Industrial Scientific VX500 Instruction Manual

P/N 1709-8773 REV 3.0 Printed 0104 1,000 Specifications Subject to Change
1001 Oakdale Road, Oakdale, PA 15071-1500 (412) 788-4353 • Toll Free 1-800-DETECTS FAX 412-788-8353 • Service Dept. 1-888-788-4353
Instruction Manual
GUARANTEED.
FOR LIFE.
P
HOTOIONIZATION DETECTOR
OUR MISSION
Design - Manufacture - Sell:
Highest quality products
for the preservation of
life and property.
Provide:
Best customer service
available.
Dear Valued Customer,
Thank you for buying and using Industrial Scientific’s VX500 PhotoIonization Detector.
Your VX500 can be relied upon for dependable service, day after day. It has been designed, manufactured, tested and proven under the most scrutinizing conditions possible. With the minimal care and maintenance described in this Instruction Manual, it will provide you with years of reliable monitoring.
I am most concerned that you be pleased with the performance of your VX500 in the months and years ahead. I urge you to call us with any questions or comments you may have. Often times a phone call and a question can save you hours of frustration. Please never hesitate to contact me at 1-800-DETECTS (338-3287).
All of us at Industrial Scientific appreciate the opportunity to serve you.
Sincerely,
Kent D. McElhattan President & CEO Industrial Scientific Corporation
3
W
ARNINGS ANDCAUTIONARYSTATEMENTS
Failure to perform certain procedures or note certain conditions may impair the performance of the instrument. For maximum safety and performance, please read and follow the procedures and conditions outlined below.
Recharge battery only in a non-hazardous location. Instrument tested for intrinsic safety in explosive gas/air (21.%
oxygen) mixtures only.
UV lamp requires periodic cleaning to function properly. The frequency of cleaning is dependent upon the environment in which the VX500 is used. Industrial Scientific recommends cleaning the UV lamp after every 40 hours of use.
Any new or freshly cleaned 10.6 eV lamp requires a burn-in period of 24 hours in order for the lamp’s output to stabilize. This is done by placing the lamp into the lamp conditioner found on the charger/datalink accessory.
Always check the ionization potential of the target gas prior to sampling to ensure proper lamp selection. For example, a 10.6 eV lamp can only detect gases with an ionization potential below
10.6 eV.
The presence of water vapor in the sample stream can cause quenching of the detector signal due to UV absorption.
Oxygen and methane are UV absorbers. Significant changes in
their concentration can cause a change in the PID signal.
Ammonia gas causes degradation of the VX500 PID module’s performance.
Ethylene gas does not provide a consistent signal therefore the VX500 should not be relied upon to provide consistent readings when monitoring ethylene.
A 5 minute warm-up time is required for VOC readings to fully
stabilize after power-up of the VX500.
!
!
!
!
!
!
!
!
!
!
2
W
ARNINGS ANDCAUTIONARYSTATEMENTS
3
I
NTRODUCTION
4
T
HEORY OFOPERATION
4
U
NPACKINGTHEINSTRUMENT
6
I
NSTRUMENTOPERATION
8 Turning the VX500 On and Off 8 VX500 Gas Reading Mode 9
C
ONFIGURATION
I
CONS
9
V
IEWING THE
VX500 O
PERATINGMODES
10 Changing Gas Type 10 User Identification 10 Site Identification 11 Zeroing and Calibration 11 Peak Readings 11 Snapshot Mode 12 TWA/STEL Readings 12 Tick Mode 12 Datalogging Session Indicator 13 Datalog Session Pause 13 Reset Datalog Session 13 Clear Datalogging Memory 13
C
ALIBRATING THE
VX500 14
C
ONFIGURING THE
VX500’SC
USTOMSETTINGS
15 Security Code 15 Change Security Code 16 Change Gas Data 16 Edit Favorite Response Factors List 16 Using the Response Factors List 16 Edit User Identification List 17 Edit Site Identification List 17 Allow Viewing of Peak Reading in the Field 18 STEL/TWA Configuration 18 Allow Viewing of Ticks Screen in the Field 18 Select Backlight Mode 19 Allow Viewing of Site and User in Field 19 Allow Selection of Gas in Field 19 Change Alarm Options 19 Edit Time and Date 20 Datalog Configuration 20 Calibration Configuration 21
M
AINTENANCE
22 Charging the Batteries 22 Changing the Internal Filter 22 Cleaning and Replacing the PID Lamp 23 Lamp Removal 24 Lamp Cleaning Procedure 24
S
PECIFICATIONS
26
I
ONIZATIONPOTENTIALS FORCOMMONINDUSTRIALGASES
27
VX500 R
ESPONSEFACTORSLIST
32
E
XPLODEDVIEW
& R
EPLACEMENTPARTS
35
VX500 R
EPLACEMENTPARTSLIST
39
W
ARRANTY
41
TABLE OF CONTENTS
P
HOTO
I
ONIZATION
D
ETECTOR
54
I
NTRODUCTION
PhotoIonization detectors (PIDs) are a proven and reliable means of detecting volatile organic compounds (VOCs) in a variety of applications. Although the PID in general is not capable of distinguishing one gas species from another, it is a useful tool for detecting potentially hazardous gas vapors such as benzene, toluene, xylene and hundreds of other compounds at parts per million (ppm) or sub-ppm levels. The low resolution of the PID makes this instrument ideal for detecting traces of hydrocarbon gases in confined spaces, hazardous materials sites, areas under arson investigation, locations prone to releasing fugitive emissions or in environments where prolonged gas exposures may present a threat to workers’health and safety.
When used within the guidelines set forth in this manual the VX500 PhotoIonization Detector can be relied on to provide years of dependable service. These instructions will guide you through the set-up, operation, maintenance and calibration necessary for you to be assured that your VX500 is operating properly.
T
HEORY OFOPERATION
The VX500 PhotoIonization Detector operates on the principle of absorption of ultraviolet light energy by a target gas. The internal sampling pump of the VX500 draws a gas stream into the instrument’s ionization chamber. This chamber holds an ultraviolet light source with a known energy potential. This potential is referred to as the photon energy of the lamp. If the energy from the lamp is at a level greater than the amount of energy required to excite the gas (the ionization potential) it will cause it to ionize, or release electrons. These electrons gather on the electrodes within the PID module and the instrument’s electronics interpret them as a quantitative concentration of gas. The number of electrons released is proportional to the concentration of the gas in the chamber.
For example: A PID with an energy source or lamp with a
10.6 electron-volt (eV) photon energy is capable of detecting any compound which has an ionization potential less than or equal to 10.6 eV. Once the gas exits the ionization chamber it reclaims its lost electrons, returns to its original state and is returned to the atmosphere.
Display
Gas In
Gas Out
(No Longer Ionized)
Gas Out
(No Longer Ionized)
UV Lamp
100 PPM
Amplifier
Electrode Electrode
UV Lamp’s Photon Energy is expressed in eV or Electron Volts
Graphic representation of
PID theory of operation
ENTER
ON/OFF
MODE
PPM VOC
PHOTOIONIZATION DETECTOR
76
Gas Sample Overflow
Visual Alarm
Gas Inlet
Lamp Cover
Charger/ Datalink Port
Light Sensor for Display Backlighting
Display
Up or “+”
Ionized Sample Output
Battery Cover
FRONT
BACK
U
NPACKING THE
I
NSTRUMENT
The shipping box should contain the following items. Account for each item before discarding the box.
