inCourage RespirTech Instruction Manual

This manual contains words and symbols that bring attention to specific statements.
WARNING Indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation which, if not
The following symbols are used in this manual and on the inCourage
®
Airway
Clearance Therapy. Their meanings are as follows:
2
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
RespirTech Business Hours: Monday – Friday 8 am – 5 pm CT Toll Free: 800–793– 1261 www.respirtech.com
2896 Centre Pointe Drive, St. Paul, MN 55113 TF 800.793.1261 | Fax 800.962.1611
© 2016 Respiratory Technologies Inc. 900141-000 Rev B
Attention: Consult
Accompanying
Documents
Warning: Risk of
Electric Shock
Upper and Lower
Temperature Limits
Typ e BF
Applied Part
Manufacturer
Lot Number Reference
Number
Date of
Manufacture
Federal law restricts this
device to sale by
or on the order
of a physician
Serial
Number
Protective
Earth
Terminal
TUV Mark
SN
RX Only
TABLE of CONTENTS
Introduction ..................................................................................................... 4
Components and Controls ................................................................................ 5
Safety Instructions ............................................................................................. 7
System Setup .................................................................................................... 8
Fitting the inCourage Comfort Vest .............................................................8
Treatment Definitions ...................................................................................... 12
Starting a Treatment Session ............................................................................ 13
Turn On the inCourage System ................................................................. 14
Quick Start Mode .................................................................................... 14
Auto Pause Mode .................................................................................... 15
Multi Step Mode ......................................................................................16
Modify a Multi Step Session ............................................................................. 17
Fast Set Programming (FS) ........................................................................ 17
Review and Adjust Multi Step Session Settings ............................................ 18
Set Session to Factory Default ................................................................... 19
System Utilities ...............................................................................................20
Information Menu (i) ................................................................................ 20
Retrieve Hour Meter Reading (HMR) .......................................................... 20
Retrieve Software Version .......................................................................... 20
Retrieve Device Average Use Information ...................................................21
Setup Menu ............................................................................................ 21
Beep Control ................................................................................... 22
Quick Start/Auto Pause Session Time .................................................22
Pause Time ...................................................................................... 22
Pressure Limit ................................................................................... 23
Pressure Increment ............................................................................ 23
Reset Defaults................................................................................... 23
Clear Average Use ........................................................................... 23
Using the Button Lock ..................................................................................... 24
Cleaning the System ....................................................................................... 25
Filter and Fuse Replacement ............................................................................ 26
Troubleshooting Guide ................................................................................... 27
Specifications ................................................................................................. 28
Contraindications ........................................................................................... 29
Electromagnetic Emissions ...............................................................................30
Electromagnetic Immunity ............................................................................... 31
Patient Bill of Rights and Responsibilities ...........................................................32
Notice of Privacy Practices ............................................................................... 36
Medicare DMEPOS Supplier Standards ............................................................ 42
3
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
INTRODUCTION
Thank you for selecting the inCourage® Airway Clearance Therapy from RespirTech. RespirTech’s vision is to improve lives by helping people breathe better.
How does the inCourage Airway Clearance Therapy work?
The inCourage System consists of an inflatable vest, interconnecting hoses, and a Pulsating Therapy Unit (PTU) that creates compressions to the chest to help loosen, thin, and move mucus throughout the lungs.
Indications for Use
The inCourage Airway Clearance Therapy is indicated when external manipulation of the chest is the physician’s treatment of choice for increasing the clearance of mucus in patients with pulmonary disorders. The system promotes airway clearance and improves bronchial drainage utilizing High Frequency Chest Compression (HFCC).
This manual includes instructions for setup, use, and maintenance of the inCourage System. It is to be used as a reference guide. Please review all sections carefully prior to using the inCourage System.
The device operating instructions contained herein apply to devices with V4.0 software.
Warranty provided separately with shipment of device.
CAUTION: Federal law restricts this device to sale by or on the order of a physician!
