General Information .................................................................................................................................................................................................. 1
First Use ............................................................................................................................................................................................................... 9
Wearing the Instrument ........................................................................................................................................................................................ 9
Unpacking the Instrument ................................................................................................................................................................................... 11
Start-up and Shutdown ....................................................................................................................................................................................... 14
Instrument Preparation and Use ............................................................................................................................................................................. 17
Event Causes and Recommended Actions ........................................................................................................................................................ 32
Service and Warranty ............................................................................................................................................................................................. 37
Service Instructions ............................................................................................................................................................................................ 37
Service Tasks ................................................................................................................................................................................................. 40
Limitation of Liability ....................................................................................................................................................................................... 44
Figure 1. Probability of sensor failure by bump test frequency ................................................................................................................................. 9
Table 2.2 Securing the garment or belt clip .............................................................................................................................................................. 9
Table 4.3 Zero, calibration, and bump test ............................................................................................................................................................. 28
Table 5.3 Recommended action by event type ...................................................................................................................................................... 32
Figure 6.2 Disassembled Tango TX1 case top assembly ...................................................................................................................................... 38
Table 6.1 Key for the Tango TX1 diagram .............................................................................................................................................................. 39
Table 6.2 Service tasks ........................................................................................................................................................................................... 40
Table A.2. ATEX and IECEx marking requirements ............................................................................................................................................... 45
iv
1
General Information
Introduction
Certifications
Product Overview
Product Specifications
Introduction
C
ERTIFICATIONS
Each Tango TX1TM is certified by one or more certifying bodies (CBs). The approved uses for which a unit is certified
appear on label(s) affixed to the instrument.
When a new certification is received, it is not retroactive to any unit that does not bear the new marking on its label.
Instrument certifications at the time of this document's publication are noted below (see Table 1.1). To determine for
which uses a unit is certified, always refer to the unit's labels.
CBs issue warnings and cautionary statements to notify the safety team and instrument operators of important
information, or to restrict instrument use or service (see Table 1.2). Those items listed under the heading, "General",
are issued by more than one CB or by Industrial Scientific Corporation (ISC); these apply to each unit regardless of
its certifications. Additionally, those items listed under the heading of a specific CB apply to units that bear its
markings.
Table 1.1 Certifications
Directive or CB Area Classifications Approved Temperature Range
Americas
CSA Ex ia IIC T4; Class 1; Groups A, B, C, D; T4 -40 °C to +50 °C (-40 °F to +122 °F)
IECEx Ex ia I Ma
Ex ia IIC T4 Ga
INMETRO
UL (C-US) Class I, Groups A, B, C, and D;
Ex ia IIC T4 Ga
Ex ia I Ma
Class II, Groups E, F, and G; T4; Exia
Class I, Zone 0, AEx ia IIC T4
-40 °C to +50 °C (-40 °F to +122 °F)
-40 °C ≤ T amb ≤ +50 °C (-40 °F ≤ T amb ≤ +122 °F)
-40 °C to +50 °C (-40 °F to +122 °F)
General Information
Table 1.1 Certifications
Directive or CB Area Classifications Approved Temperature Range
Europe and Russia
ATEX Ex ia I Ma
Ex ia IIC T4 Ga
Equipment Groups and Categories: I M1 and II 1G
GOST EAC PO Ex ia I X
0 Ex ia IIC T4 X
Asia and Pacific
China Ex Ex ia IIC T4 Ga -20 °C to +50 °C (-4 °F to +122 °F)
KOSHA Ex ia IIC T4 -20°C ≤ Ta ≤ +50°C
Note: See the Appendix for ATEX and IECEx marking requirements.
-40 °C to +50 °C (-40 °F to +122 °F)
-40 °C to +50 °C (-40 °F to +122 °F)
Failure to perform certain procedures or note certain conditions may impair the performance of this product. For
maximum safety and optimal performance, please read and follow the procedures and conditions listed below.
Table 1.2 Warnings and cautionary statements
For maximum safety and optimal performance, read and understand the manual before operating or servicing the unit. Failure
to perform certain procedures or note certain conditions may impair the performance of this product.