QUANTITY
PART NUMBER DESCRIPTION
1 18104034 VX500 PID Monitor 1 17098773 VX500 Instruction Manual 1 17095746 Maintenance Tool 1 17104407 Start-Up Card 1 18104364 Nylon Carrying Case 1 17065970 Urethane Tubing for
Calibration
After unpacking, if any listed item is missing, contact either your local distributor of Industrial Scientific products, or call Industrial Scientific Corporation at 1-800-DETECTS (338-3287) in the United States and Canada, or 412-788-4353.
Enter/ backlight
i-Button
®
Contact
Audible Alarm
Down or “-”
On/ Off Mode
External Alarm Socket
BOTTOM
98
I
NSTRUMENT
O
PERATION
TURNING THE
VX500 ON AND OFF
To turn the VX500 on, press and hold on the unit’s membrane keypad until the instrument emits a short beep and the VX500 welcome screen appears on the instrument display. The current revision of instrument operating software is shown below the instrument model name. Pressing while this screen is shown will rotate the display 180 degrees to make viewing easier in applications where the VX500 may be hung upside-down.
After the welcome screen, the VX500 configuration screen will be displayed. This screen shows all parameters currently set in the instrument including the calibration gas concentration and all current alarm level settings.
Following the configuration screen, the instrument will display a brief countdown timer while the VX500 completes the automatic warm-up cycle. During this countdown, if and are pressed simultaneously, the VX500 will enter into the configuration mode giving you the opportunity to change the instrument’s custom settings. If no keys are pressed during warm-up, the display will time out and begin detecting gas in the normal operation mode.
To turn the VX500 off, press and hold at any time during operation until the instrument display shows RELEASE. After releasing the instrument’s display will blank and all instrument operation will cease.
VX500 GAS READING MODE
Once the VX500 enters the Gas Reading Mode the instrument display will show the concentration of gas currently being detected along with the corresponding gas type. For general VOC detection this gas type will be isobutylene. You may select another gas in the configuration mode by editing the “favorites” list (see page 16). Gases selected from the favorites list will use response factors to determine actual gas concentrations. These response factors are referenced to isobutylene unless the VX500 is directly calibrated to the specific gas.
If the monitored concentration of gas exceeds a preset limit for low alarm, high alarm, STEL, or TWA the VX500 will alert you by sounding an audible tone in conjunction with flashing an ultra bright red LED. Alarm conditions are also activated whenever a sensor fails calibration, the pump fails or becomes blocked, or the battery power is too low to operate the VX500.
CONFIGURATION ICONS
The top line of the instrument display will show a series of icons which identify the instrument’s current operating configuration. These icons are identified below and are explained in greater detail in later sections of this manual.
Datalogging On
Battery Type and charge status (Shading indicates charge status) R = rechargeable A = alkaline
Tick Mode
Snapshot Mode
ENTER
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
Rev1.0
E To Rotate 180
ISOBUTYLENE Ratio = 1.00 Cal 100ppm Low 100ppm High 200ppm STEL 300ppm TWA 8H 1000ppm
9
0.1ppm
Direct Cal
ISOBUTYLENE
R
11
SITE ID:
Press until "Site ID" appears. The display will show the current site id that is entered into the instrument. If "Change Site and User ID in the Field" is enabled in the VX500 configuration modes, pressing enter will allow you to change to current site. If enter is pressed, the VX500 will look for the presence of an iButton on the iButton contact. If an iButton is detected, the site information will automatically be loaded into the instrument. If no iButton is detected, the user is prompted to select a site from the site id list, or manually enter the site name using the arrow keys and .
ZEROING AND
CALIBRATION:
If field calibration is enabled in the set up menus, press until the Calibration Screen is reached. This screen shows you the last/next time calibration has occurred/is to occur. Pressing will enter the instrument into the zeroing sequence. Once zeroing is complete, pressing the key will allow the user to perform a calibration. (see Calibrating the VX500). If no buttons are pressed after the zeroing sequence, the instrument will return to normal operation.
PEAK READINGS:
If the Peak Hold feature is enabled, press until "Peak" appears. The peaking reading displayed represents the highest VOC concentration measured since the peak reading was last cleared. In the instrument configuration menus, if clearing peaks in the field is enabled, pressing the key will clear the peaks. If this mode is not enabled, the peaks will not be able to be cleared.
10
CHANGING G
AS TYPE:
If changing gas type in the field is enabled, pressing one time during the normal operating screen will advance the instrument into the change gas type screen. This operating mode allows the user to select a response factor from either the Favorites List or the entire list of preprogrammed response factors. Using the arrow keys, scroll until the desired gas type is highlighted. Once highlighted, press the enter key. The instrument will enter into a gas set up screen where the calibration gas and alarm set points can be changed. Use the arrow keys to scroll to the desire field to be modified. Once highlighted, press to highlight the value. Using the arrow keys increase or decrease the value until desired setting is reached. Press to accept the new value. Once all changes are made, press to enter the gas readings screen. The display will now show the new response factor that the VX500 is set to. For a complete list of available response factors, refer to the table on page XX.
U
SER ID:
Press until "User ID" appears. The display will show the current user id that is entered into the instrument. If "Change Site and User ID in the Field" is enabled in the configuration, pressing enter will allow you to change to current user. If enter is pressed, the VX500 will look for the presence of an iButton on the iButton contact. If an iButton is detected, the user information will automatically be loaded into the instrument. If no iButton is detected, the user is prompted to select a user from the user id list, or manually enter the user name using the arrow keys and
.
User ID
E To Change
M To Continue
scan iButton
V
IEWING THE
VX500 O
PERATINGMODES
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
Mike Labans
Site ID
E To Change
M To Continue
scan iButton
Vessel 63
Next Cal Date
06/26/2003
Calibrate
Instrument? Press E To Zero Press M To Skip
ZEROING
Peak
0.0ppm
ISOBUTYLENE
Press E To Clear
R
13
12
DATALOGGING SESSION INDICATOR
If the Datalogging feature is enabled press until the “Datalog Session” screen appears. The datalogging session indicator displays the elapsed time that has passed since the datalogging session was last reset along with the current date and time. At one minute recording intervals the VX500 is capable of storing up to 150 hours of gas readings.
DATALOG SESSION PAUSE
If the Datalogging Pause feature is enabled, press until the “Datalog Session Enabled/Paused” screen appears. Pressing during this mode will toggle the datalogger from paused to enabled, allowing the user to control when the VX500 is datalogging. When paused, the datalogging icon flashes.
RESET DATALOGGING SESSION
If the Datalogging feature is enabled, press until the “Reset Datalog Session” screen appears. Pressing during the mode will start a new datalogging session and reset the elapsed session timer to 00:00. Pressing one time from the datalogging session indicator will return the instrument to the normal operating screen.
CLEAR DATALOGGING MEMORY:
If the Clear Datalogging feature is enabled, press until "Clear All Data" appears. Pressing the during this mode will clear all of the current datalogging memory. Pressing will return the instrument to the normal operating screen.
SNAPSHOT MODE:
If the Snapshot mode is enabled, press until the Snapshot menu screen appears. The Snapshot operating mode allows the user to capture the current instrument reading and store it in the datalogging memory along with the current time and date stamp. Pressing at any time while this mode is displayed will automatically result in the reading, time and date to be recorded into the instrument memory. When datalogging is turned off, the snapshot mode can be enabled. For more information on setting up the snapshot mode, refer to the Datalog Configuration section on page 20.
TWA/STEL READINGS
If the STEL and TWA features are enabled, press until the TWA/STEL screen appears. This screen will display the current 8-hour time weighted average (TWA) and 15 minute short term exposure limit (STEL) average readings measured since these readings were last cleared and reset. Pressing while in this mode will reset the current TWA and STEL readings to zero. Resetting the STEL and TWA readings will automatically create a new datalogging session if the VX500’s datalogging feature is enabled.
