IFR AN930, AN930A, AN930A OPT08, AN930A OPT09 technical manual

T.O. 33K3-4-3299-1

TECHNICAL MANUAL

CALIBRATION PROCEDURE

FOR

SPECTRUM ANALYZER

AN930, AN930A, AN930A OPT08/OPT09

(IFR)

This publication replaces T.O. 33K3-4-3299-1 dated 30 June 2011.

Distribution Statement C - Distribution authorized to U. S. Government agencies and their contractors for official use or for administrative or operational purposes only, 28 February 2001. Requests for this document shall be referred to AFMETCAL, 813 Irving-Wick Dr W, Heath, OH 43056-1199.

Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent disclosure of the contents or reconstruction of the document.

Published under Authority of the Secretary of the Air Force

30 DECEMBER 2012

IFR AN930, AN930A, AN930A OPT08, AN930A OPT09 technical manual

T.O. 33K3-4-3299-1

LIST OF EFFECTIVE PAGES

INSERT LATEST CHANGED PAGES. DESTROY SUPERSEDED PAGES.

NOTE:

The portion of the text and illustrations affected by the changes is indicated by a vertical

 

line in the outer margins of the page.

Date of issue for original and changed pages are:

Original......................

0 ..................

30 December 2012

 

 

 

TOTAL NUMBER OF PAGES IN THIS PUBLICATION IS 48, CONSISTING OF THE FOLLOWING:

Page

* Change

Page

* Change

Page

* Change

No.

No.

No.

No.

No.

No.

Title ...................................

0

 

 

 

 

A ........................................

0

 

 

 

 

1 - 44..................................

0

 

 

 

 

A-1.....................................

0

 

 

 

 

A-2 Blank ..........................

0

 

 

 

 

*Zero in this column indicates an original page

A

USAF

T.O. 33K3-4-3299-1

SPECTRUM ANALYZER

AN930, AN930A, AN930A OPT08/OPT09 (IFR)

1 CALIBRATION DESCRIPTION:

Table 1.

Test Instrument (TI)

Performance

Test

Characteristics

Specifications

Method

 

 

 

Frequency Accuracy

 

 

Reference Oscillator

Range: 100 MHz

Verified during Ref

 

Accuracy: (STD)

Out Calibration

 

 

 

Accuracy: 3 X 10-6 in the first year;

*

 

1 X 10-6 every year thereafter;

 

 

Aging/year: 3 X 10-6 in the first year;

 

1 X 10-6 every year thereafter;

 

 

Temperature: <5 X 10-7 **

 

 

Accuracy: (AN930A OPT 09)

 

 

Accuracy: 1 X 10-6 in the first year;

*

 

5 X 10-7 every year thereafter;

 

 

Aging/year: 1 X 10-6 in the first year;

 

5 X 10-7 thereafter;

 

 

Temperature: 2 X 10-8 **

 

Ref Out

Range: Frequency: 100 MHz

Compared to a Frequency

 

 

Standard

 

Accuracy: Same as Reference Oscillator

 

Range: Amplitude: -30 dBm

Measured with Power

 

Accuracy: 0.5 dB

Meter and Power Sensor

 

 

Frequency Readout

Range: 0 Hz to 22 GHz

Compared to a known

Accuracy

 

frequency

 

Accuracy: (3% of Span Width + Reference

 

Oscillator Accuracy + 50% of RBW)

 

Frequency Span

Range: 0 Hz; 10 Hz to 2 GHz/div

 

Width

(2.2 GHz/div for AN930A) in 1-2-5 sequence;

2.2 GHz/div in FULL SPAN;

0 Hz to 22 GHz in Start/Stop Accuracy: 5% of indicated Span

See footnotes at end of Table.

1

T.O. 33K3-4-3299-1

Table 1. (Cont.)

Test Instrument (TI)

Performance

Test

Characteristics

Specifications

Method

 

 

 

Frequency

Range: 0 Hz to 22 GHz

Frequency compared to a

Counter

Accuracy: (Frequency Standard Accuracy +

known frequency.

 

Sensitivity compared to a

 

Counter Resolution);

known signal level

 

Sensitivity: (at 100 MHz with 0 dB attenuation)

 

 

-85 dBm with 3 kHz BW, ( 10 dB gain, AN930);

 

 

-65 dBm with 5 MHz BW, ( 0 dB gain, AN930)

 

Average Noise Level

Range: 9 kHz to 22 GHz

Measured on TI

 

Accuracy: (AN930) (300 Hz RBW, 10 Hz VBW)

 

 

-70 dBm, 9 to 100 kHz;

 

 

-100 dBm, 100 kHz to 10 MHz;

 

 

-120 dBm, 10 MHz to 2.9 GHz;

 

 

-110 dBm, 2.9 to 12 GHz;

 

 

-105 dBm, 12 to 22 GHz

 

 

Accuracy: (AN930A) (300 Hz RBW, 10 Hz VBW)

 

 

-95 dBm, 9 to 100 kHz;

 

 

