INTRODUCTION
Thank you for purchasing the Model 2100 Conductivity Meter from ILIUM Technology,
Inc. The Model 2100 is intended for use as a laboratory instrument and is capable of
providing accurate measurements of the temperature and solution conductivities from
1 fS/cm to 1 mS/cm. The meter is completely auto-ranging, with the accessible
conductivity range determined by the probe chosen, with each probe spanning over 8
decades of conductivity. Most measurements require simply placing the probe in the
sample and reading the conductivity and temperature on the meter’s display. The
simple-to-use menu system allows control of the display brightness and contrast, a
choice of the conductivity and temperature units displayed, information about the meter
and the probe currently in use, and the ability to lock the display into a specific
conductivity range if desired. Additional capabilities are available with computer
communication with the meter through its USB port.
Conductivity Probes
The Model 2100 meter is designed to work with a family of conductivity probes, each of
which allows the user to access eight decades of conductivity. The standard meter
package includes a Model 1020 fully immersible, mini-dip probe, which covers the
conductivity range from 25 fS/cm up to a few µS/cm. The parts of the probe that come in
contact with the user’s samples are made of 316 stainless steel, PEEK, PTFE, and
FEP, depending on the depth that the probe and cable are immersed.
The mini-dip probes are designed to be robust, fully immersible, easy to use, and
extremely easy to clean. Many solvent based dispersions, such as paints, inks, or
water/oil emulsions are difficult to remove from electrode surfaces simply by flushing
with additional solvents. The ILIUM dip probes allow the user to remove the outer
cylinder, providing access to the electrode surfaces so they can be wiped clean with an
appropriate, non-abrasive swab or wipe.
Features
Auto-ranging meter with ability to lock range
Simultaneously capture conductivity and temperature measurements
User selectable units for conductivity and temperature
Interchangeable, factory-calibrated probes ensure accuracy and accommodate
switching between applications
Fully-guarded probes to ensure accurate measurements
Easy to clean modular design
Easy reassembly with no change in cell constant
Small sample size (< 3mL) required
USB port for control of meter and data acquisition
Included data acquisition software tracks data from up to 3 instruments
simultaneously
1
Instruction Manual Model 2100 Conductivity Meter
Unpacking
Check for any damage on the shipping container that might suggest inappropriate
handling during shipping. When you first unpack the Model 2100, check to ensure that
you find all the items listed on the packing list and look for any signs of damage that
may have occurred during shipping. Report any damage to the shipper immediately.
Note that there is a protective film over the display window that should be removed
before using the meter.
The anti-static bag and shipping container should be saved for future shipments,
including returns for annual calibration or other service.
SYMBOLS AND HEADINGS
WARNING
This heading in the user documentation explains dangers that might result in personal
injury or death.
CAUTION
This heading in the user documentation explains hazards that could damage the
instrument. Such damage may invalidate the warranty.
GENERAL INFORMATION
CAUTION: Please read this user guide thoroughly before using your meter. Any use
outside of these instructions may invalidate your warranty and cause permanent
damage to the meter.
CAUTION: To clean the instrument, use a damp cloth. Clean the exterior of the
instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter
or spill on the instrument. To clean the probe, remove the outer cylinder, flush all
contaminated surfaces with an appropriate solvent, carefully wipe the electrode
surfaces clean with an appropriate, non-abrasive swab or wipe, rinse again with the
solvent, then replace the outer cylinder.
WARNING: There are no user serviceable parts in this meter. Please return the meter
to ILIUM if service is required.
CAUTION: This meter is intended for use only with the ILIUM Power Supply included
with the meter (ILIUM part no. 01-262-0500001). Use of an AC Adapter other than the
one supplied by ILIUM may damage the meter.
2
Instruction Manual Model 2100 Conductivity Meter
WARNING: Ensure that the AC Adapter is readily accessible when positioning the
device, as it is the main AC disconnect. To fully disconnect the unit from the AC supply,
remove the AC Adapter from the wall outlet.
CAUTION: This meter is designed to be used only with ILIUM model 2100 compatible
probes, adapters, and accessories. The use of probes, adapters, or accessories not
specifically listed by ILIUM as compatible with this meter may damage the meter and
will void the warranty. A list of all currently available compatible probes and accessories
is available from ILIUM.
WARNING: If the equipment is not used in accordance with the instructions provided in
the User Manual, the protection provided by the conductivity meter may be impaired.
WARNING: This manual does not profess to address all of the safety problems
associated with the use of the Model 2100 and the samples it is used to measure. It is
the responsibility of the user of the meter to establish safety and health practices and
determine the applicability of regulatory limitations prior to use.