For safety reasons, this equipment must be operated and serviced by qualified personnel only.
Substitution of components may impair intrinsic safety and may cause an unsafe condition.
Do not replace battery in hazardous locations. Only certified for use with one Tadiran TL-5955 battery cell.
Obstruction of sensor openings – due to dust, dirt, water, or another cause – can inhibit the unit’s ability to measure gas
concentrations accurately. When this occurs, readings may appear lower than the actual gas concentration. Keep sensor
openings clean, dry, and properly exposed to the ambient air.
Obstructed, contaminated, or damaged sensor water barriers (or their gaskets) can inhibit the unit’s ability to measure gas
concentrations accurately. When this occurs, readings may appear lower than the actual gas concentration. Replace the
sensor water barriers and gaskets as needed (see “Service” for instructions).
Service the unit, use its communications port, and change its battery cell only in nonhazardous locations. Not for use in
oxygen-enriched atmospheres.
Contact your service representative immediately if you suspect that the unit is working abnormally.
Product Overview
The Tango TX1 is a portable, long-life, single-gas monitor (instrument) for personal protection. It is a diffusion
instrument for use in detecting and measuring gas present in open space. Based on the customer’s order, two
redundant sensors – sensors of the same type – are factory installed. Four sensor types are available (
see Table 1.3).
2
Table 1.3 Sensor-type options
Product Overview
Sensor Category Number of sensors
available per
instrument
Toxic Two of the same type Carbon Monoxide (CO) only, Hydrogen Suflide (H
Note: See Table 1.5 for sensor specifications.
Sensor type
only, or Sulfur Dioxide ( SO
) only
2
S) only, Nitrogen Dioxide (NO2 )
2
When two working sensors are installed, they measure gas simultaneously and independently and the Tango TX1
operates on DualSense
TM
Technology from ISC. If only one sensor is installed or only one installed sensor is
operational, the unit operates as a single-sensor or non-DualSense unit.
If two working sensors are installed, the data log contains a data set for each sensor, plus a third data set. This third
data set – labeled "VIRTUAL" or "3" – is created by the DualSense technology. Each VIRTUAL data point is an
algorithm-calculated value that is based on sensor 1 and sensor 2 data. The VIRTUAL data are saved to and
downloaded from the data log just as data are for the installed sensors. VIRTUAL gas readings are displayed to the
instrument operator.
The Tango TX1 multiplane sensor ports each sample air from three directions; this promotes continuous operation if
one or two planes become obstructed.
Tango TX1 measures gas at two second intervals, and continuously logs dataevery ten seconds. The data log can
store approximately three months of data for a unit that is on 24 hours a day and has two installed, operational
sensors.As the newest data are logged to memory, the oldest data are overwritten. The data log's date- and timestamped event log records and stores event data for 60 alarm events and 30 error events. It also stores the data for
up to 250 manual calibrations and bump tests. The data log is downloaded when the unit is docked in a compatible
docking station.
The instrument has two modes: configuration and operation. When in configuration mode , a unit's settings can be
manually edited. Entry to configuration mode can be password protected. When the instrument is on and is not in
configuration mode, it is said to be in operation mode.
The instrument features an always-on option that can be enabled or disabled from configuration mode.
The Tango TX1 has a multisensory (audible, visual, and vibration), multilevel warning and alarm system. Warnings
indicate a service need (calibration due) or an operating condition (confidence indicator). Alarms indicate potentially
hazardous gas concentrations or system faults. Alarms can be set to latch. The instrument also features a country-of-origin option that automatically sets the values for the low-gas and high-gas alarm set points for each of five different
countries or regions; each alarm set point value can also be manually edited. An optional AlarmAmp
TM
is available;
when used, the audible alarm volume is increased by approximately 10 decibels (dB).
The user interface consists of two buttons and anLCD (liquid crystal display). The buttons are used to power on and
power off the instrument, navigate the operation and configuration loops, perform tasks, and access information.The
unit can be set to display select information in English or French.
