Imed 960, 965 Service manual

r
Cut
and
insert
label
holder
on
spine
,11
binder
of
960A/
965A
|P
pi
T
WARNINGS, CAUTIONS AND NOTICES
USE ONLY
PROPER
HOSPITAL
GROUNDING.
GRADE
POWER
GROUNDING
ACHIEVEDBYCONNECTIONTOA
GRADE".
DANGER:
DANGER:
CAUTION:
WARNING: REPLACE
EXPLOSION
OF
FLAMMABLE
RISQUE
PRESENCE D'ANESTHESIQUES INFLAMMABLES.
TO
REDUCE
REMOVE
QUALIFIED
HAZARD,
ANESTHETICS.
D'EXPLOSION,
RISK
COVER
SERVICE
FUSE
OR
AS
SUPPLY
RELIABILITY
RECEPTACLE
DO
NOT
NE
PAS
OF
ELECTRICAL
BACK.
PERSONNEL.
MARKED.
CORD
USEINTHE
EMPLOYER
SHOCK,
REFER
TO
INSURE
CAN
ONLY
MARKED
PRESENCE
EN
DO
SERVICING
BE
"HOSPITAL
NOT
TO
JSJ
IPR
CAUTION:
TO
MAXIMIZE
RECOMMENDED
OPERATED
THE
IN
CONTROLLED
IN
DROPPED AT
CHECKED
PRIOR
BEFORE
SERVICE
THAT
AN
BETWEEN
THE
TO
CONNECTING
NOTE
LIFEOFTHE
THIS INSTRUMENT
ENVIRONMENT
68°F
EVENT
BY
A
USE
(20°C)
WARNING
ANY
BIOMEDICAL
FOR
AND 77°F
THE
INSTRUMENT
TIME, IT MUST BE
PATIENT
"REFER
INSTALLED
BE
THAT
IS
(25°C).
TECHNICIAN
CARE.
TO
MANUAL'
BATTERY,ITIS
STORED
TEMPERATURE
IS
AND
pi
NOTICE
JIPJ
r
fw
Product
to
design
change
without
containedinthis
of
issue.
This
publication
provided
in
repairing
solely
IMED
and/or
contains
for
specifications
notice.
manualiscurrent
the use of
infusion pump/controllers.
The
IMED
technical
information
as ofthe
proprietary
are
subject
personnel
date
data
Noneofthe
duplicated
other
than
IMED
infusion
component parts thereof.
of
the
information
the
usertosubstantial
information
nor
mayitbe
for
the
repair
pump/controllers
contained
contained
utilizedinany
and
maintenance
and
Any
unauthorized use
herein
liability,
herein
manner
the
may
may
of
subject
be
r
f^
Copyright0
9775 Businesspark Avenue
San
Diego, CA 92131-1699
(619)
All Rights
PrintedinU.S.A.
U.S.
Other Patents Pending
1990 by
566-9000
Reserved
Patents
IMED
3,985,133
Corporation
PREFACE
imed
'
(IW
This manual contains operation and maintenance
instructions for
of
Volumetric/Microvolumetric
These instruments are grouped
configurations
figure:
CONFIGURATION
Foster
Hightower
Product
PI
Inprovement
Technical
particular configuration is
reference
the
which
REFERENCE
information
code
and/or
IMED Model 960A/965A
Infusion
into
are definedinthe
SOFTWARE
F
HT
PIP
that is
unique
identified
the
software
Pumps.
three
following
DISPLAY
None
P0.6/P1.4/P1.5
P4.xx
to a
by the
version
series
displayed during initialization.
The information provided herein is intended for
use by technical personnel responsible for
servicing
into six sections and is presented
these
products.
The
material is divided
as
follows:
Section 1 - Descriptive Information; Section 2 -
Preparation for Use; Section 3 - Operating
Instructions;
Section 4 -
Functional
Description
and Schematic Diagrams; Section 5 -
Maintenance Instructions;
Section
6 -
Parts
Lists
and Assembly Drawings.
This manual
Additional by contacting
ment
and
The
960A/965A
supersedes
IMED
P/N 960-9010-00.
copies of this manual may be obtained
IMED's
ordering IMED
Customer Service Depart
Part
No. 960-9200-00.
International
Addendum
P/N
960-
9208 supplements this manual for support of the
220V
versions
of
these
instruments.
€$Sl
7^
ips
RECORD
OF
CHANGES
imecl
•PT
Date
05/01/90
DescriptionofChange
Initial
Issue
By
JD/MOW
(
F
<BB
imed
TABLE
section
•pj
SECTION
1 -
description
1.0
1.1
1.2
1.3
1.4
1.5
2 -
2.1
2.2
2.3
2.3.1
2.3.2 Pre-operational
2.3.2.1 Electrical Leakage Test 2-3
2.3.2.2 Electrical Ground Test 2-3
2.3.2.3 Dielectric Test 2-3
2.3.3
2.3.3.1 Test Requirements 2-3
2.3.3.2
introduction
operating
PHYSICAL
USER
CHARACTERISTICS
INTERFACE
OPERATING
ADMINISTRATION
PREPARATION
FOR
INTRODUCTION
PRE-OPERATIONAL
OPERATIONAL
Pre-operational Check
Abbreviated Test
Procedures
characteristics
SPECIFICATIONS
SETS
USE
MECHANICAL
PERFORMANCE
Battery
Electrical
Operational
OF
CONTENTS
!.!!!!."!.'
'.'.'.'.'.'.
!!!!!!!
INSPECTION
CHECK
Charge 2-3
Inspection 2-3
Performance
Test
'.
!
'...'.
n
1-1
i.i
i.i
i_2
1-2
1-5
2-1
2-1
2-1
2-1
2-3 2-3
f
SECTION
^pi
t
3 -
OPERATION
3.1
3.2 CONTROLS
3.3
3.3.1 Setup Procedure 3-4
3.3.2 Starting an Infusion 3-4
3.3.3 During the Infusion 3-5
3.3.4 Infusion Complete 3-5
3.3.5 Stopping an Infusion 3-5
3.3.6 Changing Rate or
3.3.7 Resetting Volume Infused 3-5
3.3.8 Clearing Alarms 3-6
INTRODUCTION
AND
INDICATORS
NORMAL
OPERATION 3.4
VTBI
during an
' "
Infusion
3-1
3.1
3-1
3-5
r
r
r
f™
SECTION 4 - PRINCIPLES OF OPERATION
4.1 INTRODUCTION 4-1
4.2
4.2.1 Physical Description
4.2.1.1 Pumping Mechanism
4.2.2 Functional Operation
4.2.2.1 Pumping Mechanism 4-2
4.3
4.3.1 Functional Description 4-2
4.3.1.1 Microprocessor System 4-2
4.3.1.2 Prime Mode Logic 4-2
4.3.1.3
4.3.1.4 Photosensor System 4-5
4.3.1.5 Alarm
4.3.1.6 Volume and Rate Inputs 4-6
4.3.1.7
4.3.1.8 Power Supply 4-6
MECHANICAL OPERATION 4-1
ELECTRICAL/ELECTRONIC OPERATION 4-2
Operate
Malfunction Alarm 4-6
Logic 4-5
System
4-1
4-1 4-1
4-1
4-5
in
imecl
4.3.2 Detailed Circuit Description 4-6
4.3.2.1
4.3.2.2 Microprocessor
4.3.2.3 Keypad Entry and 7-Segment
4.3.2.4 Circuit Enabling 4.9
4.3.2.5
4.3.2.6 Prime Mode Logic . . , 4-10
4.3.2.7 Operate
4.3.2.8 Rate Compensation 4-11
4.3.2.9 Volume To Be Infused Logic
4.3.2.10
4.3.2.11 Forward/Reverse Logic . 4-11
4.3.2.12 Alarm
4.3.2.13
4.3.2.14 Pressure Adjustments 4-19
4.4
4.4.1 Functional Descriptions 4-20
4.4.1.1 Microprocessor System 4-20
4.4.1.2 Prime Mode Logic 4-20
4.4.1.3 Operate Logic , 4-20
4.4.1.4
4.4.1.5 Alarm
4.4.1.6 Volume and Rate Inputs
4.4.2 Detailed
4.4.2.1 Microprocessor System . , 4-22
4.4.2.2
4.4.2.3 Displays . . . . 4-23
4.4.2.4
4.4.2.5 Battery Charger 4-23
4.4.2.6 Ultrasonic AIL Detector 4-24
Main
Circuit Board Logic 4-7
System
Power
Motor
Liquid
ELECTRICAL7ELECTRONIC
Photo-sensor
Keypad Back
Supply
Logic
Direction Control
Systems
Crystal Display
System
System
Circuit
Descriptions 4-22
Entry
Up
Battery
(LCD)
OPERATION
Display
Logic 4-9
[ [
..