TICK MODE
If the Tick mode feature is enabled, press until the tick screen appears. With the tick mode enabled the VX500 will sound an audilble tick which increases in rate as the gas concentration detected is increased. This function may be useful in tracking leaking gas concentrations or while detecting fugitive emissions.
ENTER
ENTER
ENTER
ENTER
ENTER
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
ON/OFF
MODE
6/20/2002 08:22
+25.3 C
Datalog Session
Elapsed Time
00:01
Datalog Session
Press E To Reset
Datalog Session
> Enabled
E To Change
TWA
0
STEL
0
E to reset both
M to Continue
0.1ppm
& to Change
Tick Rate
E to Accept
R
0ppm
0 ppm 0 ppm E To Update
R
15
14
Gas detection instruments are potential life saving devices. Recognizing this fact, Industrial Scientific Corporation recommends that a functional “bump” test be performed on every instrument prior to each days use. A functional test is defined as a brief exposure of the monitor to a concentration of gas(es) in excess of the lowest alarm set­point for the sensor for the purpose of verifying sensor and alarm operation and is not intended to be a measure of the accuracy of the instrument.
Industrial Scientific Corporation further recommends that a full instrument calibration be performed using a certified concentration(s) of Industrial Scientific branded calibration gas(es) monthly to ensure maximum accuracy.
Use of calibration gases from manufacturers other than Industrial Scientific may void product warranties and limit liability claims against the manufacturer.
If an instrument fails to operate properly following any functional “bump” test, a full instrument calibration should be performed prior to use.
To calibrate the VX500 press in the Gas Reading mode once to advance to the calibration screen. Press to begin the zeroing process. During the zeroing process, the words “Zero in progress” appear on the display. When the zeroing process is complete, the instrument will display “Apply XXXPPM” of the gas selected to be monitored. If you wish to perform the span calibration, apply the calibration gas thru a demand flow regulator and press to begin calibration. The VX500 will automatically recognize the presence of the calibration gas and the instrument will be calibrated automatically when the sensor response to the calibration gas becomes stable. Upon completion of calibration the VX500 will display a full span value and calibration status. A full span value that is greater than 70% of the applied gas concentration is considered good. A full span value that falls between 70% and 50% of the applied gas value is considered marginal.
The VX500 has many user configurable options and features. These features may be accessed through the configuration software utility and a PC, or adjusted by pressing and simultaneously during the warm-up cycle when the count down timer is displayed.
Configuring your VX500 is very intuitive. Every option is highlighted on the display. In general:
Pressing or will step you through the modes. Pressing selects a function or accepts (saves) a value. Pressing moves you backward in the configuration or
steps you completely out of set up to the “Gas Reading Mode.”
SECURITY CODE
Once the configuration mode has been entered, the instrument will prompt you to enter a three digit security code access code (if this feature is enabled). If the security code feature is activated on the VX500, no configuration changes can be made without entering the proper code. The security code is entered by using and to scroll through the values. When the desired security code is reached, press to accept the value. The VX500's display will step to the next availble configuration screen if the proper code has been entered.
ZERO
COMPLETE
Apply Cal Gas
To Instrument Press E to Span Press M to Skip
Instrument
is Calibrating
Calibration
Complete
62 Span
ENTER
ENTER
ON/OFF
MODE
ENTER
ENTER
ON/OFF
MODE
C
ALIBRATING THE
VX500
C
ONFIGURING THE
VX500’SC
USTOMSETTINGS
9
Enter Security
Code Using ▲ &
OO
E To Accept
M to Exit
O
A full span value below 50% of the applied gas value will fail calibration. For marginal or failed calibration the PID lamp may need to be cleaned (refer to Cleaning and Replacing the PID Lamp). If a VX500 fails either zero or span functions, the VX500 notes this condition on the display and the unit will not operate until the condition is corrected.
17
16
EDIT USER IDENTIFICATION LIST
The
Edit User Identification List” screen allows you to manually enter instrument user information with up to 16 alpha/numeric characters. This list can then be used to imprint the datalogger with the current user as selected in the User ID screen in the Gas Readings Mode.
Press to modify the user ID. allows changes to be made to the user list while allows changes to the current user. When making changes to the user list, use the arrow keys to scroll through current users to select a name to delete. When prompted, pressing allows users to be added to the current user lists. User names can be added automatically via an iButton, or manually entered using the arrow keys to select letters and key to accept. A maximum of 5 users can be stored on the current user list.
When making changes to the current user, the VX500 will initially look for an iButton. If an iButton is not detected on the iButton reader, the VX500 will allow a new user name to be manually entered using the arrow keys to select letters and key to accept, or by using the iButton.
EDIT SITE IDENTIFICATION LIST
The“Edit Site Identification List” screen allows you to manually enter instrument site (location) information with up to 16 alpha/numeric characters. This list can be used to imprint the datalogger with the current location as selected in the Site ID screen in the Gas Readings Mode.
Press to modify the site ID. allows changes to be made to the site list while allows changes to the current site. When making changes to the site list, use the arrow keys to scroll through current sites to select a site to delete. When prompted, pressing allows sites to be
CHANGE SECURITY CODE
The “Security Features” screen allows you to establish a security code to protect your custom configurations. The security code is a three digit number that once established cannot be bypassed unless entered correctly. It is important that you record your security code and keep it in a safe location in the event you need recall the number.
CHANGE G
AS DATA
The “Change Gas Data” screen allows you to edit gas specific setpoints such as calibration gas values and low, high, STEL, and TWA gas concentration alarm values. This screen also allows you to configure the gas range giving you control over whether your VX500 reads in tenths of a ppm or whole ppm’s.
E
DIT FAVORITE
R
ESPONSE FACTORS LIST
The “Edit Favorites Response Factors List” screen allows you to add/subtract gases on your favorite list of response factors. Response factors allow you to accurately monitor a gas, other than the gas the VX500 was calibrated to. Response factors are preprogrammed into the VX500 and are created by comparing the PID module’s response to the target gas versus the previously used calibration gas.
USING THE RESPONSE FACTORS LIST
Once selected, the response factor will automatically do the math and reference the signal to the specific compound of choice. For example, if you calibrated your VX500 using isobutylene and you wish to accurately monitor benzene, you would press and in the Gas Reading Mode, then choose between the Favorites List or the list of 100 preprogrammed compounds until you highlight benzene. Pressing here would select benzene and automatically use the associated response factor. For a list of available compound specific response factors refer to page 32 of this manual.
Change
Security Code
E to configure
for next menufor prev menu
M to exit config
Change
Gas Data
E to configure
for next menufor prev menu
M to exit config
Edit Favorite
Response
Factors List
E to configure
for next menufor prev menu
M to exit config
Edit User
Identification
List
E to configure
for next menufor prev menu
M to exit config
Edit Site
Identification
List
E to configure
for next menufor prev menu
M to exit config
ENTER
ENTER
ENTER
ENTER
ENTER
18 19
the tick feature of the VX500. The ticks mode sounds an audible tick that increases in frequency as the concentration of VOC’s detected increases. This is a useful tool when using the VX500 to detect leaks.
SELECT BACKLIGHT MODE
The “Select Backlight Mode” screen allows you to control whether your display backlight will be manually or automatically activated. If you choose manual, your backlight will only activate whenever is pressed. Choosing automatic backlight control will activate the display backlight whenever ambient light conditions diminish. This is accomplished through the use of a light sensor located on the VX500’s faceplate.
ALLOW VIEWING OF SITE AND USER IN FIELD
The “Allow Viewing of Site and User in Field” screen gives you control over whether the Site and User ID screens will be viewed in the field.