The unit's suspender clip is intended for attachment to a garment; it is not intended for attachment to a belt or hard
hat. An optional belt clip is available from ISC (see "Service").
The Tango TX1 is iNet ready and compatible with the Tango TX1 iNet DS (docking station).
Product Specifications
Effective use of the Tango TX1 includes knowledge of the instrument's specifications and its sensor and battery
specifications (see Tables 1.4 through 1.6).
3
General Information
A
A
r
Table 1.4 Instrument specifications
Item Description
Display Segment LCD
Keypad buttons Two buttons
Case materials Case top: polycarbonate with a protective rubber over-mold
Case bottom: conductive polycarbonate
Alarms Three strobe-emitting visual alarm LEDs (two red; one blue)
100 dB audible alarm at a distance of 10 cm (3.94"), typical
Vibration alarm
Dimensions99 x 51 x 35 mm (3.9" x 2.0" x 1.4")
Weight 126 g (4.4 oz.), typical
Ingress protection IP66 and IP67
Operating temperature range
Operating humidity range 15 to 95% relative humidity (RH) noncondensing (continuous)
1
Operating temperatures above 50 °C (122 °F) may cause reduced instrument accuracy. Operating temperatures below -20 °C (-4 °F) may cause reduced
instrument accuracy and affect display and alarm performance.
1
-20 °C to +50 °C (-4 °F to +122 °F)
Table 1.5 Sensor specifications
1
2
Sensor
Measurement
Category
Gas Names
bbreviation
Sensor technology
Toxic
Carbon Monoxide
with H2 low
interference
CO/
Low
E
H2
Carbon Monoxide CO E 0-1000
Hydrogen Sulfide H2S E 0-200
Nitrogen Dioxide NO2 E 0-150
range
(% vol, % LEL, or ppm)degrees (°)(%)
0-1000
ppm
ppm
ppm
ppm
Sensor properties Accuracy Response
time
(typical)
2
ture and
Measurement
resolution
Sensor
temperature
range
1 ppm-20 to +50 °C
(-4 to +122 °F)
range
t time and
temperature of
calibration
(%) (%)
Sensor RH
15-95± 5% (0-300
ppm)
Over full sensor
± 15% 9 s18 s
tempe
RH ranges
T50T90
Seconds (s)
± 15% (301-1000 ppm)
1 ppm-40 to +50 °C
15–95 ± 5± 1512 s48 s
(-40 to +122 °F)
0.1 ppm-40 to +50 °C
15–95 ± 5± 157s14 s
(-40 to +122 °F)
0.1 ppm-40 to +50 °C
15–95± 10± 1510 s30 s
(-40 to +122 °F)
Sulfur Dioxide SO2 E 0 to 150
4
ppm
0.1 ppm -20 to +50 °C
(-4 to +122 °F)
15 – 95 ± 10± 15 20 s 80 s
Product Overview
1
"E" stands for electrochemical.
2
ppm stands for parts per million; vol for volume; and LEL for lower explosive limit. Note: See the Appendix for
supplemental information about sensor types and gases.
Table 1.6 Battery properties
Battery Pack Properties
3.6 V Primary Lithium-thionyl
chloride (Li-SOCl2), 1.5AH, 2/3AA
* See "Service" for instructions. Some restrictions may apply (see "Introduction", Table 1.2 Warnings and cautionary statements).
Replaceable*
Nonrechargeable
Run time of 36 months depending on operating conditions; the amount of time the unit is in
alarm; and the enablement of unit's confidence indicator, bump test due alert, or calibration
due alert.
5
6
Recommended Practices
Introduction
Procedures
First Use
Wearing the Instrument
Introduction
2
Gas detection instruments are potentially life-saving devices. When completed regularly, the procedures defined
below help to maintain proper instrument functionality and enhance operator safety.
Procedures
Configuration. The configuration process allows qualified personnel to review and adjust a unit's settings.
Self-test. The self-test verifies the functionality of the instrument’s memory operations, battery, and each alarm
indicator (audible, visual, and vibration).
