Infopia Evolution Operation Manual

Operation Manual
Operation Manual
4
5
www.infopia21.com
Dear EVOLUTIONTMBlood Glucose Monitoring System Owner
Thank you for choosing the
EVOLUTIONTMBlood Glucose Monitoring System. This booklet has important information you must know about the EVOLUTIONTMBlood Glucose Monitoring System. Please read it carefully.
We understand that self-testing of blood glucose level provides a way to control your diabetes and may give you peace of mind by testing regularly. As a result
EVOLUTIONTMwas developed to provide you with a fast and accurate reading with a convenient and simple process. Our goal at EVOLUTIONTMis to provide the best quality healthcare products coupled with superior customer service. Always consult with your healthcare profes­sional before making any changes to your diabetes management. The EVOLUTIONTMBlood Glucose Monitoring System is for in vitro diagnostic use only.
Important information
The
EVOLUTIONTMBlood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use only). It should be used only for testing blood glucose and only with fresh capillary whole blood samples. It should not be used for the diagnosis of diabetes.
Consult your physician or diabetic healthcare professional about daily man­agement of your diabetes and proper use of the glucometer.
Please pay close attention when handling blood. Improper procedures may cause serious hazards to your health.
The
EVOLUTIONTMBlood Glucose Monitoring System contains small
parts. Please keep your monitoring system out of reach of children.
T est Principle
The enzyme glucose oxidase on the blood glucose test strip reads specifi­cally with the blood glucse. The current generated is converted and dis­played as blood glucose value. The
EVOLUTION
TM
Blood Glucose Monitoring System is plasma-calibrated
to allow easy comparison of results with laboratory methods. Blood glucose meter which is calibrated against a whole blood method
may have different results in comparison to
EVOLUTION
TM
.
The laboratory system used for calibration of the
EVOLUTION
TM
Blood Glucose Monitoring System is YSI 2300 STAT plus which is equipped with a glucuse oxidase system.
Note
< Maintaining the
EVOLUTIONTMblood glucose test meter >
1. Do not assemble or disassemble the blood glucose test meter.
2. It will have to keep in the place where the water does not reach.
3. Avoid getting dirt and moisture.
4. It does not have to keep in storage place of the chemical chemicals or
the place where gas etc occurs.
5. Keep out of direct sunlight.
CAUTION
1. If you do not check the code number, pull the blood glucose test strip out, conduct procedures from beginning.
2. If the code of the blood glucose test meter and of the blood glucose test strip vial does not match, you must set the code number.
3. If the code on the LCD window and on the blood glucose test vial does not match, follow the coding procedure for accurate results.
6
7
www.infopia21.com
1.
Learning the System
Intended use of Blood Glucose Test Meter
EVOLUTIONTMBlood Glucose Monitoring System Kit Contents EVOLUTION
TM
Blood Glucose Test Meter
EVOLUTION
TM
Blood Glucose Test Meter Display
EVOLUTION
TM
Blood Glucose Test Strip Information
2.Before Testing
Setting the blood glucose test meter Auto-Coding Function Performing a Glucose control solution test
3.Testing Your Blood Glucose
Testing Your Blood-Preparation Collecting a Drop of Blood Testing Procedure Testing Your Blood - Case of Setting user Option and ID Blood Glucose Test Strip Ejector Funtion
4.Reviewing Your Results
5. Deleting Test Results
6.Alternate Site Testing
7. Performing a Blood Test using and Alternate Site
8. EVOLUTION
TM
Blood Glucose Test Meter Software
9. Taking Care of Your Blood Glucose Test Meter
Cleaning Your Blood Glucose Test Meter and Maintenance
Replacing the Batteries
10. Troubleshooting
11. Specifications
12. Conversion table mmol/L mg/dL
13. Warranty
14. Symbol reference
Contents
The
EVOLUTIONTMBlood Glucose Monitoring System is intended for in
vitro diagnostic use only.
EVOLUTION
TM
Blood Glucose Monitoring System is for self-testing. It should not be used for the diagnosis of diabetes or for the testing of new­borns. The
EVOLUTIONTMwas designed to test a blood glucose (sugar) level and can be used with a capillary whole blood sample. Do not use the
EVOLUTIONTMBlood Glucose Monitoring System for any
other than for blood glucose testing.
CAUTION : Before using any product, please read all instructions.
Consult your physician or diabetic healthcare professional about daily management of your diabetes and proper use of
the glucometer.If you have any questions about the EVOLUTIONTMproduct, please contact your local represen tative.
