ILLKO REVEXplus USB, REVEXplus User Manual

Version 5.0-R2 ENG
04/2019
REVEX plus
USERS
MANUAL
REVEX plus USB
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
2
CONTENTS
1. INTRODUCTION .............................................................................4
1.1. Safety ................................................................................................................................ 4
1.2. Explanation of symbols on instrument .......................................................................... 4
1.3. General description ......................................................................................................... 5
1.4. Applied standards............................................................................................................ 6
1.5. Symbols used in this manual .......................................................................................... 6
2. DESCRIPTION OF THE INSTRUMENT .....................................7
2.1. Front panel and rear panel ............................................................................................. 7
2.2. Terminals.......................................................................................................................... 8
2.3. Keyboard and indicators ................................................................................................ 8
2.4. Included in the set ............................................................................................................ 9
2.5. Optional accessories ........................................................................................................ 9
3. PUTTING INTO OPERATION ................................................... 11
3.1. Connecting to mains outlet ........................................................................................... 11
3.2. Functionality test ........................................................................................................... 11
4. MEASUREMENTS ........................................................................ 13
4.1. Measurement mode control - the START key ......................................................... 13
4.1.1. Measurement mode ....................................................................................................... 13
4.1.2. HOLD function ............................................................................................................. 14
4.2. Earth bond 200 mA - R
PE
........................................................................................... 14
4.2.1. Test lead resistance compensation ............................................................................... 14
4.2.2. Earth bond measurement .............................................................................................. 15
4.3. Insulation resistance - R
ISO
........................................................................................ 16
4.4. Substitute leakage current - I
SUB
............................................................................... 17
4.4.1. Protective class I appliances ........................................................................................ 17
4.4.2. Protective class II appliances – substitute touch leakage current ............................... 18
4.5. PE current during operation - I
PE
and differential leakage current - I
............ 19
4.5.1. Protective class I one-phase appliances ....................................................................... 19
4.5.2. Three-phase and hard-wired appliances - EXT ......................................................... 21
4.6. Touch leakage current - I
D
......................................................................................... 22
4.6.1. Protective class II appliances ....................................................................................... 23
4.6.2. Protective class I appliances ........................................................................................ 24
4.7. Mains voltage and current consumption – U, I ........................................................ 25
4.7.1. Mains voltage measurement ......................................................................................... 26
4.7.2. Current consumption measurement .............................................................................. 26
4.7.3. Current measurement by means of current clamp - EXT .......................................... 27
4.8. Active power consumption, apparent power consumption and cos  P, S ......... 28
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
3
4.8.1. Power consumptions and cos  measured in test socket 1 ........................................... 28
4.8.2. Power consumptions and cos  measured by means of current clamp ........................ 29
4.9. Utilization of other external adapters – EXT ........................................................... 30
5. ADDITIONAL INFORMATION ................................................. 31
5.1. Displayed messages ........................................................................................................ 31
5.2. Reset of the instrument ................................................................................................. 32
5.3. Communication with PC (REVEXplus USB only) ..................................................... 32
5.4. Maintenance ................................................................................................................... 32
5.5. Calibration and service ................................................................................................. 32
5.6. Ecology ........................................................................................................................... 33
6. TECHNICAL SPECIFICATION ................................................. 34
6.1. General data ................................................................................................................... 34
6.2. Functions ........................................................................................................................ 34
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
4

1. INTRODUCTION

1.1. Safety

Read this User’s Manual carefully and completely and follow all instructions
contained therein. Otherwise using of the instrument may be dangerous for operator, for appliance under test or for the instrument!
Observe the following safety precautions:
Make sure that the instrument, measuring cables and all other accessories are in flawless
condition, e.g. no damaged insulation, no broken cables or plugs etc.
The instrument may be powered only from 230 V / 50 ÷ 60 Hz grounded mains outlet,
which is protected with a fuse or circuit breaker with a maximum rating of 16 A.
Only a trained, skilled person, who is familiar with hazardous voltage operations, can
handle the REVEXplus. Unexpected hazardous voltages can occur at appliance under test (dangerously charged capacitors etc.).
It is necessary to respect all safety regulations applicable to particular measurement.
Test Socket 1 of the REVEXplus must not be used for continuous power supply of
appliance. It is intended for test purposes only; max. test duration is 60 s @ 16A.
Dangerous voltage or mains voltage may be present on Test Socket 1 during some tests.
Appliance under test connected to Test Socket 1 of the REVEXplus can be powered by
mains voltage during some tests. The appliance can thus be activated.
The PE terminal is connected to PE terminal of Test Socket 1. Similarly, the I
D
terminal is electrically connected to PE terminal of Test Socket 1. When using the REVEXplus, do not connect any voltage to the terminals PE or ID. Otherwise there is a danger of electric shock or damage of the REVEXplus.
Use only standard or optional accessories supplied with the REVEXplus by your
distributor.
If there is reason to believe that safe operation has become impossible, put the instrument out of operation and secure it against any unintended operation. Safe operation must be presumed to be no longer possible, if:
The instrument does not operate properly any longer.
The instrument, cables, connectors, plugs or accessories exhibits visible damages.
The instrument was stored under unfavourable conditions for a long period.
The instrument was exposed to extraordinary stress caused by transport.

1.2. Explanation of symbols on instrument

Warning concerning a point of danger! Read User’s Manual and observe all precautions!
Protection class (double insulation).
All information, instruction and warnings stated in this User’s Manual are applicable
both for REVEXplus and REVEXplus USB unless otherwise stated.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
5

1.3. General description

The REVEXplus is professional portable appliance tester with many features packed into very compact case. The REVEXplus has extraordinary wide scope of use thanks to a lot of optional accessories, which enables to test practically any portable or hard-wired, one-phase or three­phase appliance.
FEATURES:
Earth bond test 200 mA AC.
Insulation resistance test 500 V DC.
Measurement of substitute leakage current and touch leakage current.
Measurement of differential leakage current and touch leakage current, possibility of L-N
reversal in test socket.
Measurement of PE current during operation and touch leakage current, possibility of L-N
reversal in test socket.
Functional test of tested appliance: active power consumption (W), apparent power
consumption (VA), current (A), cos φ.
Mains voltage.
Measurement of PE current during operation by means of optional current clamp.
Measurement of differential leakage current by means of optional current clamp.
Measurement of current by means of optional current clamp.
Functional test of tested appliance by means of optional current clamp: active power
consumption (W), apparent power consumption (VA), and cos φ.
Measurement of PE current during operation and differential leakage current of three-phase
portable appliance by means of optional adapters.
USB communication (REVEXplus USB only). The REVEXplus has furthermore additional features, which improves the safety of both
operator and the equipment under test:
Automatic check of dangerous contact voltage on PE of power outlet, from which the
REVEXplus is powered.
Automatic check, if PE of power outlet (from which the REVEXplus is powered) is
grounded.
Automatic check of leakage current amplitude of equipment under test.
Automatic check of external voltage absence during earth bond test.
The possibility to promptly test the basic functionality of the REVEXplus (built-in normal).
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
6