ALLOW SELECTION OF GAS IN FIELD
The “Allow Selection of Gas in Field” screen gives you control over whether the user will be able to access the “Favorites Response Factors” list from the Gas Readings Mode. Using response factors is explained in the “Edit Favorite Response Factors List” section of this manual.
CHANGE ALARM OPTIONS:
The change alarm options menu allows control over muting the alarms and latching the alarms. The audible alarm on the VX500 can be turned on and off. Using the arrow keys to toggle this feature between "on" and "off".
added to the current site lists. Site names can be added automatically via an iButton, or manually entered using the arrow keys to select letters and key to accept. A maximum of 5 sites can be stored on the current site list.
When making changes to the current Site, the VX500 will initially look for an iButton. If an iButton is not detected on the iButton reader, the VX500 will allow a new site name to be manually entered using the arrow keys to select letters and key to accept, or by using the iButton.
ALLOW VIEWING OF PEAK READING IN
THE FIELD
The “Allow Viewing of Peak Reading in the Field” screen gives you control over whether the peak (highest concentration of VOC’s monitored since the peaks were last reset) readings will be viewed by users in the field. Choosing “no” blocks this display for the Gas Readings Mode while “yes” permits it to be viewed.
STEL/TWA CONFIGURATION:
Pressing allows the STEL/TWA to be configured. This mode allows the user to have control over displaying and reseting the STEL (short term exposure limit: a 15 minute running average of VOC concentrations) and TWA (time weighted average: a user defined average, usually 8 hours, of VOC concentrations) values in the field. The arrow keys toggle between turning these modes on and off, while the
key accept the changes. The TWA time base can be changed using the arrow keys to increase or decrease the time base value and using to accept the new value.
ALLOW VIEWING OF TICKS SCREEN IN
THE FIELD
The “Allow Viewing of Ticks Screen in the Field” screen gives you control over whether the ticks mode screen will be viewed in the field. Choosing “no” in this situation blocks the user for having control over turning on or off
Allow Viewing
of Peak Reading
in the Field
E to Configure
for next menufor prev menu
M to exit config
Allow Viewing of STEL/TWA
in the Field
E to Configure
for next menufor prev menu
M to exit config
Allow Viewing
of Ticks Screen
in the Field
E to Change
for next menufor prev menu
M to exit config
Select
Backlight Mode
E to Change
for next menufor prev menu
M to exit config
Allow Viewing of
Site and User
in Field
E to Change
for next menufor prev menu
M to exit config
Allow Selection
of Gas in Field
E to Configure
for next menufor prev menu
M to exit config
Change Alarm Options
E to Configure
for next menufor prev menu
M to exit config
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
2120
When the alarms are muted, a message will be displayed across the screen of the VX500 to alert the user that the alarms are off. The alarm latching option allows the VX500's alarms to be latching or non-latching. A latching alarm alarm is an alarm that once activated, is not automatically cleared when the gas concentration drops below the alarm setpoint. To clear a latching alarm you must press once the gas concentration drops below the alarm setpoint.
EDIT
TIME AND DATE
The “Edit Time and Date” screen allows you to set the current time and date in your VX500. Accurate setting of the time and date are important for datalogging and calibration records. Use the arrow keys to increase or decrease the values for the time and date. Once a desired value is reached, use to move to the next value to be changed.
DATALOG CONFIGURATION
The “Datalog Configuration” screen gives you control over the operation of your datalogger. If you choose to make your datalogger inactive then the VX500 continues on with the custom configuration. If you choose to activate the datalogger, then a subloop appears giving you control over the following:
• Datalog Interval: Allows you to control the amount of
time (in seconds) the unit will accumulate data before averaging and writing it to the datalogger. The range is 1 to 300 seconds.
• Allow Datalog Reset in Field:Allows you to reset a
logging session in the field.
• Allow Datalog Pause in Field: Allows you to
pause/resume datalogging in the field.
• Allow Datalog Clear in Field: Allows you to purge
logged data in the field.
If datalogging is not activated, the VX500 can have the "log on alarm" feature activated or be put into the Snapshot mode. The log on alarm feature allows the VX500 to automatically start logging gas readings whenever the instrument goes into an alarm condition. This feature is useful in capturing data only when the VX500 is in alarm.
The Snapshot operating mode allows the user to capture the current instrument reading and store it in the datalogging memory instantly along with the current time and date. Pressing at anytime while this mode is active will result in a reading, time and date to be recorded into the instrument's memory.
CALIBRATION CONFIGURATION:
The "Calibration Configuration" screen gives the user control over the calibration of the VX500. When entered, the following sub-menus are availble:
• Allow Calibration in Field: The “Allow Calibration in
Field” screen gives you control over whether the VX500 can be calibrated in the field
• Vielw Calibration Date in Field: The “View
Calibration Date in Field” screen allows you to choose whether last or due calibration dates are able to be viewed in the field.
Choose Next/Last Calibration Date: The “Choose
Next/Last Calibration Date” screen allows you to choose whether your calibration dates are expressed as the last time the VX500 was calibrated or when the calibration is due. You have two options for this screen: last and next.
Edit Interval of Calibration: The “Edit Interval of
Calibration” screen activates if you chose to display the calibration date in field. This screen allows you to choose the duration. The range is from 0-365 days.
Datalog
Configuration
E to Configure
for next menufor prev menu
M to exit config
Datalog
Active
or change
E to accept
M to exit
>Yes
Allow
Calibration
In Field
E to Configure
for next menufor prev menu
M to exit config
View Calibration
Date In Field
E to Configure
for next menufor prev menu
M to exit config
Choose
next/last
Calibration Date
E to Configure
for next menufor prev menu
M to exit config
ENTER
ENTER
ENTER
Edit Time and Date
E to Configure
for next menufor prev menu
M to exit config
2322
With just normal routine maintenance the VX500 can be depended upon to provide years of reliable service. The following guidelines should be followed when performing maintenance on the VX500.
CHARGING THE
BATTERIES
The lithium-ion battery pack should be fully charged before using the VX500. To charge the battery pack, plug the flying lead from the VX500 battery charger into the socket on the back of the instrument. The VX500 battery pack will be fully charged within 5 hours. With a fully charged lithium-ion battery pack, the VX500 should function for up to 18 hours of continuous operation. As battery life decreases, the shaded area of the battery icon will also decrease. With a minimum of 15 minutes of battery life remaining the VX500 will emit a periodic beep alerting you to charge or replace the battery.
A replaceable cell alkaline battery pack (7 hours run-time) is also available for use with the VX500. To remove the battery pack and replace the 3 AA battery cells, loosen the two screws from the battery cover on the back of the instrument. Remove the battery pack and replace the AA battery cells with fresh alkaline batteries. Replace the battery pack and battery cover in the same fashion.
CHANGING THE INTERNAL FILTER
The VX500 sample inlet is protected by an internal dust filter/water stop. When this filter becomes obstructed the VX500 will display a PUMP FAULT condition and the alarm will sound continuously to indicate that inadequate
!
CAUTION
DO NOT REPLACE OR CHARGE INSTRUMENT
BATTERIES IN HAZARDOUS LOCATIONS!
M
AINTENANCE
sample flow is being delivered to the instrument. To replace the filter, unscrew the sample inlet housing from the top of the instrument. Remove the dust filter/water stop and replace it with a new, fresh one. During replacement, ensure that the filter is inserted with the clear end down. Replace and tighten the sample inlet fitting on the top end of the VX500. If the VX500 still shows PUMP FAULT, the filter is not properly seated, the inlet cap is not tight, or the pump has failed.
