is
maintained. It varies from 15-120 sec with increments of 5 sec.
Elapsed Time
This indicator shows elapsed treatment time.
System Mode
The System has three treatment modes: Standby, Ready and
Active.
Standby mode allows the user to set treatment parameters.
Activation of energy is unavailable in Standby mode.
In Ready mode, the system waits for a signal from the
footswitch to activate the energy. Any change to the treatment settings switches the system to Standby mode.
Active mode is entered during RF energy delivery.
BodyTite Operator Manual
Section 4: Device Description
18 of 30
4.3.1 Sound Indicator
Periodic beeping signal is emitted when RF energy is delivered. Warning sound tone indicates Bad Coupling. When the measured temperature approaches the Cut-off temperature, the tone
beeps double in speed. It becomes faster when the Cut-off temperature is reached.
4.3.2 External Cut-Off Temperature Control
The selected Cut-Off temperature in the range of 35-42oC is constantly maintained and when reached, RF delivery is adjusted automatically to maintain required temperature. Temperature is monitored by temperature sensors in the Handpiece (in the external electrodes).
4.3.3 Internal Cut-Off Temperature Control
The selected Cut-Off temperature in the range of 50-70oC is constantly maintained and when reached, RF delivery is adjusted automatically to maintain required temperature. Temperature is monitored by temperature sensors in the Handpiece (in the internal electrodes).
4.3.4 Treatment Time Control
The treatment time is selected in the range of 15-120sec and indicates time that Cut-Off temperature is maintained.
4.4 Handpieces
Handpiece comprises internal active electrode, external return electrode, both with temperature sensor, handle, cable and connector.
There are three types of handpieces, both with external and internal electrodes.
4. Handpiece HP101306A has an external electrode diameter of 13.5mm and internal
electrode of 1.2mm diameter and 10cm long. Maximal RF power 20W
5. Handpiece HP172206A has an external electrode diameter of 13.5mm and internal
electrode of 2.2mm diameter and 17cm long. Maximal RF power 20W
6. Handpiece HP172246A has an external electrode diameter of 22mm and internal
electrode of 2.2mm diameter and 17cm long. Maximal RF power 40W
BodyTite Operator Manual
Section 4: Device Description
19 of 30
Figure 4-5: Handpiece HP101306A
Figure 4-6: Handpiece HP172206A
Figure 4-7: Handpiece HP172246A
Internal Electrode
The Handpiece active internal electrode is an insulated metal
tube with conductive area at the distal part, but not at the tip. For added safety, at the end of the electrode there is an isolated blunt plastic tip which is not conductive.
The internal electrode has an embedded temperature sensor
to monitor the tissue temperature.
Return Electrode
The handpiece has a large area return electrode with
embedded temperature sensor to monitor the skin temperature.
Cable
Has a length of 250cm (100”).
Connector
It is connected to the front control panel of the system.
BodyTite Operator Manual
Section 5: System Operation
20 of 30
Section 5: System Operation
This section of the manual explains how to start the device, operate it and turn it off.
Prior to using or connecting the Handpiece, inspect the System and Handpiece
for possible mechanical damage.
5.1 Device Start-Up
1. Connect the Handpiece to the Handpiece connector socket on the front panel.
2. Turn on Main Power switch at the rear panel.
3. Press the On-Off button on the control panel to turn the device on. The System
loads the software and enters the Login Screen.
4. Enter password to get access to the treatment screen.
5. The system loads the software and enters a self-test mode. If any problem is
detected during the test, an error message will appear (See Troubleshooting Section in this manual). If the test is passed correctly then the system automatically enters the Treatment Screen.
6. Select treatment parameters using touch screen.
7. Enter Ready mode to confirm the selected parameters. The System is ready for RF
delivery.
8. Press footswitch to start procedure.
5.2 System Shutdown
To shut down the System turn the On-Off switch on the front panel off. Turn the Main Power switch off at the end of the day.
BodyTite Operator Manual
Section 6: Treatment Information
21 of 30
Section 6: Treatment Information
6.1 Training Requirements
This manual is not intended to be a complete clinical guide to the use of the System. All users should be trained prior to operating the BodyTite. Updated treatment protocols and recommendations are presented in the Quick Reference Guide (QRG).