Intended use of EVOLUTIONTMBlood Glucose T est Meter
7 8 9
10
11-12
13-14
15
16-17
18 19
20-22
23 24
25-26
27 28
29-30
31
32 33
34-36
37 38 39 40
9
8
www.infopia21.com
EVOLUTIONTMBlood Glucose Monitoring System KIT Contents
Your EVOLUTIONTMBlood Glucose Monitoring System has been sealed and it should contain all of the parts shown above. Please return the system to place of purchase or contact your nearest EVOLUTION
TM
representative if the container seals are broken or parts
are missing.
1.
EVOLUTIONTMBlood Glucose Test Meter
2.
EVOLUTIONTMBlood Glucose Test Strips
3.
Lancing Device
4.
Lancets
5. Operatrion
Manual
6. Warranty Registration Card
7. Patient Logbook
8. Carrying Case
9. Two 3V Li-CR2032 Batteries
EVOLUTION
TM
Blood Glucose T est Meter
Test Strip Port: Insert the Evolution test strip here
Test Strip Ejector
Display Screen: Displays your test results, sym­bols and messages.
Function Button: Powers the meter on/off, prompts memory data and acts as an enter button during setup and after tests.
Up/Down Arrow Buttons: Adjusts setting, scrolls through memory
Serial Number
Battery Cover
CAUTION
1. Do not use EVOLUTIONTMblood glucose test meter in a dry environment, especially if synthetic materials are present. Synthetic clothes, carpets, etc., may cause damaging static discharges in a dry environment.
2. Do not use EVOLUTIONTMblood glucose test meter near cellular or cord­less telephones, walkie talkies, garage door openers, radio transmitters, or other electrical or electrical equipment that are sources of electromagnetic radiation, as these may interfere with the proper operation of the blood glu­cose test meter.
11
10
www.infopia21.com
Low Battery Warning Symbol User ID Symbol Strip Code Symbol
Blood drop symbol for test Alarm Symbol
Before Having a Meal After Having a Meal After Sports Activity
After Taking Medication Temperature / Date Time / Test Result Amount
Test Result Unit Symbol Control Solution Symbol Average Test Result System
Delete results
EVOLUTIONTMBlood Glucose T est Meter Display
Use only
EVOLUTIONTMbrand blood glucose test strip with your
EVOLU-
TION
TM
blood glucose test meter.
Apply Blood Target
Confirmation Window
Contact Bar
EVOLUTION
TM
Blood Glucose T est Strip
Information
Storage and Handling Caution
1. Keep out of direct sunlight.
2. Store the
EVOLUTION
TM
Blood Glucose
Test Strip vials in a cool, dry
place between 2-30°C(36-86). Do not freeze.
3. Make a notation of the discard date on the vial label when you first open it. Discard remaining
EVOLUTION
TM
blood glucose test
strips 3
months after first opening the vial.
4. Close the vial cap tightly immediately after removing a
EVOLUTION
TM
blood glucose
test strip.
5. Do not use test strips after the expiration date printed on the pack­age or vial since it may cause inaccurate results.
6. Avoid getting dirt, food or water on the
blood glucose
test strip. Do not
bend, cut, or alter the
EVOLUTION
TM
blood glucose test
strip.
7.
EVOLUTION
TM
blood glucose
test strips are for single use only.
8. Keep the
blood glucose
test strip vial away from children.
9. Please refer to the user’s manual for
EVOLUTION
TM
blood glucose
test strip for additional information.
10. Discard the used
blood glucose
test strip carefully, to prevent any
infection.
12
13
www.infopia21.com
EVOLUTION
TM
Blood Glucose T est Strip
Information
Limitations of System :
EVOLUTION
TM
blood glucose test strips provide accurate results when the following
constraints are observed:
• Use only the EVOLUTION
TM
blood glucose test strips with the EVOLUTIONTMblood
glucose test meter.
• Use fresh capillary whole blood only. Do not use plasma or serum.
• Do not use neonate samples.
• The blood glucose test strips are for single use only. Do not reuse.
• Dehydration may lower test results. If you are severely dehydrated, contact you physician immediately.
• Inaccurate results may occur when in shock, hypotensive individuals, hyperglycemic, or hyperosmolar state, with or without ketosis.
• EVOLUTION
TM
blood glucose test strips may be used at altitude up to 10,000 feet
without an effect on test results.
Physicians - Please note the following interferences that may affect test results :
• EVOLUTION
TM
test strip results can be used with hematocrit levels in the range of 20% to 60%. If the hematocrit range is out of the range(20%~60%), then the test results may be lower or higher than actual value. If you do not know your hematocrit level, consult your healthcare professionals.
