ILLKO MDtest, IL2550 User Manual

USER’S
MANUAL
MDtest
(100-140 V)
MDtest (100-140 V) - User’s Manual Contents
Contents
1.1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Performance MDtest . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 Applied standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4 Terms and definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Description of the instrument 8
2.1 General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.4 Additional functions . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.5 Display of information . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.6 Graphic OLED display . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.7 Selection of the items in Menu or List . . . . . . . . . . . . . . . . . . 12
2.8 Users interface symbols . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.9 Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.10 Included in the set . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.11 Optional accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3 Putting into operation 16
3.1 Connecting to mains outlet . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2 Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3 Help screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4 Measurements 18
4.1 Selection of the function . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2 Measurement mode control . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2.1 Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . 18
4.2.2 HOLD function . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.2.3 Measurement of apparent power consumption . . . . . . . . . 19
4.3 Earth bond - Rpe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.3.1 Test lead resistance compensation . . . . . . . . . . . . . . . . 21
4.3.2 Protective earth resistance . . . . . . . . . . . . . . . . . . . . 21
4.4 Insulation resistance - Riso . . . . . . . . . . . . . . . . . . . . . . . . 22
4.5 Enclosure leakage current according to EN 62353 . . . . . . . . . . . 25
4.5.1 Alternative method - IaltEq . . . . . . . . . . . . . . . . . . . 25
4.5.2 PE current during operation - IdirEq . . . . . . . . . . . . . . 26
4.5.3 Differential leakage current - IdifEq . . . . . . . . . . . . . . . 28
4.6 Leakage current from the Application parts according to EN 62353 . 30
4.6.1 Alternative method - IaltPat . . . . . . . . . . . . . . . . . . . 30
4.6.2 Direct method - IdirPat . . . . . . . . . . . . . . . . . . . . . 31
4.6.3 Direct method, DUT with an internal electrical power source
- IdirPatB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.7 Leakage currents according to IEC EN 60601-1 . . . . . . . . . . . . 34
4.7.1 Earth leakage current - I60601Earth . . . . . . . . . . . . . . 35
c
2018 ILLKO, s.r.o. www.illko.cz Page 2
MDtest (100-140 V) - User’s Manual Contents
4.7.2 Enclosure (chassis) leakage current - I60601Touch . . . . . . . 36
4.7.3 Lead to ground (application part) leakage current - I60601Pat 37
4.7.4 Lead isolation – Mains on Applied Parts leakage current -
I60601MAP . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.8 Other leakage currents . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.8.1 Alternative method - IaltEq . . . . . . . . . . . . . . . . . . . 39
4.8.2 PE current during operation – IdirEq / differential leakage
current - IdifEq . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.8.3 Enclosure leakage current - IdirTouch / IdifTouch . . . . . . . 41
4.9 Apparent power, mains voltage and current consumption - Power . . 43
4.9.1 Power, mains voltage and current consumption in test socket . 43
4.9.2 Measuring of power and current consumption by means of
current clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5 Database operations 45
5.1 Storing measurement data . . . . . . . . . . . . . . . . . . . . . . . . 45
5.2 The structure of the data memory . . . . . . . . . . . . . . . . . . . . 45
5.3 Entering characters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.4 Identification number of device under test (ID) . . . . . . . . . . . . . 46
5.5 Working with test results . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.5.1 Viewing a list of tested DUTs . . . . . . . . . . . . . . . . . . 47
5.5.2 Searching for stored DUTs . . . . . . . . . . . . . . . . . . . . 48
5.5.3 Creating a new DUT file . . . . . . . . . . . . . . . . . . . . . 48
5.5.4 Deleting a DUT from the main memory . . . . . . . . . . . . 48
5.5.5 Modification of the ID number . . . . . . . . . . . . . . . . . . 49
5.5.6 Viewing measured values and evaluation . . . . . . . . . . . . 49
5.5.7 Modification of measured values and evaluation . . . . . . . . 50
5.6 Test sequences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.6.1 Description of the test sequence . . . . . . . . . . . . . . . . . 50
5.6.2 Creating a test sequence . . . . . . . . . . . . . . . . . . . . . 50
5.6.3 Assigning a test sequence to appliance . . . . . . . . . . . . . 51
5.6.4 Using the test sequence . . . . . . . . . . . . . . . . . . . . . . 51
6 Instrument main menu 52
6.1 Main menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.2 Menu - DUT properties . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.2.1 Menu - Measured values . . . . . . . . . . . . . . . . . . . . . 52
6.2.2 Menu - Visual inspection result . . . . . . . . . . . . . . . . . 52
6.2.3 Menu - Functional test result . . . . . . . . . . . . . . . . . . 52
6.2.4 Menu - Measurement result . . . . . . . . . . . . . . . . . . . 53
6.3 Menu - Test sequences . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.4 Menu - Empty file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.5 Menu - Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.5.1 Menu - System info . . . . . . . . . . . . . . . . . . . . . . . . 53
6.5.2 Menu - Measure DC values . . . . . . . . . . . . . . . . . . . . 53
6.5.3 Menu - Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.5.4 Menu - Date of next calibration . . . . . . . . . . . . . . . . . 54
c
2018 ILLKO, s.r.o. www.illko.cz Page 3
MDtest (100-140 V) - User’s Manual Contents
6.5.5 Menu - Erase database . . . . . . . . . . . . . . . . . . . . . . 54
6.5.6 Menu - Measurement locking . . . . . . . . . . . . . . . . . . . 54
6.5.7 Menu - Save last method . . . . . . . . . . . . . . . . . . . . . 55
6.5.8 Menu - Date and Time . . . . . . . . . . . . . . . . . . . . . . 55
6.5.9 Menu - Keys beep . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.5.10 Menu - Volume . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.5.11 Menu - Language . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.5.12 Menu - Nominal voltage . . . . . . . . . . . . . . . . . . . . . 55
7 Additional informations 56
7.1 Communication – data transfer . . . . . . . . . . . . . . . . . . . . . 56
7.2 Messages on the display . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.2.1 Warning messages . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.2.2 Operational information . . . . . . . . . . . . . . . . . . . . . 62
7.2.3 Setting information . . . . . . . . . . . . . . . . . . . . . . . . 64
7.2.4 Database information . . . . . . . . . . . . . . . . . . . . . . . 65
7.3 Reset of the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . 65
7.4 Utilization of other external adapters . . . . . . . . . . . . . . . . . . 66
7.4.1 Adapter for testing extension power supply cords (P 8030) . . 66
7.4.2 Three-phase adapter (P 8015-8018) . . . . . . . . . . . . . . . 66
7.5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7.6 Recurrent tests and testing after repair . . . . . . . . . . . . . . . . . 67
7.7 Calibration and service . . . . . . . . . . . . . . . . . . . . . . . . . . 68
7.8 Ecology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
8.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.2 Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.2.1 Earth bond . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.2.2 Insulation resistance . . . . . . . . . . . . . . . . . . . . . . . 70
8.2.3 Alternative leakage current . . . . . . . . . . . . . . . . . . . . 70
8.2.4 Direct leakage current (TRMS AC+DC value) . . . . . . . . . 71
8.2.5 Direct leakage current (DC component) . . . . . . . . . . . . . 71
8.2.6 Differential leakage current . . . . . . . . . . . . . . . . . . . . 71
8.2.7 Leakage current by means of three-phase adapter . . . . . . . 72
8.2.8 Apparent power . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.2.9 Mains voltage UL-N . . . . . . . . . . . . . . . . . . . . . . . . 72
8.2.10 Current consumption . . . . . . . . . . . . . . . . . . . . . . . 72
8.2.11 Current consumption by means of current clamp . . . . . . . . 72
8.2.12 Apparent power by means of current clamp . . . . . . . . . . 72
c
2018 ILLKO, s.r.o. www.illko.cz Page 4
MDtest (100-140 V) - User’s Manual 1 Introduction
1 Introduction
1.1 Safety
WARNING Read this User’s Manual carefully and completely and follow all instructions contained therein. If not, the use of the in­strument may be dangerous for the operator, devices under test or for the instrument!
If there is reason to believe that safe operation has become impossible, put the instrument out of operation and secure it against any unin­tended operation. Safe operation must be presumed to be no longer possible, if:
The instrument does not operate properly any longer.
The instrument, cables, connectors, plugs or accessories exhibits
visible damages.
The instrument was stored in unfavourable conditions for a long period.
The instrument was exposed to extraordinary stress caused by transport.
WARNING
Make sure that the instrument, measuring cables and all other accessories are in flawless condition, e.g. no damaged insulation, no broken cables or plugs etc.
Make sure that the AC outlet is easily accessible. If some trou­ble or malfunction occurs, immediately disconnect the MDtest adaptor from the outlet.
If the equipment is used in a manner not specified by the manu­facturer, the device’s protection may be compromised.
The instrument may only be powered from a 100÷140 V / 50 or 60 Hz grounded mains outlet, which is protected with a fuse or circuit breaker with a maximum rating of 16 A.
Voltage 127 V / 50/60 Hz will be used in this manual hereinafter.
In order to check the proper PE connection to a mains outlet,
MDtest generates a continuous current of ca. 1.5 mA / 50 or 60 Hz to the PE conductor. This test current can activate the insulation monitoring device in IT systems.
Only a trained, skilled person, who is familiar with hazardous voltage operations, can handle the MDtest. Unexpected haz­ardous voltages can occur on devices under test (dangerously charged capacitors, etc.).
c
2018 ILLKO, s.r.o. www.illko.cz Page 5
MDtest (100-140 V) - User’s Manual 1 Introduction
WARNING
It is necessary to respect all safety regulations applicable to the particular measurement.
Test Socket 1 of the MDtest must not be used for continuously powering the appliance. It is intended for test purposes only; max. test duration is 25 min @ 12 A and 60 s @ 12÷16 A.
The functional test may only be performed after the DUT has successfully passed the safety test.
Dangerous voltage or mains voltage may be present on Test Socket 1 during some tests.
Devices under test which are connected to Test Socket 1 of the MDtest can be powered by mains voltage during some tests. The appliance can thus be activated.
The PE terminal is connected to the PE terminal of Test Socket
1. When using the MDtest, do not connect any voltage to the
terminal PE. If connected, there is a danger of electric shock or damage to the MDtest.
Use only standard or optional accessories supplied with the MDtest by your distributor.
The figures in this manual are only illustrative and may vary slightly from the actual state or the version on your instrument.
1.2 Performance MDtest
The MDtest is a professional, compact, portable analyzer, designed to verify the electrical safety of medical devices. The MDtest has an extraordinary wide scope of use thanks to a lot of optional accessories, which enable testing practically any portable or hard-wired, one-phase or three-phase medical device.
