Infa Warmer ATOM User Manual

ATOM Infant Warmer
From the delivery room to NICUMeets wide needs with substantial functions and features.
Superior operation performance and functionality will support daily nursing care.
Pre-heating function
Pre-heating condition enables to treat a baby under warm condition right after delivery.
CPR timer
Integrated CPR timer in compliance with the international resuscitation consensus. APGAR timer is also integrated.
Note: Information in this document is subject to change without notice.
Far-infrared radiation
High warming efficiency thanks to Far-infrared radiation.
Adjustable LED light
Even if the canopy is tilted, baby’s skin color can be clearly
checked thanks to the LED light. Light intensity can be adjusted.
Drawer
Drawer is available as an option. The drawer is closed very silently in order to minimize the sound stress to a baby.
Baby Guard
Thanks to rotary damper, the baby guard can be opened very silently without touching.
LED Light
Canopy
Add-on Unit
As add-on unit for Infant Warmer, two kinds of resuscitation unit are available depending on user’s demand.
Resuscitation Unit 104
Resusciflow is integrated for safety resuscitation treatment.
Manometer
Resusci– flow
Suction unit
O2 Blender
Resuscitation Unit 105
Oxygen Blender and manometer is integrated.
Manometer
O2 Blender
Suction Unit
X-ray tray
Hi-Low Stand
Specifications
Note: Information in this document is subject to change without notice.
Atom Infant Warmer Model 103
Power requirements Customer specified
Power consumption : 700VA (Maximum)
Heater capacity 500W
Classification Class I, Type BF
Dimensions 84(W) X 110(D) X 178-218(H)cm (Mattress surface : 80-120cm)
Weight 105kg
Temperature control Servo control (Manual control also available)
Skin temperature Setting range : 34.0-38.0
Display range : 30.0-42.0
Heater output Setting range : 0-100% (in 5% increments)
Indication range : 0-100 (in 10 levels)
Alarms Skin temperature, skin temperature probe, check baby, sysytem failure, main failure, canopy state
Timer Time display range : 0-60minutes 00 seconds (in 1 second increments)
APGAR mode /CPR mode switchable
Lighting LED lamp Adjustable illumination intensity 0-1200lux
Mattress platform angle
Inclination angle (back and forth) : +13°-13°
Communication port RS-232
Accessories Skin temperature probe (yellow) ……………………………1
Dust cover ………………………………...………………….1
Mattress sheet ……………………………..…………………1
Pulse Oximeter Unit
Display range 1-100%SpO2
Averaging time 8, 4, 16 seconds (Selectable)
SpO2 alarm setting range Upper limit : 50-99% (in 1% increments)
Lower limit : 45-95% (in 1% increments)
Pulse rate display range 25-240bpm
Pulse rate alarm setting range Upper limit : 80-240bpm, OFF
Lower limit : 35-180bpm, OFF (in 5bpm increments)
Alarms SpO2 high/low limit, pulse rate high/low limit, SpO2 sensor
Resuscitation unit 104
Manometer
Pressure indication : -2 8 kPa (-20 80cmH2O)
Suciton unit
Suction pressure control range : 0 26.7 kPa (0 200mmHg)
O2 Blender
Oxygen concentration control range : 21 100%
Oxygen flow rate control range : 0 15L/min.
Resusci-flow unit
PMAX setting range : 19 58 kPa (20 60cmH2O)
Default setting range : 39 kPa (40cmH2O)
Resuscitation unit 105
Manometer
Pressure indication : -2 8 kPa (-20 80cmH2O)
Suciton unit
Suction pressure control range : 0 26.7 kPa (0 200mmHg)
O2 Blender
Oxygen concentration control range : 21 100%
Oxygen flow rate control range : 0 15L/min.
Power Requirement AC120V AC220-240V
SpO2 Masimo
SpO2 Nellcor
Weight Monitor
Optional Add-on Unit
Resuscitation unit 104 PV
Resuscitation unit 104 EV
Resuscitation unit 105 PV
Resuscitation unit 105 EV
Manufacturer: Atom Medical Corporation Exporter: Atom Medical International, Inc. Iwakata Bldg., 3-18-16, Hongo, Bunkyo-ku, Tokyo, JAPAN Phone: + 81 3 3815 2941 Fax: + 81 3 3812 9670/4080 URL: http://www.atom-ami.co.jp
Model Number
Model 103
<Note> 104 PV: T-Piece Resuscitation Unit with Blender and Suction Unit (connect to central piping system) 104 EV: T-Piece Resuscitation Unit with Blender and Suction Unit of ejector type to make suction operation possible where no suction line is available 105 PV: O2 Blender Unit and Suction Unit (connect to central piping system) 105 EV: O2 Blender Unit and Suction Unit of ejector type to makes suction operation possible where no suction line is available
: Add-on Unit