• Interferences: Acetaminophen, uric acid, ascorbic acid(vitamin C), and other reducing substances (when occurring in normal blood or normal therapeutic concentrations) do not significantly affect results. However, abnormally high concentrations in blood may cause inaccurately high results.
• Lipemic samples; Cholesterol up to 500 mg/dL or triglyceride up to 3000 mg/dL do not significantly affect the results. Glucose values, however, in specimens beyond these lev­els should be interpreted with caution.
• Blood samples that contain a high concentration of dissolved oxygen may lower the test result.
• EDTA containing tube is recommended as an anticoagulant tube.
Performance Characteristics :
The performance of EVOLUTION
TM
blood glucose test strips has been evaluated in
laboratory and in clinical tests.(please refer to strip manual for more details) Measurement Range: The measurement range of the Blood Glucose Monitoring
System is 10 to 600 mg/dL (0.6-33.3 mmol/L).
The EVOLUTION
TM
Blood Glucose Test Meter has a variety of excellent functional modes; beep sound to alert when function is set, date, time, five daily alarms, unique user ID, ability to designate activities related to the test result, and the ability to designate three unique data averages.
Power On
Press for 3 seconds at least
Date / Time
After pressing and releasing either or buttons to set the date and time, confirm each with .


Year
After pressing and releasing either or buttons to set the year (from 2007 to 2099), confirm with .
Test Unit
Press and release either or buttons to set the test unit (mg/dL or mmol/L), confirm with
.
User Activity Option
After pressing and releas­ing either or buttons to turn the user Activity option on/off, confirm with
.
User ID Option
After pressing and releas­ing either or buttons to turn User ID option on/off, confirm with .
Setting Your Blood Glucose Test Meter
14
15
www.infopia21.com
Setting Your Blood Glucose Test Meter
Temperature unit
Press and release either
or ▼buttons to set
or . Enter button.
Number of days for average
After pressing and releas­ing either or buttons to set the number of days for average calculation (3 different possible), confirm with .
Alarm on/off
After pressing and releas­ing either or buttons to turn the alarm on/off, confirm with .
Alarm time
After pressing and releas­ing either or buttons to set the alarm hour and minute, confirm with
.
Alarm
After programming the number of desired alarms (5 alarms possible), con­firm with
.
CAUTION
1. **VERY IMPORTANT: Correct test unit. Your Evolution
TM
blood glucose test meter is pre-set to mg/dL as test unit. A wrong test unit will lead to wrong test result display! Please consult your healthcare professional if you’re not sure which test unit to use. Stored test results are displayed correctly after re-setting the test unit. See also conversion table on page 38.
2. Without setting the date properly, the average glucose level and the results in memory will not show proper values. It is recommended to set the meter before use and when new batteries are installed.
3. Yon can’t test your blood glucose while in the setting mode.
4. To turn the meter off during setup, press for at least 3 seconds

Auto-Coding Function
1. EVOLUTIONTMhas automatic code recognition function.
2. It recognizes automatically the code number. This is a very convenient function preventing the inconven­ience of setting the code number on the blood glucose test meter every time.
CAUTION
1. Do not bend the blood glucose test strip to prevent the Automatic code recognition failure.
2. If the code recognition label is damaged, the code recognition may fail. Please check the code number on the LCD display window with the code number on the blood glucose test strip vial.
17
16
www.infopia21.com
Performing a Glucose Control Solution T est
1. Shake the glucose control solution before use.
2. Discard the 1st drop of glucose control solution, this will eliminate any residue.
3. Place a drop of glucose control solution on a clean dry surface (ex. The lid of the blood glucose test strip vial).
4. Insert a blood glucose test strip into the port and press the up arrow. A glucose control solution vial will appear on the bottom left hand side of the screen, above the date and time. This will allow you to differentiate between a glu­cose control solution test and an actual blood test for future reference.
5. Dip the blood glucose test strip into the glucose control solution.
6. Results appear in 3 seconds.
7. Compare the result to the range printed on the blood glucose test strip vial, the results should fall within that range.
8. If your glucose control solution falls out of range, do not perform a glucose test, please contact your local representative.
INFC16A
INFC16A
INFC16A
GL-3B0-3C1-01 Rev.2008-10
GL-3B0-3C0-01 Rev.2008-10
GL-3B0-3C2-01 Rev.2008-10
Glucose Control Solution
(LOW)
Glucose Control Solution
(NORMAL)
Glucose Control Solution
(HIGH)
For use with EVOLUTIONTM blood glucose test meter and blood glucose test strips. For in vitro diagnostic use. For self-testing. Shake well before using. Do not freeze. Store in cool, dry place between 8~30ŭ(46~86).
Use with in 3 months after first opening. Do not use if expiry date has passed.