[ . [
4-7
'
4-9
4-10
4-11 4-11
4-12
4-19
4-20
'
4-22
4-22
4-22
.'
4-22 4-23
SECTION
5.1
5.2
5.2.1 Cleaning Instructions
5.2.3
5.2.3.1
5.3
5.4
5.4.1
5.4.2
5.4.2.1
5.4.2.2
5.4.2.1
5.4.2.2
5.4.2.3 Angle
5.4.2.4
5.4.2.5 Cassette Nest Removal 5-9
5.4.2.6 AirInLine
5.4.2.7
5.4.2.8
5.4.3
5.4.3.1
5.4.3.2
5.4.3.3
5.4.3.4
5 -
MAINTENANCE
INTRODUCTION
PREVENTIVE
Electrical
Battery
TROUBLESHOOTING
DISASSEMBLY
Separating
Front
Main Front
Power
Keypad Door
Invertor
Hall
Disassembly
Valve Shaft
Shuttle
Angle
Shuttle
MAINTENANCE
!!!!!!!
Inspection
Voltage
Check
'.'.'.'.'.'.'.'.'.
the
Case
Pump
Logic
Panel
Assembly
Effect
Assembly
Board
Assembly
and
Display
Assembly
Plate
Assembly
Removal
Detector Removal 5-9
Removal . . . 5-10
Sensor
of the
Drive
Drive
Motor
Plate
Circuit
Assembly
Disassembly
Removal
Removal
Board
Removal
Removal
Harness
Angle
Motor
Removal 5-10
Removal
Card
Removal
Removal
Removal 5-9
,[
' "
Removal
Plate
Assembly
!!!!!!!!!
!!!!!!!
5-1
5-1
5-1
5-1
^
5-2
5-2
5-8
'.
5-8
5-8
5-8 5-8
5-9
5-9
5-9
5-10
5-10
5-10
5-10
5-11
imecl
5.4.4
wn
5.4.5
5.4.5.1
5.4.5.2
w\
5.4.5.3
5.4.5.4
5.4.5.5
5.4.5.6
5.4.5.7
5.4.6 Chassis Assembly Disassembly
5.4.6.1 Battery Removal
5.4.6.2
5.4.6.2
5.4.6.3 Transformer Removal 5-12
5.5 REASSEMBLY
5.6 COMPREHENSIVE OPERATIONAL PERFORMANCE TEST 5-15
5.6.1
5.6.2 Qualitative Operational Performance Test 5-15
5.6.3 Quantitative Operational Performance Test 5-15
5.6.3.1 Equipment Requirements 5-15
5.6.3.2 Test Setup 5-15
5.6.3.3 Rate/Volume Inhibit
5.6.3.4 Prime Inhibit
5.6.3.5 Volume Infused
5.6.3.6 Normal Operation
5.6.3.7 Walk-Away Alarm
5.6.3.8 Door Alarm
5.6.3.9 Power Off
5.6.3.10 Purge Pressure
5.6.3.11 Operate (up stroke) Pressure Test Procedure 5-18
5.6.3.12
5.6.3.13 Downstroke Occlusion Pressure
5.6.3.14 Operate
5.6.3.15 AIL Detector
5.7
5.7.1
5.7.2 Battery Alarm Check 5-20
5.7.3 Occlusion Calibration 5-21
5.7.4 Battery Charger Adjustment
5.7.5 Plunger
5.7.6 Valve
5.7.7
5.7.8 Occtusion
Front
Panel
Rear
Pump
Chassis Assembly Removal
Pole
Clamp
AC
Power
Sonalert
Audio
Nurse
Line
Cord
Assembly Disassembly
Assembly
Assembly
Entry
Module
Removal
Silence
Call
Switch
Receptacle
Clip
Removal
Disassembly
Removal
Removal
Removal
Removal
."!!!!
!!!!!!!
\\\
5-11
5-11
5-11
5-11
.'
5-11
''
5-11
5-12
'
5-12
.'
5-12
5-12
5-12
Power
Battery
Electrical
Volume-Rate
ADJUSTMENT AND CALIBRATION 5-20
Equipment
Ultrasonic AIL Detector Calibration Procedure 5-22
Supply
Backup
Board
Board
Inspection
Test
Reset
Test
Memory
Test
Test
(Upstroke)
Test
Removal
Removal
Test
Procedure 5-17
Procedure 5-17
Inhibit
Test
Procedure 5-17
Test
Procedure 5-17
Procedure 5-17
Test Procedure 5-18
Procedure 5-18
Procedure 5-18
Pressure Test
Procedure 5-19
Requirements
Limit
Adjustment 5-22
Drive
Shaft Check and Adjustment
Calibration
(Alternate
(PIP)
Test
Procedure 5-17
Test
Procedure 5-18
Procedure
Method)
(Alternate)
.[
5-12
5-12
5-13
5-15
5-19
5-20
5-21
5-22 5-23
SECTION 6 - ILLUSTRATED PARTS BREAKDOWN
6.1 INTRODUCTION 6-1
INFUSION
WARRANTY
IMED INTERNATIONAL
DEVICE
SERVICE
DISCREPANCY
CENTERS
OFFICES
REPORT
(IMED
Form
9000-x
6-1
1/90)
imed
LIST
Figure
1-1
1-2
2-1
2-2 Nurse
3-1
4-1
4-1
a
4-2 Stepper
4-3 4-4
4-4a Ultrasonic AIL Detector Schematic 4-24
4-5
4-5a Power/Display Board (PIP/P4.x) Sheet 1of5 . . . . 4-27
4-5a Power/Display Board
4-5a Power/Display Board
4-5a
4-5a
4-6
4-6 4-6
4-6a
4-7 Angle
4-7a
4-8 Power Supply
4-9 Battery Backup
4-10
4-10a Interconnect 4-10b Interconnect
4-10c
4-1
Od
4-11 4-11a
5-1
5-2
5-3
5-4
5-5
6-1
6-2
6-3a
6-3b
6-4a Parts
6-4b
6-5
6-6
6-7 6-7a
6-8
6-9
6-10
6-11
6-12
Title
IMED
Audio
96QA/965A
960A/965A
960A/965A
960AJ965A
Stepper
960A/965A
Main
Power/Display
960A/965A Volumetric
Characteristics
Front
Call
Test
Controls
Functional
Functional
Motor
Motor
Functional
Logic
Board
Power/Display
Front
Front Front
960A/965A
Angle
Interconnect
Interconnect
Interconnect
960A
965A
Universal
Open-ended
AIL
Typical
960A/965A
Parts
Parts
Parts
Parts
Parts
Parts
Parts
Parts Parts
Parts
Parts
Parts
Parts
Parts
Panel
Panel Panel
Plate
Plate
Assy.
Assy. Assy.
Front
Board
Board
Diagram
Diagram Diagram
Diagram
Diagram
AIL
Detector
AIL
Detector
Test
AIL
PCB
Board
Pressure
Test
Identification
Identification
Identification
Identification
Identification
Identification
Identification
Identification
Identification Identification
Identification
Identification
Identification
Identification
Identification
Infusion
and Rear Operating Features 2-2
Light
and
Indicators
Block
Block
Pulse Phasing 4-12
Phase
Sequencing
Block
Schematic . . . 4-26
(PIP/P4.x)
(PIP/P4.x)
Board
Board
(PIP/P4.x)
(PIP/P4.x)
(Display
(Display (Display
Panel
Schematic 960A 4-36
Schematic
Board
Schematic 960A/965A 4-38
Board
(PIP/P4.x)
960A
960A
965A (Foster) . . . . 4-42
965A
960A
(Optical) (Optical)
Station
Setup
Diagram
Diagram
Diagram
Board)
Board) Board)
Assy.
(Display
965A
(Foster)
(Hightower)
(Hightower)
(PIP/P4.X)
Schematic Schematic
Simulator Test
Station
Data
Sheet
960A/965A
960A/965A
960A
965A
960A Angle
965A
960A/965A
960A/965A
960A/965A
960A/965A
9S0A/965A
960A/965A
960A/965A
960A/965A
9S0A/965A
Setup
Final
Front
Angle
Angle
Angle
Door
Shuttle
MLB Power&Display
Front
Front
Rear
Power
Battery
OF
FIGURES
Page
Pump 1-0
.'