1.4. Applied standards

* Safety .................................................. EN 61010-1 + A2
* EMC .................................................... EN 55 022
EN 61000-4-2 EN 61000-4-3 EN 61000-4-4 EN 61000-4-5 EN 61000-4-6 EN 61000-4-11
* Equipment for testing, measuring or monitoring of protective measures EN 61557, part 2
EN 61557, part 4

1.5. Symbols used in this manual

Warning concerning a point of danger! Read User’s Manual and observe all precautions!
Hint, notice.
--
Control key – meaning in the text is “Press the key”.
(--)
Indicator – it is turned on if particular function is activated. The (START) indicator can blink or it can be turned on – it depends on
selected measuring mode.
Press and release the relevant key.
Press and release the relevant key repetitively for the purpose of switching between its several functions.
Press the relevant key and hold it pressed.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
7

2. DESCRIPTION OF THE INSTRUMENT

2.1. Front panel and rear panel

1
19
20
30
30
234
7 8
31-33
21-29
9-18
6
5
P,S
Ext
START
RR
PE R
I
I
I
I
I
Cal
LOCK
HOLD
ISOPE
PE
SUB
PROTECT
EXT
INV
PE
D
D
U,I
Fig. 1 – Front panel and rear panel
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
8

2.2. Terminals

1 Test Socket for connection of tested appliance’s power supply cord. 2 ID – touch leakage current test terminal. 3 RPE – earth bond test terminal. 4 PE – terminal for connecting of exposed conductive part of tested appliance (the
terminal is connected to PE terminal of Test Socket 1).
5 Power supply cord. 6 R
ISO+ISUB
– insulation resistance / substitute leakage current test terminal.
7 EXT INPUT – current clamp / three-phase adapter / other adapters connector. 8 USB connector (REVEXplus USB only).

2.3. Keyboard and indicators

Keyboard – each key selects one or more functions
9 RPE – Earth bond test. 10 I
SUB
Substitute leakage current and touch leakage current.
11 IPE – PE current during operation and touch leakage current. 12 I – Differential leakage current and touch leakage current. 13 ID – Touch leakage current on protective class I appliances. 14 R
ISO
– Insulation resistance.
15 U,IMains voltage and appliance’s current. 16 EXT – EXT INPUT for current clamp / three-phase adapter activation. 17 P,S – Active power / apparent power / cos φ. HOLD – Holding of displayed data. 18 CAL – Test lead resistance compensation on RPE function / basic functionality test. LOCK - Locks the START key. 19 START – Starts measurement.
Indicators
20 LED display.
 (M) − Insulation resistance in   (k) − Insulation resistance in k  () − Earth bond resistance in 
24 (mA) Leakage or touch current in mA. 25 (A) Supply current of tested application in A. 26 (V) Mains voltage in V. 27 (W) Active power in W. 28 (VA) Apparent power in VA. 29 (cos φ) cos φ.
Do not connect any external voltage to any terminal or connector. The only
exception is power supply cord 5 which is intended for powering of the REVEXplus from 230 V / 50 ÷ 60 Hz mains outlet.
Max. load on test socket 1 is 16 A, max test duration is 60 s.
Use original accessories only.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
9
Status indicators
30 Indicators of active status of the relevant key. 31 (INV) indicator of L-N reversal in the test socket 1. 32 (EXT) EXT INPUT is activated. 33 (PROTECT) Protection (“electronic fuse”) is activated.

2.4. Included in the set

The set includes all accessories necessary for testing of 230 V / 50 ÷ 60 Hz appliances equipped with movable power supply cord.
The set includes:
REVEXplus or REVEXplus USB.
P 2011 – Test lead, black, 2 m.
P 3011 – Test tip, black.
User’s manual.
Calibration certificate.
USB cable A-B (REVEXplus USB only).
CD with USB drivers (REVEXplus USB only).

2.5. Optional accessories

Carrying bag
P 6080 – bag for instrument and some accessories.
It enables measurement with instrument hung around operator’s neck.
Measuring cables and adapters
P 2012 – Test lead, blue, 2 m.
P 4012 – Crocodile clip, blue.
P 4011 – Crocodile clip, black.
P 3012 – Test tip, blue.
P 2021 – Extension lead, black, 5 m (for P 2011).
P 8015 – Three-phase adapter, 16 A socket, 5 contacts.
P 8016 – Three-phase adapter, 16 A socket, 4 contacts.
P 8017 – Three-phase adapter, 32 A socket, 5 contacts.
P 8018 – Three-phase adapter, 32 A socket, 4 contacts.
P 8030 – Adapter for testing of extension power supply cords.
Special accessories
P 8010 – Current clamp transformer.
WELDtest – adapter, which enables measuring of arc welding equipments output
voltage in accordance with EN 60974-4
PL 2051 – adapter, which enables measuring of medical instruments leakage current.
P 9080 – Stick-on label with appliance next test term marking.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
10
Special accessories for REVEXplus USB only
P 9020 – Barcode reader with PS/2 connector; barcode reader connects to a PC (USB
barcode reader is available, too).
P 9060 – Stick-on label with barcode.
P 9092 – Barcode printer PT-E300VP.
P 9100 – Cartridge TZ-241 for barcode printer; white, width 18 mm, length 8 m.
P 9101 – Cartridge TZ-231 for barcode printer; white, width 12 mm, length 8 m.
P 9102 – Cartridge TZ-221 for barcode printer; white, width 9 mm, length 8 m.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
11

3. PUTTING INTO OPERATION

3.1. Connecting to mains outlet

Use grounded mains outlet only to supply the instrument!
If mains outlet, mains cable, instrument’s case or accessories are damaged, do
not connect the REVEXplus to mains outlet!
The REVEXplus can be powered only from 230 V / 50 ÷ 60 Hz grounded mains, which is protected with a fuse or circuit breaker with a maximum rating of 16 A. The instrument is automatically switched on after plugging in to the mains outlet. Disconnect any equipment under test while connecting/disconnecting the REVEXplus to/from the mains outlet!
The REVEXplus automatically tests protective PE (ground) pin of mains outlet:
1) If PE connection is inadequate, then the (PROTECT) indicator lights up after the REVEXplus is plugged in to mains outlet.
2) If dangerous contact voltage arises on PE, then the (PROTECT) indicator lights up when operator's finger touches the key START. If operator presses the START key for all that, the measurement will not start and message „Prot“ appears on the display.
If test result of ground pin of mains outlet is unsatisfactory, the wrong outlet must be repaired before you plug in the REVEXplus to mentioned outlet again!