1.3 Applied standards
EMC . . .. .. . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . .EN 55011:2009+A1:2010
EN 61326-1:2013
EN 61000-4-2:2009
EN 61000-4-3:2006+A1:2007+A2:2010
EN 61000-4-4:2004+A1:2010
EN 61000-4-5:2006 EN 61000-4-6:2009
EN 61000-4-11:2004
LVD . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . .. . . . . . EN 61010:2010
EN 61010-2-030:2010
EN 61557-16:2014
EN 62353:2014
c
2018 ILLKO, s.r.o. www.illko.cz Page 6
MDtest (100-140 V) - User’s Manual 1 Introduction
The MDtest is designed to perform electrical safety testing based on:
IEC 62353:2014
IEC 60601-1 (only some tests)
1.4 Terms and definitions
Accessible conductive part
Applied part
The part of the ME equipment that (in normal use) necessarily comes into physical contact with the patient for ME equipment or an ME system to perform its function.
Mains part
The part of electrical equipment forming a circuit that is intended to be connected to the supply mains.
Another terms and definitions – see IEC 62353:2014
ID number (ID)
Unique ID number for the DUT.
Hazardous voltage.
Warning concerning a point of danger! Read User’s Manual and observe all precautions!
Hint, notice.
c
2018 ILLKO, s.r.o. www.illko.cz Page 7
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
2 Description of the instrument
2.1 General description
The MDtest instrument is designed to verify the safety of electrical appliances and medical devices in general. All AC measurements are measured by the method TRMS (true rms), and for some measurements the DC component value can be displayed. Measured circuits meet the requirements of EN 60601 and EN 62353 (requirements for measuring circuits), including the required input characteristics.
The MDtest performs the following tests:
Earth bond test >1 A AC
Insulation resistance test 250 V DC
Alternative (substitute) equipment and applied part leakage current
Direct equipment and applied part leakage current
Differential leakage current
Ground (Earth) leakage current
Chassis (Enclosure) leakage current
Lead to Ground (Patient) leakage current
Lead isolation (Mains on applied parts leakage current)
PE current during operation and differential leakage current
Apparent power, mains voltage and current consumption
With additional equipment can be measured:
measurement of PE current during operation and differential leakage current
of three-phase portable appliance by means of optional adapters
measurement of apparent power and current consumption by means of current
clamp
check for dangerous contact voltage on the PE of power outlet (before the
start of the measurement)
check if the PE of the power outlet is grounded (before the start of the mea-
surement)
automatic check for leakage current amplitude of the device under test
check of the appliance’s current with automatic switch off the Test Socket 1:
0 ÷ 12 A . . . . . . . . . . . . . . . . . . .. max. 25 min
12 ÷ 16 A . .. . . . . . . . . . . . . . . . . max. 60 sec
manual self-test
nominal voltage 127 V or 120 V can be selected
c
2018 ILLKO, s.r.o. www.illko.cz Page 8
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
Features and highlights:
test results can be saved under an ID number which is unequivocally assigned to the respective DUT (barcode, RFID chip)
the test structures created at the PC software can be loaded into the MDtest
measurement data saved to the test instrument can be imported to PC software
via the USB port and reports can be generated
user-defined test sequences created at the PC software
user-defined report templates created at the PC software
2.2 Terminals
1. Test Socket for connection of tested appliance’s power supply cord
2. PE – terminal for connecting the exposed conductive part of the tested appli­ance (the terminal is connected to the PE terminal of Test Socket 1)
3. ITOUCH 2k– touch leakage current test terminal
4. RISO / IALT – insulation resistance / alternative leakage current test terminal
5. PAT – terminal for connecting the patient parts of tested appliance
6. RPE – earth bond test terminal
7. PEMAINS – terminal connected with PE of mains
8. USB - socket type A for connection USB 2.0 Full Speed HID (Human Interface Device), e.g. barcode scanner
9. USB – socket type B, PC connection
10. EXT input - current clamp / three-phase adapter / other adapters connector
11. Power supply cord
WARNING
Do not connect any external voltage to any terminal or connector. The only exception is the power supply cord 11 which is intended for powering the MDtest from the 127 V / 50 or 60 Hz mains outlet.
Max. load on Test Socket 1 is 16 A (limited in time).
Use original accessories only.
Do not connect any external voltage between RPE terminal and
the PE in Test Socket 1 or PE terminal.
c
2018 ILLKO, s.r.o. www.illko.cz Page 9
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
Figure 1: Front panel and rear panel
c
2018 ILLKO, s.r.o. www.illko.cz Page 10
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
2.3 Keyboard
12. [1 IALT] – Alternative leakage current of the device
13. [2 IALT] – Alternative leakage current of the patient parts
14. [3 RPE] – Earth bond test
15. [4 IDIR] – Direction leakage current of the accessible part
16. [5 IDIR] – Direction leakage current of the patient part
17. [6 RISO] – Insulation resistance, test voltage 250 V DC
18. [7 IDIF] – Differential leakage current of the accessible part
19. [8 IDIR Battery] – Leakage current of the patient part, battery powered device
20. [9 Power] – Apparent power consumption, current, mains voltage
21. [* UTIL] – Leakage current according to EN 60601-1:
- Leakage current of instrument part
- Touch leakage current of instrument part
- Leakage current of the patient part
- Leakage current of the patient part using the MAP method (mains on applied parts)
22. [TEST] – Starts measurement
2.4 Additional functions
23. [0 HELP] – display the circuit diagram for selected measurement
24. [# ARET] – lock the TEST key
25. [N] – scroll up
26. [H] – scroll down, hold displayed data
27. [save] – save measured values
28. [menu] – main menu
29. [ent] – ENTER, find DUT according to ID, other functions depending on context
30. [esc] – ESCAPE, other functions depending on context
31. Context menu buttons - actual function is displayed above the button
2.5 Display of information
32. Graphic OLED display
33. LED indicator for measuring status
c
2018 ILLKO, s.r.o. www.illko.cz Page 11
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
2.6 Graphic OLED display
a) Measured function b) Real time clock / Measuring is in progress c) Selected method and ID of DUT d) Measured value e) Actual function of context button (31)
f) Warning, notice, additional information
Figure 2: Graphic OLED display
2.7 Selection of the items in Menu or List
The item is selected using buttons:
[N] and [H]
Faster scroll in the lists:
[Page-] and [Page+]
c
2018 ILLKO, s.r.o. www.illko.cz Page 12
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
2.8 Users interface symbols
ARET
lock the TEST key
AUT.ARET
automatic lock the TEST key
HOLD
displayed value is held
L-N x
L-N reversal in Test Socket 1
CP PAT
plug the test lead into the PAT terminal
CP 2k
plug the test lead into the ITOUCH 2kterminal
x PATIENT unplug / do not plug patient parts
SEQ
test sequence is active
CAL
test lead resistance compensation is finished
CAL
the value of compensation is higher than measured value
ESC press button [esc] TEST press button [TEST]
the selection is active, self-test passed
the selection is not active, self-test failed
measuring is in progress
warning – Test Socket 1 energizing
plug external accessories
inner temperature is increasing when measuring PE resistance
notice - DUT must be isolated from ground
test voltage 250 V DC
menu Properties of DUT
menu Test sequences
menu Empty file of DUT
menu Setup
c
2018 ILLKO, s.r.o. www.illko.cz Page 13
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
2.9 Abbreviations
L phase conductor N neutral conductor PE protect earth conductor ALT alternative (substitute) method DIF differential method DIR direct method 60601Earth protect conductor current – direct method (EN 60601-1) 60601Touch touch current – direct method (EN 60601-1) 60601Pat patient leakage current – direct method (EN 60601-1) 60601MAP patient leakage current Mains on applied part – direct method
(EN 60601-1)
Eq equipment R L-N reversal Ext external Touch touch Pat patient B battery M mains part CP conductive part AP applied part MD measuring device
FE functional earth Mains mains voltage supply DUT device under test DUT PE device under test protective earth Open PE open protective earth Open N open neutral
socket
connection: Test Socket 1
RPE
connection: terminal RPE (Note: also applies to the other terminals)
c
2018 ILLKO, s.r.o. www.illko.cz Page 14
MDtest (100-140 V) - User’s Manual 2 Description of the instrument
2.10 Included in the set
The set includes all accessories necessary for testing 127 V / 50 or 60 Hz appliances equipped with a movable power supply cord. The set includes:
The measuring instrument MDtest
P 2011 - Test lead, black, 2 m
P 3011 - Test tip black
USB cable A-B
CD (User’s Manual + SW for PC)
User’s Manual
Calibration certificate
Warranty card
2.11 Optional accessories
Carrying bag:
P 6140 - Bag for instrument and some accessories
Measuring cables and adapters:
P 2012 - Test lead, blue, 2 m
P 2021 - Extension lead, black, 5 m
P 2030 - Test lead - connects up to 10 patient parts to MDtest
P 3012 - Test tip, blue
P 4011 - Crocodile clip, black
P 4012 - Crocodile clip, blue
P 8030 - Adapter for testing extension power supply cords
Special accessories:
P 8010 - Current clamp transformer
P 8015 - Three-phase adapter, 16 A socket, 5 contacts
P 8016 - Three-phase adapter, 16 A socket, 4 contacts
P 8017 - Three-phase adapter, 32 A socket, 5 contacts
P 8018 - Three-phase adapter, 32 A socket, 4 contacts
P 9022 - Barcode scanner (CCD) USB 2.0 Full Speed
P 9060 - Self-adhesive barcode labels
P 9080 - Stick-on label with appliance next test term marking
P 9092 - Barcode and label printer Brother PT-E300VP
P 9100 - Label tape TZ-241 for printer Brother PT-E300VP, white, 18 mm
width, 8 m length
P 9101 - Label tape TZ-231 for printer Brother PT-E300VP, white, 12 mm width, 8 m length
P 9102 - Label tape TZ-221 for printer Brother PT-E300VP, white, 9 mm width, 8 m length
c
2018 ILLKO, s.r.o. www.illko.cz Page 15
MDtest (100-140 V) - User’s Manual 3 Putting into operation
3 Putting into operation
3.1 Connecting to mains outlet
If the mains outlet, mains cable, instrument’s case or acces­sories are damaged, do not connect the MDtest to the mains outlet! If Idir and Idif is being measured, use the grounded mains outlet only to supply the instrument! Make sure the correct nominal voltage (120 V or 127 V) is selected (see chapter 6.5.12)
The MDtest can only be powered from 100÷140 V / 50 or 60 Hz, which is protected with a fuse or circuit breaker with a maximum rating of 16 A. The instrument is automatically switched on after plugging into the mains outlet. Disconnect any DUT while connecting/disconnecting the MDtest to/from the mains outlet!
The MDtest automatically tests the protective PE (ground) pin of the mains outlet. If the PE connection is inadequate, then the message (PE error) is displayed. In order to check the proper PE connection to a mains outlet, the MDtest generates a continuous current of ca. 1.5 mA / 50 or 60 Hz to the PE conductor. This test current can activate the insulation moni­toring device in IT systems. If the MDtest is connected to an IT earthing system, it is possible to measure – e.g. Rpe, Riso, Ialt and IdirPatB. If dangerous contact voltage arises on the PE, then the (PE error) indicator lights up when the operator presses the key [TEST]. If op­erator presses the [TEST] key after this, the measurement will not start.