1.4
2-3
'.'
3-2
(Foster)
(Hightower)
(PIP/P4.x)
Sheet 2 of 5 4-28
Sheet 3 of 5 4-29
Sheet
4 of 5 4-30
Sheet
5 of 5 .'.' '
Schematic
Schematic Schematic
Board)
Schematic
Sheet 1 of 3
Sheet 2 of 3 Sheet 3 of 3
Schematic
(F)
(F) (F)
(HT)
...'.
' ' ' '
'.,.'
*
4-13
.'.
4-21
4-31
4-32
4-33 4-34
4-35
4-37
4-39 m
4-3
4-4
4-40
4-41
4-43
4-44 )
' '
4-45 4-46
5-16
5-23
Plate
Plate
Plat
Plate
Assembly
Pump
Assembly
Assembly
Assembly
Circuit
Circuit
Card
Card
Assembly
Assembly
Circuit
Case
Panel
Panel
Card
Assembly
CCA
Assembly
Supply
Back-Up
Assembly
CCA
Keyboard
PCB
PCB
(PIP/P4.X)
and
Assembly
Display
(HT)
(PIP/P4.X)
'.'.'.'.'.'.
. . . . . .
'.
5-23
5-24
5-25
6-3
6-6
6-9
6-10
6-13
6-17
6-19
6-21
6-25 6-31
6-33
6-36
6-40
6-43
6-45
J
f^?,
1
f*$$
]
1
m
-J
{
^
m
J
'
m
*!
j
*.
I
-J
<m
J
1
ft»'
VI
imed
Table
1-1
1-2
1-3
3-1
3-2
4-1
5-1
5-2
5-3
5-4
6-1
6-2 6-3
6-4a Parts
6-4b Parts
6-5 Parts
6-6 6-7 Parts
6-7a Parts
6-8
6-9
6-10
6-11
6-12 Parts
Title
Operating
Performance
Administration
960A/965A
Alarm/Prompt
Microprocessor
Troubleshooting/Fault
Malfunction
Malfunction
TableofTorque
Parts
Parts Parts
Parts
Parts Ust
Parts
Parts
Parts
Requirements
specifications
Sets
Controls
and
Messages
Signal
Isolation
Alarm
Alarm
Codes
Codes
Values
List-960A/965A
List-960A/965A
List-960A/965A
List
960A
Angle
List
965A
Angle
List
960A/965A
List
9S0A/965A
List
960A/965A
List
9S0A/965A
960A/965A
List
96QA/965A
List
9S0A/965A
List
960A/965A
List
9S0A/965A
Door
Shuttle MLB
Power
Front
Front
Rear Case
Power
Battery
LIST
OF
TABLES
Indicators-Functional
Functions
Guide
(P0.x/1.x)
(PIP/P4.X)
Final
Assembly
Front
Angle
Plate
Plate
Pump
Circuit
Circuit
Assembly
Assembly
Circuit
and
Panel
Panel
Supply
Back-Up
Plate
Card
Display
Assembly
Circuit
Assembly
Assembly Assembly
Board
Card
Assembly
CCA
Card
PCB
Assembly
PCB
Description
(HIGHTOWER)
Assembly
Assembly
Keyboard
...[..........
'
.'
!
and
Display
(HT)
.
Page
1_3
1-3
1-5
3-3
3.6
4-8
5.3
5-4
5-6
5.14
6-2
6-4 6-8
6-11
6-15
6-18
6-20 6-22
6-26
6-32
6-34
6-37
6-42
6-44
•^w
VII
imed
Figure
============
1-1.
IMED
9G0A/965A
1-0
Volumetric
======
Infusion
Pump
imed
r
SECTION
1.0
INTRODUCTION
This section contains a description of the operating and physical characteristics plus the
operating requirements
specifications for the
965A
Micro
Volumetric
administration sets compatible
instruments
are
listedatthe
and
IMED
Infusion
performance
960A Volumetric
with
end
Pumps.
of
IMED
each of these
this
section.
and
The physical and operational characteristics of
the
960A
and
965A
are
identical in
respects.
differ
face italic print.
1.1
The
Infusion
electro-mechanically
Those
from
OPERATING
IMED
characteristicsofthe
the 960A
Model
willbeidentified
CHARACTERISTICS
960A&965A
devices are electronically controlled,
driven,
piston
pumps. The 960A is designed delivery
965A
cations.
sets configured
of drugs and
fluids
to adult patients. The
is designed for pediatric/neonatal appli
Both
instruments
with
utilize
a snap-in cassette pump.
most
965A
by
Volumetric
(syringe)
primarily
administration
for the
that
bold
type
An
automatic priming feature is incorporated to facilitate preparing the administration set. A tactile
keypad allows the user to enter infusion rate and
volume to be infused
and
VTBI
are shown on seven-segment
displays. A backlighted
Volume
Infused, operator prompts and alarm
(VTBI).
LCD
The selected rate
LED
panel displays
messages. Table 1-1 defines the operating range
of the rate and instrument. Once
VTBI
parameters for each
set
up and started, both
instruments operate on a semi-automatic basis.
While
microprocessor monitors the
in-line
selected
an
infusion
is in progress, the instrument's
fluid
column for air-
and occlusion pressure. When the
volume
has
been
infused,
the
instrument
automatically changes over to the "keep vein
open
visual
(KVO)"
INFUSION
rate
and
COMPLETE
initiatesanaudio
alarm.
and
1 -
DESCRIPTION
In addition to controlling and monitoring the
infusion, the instrument's microprocessor
diagnostic program
appropriate action in the event of an open door,
the battery not chargingoreither a hardware or
software
The instruments can be operated on battery
power for a limited period of time. The length of
operating time available on battery power is
dependent upon the delivery rate used.
operating on AC power the battery charger is
recharging or maintaining the battery at
charge.
1.2
The 9S0A/965A instruments have the
physical characteristics:
Height: 11.75 inches (29.6 cm)
Width: 10.00 inches (25.4 cm)
Depth: 6.00 inches (15.2 cm)
Weight -14.75 pounds (6.7 kg) The
retardant plastic material housing and support for
components
is molded in two
rear
system
Although
are
built
malfunction.
the
and
specifications
not
intendedtoreplace
by
medical
We
urge
professional
use
of
PHYSICAL
case
is molded from a high-impact, flame
case
assemblies.
personnel
the
user
standard
this
equipment.
CHARACTERISTICS
and
pole clamp assembly. The case
pieces
will
NOTE:
IMED
tested
and
detect
Models
to
exacting
and
960AJ965A
standards,
the
supervision
of
IV
infusions.
to
exercise
of
careinthe
and
provides the
the
internally mounted
which form
initiate
they
a
following
the
While
full
are
front and
1-1
imocl
'
FRONT
Software
The front
CASE
Release
case
ASSEMBLY
P-0.x/1.x
Instruments
assembly contains the front panel
assembly, main logic board, invertor, angle plate assembly,
cassette
air in line detector
Software
Release
Thefront case assembly contains the
display
and
boot,
door assembly.
REAR
Software
cassette
CASE
Release
boards,
nest,
The rear case assembly contains the battery,
power input module, sonalert, audio control
switch,
assembly, power supply board, transformer line cord clip.
Software
The
power
battery
assembly and line cord clip.
1.3
The
panel assembly,
nurse call connector, pole clamp
Release
rear case assembly contains the
input
module,
backup
USER
960AJ965A
INTERFACE
nest, shuttle boot assembly,
and
door assembly.
P-4.x
Instruments
angle plate
air-in-line
ASSEMBLY
P-0.x/1.x
P-4.x
Instruments
sonalert, audio control
board,
transformer,
user
interface
the
administration
assembly,
detector
Instruments
pole
includes
keypad,
shuttle
assembly
battery,
clamp
the
set
installation
power
and
AC
and
AC
switch,
front
area and the rear panel assembly.
FRONT
PANEL
The front panel assembly employs four controls
to
operate
the
instrument.
The
ON/STOP
control
is an alternate action type switch used to turn the
instrument
On
and
Off.
The
PRIME
ON/OFF
control initiates operation of the pumping mechanism for approximately 1 minute,
is
setto"000M.
The
VOL INF
RESET
clear the volume infused register and the
panel display to "0",ifactuated
setto"0000".