Note: a self-test does not verify sensor functionality (see “Bump Test”) or instrument accuracy (see “Calibration”).
Bump Test (or "functional test"). Bump testing checks for sensor and alarm functionality. The installed sensors are
briefly exposed to an expected concentration of calibration gas that is greater than the sensors’ low alarm set point.
When one or more sensors “pass” the test, they are “functional” and the unit will alarm. Each sensor’s “pass” or “fail”
result is indicated on the unit’s display.
Note: a bump test does not measure for accuracy (see “Calibration”).
Zero. Zeroing sets each installed sensor to recognize the ambient air as clean air. If the ambient air is not truly clean
air, any gases that are present and relevant to the installed sensor types will be measured and displayed as zero.
Readings will be inaccurate until the unit is correctly zeroed in truly fresh air or with a zero air cylinder.
Calibration. All sensors gradually degrade over time. This diminishes a sensor's ability to measure gas
concentrations accurately; however, regular calibrations adjust the instrument to compensate for this decline in
sensitivity. During calibration, the installed sensors are exposed to an expected concentration of calibration gas and,
when needed, the instrument will self-adjust to ensure the accurate measurement and display of gas concentration
values.
Note: when a sensor has degraded beyond an acceptable level, no further adjustment is possible and the sensor will no longer pass
calibration.
Recommended Practices
Peak Readings. The instrument stores the highest detected gas readings, the "peak readings" or "peaks". Bump
testing and calibration will often register new peak readings. Therefore, the clearing of the peak readings should
follow each calibration. The instrument operator may also wish to clear the peak readings after a bump test, before a
change in location, or after an alarm is addressed and cleared.
Note: The peak readings and the data log readings are stored independently of one another; therefore, clearing the peak reading does not
affect the data log. Powering the instrument off or changing its battery does not affect the peak reading. These checks and balances help
promote operator safety, and serve to contain the peak readings in a "black-box" manner. In the event of a gas-related incident, this black-box
record can be useful to the safety team or a prospective investigator.
Recommendations
Industrial Scientific Corporation (ISC) minimum frequency recommendations for each procedure are summarized in
the table below. These recommendations are based on field data, safe work procedures, industry best practices, and
regulatory standards to help ensure worker safety. Industrial Scientific is not responsible for setting safety practices
and policies.
ISC recommendations include a daily bump test for any ISC instrument that is not operating on DualSense
technology. This includes the Tango TX1 when it is operating with only one working sensor.
When two redundant, working sensors are installed in the Tango TX1, the instrument is operating on DualSense and
the probability of sensor failure – compared with a single-sensor instrument – is diminished regardless of bump test
frequency (see Figure 1). The frequency of bump testing for DualSense instruments, between monthly calibrations, is
best determined by a company's safety policies. These policies may be affected by the directives and
recommendations of regulatory groups, environmental conditions, operating conditions, instrument use patterns and
exposure to gas, and other factors.
Table 2.1 ISC recommended practices
Procedure Recommended minimum frequency
DualSense instruments
Configuration Before first use, when there is a change in the installed sensor type, and as otherwise needed.
Calibration
Bump test
Self-test
Non-DualSense instruments
Configuration Before first use and as needed thereafter.
Calibration
Bump test
Self-test
a
Between regular calibrations, ISC also recommends a calibration be performed immediately following each of these incidences: the unit falls,
is dropped, or experiences another significant impact; is exposed to water; fails a bump test; or has been repeatedly exposed to an over-range
(positive or negative) gas concentration. A calibration is also recommended after the installation of a new (or replacement) sensor.
b
If conditions do not permit daily bump testing, the procedure may be done less frequently based on company safety policy.
c
The Tango TX1 self-test is performed automatically during the start-up process. It can be user-initiated from operation mode.
Note: The use of calibration gases not provided by ISC may void product warranties and limit potential liability claims.
a
Before first use and monthly thereafter.
b
As desired between monthly calibrations.
c
Prior to each day’s use for a unit that is always on or is left on.
a
Before first use and monthly thereafter.
b
Prior to each day’s use.
c
As desired between daily bump tests.