WARNING If the test result for the ground pin of the mains outlet is unsatisfactory (in TT or TN systems), the outlet must be repaired before you plug in the MDtest to the outlet again!
c
2018 ILLKO, s.r.o. www.illko.cz Page 16
MDtest (100-140 V) - User’s Manual 3 Putting into operation
3.2 Self-test
It can sometimes be difficult to identify if the measuring instrument does not work properly, especially when measuring current leakage. The MDtest therefore has built in the possibility to promptly test the basic functionality. The functionality of the following functions can be tested:
Insulation resistance
Measurement of alternative (substitute) leakage current
Measurement of PE current during operation
Measurement of differential leakage current
WARNING
Disconnect DUT before the MDtest self-test!
This test does not substitute for calibration!
The MDtest has to be connected to TT or TN mains.
Disconnect the DUT.
Press the [menu] key and release it.
Use the [H] key to select (Setup), press the [ent] key for confirmation.
Use the [H] key to select (Self-test), press the [ent] key for confirmation.
Press the [TEST] button. The test will then start; after it is finished, the
results are displayed: (self-test passed) or (self-test failed).
The correct values are:
RisoM-PE: 1.000 ± 0.050 MIaltEq: 0.100 ± 0.010 mAIdirEq: 0.100 ± 0.010 mAIdifEq: 0.100 ± 0.010 mA
Contact service if the test results differ from the ones described above.
3.3 Help screens
The help screens are available in all functions. The Help screens contain schematic diagrams for illustrating how to properly connect the instrument for electric instal­lation, applied standards and the number of a picture according to the standard. After selecting the measurement you want to perform, press the [0 HELP] (once or repetitively) key in order to see the associated Help screens. Press the [0 HELP] or [esc] to leave the Help.
c
2018 ILLKO, s.r.o. www.illko.cz Page 17
MDtest (100-140 V) - User’s Manual 4 Measurements
4 Measurements
4.1 Selection of the function
The measured function can be selected automatically in sequence mode (see chap­ter 5.6) or manually (see Figure 3).
Figure 3: Keyboard
4.2 Measurement mode control
The beginning, process and ending of measurements are controlled by the [TEST] key. If the [esc] key was pressed during the measurement, then it will be cancelled (the measured value is not saved).
4.2.1 Measurement mode
The measurement can start in four different measurement modes:
One measuring cycle – after the [TEST] key is pressed and released, one measuring cycle starts and is completed after a few seconds. The result of the measurement is displayed.
Continuous measurement – press the [TEST] key; measurement is in progress and current measured values are displayed until the [TEST] key is released. The result of the measurement is displayed.
Locking the measurement - press the [# ARET] key and then press the [TEST] key. Then release both keys. The measurement is in progress and the
current measured values are displayed until the [TEST] key is again pressed and released. Reverse steps are possible too: press the [TEST] key and then press the the [# ARET] key.
c
2018 ILLKO, s.r.o. www.illko.cz Page 18
MDtest (100-140 V) - User’s Manual 4 Measurements
Automatically locking the measurement – this function can be set in the menu Setup. Measuring is similar to Locking the measurements but the [# ARET] key is not used.
Measuring in process is indicated by the symbol .
4.2.2 HOLD function
4.2.3 Measurement of apparent power consumption
During some measurements, the apparent consumption is displayed in the upper display.
c
2018 ILLKO, s.r.o. www.illko.cz Page 19
MDtest (100-140 V) - User’s Manual 4 Measurements
4.3 Earth bond - Rpe
Figure 4: Rpe measurement
WARNING
Do not connect any external voltage between the RPE terminal and the PE in Test Socket 1 or PE terminal.
If you are going to test equipment which can be charged to volt­age >25 V (e.g. filter capacitors), discharge it before the test!
If external voltage is present on the device under test, the message (Interference voltage at the input) is displayed. This status is a dangerous fault condition; disconnect the DUT from the MDtest im­mediately to avoid damage! Resistance measurement is also possible if the MDtest is connected to the IT mains. We recommend performing the measurements for a max. of 30 secs. because of power dissipation inside the instrument (only for value <10.00 ). If the temperature inside exceeds the allowed limit, the
symbol
is displayed and the acoustic signal sounds.
c
2018 ILLKO, s.r.o. www.illko.cz Page 20
MDtest (100-140 V) - User’s Manual 4 Measurements
4.3.1 Test lead resistance compensation
Compensation means that the test lead resistance is measured and then automatically memorized and subtracted from all Rpe measurements. The real earth bond resistance is thus displayed. The compensation constant remains in the instrument’s memory even if the instrument is switched off. Therefore, the lead compensation is valid until a new compensation is carried out.
Procedure:
Press the [3 RPE] key and release it.
Plug the test lead with the test tip into the RPE terminal.
Plug the test tip of the test lead either into the PE terminal or press it
against the ground pin of the Test Socket 1. If you use two test leads (which are connected to the PE terminal and the RPE terminal), then make a short­circuit of their test tips.
Press the [TEST] key and release it.
Press the [Cal] key and release it. The measured value is saved.
The compensation is effective for test lead resistance <2.00 . If the test lead resistance is >10.00 and the compensation procedure has been carried out, com­pensation is cancelled and all the following displayed results are the sum of the earth bond + test lead resistance. The CAL symbol is not displayed. If compensation is active (it means that test lead resistance <2.00 was memo­rized), then the CAL symbol is displayed during Rpe measurements. If the measured value is less than the saved compensation value during measure­ment (after subtraction, the memorized resistance will result in a negative number), the display shows the absolute value and the CAL symbol appears inversely. The cause may be replacing test leads for shorter ones with less resistance. In this case, it is necessary to re-compensate the test leads resistance.
4.3.2 Protective earth resistance
Procedure:
Press the [3 RPE] key and release it.
Plug the test lead with the test tip into the RPE terminal.
Connect the tested appliance:
– One-phase appliance with a movable power supply cord – plug in its
mains plug to Test Socket 1.
– Three-phase appliance with a movable power supply cord – connect the
PE pin of its plug to the instrument’s PE terminal by means of the optional test lead + crocodile clip.
Hard-wired appliance – firstly disconnect the appliance from the
mains! Connect its point of PE connection to the instrument’s PE ter-
minal by means of the optional test lead + crocodile clip.
c
2018 ILLKO, s.r.o. www.illko.cz Page 21
MDtest (100-140 V) - User’s Manual 4 Measurements
Press the test tip against the exposed conductive part of the tested appliance.
Press the [TEST] key. If the CAL indicator lights up during measurement,
it means that the test lead resistance is compensated.
After finishing the measurement, the last measured value is displayed.
The lowest measured value can be displayed - press the [3 RPE] key during the measuring. The current and the lowest value are displayed now. When the measurement is finished, only the lowest value is dis­played. To deactivate the function, press the [3 RPE] key again.
WARNING
If only >10.00 is displayed it means that the measured earth bond resistance is higher than the maximal value which can be measured by MDtest. The maximal value is the sum resistance of the test
leads and the resistance of the earth bond!
4.4 Insulation resistance - Riso
The three insulation resistance tests take measurements on (250 V):
Mains – Protective earth [M-PE]
Applied part - Protective earth [AP-PE]
Mains – Applied part [M-AP]
Figure 5: RisoM-PE measurement
c
2018 ILLKO, s.r.o. www.illko.cz Page 22
MDtest (100-140 V) - User’s Manual 4 Measurements
Figure 6: RisoAP-PE measurement
Figure 7: RisoM-AP measurement
c
2018 ILLKO, s.r.o. www.illko.cz Page 23
MDtest (100-140 V) - User’s Manual 4 Measurements
Do not touch the exposed conductive parts of the tested appliance while measurement is in progress – RISK OF ELECTRIC
SHOCK!
Do not disconnect the test leads while measurement is in progress or immediately after finishing the measurement. The capacitive component of the tested appliance can be charged to a voltage of up to 300 V; after the [TEST] key is released, this voltage is automatically discharged.
Make sure the tested appliance is de-energized – its mains voltage must be disconnected before the start of the measurement!
Before measuring resistance, the MDtest device checks if the DUT is de-energized. If so, then an acoustic signal sounds and the display shows a warning message (Interference voltage at the input).
Resistance measurement is also possible if the MDtest is con­nected to the IT mains.
Procedure:
Press the [6 RISO] key and release it.
Press one of the function keys to select the type of measuring you require.
Connect the appliance for testing:
– For a one-phase appliance with a movable power supply cord – plug in
its mains plug to Test Socket 1.
– For a three-phase appliance with movable power supply cord – connect
the PE pin of its plug to the instrument’s PE terminal by means of the optional test lead + crocodile clip. Then connect the remaining current-carrying conductors by means of the test lead + test tip to the instrument’s RISO / IALT terminal.
For a hard-wired appliance – first disconnect the appliance from
the mains! Connect the point of the PE connection to the instrument’s PE terminal by means of the optional test lead + crocodile clip. Then
connect the remaining current-carrying conductors by means of the test lead + test tip to the instrument’s RISO / IALT terminal.
If an applied part is measured, plug it into the PAT terminal.
Switch on the tested appliance’s mains switch.
Press the [TEST] key. It is recommended that you hold the [TEST] key
down for 5-10 seconds (or even longer if the displayed result is not stabilized); the influence of the tested appliance’s internal capacitance can be eliminated by this method.
After finishing the measurement the last measured value is displayed.
c
2018 ILLKO, s.r.o. www.illko.cz Page 24
MDtest (100-140 V) - User’s Manual 4 Measurements
4.5 Enclosure leakage current according to EN 62353
4.5.1 Alternative method - IaltEq
Figure 8: IaltEq measurement
Outlet conditions:
- Normal
- Open PE
Do not touch exposed conductive parts of the tested appliance while measurement is in progress – RISK OF ELECTRIC SHOCK!
The test voltage is max. 127 V / 50 or 60 Hz, and the short circuit current is approximately 1.5 mA. Leakage current measurement is also possible if the MDtest is con­nected to the IT mains.
Procedure:
Press the [1 IALT] key and release it.
If the measurement with open PE can be performed, press the [OpenPE] key.
For a one-phase appliance with movable power supply cord, plug in its mains
plug to Test Socket 1 and plug the applied part into the PAT terminal. If necessary (e.g. protective class II appliances), plug the test lead into the PE terminal (if Open PE is selected plug into the PAT terminal).
Switch on tested appliance’s mains switch.
Press the [TEST] key.
Press the test tip against the exposed conductive part of the tested appliance.
The last measured value is displayed after finishing the measurement.
Switch off the tested appliance’s mains switch.
c
2018 ILLKO, s.r.o. www.illko.cz Page 25
MDtest (100-140 V) - User’s Manual 4 Measurements
4.5.2 PE current during operation - IdirEq
Figure 9: IdirEq measurement
Outlet conditions:
- Normal Polarity - Reversed Polarity
- Normal Polarity, Open PE - Reversed Polarity, Open PE
If the test voltage is present in Test Socket 1, then the tested equip­ment starts to run after its mains switch is switched on! Take into consideration all safety warnings referred to by the DUT! Do not touch the exposed conductive parts of the tested object after the mains switch is switched on!