The
START
when the
control
control
initiates
if
the rate LCD
VTBI
will
operation of the pumping mechanism,ifthe rate
and
VTBI
are
set
toavalue
>0.
Fourteen arrow ®(2) controls are used to
program the instrument for a specific infusion. Six
arrow controls
which
displays in three seven-segment
are
used
to input the infusion rate
LED
windows.
the
LED
Eight
arrow
VTBI
which displays in four seven-segment
controls are used to
windows. An amber backlighted "PRESS"
input
indicator (above the START control) illuminates with
an accompanying audio alarm if the START
control is not pressed
and/or
VTBI
while
The backlighted
"VOLUME INFUSED"
after the power
has
the instrument is operating a series of 5
sequence
across
following
an
infusion
LCD
panel displays the
since
been
a change in rate
is in progress.
the
last
reset
off for 4 hours.
or "0"
While
dashes
the LCD panel. Eleven visual
messages are displayed on the panel in response
to
instrument
operator. In addition to the
LED indicator
OPERATING
when
the
ADMINISTRATION
The
set
plastic door
shuttle
arm
must be
fitting
engaged
will
close. The
alarm
flashes
andared
instrument
installation
and
includes
and
air in line
properly
to the shuttle arm before the door
set
conditions
when
LED indicator
is in
SET
INSTALLATION
areaisbehind
or to
LCD
the
ALARM.
the
cassette
detector.
prompt
panel a green
instrument
the
The
is
flashes
AREA
clear
nest,
cassette
the
seated in the nest and plunger
tubing must be inserted into
the air in line detector to enable starting the
will
pump. The pump
is
opened
REAR
during an infusion.
PANEL
stop and alarm if the door
The rear panel controls include the AUDIO
silence switch. When permitted the audio alarm can be silenced temporarily by pressing the
AUDIO
Software
A
switch.
NURSE
Release
CALL
P-0.x/1.x
instruments
connector is provided. When an
external Nurse Call system is connected to the
instrument,
facility's
enters
is
1.4
OPERATING
the
960A/965A
will
activate
Nurse Call systemifthe instrument
an
alarm
condition.
SPECIFICATIONS
the
The operating specifications for the 960A & 965A are divided into operating requirements and
performance specifications. These requirements
and
specifications
and
1-2 respectively.
are
delineated in
Tables
1-1
1-2
_
Table 1-1. Operating Requirements
imed
Parameter
Power Required:
Battery:
Nurse Call System Power
Limitations: (P-O.x/1 .x Software
Instruments
ONLY)
Operating Temperature Range:
Operating
Storage Temperature: * Storage
Humidity
Humidity:
*
Parameter
Range:
See
NOTE
108-132
(secondary),
Sealed
Maximum
24
VAC,
0.3
Amps
50-60
Hz,
nominal,
Single
Lead-Acid,6VDC,10Amp-Hr
recommended Voltage/Current
VDC/500
mA
(Leakage current 100
40°F (5°C) to 104°F (40°C)
40% to 95% Relative
Humidity,
-40°F (-40°C) to 158°F (70°C)
on
Warning,
40% to 95% Relative
Caution
and
Notice
Humidity,
page inside of the
Table 1-2. Performance Specifications
Specification
Fusedat0.4A
<|>t2wire
non condensing
non condensing
Specification
plus
uA
maximum)
front
(primary)
Ground
cover.
5.0A
Operating Principle:
Operating
RATE
range:
VOLUMETOBE
Keep
Vein
Open
Occlusion
Air In
Battery Operation:
Line
Pressure:
Detection:
INFUSED
(KVO)
rate:
range:
Electronically
type
pump
1-999
0.1-99.9
1-9999
0.1-999.9
1
0.1
18 psi nominal
Software
Optical (5 Elements)
Software
Ultrasonic
mL/hr
mL/hr
mL/hr
10 hours at 125
mL/hr in
mL
in 1
mL in
Release
Release
controlled,
in 1
mL/hr
0.1
mL
0.1
P-4.x
mL/hr
electro-mechanically
increments
mL/hr
increments
mL
P-0.x/1.x
Instruments
increments
increments
Instruments
(with
new,
fully
charged battery)
driven
piston/cylinder
10
hours
at 50
=
1-3
mL/hr
(with
new,
fully
charged
battery)
imecl
Battery Charging:
_/
Instrument Operating at 125
mL/hr:
Instrument-Off:
Automatic
ALARMS/PROMPTS:
Audio
Prime:
Characteristics:
AUDIO
MALFUNCTION
ALARM
PROMPT (F)
PROMPT
WALKAWAY
"PRESS"
ADVISORY
TYPE
(HT)
20
hours
7
hours
Stops
second
after
approximately1minute;
press
AIRINLINE,
BATTERY, RATE
AUDIO
CONTINUOUS
INTERMITTENT
CONTINUOUS CONTINUOUS
DOOR
000,
NOT
OFF
AUDIBLE
manual
of PRIME ON/OFF control
OCCLUSION,
OPEN,
CHARGING,
and
PRESS
SOUND
WARBLE
WARBLE
WARBLE
BEEP
INFUSION
NOT
OPERATING,
SET 0000
stop
capability
COMPLETE,
RATE
VOLUME,
AUDIO
SILENCE
NO
YES
YES
YES
LOW
NOT
SET
with
000,
VOLUME,
The
960A
and
Laboratory's
the
Canadian
Equipment
965A
Volumetric Infusion Pumps comply with the
544,
Medical
and
Standard Association's
specifications.
Dental
NOTE
Equipment
C22.2
No.
(Second
125-M1984,
Underwriter
Edition,asamended)
Electromedical
and
1
fHS
1
1-4
p
imed
•fR>
1.5
The following administration
with
965A
in
ADMINISTRATION
the
960A
Micro
Table
Reorder
Number
1-3.
9200
9201
9205
9210
9211
9214
Volumetric
Volumetric
Infusion
ACCUSET™
ACCUSET
ACCUSET Administration
ACCUSET
ACCUSET Closed System Vented Administration
ACCUSET
SETS
sets
available for
Infusion
Pump
and
Pump are listed
Table
Administration
1-3.
Set
Administration Set
Set
Closed System Vented
Closed
System
Nonvented
use
Administration
Description
960A
with
Microbore
with
Microbore
with
Macrobore Patient Extension
Administration
Sets
Patient Extension
Patient Extension and 0.22
Set
Set
Administration
with
0.22
Set
with2Injection
Micron
Micron
Filter
Sites
Filter
9215
9216
9219
9224
9230
•MB
9235
9251
9252
9381
9382
9510
9512
ACCUSET
ACCUSET Closed System Nonvented Administration
ACCUSET
Chamber
ACCUSET
Chamber
ACCUSET
ACCUSET
Set
ACCUSET
ACCUSET
SECONDARY20Vented
SECONDARY20Nonvented
GRAVISET™20Vented
GRAVISET20Nonvented
Closed System Nonvented Administration Set
Closed
Closed
Closed
Closed
Closed System
Closed System
System
System
System
System
Primary
Vented
Administration
Nonvented
Vented
Quick
Nonvented
ADD-i-MED™
ADD-i-MED
Secondary
Administration
Secondary
Administration
Primary
Administration
Set
Administration
Spike
Quick
Vented
Nonvented Administration
Administration
(without
spike
Administration
Set
Set
Set
with
with
Set
with
Drip
(without
Set
Set
0.22
Metered
Metered
Chamber)
Drip
Chamber)
Set
Set
Micron
Filter
Chamber
Chamber
Administration
Administration
and
Drip
and
Drip
Set
9515
9517
GRAVISET
GRAVISET
20 Vented
Primary
20 Nonvented
with
Primary
0.22
with
1-5
Micron
0.22
Filter
Micron
Administration
Filter
Administration
Set
Set
imecl
9550
9552
9555
9557
9630
9635
9260
9261
9269
9271
GRAVISET20Vented
ADD-i-MED
GRAVISET20Nonvented
GRAVISET20Vented
Administration
Set
ADD-i-MED
GRAVISET20Nonvented
Administration
ACCUSET
Administration
ACCUSET
Administration
MICROSET™
MICROSET
MICROSET
Chamber
MICROSET
Micron
filter
Set
Closed
Closed
Administration
Closed
System
Set
System
Set
Administration
System
Closedsystem
Primary
ADD-i-MED
Primary
ADD-i-MED
NON-PVC
NON-PVC
Set
Set
with
Vented
Vented
Vented
Nonvented
965A
with
Patient
Patient
Administration
Administration
with
Primary
with
Primary
Extension
Extension
Check
with
Check
with
Quick
Quick
Set
Check
Check
Spike
and 0.22
Set
with
Valve
Valve
Spike
with
Administration
Valve
Valve
(without
Administration
and
0.22
and
Drip
(without
Micron
Metered
Metered
Chamber
Set
Micron
0.22
Micron
Chamber)
Drip
Chamber)
Filter
Chamber
Filter
and
and
Set
Filter
Drip
022
9274
9276
MICROSET
Chamber
MICROSET
Micron
Filter
Closed
Closed
System
System
Nonvented
Nonvented
Administration
Administration
Set
Set
with
with
Metered
Metered
Chamber
Chamber
and
and
Drip
r
0.22
1-6
p*
imed
2.1
INTRODUCTION
This
section
initial
the
the
inspection and pre-operational checkout of
IMED
965A
contains
960A
Micro
information relative to
Volumetric
Volumetric
Infusion
Infusion
SECTION
Pump and
Pump.