8
1 in 1
1 in 10
1 in 100
1 in 1,000
use-days
1 in 10,000
1 in 100,000
First Use
Average number of failures per instrument
1 in 1,000,000
0 5 10 15 20 25 30
Bump test interval (days)
Non-DualSense instruments
DualSense instruments
Figure 1. Probability of sensor failure by bump test frequency
First Use
To prepare the Tango TX1 for first use, qualified personnel should configure and calibrate the unit.
Wearing the Instrument
Based on the U.S. Department of Labor's Occupational Safety and Health Administration (OSHA) definition of the
breathing zone, it is recommended that the unit be worn within a 25.4 cm (10") radius of the nose and mouth. Refer
to OSHA and to other agencies or groups as needed for additional information.
ISC also recommends that the unit be worn within the instrument operator's sight line.
The instrument operator may wear the unit with its factory-installed suspender clip or with the optional belt clip. The
suspender clip is solely intended for attachment to a garment. The belt clip may be attached to a hard hat, belt, or
garment.
The clips should be securely fastened and attached in a manner that ensures the unit's sensor portals are fully
exposed to the air. No part of the unit should be covered by any garment, part of a garment, or other item that would
restrict the flow of air to the sensors or impair the operator's access to the audible, visual, or vibration alarms.
Attach the desired clip as shown below.
Table 2.2 Securing the garment or belt clip
Suspender clip
Lift the clip cover.
Position the garment between the clip's upper and lower teeth.
Press down on the clip cover to secure the clip in place.
9
Recommended Practices
Table 2.2 Securing the garment or belt clip
Belt clip
Position the hard hat or
garment between the clip and
clip back.
Slide the clip to secure it in place.
The instrument can be worn right side up or upside down.
10
Instrument Basics
Unpacking the Instrument
Hardware Overview
Display Overview
Start-up and Shutdown
3
Unpacking the Instrument
The items that are shipped with the unit are listed below (see Table 3.1); each item should be accounted for in the
unpacking process.
Table 3.1 Package contents
Quantity Item Notes
1 as ordered Tango TX1 Part number 18109075
1 Suspender clip (installed) Optional belt clip sold separately.
1 Calibration cup
1 Calibration and bump test tubing 60.96 cm (2 ') of urethane tubing; 4.762 mm (3/16 ”) ID
1 Reference Guide Companion to the Tango TX1 Product Manual
1 Final Inspection & Test Report and
Declaration of Conformity
1
At the time of shipment.
Note: If any item is missing or appears to have been damaged, contact ISC (see "Contact Information") or a local distributor of ISC products.
—
The Final Inspection & Test Report includes these values:
Unit setup date
Unit part number
Unit serial number
For each sensor:
Part number
Serial number
Type
Alarm settings
Span information
1
1
Instrument Basics
Hardware Overview
The instrument's main hardware components are identified below (see table 3.2).
Table 3.2. Hardware overview
Tango TX1
Visual alarm (or alert) indicator
IrDA (infrared data exchange)
window
Sensor port 1 and dust filter Sensor port 2 and dust filter
Case top
On-off-mode button
Case bottom
Enter button
LCD Audible alarm (or alert) indicator and
Garment clip (closed)
Garment clip (open)
Visual alarm (or alert) indicators
dust filter
Display Overview
The visual test screen shown below contains all the indicators that can appear on the display screen. Each indicator
is stationary and appears only when relevant to the task being performed. For example, in the gas-monitoring screen
shown below (numeric display), the following apply: the check mark indicates there are no sensor faults; the sensortype icon indicates that H2S sensors are installed; the numeric display shows a gas reading of 5.1 ppm.
Table 3.3 Display screen indicators and abbreviations
Display screens
Visual test screen
Status indicators
only
12
Two sensors are installed and neither is in fault.
Two sensors are installed and one is in fault; a
in fault.
Gas-monitoring screen
(numeric display)
Gas-monitoring screen
(text display)
sensor location icon also displays to indicate which sensor is
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.