The MDtest is equipped with an electronic check for excessive leakage current during measurement. If the leakage current exceeds the limit value of approximately 10 mA, then the mains voltage is disconnected from the Test Socket 1. This status is indicated by displaying (Cur- rent >10.00 mA) and the acoustic signal sounds. The value of the DC component can be measured and displayed (see chapter 6.5.2).
c
2018 ILLKO, s.r.o. www.illko.cz Page 26
MDtest (100-140 V) - User’s Manual 4 Measurements
WARNING
The earth bond resistance (chapter 4.3) must be first tested; it is recommended that you also test the insulation resistance if it is possible too (chapter 4.4). Then you can carry out the leakage current measurement!
The DUT must be isolated from the ground (from the earth potential) during measurement!
Test Socket 1 of the MDtest must not be used for continuous power supply of the appliance!
Procedure:
Press the [4 IDIR] key and release it.
If the measurement with open PE can be performed, press the [OpenPE] key.
Press the [Eq] or [OpenPE] keys repetitively in order to measure with L-N
reversal. Symbol L-N reversed is displayed
Connect the tested appliance: plug in its mains plug into Test Socket 1 and plug the applied part into the PAT terminal. If necessary (e.g. protective class II appliances), plug the test lead into the PE terminal (if Open PE is selected plug into the PAT terminal).
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
Press the test tip against the exposed conductive part of the tested appliance.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 27
MDtest (100-140 V) - User’s Manual 4 Measurements
4.5.3 Differential leakage current - IdifEq
Figure 10: IdifEq measurement
Outlet conditions:
- Normal Polarity - Reversed Polarity
- Normal Polarity, Open PE - Reversed Polarity, Open PE
If the test voltage is present in Test Socket 1, then the tested equip­ment starts to run after its mains switch is switched on! Take into consideration all safety warnings referred to by the DUT! Do not touch the exposed conductive parts of the tested object after the mains switch is switched on!
The MDtest is equipped with an electronic check for excessive leak­age current during measurement. If the leakage current exceeds the limit value of approximately 10 mA, then the mains voltage is discon­nected from the Test Socket 1. This status is indicated by displaying
(Current >10.00 mA) and the acoustic signal sounds.
WARNING
The earth bond resistance (chapter 4.3) must be first tested; it is also recommended that you also test the insulation resistance if it is possible (chapter 4.4). Then you can carry out the leakage current measurement!
Test Socket 1 of the MDtest must not be used for continuous power supply of the appliance!
c
2018 ILLKO, s.r.o. www.illko.cz Page 28
MDtest (100-140 V) - User’s Manual 4 Measurements
Procedure:
Press the [7 IDIF] key and release it.
If the measurement with open PE can be performed, press the [OpenPE] key.
Press the [Eq] or [OpenPE] keys repetitively in order to measure with L-N
reversal. Symbol L-N reversed is displayed.
Connect the tested appliance: plug in its mains plug into Test Socket 1 and plug the applied part into the PAT terminal. If necessary (e.g. protective class II appliances), plug the test lead into the PE terminal (if Open PE is selected plug into the PAT terminal).
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
Press the test tip against the exposed conductive part of the tested appliance.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 29
MDtest (100-140 V) - User’s Manual 4 Measurements
4.6 Leakage current from the Application parts according to EN 62353
4.6.1 Alternative method - IaltPat
Figure 11: IaltPat measurement
Do not touch the exposed conductive parts of the tested appliance while measurement is in progress – RISK OF ELECTRIC SHOCK!
The test voltage is max. 127 V / 50 or 60 Hz and the short circuit current is approximately 1.5 mA. Leakage current measurement is also possible if the MDtest is con­nected to the IT mains.
Procedure:
Press the [2 IALT] key and release it.
Connect the tested appliance: plug in its mains plug into the Test Socket 1
and plug the applied part into the PAT terminal. If necessary (e.g. protective class II appliances), plug the test lead into the PE terminal.
Switch on tested appliance’s mains switch.
Press the [TEST] key.
Press the test tip against the exposed conductive part of the tested appliance.
The last measured value is displayed after finishing off the measurement.
Switch off the tested appliance’s mains switch.
If the measured value is <0.100 mA, then the resolution can be in­creased. For higher resolution press the [4/5 D] key. This setting is not saved.
c
2018 ILLKO, s.r.o. www.illko.cz Page 30
MDtest (100-140 V) - User’s Manual 4 Measurements
4.6.2 Direct method - IdirPat
Figure 12: IdirPat measurement
If the test voltage is present in Test Socket 1, then the tested equip­ment starts to run after its mains switch is switched on! Take into consideration all safety warnings referred to by the DUT! Do not touch the exposed conductive parts of the tested object after the mains switch is switched on!
The MDtest is equipped with an electronic check for excessive leak­age current during measurement. If the leakage current exceeds the limit value of approximately 10 mA, then the mains voltage is discon­nected from the Test Socket 1. This status is indicated by displaying
(Current >10.00 mA) and the acoustic signal sounds.
WARNING
The earth bond resistance (chapter 4.3) must be first tested; it is recommended that you also test the insulation resistance if it is possible too (chapter 4.4). Then you can carry out the leakage current measurement!
Test Socket 1 of the MDtest must not be used for continuous power supply of the appliance!
c
2018 ILLKO, s.r.o. www.illko.cz Page 31
MDtest (100-140 V) - User’s Manual 4 Measurements
Procedure:
Press the [5 IDIR] key and release it.
Press the [Rever] key repetitively in order to measure with L-N reversal. The
symbol L-N reversed is displayed.
Connect the tested appliance: plug in its mains plug into Test Socket 1 and plug the applied part into the PAT terminal. If necessary (e.g. protective class II appliances), plug the test lead into the PE terminal.
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
Press the test tip against the exposed conductive part of the tested appliance.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 32
MDtest (100-140 V) - User’s Manual 4 Measurements
4.6.3 Direct method, DUT with an internal electrical power source ­IdirPatB
Figure 13: IdirPatB measurement
Do not touch the exposed conductive parts of the tested appliance while measurement is in progress – RISK OF ELECTRIC SHOCK!
The test voltage is max. 127 V / 50 or 60 Hz and the short circuit current is approximately 1.5 mA. Leakage current measurement is also possible if the MDtest is con­nected to the IT mains.
Procedure:
Press the [8 IDIR BATTERY] key and release it.
Plug the applied part into the PAT terminal.
Plug the test lead into the PE terminal.
Switch on tested appliance’s mains switch.
Press the test tip against the exposed conductive part of the tested appliance.
Press the [TEST] key.
The last measured value is displayed after finishing the measurement.
Switch off the tested appliance’s mains switch.
c
2018 ILLKO, s.r.o. www.illko.cz Page 33
MDtest (100-140 V) - User’s Manual 4 Measurements
4.7 Leakage currents according to IEC EN 60601-1
WARNING
Leakage currents (according to EN 60601-1) are measured by the same measurement methods as leakage current, which is specified in chapter
4.5.2. Therefore, all safety warnings and operating instructions apply for these measurements as to those measured leakage currents listed in chapter 4.5.2.
Some measurements are possible to perform with the DC com­ponent measurement (see chapter 6.5.2).
To perform the SFC (Single Fault Condition) – the Neutral or PE conductor is interrupted. During the measurement a warning beep is emitted.
c
2018 ILLKO, s.r.o. www.illko.cz Page 34
MDtest (100-140 V) - User’s Manual 4 Measurements
4.7.1 Earth leakage current (direct method) - I60601Earth
Figure 14: I60601Earth measurement
Outlet conditions:
- Normal Polarity - Reversed Polarity
- Normal Polarity, Open Neutral - Reversed Polarity, Open Neutral
Procedure:
Press the [* UTIL] key and release it. Then press the [1] key and release it.
If the measurement with open N can be performed press the [OpenN] key.
Press the [Default] or [OpenN] keys repetitively in order to measure with
L-N reversal. The symbol L-N reversed is displayed.
Connect the tested appliance: plug in its mains plug into Test Socket 1.
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 35
MDtest (100-140 V) - User’s Manual 4 Measurements
4.7.2 Enclosure (chassis) leakage current (direct method) - I60601Touch
Figure 15: I60601Touch measurement
Outlet conditions:
- Normal Polarity - Reversed Polarity
- Normal Polarity, Open Neutral - Reversed Polarity, Open Neutral
- Normal Polarity, Open PE - Reversed Polarity, Open PE
Procedure:
Press the [* UTIL] key and release it. Then press the [2] key and release it.
If the measurement with open N can be performed, press the [OpenN] key.
If the measurement with open PE can be performed, press the [OpenPE] key.
Press the [Default], [OpenN] or [OpenPE] keys repetitively in order to
measure with L-N reversal. Symbol L-N reversed is displayed.
Connect the tested appliance: plug in its mains plug into Test Socket 1. Plug the test lead into the PAT terminal. Do not plug the applications parts!
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
Press the test tip against the exposed conductive part of the tested appliance.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 36
MDtest (100-140 V) - User’s Manual 4 Measurements
4.7.3 Lead to ground (application part) leakage current (direct method)
- I60601Pat
Figure 16: I60601Pat measurement
Outlet conditions:
- Normal Polarity - Reversed Polarity
- Normal Polarity, Open Neutral - Reversed Polarity, Open Neutral
- Normal Polarity, Open PE - Reversed Polarity, Open PE
Procedure:
Press the [* UTIL] key and release it. Then press the [3] key and release it.
If the measurement with open N can be performed, press the [OpenN] key.
If the measurement with open PE can be performed, press the [OpenPE] key.
Press the [Default], [OpenN] or [OpenPE] keys repetitively in order to
measure with L-N reversal. Symbol L-N reversed is displayed.
Connect the tested appliance: plug in its mains plug into Test Socket 1. Plug the applications parts into the PAT or PEMAINS terminals.
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 37
MDtest (100-140 V) - User’s Manual 4 Measurements
4.7.4 Lead isolation – Mains on Applied Parts leakage current (direct method) - I60601MAP
Figure 17: I60601MAP measurement
Outlet conditions:
- Normal Polarity, Normal Polarity Isolation transformer
- Normal Polarity, Reversed Polarity Isolation transformer
- Reversed Polarity, Normal Polarity Isolation transformer
- Reversed Polarity, Reversed Polarity Isolation transformer
Procedure:
Press the [* UTIL] key and release it. Then press the [4] key and release it.
Select the polarity of the isolated AC test voltage:
[MAP1] => normal, [MAP2] => reversed.
Connect the tested appliance: plug in its mains plug into Test Socket 1. Plug the applications parts into the PAT terminal. If necessary (e.g. protective class II appliances), plug the test lead into the PE or PEMAINS terminals.
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
Press the test tip against the exposed conductive part of the tested appliance.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 38
MDtest (100-140 V) - User’s Manual 4 Measurements
4.8 Other leakage currents
4.8.1 Alternative method - IaltEq
The measuring method is similar to Alternative method - IaltEq (chapter 4.5.1), but the applications parts are not connected. The test lead connect to the ITOUCH 2k terminal.