CLEANING AND
REPLACING THE PID LAMP
The function of the PID’s detector is dependent upon exposure of the gas sample to UV light via a precision lamp. A critical component of this lamp is its wavelength filter, or window. This window is the portion of the lamp directly exposed to the gas sample, and is therefore potentially exposed to water vapor, dirt, debris and oil residues. Because this lamp is used as a UV light source, it is critical to understand that degradation of the lamp’s output will cause the detector to lose sensitivity, and in extreme cases can limit the range of the detector. In order to retain peak performance of the VX500 PID detector, it is vital that the unit’s UV lamp window remain free of dirt debris and oil residues. When the window is contaminated you must periodically remove these contaminants. Refer to Figure 1 detailing the lamp’s window.
The first step to preventing or decreasing lamp window contamination is repetitive replacement of the VX500’s sampling filter (part number 17058157). Also be careful when storing a spare lamp to avoid finger contact with the window and do not attempt to clean the lamp with compressed air (unless air source is oil and contaminant free). It is virtually impossible to prevent all factors that obstruct or degrade the UV lamp’s window surface, therefore as a recommendation a 10.6 eV lamp should be cleaned after every 40 hours of use, at any time when readings are suspect, or if a calibration failure occurs.
25
24
Industrial Scientific Corporation recommends an abrasive lamp cleaning, which will restore the lamp’s window to a pristine condition. A lamp cleaning kit is available (part number 17090721). Because the output of a freshly cleaned lamp is not fully stable until after a 24 hour burn­in, please use a VX500, or one of the built in lamp conditioners contained in an Industrial Scientific charger or Docking Station™ to burn-in the lamp. If the unit is calibrated with a freshly cleaned lamp, sensitivity of the detector will gradually decrease until the lamp’s output stabilizes. This should occur within 24 hours.
LAMP
REMOVAL
To remove the lamp from the VX500, unthread the unit’s lamp cap (cap is a right hand thread). After removing the lamp cap, grasp the bulb’s tubing grip, and pull the lamp straight out of its compartment. Inspect the lamp’s window for dirt, debris or oil residue. DO NOT TOUCH
THE LAMP’S WINDOW !!!
LAMP CLEANING PROCEDURE
To clean the lamp, grasp the lamp by its body, not its tubing grip. Place a grey abrasive pad from cleaning kit 17090721 on a flat surface. (The adhesive on the back of the gray abrasive pad may be exposed via removal of its protective covering to affix the pad if desired). Position the UV lamp on the abrasive surface of the pad so that the window is flat against the surface of the abrasive pad. Apply light pressure to the lamp. Turn the lamp and lift. Reseat the lamp in a fresh position on the pad, and repeat this process 10-12 times to fully clean the lamp.
To remove any abrasive residue from the lamp, grasp the lamp by its body again. Please use the brown felt cleaning pad from cleaning kit 17090721 on a flat surface, exactly like the abrasive pad. The adhesive on the back of the
brown cleaning pad may also be exposed via removal of its protective covering to affix the pad also if desired. As with the abrasive pad, position the UV lamp on the felt surface of the pad and make sure it is flat. Apply light pressure to the lamp and turn. Lift the lamp from the pad and reseat in a fresh position. Repeat this process 5-6 times.
The lamp should be burned-in for a 24 hour period after cleaning. The VX500 should then be recalibrated once the lamp is re-installed in the unit. During lamp installation, insure that the lamp is fully seated in the VX500.
S
PECIFICATIONS
S
IZE: (Maximum dimensions including case filter
extension): 10"L x 2.9"W x 1.9"H (253mm x 75mm x 50mm)
C
ASE: EMI/RFI static resistant, nickel plated,
composite case.
W
EIGHT: with Li-Ion rechargeable battery pack --
26 oz. (737 g) with AA disposable cell battery pack --
23.8 oz. (675.9 g)
D
ISPLAY
: 128 X 64 Graphic Dot-Matrix LCD with
built-in backlighting for low light conditions.
A
UDIBLE ALARM
OUTPUT: 90 dB typical @ 12 inches/ 1 foot. V
ISUAL ALARM: Two (2) Red Ultra-bright LED’s (Light
Emitting Diodes)
T
EMPERATURE RANGE
OF
OPERATION: -10 deg. C to 40 deg. C (14 deg. F to 104
deg. F)
H
UMIDITY RANGE
OF
OPERATION: 15 to 90% Relative Humidity
(noncondensing).
C
ONTINUOUS NON-ALARM RUN TIME
AT
ROOM
TEMPERATURE
Fully charged Li-Ion battery, in good condition 18 hours Three (3) fresh AA cell Alkaline batteries 7 hours Maximum recommended power down storage time for fully
charged Li-Ion before recharging < 50 days.
S
AMPLE PUMP GAS
FLOW RATE
1.0 SCFH (.5 LPM)
Many compounds not appearing in this list with an ionization potential of
10.6 eV or less may be detected. Other sources of ionization potential data:
CRC Handbook of Chemistry and Physics NIOSH Pocket Guide to Chemical Hazards HYPERLINK http://w
ebbook.nist.gov/
I
ONIZATIONPOTENTIALS FORCOMMONINDUSTRIALGASES