3.2. Functionality test

It can be sometimes difficult to identify that the measuring instrument does not work properly, especially during leakage current measurement. The REVEXplus has therefore built in the possibility to promptly test the basic functionality.
Functionality of following functions can be tested:
Insulation resistance.
Measurement of substitute leakage current.
Measurement of differential leakage current (only one L-N polarity).
Measurement of PE current during operation. (only one L-N polarity).
Disconnect equipment under test before the REVEXplus functionality test!
Functionality test procedure:
Disconnect equipment under test.
Use respective key to select the function you are going to test.
Press the CAL key and release it. The (CAL) indicator lights up.
Press the START key and release it. The test starts. After it is finished, the result is
displayed and the (CAL) indicator turns off.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
12
If you test I
PE
or I functions and displayed value is close to 0.000 mA, you must repeat
the test once more after switching the L-N polarity by repetitive press and release of either the IPE or I key. The (INV) indicator lights up in such case.
If the value 2.000 M ÷ 2.600 M is displayed, the insulation resistance measurement is correct. If the value 0.090 mA ÷ 0.110 mA is displayed, the leakage current measurement is correct.
Contact service if the test results differ from the ones described above.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
13

4. MEASUREMENTS

4.1. Measurement mode control - the START key
Beginning, process and ending of measurements is controlled by the START key.
START
START START
One measuring cycle
One measuring cycle starts an it is
completed.
START
START START
Continuous measurement
Measurement is in progress until
the START key is released.
1
START
Cal
LOCK
START START
Locking of measurement
Measurement continues though
both keys are released.
2
START
Ending of measurement.
1
START
START START
Holding of displayed data
Measurement is in progress and
displayed result is periodically
updated.
2
P,S
HOLD
START
Measurement is in progress and last
displayed result is displayed.
To cancel the HOLD mode press
and release the [HOLD] key again.
Fig. 2 – Measurement mode control

4.1.1. Measurement mode

The instrument enables to accomplish measurements in following three modes:
One measuring cycle – after the START key is pressed and released, one measuring
cycle starts and it is completed after few seconds. The result of measurement is displayed.
Continuous measurement – press the START key; measurement is in progress and
current measured values are displayed until the START key is released.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
14
Locking of measurement – press the START key and then press the LOCK key.
Release both keys. Measurement is in progress and current measured values are displayed until the START key is again pressed and released.
Measuring process is indicated by (START) indicator.

4.1.2. HOLD function

When measurement is in progress then current measured values are displayed. By press and release of the HOLD key the last measured result is displayed on display even if measurement either ended or continues. The HOLD function is indicated by such way that blinking of the (START) indicator changes to unremitting light. HOLD function can be cancelled (measurement is still in progress) by press and release of the
HOLD key. The (START) indicator then blinks again.
4.2. Earth bond 200 mA - RPE
Make sure that tested equipment is deenergized – its mains voltage must be
disconnected before starting the measurement!
Do not connect any external voltage between R
PE
terminal 3 and either PE in
test socket 1 or PE terminal 4.
If you are going to test equipment which can be charged to voltage > 25V (e.g.
filter capacitors), discharge it before test!
In case that external voltage is present on appliance under test, the
(PROTECT) indicator turns on and „Prot“ message is displayed. Such status is dangerous fault condition; disconnect equipment under test from the REVEXplus immediately to avoid its damage!

4.2.1. Test lead resistance compensation

Compensation means that the test lead resistance is measured and then automatically memorized and subtracted from all RPE measurements. The real earth bond resistance
is thus displayed. The compensation constant remains in instrument’s memory even if
the instrument is switched off. And so the lead compensation is valid until new compensation is carried out.
1
RPE
M
mA
cos
A V
VAW
K
RPE
PE
ID
2
Cal
LOCK
3
START
Fig. 3 – Test lead resistance compensation
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
15
Pres the R
PE
key and release it. Symbol „- - is displayed and indicators (RPE) and
() light up.
Connect test lead P2011 with test tip P3011 to the R
PE
terminal.
Connect test tip of the test lead either to the PE terminal or press it against the ground
pin of the test socket 1. If you use two test leads (which are connected to the PE terminal and the RPE terminal) then make short-circuit of their test tips.
Press the CAL key and release it. Now the indicators (CAL), (R
PE
) and () shine.
Press the START key and release it. Test lead resistance is shortly displayed, then the
(CAL) indicator turns off and 0.000“ is displayed. Test lead resistance compensation is
finished.
The compensation is effective for test lead resistance < 2 . If test lead resistance is ≥ 2  and compensation procedure is carried out, compensation is cancelled and all following displayed results are a sum of earth bond + test lead resistance. If compensation is active (it means that test lead resistance < 2 was memorized) then the (CAL) indicator lights up during RPE measurements.