2 -
the
These
procedures include a mechanical inspection,
electrical inspection, pre-operational battery
charge and a performance check to ensure that the instrument operates properly and has not
been damaged during shipment or storage.
2.2
PRE-OPERATIONAL
INSPECTION
The
960A/965A
thorough testing shipping
production
prior
container has been designed to protect
instruments have undergone
control
to shipment
MECHANICAL
and
quality
from
the
factory.
assurance
The
the instrument against damage under normal
shipping conditions; nevertheless, internal
physical
and/or electronic component damage could have
occurred without leaving a visible signature.
Therefore, it is recommended that the
following
inspection procedure be performed upon receipt
of the instrument at the user's
1. Carefully remove the instrument from the shipping container. (It is recommended the
shipping material
be
saved
facility.
in the event the
instrument has been damaged and must be
returned to the factory for service or repair.)
2. Inspect the exterior
case,
front
and
rear, for
holes, cracks, scratches, broken or damaged
controls, missing components and/or screws.
PREPARATION
5. Check the cassette nest for damage or
distortion.
6. Inspect proper
7. Inspect the
8. Install an approved administration
instrument
correctly
9. Actuate
front panel
the rear panel to ensure proper operation
(see
10. Inspect the power cord for damage, bent
prongs
11. Exercise the pole clamp mechanism to
ensure
In
evidence
the
not
the
has
for
instrument
liability for repair
2.3
FOR
the
carrier's
returnadamaged
factory
authorized
authorization
USE
the
shuttle
installation.
Air-in-line
to
and
each
and
Figure 2-1 for location of controls).
or deformed connector.
freedom
event
of
shipping
before
boot
for
cutsortears
detector for damage.
ensure
the
door
the
cassette
closes.
of the keypad controls on the
the
Audio silence switch on
of
movement.
NOTE
the
instrument
damage,
agent
immediately. Do
instrument
the
carrier's
repairs.
to
Contact
return
the
for repair regardless
costs.
OPERATIONAL
PERFORMANCE
shows
notify
set
seats
to
agent
IMED
of
CHECK
and
for the
3. Inspect the green tinted covering the Rate and
clear window on the
or
cracks.
4.
Ensure
the
the top, bottom,
door
fits
and
LCD
flush
side.
Lexan®
VTBI
windows
displays and the
panel for scratches
with
the
case
at
Prior to the first operational use and
routine maintenance or servicing of the
960A/965A, it is strongly
operational performance
recommended
checkbeperformed.
The operational performance check consists of
two
phases;
a Pre-operational Electrical
Inspection to check the electrical integrity of the
instrument for compliance with regulatory agency
2-1
following
that an
any
irned
"
CASSETTE
AIR-IN-LINE
ON
STOP
VALVE
SHUTTLE
DETECTOR-
CONTROL
SHAFT-
NEST.
AfU
DOOR-
-OPERATING
-ALARM
LCD
DISPLAY
RATE (ML/HR)DISPLAY AND
'
INDCATOR
VOLUME
DISPLAY
PRESS
.131111110
-
START
VOLUME
PRIME
INDCATOR
INDCATOR
ALARM/PROMPT
TOBE
INFUSED
AND
INDCATOR
INDCATOR
CONTROL
INFUSED
ON/OFF
CONTROL
(ML)
CONTROL
FRONT
CORDCLIP
AC INPUT MODULE
A
FUSE
HOLDER
Figure
2-1.
960A/965A
REAR
Front
and Rear Operating Features
2-2
POLE
CLAMP
SONALERT
AUDIO
SILENCE
SWITCH
NURSE
CALL
IP-OxSoftwareInttrunwntt
CONNECTOR
ONLY)
imecl
requirements and an operational performance test
to verify instrument performance.
2.3.1
The
fully
able
Pre-operational Check Battery Charge
instrument
charged condition. However, since consider
time
could
is shipped
elapse
between
with
the
batteryina
shipping
and
first
use, a pre-operational battery charge is
recommended. Connect the
suitable
for
24
AC
hours.
outlet
and
allow
AC
power cord to a
the
batterytocharge
2.3.2 Pre-operational Electrical Inspection
The pre-operational electrical inspection
an
electrical
check
andadielectric
leakage test, a
test.
CAUTION
ground
includes
continuity
Some of these tests are inherently
hazardous. Safeguards for personnel and
property
should
be
employed
when
conducting such tests. Tests should only
be performed by qualified personnel.
CSA Standard C22.2 No. 125 for Risk Class 2
Equipment.
2.3.3 Abbreviated Operational Performance
Test
The
following
designedtoensure
controls and indicators is
operational performance test is
each
of the 9Q0A/965A
functioning
properly
and
check the operability of all the features available in the normal operating modes.
2.3.3.1
The
Test
following
Requirements
items of laboratory equipment and
supplies are required to conduct the operational
performance
1. ACCUSET Closed System
administrative
2. IV
3.
Standard
4. Nurse Call Test (Software Release P-0.x/1.x
Solution
tests:
Container
IV
Pole
set
Light
ADD-i-MED
(see
figure 2-2)
ONLY)
2.3.2.1 Electrical Leakage
Test
Perform an electrical leakage current measure
ment in compliance with Underwriters Labora
tories Canadian Standards Association
C22.2
Leakage
microamperes.
2.3.2.2
Perform
measurement
(UL)
544
for
Patient Care
Equipment
(CSA)
No.
125 for
currents
Electrical
Risk
Class 2 Equipment.
aretobe
Ground
less
Test
than
an electrical ground impedance
in compliance with UL 544 for
Standard
100
Patient Care Equipment and/or CSA Standard
C22.2 impedance between the
power cord plug and the grounding
rear
2.3.2.3
Performadielectric
with
No.
125
for
Risk
case
should
Dielectric
not
Class 2
grounding
exceed
Test
Equipment.
pin
100
milliohms.
on the
point
on the
withstand testincompliance
UL 544 for Patient Care Equipment and/or
and/or
The
5.
Hemostat
6. Empty fluid container
7. Gas
8. Fluid
9. Fluid simulator,
10.
a—OE
2.3.3.2
tight
syringe: 100
simulator,
Test
magnet (door alarm override).
NURSE
CALL
PUJC
clear
opaque
•V
ALKALINE
+ l I
—HI
I—
fluid
Figure 2-2. Nurse Call Test
Test
Procedures
microliter
fluid
390 OHU,
capacity
I/4W.3X
WV
Light
The following tests and associated procedures
are
presented
efficient, qualitative
operability.
in a
sequence
check
of instrument
that
provides an
LLD
2-3
\
imed
INITIAL
SETUP
1. Mount pump onIVpole (leave AC power
cord
unplugged).
2. Spike
3. Install Nurse Call
and
hangIVfluid container above the
instrument, then
fill
Test
connector. (Software
drip
Light
Release
chamber
2/3
in the Nurse Call
P-0.x/1.x
full.
ONLY)
NOTE
Test
procedures
that
refer to operation
the Nurse Call Test Light are applicable
ONLY
Release
TEST
to
instruments
P-O.x
SEQUENCE
installed.
with
Software
1. Press ON/STOP key and check:
The
valve
position
the down position
and
shaft
the
will
hometothe
shuttle
arm
will
10 o'clock
home
NOTE
The
"PRESS"
indicators
test
check.