After use, replace cap immediately. The expected range is printed on the blood glucose test strip vial. Discard the used glucose control solution carefully, according to the local regulations. Test at 20-25( 62-77䲵)
For use with EVOLUTION
TM
blood glucose test meter and blood glucose test strips. For in vitro diagnostic use. For self-testing. Shake well before using. Do not freeze. Store in cool, dry place between 8~30ŭ(46~86).
Use with in 3 months after first opening. Do not use if expiry date has passed.
After use, replace cap immediately. The expected range is printed on the blood glucose test strip vial. Discard the used glucose control solution carefully, according to the local regulations. Test at 20-25( 62-77䲵)
For use with EVOLUTION
TM
blood glucose test meter and blood glucose test strips. For in vitro diagnostic use. For self-testing. Shake well before using. Do not freeze. Store in cool, dry place between 8~30ŭ(46~86).
Use with in 3 months after first opening. Do not use if expiry date has passed.
After use, replace cap immediately. The expected range is printed on the blood glucose test strip vial. Discard the used glucose control solution carefully, according to the local regulations. Test at 20-25( 62-77䲵)
Performing a Glucose Control Solution T est
The glucose control solution should be used when there is suspicion that the blood glu­cose test meter or the blood glucose test strips are not working properly. For example: if strips have been exposed to extreme conditions or if you drop the blood glucose test meter.
Check the expiration date before performing a glucose control solution test. Do not used if expired. Once opened, solution expires after 3 months.
The Glucose control solution should be used
Whenever you suspect the blood glucose test meter or blood glucose test strip is not functioning properly.
If your blood glucose test results are not consistent with your symptoms or if you think they are not accurate.
If you have dropped the blood glucose test meter.
For quality control in the point of care usage.
For teaching or learning the system.
CAUTION
If your glucose control solution falls out of range, do not perform a glucose test, please
contact your local representative.
The Glucose Control Solution range is not the recommended range for your blood glucose level.
For example If you use a normal glucose control solution, you must use the normal range printed on the side of the blood glucose test strip vial.
Results that fall outside the expected range may indicate:
Test not performed at room temperature.
An error in the way you are doing the test.
Expired or contaminated glucose control solution.
A problem with the blood glucose test meter.
Glucose control solution that is too warm or cold.
Failure to discard the first drop of glucose control solution and to wipe the bottle tip
clean.
Failure to shake the glucose control solution adequately.
Note: - EVOLUTIONTMglucose control solution is sold separately.
- It is recommended that glucose control solution is stored at room temper­ature 20~25℃(68〜77) before testing.
19
18
www.infopia21.com
Collecting a Drop of BloodT esting Your Blood - Preparation
Before testing, be sure to read this section and
blood glucose
test strip pack-
age insert found in the
blood glucose
test strip box carefully, Make sure you
have all the necessary items to begin testing:
EVOLUTION
TM
blood glucose
test
meter
EVOLUTION
TM
blood glucose
test
strip
• Lancing Device
• Sterile Lancets
CAUTION
To reduce the chance of infection.
• The Lancing Device and sterile lancets should NOT be shared with others.
• Always use a new, sterile lancet and new blood glucose test strip.
• Lancets are for single use only. DO not reuse.
• Avoid getting hand lotion, oils, dirt, or debris in or on the Lancets and the Lancing Device.
• Dry your hands completely before testing.
Practice using the lancing device and become accustomed with its use. Wash your hands with warm, clean water and soap. Dry your hands completely before testing.
The lancing device is a convenient medical tool for collecting capillary blood samples for glucose monitoring or other tests which require one or two drops of blood. As a more advanced lancing device, it provides ultimate safety and comfort for obtaining blood samples. An adjustable tip offers 5 levels of skin penetration for individual user’s comfort and a lancet ejector enables the safe disposal of the used lancet. As an added bonus, the also comes with a transparent tip for AST (Alternative Site Testing).
PRECAUTIONS
Never use a lancet that has been used by someone else. This could lead to con­tamination.
If the lancing device is to be used by another person, the unit must be properly dis-
infected and a new tip and lancet must be used.
Do not leave the lancet in the device after use.
A new lancet must only be placed into the lancing device directly before testing.
This device has many small parts and could be a choking hazard for children if
swallowed.
In the case of hospital use, hospitals need to consult their own infection control
protocols in order to avoid any contamination.
Always dispose of the used lancet in a biohazard garbage container.