Acetaldehyde 10.22 Acetone 9.69 Acetylene 11.40 Acrolein 10.13 Allene 9.83 Allyl Alcohol 9.63 Allyl Chloride 10.05 Aminoethanol 9.87 2-Amino Pyridine 8.00 Ammonia 10.18 Aniline 7.70 Arsine 9.89 Benzaldehyde 9.53 Benzene 9.24 Benzenethiol 8.33 Bromobenzene 8.98 1-Bromobutane 10.13 2-Bromobutane 9.98 1-Bromobutanone 9.54 1-Bromo-2-Chloroethane 10.63 Bromoethane 10.28 Bromoethene 9.80 Bromoform 10.48 1-Bromo-3-Hexanone 9.26 Bromomethane 10.53 Bromomethyl Ethyl Ether 10.08 1-Bromo-2-Methylpropane 10.09 2-Bromo-2-Methylpropane 9.89 1-Bromopentane 10.10 1-Bromopropane 10.18
2-Bromopropane 10.08 1-Bromo8propene 9.30 2-Bromopropene 10.06 3-Bromopropene 9.70 2-Bromothiophene 8.63 o-Bromotoluene 8.79 m-Bromotoluene 8.81 p-Bromotoluene 8.67 1,3-Butadiene 9.07 2,3-Butadione 9.23 n-Butanal 9.83 s-Butanal 9.73 n-Butanol 10.04 s-Butanol 10.23 t-Butanol 10.25 2-Butanone 9.54 1-Butene 9.58 cis-2-Butene 9.10 3trans-2-Butene 9.13 n-Butyl Acetate 10.00 s-Butyl Acetate 9.91 t-Butyl Acetate 9.90 n-Butyl Alcohol 10.04 n-Butylamine 8.71 s-Butylamine 8.70 t-butylamine 8.64 n-Butylbenzene 8.69 t-Butylbenzene 8.68 Butyl Cellusolve 8.68 n-Butyl Mercaptan 9.15
Chemical Name IP (eV)
Chemical Name IP (eV)
27
26
t-Butyl Mercaptan 9.03 p-tert-Butyltoluene 8.28 1-Butyne 10.10 82-Butyne 9.85 n-Butyraldehyde 9.83 Carbon Disulfide 10.08 Chlorobenzene 9.07 1-Chlorobutanone 9.54 1-Chloro-2,3-Epoxypropane 10.60 Chloroethene 9.99 1-Chloro-2-Fluorobenzene 9.16 1-Chloro-3-Fluorobenzene 9.21 cis-1-Chloro-2-Fluoroethene 9.87 trans-1-Chloro-2-Fluoroethene 9.87 o-Chloroiodobenzene 8.35 Chloromethylethyl Ether 10.08 Chloromethylmethyl Ether 10.25 3-Chloropropene 10.04 2-Chlorothiophene 8.68 o-Chlorotoluene 8.83 m-Chlorotoluene 8.83 p-chlorotoluene 8.70 o-Cresol 8.93 m-Cresol 8.98 p-Cresol 8.97 Crotonaldehyde 9.73 Cumene 8.75 3-Cyanopropene 10.39 Cyclobutane 10.50 Cyclohexane 9.88 Cyclohexanol 10.00 Cyclohexanone 9.14 Cyclohexene 8.95 Cyclo-Octatetraene 7.99 Cyclopentadiene 8.56 Cyclopentane 10.52 Cyclopentanone 9.26 Cyclopentene 9.01
Cyclopropane 10.06 2-Decanone 9.40 Dibromochloromethane 10.59 1,1-Dibromoethane 10.19 Dibromomethane 10.49 1,2-Dibromopropane 10.26 Dibutylamine 7.69 1,2-Dichlorobenzene 9.07 cis-1,2-Dichloroethene 9.65 trans-1,2-Dichloroethene 9.66 1,1-Dichloropropanone 9.71 2,3-Dichloropropene 9.82 Dicyclopentadiene 7.74 Diethoxymethane 9.70 Diethylamine 8.01 Diethylamino Ethanol 8.58 Diethyl Ether 9.53 Diethyl Ketone 9.32 Diethyl Sulfide 8.43 1,2-Difluorobenzene 9.31 1,4-Difluorobenzene 9.15 Difluoromethylbenzene 9.45 Diiodomethane 9.34 Diisobutyl Ketone 9.04 Diisopropylamine 7.73 1,1-Dimethoxyethane 9.65 Dimethoxymethane 10.00 Dymethylamine 8.24 Dimethylaniline 7.14 2,3-Dimethylbutadiene 8.72 2,2-Dimethylbutane 10.06 2,3-Dimethylbutane 10.02 2,2-Dimethylbutan-3-one 9.18 3,3-Dimethylbutanone 9.17 2,3-Dimethyl-2-Butene 8.30 Dimethyl Disulfide 8.46 Dimethyl Ether 10.00 3,5-Dimethyl-4-Heptanone 9.04
Chemical Name IP (eV)
Chemical Name IP (eV)
1,1-Dimethylhydrazine 8.05 2,2-Dimethyl-3-Pentanone 8.98 2,2-Dimethylpropane 10.35 Dimethyl Sulfide (DMS) 8.69 Di-n-Propylamine 7.84 Di-n-Propyl Disulfide 8.27 Di-n-Propyl Ether 9.27 Di-i-Propyl Ether 9.20 Di-n-Propyl Sulfide 8.30 Epichlorohydrin 10.60 Ethanethiol (Ethyl Mercaptan) 9.29 Ethanolamine 9.87 Ethene (Ethylene) 10.52 Ethyl Acetate 10.01 Ethylamine 8.86 Ethyl Amyl Ketone 9.10 Ethylbenzene 8.76 Ethyl Bromide 10.29 Ethyl Butyl Ketone 9.02 Ethyl Chloroacetate 10.20 Ethyl Disulfide 8.27 Ethyl Ethanoate 10.10 Ethyl Ether 9.41 Ethylene Dibromide 9.45 Ethylene Oxide 10.56 Ethyl Iodide 9.33 Ethyl Isothiocyanate 9.14 Ethyl Methyl Sulfide 8.55 Ethyl Propanoate 10.00 Ethyl Trichloroacetate 10.44 mono-Fluorobenzene 9.20 mono-Fluoroethene 10.37 o-Fluorotoluene 8.92 m-Fluorotoluene 8.92 p-Fluorotoluene 8.79 Furan 8.89 n-Heptane 9.90 2-Heptanone 9.33
4-Heptanone 9.12 n-Hexane 10.18 2-Hexanone 9.44 1-Hexene 9.46 Hydrogen Selenide 9.88 Hydrogen Sulfide 10.46 Hydrogen Telluride 9.14 Iodobenzene 8.73 1-Iodobutane9.212-Iodobutane 9.09 Iodoethane (Ethyl Iodide) 9.33 Iodomethane (Methyl Iodide) 9.54 1-Iodo-2-Methylpropane 9.23 1-Iodopentane 9.19 1-Iodopropane 9.26 2-Iodopropane 9.17 o-Iodotoluene 8.62 m-Iodotoluene 8.61 p-Iodotoluene 8.50 Isoamyl Acetate 9.90 Isoamyl Alcohol 10.16 Isobutanol 10.12 Isobutyl Acetate 9.97 Isobutyl Alcohol 10.12 Isobutylamine 8.70 Isobutylbenzene 8.68 Isobutylene 9.43 Isobutyl Ethanoate 9.95 Isobutyl Formate 10.46 Isobutyl Mercaptan 9.12 Isobutyl Methanoate 10.46 Isobutyraldehyde 9.74 Isopentane 10.32 Isoprene 8.85 Isopropyl Acetate 9.95 Isopropyl Alcohol 10.10 Isopropylamine 8.72 Isopropylbenzene 8.75 Isopropyl Ether 9.20
Chemical Name IP (eV)
Chemical Name IP (eV)
29
28
Isovaleraldehyde 9.71 Ketene 9.61 Mesitylene 8.40 Mesityl Oxide 9.08 Methyl Acetate 10.27 Methylamine 8.97 Methyl Bromide 10.54 2-Methyl-1,3-Butadiene 8.85 2-Methylbutanal 9.71 2-Methylbutane 10.31 2-Methyl-1-Butene 9.12 3-Methyl-1-Butene 9.51 3-Methyl-2-Butene 8.67 Methyl n-Butyl Ketone 9.34 Methyl Butyrate 10.07 Methyl Chloroacetate 10.35 Methylcyclohexane 9.85 Methylcyclohexanol 9.80 Methylcyclohexanone 9.05 4-Methylcyclohexene 8.91 Methylcyclopropane 9.52 Methyl Dichloroacetate 10.44 Methyl Ethanoate 10.27 Methyl Ethyl Ketone 9.53 Methyl Ethyl Sulfide 8.55 2-Methyl Furan 8.39 Methyl Iodide 9.54 Methyl Isobutyl Ketone 9.28 Methyl Isobutyrate 9.98 Methyl Isopropyl Ketone 9.32 Methyl Mercaptan 9.44 Methyl Methacrylate 9.70 2-Methylpentane 10.12 3-Methylpentane 10.08 2-Methylpropanal 9.74 2-Methylpropane 10.56 2-Methylpropene 9.23 Methyl n-Propyl Ketone 9.39