Figure 18: IaltEq measurement
4.8.2 PE current during operation – IdirEq / differential leakage current
- IdifEq
Figure 19: IdirEq measurement
c
2018 ILLKO, s.r.o. www.illko.cz Page 39
MDtest (100-140 V) - User’s Manual 4 Measurements
Figure 20: IdifEq measurement
WARNING
These leakage currents are measured by the same measurement meth­ods as leakage currents, which are specified in chapters 4.5.2 or 4.5.3. Therefore, all safety warnings and operating instructions apply for these measurements as to those measured leakage currents listed in chapters 4.5.2 or 4.5.3.
Procedure:
Press the [4 IDIR] key (direct method) or the [7 IDIF] key (differential method).
Press the [Eq] key repetitively in order to measure with L-N reversal. The symbol L-N reversed is displayed.
Connect the tested appliance: plug in its mains plug into Test Socket 1.
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
Three-phase and hard-wired appliance:
Optional accessories are intended for leakage current measurement of both three­phase and hard-wired appliances.
Procedure:
Press the [4 IDIR] or [7 IDIF] key repetitively in order to select the external
method. The symbol is displayed.
Connect the optional three-phase adapter to the terminal 10.
Switch on the tested appliance’s mains switch.
Press the [TEST] key.
The last measured value is displayed after finishing the measurement.
Switch off the tested appliance’s mains switch.
c
2018 ILLKO, s.r.o. www.illko.cz Page 40
MDtest (100-140 V) - User’s Manual 4 Measurements
4.8.3 Enclosure leakage current (direct method or differential method)
- IdirTouch / IdifTouch
Figure 21: IdirTouch measurement
Figure 22: IdifTouch measurement
WARNING
Enclosure leakage currents are measured by the same measurement methods as leakage currents which are specified in chapters 4.5.2 or
4.5.3. Therefore, all safety warnings and operating instructions apply for these measurements as to those measured leakage currents listed in chapters 4.5.2 or 4.5.3.
IdirTouch measurement: Only the component of the current that leakages from the ungrounded parts is measured (see Figure 21). IdifTouch measurement: Both components of the leakage current (PE lead and ungrounded parts) are measured (see Figure 22).
c
2018 ILLKO, s.r.o. www.illko.cz Page 41
MDtest (100-140 V) - User’s Manual 4 Measurements
Procedure:
Press the [4 IDIR] key (direct method) or the [7 IDIF] key (differential method) and release it. Then press the [Touch] key and release it.
Press the [Touch] key repetitively in order to measure with L-N reversal. The symbol L-N reversed is displayed.
Connect the tested appliance: plug in its mains plug into Test Socket 1.
Plug the test lead into the ITOUCH 2kterminal.
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
Press the test tip against the exposed conductive part of the tested appliance.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
c
2018 ILLKO, s.r.o. www.illko.cz Page 42
MDtest (100-140 V) - User’s Manual 4 Measurements
4.9 Apparent power, mains voltage and current consump­tion - Power
4.9.1 Power, mains voltage and current consumption in test socket
If test voltage is present in Test Socket 1 – it is indicated by the (TEST) indicator, then the tested equipment starts to run after its mains switch is switched on! Take into consideration all safety warnings referred to by the DUT! Do not touch the exposed conductive parts of the tested object after the mains switch is switched on – RISK OF ELECTRIC SHOCK!
WARNING
Test Socket 1 of the MDtest must not be used for continuous power supply of appliance!
Procedure:
Press the [9 Power] key and release it.
Connect the tested appliance: plug in its mains plug into Test Socket 1.
Press the [TEST] key.
Switch on the tested appliance’s mains switch.
The HOLD function can be used (press the [H] key).
Switch off the tested appliance’s mains switch.
Finish the measurement (see chapter 4.2.1).
If the measured value of power is higher than the highest measurable value the symbol >>>> appears.
c
2018 ILLKO, s.r.o. www.illko.cz Page 43
MDtest (100-140 V) - User’s Manual 4 Measurements
4.9.2 Measuring of power and current consumption by means of current clamp
WARNING
It is important to observe these principles for proper measurement: Both the MDtest and the tested appliance must be powered from iden­tical phase.
Procedure:
Press the [9 Power] key repeatedly in order to select the external method.
The symbol is displayed.
Connect the optional current clamp to the terminal 10.
Switch on the tested appliance’s mains switch.
Enclose with current clamp’s jaws live conductor L of the measured appliance.
Press the [TEST] key.
The last measured value is displayed after finishing the measurement.
Switch off the tested appliance’s mains switch.
If the measured value of power is higher than the highest measurable value the symbol >>>> appears.
c
2018 ILLKO, s.r.o. www.illko.cz Page 44
5 Database operations
5.1 Storing measurement data
A set of data can be stored for each appliance. Data can be created and transferred to the MDtest from a PC and other data is created during the measurement. Each set of data (file) has its unique identification number (ID) assigned.
The structure of the file:
Item Transfer from PC Description Created ID yes 1-20 ASCII
code charts
barcode scanner, ID editor
test sequence yes sequence
of test steps
transfer from PC,
assign of the test sequence measurement data
no measurement and store
of the value final evaluation
no initial evaluation (passed) is
possible to change manually date of measurement
no editable on
a PC only
saving the measured value
5.2 The structure of the data memory
Measured values and other possible data are stored as a file in a temporary memory where the ID is assigned and finally the file is transferred to the main memory. Conversely you can load a set of measurements of a particular appliance from the main memory to the temporary memory and then edit it.
Measuring up to 4 000 DUTs can be stored in the main memory.
To perform the whole procedure:
Read the device ID number (use the barcode scanner) or enter the ID from the keyboard.
Perform all required measurements.
Check the visual inspection, functional test and measured values (PASSED
is preset by default).
Press the [save] key, the data will be transferred from the temporary memory to main memory.
c
2018 ILLKO, s.r.o. www.illko.cz Page 45
The data stored in the temporary memory can be displayed after se­lecting the appropriate quantity. If a memory cell in the temporary memory is empty, the symbol − − − − is displayed.
5.3 Entering characters
In this mode the text, setting, etc. can be entered and edited. When the keys are pressed repetitively, the alphanumeric symbols are entered:
[0] 0 space [1] 1 + − . , _ [2] 2 A B C a b c [3] 3 D E F d e f [4] 4 G H I g h i [5] 5 J K L j k l [6] 6 M N O m n o [7] 7 P Q R S p q r s [8] 8 T U V t u v [9] 9 W X Y Z w x y z [*] : ; = ! ? @ $ % & [#] # ( ) < > { } [ ] / \
Context button:
[< <] move the cursor left [> >] move the cursor right [< Del] backspace
Entry mode can be abandoned:
[ent] confirm entry and returns back [esc] discard modifications and returns back
5.4 Identification number of device under test (ID)
The ID number is an individual number, which is assigned to only one DUT. An appliance ID of up to 20 alphanumeric characters can be entered. The appliance ID can be entered with:
keypad of MDtest
barcode scanner
external USB keyboard (necessary to confirm it by pressing the ENTER key)
c
2018 ILLKO, s.r.o. www.illko.cz Page 46
The ID number should be entered before the start of the measurement. If the ID number already exists in the main memory (e.g. was trans­ferred from a PC), then the data is loaded into temporary memory. If the user forgot to enter the ID number before the start of the mea­surement, it is possible to enter it before saving the data file into the main memory of the appliance. However, if the appliance with this ID exists already, the data will be rewritten completely!
Barcode scanner, external keyboard
Any device which has an USB 2.0 Full Speed HID (Human Interface Device) can be used with MDtest.
Plug the barcode scanner or keyboard into the USB connector 8.
A confirmation beep confirms the barcode has been received. If an external
keyboard is used, press the ENTER key to confirm the ID and the ID number will be displayed.
The ID number is read only when measurement is not in progress and/or no menu / warning is displayed.
The ID number can be inserted via barcode scanner or keyboard in the ID editor mode too.
5.5 Working with test results
5.5.1 Viewing a list of tested DUTs
Use the [N] or [H] keys to display a list of tested DUTs, which are stored in the main memory. The data stored for each appliance can be viewed or edited. To modify/edit stored data, it is necessary to transfer the data to temporary memory and after modification send the data back to the main memory. The list of appliances contains the following information:
ID number
measured values - if the device has been loaded onto the temporary memory
and then stored in the main memory the symbol (V) – value is displayed
test sequence – if a test sequence is made for the device, the symbol (S) – sequence is displayed
The symbol (V) next the ID of DUT is displayed even if no measure­ment has been performed. The symbol only informs the user about a data transfer from temporary memory to the main memory.
c
2018 ILLKO, s.r.o. www.illko.cz Page 47
5.5.2 Searching for stored DUTs
The device allows the user to search for a DUT according to their ID numbers:
Insert the ID number via the barcode scanner, external USB keyboard or (internal) keypad.
MDtest starts searching.
If the ID number is found, then it is confirmed by a long beep, the data file is
transferred to the temporary memory and the ID number is displayed.
If the ID number is not found, the message will be displayed and triple beep sounds.
For manual searching press the [ent] key and insert the ID number via the MDtest keypad and then press the [ent] key again.
DUT can be also selected directly from the list (database). Press the [N] or [H] keys to open the database. Then select the DUT via [N] or [H] keys and press the
[Load] key.
5.5.3 Creating a new DUT file
Creating a new DUT file is performed as follows:
Create a new ID number (see chapter 5.5.2).
Confirm that it is a new device.
Make a note of the device (type, serial number etc.) for archiving on a PC.
Perform all measurements (if required).
Evaluate the results of the visual inspection, functional test and measurement
(default value is "PASSED") – see chapter 6.2.
Press the [save] key to save the new record in the main memory.
The data file is created in temporary memory (see chapter 5.2) and then is transferred to the main memory. Therefore, it is flagged that a measurement was performed – the symbol (V) is displayed even if no values were measured.
5.5.4 Deleting a DUT from the main memory
A stored DUT can be deleted from the main memory one by one or by selecting DUTs. Deleting a DUT is performed as follows:
Press the [N] or [H] keys to open the database.
Use the [N] or [H] keys to highlight the ID number which should be deleted.
Press the [menu] key.
Press the [Delete] key and confirm [Yes].
c
2018 ILLKO, s.r.o. www.illko.cz Page 48
When removing more records at once follow in a similar way - use the [Mark] key.
WARNING
Deleting records will not free up more space in the main memory. The memory is released only when the complete memory is erased (manually in the menu or before loading new data from the PC).
5.5.5 Modification of the ID number
Modification of ID number is performed as follows:
Press the [N] or [H] keys to open the database.
Use the [N] or [H] keys to highlight the ID number which should be modified.
Press the [menu] key.
Press the [Rename] key and insert a new ID – see chapter 5.3.
Press to the [ent] key to confirm or press [esc] to leave.
During the modification, the ID is checking if the new ID contains up to 20 characters (a longer text is automatically reduced). Furthermore, it is checked whether an appliance with the same ID is already present in the database. If this situation occurs, the new appliances’ ID will not be saved and a message will be displayed (ID
is already in database).