COMPONENTS
Clear Tip
Lancet Holder
Release Button
Ejection System
Sliding Barrel
Depth Indicator/Adjustable Tip
21
20
www.infopia21.com
1. Unscrew the tip of the lancing device by turning it counter-clock­wise while holding the base firm­ly.
2. With two fingers, pull out the lancet carrier and hold it in place.
3. While holding the lancet carri­er, insert a new sterile lancet into the bottom of lancet carri­er.
4. Twist the cap of the lancet off.
5. Put the tip of the lancing device back on and turn it clockwise.
6. Hold the tip firmly in one hand then pull out the sliding barrel with the other hand. This will cock the lancing device.
7. Place the lancing device in place. Hold the firmly against the side of the finger, with the cap resting on the finger. (The harder it is pressed the deeper the punc­ture.) Press the release button to take a sample.
8. Unscrew the tip of the lancing device by turning it counter­clockwise.
9. Push the lancet ejector forward with the thumb and simultane­ously pull out the sliding barrel to dispose of the used lancet in a proper biohazard container.
Click
RECOMMENDED AVAILABLE LANCETS
Some general sterilized lancets are BD-Ultra Fine, Cleanlet Fine, E-Z Ject, G-P Lite, Microlet, Monolet, One Touch, Soft Touch, Therasense, Ultra TLC and Unilet GP. Note: Some less known lancets may not work properly with the due to their dimen­sions. Please, contact your supplier to make sure you are using proper lancets.
USING THE CLEAR TIP (AST)
1. Place the on the skin in the chosen area, press and hold the continuously for a few seconds, then push the release button to take a sample.
2. Watch through the clear tip until a sufficient blood sample is taken. If there is not enough blood, gently massage the area until a sufficient sample has been collect­ed.
STORAGE
Products must be stored at room temperature protected from sunlight or moisture.
CLEANING AND DISINFECTION
Clean and wipe the outside of the lancing device a minimum of once a week with a soft cloth, dampened with mild soap and water. Wash the adjustable tip and the clear tip once a week with mild soap. DO NOT immerse the in liquid.
WARRANTY
Lancing device has a 2-year warranty from the date of purchase. A malfunctioning device should be returned to the following address:
Manufacturer
: GMMC Room No. 1112, Ace Tower 9th Bldg., 345-30, Kasan-Dong, Kumchon-Ku, Seoul, South Korea Tel : +82 2 2104 0470 Fax : +82 2 2104 0472 EU
Authorized Representative : GMMC, S.L. C/Vinalopo' 4 - 1° - Pta 4 Valencia, Spain Tel : +34 963 691 484 Fax: +34 961 125 949
Collecting a Drop of Blood Collecting a Drop of Blood
23
22
www.infopia21.com
Correct
Completely Filled
Place the top edge of the blood glucose test strip to the area you
prepared for testing.
The blood will automatically be “drawn” into the blood glucose test
strip channel.
When the chamber has enough blood, your blood glucose test meter
will automatically count down.
Incorrect
T esting Procedure
CAUTION
1. If the countdown does not start, do not add more blood to the blood glucose test strip! Discard the blood glucose test strip an restart testing.
2. If you do not conduct the test within 3 minutes, the blood glucose test meter will automatically power off to save battery life. In this case the test procedure should start again from the beginning.
Sample size
Evolution
TM
blood glucose
test meter only requires 0.3
sample volume.
Step 2
Applying Blood
Incorrect
Correct
Poorly filled
T esting Procedure
Firmly insert the blood glucose test strip into the blood glucose test meter port with the “3.3” facing towards you.
The blood glucose test meter automatically turns on displaying the code, ambient temperature and the date and time.
The EVOLUTIONTMwill automatically recognize the blood glucose test strip code number and adjust the blood glucose test meter accordingly.
A blinking
blood glucose
test strip will appear at the top of the screen indi-
cating the
blood glucose test
meter is ready for testing.
CAUTION
1. If the code is not displayed after for a while, pull the blood glucose test strip out of the port, and re-start the procedure from the beginning.
2. If the code on the LCD display window and on the test vial does not match, try another new blood glucose test strip. If the mismatch persists, please contact your local representative for help.
3. If the meter does not power on, pull the blood glucose test strip out of the port, and re-insert the blood glucose test strip.
4. Avoid testing under direct sunlight, for a more accurate test result.
5. If you apply your blood sample too early, ERROR5 message will appear on LCD ( refer to p35).
• For a more accurate test result avoid testing under direct sunlight.
Step 1
Note : Inserting the test strip in the wrong direction will turn the meter on, but you will
be unable to test.
25
24
www.infopia21.com
T esting Procedure
Step 3
After beeping sound, the test will begin automatically and your results will appear in 3 seconds. It should begin counting down from 3 to 1 second on the LCD window. The LCD window will display the result of your blood glucose level, including temperature and time.