6.2 Indications for Use
The BodyTite system is intended for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
6.3 Contraindications
DO NOT USE in patients who have electronic implants such as cardiac pacemakers or
internal defibrillators without first consulting a qualified professional (e.g., cardiologist). A possible hazard exists because interference with the action of the electronic implant may occur, or the implant may be damaged.
The Handpiece should be used at least 1cm away from cochlear implants in the ear. Permanent implant in the treated area such as metal plates and screws, silicone
implants or an injected chemical substance, unless deep enough in the periosteal plane.
Current or history of skin cancer, or current condition of any other type of cancer, or
pre-malignant moles.
Severe concurrent conditions, such as cardiac disorders, sensory disturbances,
epilepsy, uncontrolled hypertension, and liver or kidney diseases.
Pregnancy and nursing. Impaired immune system due to immunosuppressive diseases such as AIDS and HIV,
or use of immunosuppressive medications.
Patients with history of diseases stimulated by heat, such as recurrent Herpes
Simplex in the treatment area, may be treated only following a prophylactic regimen.
Poorly controlled endocrine disorders, such as diabetes or thyroid dysfunction and
hormonal virilization.
BodyTite Operator Manual
Section 6: Treatment Information
22 of 30
History of skin disorders, keloids, abnormal wound healing. History of bleeding coagulopathies. Any surgical procedure in the treatment area within the last 3 months or before
complete healing.
Any therapies or medications which may interfere with treatment. As per the practitioner's discretion, refrain from treating any condition which might
make it unsafe for the patient.
6.4 Possible Adverse Effects
The patient must understand the importance of pre-treatment and post-treatment instructions and failure to comply with these instructions may increase the probability of complications.
Possible adverse effects include, but are not limited by: discomfort or pain, excessive skin redness (erythema) and/or swelling (edema), ecchymosis, burns, damage to natural skin texture (crust, blister, and burn), change of pigmentation (hyper- and hypo­pigmentation), scarring, temporary injury to nerves (neuropraxia) where nerve branches are superficial, and very slight risk of infection.
Make sure that Handpiece is dry and intact before use. Do not use the Handpiece if it is damaged.
6.4.1 Gamma sterilized - Instruction Prior to Use
The Gamma sterilized Handpieces are sent sterile by gamma-rays in double sterilization bags. Therefore, the gamma sterilized handpieces SHOULD NOT be autoclaved, as the thermal fuse of the hand piece is destroyed and the handpieces are no longer functional.
There are a few signs to recognize the gamma-sterilized handpieces:
1. There is a white sticker on the large box, stating "Gamma Sterilized" and on it is a
small red sticker. This sticker is not on the smaller individual boxes.
2. The hand pieces are packed in double sterilization bags inside a small cardboard
box, two in each box.
3. There is a silver sticker on each large and small box carrying the sign “Sterile R”
that stands for Sterilization by Radiation.
4. The color of the handpiece is green (unlike the white color of the autoclavable
handpiece).
BodyTite Operator Manual
Section 6: Treatment Information
23 of 30
5. The internal electrode is covered by a transparent sleeve (unlike a white sleeve of
the autoclavable handpiece).
6.5 Pre-Treatment Recommendations
Make sure that the Handpiece is intact and sterile.
There are a few signs to recognize the gamma-sterilized hand pieces as
specified in items 1-5 above.
During the patient’s first visit treating attendant should:
Exclude from treatment anyone who may be affected by the listed contraindications. Instruct the patient about the safety warnings. Have the patient sign an informed consent form. Make sure that the hand-pieces are intact, sterile and dry.
6.6 Treatment Recommendations
The device can be used in a certified operating room or in a clean procedure room in
an office setting, using a sterile technique.
Apply sterile water–based gel, such as ultrasound gel to the skin surface to ensure
good contact of external electrode.
Select treatment parameters. Insert cannula to the intended tissue and ensure good contact of return electrode
with skin surface
Enter Ready mode, press the footswitch and apply the RF energy. Release the footswitch after treatment time is elapsed. Move to the next site or terminate the treatment.
6.7 Post-Treatment Recommendations
Post treatment recommendations should be provided by doctor. The Handpiece should be discarded.
BodyTite Operator Manual
System Maintenance
24 of 30
System Maintenance
7.1 Cleaning the Device
Wipe the device, including the footswitch with a damp, soft cloth. Liquid disinfecting solution may be used. Avoid using flammable solutions. Do not immerse any part of the system. The Handpiece should be discarded after use.