Methyl Styrene 8.35 Napthalene 8.12 Nitric Oxide 9.27 Nitrobenzene 9.92 p-Nitrochlorobenzene 9.96 5-Nonanone 9.10 3-Octanone 9.19 4-Octanone 9.10 1-Octene 9.52 cis-1,3-Pentadiene 8.59 trans-1,3-Pentadiene 8.56 n-Pentanal 9.82 n-Pentane 10.34 2,4-Pentanedione 8.87 2-Pentanone 9.39 3-Pentanone 9.32 1-Pentene 9.50 Perfuoro-1-Heptene 10.48 n-Perfuoropropyl Iodide 10.36 n-Perfuoropropyl-Iodomethane 9.96 n-Perfuoropropyl-Methyl Ketone 10.58 Phenol 8.50 Phenyl Ether 8.09 Phenyl Isocyanate 8.77 Phosphine 9.96 Pinene 8.07 Propadiene 10.19 n-Propanal 9.95 1-Propanethiol (n-Propyl Mercaptan) 9.20 n-Propanol 10.51 Propanone 9.69 Propene 9.73 Prop-1-ene-2-ol 8.20 Prop-2-ene-1-ol 9.67 Propionaldehyde 9.98 n-Propyl Acetate 10.04 n-Propyl Alcohol 10.15 n-Propylamine 8.78
Chemical Name IP (eV)
Chemical Name IP (eV)
n-Propylbenzene 8.72 Propylene 9.73 Propylene Imine 9.00 Propylene Oxide 9.81 n-Propyl Ether 9.27 n-Propyl Formate 10.54 Propyne 10.36 Pyridine 9.27 Styrene 8.40 Tetrachloroethylene (PCE) 9.32 Tetrafluoroethene 10.12 Tetrahydrofuran 9.45 Thioethanol 9.29 Thiomethanol 9.44 Thiophene 8.86 1-Thiopropanol 9.20 Toluene 8.82 o-Toluidine 7.44 Tribromoethene 9.27 1,1,1-Trichlorobutanone 9.54 Trichloroethylene (TCE) 9.45 Trichloromethyl Ethyl Ether 10.08 Triethylamine 7.50
1,2,4-Trifluorobenzene 9.37 1,3,5-Trifluorobenzene 9.32 Trifluoroethene 10.14 1,1,1-Trifluoro-2-Iodoethane 10.10 Trifluoroiodomethane 10.40 Trifluoromethylbenzene 9.68 Trifluoromethylcyclohexane 10.46 Trimethylamine 7.82 2,2,4-Trimethyl Pentane 9.86 2,2,4-Trimethyl-3-Pentanone 8.82 n-Valeraldehyde 9.82 Vinyl Acetate 9.19 Vinyl Bromide 9.80 Vinyl Chloride 10.00 4-Vinylcyclohexene 8.93 Vinyl Ethanoate 9.19 Vinyl Fluoride 10.37 Vinyl Methyl Ether 8.93 o-Vinyl Toluene 8.20 o-Xylene 8.56 m-Xylene 8.56 p-Xylene 8.45 2,4-Xylidine 7.65
Chemical Name IP (eV)
Chemical Name IP (eV)
31
30
The response factors listed below in italics were derived from experimental data and are considered accurate +/- 25%. All other response factors are theoretical and no accuracy is published. All response factors are for the 10.6 eV lamp. Whenever possible, calibrate to the same gas being monitored.
33
32
VX500 R
ESPONSE
F
ACTORSLIST
1,4-BUTANEDIOL 1,4-BUTANEDIOL 37.20 1,4-DIOXANE 1,4-DIOXANE 1.48 124(CH3)C6H5 1,2,4-TRIMETHYLBENZENE 0.43 123(CH3)C6H5 1,2,3-TRIMETHYLBENZENE 0.49 12C2H2BR2 1,2-DIBROMOETHANE 11.66 12CL2C6H6 1,2-DICHLOROBENZENE 0.62 135(CH3)C6H5 1,3,5-TRIMETHYLBENZENE 0.34 1-BUTANOL 1-BUTANOL 6.02 1MTH02PROPOL 1-METHOXY-2-PROPANOL 1.85
1-PROPANOL 1-PROPANOL 13.85
1XACETATE METHYLACETATE 8.50
1XACRYLAC METHYLACRYLATE 7.16
1XACTOACETAT METHYLACETOACETATE 1.30 1XBENZOATE METHYLBENZOATE 0.93 1XMTHACRYLAT METHYMETHACRYLATE 2.20 2-BUTANONE 2-BUTANONE 0.90 2-METHYLFORMAMIDE DIMETHYLFORMAMIDE 1.10 2MTHOXYETOH 2-METHOXYETHANOL 3.64 2-PENTANONE 2-PENTANONE 0.87 2-PICOLINE 2-PICOLINE 0.72
2-PROPANOL 2-PROPANOL 12.52
2XFORMAMIDE N,N-DIMETHYLFORAMIDE 1.13 2XMTACETAMID N,N-DIMETHYLACETAMIDE 0.66 3-PICOLINE 3-PICOLINE 0.92 4HYD4MTH2PNT 4-HYDROXY-4-METHYL-2-PENTANONE 0.73 ACETALDEHYDE ACETALDEHYDE 11.00
ACETONE ACETONE 1.51
ACETOPHENONE ACETOPHENONE 0.59 ALL YL ALCOHOL ALL YL ALCOHOL 2.92 AMMONIA AMMONIA 16.62 AMYL ACETATE AMYL ACETATE 5.31
BENZENE BENZENE 0.45
Abbreviated Name Chemical Name Response Factor
BROMOMETHANE BROMOMETHANE 2.72
BUTADIENE BUTADIENE 0.73
BUTOXYETHANOL BUTOXYETHANOL 1.44 BUTYLACETATE BUTYLACETATE 8.03
C2CL4 TETRACHLOROETHYLENE 0.32 (C2H5)C6H5 ETHYLBENZENE 0.65 C2HCL3 TRICHLOROETHYLENE 0.30
C6H1003 ETHYLACETOACETATE 1.14 CHLOROBENZEN CHLOROBENZENE 0.49 CUMENE CUMENE 0.54
CYCLOHEXANE CYCLOHEXANE 1.93
CYCLOHEXANON CYCLOHEXANOE 0.82 DECANE DECANE 3.66 DIETHYLAMINE DIETHYLAMINE 0.89 DIMETHOXMETH DIMETHOXYMETHANE 11.27 EPICHLOROHYDRIN EPICHLOROHYDRIN 13.00
ETHANOL ETHANOL 23.59
ETHYGLYCOL ETHYLENEGLYCOL 15.30
ETHYLACETATE ETHYLACETATE 11.09 ETHYLENE ETHYLENE 20.27 ETHYLENE OXIDE ETHYLENE OXIDE 26.92
G-BUTYROLACTONE GAMMA BUTYROLACTONE 3.01 H2S HYDROGEN SULFIDE 6.25
HEPTANE HEPTANE 5.91
HEXANE HEXANE 6.23 HYDRAZINE HYDRAZINE 2.60 IAMYLACETATE ISOAMYLACETATE 6.06 IPROPYLAMINE ISOPROPYLAMINE 1.28
IPROPYLETHER ISOPROPYLETHER 0.93
ISOBUTANOL ISOBUTANOL 4.99 ISOBUTYLENE ISOBUTYLENE 1.00 ISOOCTANE ISOOCTANE 1.86 ISOPHORONE ISOPHORONE 0.74
ISOPROPANOL ISOPROPANOL 12.52
JET A FUEL JET A FUEL 3.00 JET A 1 FUEL JET A 1 FUEL 3.80 JP 5 & JP 8 JP 5 & JP 8 FUEL 2.90
Abbreviated Name Chemical Name Response Factor
35
34
FIGURE 4
MEK METHYL ETHYL KETONE 1.20
MESITYLOXIDE MESITYLOXIDE 0.54
MIBK METHYLISOBUTYLKETONE 1.14
MONOMETHYLAMINE MONOMETHYLAMINE 2.07
MTBE METHYLTERTIARYBUTYLETHER 1.08
MTHLBENZLALCOHOL METHYL BENZYL ALCOHOL 7.12 M-XYLENE META-XYLENE 0.45 N-METHLPYROLIDONE N-METHYL PYRROLIDONE 1.81 OCTANE OCTANE 2.71 O-XYLENE ORTHO-XYLENE 0.54 PHENLETHLALCOHOL PHENYL ETHYL ALOCHOL 9.04 PHENOL PHENOL 8.90 PHOSPHINE PHOSPHINE 3.02 PROPYLENE PROPYLENE 1.41 PROPYOX PROPYLENE OXIDE 16.02 P-XYLENE PARA-XYLENE 0.47 PYRIDINE PYRIDINE 0.78 QUINOLINE QUINOLINE 0.97 STYRENE STYRENEMONOMER 0.47 T-BUTYLAMINE TERTIARBUTYLAMINE 1.01 T-CLC2H2CL TRISDICHLOROETHENE 0.45 TERTBUTLMERCAPTAN TERT BUTYL MERCAPTAN 0.78 TERBUTYLALCOHOL TERTIARY BUTYL ALCOHOL 4.18 THF TETRAHYDROFURAN 2.06 THIOPHENE THIOPHENE 0.41
TOLUENE TOLUENE 0.53
TURPENTINE TURPENTINE PURE GUM SPIRITS 0.50 VCH VINYLCYCLOHEXONE 0.54 VINYL ACETATE VINYL ACETATE 1.40
VINYL CHLORIDE VINYL CHLORIDE 2.47
Abbreviated Name Chemical Name Response Factor
FIGURE 7
37
36
FIGURE 6
FIGURE 5
38
NOTE: Refer to Figure 4, 5, 6, 7 And 8 for each Item number listed.