2. Plug the AC power cord into a suitable
outlet
• The
and
"NOT
check:
and
OPERATING
do
not
flash during
CHARGING"
message
LED
the
self-
wall
extinguishes
The
LCD panel backlight illuminates.
of
3. If
4. If VTBI not "OOOO7"000.0",
5.
6. Open the
to
rate
not
MOOO7"00.0"I
"OOOO7"000.0".
If
"VOLUME
VOL
INF
display on the
INFUSED"
RESET.
door.
LCD
"DOOR
panel (red
indicator and alarm audio
Nurse Call
test
light remains illuminated.
set
to "OOO7"00.0".
set
to
is not
OPEN"
"O'TO.O",
message
ALARM
will
remain active).
press
will
LCD
Panel-"VOLUME
"8888"/"888.8" illuminates for 3
then
reverts
to "VOLUME
INFUSED"
seconds,
INFUSED"
last volume infused value or "07*0.0", if
the
power
hours
has
been
off for more than 4
• Operating/Rate graphics - illuminate
steadily for 3 seconds, then go blank
• Alarm/Prompt messages display
for
seconds, then go blank except for
OPERATING"
which display steadily and
which cator
flashes.
flashes
The Nurse Call test
RATE
•,8887"88.8H
and
momentarily, then display
in the Rate and
and
The
and
VTBI
Displays show
and
"88887"888.8"
"P-X.X"
"NOT CHARGING"
"AIRINLINE"
red
ALARM LED indi
the
alarm
audio
light
illuminates
"96.0"or"96.5"
(X.X
refers to the
installed software revision) in the =2 seconds
followed
by reversion to the
last confirmed infusion parameters or
"OOO7"00.0"
power
and "00007"000.0", if the
has
been
off for >4
hours.
and
3
"NOT
sounds.
VTBI
for
7.
Install
cassetteinthe nest and
affix
to shuttle arm. Press tubing into the Air In
Line
detector.
Do
not
connected
8. Ensure
9. Press
priming.
"AIRINLINE"
"NOT
steadily on the
and
• Rate and
and
10.
Checktoensure
tubing
finish.
prime
all
PRIME
visual
—-/—.-
set.
Close
the
WARNING
with
the
door.
administration
to a patient.
clamps are open.
ON/OFF key once to initiate
message
OPERATING"
LCD
alarm
VTBI
indications
displays
will
message
panel and the audio
will
respectively for 960
Allow
all air is
the auto-prime
cleared
flash and
will
show ---/-.-
plunger
will
display
continue.
A/965A.
from the
cycle
set
to
2-4
11.
Check
allowing instrument to remain idle
observe:
• After 45/80
12.
Set
25/70.0
walk
away
alarm
seconds
flashes,
flashes, the audio alarm
"NOT
OPERATING"
Nurse Call test
rate to 125/50.0
mL,
then press START and check:
All
indicators reflect
light
mLThr
red
illuminates.
normal
(operator
"ALARM"
message
sounds
and
VTBI
operation.
alert)
and
indicator
and
to
by
the
"INFUSION
on
the
COMPLETE"
LCD
message
• Operating/Rate graphics sequence at
rate
Audio
• Nurse Call test light illuminates
• Rate display alternates between
"1257"50.0"
18.
Press
Audio
"AUDIO
LCD panel.
Alarm
AUDIO
alarm
sounds
and
Silence
silences
"17"0.r
switch
mL/hr.
and
OFF" message appears on the
imecl
flashes
KVO
check:
13. Place hemostat on the
check:
• ALARM
OPERATING
indicator
indicator extinguishes
• Operating/Rate graphics
"OCCLUSION"
OPERATING"
panel
Audio
• Nurse Call test
Alarm
displays steadily on the
sounds
distal
tubing
flashes
extinguish
message flashes and
light
illuminates
• Pumping mechanism stops.
14.
Remove
check:
All
15.
Place
check:
• ALARM
OPERATING
• Operating/Rate graphics
"OCCLUSION"
OPERATING"
panel
Audio
• Nurse Call test
Pumping
16. Remove the hemostat, press
check:
All
17. Set
await Infusion Complete alarm, then check:
• ALARM
OPERATING
the hemostat, press
indicators
reflect
hemostat on the
indicator
indicator extinguishes
Alarm
sounds
mechanism stops.
indicators
VTBI
to 5/2.0
reflect
indicator
indicator
START
normal
proximal
flashes
operation.
tubing
extinguish
message flashes and
displays
light
steadily
illuminates
START
mL,
flashes
normal
press
operation.
START
extinguishes
and
and
on the
and
and
"NOT
LCD
and
"NOT
LCD
19.
Press
20.
Open
set, install the test magnet over the door
sensor
fluid
OPTICAL AIR
configured
step
21. Press
START
22.
Slide
the
23.
Slide
the
24. Expose the third
• ALARM
ON/STOP to
the door, remove the administration
and
put the filled portion of the clear
simulatorinthe
IN
with
Ultrasonic
#32).
ON/STOP,
and
check:
All
indicators reflect normal operation.
the clear
first
sensor
All
indicators
the clear
second
All
OPERATING
Audio
sensor
indicators reflect
indicator
Alarm
shut
down
AIL
detector.
LINE
set
fluid
to
air
reflect
fluid
TEST
simulator
simulator
to
(If
AIL
detector proceed to
VTBI
to 100/40.0, press
and
check:
normal
air
and
normal
sensor
to air
flashes
indicator extinguishes
sounds
• Operating/Rate graphic extinguishes
"AIRINLINE"
OPERATING"
• Nurse Call test light illuminates
• Pumping mechanism stops.
25.
Remove
and
all sensors
check:
All
indicators reflect normal operation.
message
flashes and
message displays steadily
invert
see
fluid,
simulator
press
then
START
the
instrument.
instrument is
downtoexpose
operation.
downtoexpose
check:
operation.
and
check:
"NOT
install
so
and
2-5
imed
26. Slide simulator up to expose the lower 3
sensors
to
air
and
check:
All
indicators reflect normal operation.
27. Slide simulator up to expose the lower 4
sensors
• OPERATING indicator extinguishes
ALARM
Audio
to
air
indicator
Alarm
and
sounds
check:
flashes
• Operating/Rate graphic extinguishes
"AIRINLINE" OPERATING"
• Nurse Call
message
message
test
light illuminates
flashes and
"NOT
displays steadily
• Pumping mechanism stops.
28.
Press
simulator
simulator
START
ON/STOP,
and
so
and
All
indicators reflect normal operation.
remove
replace with the
all
sensors
check:
see
the
clear
opaque
fluid,
fluid
then
press
29. Slide the simulator up to expose the lower 2
sensors
«
• OPERATING indicator extinguishes
ALARM
Audio
to
air
indicator
Alarm
and
sounds
check:
flashes
• Operating/Rate graphic extinguishes
"AIRINLINE"
OPERATING"
• Nurse Call
message
message
test
light illuminates
flashes and
"NOT
displays steadily
• Pumping mechanism stops.
30.
Press
simulator
START
ON/STOP, reposition the
so
all
and
sensors
check:
see
fluid, then
opaque
press
ULTRASONIC
32.
Press
START
All
33. Slide the fluid simulator down to
sensor
ALARM
AIR
IN
LINE
TEST
ON/STOP,
and
check:
set
VTBI to 100/40.0,
indicators reflect normal operation.
expose
to
air
and
check:
indicator
flashes
• OPERATING indicator extinguishes
Audio
Alarm
sounds
• Operating/Rate graphic extinguishes
"AIRINLINE"
OPERATING"
• Pumping mechanism
• Nurse Call
34. Reposition liquid,
All
35.
Press
Instrument
36.
Remove
sensor
detector.
fluid
press START and check:
indicators reflect normal operation.
ON/STOP
the
and
Remove
message
message
stops
test
light illuminates.
simulator so sensor
and
check:
shuts
down.
test
the
magnet
fluid
the
from
simulator
Nurse
flashes and
displays steadily
the
door
from
Call
Test
from the Nurse Call receptacle, if appropriate.
The instrument is now ready for patient
service.
See
Section
3 -
OPERATION
for
standard operating procedures.
will
the
care
1
press
"NOT
see
AIL
Light 6
(w?k
All
indicators reflect normal operation.