WARNING: When handling and measuring flammable samples and cleaning materials:
Ensure that the Model 2100 is located in a sufficiently ventilated area or fume
hood.
Do not store flammable material near the Model 2100.
Do not leave sample containers uncovered.
Clean all spills immediately.
The specifications, descriptions, drawings, ordering information and part numbers within
this document are subject to change without notice.
3
Instruction Manual Model 2100 Conductivity Meter
r
y
CONTROLS, CONNECTIONS, AND DISPLAY
Front Panel
The front panel contains the 16 x 2 character display, the probe connector, and seven
control keys as shown below in Figure 1. Table 1 lists the control keys and describes
the function of each.
16 x 2 Character Displa
Figure 1
Probe Connecto
Table 1
Key Name Function
On – Off Pressing turns on meter. If on, pressing and holding for 4
seconds turns off meter.
Range (Escape) If meter is in normal operating mode, pressing toggles
between Auto Ranging (AR) and Display Lock (DL) range
Menu (Enter) If meter is in normal operating mode, pressing enters the
Left In menu mode, moves to the item or menu selection to
Right In menu mode, moves to the item or menu selection to
Down In Display Lock (DL) operating mode, decreases display
Up In Display Lock (DL) operating mode, increases display
mode. If in Menu mode, pressing jumps up one menu
level without saving any changes.
menu mode. If in menu mode, pressing enters the
current selection.
the left.
the right.
resolution. In menu mode, adjusts a setting down or
moves down one menu level.
resolution. In menu mode, adjusts a setting up or moves
up one menu level without saving any changes.
All ILIUM Model 2100 compatible probes, adapters, and accessories plug into the front
panel Probe Connector. A list of all currently available compatible probes and
accessories is available from ILIUM.
4
Instruction Manual Model 2100 Conductivity Meter
Display
The meter has a 16 character by 2 line, backlit, LCD display. During normal
measurement operation the main display screen will appear as shown in Figure 2.
Conductivity
Figure 2
Temperature
The conductivity will be presented in up to four digits, using 0, 1, or 2 digits to the right
of the decimal place. The available units include femtoSiemens (fS), picoSiemens (pS),
nanoSiemens (nS), microSiemens (µS), and milliSiemens (mS). As described below in
the Getting Started section, the user can choose to see the conductivity values in
Siemens per centimeter or Siemens per meter. Note that this setting is persistent over
on/off cycles.
The temperature is displayed on the second line of the display to the nearest 0.1°. As
described below in the Getting Started section, the temperature units can be set to
either °C or °F. Note that this setting is persistent over on/off cycles. If the connected
probe does not have a temperature sensor, this area will remain blank. If the connected
probe does have a temperature sensor, but the meter cannot obtain a reading, it will
display “ - - - . -°C” or “ - - - . -°F”. This will occasionally happen when the instrument
momentarily cannot get an accurate temperature reading. In this case, the temperature
will return shortly. If the temperature does not return, or is never available when you
connect the probe, the temperature capability has been damaged and the probe should
be returned for repair.
The Status is shown just to the right of the conductivity units. An asterisk is shown when
the meter has not reached a completely stable reading. Once a stable reading is
reached, the asterisk will turn off. This status does not imply that your sample has
reached a steady value, but that the instrument is satisfied with the immediate value
reported. Occasionally, the value displayed may appear to be stable even though the
asterisk is displayed. This indicates that the instrument is not confident in the precision
of the answer.
In the upper right hand corner of the display, the Mode is designated as either AR for
Auto Range mode or DL for Display Lock mode as described in the Making Measurements section below.
Conductivity
Units
Status
35.24pS/cm* AR
25.0ºC
Temperature
Units
Mode
5
Instruction Manual Model 2100 Conductivity Meter
p
Rear Panel
The rear panel contains the connections for the power supply and USB communications
cable as shown below in Figure 3.
Figure 3
Power Supply Input
USB cable in
ut
Power Supply Connection
The ILIUM power supply is designed to work with inputs of 100 to 240 VAC at 50 to
60Hz. Units sold in the US are equipped with US standard plugs. However, the plug
type is changeable, and a kit is available from ILIUM that provides a selection of plug
types for use in other countries.
CAUTION: This meter is intended for use only with the ILIUM Power Supply included
with the meter. Use of an AC Adapter other than the one supplied by ILIUM may
damage the meter.
USB Cable Input
The USB cable connector accommodates a standard USB Type B plug. The USB
connection is used for control of the meter and data acquisition.