4.2.2. Earth bond measurement

1
RPE
M
mA
cos
A V
VAW
K
Function selection
2
Connect tested appliance
Appliance connecting
3
Press the test tip connected to the RPE terminal against exposed conductive part of tested appliance
4
START
Cal
LOC K
Measurement
Fig. 4 – Earth bond measurement
Step 1 - preparation for measurement
Select the R
PE
function: pres the RPE key and release it. Symbol „- -“ is displayed and
indicators (RPE) and () light up.
Connect test lead P2011 with test tip P3011 to the R
PE
terminal.
If test lead compensation was not carried out yet, compensate it now – see details in
chapter 4.2.1.
Connect tested appliance:
- One-phase appliance with movable power supply cord – plug in its mains plug
to test socket 1.
- Three-phase appliance with movable power supply cord – connect PE pin of its
plug to the instrument’s PE terminal by means of optional test lead P2012 + crocodile clip P4012.
- Hard-wired appliance – firstly disconnect appliance from mains! Connect its
point of PE connection to the instrument’s PE terminal by means of optional test lead P2012 + crocodile clip P4012.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
16
Step 2 - measurement
Press test tip against exposed conductive part of tested appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator. If the (CAL) indicator lights on during measurement, it means that test lead resistance is compensated (see chapter 4.2.1. for details).
After finishing of measurement (the (START) indicator turns off) the lowest measured
value of earth bond resistance is displayed. If only „1“ is displayed (see chapter 5.1 for
details) it means that measured earth bond resistance is higher than maximal value which can be measured by the REVEXplus.
4.3. Insulation resistance - R
ISO
Make sure tested appliance is deenergized – its mains voltage must be
disconnected before starting of measurement!
Do not touch exposed conductive parts of tested appliance while measurement
is in progress – RISK OF ELECTRC SHOCK! Measurement in progress is indicated by the (START) indicator.
Do not disconnect test leads while measurement is in progress or immediately
after finishing the measurement. Capacitive component of tested appliance can be charged to voltage up to 750 V; after the key START is released, this voltage is automatically discharged! If the capacitive component would be so extraordinary high that the discharging circuitry in the REVEXplus would not be able to discharge it to voltage < 50 V, the message „Prot“ would be displayed. Tested appliance must be disconnected from the instrument in this case and its capacitance must be discharged by another alternative safe method.
In case that external voltage is present on tested appliance, measurement will not be carried out. Indicator (PROTECT) turns on and message „Prot“ is displayed.
1
RISO
M
mA
cos
A V
VAW
K
Function selection
2
Connect tested appliance; its mains switch must be switched on
Appliance connecting
3
START
Measurement
Fig. 5 – Insulation resistance measurement
Step 1 - preparation for measurement
Select the R
ISO
function: pres the R
ISO
key and release it. Symbol „- -“ is displayed
and indicators (R
ISO
) and (M) light up.
Connect tested appliance:
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
17
- One-phase appliance with movable power supply cord – plug in its mains plug
to test socket 1.
- Three-phase appliance with movable power supply cord – connect PE pin of its
plug to the instrument’s PE terminal by means of optional test lead P2012 + crocodile clip P4012. Then connect remaining current-carrying conductors by means of test lead P2011 + test tip P3011 to the instrument’s R
ISO
+ I
SUB
terminal (placed on rear panel).
- Hard-wired appliance – firstly disconnect appliance from mains! Connect its
point of PE connection to the instrument’s PE terminal by means of optional test lead P2012 + crocodile clip P4012. Then connect remaining current­carrying conductors by means of test lead P2011 + test tip P3011 to the
instrument’s R
ISO
+ I
SUB
terminal (placed on rear panel).
Step 2 - measurement
Switch on tested appliance’s mains switch.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator. It is recommended to hold the START key pressed 5-10 s (or even longer if displayed result is not stabilized) - the influence of tested appliance’s internal capacitance can be eliminated by this method.
After finishing of measurement (the (START) indicator turns off) the last measured
value of insulation resistance is displayed. At the same time either indicator (M) or (k) lights up and so the unit of measured insulation resistance is determined.
If only „1“ is displayed (see chapter 5.1 for details) it means that measured insulation
resistance is higher than maximal value which can be measured by the REVEXplus.
4.4. Substitute leakage current - I
SUB
Make sure tested appliance is deenergized – its mains voltage must be
disconnected before starting of measurement!
Firstly earth bond resistance (chapter 4.2.) must be tested; it is recommended
to test insulation resistance, too (chapter 4.3.). Then you can execute substitute leakage current measurement!
Do not touch exposed conductive parts of tested appliance while measurement
is in progress – RISK OF ELECTRC SHOCK! Measurement in progress is indicated by the (START) indicator.
Test voltage is max. 230 V / 50 ÷ 60 Hz, short circuit current is about 3.5 mA.

4.4.1. Protective class I appliances

1
ISUB
M
mA
cos
A V
VA
W
K
 
Function selection
2
Connect tested appliance; its mains switch must be switched on
Appliance connecting
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
18
3
START
Measurement
Fig. 6 – Substitute leakage current measurement
Step 1 - preparation for measurement
Select the I
SUB
function: pres the I
SUB
key and release it. Symbol „- -“ is displayed
and indicators (I
SUB
) and (mA) light up.
Connect tested appliance:
- One-phase appliance with movable power supply cord – plug in its mains plug
to test socket 1.
- Three-phase appliance with movable power supply cord – connect PE pin of its
plug to the instrument’s PE terminal by means of optional test lead P2012 + crocodile clip P4012. Then connect remaining current-carrying conductors by means of test lead P2011 + test tip P3011 to the instrument’s R
ISO
+ I
SUB
terminal (placed on rear panel).
- Hard-wired appliance – firstly disconnect appliance from mains! Connect its
point of PE connection to the instrument’s PE terminal by means of optional test lead P2012 + crocodile clip P4012. Then connect remaining current­carrying conductors by means of test lead P2011 + test tip P3011 to the
instrument’s R
ISO
+ I
SUB
terminal (placed on rear panel).
Step 2 – measurement
Switch on tested appliance’s mains switch.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
After finishing of measurement (the (START) indicator turns off) the last measured
value of substitute leakage current is displayed.

4.4.2. Protective class II appliances – substitute touch leakage current

1
ISUB
M
mA
cos
A V
VA
W
K
Function selection
2
Connect tested appliance; its mains switch must be switched on
Appliance connecting
3
Press the test tip connected to the ID terminal against exposed conductive part of tested appliance
4
STARTSTART
Measurement
Fig. 7 – Substitute touch leakage current measurement
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
19
Step 1 - preparation for measurement
Select the I
SUB
function: pres the I
SUB
key and release it. Symbol „- -“ is displayed
and indicators (I
SUB
) and (mA) light up.
Connect tested appliance:
- One-phase class II appliance with movable power supply cord – plug in its
mains plug to test socket 1.
Connect test lead P2011 + test tip P3011 to the instrument’s ID terminal.
Switch on tested appliance’s mains switch.
Step 2 – measurement
Press test tip against exposed conductive part of tested appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
After finishing of measurement (the (START) indicator turns off) the last measured
value of substitute touch leakage current is displayed.
4.5. PE current during operation - IPE and differential leakage current - I