4
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
COMPONENTS and CONTROLS
Pulsating Therapy Unit (PTU)
The PTU is the main control unit, which creates the pulsating air compressions that are transferred to the vest. It is the source of control for the system. PTU Model number: ICS-1M-US-A
inCourage Comfort Vest
The vest is made of multi-layered nylon. It is secured utilizing QuickFit straps that allow you to quickly set the vest to the recommended size to optimize the benefits of the pulsating air.
Interconnecting Hoses
The two hoses are made of high-strength, PVC. The hoses deliver the pulses of air from the PTU to the vest.
Locking Hose Connectors
The connectors are made of high strength plastic with locking levers.
Power Cord
The power cord plugs into the PTU and to a grounded, three-pronged outlet. Grounding reliability can only be achieved when the equipment is connected to a properly grounded, three-pronged outlet.
On/Off Switch
The power to the unit is controlled by the On/Off switch located on the back of the unit. On = I Off = 0
5
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
Control Panel
The control panel is located on the top/front of the PTU and provides all interactive menus and performance readouts.
Control Buttons
The control buttons are located just below the display. As the readout above each button changes, the function of the button changes.
Power Indicator
The power indicator glows green when the unit is plugged in and the On/Off button is in the On position.
Stop/Reset Button
This button ends all current functions, resets the control board, and returns to the main menu.
QuickFit Adjustable Straps
These straps allow you to size your vest to the exact, repeatable, recommended therapy spacing.
Air Ports
The air ports are located at the front of the PTU and on the front of the vest. The interconnecting hoses attach to the ports to complete the connection from the vest to the PTU.
COMPONENTS and CONTROLS (cont’d)
6
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
WARNING: To reduce the risk of electrocution, ALWAYS unplug this
product immediately after using. Failure to do so could result in personal injury or damage to equipment!
WARNING: To reduce the risk of electrocution, fire, or personal injury,
follow these instructions!
1. Be sure to read all information and instructions prior to setting up and using the
inCourage System. Failure to do so could result in equipment damage, personal
injury, or death.
2. Close supervision is necessary when this product is used by or near children or
individuals incapacitated by a chronic illness or injury.
3. Product is to be used only as described in this manual and only as prescribed by a
physician. Failure to do so could result in serious injury or death.
4. All attachments must be recommended by the manufacturer. Use only
components specified in the manual.
5. Do not use device near flammable substances.
6. Inspect power cord and unit prior to use. Do not use this product if any damage
has occurred to power cord or any part of the system. Call a RespirTech Customer
Support Representative at 1-800-793-1261 if any damage is suspected.
7. Keep all components of the inCourage System away from heated surfaces to avoid
damage and possible injury.
8. Do NOT insert any object into any opening on the system. Do NOT place objects
on or over the top of the inCourage System.
9. Always observe ALL warnings, cautions, and notes listed in this manual and posted
on the inCourage System.
10. Be sure to plug the inCourage System DIRECTLY into a properly grounded,
three-pronged electrical outlet. Grounding reliability can only be achieved when
the equipment is connected to a properly grounded, three-pronged outlet.
11. Do not use the inCourage System near water or ANY wet surface. Allow the
inCourage System to dry completely before plugging it into an electrical outlet if
the system becomes wet.
12. Place the inCourage System only on a flat and stable surface away from any
materials that could block unit air intake or exhaust.
13. When traveling by air, the inCourage System should always be
considered carry-on baggage and NEVER checked baggage.
14. It is recommended that the owner or renter of the inCourage Airway Clearance
Therapy insure the device under their renters or homeowner insurance policy
to cover non-warranty damage expenses.