31. Slide the simulator down to
4
sensors
ALARM
to
air
indicator
and
check:
flashes
expose
• OPERATING indicator extinguishes
Audio
Alarm
sounds
• Operating/Rate graphic extinguishes
"AIRINLINE"
OPERATING"
Nurse
Call
• Pumping mechanism
message
message
test
light illuminates
flashes and
displays steadily
stops.
the upper
"NOT
1
2-6
imed
SECTION
3.1
INTRODUCTION
This
section
procedures
Vo
umetric
Volumetric
intended to provide maintenance technicians
a basic understanding of
including
system.
Although
and tested to exacting specifications,
the
replace
in
the
The
professional
utilization
describes
for
operationofthe
Infusion
Infusion
the
audio
the
instrument
the
supervision
user
is
of
the
recommended
IMED
960A
Pump
Pump.
and
960A&965A
role of medical personnel
urged
standards
the
and
The
instrument
visual
NOTE
is
not
intended
ofIVinfusions.
to
exercise
of
instruments.
the
965A
Micro
information
operation
alarm/prompt
are
built
to
care
in
the
is
with
3 -
OPERATION
3.2
The
operate
Figure
3-1.
italics,
CONTROLS
controls
Items
and
the
960A
3-1
and
uniquetothe
listed
AND
INDICATORS
indicators
and
965A
with
descriptionsinTable
965A
usedtosetupand
are
illustrated
are
presented
in
in
3-1
imed
ED
OPERATING
VOLUME
tssmmm
AIRINLINE OCCLUSION
INFUSION LOW
BATTERY DOOR MALFUNCTION
V
RATE
tt
-0-
ON
STOP
OPEN
(ML/HR)
tr
43-
INFUSED
tasaaaai
COMPLETE
-L? 43-
ON/OFF
PRIME
laataaaa
j
960A
niff™
NOT
OPERATING
RATE NOT
CHARGING
SET
0000
AUDIO
VOLUME
tr
-0-V43-
VOL
INF
RESET
^. J
88
as&ssaaa
NOT
000
VOLUME
OFF
TOBE INFUSED
u
tt
J3-
PRESS
START
ALARM
88]
J
(ML)
-a
43-
IPFR&TIMfti'!'|V^Il\/
VOLUME
AIR IN OCCLUSION INFUSION LOW
DOOR
MALFUNCTION
RATE
tr
43- 43-
INFUSED
LINE
COMPLETE
BATTERY
OPEN
0.1 ML/HR U
(ML/HR)
tr
ON
STOP
NOT
OPERATING
RATE NOT
CHARGING
SET
0000
AUDIO
VOLUMETO BE INFUSED (ML)
-t?
43-
ON/OFF
PRIME
•a
43-
VOL
RESET
965A
8 8
NOT
000
VOLUME
OFF
-cr
tr
43- 43- 43-
^RESS
INF
START
fti
ARM
8.8
tr
Figure
3-1.
960A/965A
Controls
3-2
and
Indicators
Wl
imecl
Control/Indicator
OPERATING
ALARM
Indicator
Indicator
ALARM/PROMPT
RATE
(ML/HR)
Controls/Display
VOLUMETOBE
Controls/Displays
ON/STOP
Control.
Display
INFUSED
Table
(ML)
3-1
960A/965A
- Functional Description
Green
Red
condition.
Backlighted
INFUSED"
on and either the last Volume Infused value or
Operating/Rate sequencing graphics; and eleven
alarm/prompt messages, when appropriate.
Six
into
separately controlled. Digits do not rollover.
Eight
infused
separately controlled. Digits do not rollover.
Applies
mode. When in
powered down except for the microprocessor
remains powered for 4 hours.
Controls
and
Indicators
Functional Description
LED
that
LED
that flashes when the instrument isinan
arrow
(0)0
the three
arrow
into
flashes
LCD
panel that displays:
message
controls
RATE
®@
controls
the
four
when
the
instrumentisinfusing.
"VOLUME
whenever
the
instrument
usedtoinputaninfusion
digit
displays. Each
usedtoinputavolumetobe
VTBI
digit
displays. Each
system power to the pump when in the
the
STOP
mode, the instrument is
digit
power
"0";
is
digit
ON
which
alarm
is
rate
is
PRIME
START
VOLUME
RESET) Control
•PRESS"
AUDIO
fpl
ON/OFF
Control
INFUSED
Indicator
Switch
Control
RESET
(Rear Panel)
(VOL
INF
Operates the
±3) when the rate is set to 000
manually stopped by actuating the PRIME ON/OFF
control during the automatic
Initiates
rate and/or
restarts an infusion
condition.
Clears the
INFUSED"
"00007*000.0".
Illuminates if
within
VTBI
Temporarily
When silencing is invoked,
conditions will be
10 seconds
during an infusion.
pumping
mechanism
(00.0).
priming
for
63 ±3 seconds
Priming
sequence.
operation of the pumping mechanism,
VTBI
changes input during an infusion and
following
Volume
display, when actuated
Infused register and
the
START
following
silences the audio alarm when
sounded
clearing of an alarm
"VOLUME
with
the
VTBI
control
has
not
been
the changing of rate and/or
permitted.
NOTE
subsequent
upon
occurrence.
alarm
can be
confirms
actuated
(72
set to
3-3
imed
3.3
NORMAL
OPERATION
5. If
Rate
not
"OOO7"00.0",
setto"0007*00.0".
The following
steps
necessary
and the 965A
procedures
to
operate
Micro
Volumetric Infusion Pumps.
describe in detail the
the 960A Volumetric
Operating instructions are also provided in the
Operator's Manual.
3.3.1
1. Attach the 960A/965A to the
2. Plug
3. Spike
Setup
power
the
outlet.
and
Procedure
AC
power
hang the
IV
pole.
cord into a suitable AC
IV
fluid
container in
accordance with accepted hospital
procedures.
4.
Press
instrument,
the
ON/STOP
then
check:
LCD
Panel-"VOLUME
"88887"SSS.S"
then
reverts
last volume infused
power
has
illuminates for 3
to
"VOLUME
been
off for
controltoturn
INFUSED"
INFUSED"
value
or 0/0.0, if the
more
than4hours
on
the
seconds,
and
• Operating/Rate graphics - illuminates goes
steadily for 3 seconds, then
blank
6. If
7.
8. Open the door.
VTBI
not "OOOO7"000.0",
"00007*000.0".
If
"VOLUME
INF
RESET.
INFUSED"
"DOOR
is not
OPEN"
display on the LCD panel (red
indicator and alarm audio
9.
Install
to shuttle arm.
Line
Do
connected
10. Ensure all clamps are open.
11.
Press
priming.
cassette in the nest and
Press
detector.
not
Close
WARNING
prime with
toapatient.
PRIME
ON/OFF
"AIRINLINE"
OPERATING"
and
the operating/rate graphic
the
LCD
panel.
alarm
indications
tubing into the
the
the
message
message
The
will
door.
administration
control
audio
continue.
set
to
"0",
press
message
ALARM
will
remain
affix
once
will
flash, "NOT
will
display steadily
will
and
visual
VOL
will
active).
plunger
Air-ln-
set
to initiate
flash on
•—
-
• Alarm/Prompt
seconds,
OPERATING"
"AIR IN LINE" ALARM
alarm
RATE
"8887*35.8"
audio
and
messages
display for 3
then go blank except for "NOT
LED
which
which
indicator
sounds
VTBI
and
displays steadily and
flashes.
flashes
Displays show
"QBQ8"/"888.8"
The
and
momentarily, then display "96.07*96.5" in
the Rate
"P-X.X" (where
X.X
and
the installed software revision) in the
for =2
last
"0000TOOO.O",
for
The
indicators do not flash during
test
seconds
set
values
>4
hours.
"PRESS"
check.
followed by reversion to the
or
"OOO7"00.0"
if the power
NOTE
and
OPERATING
has
and
been off
the
red
the
refers to
VTBI
LED
self-
• Rate and
VTBI
displays
will
and -—/--.- respectively.
12.
Checktoensure
tubing set, then
manually
stop
all
airiscleared
press
PRIME ON/OFF to
priming or allow automatic
cycle to complete.
13. Attach administration
set
to the patient's
indwelling venipuncture device.
14. Use
15. Use
3.3.2
RATE
VTBI
Starting
arrow @@ controls to
arrow
©03 controls to set
an
Infusion
1. Press START to begin the infusion, then
check:
The
OPERATING
LED
indicator
• The Operating/Rate graphic begins
sequencing
3-4
show —
from
set
VTBI.
flashes
/-.-
the
RATE.
_
imed
r
M
The
"NOT
extinguishes.