5.5.6 Viewing measured values and evaluation
The record on the stored values of the appliance can be viewed as follows:
Press the [N] or [H] keys to open the database.
Use the [N] or [H] keys to highlight the ID number which should be viewed.
Press the [ent] key to open the file with the measured values and evaluation.
Use the [N] or [H] keys to display the results.
If the data file has not transferred from the temporary memory to the main memory (the symbol V is not displayed), you cannot view the measured values and their evaluation. Therefore, when you press the [ent] key, you will hear a warning beep.
c
2018 ILLKO, s.r.o. www.illko.cz Page 49
5.5.7 Modification of measured values and evaluation
Measured values cannot be edited, but it is possible to repeat the measurement. Measured values and the evaluation status of appliances stored in the database can be changed so that the data set, which is loaded into the temporary memory, performs a new measurement of the variables whose values you want to change. You can also change the evaluation status (see chapter 6.2). A changed data file is saved again to the main memory. The change is made in the following manner:
Press the [N] or [H] keys to open the database.
Use the [N] or [H] keys to highlight the ID number which should be modified.
Press the [Load] key to transfer the set of data from the main memory to the
temporary memory.
Perform the new measurement or change the evaluation (see chapter 6.2).
Press the [save] key to store the modified data set back to the main memory.
5.6 Test sequences
5.6.1 Description of the test sequence
The test sequence is the fastest and easiest way to test appliances. During the test, preprogrammed measurements run automatically in a sequential order. The com­plete test results can be stored together with the associated appliance ID. The automatic measurement procedure consists of individual steps which are neces­sary to do step by step and each step has to be confirmed. The automatic procedure is created in the PC software. The steps of measurement are selected from the se­lection and become part of the definition of the appliance. The process can include all variables that the measuring device MDtest allows. Furthermore, you can define their order. After activation, the test sequence with the symbol SEQ is displayed in the upper right corner of the display. The automatic process can be interrupted at any time and continue with the manual measurement mode. The automatic procedure always consists of the following steps:
Confirm whether the user wishes to start the measurement procedure
The result of the visual inspections (required)
Selected measuring methods (optional item)
• ...
The result of the functional test (required)
The measurement result (required)
The question of whether the values are to be stored in the main memory
5.6.2 Creating a test sequence
The test sequence can only be created using the PC software. Here measurements can be defined, including their sequence. The procedure also automatically assigns the result of the visual inspection, functional test and measurement operation.
c
2018 ILLKO, s.r.o. www.illko.cz Page 50
Some measurements cannot be measured separately (e.g. IEq, Uline and S, etc.). If are these variables are independently selected, mea­surement is performed at the same time.
5.6.3 Assigning a test sequence to appliance
In order to use the automatic procedure, is necessary to activate it. The automatic measurement procedure can be activated in the following ways:
An automatic process is transferred from the PC individually for a particular appliance, at the same time as the file’s data. The test sequence is automati­cally activated before starting the measurement.
If the file does not contain the loaded data for an test sequence, the appliance can assign any of the procedures stored on the MDtest (see chapter 6.3).
5.6.4 Using the test sequence
If the tested appliances are assigned to the automatic measurement procedure, it is possible to start a test of the appliance using the following procedure:
Confirm the Sequence wizard measurements using the [TEST] key. If you press the [esc] key, the automatic procedure will be shut down and the mea­surement can be done without using the test sequence.
Follow the instruction shown on the display or select the result of the evalua­tion.
If you want to repeat a step, press the [Repeat] key. If the step was performed successfully and you wish to proceed, press [TEST].
If you want to interrupt the automatic process and switch to the manual mode, press the [End] key. The steps which have been performed up to now will be retained.
During the test sequence procedure, it is not possible to store the measured values with the [save] key. The results are stored after confirmation that the procedure is completed.
For some measurements the middle context button can be displayed. For example: IaltPat - change resolution 4/5 digits.
c
2018 ILLKO, s.r.o. www.illko.cz Page 51
MDtest (100-140 V) - User’s Manual 6 Instrument main menu
6 Instrument main menu
6.1 Main menu structure
DUT properties
Measured valuesVisual inspection resultFunctional test resultMeasurement result
Test sequences
Empty file
Setup
System infoMeasure DC valuesSelf-testDate – next calibrationErase databaseMeasurement lockingSave last methodDate and TimeKeys beepVolumeLanguage
The operating the menu is as follows:
Press the [menu] key to open the Main menu.
Use the [N] or [H] keys to select the desired function, confirm the selection by
pressing the [enter] key.
Use the [esc] keys to return.
6.2 Menu - DUT properties
6.2.1 Menu - Measured values
This menu can be used to display measured values with an evaluation status. All data are transferred from the temporary memory.
6.2.2 Menu - Visual inspection result
6.2.3 Menu - Functional test result
In this menu it is possible to set the result of functional test result: Failed / Passed / N/A (not performed).
c
2018 ILLKO, s.r.o. www.illko.cz Page 52
MDtest (100-140 V) - User’s Manual 6 Instrument main menu
6.2.4 Menu - Measurement result
The default setting for the visual inspection, functional test and mea­suring results is PASSED.
6.3 Menu - Test sequences
In this menu you can view a list of available measurement test sequences which can be assigned to the currently selected appliance.
Press the [menu] key.
Select the Test sequences item and press the [ent] key for confirmation.
Select required test sequence.
Press the [ent] key to check each individual step in the test sequence.
Press the [Load] key to assign the test sequence to DUT.
The list of test sequences cannot be modified in MDtest – it is trans­ferred from the PC software.
6.4 Menu - Empty file
In this menu the temporary memory is erased, i.e.:
the ID number is empty
all measured values are deleted
the results of the visual inspection, functional test and measurement operation
are set to PASSED
the test sequence is deactivated
6.5 Menu - Setup
6.5.1 Menu - System info
6.5.2 Menu - Measure DC values
This will activate the measurement and display DC components in the selected leakage currents:
c
2018 ILLKO, s.r.o. www.illko.cz Page 53
MDtest (100-140 V) - User’s Manual 6 Instrument main menu
IdirEq(R)
I60601Earth(R) + SFC, variant Open N
I60601Touch(R) + SFC, variant Open N or Open PE
I60601Pat(R) + SFC, variant Open N or Open PE
If measurement of a DC component is enabled, that measurement cycle will take longer. The value is transmitted to the PC software.
6.5.3 Menu - Self-test
This option can be performed to verify the functionality of the MDtest. Disconnect equipment under test before the self-test. The self-test does not replace calibration. The procedure is described in chapter 3.2.
6.5.4 Menu - Date of next calibration
In this menu, you can set the next calibration date. The measuring device starts to display a warning message 45 days before the expiration date, e.g., Date of the next calibration: 01.01.2018. This information is displayed only once a day. If the calibration has expired, there is the message, e.g., Calibration expired on
01.01.2018. Enter into the menu, the date of the next calibration is performed by pressing the [ent] key. The date is inserted using the keyboard device. Confirm the selection by pressing the [ent] key or press the [esc] key to return without saving the changes.
The ’.’ characters are automatically skipped in the editor.
6.5.5 Menu - Erase database
Use this option to erase the entire main memory device, i.e., database appliances. The essential instrument settings and a list of test sequences remain preserved. Erasing the memory must be confirmed by pressing the [Yes] key.
How long it takes to erasing entries depends on the memory usage. It can take up to 2 minutes when all memory locations are occupied.
6.5.6 Menu - Measurement locking
This option allows you to automatically lock every measurement. When enabled, each time measurement is displayed the symbol AUT.ARET . The switch activation is done by pressing the [ent] key.
c
2018 ILLKO, s.r.o. www.illko.cz Page 54
MDtest (100-140 V) - User’s Manual 6 Instrument main menu
6.5.7 Menu - Save last method
This option allows you to save last measuring method. When enabled, last method is selected immediately after connection the MDtest to the mains. The switch activation is done by pressing the [ent] key.
6.5.8 Menu - Date and Time
The date and time can be set in this menu. The date is attached to each stored test result and then transferred to the PC software. Press the [ent] key to enter this submenu. Insert the date and time via the MDtest keypad and confirm by the [ent] key or press the [esc] key to return without saving the changes.
The ’.’ characters are automatically skipped in the editor. When you transfer data from the PC to the MDtest, the internal clock is automatically synchronized with the PC’s time.
6.5.9 Menu - Keys beep
In this menu, an audible indication of pressing a key can be enabled / disabled. The change is done by pressing the [ent] key.
6.5.10 Menu - Volume
In this menu, the volume is set of the audible indication. Press the [ent] key to change the level.
6.5.11 Menu - Language
The instrument language can be set in this menu. Press the [ent] key to change the language.
6.5.12 Menu - Nominal voltage
The nominal voltage 120 V or 127 V can be set in this menu. Press the [ent] key to change it. The value 127 V is preset by default. This setting is important for the correct calculation of the leakage current.
c
2018 ILLKO, s.r.o. www.illko.cz Page 55
MDtest (100-140 V) - User’s Manual 7 Additional informations
7 Additional informations
7.1 Communication – data transfer
Data transfer between the MDtest and the PC is completely controlled by the PC software and it cannot be controlled directly from the measuring device. To transfer data, prepare the MDtest as follows:
Connect the MDtest to a PC using a USB cable plugged into connector 9 (drivers must be installed on the PC for the proper use of USB communica­tion).
If the MDtest is in the mode of working with databases, shut it down and set the device to measuring mode.
Before transferring data from a PC to the MDtest, memory will be erased. How long it takes to erasing entries depends on the memory usage. It can take up to 2 minutes when all memory locations are occupied. Attention: data
unsaved to the PC will be lost forever!
Do not operate with the instrument during data transfer.
An informative window is displayed during data transfer.
7.2 Messages on the display
7.2.1 Warning messages
PE error
A fault occurred in the PE circuit. The cause may be:
the PE conductor circuit is not sufficiently grounded (high impedance fault loop or IT system)
voltage on the protective wire
All device functions are blocked. If it is the IT system, the user must confirm this condition (see Figure 23). The resistance of the PE conductor, substitute leakage current and insulation resistance can be only measured in IT systems. In other cases, we cannot continue measuring (see Figure 24).
The test of the PE circuit is performed only once, immediately after connection to the mains.
Figure 23: Warning message: PE error (IT systems)
c
2018 ILLKO, s.r.o. www.illko.cz Page 56
MDtest (100-140 V) - User’s Manual 7 Additional informations
Figure 24: Warning message: PE error
Interference voltage at the input
The device has blocked the measuring. The cause may be:
Interference voltage on the DUT
The measuring instrument tests (before start of Riso and Rpe measurements) if interference voltage is not present on the DUT. If the warning message (Figure 25) appears, then check the disconnection of the DUT from the mains. Another reason could be that the measuring leads are too long. Please, use the original leads only.
Interference voltage in the mains
If during measurement a greater interference in the mains occurs, the MDtest disables the measuring and the warning message (Figure 25) appears. In this case, repeat the measurement later.