Record the result value in your EVOLUTION
TM
logbook. When you
pull out the blood glucose test strip, the meter turns off automatically.
Step 4-1
If you selected to use the User Activity option during set up, press and button to select the activity that corre­lates with your result (refer to p10), then press .
If you do not select to use user activity option during set up, press to contin­ue to step 4-2.
Step 4-2
If you selected to use the User ID option during setup, press and button to select the correct User ID, then press .
CAUTION
1.
If the test result is out of the test range, the Hi/Lo message will be shown on the LCD window.
(refer to p35)
2.
Safely discard used blood glucose test strips and lancets.
Testing Your Blood - Case of Setting User Option and ID
CAUTION
1. If you did not activate the User Activity option and/or the User ID option, Step 4-1 and/or Step 4-2 procedures are not conducted.
2. Although User Activity option and/or User ID option are activated, if you pull the blood glucose test strip out right after the test, you can not move onto step 4, The test result is stored, not under a specific User ID and only displayed after deactivating the User Id option.
3. Please refer to page 13, if you want to set the User Activity option and User ID option.
27
26
www.infopia21.com
Reviewing Y our Result
The EVOLUTIONTMblood glucose test meter stores up to 365 test results in its built-in memory, along with the average sugar level for the number of days which you had preset the blood glucose test meter to calculate. These values can be recalled in the LCD display at any time.
Press and release button. The most resent result first if user ID feature has been elected during setting the blood glucose test meter.
Check your previous result by pressing the arrow button.
Check your average data by pressing the arrow button.
The third latest The second latest The latest
7day average
from the test date
14 day average
from the test date
21day average
from the test date
Test Strip Ejector Function
1. Once the test is complete, slide the ejector button forward to remove the blood glucose test strip from the blood glucose test meter .
2. Discard the used blood glucose test strip and lancet accordingly .
Press and Push forward
Eject your blood glucose
test strip
CAUTION
1. If you push the ejector button forward too much, it may be part of the failure.
2. Do not give strong impact to blood glucose test meter.
29
28
www.infopia21.com
Deleting T est Results
Deleting individual test result
To delete any individual test result in the memory press or button for 3 seconds, while the test result to delete is displayed. is blinking. After the second beep sound the test result is deleted.
Deleting all test results
To delete all test results press and but­ton at the same time for 3 seconds, while any test result is displayed. is blinking. After the third beep sound all test results are deleted.
Caution
Only selected Individual test results are deleted when you
recalled test results in the user ID condition
The deleted test results can’t be recovered. Please be careful
in deleting the test results.
Pressing the button once and the
User ID option will appear in the LCD screen.
Average test result Test Unit
Test result amount Average test result days
Reviewing Y our Result
After entering the user ID mode, use
and button to scroll and choose the correct user ID, then press the but­ton.
If you choose the User ID to check the
stored results, only the results of the select­ed user are displayed.
31
30
www.infopia21.com
Alternate Site T esting
Important Information About Using Alternative Sites Testing :
• Alternate sites where you can test are dorsal hand, ventral palm, upper arm, forearm, calf, and thigh.
• Under certain conditions, blood glucose test results obtained using sam­ples taken from your alternate sites may differ significantly from fingertip samples
• The conditions in which these differences are more likely to occur are when your blood glucose is changing rapidly such as following a meal, as insulin dose, or associated with physical exercise.
• When blood glucose is changing rapidly, fingertip samples show these
changes more quickly than alternate sites samples.
• When your blood glucose is falling, testing with a fingertip sample may identify a hypoglycemic (low blood glucose) level sooner than a test with a alternate sites sample.
• Use alternate sites samples only for testing prior to or more than two hours
after meal, insulin doses, or physical exercise.
• Testing performed within two hours after meals, insulin doses, or physical exercise or whenever you feel that your glucose levels may be changing rapidly should be done from the fingertip.
• You should also use fingertip testing whenever you have a concern about hypoglycemia (insulin reactions) such as when driving a car, particularly if you suffer from hypoglycemic unawareness (lack of symptom to indicate as insulin reaction), as forearm testing may fail to detect hypoglycemia.
Performing a Blood T est Using and Alternate Site
Performing a Blood Test Using an Alternate Site To ensure accurate results when lancing your forearm, upper arm, hand, thigh, or calf, wash the test site with soap and water.
Make sure there is no cream or lotion on the test site.Thoroughly dry your hands and test site.
CAUTION
To ensure accurate results when lancing your arm (Upper arm or forearm), leg (calf or thigh) or palm (ventral palm or dorsal hand), wash your hand and test site with soap and water.
T o receive accurate test results, at least 0.3minimum sam- ple volume is required.
