Atom Infant Warmer Model 103
Power requirements Customer specified
Power consumption : 700VA (Maximum)
Heater capacity 500W
Classification Class I, Type BF
Dimensions 84(W) X 110(D) X 178-218(H)cm (Mattress surface : 80-120cm)
Weight 105kg
Temperature control Servo control (Manual control also available)
Skin temperature Setting range : 34.0-38.0℃
Display range : 30.0-42.0℃
Heater output Setting range : 0-100% (in 5% increments)
Indication range : 0-100 (in 10 levels)
Alarms Skin temperature, skin temperature probe, check baby, sysytem failure, main failure, canopy state
Timer Time display range : 0-60minutes 00 seconds (in 1 second increments)
APGAR mode /CPR mode switchable
Lighting LED lamp Adjustable illumination intensity 0-1200lux
Mattress platform angle
Inclination angle (back and forth) : +13°~-13°
Communication port RS-232
Accessories Skin temperature probe (yellow) ……………………………1
Dust cover ………………………………...………………….1
Mattress sheet ……………………………..…………………1
Pulse Oximeter Unit
Display range 1-100%SpO2
Averaging time 8, 4, 16 seconds (Selectable)
SpO2 alarm setting range Upper limit : 50-99% (in 1% increments)
Lower limit : 45-95% (in 1% increments)
Pulse rate display range 25-240bpm
Pulse rate alarm setting range Upper limit : 80-240bpm, OFF
Lower limit : 35-180bpm, OFF (in 5bpm increments)
Alarms SpO2 high/low limit, pulse rate high/low limit, SpO2 sensor
Resuscitation unit 104
Manometer
Pressure indication : -2 ~ 8 kPa (-20 ~ 80cmH2O)
Suciton unit
Suction pressure control range : 0 ~ 26.7 kPa (0 ~ 200mmHg)
O2 Blender
Oxygen concentration control range : 21 ~ 100%
Oxygen flow rate control range : 0 ~ 15L/min.
Resusci-flow unit
PMAX setting range : 19 ~ 58 kPa (20 ~ 60cmH2O)
Default setting range : 39 kPa (40cmH2O)
Resuscitation unit 105
Manometer
Pressure indication : -2 ~ 8 kPa (-20 ~ 80cmH2O)
Suciton unit
Suction pressure control range : 0 ~ 26.7 kPa (0 ~ 200mmHg)
O2 Blender
Oxygen concentration control range : 21 ~ 100%