BodyTite Operator Manual
Troubleshooting
25 of 30
Troubleshooting
The BodyTite System provides monitoring of all critical parameters to ensure safety of the patient and user. If any of the following faults are detected the System automatically disables RF output and goes to an Error Screen.
8.1 Description of Faults
Problem
Checks
System did not turn on
Check power cord connection Check that main switch on rear panel is on Check fuses on back panel of the System Call Technical Service if problem persists
Software plug missing
The software plug is not inserted
Checksum
The software was not loaded properly from software plug Check plug connection and reboot the System Call Technical Service if problem persists
Fault H8002 - Handpiece is not connected
Check connection of Handpiece Reboot the System Replace Handpiece Call Technical Service if problem persists
Fault H8005 – System Memory Fault
Reboot the System Call Technical Service if problem persist
RF delivery problem
No energy delivery after the footswitch is pressed. Check
that system is in READY mode and the footswitch is connected. Reboot the system. Call Technical Service if this problem persists.
BodyTite Operator Manual
Troubleshooting
26 of 30
Problem
Checks
Bad Coupling Indicator exists at all times during treatment
Check that there is a water based gel applied to the skin
surface,and ensure that proper tumescent infiltration has been done. Reboot the system.
Replace the Handpiece if this problem continues. Call
Technical Service if this problem persists
Low Temperature
Message is displayed on screen if the temperature does not
rise during the treatment.
Reboot the system. Replace the Handpiece if this problem
continues. Call Technical Service if this problem persists.
BodyTite Operator Manual
System Specifications
27 of 30
System Specifications
Category
Item
Detail
Inut Power
Main Line Frequency (nominal)
Input Voltage (nominal) Input Current (rms)
50 - 60Hz
100 - 240VAC 2A
Operating Parameters
Ambient Temperature Range
Relative Humidity Atmospheric Pressure Warm-up Time
15° – 35°C [59° – 95°F] 30% to 80%, non-condensing 90 - 110kPa If transported or stored at temperatures
outside the operating temperature range, allow one hour for the system to reach room temperature before use.
Transport and Storage
Ambient Temperature Range
Relative Humidity Atmospheric Pressure
-20° – 65°C [35° – 131°F] 0% - 80%, non-condensing 50 - 110 kPa
Dimensions
System Handpiece cable
35 cm W x 46cm D x 100cm H
250 cm L
[18.2’’ W x 18.2’’ D x 40’’ H]
[100” L]
Weight
System Handpiece
20Kg
0.2Kg
[44lb] [0.4lb]
Output Parameters
Maximal RF Power
External Cut-Off Internal Cut-Off Treatment Time RF Frequency Tissue impedance
20 W for HP101306A and HP172206A 40 W for HP172246A 35-42oC 50-70oC 15-120sec 1MHz 50 -300Ohm
BodyTite Operator Manual
System Specifications
28 of 30
9.1 BodyTite Output Power Curves
The following curves depict the BodyTite output Power vs. Range of Load Impedance.
Figure 9-1: Output Power Versus Impedance
Figure 9-2: Output Power Versus Impedance
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
40.00
0 50 100 150 200 250 300 350 400 450 500 550 600 650 700 750 800 850 900
Output Power [W]
Load [Ohm]
BodyTite Output Power vs. Impedance
BodyTite Operator Manual
System Specifications
29 of 30
9.2 EMC Safety for the BodyTite Device
The device has been tested and found to comply with the limits for the medical devices to the IEC 60601-1-2. These limits are designed to provide reasonable protection against harmful interference in a typical clinical installation. This device generates uses and can radiate radio frequency energy. If not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that the interference to other devices, which can be determined by turning the device off and on, is caused by this instrument.
The user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device. Increase the separation between the devices. Connect the device into an outlet on a circuit different from that which was
previously used.
Consult InMode MD Ltd. service personnel for help. Interference to the device may be caused by portable and mobile RF communication
equipment. In case of an interruption, beware of such a device in the vicinity.
Use of the system with any accessory, transducer or cable other than those specified
may result in increased EMISSIONS or decreased IMMUNITY than those specified.
The system should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the system should be observed to verify normal operation in the configuration in which it will be used.
BodyTite Operator Manual
System Specifications
30 of 30