4.5.1. Protective class I one-phase appliances

WARNING - DANGER!
If test voltage is present in Test socket 1 – it is indicated by the (START)
indicator, then tested equipment starts to run after its mains switch is switched on! Take into consideration all safety warnings, referred to equipment under test!
Do not touch exposed conductive parts of tested object after mains switch is
switched on!
Firstly earth bond resistance (chapter 4.2.) must be tested; it is recommended
to test insulation resistance if it is possible, too (chapter 4.3.). Then you can execute leakage current measurement!
Equipment under test must be isolated from ground (from the earth
potential) during measurement of IPE!
Test Socket 1 of the REVEXplus must not be used for continuous power
supply of appliance! It is intended for test purposes only; max. test duration is 60 s @ 16A.
The REVEXplus is equipped with electronic checking of excessive leakage current during measurement. If leakage current exceeds limit value about 12 mA, then mains voltage is disconnected from test socket 1. This status is indicated by displaying Prot“ message; the indicator (PROTECT) turns on, too.
1
IPE
M
mA
cos
A V VA
W
K  
PROTEC T
EXT
INV
Function selection
2
Connect tested appliance
Connect tested appliance to
test socket 1
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
20
3
START
Cal
LOCK
STARTSTART STARTSTART
Test socket 1 energizing
(Locking of the START
key)
4
Switch on tested appliance
Measurement
5
Switch off tested appliance
End of measurement
6
START
Test socket 1 deenergizing
(Unlocking of the START
key)
7
IPE
M
mA
cos
A V
VA
W
K
PROTEC T
EXT
INV
L-N reversal in test socket 1
8
START
Cal
LOCK
START START
Test socket 1 energizing
(Locking of the START
key)
9
Switch on tested appliance
Measurement
10
Switch off tested appliance
End of measurement
11
START
Test socket 1 deenergizing
(Unlocking of the START
key)
Fig. 8 – Leakage current measurement
Step 1 - preparation for measurement
Select the I
PE
function or the I function: pres either the IPE key or the I key and
release it. Symbol „- -“ is displayed and indicators either (IPE) or (I) and (mA) light up.
Connect tested appliance: plug in its mains plug to test socket 1.
Step 2 – measurement with L-N default
Press and release either the I
PE
or the I key. If the indicator (INV) lights up, then
press and release corresponding key once more in order to turn off the indicator (INV).
Press the START key. It is recommended to lock the measurement - press the
STARTkey and then press the LOCK key. Release both keys. The presence of
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
21
voltage in test socket 1 and measurement in progress is indicated by the (START) indicator (see chapter 4.1. for measurement control details).
Switch on tested appliance’s mains switch and put it into operation.
You can activate HOLD function by pressing and releasing of the HOLD key after
stabilizing of displayed result (see chapter 4.1.2. for details).
If it is necessary test touch leakage current – see chapter 4.6. for details.
Switch off tested appliance by its mains switch.
Disconnect voltage from test socket 1 by pressing and releasing of the START key.
The indicator (START) turns off.
Step 3 – measurement with L-N reversal
Press and release either the I
PE
or the I key. If the indicator (INV) does not light up,
then press and release corresponding key once more in order to turn on the indicator (INV).
Proceed once more with above described procedure from Step 2, 2nd item.
4.5.2. Three-phase and hard-wired appliances - EXT
Optional accessories are intended for leakage current measurement of both three-phase and hard-wired appliances. Three-phase adapter must be used for measurement of three-phase appliances with movable power supply cord. It is necessary to plug in tested appliance by means of three-phase adapter and to select required measuring method by means of three-phase adapter’s function switch (see fig. 9) before measurement. Current clamp must be used for measurement of hard-wired appliances.
The EXT key activates external input EXT INPUT. If it is activated, the (EXT) indicator lights up and test socket 1 is not functional!
Connect either optional current clamp or three-phase adapter to the EXT INPUT
terminal on the rear side of the REVEXplus.
Press and release the EXT key, the (mA) indicator has to light up. The (EXT)
indicator is turned on, too.
Switch on tested appliance’s mains switch and put it into operation.
If you use current clamp then enclose with its jaws either protective conductor (PE
current during operation measurement) or all current carrying (L+N) conductors (differential leakage current measurement) of tested appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
After finishing of measurement (the (START) indicator turns off) the last measured
value of leakage current is displayed.
Switch off tested appliance’s mains switch.
1
IEXT
M
m A
cos
A V
VA
W
K  
PROTEC T
EXT
INV
Function selection
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
22
2
Connect optional accessories – either current clamp or three-phase adapter.
Appliance connecting
3
Switch on tested appliance’s mains switch and
put it into operation.
4
START
Measurement
Fig. 10 – Leakage current measurement by means of optional accessories
EXT
INPUT
epínač měřicí metody
P,S
EXT
RR
PE R
I
I
I
I
I
Ca l
LOCK
HOLD
ISOPE
PE
SUB
PROTECT
PE
D
D
U,I
EXT
INV
START
Fig. 9 – Three-phase adapter connection
4.6. Touch leakage current - ID
Touch leakage current measurements are performed with the same measuring methods like leakage current measurements. Thus all safety and working warnings and instructions stated for leakage current measurements are valid for touch leakage current measurements, too!
Touch leakage current can be measured by methods described in chapters 4.4. and 4.5. The measurement of substitute touch leakage current is described in chapter 4.4.2.
Measuring method
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
23

4.6.1. Protective class II appliances

1
IPE
M
mA
cos
A V VA
W
K  
PROTEC T
EXT
INV
Function selection
2
Connect tested appliance
Connect tested appliance to
test socket 1
3
START
Cal
LOCK
STARTSTART STARTSTART
Test socket 1 energizing
(Locking of the START
key)
4
Switch on tested appliance
5
Press the test tip connected to the I
D
terminal against exposed conductive part of tested appliance.
Touch leakage current
measurement
7
Switch off tested appliance
End of measurement
with L-N default
8
START
Test socket 1 deenergizing
(Unlocking of the START
key)
9
IPE
M
mA
cos
A V
VA
W
K
PROTEC T
EXT
INV
L-N reversal in test socket 1
Proceed once more with above described procedure from Step 3 to Step 8.
Fig. 11 – Touch leakage current of protective class II appliances measurement
Step 1 - preparation for measurement
Connect test lead P2011 with test tip P3011 to the I
D
terminal.
Accomplish preparation for measurement described in chapter 4.5.1. Step 1.
Step 2 – measurement with L-N default
Press and release either the I
PE
or the I key. If the indicator (INV) lights up, then
press and release corresponding key once more in order to turn off the indicator (INV).
Press the START key. It is recommended to lock the measurement - press the
STARTkey and then press the LOCK key. Release both keys. The presence of
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
24
voltage in test socket 1 and measurement in progress is indicated by the (START) indicator (see chapter 4.1. for measurement control details).
Switch on tested appliance’s mains switch and put it into operation.
Press the test tip of the test lead (connected to the I
D
terminal) against exposed
conductive part of tested appliance. Measurement starts; measured values are displayed.
Switch off tested appliance’s mains switch.
Disconnect voltage from test socket 1 by pressing and releasing of the START key.
The indicator (START) turns off.
Step 3 – measurement with L-N reversal
Press and release either the I
PE
or the I key. If the indicator (INV) does not light up,
then press and release corresponding key once more in order to turn on the indicator (INV).
Proceed once more with above described procedure from Step 2, 2nd item.