SAFETY INSTRUCTIONS
7
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
STEP 1: Fitting the inCourage Comfort Vest
Inspect the Comfort Vest:
Note: Prior to use, the vest should be inspected for any sign of defects. If any of
the following defects are found, please contact a RespirTech Customer Support Representative at
1-800-793-1261 for immediate vest replacement:
1. Any size hole found in the vest (inner layer, outer layer, seams, or vest ports).
2. Any stitching that has become loose or frayed on the vest.
3. Any damage to the buckles or straps that restricts the function of the strap system.
Adjust the Comfort Vest Straps:
We have taken careful consideration in supplying you with the appropriate size vest based on your information. Please take a moment and review the fitting instructions carefully and consult your physician if any discomfort develops.
NOTE: We recommend a thin layer of cotton clothing to be worn under the vest for best comfort.
To loosen straps:
Lift the back of the buckle.
To tighten straps: Hold the buckle and pull the end of the strap
to the patient’s right.
1. These instructions refer to the
three colored QuickFit tabs:
blue, white, and yellow.
2. Before putting on the vest, preset
the QuickFit tabs to the pre-therapy position by placing the blue tabs over the yellow tabs.
SYSTEM SETUP
8
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
3. Put the vest on and connect the
buckles located on the front.
4. Adjust the shoulder straps so the
bottom edge of the vest sits at the top of the hipbone.
5. Gently tighten the front straps
until each is snug to the chest.
6. Pull each of the QuickFit blue
tabs off the yellow tabs, then place the blue tabs over the white tabs.
SYSTEM SETUP (cont’d)
9
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
7. You are now ready to connect
vest to the PTU.
STEP 2 : Plug in the PTU
1. Place the PTU on a level, sturdy,
dry surface, clear of any flammable
materials or obstructions.
2. Plug the power cord to the back of
the PTU.
3. Plug the power cord into a
properly grounded, three-pronged
outlet. Grounding reliability can
only be achieved when the
equipment is connected to a
properly grounded, three-pronged
outlet. See product specifications
for power requirements.
SYSTEM SETUP (cont’d)
10
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
STEP 3 : Connecting the Vest to the PTU
NOTE: The interconnecting hoses are interchangeable and can be flipped. You do not need to worry about which end of the hose is attached to the PTU or the vest.
1. Grasp the sides of the locking
hose connecter and push onto the
device (Image A). Blue lever on
hose connector will automatically
flip up (Image B).
2. Continue to push the hose
connector firmly against device
until hose connector has been
properly fitted against device
(lever will flip down).
3. Verify the blue lever is securely
latched by pressing down firmly.
(Image C).
NOTE: Do not force blue lever to close. If you feel resistance, push the hose connector more firmly against device.
4. Grasp the sides of the locking
hose connecter and push onto
the vest port as show (Image D).
Blue lever on hose connector will
automatically flip up.
5. Continue to push the hose
connector firmly against the vest
port until hose connector has been
properly fitted (lever will flip down).
6. Verify the blue lever is securely
latched by pressing down firmly.
(Image E).
NOTE: To disconnect the vest, lift the lever and pull hose off the vest port.
AB C
ED
SYSTEM SETUP (cont’d)
11
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
Treatment Definitions:
1. FREQUENCY (F) – Represents the number of air pulses that are delivered over
a period of time. [The inCourage System represents frequency in Hertz (Hz), or
the number of pulses per second.]
2. PRESSURE (P) – Represents the force that the system places against your body.
[The inCourage System represents pressure as a percentage (%) of the
maximum pressure created.]
3. SESSION – A full treatment of individual or combined steps.
4. STEP – A segment of frequency, pressure, and time that together with other steps
make up a complete session.
5. RAMPING – During a step, the inCourage System gradually increases or
decreases the frequency to the next step’s frequency.
6. AUTO PAUSE – Automatically pauses session every 5 minutes. Press RUN to resume.
7. HOUR METER READING (HMR) – Represents the total use time over the life of
the device. This timer cannot be reset.
8. USE INFO – Represents the average session length in minutes and average
pressure setting for the previous number of sessions displayed. Only sessions
over five minutes will be counted. The user can reset this information.
9. PAUSE TIME – Allows for automatic continuation of therapy after a pre-
programmed pause during an Auto Pause or Multi Step session.