Upper Arm
Forearm
Thigh
Calf
Dorsal Hand
Fingertips
Ventral Palm
Important :
We recommend that you test on your fingertips if you are testing for hypoglycemia (low blood glucose) or if you are suffering from hypoglycemia unawareness.
33
32
www.infopia21.com
Lance the test site with the lancing device:
1. Hold the clear cap down against the top of your test site. Press the release button. Do not lift up.
2. Continue to hold the lancing device and gradually increase pressure for several seconds.
3. While holding the lancing device on your test site, look through the clear cap; until a drop of blood appears approximate
size
4. Lift the lancing device straight up; be care­ful not to smear the blood on your testing site.
5. Place the edge of the blood glucose test strip to the drop of blood.
6. The blood will automatically draw into the channel of the blood glucose test strip.
T o bring fresh blood to the surface of the test site, rub the test site vigorously for a few seconds until you feel it getting warm.
EVOLUTIONTMBlood Glucose T est Meter Software
You can transfer test results from the EVOLUTIONTMblood glucose test meter to a computer. You can download the software from our web site.
(http://www.infopia21.com)
You can purchase the computer communication cable separately from a local representative.
cable
[User]
[PC]
Performing a Blood T est Using and Alternate Site
35
34
www.infopia21.com
Cleaning Your Meter and Maintenance
Meter
Your meter does not require special maintenance or cleaning. avoid getting dirt, dust, blood, glucose control solution, or liquids on the blood glucose test meter, the test port, or data port. Your blood glucose test meter operation temperature is 10~40(50℉〜104℉). It is recommended that you store the meter in its carrying case after each use. A cloth dampened with water and mild detergent can be used to wipe down the outside of the blood glucose test meter. Your
EVOLUTION
TM
blood glucose test
meter is a precision instrument. Please handle it with care.
Lancet Device:
Clean the lancet device and caps with soap and tepid water. To disinfect the lancet device, prepare a disinfectant solution of one part household bleach to 10 parts water. Dampen a cloth with this solution and wipe the lancet device thor­oughly. Soak only the cap for at least 30 minutes in the disinfectant solution. Do not soak the lancet device in liquid. Rinse the lance device and cap with water and dry thoroughly.
For detailed instructions concerning
EVOLUTIONTMblood glucose test strips,
refer to the test strip package insert found in the blood glucose test strip box.
Please store the
EVOLUTIONTMblood glucose monitoring system in a cool,
dry area, out of reach of children. Do not freeze. For a more accurate test result avoid testing under direct sunlight.
Do not soak the blood glucose test meter or blood glucose test strips in water or liquid.
Do not subject the blood glucose test meter or blood glucose test strips to excessive heat.
Use your blood glucose test meter according to the instructions in the manual.
If you need to purchase the lancing device, lancets or blood glucose test
strips, contact your local representative.
Discard the used Lancets carefully, to prevent any infection.
Replacing the Batteries
The low battery icon will appear in the upper left corner of the LCD screen to alert you when the battery power is running low, indicating a new battery is needed.
Your EVOLUTION
TM
blood glucose test meter uses only two 3V Lithium batteries (CR2032), which are included. When replacing the batteries, only CR2032 or equivalent lithium battery should be used.
After changing the batteries, please be sure to recheck your meter settings.
⊕⊕
<Follow-up Action>
1. Check your meter with the glucose control solution (p16-17)
2. Perform your test again (p20 - 22)
3. If you are experiencing symptoms that are not consistent with your
blood glucose test results, call your healthcare professional.
CAUTION
Remove both batteries from the battery compartment and dispoes of them according your institution's guidelines.
37
36
www.infopia21.com
T roubleshooting
The following chart may help you identify certain problems, but may not solve all problems that can occur. Contact your authorized representative or Infopia USA customer support if the problem persists.
Message
Problem Action Required
Problem with the blood glucose test meter
Place the battery again, and set the blood glucose test meter. If the problem persists, please contact your authorized repre­sentative or customer support
Caused by either used or wet blood glucose test strip
Please insert a new blood glucose test strip and perform your test again
Blood glucose test strip is dam­aged
Please insert a new blood glucose test strip and perform your test again
User applied the blood sample before the LCD display was blinking
Please wait for the LCD to display icon blinking before applying your blood sample
The ambient tem­perature is too low
Place the blood glucose test meter within the operating temperature range for 10 min­utes or more and retest.
The ambient tem­perature is too high
Place the blood glucose test meter within the operating temperature range for 10 min­utes or more and retest.
The test result is lower than 10mg/dl
Check the accuracy of your blood glucose test strip by performing a glucose control solution test. If the test results in a normal reading, re-test your blood sample two or three times. If "Lo" persists, consult your doctor immediately.