4.6.2. Protective class I appliances

There is used the same circuitry both for touch leakage current measurement and for leakage current measurement. It means the instrument measures at the same time both mentioned leakage currents. In order to separate these currents it is firstly necessary to measure PE current during operation and memorise its value by means of the I
D
key. Thereafter it is practicable to measure touch leakage current - formerly memorised value is automatically subtracted from the measured total value of leakage current and thus only touch leakage current is displayed.
Step 1 - preparation for measurement
Connect test lead P2011 with test tip P3011 to the I
D
terminal.
Accomplish preparation for measurement described in chapter 4.5.1 Step 1.
Step 2 – measurement with L-N default
Start the procedure described in chapter 4.5.1 Step 2.
As soon as displayed leakage current is stabilized, press the I
D
key and release it.
Displayed value is memorized, the (ID) indicator lights up and 0.000 mA is displayed afterwards.
Press the test tip of the test lead (connected to the I
D
terminal) against exposed conductive part of tested appliance. Measurement of the touch leakage current starts. Formerly memorised value is automatically subtracted from the measured total value of leakage current and thus only touch leakage current is displayed.
Switch off tested appliance’s mains switch.
Disconnect voltage from test socket 1 by pressing and releasing of the START key.
The indicator (START) turns off.
Step 3 – measurement with L-N reversal
Start the procedure described in chapter 4.5.1 Step 3.
Proceed once more with Step 2.
If the PE leakage current fluctuates and it is unstable during touch leakage current measurement, this fluctuation affects measured value of touch leakage current. In such case we recommend not to use the I
D
key for its memorizing, but to
remember the value which was displayed immediately before pressing of the test tip
against exposed conductive part of tested appliance and to subtract it “manually”.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
25
Fig. 12 –Touch leakage current of protective class I appliances measurement
4.7. Mains voltage and current consumption – U, I
Function U measures L-N voltage in mains outlet from which the REVEXplus is powered. Function I measures current consumption of tested appliance connected to test socket 1.
1
IPE
M
mA
cos
A V VA
W
K  
PROTEC T
EXT
INV
Function selection
2
Connect tested appliance
Connect tested appliance to
test socket 1
3
START
Cal
LOCK
STARTSTART STARTSTART
Test socket 1 energizing
(Locking of the START
key)
4
Switch on tested appliance
PE current during operation
measurement
5
ID
Memorizing of PE current
during operation.
6
Press the test tip connected to the I
D
terminal against exposed conductive part of tested appliance.
Measurement of touch
leakage current
(PE current during operation
is automatically subtracted;
displayed result is thus
correct).
7
Switch off tested appliance
End of measurement
with L-N default
8
START
Test socket 1 deenergizing
(Unlocking of the START
key)
9
IPE
M
mA
cos
A V
VA
W
K
PROTEC T
EXT
INV
L-N reversal in test socket 1
Proceed once more with above described procedure from Step 3 to Step 8.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
26

4.7.1. Mains voltage measurement

Press the U, I
key and release it repetitively in order to select voltage measurement
function U. The (V) indicator has to light up.
Press the START key and release it. Measurement is indicated by the (START)
indicator.
After finishing of measurement (the (START) indicator turns off) the last measured
value of mains voltage is displayed.
1
U,I
M
mA
cos
A V VA
W
K
 
Function selection
2
START
Measurement
Fig. 13 – Mains voltage measurement

4.7.2. Current consumption measurement

WARNING - DANGER!
If test voltage is present in Test socket 1 – it is indicated by the (START)
indicator, then tested equipment starts to run after its mains switch is switched on! Take into consideration all safety warnings, referred to equipment under test!
Do not touch exposed conductive parts of tested object after mains switch is
switched on!
Test Socket 1 of the REVEXplus must not be used for continuous power
supply of appliance! It is intended for test purposes only; max. test duration is 60 s @ 16A.
1
U,I
M
mA
cos
A V VA
W
K
 
Function selection
2
Connect tested appliance
Connect tested appliance
to test socket 1
3
START
Test socket 1 energizing
4
Switch on mains switch of tested appliance
Measurement
Fig. 14 – Current consumption measurement
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
27
Press the U, I
key and release it repetitively in order to select current consumption
measurement function I. The (A) indicator has to light up.
Connect tested appliance to test socket 1.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
Switch on tested appliance’s mains switch and put it into operation.
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
After finishing of measurement (the (START) indicator turns off) the last measured
value of tested appliance’s supply current is displayed.
Switch off tested appliance’s mains switch.
4.7.3. Current measurement by means of current clamp - EXT
The EXT key activates external input EXT INPUT. If it is activated, the (EXT) indicator lights up and test socket 1 is not functional!
Connect optional current clamp to the EXT INPUT terminal on the rear side of the
REVEXplus.
Press and release the EXT key repetitively until the (A) indicator lights up. The
(EXT) indicator is turned on, too.
Switch on tested appliance’s mains switch and put it into operation.
Enclose with current clamp’s jaws live conductor L of measured appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
After finishing of measurement (the (START) indicator turns off) the last measured
value of current is displayed.
Switch off tested appliance’s mains switch.
1
IEXT
M
m A
cos
A V
VA
W
K  
PROTEC T
EXT
INV
Function selection
2
Connect current clamp to the instrument and enclose with its jaws live conductor of measured appliance.
Connect current clamp
3
Switch on mains switch of tested appliance
4
START
Measurement
Fig. 15 – Current measurement by means of current clamp
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
28
4.8. Active power consumption, apparent power consumption and cos P, S
During measurement of any electrical quantity carried out in test socket 1 (measurements 1-3 in the following table) or by means of current clamp (measurements 4-6 in the following table) remaining quantities 1-3 or 4-6 are automatically measured and memorized. Measured results (1-3 or 4-6) can be then successively displayed by pressing/releasing of the P, S key. Currently displayed quantity is indicated by the indicator (W), (VA) or (cos ) and possibly (EXT).
Function
El. quantity
indicator
Input
indicator
1 2
Active power - tested appliance connected to test socket 1 Apparent power - tested appliance connected to test socket 1
(W)
(VA)
3
Cos - tested appliance connected to test socket 1
(cos )
4
Active power - by means of current clamp
(W)
(EXT)
5
Apparent power - by means of current clamp
(VA)
(EXT)
6
Cos - by means of current clamp
(cos )
(EXT)
4.8.1. Power consumptions and cos  measured in test socket 1
WARNING - DANGER!
If test voltage is present in Test socket 1 – it is indicated by the (START)
indicator, then tested equipment starts to run after its mains switch is switched on! Take into consideration all safety warnings, referred to equipment under test!
Do not touch exposed conductive parts of tested object after mains switch is
switched on!
Test Socket 1 of the REVEXplus must not be used for continuous power
supply of appliance! It is intended for test purposes only; max. test duration is 60 s @ 16A.
1
P,S
M
mA
cos
A V
VAW
K
Function selection
2
Connect tested appliance
Connect tested appliance to
test socket 1
3
START
Test socket 1 energizing
4
Switch on mains switch of tested appliance
Measurement
5
P,S
cos
VAW
Displaying of other
measured quantities
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
29
Fig. 16 – Power consumption and cos  measurement
Press the P, S
key and release it repetitively in order to select required function. The
(W), (VA) or (cos ) indicator has to light up.
Connect tested appliance to test socket 1.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
Switch on tested appliance’s mains switch and put it into operation.
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
After finishing of measurement (the (START) indicator turns off) the last measured
value is displayed. You can check other measured quantities press and release the P, S key repetitively.
Switch off tested appliance’s mains switch.
4.8.2. Power consumptions and cos  measured by means of current clamp
It is important to observe these principles for proper measurement:
Both REVEXplus and tested appliance must be powered from identical
phase.
Current clamp must enclose live conductor in a way that arrow on current
clamp points from power source to appliance. Mains outlet from which the instrument is powered has to have live contact located on left side. If apparent power value or cos value fluctuates randomly and even values out of specified measuring ranges are displayed then reverse either clamp’s or mains plug’s orientation.
1
IEXT
M
m A
cos
A V
VA
W
K  
PROTEC T
EXT
INV
Function selection
2
Connect current clamp to the instrument and enclose with its jaws live conductor of measured appliance.
Connect current clamp
3
Switch on mains switch of tested appliance
4
START
Measurement
5
P,S
cos
VAW
Displaying of other
measured quantities
Fig. 17 – Power consumption and cos  measurement by means of current clamp
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
30
Connect optional current clamp to the EXT INPUT terminal on the rear side of the
REVEXplus.
Press and release the P, S key repetitively in order to select required function. The
(W), (VA) or (cos ) indicator has to light up and (EXT) indicator has to light up, too.
Switch on tested appliance’s mains switch and put it into operation.
Enclose with current clamp’s jaws live conductor L of measured appliance in a way that
arrow on current clamp points from power source to appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
After finishing of measurement (the (START) indicator turns off) the last measured
value is displayed. You can check other measured quantities press and release the P, S key repetitively.
Switch off tested appliance’s mains switch.
4.9. Utilization of other external adapters – EXT
By repetitive press and release of the EXT key it is possible to select besides other functions the function, where both the (V) and (EXT) indicators are turned on. This function is intended for connecting of other adapters, e.g. the WELDtest - adapter enabling measuring of arc welding equipments output voltage in accordance with EN 60974-4.
The way of connecting of adapter to instrument, description of measuring
procedure and other necessary information are published in adapter’s user manual.
The EXT key activates external input EXT INPUT. If it is activated, the
(EXT) indicator lights up and test socket 1 is not functional.
It is prohibited to connect any external voltage to the EXT INPUT terminal! Use original accessories only!
User’s Manual REVEXplus, REVEXplus USB
Copyright © 20062019 ILLKO, s.r.o.
31