-115 dBm, 100 kHz to 2.9 GHz;

 

 

-110 dBm, 2.9 to 12 GHz;

 

 

-105 dBm, 12 to 22 GHz

 

 

Accuracy: (AN930A) (3 Hz RBW, VBW N/A)

 

 

-115 dBm, 9 to 100 kHz;

 

 

-135 dBm, 100 kHz to 2.9 GHz;

 

 

-130 dBm, 2.9 to 12 GHz;

 

 

-125 dBm, 12 to 22 GHz

 

Display Linearity

Range: 1, 2, 5 and 10 dB/div; Linear

Compared to a known

 

Accuracy: 5 or 10 dB/div:

signal level

 

 

 

0.15 dB/dB, 1.5 dB over 8 div;

 

 

1 or 2 dB/div:

 

 

0.5 dB over 8 div;

 

 

Linear: 2% of Reference Level, RBW 5 MHz;

 

 

10% of Reference Level with <10 dB IF gain,

 

 

RBW 10 and 30 MHz

 

2

 

 

T.O. 33K3-4-3299-1

 

Table 1. (Cont.)

 

 

 

 

Test Instrument (TI)

Performance

Test

Characteristics

Specifications

Method

 

 

 

IF Gain

Range: 5 MHz RBW:

Compared to a known

 

-10 to +65 dB

signal level

 

Accuracy: 0.25 dB, 0 to +20 dB;

 

 

0.5 dB, +20 to +65 dB

 

 

Range: 10 and 30 MHz RBW:

 

 

0 to +20 dB

 

 

Accuracy: 3 dB nominal for AN930;

 

 

2 dB nominal for AN930A

 

Attenuator

Range: 0 to 60 dB

 

 

Accuracy: (AN930)

 

 

0.5 dB or 2% of dB setting,

 

 

whichever is greater, 0.1 MHz to 3 GHz;

 

 

1 dB or 4% of dB setting, whichever is

 

 

greater, 3 to 18 GHz; 2 dB or 7% of setting,

 

 

whichever is greater, 18 to 22 GHz

 

 

Accuracy: (AN930A)

 

 

0.5 dB or 2% of dB setting,

 

 

whichever is greater, 9 kHz to 2.9 GHz;

 

 

1 dB or 4% of dB setting, whichever is

 

 

greater, 2.9 to 18 GHz; 1.5 dB or 7% of setting,

 

 

whichever is greater, 18 to 22 GHz

 

Frequency Response

Range: 9 kHz to 22 GHz

 

 

Accuracy: 1.5 dB, 9 kHz to 2.9 GHz;

 

 

2.0 dB, 2.9 to 12 GHz;

 

 

3.0 dB, 12 to 18 GHz;

 

 

4.0 dB, 18 to 22 GHz

 

3

T.O. 33K3-4-3299-1

Table 1. (Cont.)

Test Instrument (TI)

Performance

Test

Characteristics

Specifications

Method

 

 

 

Resolution

Range: 3, 10, 30, 100 and 300 Hz, 3, 30

Bandwidth

and 300 kHz, 5 MHz; 10 and 30 MHz,

 

Linear display and FM modes

 

Accuracy: (All specifications are at 3 dB)

 

300 Hz to 30 kHz, 20% of RBW setting;

 

100 kHz (300 kHz for AN930A) to 5 MHz,

 

30% of RBW setting;

 

10 and 30 MHz, 30% of RBW setting

 

Range: (OPT 08) Adds:

 

1 kHz and 1 MHz at 3 dB;

 

200 Hz , 9 and 120 kHz at 6 dB

 

Accuracy:

 

1 kHz, 20% of RBW setting;

 

1 MHz, 30% of RBW setting;

 

200 Hz, 9 and 120 kHz, 10% of RBW setting

Selectivity

Range: -60 dB to -3 dB ratio

 

Accuracy: (AN930)

 

<5:1 at 3 kHz to 5 MHz RBW;

 

<12:1 at 200 Hz to 1 kHz RBW

 

Range: -60 to -3 dB, -60 to -6 dB,

 

-50 to -3 dB and -50 to -6 dB ratios

 

Accuracy: (AN930A and AN930A OPT 08)

 

-60 to -3 dB ratio: <5:1 at 3, 30 and 300 kHz,

 

1 and 5 MHz RBW; <12:1 at 1 kHz RBW;

 

-60 to -6 dB ratio: <5:1 at 9 and 120 kHz;

 

-50 to -3 dB ratio: <10:1 at 300 Hz;

 

-50 to -6 dB ratio: <10:1 at 200 Hz

Resolution

Range: 3, 10, 30, 100 and 300 Hz, 3, 30

Bandwidth

and 300 kHz, 5 MHz (add 10 and 30 MHz for

Switching Error

AN930A)

 

Accuracy: (Referenced to 30 kHz RBW)

 

0.5 dB, 1 kHz to 5 MHz RBW;

 

1.0 dB, <1 kHz RBW;

 

2.0 dB, 10 and 30 MHz

-3 dB points are measured on TI

10 dB/div Bandwidth is measured, then divided by the 1 dB/div Bandwidth

Measured on TI

4

 

 

T.O. 33K3-4-3299-1

 

Table 1. (Cont.)