The test result is higher than 600mg/dL
Check the accuracy of your blood glucose test strip by performing a glucose control solution test. If the test results in a normal reading, re-test your blood sample two or three times. If "Hi" persists, consult your doctor immediately.
Message
Problem Action Required
Troubleshooting
Problem with the Auto-coding label
Please insert a new blood glucose test strip and perform your test again. If the problem persists, pleas contact your authorized representa­tive or customer support.
39
38
www.infopia21.com
Battery is dead or there is a problem with the blood glu­cose test meter
The blood glu­cose test meter does not power on
Test does not start after apply­ing blood sam­ple
The result is inconsistent
Change the battery and if the problem persist, contact your local representative
Poor blood sample
Please insert a new blood glucose test strip and perform your test again
There may be a problem with the blood glucose test strip
No readings in memory to display
No enough read­ings in memory to display designat­ed averages
Please insert a new blood glucose test strip and perform your test again
Message
Problem Action Required
Troubleshooting Specifications
Sample type Sample volume
Test Range Reading time
Calibration Altitude
Operating Temperature
Operating Humidity Strip storage temperature
Display Type Dimension Weight
Power rating Battery Life
PC communications port
Capillary whole blood
0.3ul 10~600 mg/dL
3 seconds Plasma - Equivalent
3048m (10,000 feet )
10-40(50-104) 10-90%
2-30(36-86
℉)
LCD 76 X 56 X 18 (mm) 45g ( Including Battery)
3 V Li Battery (CR2032) X 2 5,000 running tests
Communication available
41
40
www.infopia21.com
Conversion table mmol/L mg/dL
mmol/L
mg/dL
mmol/L
mg/dL
mmol/L
mg/dL
6.7 7.0 7.2 7.5 7.8 8.0 8.3 8.9 120 126 130 135 140 145 150 160
0.55 1.0 1.5 2.0 2.2 2.5 2.8 3.0 10 18 27 36 40 45 50 54
14.4 15.0 16.0 16.6 17.0 18.0 19.0 20.0 260 270 288 300 306 325 342 360
3.3 3.9 4.0 4.4 4.7 5.0 5.5 6.0 6.1 60 70 72 80 85 90 100 108 110
9.0 9.4 10.0 10.5 11.0 11.1 12.0 12.5 13.9
162 170 180 190 196 200 216 225 250
20.8 22.2 23.0 24.0 25.0 26.4 27.7 30.0 33.3
375 400 414 432 450 475 500 540 600
mmol/L
mg/dL
mmol/L
mg/dL
mmol/L
mg/dL
Three - Year Warranty
If, at any time during the first three years after purchasem the blood glucose test meter does not work for any reason with the exception of obvious abuse, misuse, or disastrous damage, Infopia Co.,Ltd. will replace your EVOLUTION
TM
blood glucose test meter or equivalence
free of charge.
The Warranty Policy of Infopia Co.,Ltd. only applies to the original pur­chaser of the Meter and will not include the batteries supplied with the blood glucose test meter set.
Please fill in the warranty card attached to the set. Mail or fax it back to us.
The EVOLUTION
TM
blood glucose test meter has a full three-year war­ranty from the original date of your purchase. For your convenience, please include your date and place of purchase.
Warranty
42
www.infopia21.com
Description
Symbol
Used By This prodect fulfills the requirements of Directive 98/79/EC
on in vitro diagnostic medical devices Caution, consult accompanying documents
For In Vitro Diagnostic Use Batch code
Catalogue Number Temperature limitation
Do not re-use
Serial number
Date of Manufacture Manufacturer
Keep away from sunlight For Self-testing
Separate collection for waste electrical and electorinic equipment
Symbol Reference
Consult Instructions for use
SN
INFM16A
GL-3B1-1B0-02 Rev. 2012-04
1155-1, Pubgot-Dong, Ilsanseo-Gu, Goyang-City, Gyeonggi-Do, Korea
Obelis S.A Bd. General Wahis 53, 1030 Brussels, Belgium Tel: +(32) 2. 732.59.54 Fax: +(32) 2.732.60.03 E-Mail : mail@obelis.net
infopia Co., Ltd.
#1603, Dongil Technotown A bldg. 889-3, Kwanyang2-Dong, Dongan-Gu, Anyang, Kyunggi, 431-716, Korea
Room No. 1112, Ace Tower 9
th
Bldg., 345-30, Kasan-Dong, Kumchon-Ku, Seoul, South Korea
Lancing Device : GMMC
Lancet : SAE HAN Medical Corp.
Loading...