Figure 25: Warning message: Interference voltage at the input
Current >10.00 mA
If the leakage current of the appliance exceeds approximately 10.00 mA, power supply is disconnected from the Test Socket 1, then an acoustic signal sounds and the display shows the warning message (Figure 26).
c
2018 ILLKO, s.r.o. www.illko.cz Page 57
MDtest (100-140 V) - User’s Manual 7 Additional informations
Figure 26: Warning message: Current >10.00 mA
Measuring - time limit
If the time of measuring exceeds the allowed limit (according to the appliance’s current - see chapter 2.1), power supply is disconnected from the Test Socket 1, then an acoustic signal sounds and the display shows the warning message (Figure 27).
Figure 27: Warning message: Measuring - time limit
Cannot be measured (calibration error / damaged data)
Figure 28: Warning message: Cannot be measured (calibration error)
Figure 29: Warning message: Cannot be measured (damaged data)
c
2018 ILLKO, s.r.o. www.illko.cz Page 58
MDtest (100-140 V) - User’s Manual 7 Additional informations
Empty database
When you try to view a list of appliances with an empty memory, then an acoustic signal sounds and the display shows the warning message (Figure 30).
Figure 30: Warning message: Empty database
Memory usage
Figure 31: Warning message: Memory usage
Measurement was cancelled
If the [esc] key was pressed during the measurement, then it will be cancelled. An acoustic signal sounds and the display shows the warning message (Figure 32). The measured value is not saved.
Figure 32: Warning message: Measurement was cancelled
c
2018 ILLKO, s.r.o. www.illko.cz Page 59
MDtest (100-140 V) - User’s Manual 7 Additional informations
ID is already in database
When you enter an ID which is already in the database, then an acoustic signal sounds and the display shows the warning message (Figure 33).
Figure 33: Warning message: ID is already in database
Test leads: Rpe >2.0
The maximum value of the test leads’ resistance is 2.0 . When you try to compen­sate for a higher value, an acoustic signal sounds and the display shows the warning message (Figure 34).
Figure 34: Warning message: Test leads: Rpe >2.0
Warning concerning of a point of danger
Read the User’s Manual and observe all precautions! Symbol (Figure 35) highlights the fact that it is necessary to follow the safety instructions in the manual. A RISK OF INJURY OR DEVICE DAMAGE!
Figure 35: Warning concerning of a point of danger
c
2018 ILLKO, s.r.o. www.illko.cz Page 60
MDtest (100-140 V) - User’s Manual 7 Additional informations
Warning concerning a DUT isolated placing
Symbol (Figure 36) highlights the fact that all conductive parts must be isolated from accidental grounding.
Figure 36: Warning concerning a DUT isolated placing
Warning concerning heating the internal circuit when Rpe is measured
When low Rpe values are measured, it may overheat the internal circuit of the MDtest. We recommend performing the measurements for a max. of 30 secs. (only for value <10.00 ). If the temperature inside exceeds the allowed limit, the symbol (Figure 37) is displayed and the acoustic signal sounds.
Figure 37: Warning concerning heating the internal circuit when Rpe is measured
c
2018 ILLKO, s.r.o. www.illko.cz Page 61
MDtest (100-140 V) - User’s Manual 7 Additional informations
7.2.2 Operational information
Measuring in progress
An animated symbol indicates that the measurement is running (Figure 38). It is displayed only during measurement and a LED indicator 33 simultaneously flashes.
Figure 38: Measuring in progress
Locking the measurement
The locking can be cancelled by pressing the [TEST] key or any key except for [H] and [esc].
Figure 39: Locking the measurement
Automatically locking the measurement
The locking can be cancelled by pressing the [TEST] key or any key except for [H] and [esc]. This setting can be enabled / disabled in Setup menu, see chapter 6.5.6.
Figure 40: Automatically locking the measurement
c
2018 ILLKO, s.r.o. www.illko.cz Page 62
MDtest (100-140 V) - User’s Manual 7 Additional informations
HOLD function
By pressing and releasing the [H] key, the last measured value is displayed even if the measurement has either ended or is in progress. The HOLD function can be cancelled (measurement is still in progress) by pressing and releasing the [H] key.
Figure 41: HOLD function
Test sequence
After activation, the test sequence symbol is displayed (Figure 42).
Figure 42: Test sequence
Measured value is out of range
If the value is higher / lower then max. / min. range, the maximal / minimal value is displayed with the symbol > "higher then" (Figure 43) or < "lower then" (Figure 44).
Figure 43: Measured value is out of range (max)
Figure 44: Measured value is out of range (min)
c
2018 ILLKO, s.r.o. www.illko.cz Page 63
MDtest (100-140 V) - User’s Manual 7 Additional informations
7.2.3 Setting information
Test leads compensation
The test leads’ resistance is measured and then automatically memorized and sub­tracted from Rpe measurement (see chapter 4.3.1).
Figure 45: Test leads compensation
Incorrect compensation for the test leads
If the measured value is less than the saved compensation value during measurement (after subtraction, the memorized resistance will result in a negative number), the display shows the absolute value and the symbol CAL on the display appears inversely (Figure 46). In this case, it is necessary to perform the compensation again (see chapter 4.3.1).
Figure 46: Incorrect compensation for the test leads
External input activation
After activation of the external input 10, a symbol is displayed (Figure 47). The Test Socket 1 is not active.
Figure 47: External input activation
c
2018 ILLKO, s.r.o. www.illko.cz Page 64
MDtest (100-140 V) - User’s Manual 7 Additional informations
7.2.4 Database information
The appliance has associated data about a performed measurement
When data transfer from the temporary memory to the main memory is performed, then the symbol is displayed. The symbol is displayed even if no measurement has been performed.
Figure 48: The appliance has associated data about a performed measurement
The appliance has associated data about a test sequence
Figure 49: The appliance has associated data about a test sequence
7.3 Reset of the instrument
It is advised to execute an instrument RESET in the event of a malfunction. Dis­connect the MDtest from mains, wait 30 s and plug it in again. If problems occur with the database system of MDtest, please reset the device and erase the memory (see chapter 6.5.5).
Attention! All data stored in the main memory will be lost.
If malfunction persists, contact your local distributor.
c
2018 ILLKO, s.r.o. www.illko.cz Page 65
MDtest (100-140 V) - User’s Manual 7 Additional informations
7.4 Utilization of other external adapters
7.4.1 Adapter for testing extension power supply cords (P 8030)
The adapter is designed to measure the continuity of the protective circuit extension cords. It facilitates the work and significantly improves the contact of their PE pins. Connect the banana plug adapter into the RPE terminal, plug the adapter into Test Socket 1 on the front panel MDtest and perform the test lead resistance compensa­tion of the adapter (see chapter 4.3.1). Then connect the plug of the measured extension cord into Test Socket 1, lock the measurement (see chapter 4.3) and plug the adapter into the socket of the tested extension cord. The resistance of the PE circuit is now measured.
7.4.2 Three-phase adapter (P 8015-8018)
The three-phase adapter must be used for measuring three-phase appliances with a movable power supply cord. It is necessary to plug in tested appliance by means of the three-phase adapter and to select the required measuring method by means of the three phase adapter’s function switch before measurement.
The three-phase adapters are produced in different versions:
P 8015 - Three-phase adapter, 16 A socket, 5 contacts P 8016 - Three-phase adapter, 16 A socket, 4 contacts P 8017 - Three-phase adapter, 32 A socket, 5 contacts P 8018 - Three-phase adapter, 32 A socket, 4 contacts
7.5 Maintenance
Regularly check the technical safety and integrity of the insulation on the power supply cord, plastic case and all accessories. The plastic case should not be polluted with substances which can noticeably degrade the case’s insulation quality. Use soft cloth, slightly moistened with lukewarm soap water to clean the plastic case. Do not spill cleaning liquid over the instrument! Do not use cleaning liquids based on petrol, hydrocarbons, etc. Wait until the instrument becomes totally dry before using it!
WARNING Do not expose the instrument to temperatures over 70◦C. If this temperature is exceeded, the plastic case can be dam­aged!
c
2018 ILLKO, s.r.o. www.illko.cz Page 66
MDtest (100-140 V) - User’s Manual 7 Additional informations
7.6 Recurrent tests and testing after repair
The MDtest is powered with a mains voltage of 127 V / 50 or 60 Hz and therefore all recurrent tests and testing after repair should be done in accordance with local regulations.
c
2018 ILLKO, s.r.o. www.illko.cz Page 67
MDtest (100-140 V) - User’s Manual 7 Additional informations
7.7 Calibration and service
Measuring instruments should be regularly calibrated. We recommend an inter­val of calibration 1 year.
Furthermore, we recommend carrying out calibration after each repair.
Please contact us to fix an appointment for the inspection and calibration of your test instruments, for the address see below:
ILLKO, s.r.o. Masarykova 2226 678 01 Blansko Czech Republic
tel./fax: (+420) 516 417 355 e-mail: illko@illko.cz http://www.illko.cz
WARNING
Unauthorized persons are not allowed to open the MDtest. There are no replaceable components inside the instrument.
Please remember to include your address, email contact, phone number and de-
scription of any faults with the package.
7.8 Ecology
Transportation packing
Transport packaging is made of corrugated cardboard; it is recyclable.
Instrument
This symbol signifies that the product should not be thrown away into municipal waste at its end-of-life. Please dispose of this product according to the relevant statutory re­quirements.