10. PRESSURE LIMIT – The maximum allowable pressure setting for all therapy
sessions and programming.
11. PRESSURE INCREMENT – The amount of pressure percent (%) changed
when increasing or decreasing pressure.
12. FAST SET PROGRAMMING (FS) – A method for creating a customized Multi
Step session of up to 12 steps with alternating frequency with an input of 6
settings: Number of Steps, Low Frequency, High Frequency, Pressure Setting,
Total Program Time, and Auto Pausing On or Off.
CAUTION: Use of the system other than as prescribed could result in
ineffective treatment. Prescribed settings must be utilized!
TREATMENT DEFINITIONS
12
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
The inCourage® Airway Clearance Therapy is programmed to provide easy-to-follow instructions with an easy-to-understand menu. It has three types of sessions, each with unique advantages. The Quick Start session is a pre-programmed session that starts immediately. The Auto Pause session is the Quick Start program with an automatic feature to pause the session every 5 minutes
1
. The Multi Step sessions are for custom
programs that can be saved for future use.
NOTE: The PTU will generate an audible tone as feedback when pressing any of the front panel buttons.
2
1. Quick Start and Auto Pause default duration of 30 minutes can be adjusted by 5 minute intervals from a minimum of 5 minutes to a maximum of 90 minutes per session, the number of minutes per step will vary.
2. By default, the audible tone is only generated when pressing one of the front panel buttons. Setting Beep Control to ‘On’ will also generate audible feedback at the end of a step.
QUICK START AUTO PAUSE MULTI STEP
Key Feature
Program
Session Length
Pressure
Frequency
Memory
Instant 30-minute1 ramping session
Pre-programmed
30-minute session divided into 12 steps of 2.5 minutes each
1
Default set at 60% and adjustable during session
Ramps from 6Hz to 15Hz over a 2.5 minute period – then ramps from 15Hz to 6Hz over the next 2.5 minutes and repeats for a total of 30 minutes
1
Last pressure setting used is retained for the start of the next session
Pauses session every 5 minutes
1
Pre-programmed
Same as Quick Start with automatic pause every 5 minutes
1
Default set at 60% and adjustable during session
Ramps from 6Hz to 15Hz over a 2.5 minute period – then ramps from 15Hz to 6Hz for the next
2.5 minutes followed by a pause. Repeats for a total of 30 minutes
1
Last pressure setting used is retained for the start of the next session
Customized therapy with multiple steps and ramping
Programmable up to 12 steps
12 User­programmable steps up to 99 minutes each
Set at desired pressure per step and adjustable during session
User defined per step between 5Hz and 30Hz Ramping
3 programmable memory buttons
STARTING A TREATMENT SESSION
13
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL
NOTE: The power indicator light on the control panel will light green to show that the unit is in use.
A U T O M U L T I Q U I C K P A U S E ( i ) S T E P S T A R T
STEP 1: Turn On the inCourage Airway Clearance PTU
1.
Turn the power switch on the back of the PTU to
the On position. (Ensure unit is plugged in to a
properly grounded three-pronged outlet.)
2. Adjust pressure by pressing either DOWN
or UP. Press PAUSE to temporarily stop the
session. Press RUN to resume the session
or END to return to the startup screen
.
3. When the session pauses, press RUN
to resume the session or END to return
to the startup screen.
STEP 2: Choose a Session Mode (Quick Start, Auto Pause, Multi Step)
To run Quick Start mode:
1. From the startup screen, select QUICK
START. Your QUICK START session
will begin.
A U T O M U L T I Q U I C K P A U S E ( i ) S T E P S T A R T
1 8 . 9 6 0 % 2 9 : 3 4 P A U S E D O W N U P E N D
S E S S I O N P A U S E D R U N E N D
STARTING A TREATMENT SESSION (cont’d)
14
inCourage® Airway Clearance Therapy | INSTRUCTION MANUAL