5. ADDITIONAL INFORMATION

5.1. Displayed messages

Operational state
The instrument is ready for measurement.
Overflow
Measured value is higher than the highest measurable value.
Protection activated
Protective circuitry is activated. Measurements will be possible after removal of fault.
PROTEC T
EXT
INV
Possible causes of protective circuitry activation:
Error indication
Action statement
Cause of fault condition
(PROTECT)
(+ eventually
“Prot” displayed)
The (PROTECT) indicator lights up instantly after the instrument is plugged in to mains outlet. If operator presses the START key for all that, the measurement will not start and message „Prot“ appears on the display.
PE connection in mains outlet from which the REVEXplus is powered is inadequate.
(PROTECT)
(+ eventually
“Prot” displayed)
The (PROTECT) indicator
lights up when operator’s
finger touches the START key. If operator presses the START key for all that, the measurement will not start
and message „Prot“ appears
on the display.
Dangerous contact voltage arises on PE.
(PROTECT)
+ “Prot” displayed
RPE measurement – at any time.
External voltage > 10 V is present on tested appliance.
(PROTECT)
+ “Prot” displayed
R
ISO
measurement – after the
START key is pressed.
External voltage > 10 V is present on tested appliance.
(PROTECT)
+ “Prot” displayed
IPE, I, ID measurement – at any time.
Leakage or touch leakage current exceeds about 12 mA.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
32

5.2. Reset of the instrument

It is advised to execute instrument’s RESET in case of any malfunction. Disconnect the
REVEXplus from mains, wait 30 s and plug it in again. If malfunction persists, contact your local distributor.

5.3. Communication with PC (REVEXplus USB only)

The Hyperterminal program (which is standard component of Windows installation) is utilized for communication with PC. Information related to USB driver installation, Hyperterminal setup and communication are on enclosed CD-ROM.

5.4. Maintenance

Regularly check the technical safety and integrity of insulation of power supply cord, plastic case and all accessories. Plastic case should not be polluted with substances, which can noticeably degrade case’s insulation quality. Use soft cloth, slightly moistened with lukewarm soap water for plastic case cleaning. Do not spill cleaning liquid over the instrument! Do not use cleaning liquids based on petrol, hydrocarbons etc. Wait until the instrument becomes totally dry before using it!
Do not expose the instrument to temperatures over 70 C, otherwise the plastic case can be damaged!

5.5. Calibration and service

Calibration
Measuring instruments should be regularly calibrated. We recommend interval of calibration 2 years. Furthermore we recommend carrying out calibration after each repair.
Service
Unauthorized persons are not allowed to open the REVEXplus. There are no replaceable components inside the instrument. Any claims for guarantee will become invalid in the event of damage that results from non-observance of the User’s Manual. We do not accept responsibility for such damage.
Contact your local distributor for further information.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
33
Manufacturer
ILLKO, s.r.o. Masarykova 2226 678 01 Blansko Czech republic
tel./fax: +420 516 417 355 e-mail: illko@illko.cz http://www.illko.cz

5.6. Ecology

Transportation packing
It is made of corrugated board; it is recyclable.
Instrument
This symbol signifies that the product should not be thrown away to municipal waste at end-of-life. Please dispose of this product according to the relevant statutory requirements.
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
34

6. TECHNICAL SPECIFICATION

6.1. General data

Protective class .......... II (double insulation)
Over voltage class ............................. CAT II
Pollution degree .......................................... 2
Degree of protection ............................. IP 20
Reference condition:
Line voltage . 230V1%/50Hz,60Hz 1%
Ambient temperature ............. (23 2) C
Humidity ............ 45 55 % RH @ 23 C
Position ...................................... arbitrary
Operating condition:
Line voltage 230V10%/50Hz,60Hz 1%
Ambient temperature ............ 0C 40C
Humidity ............ max. 75 % RH @ 23C
(condensation not allowed)
Elevation..............................max. 2000 m
Max. supply current .............................. 16 A
(it depends on tested appliance)
Quiescent current ............................  50 mA
Dimensions (w x d x h) . 150 x 195 x 50 mm
Mass (without accessories) .................... 1 kg
Storage temperature / storage humidity.........
.......................-10 50C / max. 75% RH
Case resistance to temperature ............. 70C
PE test ........................................... automatic
PC connection (REVEXplus USB only).USB

