12
13
www.infopia21.com
EVOLUTION
TM
Blood Glucose T est Strip
Information
Limitations of System :
EVOLUTION
TM
blood glucose test strips provide accurate results when the following
constraints are observed:
• Use only the EVOLUTION
TM
blood glucose test strips with the EVOLUTIONTMblood
glucose test meter.
• Use fresh capillary whole blood only. Do not use plasma or serum.
• Do not use neonate samples.
• The blood glucose test strips are for single use only. Do not reuse.
• Dehydration may lower test results. If you are severely dehydrated, contact you
physician immediately.
• Inaccurate results may occur when in shock, hypotensive individuals, hyperglycemic,
or hyperosmolar state, with or without ketosis.
• EVOLUTION
TM
blood glucose test strips may be used at altitude up to 10,000 feet
without an effect on test results.
Physicians - Please note the following interferences that may affect
test results :
• EVOLUTION
TM
test strip results can be used with hematocrit levels in the range of 20%
to 60%. If the hematocrit range is out of the range(20%~60%), then the test results may
be lower or higher than actual value. If you do not know your hematocrit level, consult
your healthcare professionals.
• Interferences: Acetaminophen, uric acid, ascorbic acid(vitamin C), and other reducing
substances (when occurring in normal blood or normal therapeutic concentrations) do
not significantly affect results. However, abnormally high concentrations in blood may
cause inaccurately high results.
• Lipemic samples; Cholesterol up to 500 mg/dL or triglyceride up to 3000 mg/dL do not
significantly affect the results. Glucose values, however, in specimens beyond these levels should be interpreted with caution.
• Blood samples that contain a high concentration of dissolved oxygen may lower the test
result.
• EDTA containing tube is recommended as an anticoagulant tube.
Performance Characteristics :
The performance of EVOLUTION
TM
blood glucose test strips has been evaluated in
laboratory and in clinical tests.(please refer to strip manual for more details)
Measurement Range: The measurement range of the Blood Glucose Monitoring
System is 10 to 600 mg/dL (0.6-33.3 mmol/L).
The EVOLUTION
TM
Blood Glucose Test Meter has a variety
of excellent functional modes; beep sound to alert when
function is set, date, time, five daily alarms, unique user ID,
ability to designate activities related to the test result, and the
ability to designate three unique data averages.
Power On
Press for 3 seconds
at least
Date / Time
After pressing and
releasing either ▲ or ▼
buttons to set the date
and time, confirm each
with .
Year
After pressing and
releasing either ▲ or ▼
buttons to set the year
(from 2007 to 2099),
confirm with .
Test Unit
Press and release either
▲ or ▼ buttons to set the
test unit (mg/dL or
mmol/L), confirm with
.
User Activity Option
After pressing and releasing either ▲ or ▼ buttons
to turn the user Activity
option on/off, confirm with
.
User ID Option
After pressing and releasing either ▲ or ▼ buttons
to turn User ID option
on/off, confirm with .
Setting Your Blood Glucose Test Meter