c
2018 ILLKO, s.r.o. www.illko.cz Page 68
MDtest (100-140 V) - User’s Manual 8 Technical specification
8 Technical specification
8.1 General data
Protective class . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . II (double insulation)
Over voltage class . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . CAT II/300V
Pollution degree . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . 2
Degree of protection . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . device: IP 40
test socket: IP 20
Reference condition:
Line voltage . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . . . . . 127 V ± 1 % / 60 Hz ± 1 %
Ambient temperature . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . (23 ± 2)◦C
Humidity . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . 45 ÷ 55 % @ 23◦C
Position . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . arbitrary
Operating condition:
Line voltage . .. . . . . . . . . . . . . . . . .. . . 100÷140 V / 50 Hz ± 1 % or 60 Hz ± 1 %
Ambient temperature . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . 0 ÷ 40◦C
Humidity . . . . . . . . . . . . . . . .. . . .. . . . . . 75 % @ 23◦C (condensation not allowed)
Elevation . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . max. 2 000 m
Supply current . . . . . . . . . . . max. 16 A (it depends on tested appliance, time limit)
Quiescent current . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . < 25 mA
Dimensions (w x d x h) . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . 250 x 195 x 50 mm
Mass (without accessories) . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . 2.1 kg
Storage temperature / storage humidity . . . . . . . . . . . . . . . -10 ÷ 50◦C / max. 70 %
Case resistance to temperature . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . 70◦C
PE test . . .. . . . . . . . . . . . . . . . . . . . .. . . .. . . . . . . . . before the start of the measurement
PC connection . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . USB cable
8.2 Functions
Notes: R - reading, D - digit
8.2.1 Earth bond
Entire measuring range . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . . . . . .. . 0.000 ÷ 10.00
Measuring sub range 0.000 ÷ 0.999
Resolution . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . 0.001
Reference error (LS< 200 µH) . . . . . . . . . . . . . . . . . . . .. . . .. . ± (2 % of R + 6 D)
Operating error (LS< 200 µH) . . . . . . . . . .. . . .. . . . . . . . . . . . ± (3 % of R + 8 D)
Measuring sub range 1.00 ÷ 10.00
Resolution . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . 0.01
Reference error (LS< 200 µH) . . . . . . . . . . . . . . . . . . . .. . . .. . ± (2 % of R + 5 D)
Operating error (LS< 200 µH) . . . . . . . . . .. . . .. . . . . . . . . . . . ± (3 % of R + 7 D)
Operating range of use (according EN 61557-4) . . . . . . . .. . . . . . . . . 0.030 ÷ 10.00
Open-circuit voltage . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . 5 ÷ 12 V AC
Test current (R 1 ) . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . 0.8 A AC
Test lead compensation . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . yes (up to 2 )
c
2018 ILLKO, s.r.o. www.illko.cz Page 69
MDtest (100-140 V) - User’s Manual 8 Technical specification
8.2.2 Insulation resistance
Entire measuring range . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. 0.100 ÷ 100.0 M
Measuring sub range 0.100 ÷ 3.999 M
Resolution . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . 0.001 M
Reference error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . ± (2 % of R + 8 D)
Operating error . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. ± (3 % of R + 8 D)
Measuring sub range 4.00 ÷ 19.99 M
Resolution . . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . 0.01 M
Reference error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . ± (2 % of R + 8 D)
Operating error . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. ± (3 % of R + 8 D)
Measuring sub range 20.0 ÷ 100.0 M
Resolution . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . 0.1 M
Reference error . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . ± (4 % of R + 15 D)
Operating error . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . ± (5 % of R + 15 D)
Operating range of use (according EN 61557-2) . . .. . . . . . . . . . . . 0.100 ÷ 100.0 M
Nominal output voltage Un . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . 250 V DC
Open-circuit voltage . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . (-0 %, +50 %) Un
Nominal test current . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. 1 mA DC
Short-circuit current .. . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . < 10 mA DC
Self-test . . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . 1.000 ± 0.050 M
8.2.3 Alternative leakage current
Entire measuring range . . . . . . . . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . 0.000 ÷ 20.00 mA
Measuring sub range 0.0000 ÷ 0.0999 mA
1
Resolution . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . 0.0001 mA
Reference error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . ± (2 % of R + 9 D)
Operating error . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. ± (3 % of R + 9 D)
Measuring sub range 0.000 ÷ 3.999 mA
Resolution . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . 0.001 mA
Reference error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . ± (1 % of R + 8 D)
Operating error . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. ± (2 % of R + 8 D)
Measuring sub range 4.00 ÷ 20.00 mA
Resolution . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . 0.01 mA
Reference error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . ± (1 % of R + 5 D)
Operating error . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. ± (2 % of R + 5 D)
Operating range of use . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . 0.028 ÷ 20.00 mA
Nominal open-circuit test voltage . . . .line voltage ± 15 %, line frequency ± 0.5 Hz
Short-circuit current . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. < 3.5 mA
Internal resistance of test voltage source . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . > 80 k
Self-test . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . 0.100 ± 0.010 mA
1
When measuring patient leakage current and higher resolution (5 digits) is set.
c
2018 ILLKO, s.r.o. www.illko.cz Page 70
MDtest (100-140 V) - User’s Manual 8 Technical specification
8.2.4 Direct leakage current (TRMS AC+DC value)
Entire measuring range . . . . . . . . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . 0.000 ÷ 10.00 mA
Measuring sub range 0.000 ÷ 1.499 mA
Resolution . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . 0.001 mA
Reference error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . ± (1 % of R + 3 D)
Operating error . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . ± (1.5 % of R + 4 D)
Measuring sub range 1.50 ÷ 10.00 mA
Resolution . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . 0.01 mA
Reference error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . ± (1 % of R + 5 D)
Operating error . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. ± (2 % of R + 7 D)
Operating range of use . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . 0.030 ÷ 10.00 mA
Test voltage . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . line voltage
Frequency characteristic of internal test circuitry . . . . . . according IEC 62353:2014
Self-test . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . 0.100 ± 0.010 mA
8.2.5 Direct leakage current (DC component)
Entire measuring range . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . 0.00 ÷ 10.00 mA
Resolution .. . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . 0.01 mA
Reference error . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . ± (2 % of R + 5 D)
Operating error . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . ± (3 % of R + 8 D)
Operating range of use . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . 0.30 ÷ 10.00 mA
Test voltage . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . line voltage
Frequency characteristic of internal test circuitry . . . . . . according IEC 62353:2014
8.2.6 Differential leakage current
Entire measuring range . . . . . . . . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . 0.000 ÷ 10.00 mA
Measuring sub range 0.000 ÷ 0.999 mA
Resolution . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . 0.001 mA
Reference error . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . ± (2 % of R + 5 D + *X)
Operating error . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . ± (3 % of R + 8 D + *X)
Measuring sub range 1.00 ÷ 10.00 mA
Resolution . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . 0.01 mA
Reference error . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . ± (2 % of R + 5 D + *Y)
Operating error . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . ± (3 % of R + 5 D + *Y)
Operating range of use . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . 0.036 ÷ 10.00 mA
Test voltage . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . line voltage
Frequency characteristic of internal test circuitry . . . . . . according IEC 62353:2014
Self-test . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . 0.100 ± 0.010 mA
*X = S (0.01 I+ 0.006) [D] *Y = S (0.001 I+ 0.0006) [D]
S ..... Apparent power of tested appliance [VA]
I... Displayed value of leakage current [mA]
c
2018 ILLKO, s.r.o. www.illko.cz Page 71
MDtest (100-140 V) - User’s Manual 8 Technical specification
8.2.7 Leakage current by means of three-phase adapter
Entire measuring range . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . 0.00 ÷ 20.00 mA
Resolution .. . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . 0.01 mA
Reference error . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . ± (3 % of R + 10 D)
2
Operating error . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . ± (4 % of R + 15 D)
2
Operating range of use . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . 0.58 ÷ 20.00 mA
8.2.8 Apparent power
Entire measuring range . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . 10 ÷ 2240 VA
Resolution . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . 1 VA
Reference error . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . ± (3 % of R + 2 D)
Operating error . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . ± (5 % of R + 5 D)
8.2.9 Mains voltage UL-N
Entire measuring range . . . . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . 100.0 ÷ 140.0 V
Resolution .. . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . .. . . . . . 0.1 V
Reference error . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . ± (1 % of R + 2 D)
Operating error . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . ± (2 % of R + 2 D)
8.2.10 Current consumption
Entire measuring range . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . 0.0 ÷ 16.0 A
Resolution .. . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . .. . . . . . 0.1 A
Reference error . . . .. . . .. . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . ± (2 % of R + 1 D)
Operating error . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . ± (3 % of R + 2 D)
8.2.11 Current consumption by means of current clamp
Entire measuring range . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . 0.0 ÷ 100.0 A
Resolution .. . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . .. . . . . . 0.1 A
Reference error . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . ± (3 % of R + 4 D)
2
Operating error . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . ± (4 % of R + 4 D)
2
8.2.12 Apparent power by means of current clamp
Entire measuring range . . . . . . . . . .. . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . 10 ÷ 9999 VA
Resolution . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . . 1 VA
Reference error . . . . . . . . .. . . .. . . . . . . . . . . . . . . . .. . . .. . . . . . . . . . ± (5 % of R + 5 D)
2
Operating error . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . .. . ± (8 % of R + 5 D)
2
2
Current clamp or three-phase adapter error has to be considered additionally.
c
2018 ILLKO, s.r.o. www.illko.cz Page 72
ILLKO, s.r.o., Masarykova 2226, 678 01 Blansko, Czech Republic
DECLARATION OF CONFORMITY
Apparatus Identification: MDtest (IL2550) Apparatus Classification: Test and Measurement Equipment
Statement of Conformity:
Based on sample product test result using appropriate standards, and in accordance with the folloving EC directives ILLKO, s.r.o. hereby declares the MDtest (IL2550) to be in conformity with:
EC Directive LVD 2006/95/EC EC Directive EMC 2004/108/EC
Sample Product Testing for EMC:
Tested by: INSTITUTE FOR TESTING AND CERTIFICATION, a.s.
Testing Laboratory No. 1004.3 Division 4 - MESIT QM Sokolovská 573 Uherské Hradiště Czech Republic
Standard used As Czech implementation of
ČSN EN 55011 ed.3:2010 +A1:2011 EN 55011:2009 +A1:2010 ČSN EN 61326-1 ed.2:2013 EN 61326-1:2013 ČSN EN 61000-4-2 ed.2:2009 EN 61000-4-2:2009 ČSN EN 61000-4-3 ed.3:2006+A1:2008 +A2:2011 EN 61000-4-3:2006+A1:2007+A2:2010 ČSN EN 61000-4-4 ed.2:2005+A1:2010 EN 61000-4-4:2004+A1:2010 ČSN EN 61000-4-5 ed.2:2007 EN 61000-4-5:2006 ČSN EN 61000-4-6 ed.3:2009 EN 61000-4-6:2009 ČSN EN 61000-4-11 ed.2:2005 EN 61000-4-11:2004
Report ID: EMC: Test Report No. 41410274AE1, issued June 11, 2015 Test Report about Electromagnetic Test on the Measuring Instrument MDtest (IL2550)
Sample Product Testing for safety:
Tested by: INSTITUTE FOR TESTING AND CERTIFICATION, a.s.
Testing Laboratory No. 1004.3 Division 4 - MESIT QM Sokolovská 573 Uherské Hradiště Czech Republic
Standard used As Czech implementation of
ČSN EN 61010-1 ed.2 :2011 6, A13: 2009, A14: 2010, Opr.4: 2010, A15: 2012 EN 61010:2010 ČSN EN 61010-2-030:2011 EN 61010-2-030:2010
Report ID: LVD: Test Report No. 414102746AL1, issued June 30, 2015 Test Report about Electric Safety Test on the Measuring Instrument MDtest (IL2550)
Ivo Lipovský
Issue Date: 2015-07-12 Product Manager
Manufacturer, service:
ILLKO, s.r.o. tel./fax: (+420) 516 417 355
Masarykova 2226 illko@illko.cz
678 01 Blansko www.illko.cz
Czech Republic
Last modification: January 8, 2018 Document revision: 1.0
This manual is current when is delivered together with the device, but may not correspond to the older or newer version of firmware. The firmware can be upgraded during calibration, repair, or by ap­pointment. When the firmware was upgraded, please check our website www.illko.cz for modified User’s manual.
Copyrightc 2018, ILLKO, s.r.o.
Copying or any other way of duplication of this document or its parts is
allowed with antecedent written permission of ILLKO, s.r.o. only!
Loading...