6.2. Functions

Earth bond 200 mA
Entire measuring range
0.000 ÷ 20.00
Measuring sub range
0.000 ÷ 0.999
Resolution
0.001
Reference error
(1 % of R + 5 D)
Operating error
(2 % of R + 5 D)
Measuring sub range
1.00 ÷ 20.00
Resolution
0.01
Reference error
(1 % of R + 3 D)
Operating error
(2 % of R + 3 D)
Operating range of use (according to EN 61557-4)
0.018 ÷ 20.00 Open-circuit voltage
AC 5 ÷ 12 V
Test current (R ≤ 2 )
≥ AC 200 mA
Test lead compensation
yes (up to 2 )
PE current during operation
Entire measuring range
0.000 ÷ 10.00 mA
Measuring sub range
0.000 ÷ 1.499 mA
Resolution
0.001 mA
Reference error
(1 % of R + 5 D)
Operating error
(2 % of R + 8 D)
Measuring sub range
1.50 ÷ 10.00 mA
Resolution
0.01 mA
Reference error
(1 % of R + 5 D)
Operating error
(2 % of R + 5 D)
Test voltage
line voltage
Frequency characteristic of internal test circuitry
according to EN 61010 Functionality test
0.090 ÷ 0.110 mA
Insulation resistance
Entire measuring range
0.000 k ÷ 100.0 M
Measuring sub range
0.0 k ÷ 999.9 k
Resolution
0.1 k
Reference error
(2 % of R + 8 D)
Operating error
(3 % of R + 8 D)
Measuring sub range
1.000 M ÷ 3.999 M
Resolution
0.001 M
Reference error
(2 % of R + 8 D)
Operating error
(3 % of R + 8 D)
Measuring sub range
4.00 M ÷ 19.99 M
Resolution
0.01 M
Reference error
(2 % of R + 8 D)
Operating error
(3 % of R + 8 D)
Measuring sub range
20.0 M ÷ 100.0 M
Resolution
0.1 M
Reference error
(4 % of R + 15 D)
Operating error
(5 % of R + 15 D)
Operating range of use (accord. to EN 61557-2)
3.0 k ÷ 100.0 M Nominal output voltage Un
DC 500 V
Open-circuit voltage
(-0%, + 50%) Un
Nominal test current
DC 1 mA
Short-circuit current
< DC 10 mA
Functionality test
2.000 ÷ 2.600 M
Mains voltage U
L-N
Measuring range
207.0 ÷ 253.0 V
Resolution
0.1 V
Reference error
(1 % of R + 2 D)
Operating error
(2 % of R + 2 D)
User’s Manual REVEXplus, REVEXplus USB
Copyright © 2006–2019 ILLKO, s.r.o.
35
Differential leakage current
Entire measuring range
0.000 ÷ 10.00 mA
Measuring sub range
0.000 ÷ 0.999 mA
Resolution
0.001 mA
Reference error
(1 % of R + 5 D + X*)
Operating error
(2 % of R + 10 D+ X*)
Measuring sub range
1.00 ÷ 10.00 mA
Resolution
0.01 mA
Reference error
(1 % of R + 5 D + Y*)
Operating error
(2 % of R + 5 D + Y*)
Test voltage
line voltage
Frequency characteristic of internal test circuitry
according to EN 61010 Functionality test
0.090 ÷ 0.110 mA
*X = S (0.01 IΔ + 0,005) *Y = S (0.001 IΔ + 0,005)
S….. Apparent power of tested appliance [VA] IΔ… Displayed value of leakage current [mA]
Substitute leakage current
Entire measuring range
0.000 ÷ 20.00 mA
Measuring sub range
0.000 ÷ 3.999 mA
Resolution
0.001 mA
Reference error
(1 % of R + 8 D)
Operating error
(2 % of R + 8 D)
Measuring sub range
4.00 ÷ 20.00 mA
Resolution
0.01 mA
Reference error
(1 % of R + 5 D)
Operating error
(2 % of R + 5 D)
Nominal open-circuit test voltage
230 V 15 % 50 Hz ÷ 60 Hz
Short-circuit current
3,5 mA
Internal resistance of test voltage source
80 k Functionality test
0.090 ÷ 0.110 mA
Leakage current by means of current clamp or three-phase adapter
Measuring range
0.00 ÷ 20.00 mA
Resolution
0.01 mA
Reference error
(3 % of R + 10 D) (1)
Operating error
(4 % of R + 15 D) (1)
Touch leakage current
Specifications are identical with measuring method selected for touch leakage current measurement.
Notes:
R - Reading D - Digit Some functions have more measuring sub ranges. These sub ranges are switched over automatically.
Active power, apparent power
Measuring range
5 ÷ 3680 W, VA
Resolution
1 W, VA
Reference error
(3 % of R + 2 D)
Operating error
(5 % of R + 5 D)
Cos φ
Displayed range
0.00 ÷ 1.00
Measuring range
0.40 ÷ 1.00 (2)
Resolution
0.01
Reference error
(1 % of R + 2 D)
Operating error
(2 % of R + 4 D)
Current consumption
Measuring range
0.00 ÷ 16.00 A
Resolution
0.01 A
Reference error
(2 % of R + 5 D)
Operating error
(3 % of R + 5 D)
Current by means of current clamp
Entire measuring range
0.00 ÷ 100.0 A
Measuring sub range
0.00 ÷ 4.99 A
Resolution
0.01 A
Reference error
(3 % of R + 4 D) (1)
Operating error
(4 % of R + 10 D) (1)
Measuring sub range
5.0 ÷ 100.0 A
Resolution
0.1 A
Reference error
(3 % of R + 4 D) (1)
Operating error
(4 % of R + 4 D) (1)
Cos φ by means of current clamp
Displayed range
0.00 ÷ 1.00
Measuring range
0.40 ÷ 1.00 (2)
Resolution
0.01
Reference error
(2 % of R + 2 D) (1)
Operating error
(5 % of R + 5 D) (1)
Active power, apparent power by means of current clamp
Measuring range
10 ÷ 9999 W, VA
Resolution
1 W, VA
Reference error
(5 % of R + 5 D) (1)
Operating error
(8 % of R + 5 D) (1)
(1) Current clamp or three-phase adapter error has to
be considered additionally
(2) Measuring error is not specified out of this range
Copyright © 2006-2019, ILLKO, s.r.o. No part of this publication may be reproduced or utilized in any form or by any means without
permission in writing from ILLKO, s.r.o.