ITEM PART NO. DESCRIPTION
1 17090408 Filter Cap (Includes Item 2, Filter) 2 17058157 Replacement Dust/Water Stop Filter 3 17098971 Case Bottom, Basic
(Includes Items 3 & 16 Must also order an Item 11 to re-use Items 8 - 10 in case bottom)
3 17098930 Case Bottom, Complete
(Includes Items 3 - 17)
4 17051513 Fitting, Inlet, Swivel, 1/16" tube to 10-32
thread. (two required per VX500). 5 17089269 O-ring Seal, Input/Output Connector 6 17098948 Input/Output Connector
(Includes Item 5) 7 17052628 Mounting Screws I/O connector
(two required per VX500) 8 17089319 C-clip, i
-Button
®
Contact
9 17089285 Seal, i
-Button®, Contact
10 17087735 Contact, i
-Button
®
11 17092727 Case Insulator, i-Button®, Contact 12 17089004 Gasket, Instrument, Conductive 13 17049876 Instrument Strap 14 17115205 Case Screws, Long, VX500
(two required per VX500) 15 17089095 Case Screws, Short, VX500
(four required per VX500) 16 17050245 Barrier, PID Vent 17 17099011 PID Seal/Barrier Kit 18 17091380 UV Lamp, 10.6 eV 19 17098823 Lamp Cap, Replacement, VX500 20 17088618 Battery Pack, Rechargeable, Lithium-Ion 20 17089376 Battery Pack, Alkaline Replaceable 21 17089012 Gasket, Battery Cover, Conductive 22 17098815 Battery Cover, Replacement, VX500
(Includes Items 22 - 26) 23 17089079 Battery Cover, Screw, VX500, Short 24 17099565 Battery Cover Screw Compression Stop,
Short
VX500 R
EPLACEMENT
P
ARTSLIST
39
FIGURE 8
W
ARRANTY
Industrial Scientific Corporation portable gas monitoring instruments are warranted to be free from defects in material and workmanship for as long as the instrument is in service.
The above warranty does not include sensors, battery packs, internal pumps or filters, all of which are warranted to be free from defects in material and workmanship for 18 months from the date of shipment, or 1 year from the date of first use, whichever occurs first, except where otherwise stated in writing in Industrial Scientific literature accompanying the product.
All other Industrial Scientific products are warranted to be free from defects in material and workmanship for a period of 18 months from the date of shipment, 1 year from the date of first use, whichever occurs first, except where otherwise stated in writing in Industrial Scientific literature accompanying the product.
LIMITATION OF LIABILITY
INDUSTRIAL SCIENTIFIC MAKES NO OTHER WARRANTIES, EITHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.
SHOULD THE PRODUCT FAIL TO CONFORM TO THE ABOVE WARRANTY, BUYER’S ONLY REMEDY AND INDUSTRIAL SCIENTIFIC’S ONLY OBLIGATION SHALL BE, AT INDUSTRIAL SCIENTIFIC’S SOLE OPTION, REPLACEMENT OR REPAIR OF SUCH NON-CONFORMING GOODS OR REFUND OF THE ORIGINAL PURCHASE PRICE OF THE NON-CONFORMING GOODS.
40
41
ITEM. PART NO. DESCRIPTION
25 17089053 Battery Cover, Screw, VX500, Long 26 17099581 Battery Cover Screw Compression Stop, Long 27 17050453 Screw, Pump/Display and Main PC Board Mount
(two each required for Pump/Display Mounting
(two each required for Main PC Board Mounting) 28 17082025 Main Board, VX500 with programmed EPROM 28 17094855 Programmed EPROM for VX500 (Not Shown) 29 17091141 Module, PID Detector 30 17095183 PID Flow Control Valve 31 17095233 Flow Control Mounting Bracket 32 17097122 PM-7000, VX500 Sampling Pump 32 17097304 PM-7000 Sampling Pump Repair Kit 33 17098831 Tubing, VX500 Replacement Assembly
(Not Shown - See Figure 8 for placement) 34 17086695 Left Display Mounting Bracket 35 17094673 Display Assembly 36 17084377 Right Display Mounting Bracket 37 17051845 Screws, PID Detector Mounting
(two required per VX500) 38 17098914 Case Top, Basic,VX500
(Includes Items 38 - 40, 44 & 45) 38 17098906 Case Top, Complete,VX500
(Includes Items 38 - 48) 39 17088998 Faceplate, VX500 40 17084435 Keypad, VX500 41 17028374 External/Vibrating Alarm Jack 42 17029273 External Alarm Cap Plug 43 17050277 O-ring Seal, External Alarm Jack 44 17058918 Alarm Water Barrier 45 17097296 Alarm Seal 46 17057118 Alarm 47 17057027 Alarm Retaining Clips (three required per VX500) 48 17083585 Screws, Alarm Retaining
(three required per VX500)
M
AINTENANCE ACCESSORIES:
17090721 Lamp Cleaning Kit
A
DDITIONAL ACCESSORIES:
17096348 ISC i
-Button®memory device with card mount
18104729 ISC i
-Button®programming kit with software, 120 VAC
18105080 ISC i
-Button®programming kit with software, 230 VAC
42
IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING LOSS OF PROFIT OR LOSS OF USE,ARISING OUT OF THE SALE, MANUFACTURE OR USE OF ANY PRODUCTS SOLD HEREUNDER WHETHER SUCH CLAIM IS PLEADED IN CONTRACT OR IN TORT, INCLUDING STRICT LIABILITY IN TORT.
It shall be an express condition to Industrial Scientific’s warranty that all products be carefully inspected for damage by Buyer upon receipt, be properly calibrated for Buyer’s particular use, and be used, repaired, and maintained in strict accordance with the instructions set forth in Industrial Scientific’s product literature. Repair or maintenance by non-qualified personnel will invalidate the warranty, as will the use of non-approved consumables or spare parts. As with any other sophisticated product, it is essential and a condition of Industrial Scientific’s warranty that all personnel using the products be fully acquainted with their use, capabilities and limitations as set forth in the applicable product literature.
Buyer acknowledges that it alone has determined the intended purpose and suitability of the goods purchased. It is expressly agreed by the parties that any technical or other advice given by Industrial Scientific with respect to the use of the goods or services is given without charge and at Buyer’s risk; therefore, Industrial Scientific assumes no obligations or liability for the advice given or results obtained.
Loading...