If START is not pressed,
OPERATING"
illuminated.
Rate and/or VTBI entry
indicator
display(s) will flash,
an
audio
not
pressed
data entry,
3.3.3 During
1.
The
VOLUME
2. The
OPERATING
graphic
OPERATING"
NOTE
message
Within
plus
prompt
within5minutes
the
the
ten
the
Rate
will
pump
Infusion
INFUSED
LED
will
flash and sequence respectively.
message
the
"NOT
will remain
seconds
the
"PRESS"
and/or VTBI
as
appropriate, and
sound.IfSTART
of
will
shut
down.
value
will
and Delivery/Rate
of
the
increment.
3. Various alarm conditions may arise require operator attention. An alarm message
will
be displayed on the
ALARM LED will
will
sound.
Alarm/Prompt
recommended
See
messages
actions.
flash
Table 3-2 for a
LCD
and
and the
panel, the red
the
alarm
listing
the
last
which
audio
of the
last
is
will
LCD
panel
battery
will
remain backlighted and
will
continue to charge as
the AC power cord is connected to an
power
source.
3.3.6 Changing Rate or VTBI during an
Infusion
1.
Use
the
arrow
controlstoinputanew
rate
and/or
• New rate and/or
VTBI.
VTBI
value
will
appear
the respective display.
2.
Press
the
START
controltoconfirm
initiate the new parameter(s).
• If START is not
seconds,
an audio prompt
and/or
The
instrument
the
previous
START
3.3.7
control is
Resetting
the
VTBI
"PRESS"
display
NOTE
will
rate
Volume
pressed
will
sound
will
continue
and/or
pressed.
Infused
within
indicator will flash,
and
10
the rate
flash.
to
infuse
VTBI until
long
and
the
as
AC
in
at
JPB
3.3.4 Infusion Complete
1. When the programmed volume to be infused
has
been
delivered,
an
INFUSION
KVO
infusion.
• The
"INFUSION
flash
LED
will
sound.
on the
flash
• The infusion rate
mL/hr or 0.1
the
instrument
COMPLETE
COMPLETE"
LCD
panel, the red
and
an
audio
will
change to
mL/hr)
and
will initiate
alarm and begin a
message
ALARM
alarm
"1" or "0.7"
will
KVO
(1
will
flash alternately with the programmed rate
on the Rate display.
3.3.5
1.
Stopping
Press
the
an Infusion
ON/STOP
control:
• Pumping mechanism
All
indicators and displays
will
stop
will
extinguish
will
1. Use the arrow
display to "0000" or "000.0".
2.
Press
VOL
INF
"VOLUME
"0"or"0.0"
RESET
INFUSED"
"INFUSION COMPLETE"
display
"OPERATING"
flashing
LED
Operating/Rate graphic
rate
ALARM
audio
3. Use the arrow
display to
4.
Press
"OPERATING"
the
LED
alarm
will
the
desired value.
START
indicator
control:
LED
• "INFUSION COMPLETE"
extinguish
controls to
control:
value
indicator
will
will
sound.
controls to
indicator
set
will
reset
message
will
reflect
flash
and
set
the
will
message
the
will
stop
KVO
flash
will
VTBI
to
the
VTBI
3-5
imed
"ALARM"
alarm
• Operating/Rate graphic
programmed delivery rate.
If
the
instrumentisstopped
infusion, the valve shaft
audio
LED
will
indicator
silence
NOTE
will
extinguish and
will
reflect
during an
will
stop at its
present position. To load a
turn
"home"tothe
3.3.8 Clearing
1.
RefertoTable
and
Table 3-2. Alarm/Prompt Messages
NOTE
the
Response
instrument
loading position.
Alarms
on.
The
3-2.
Alarm/Prompt
Procedures.
new
valve
cassette,
shaft
Message
first
will
Alarm
applicable ONLY to instruments with Software Release
Message/Presentation
VOLUME
AIR
Audio
OCCLUSION
Audio
Nurse
INFUSION
Visual - Flashing
Audio-Alarm Nurse
LOW
Visual - Flashing
Audio-Alarm Nurse
INFUSED
Visual-Continuous Audio-None
IN
LINE
Visual - Flashing
-
Alarm
Nurse
Call-On
Visual - Flashing
-
Alarm
Call-On
COMPLETE
Call-On
BATTERY
Call-On
responses
XXXX
that refer to operation of the Nurse Call feature are
Cause
Normal display of value in volume
infused register.
Air bolus detected by the
AIL
sensor. Pumping stops.
Tubing in line pressure exceeds
occlusion pressure
stops. Can occur during
normal operation.
Volume
switches to
Infused
KVO
limit.
Pumping
priming
equals
VTBI.
delivery rate.
or
Pump
Battery charge has dropped to 5.8 V
threshold. Pumping continues.
voltage drops to 5.4 V, a
MALFUNCTION alarm will
and
pumping
will
stop.
If
activate
P-0.X/1.X
None. Reset to
Clear air
START.
Clear
Set
instrument
Plug instrument AC power cord
into
installed.
Response
from
occlusion;
up another infusion or turn
off.
a suitable
"0"
as required.
tubing; Press
Press
AC
START.
power source.
DOOR
OPEN
Visual - Flashing
Audio-Alarm
Nurse
MALFUNCTION
Call-On
Visual - Flashing
Audio-Alarm
Nurse
Call-On
Door
is opened
during
Malfunction
the
stops (except Malfunction
an
infusion.
detected in any one of
monitored
code displays in
LOW
Infused quantity.
while
priming
Pumping
parameters.
BATTERY).
3-6
or
stops.
Pumping
Volume
If
priming,
PRIME
close door and press
ON/OFF
twice.Ifinfusing,
close door and press START.
Instrument must be repaired by
biomedical
returningtoservice.
BATTERY,
into
AC
technician
before
If LOW
plug AC power cord
outlet.
Table 3-2. Alarm/Prompt Messages
imed
jpi
IB
Message/Presentation
NOT
OPERATING
Visual
Audio•None Nurse
RATE
Visual-Flashes
Audio
RATE
Visual-Flashes
Audio
NOT
- Steady
Call-On
NOT
000
- Prompt
000
- Prompt
CHARGING
Visual - steady
Audio-None
SET
0000
VOLUME
Visual -
Audio
flashes
- Prompt
twice
twice
twice
Cause
Instrument power on, but has not pressed.
message
sounds.
PRIME
Rate
flashes
ON/OFF
set
>000 (00.0).
After
and
control pressed
START control pressed Rate
set
to 000 (00.0;.IfSTART
pressed
"ALARM" indicators flash
again, "PRESS"
START
=45 seconds,
alarm audio
with
the
and
and
alarm
audio sounds; otherwise, after 45
seconds
OPERATING"
alarm
"ALARM"
audio
indicator&"NOT
message flash and
sounds.
Instrument power on, with the AC
power
power
VOL
with
cord not
source.
INF
VTBI
connectedtoan
RESET
set
control
>0000 (000.0).
pressed
with
AC
Response
Press
START, or turn instrument
power off.
Set
Rate to 000 (00.0;, then press
PRIME
Set
then
ON/OFF.
Rate to a value >000
press
START.
(00.0),
Plug AC power cord into an AC
power
Set
press
outlet.
VTBI
to 0000 (000.0;, then
VOL INF RESET.
SET
'p
VOLUME
Visual -
Audio
flashes
- Prompt
twice
START
set pressed
"ALARM" indicators flash
control pressed
to 0000 (000.0/ If START
again, "PRESS"
with
and
and
VTBI
alarm
Set
VTBI to a
(000.0;, then
value
press
>0000
START.
audio sounds; otherwise, after 45
seconds
"NOT
jpi
AUDIO
OFF
Visual - Steady
Audio-Silenced
and
AUDIO
while
"ALARM"
OPERATING"
audio
alarm
switch
indicator
and
message flash
sounds.
(rear panel) pressed
an audio alert is sounding.
After temporary delay, audio
will
resume. May be silenced again,
permissible.
if
temporarily
PRESS
Visual•Flashes
Audio
- Prompt
Rate and/or
infusion,
and
START
rate
10 seconds have elapsed
START
control
and/or
VTBI
control
pressed
VTBI
changed
not
actuated
twice
setto000
during
with
or
an
Press
VTBI
START.
START,orset
to a value >0, then press
rate and/or
(00.0;/oooo (000.0;.
3-7
Loading...
+ 96 hidden pages