 

 

 

 

Test Instrument (TI)

Performance

Test

Characteristics

Specifications

Method

 

 

 

Sweep Time

Range: 200 ns to 10 s/div in a 1-2-5 sequence,

Compared to a known

 

Span = 0, resolution reduced <2 s/div

signal

 

Accuracy: 1% of FS; 1% of FS, 2 s/div

 

 

for AN930A

 

Residual Responses

Range: 9 kHz to 22 GHz (Input

Measured on TI

 

terminated with 50 )

 

 

Accuracy: -90 dBm, 9 to 100 kHz;

 

 

-100 dBm, 100 kHz to 22 GHz

 

Scope Mode

 

 

Amplitude

Range: 5 mV to 5 V/div in 1-2-5 sequence

Compared to a known

 

Accuracy: 3% of setting

signal level

 

 

Bandwidth

Range: DC to 5 MHz

Apply a constant amplitude

 

 

signal while changing

 

Accuracy: Down not more than 3 dB

frequency. Vertical

 

at sweeptime 2 s/div

deflection compared to

 

 

deflection at reference

 

 

frequency

*The accuracy is the manufacturers calculated specification after one year. The accuracy specification is found by multiplying the longest term aging rate by the appropriate time interval to obtain one year.

**Typical or operational specification. Not calibrated.

2 EQUIPMENT REQUIREMENTS:

 

 

Minimum Use

Calibration

Sub-

 

Noun

Specifications

Equipment

Item

 

 

 

 

 

2.1

FREQUENCY

Range: 10 MHz

Arbiter

As available

 

STANDARD

Accuracy: 1.25 X 10-7

1083B

 

 

 

 

 

2.2

ELECTRONIC

Range: 100 MHz

Hewlett-Packard

 

 

COUNTER

Accuracy: 1 count of LSD

5345A

 

 

 

 

 

2.3

SYNTHESIZED

Range: 50 MHz to 22 GHz,

Hewlett-Packard

 

 

SWEEPER

-35 to +3 dBm

8340B

 

Accuracy: 1 count of LSD

5

T.O. 33K3-4-3299-1

 

 

Minimum Use

Calibration

Sub-

 

Noun

Specifications

Equipment

Item

 

 

 

 

 

2.4

POWER SENSOR

Range: 100 MHz, -31 to -29 dBm

Hewlett-Packard

 

 

 

Accuracy: 3.5% of charted

8484A

 

 

 

 

 

 

 

value *

 

 

2.5

POWER SENSOR

Range: 50 MHz to 22 GHz,

Hewlett-Packard

 

 

 

-30 to -10 dBm

8485A

 

 

 

Accuracy: (all % are of charted

 

 

 

 

value)

 

 

 

 

3.5%, 50 to 100 MHz;

 

 

 

 

3.3%, 100 MHz to 2 GHz;

 

 

 

 

3.4%, 2 to 8 GHz;

 

 

 

 

3.8%, 8 to 12 GHz;

 

 

 

 

4.1%, 12 to 18 GHz;

 

 

 

 

4.4%, 18 to 22 GHz

 

 

2.6

POWER SPLITTER

Range: 9 kHz to 22 GHz

Hewlett-Packard

 

 

 

Accuracy: Tracking:

11667B

 

 

 

 

 

 

 

0.25 dB, 9 kHz to 18 GHz;

 

 

 

 

0.40 dB, 18 to 22 GHz

 

 

2.7

FREQUENCY

Range: 600 Hz to 14 MHz;

Hewlett-Packard

 

 

SYNTHESIZER

-30 dBm

3325B

 

 

 

Accuracy: 1.25% of frequency

 

 

 

 

setting

 

 

2.8

SYNTHESIZED

Range: -80 to +13 dBm, 50 MHz

Hewlett-Packard

 

 

LEVELED

 

3335A

 

 

GENERATOR

Accuracy:

 

 

 

 

0.025 dB, +3.02 to +13 dBm;

 

 

 

 

0.0375 dB, -6.98 to +3.01 dBm;

 

 

 

 

0.1 dB, -80 to -6.99 dBm

 

 

 

 

Range: 9 kHz to 50 MHz

 

 

 

 

(Flatness)

 

 

 

 

Accuracy: 0.375 dB

 

 

2.9

POWER

Range: -31 to -29 dBm, 100 MHz;

Hewlett-Packard

 

 

METER

-30 to -10 dBm, 50 MHz to 22 GHz

436A

 

Accuracy: 2% of indication *

See footnote at end of Equipment Requirements.

6

 

 

 

T.O. 33K3-4-3299-1

 

 

 

 

 

Minimum Use

Calibration

Sub-

Noun

Specifications

Equipment

Item

 

 

 

 

2.10 TERMINATION

Range: 50

Hewlett-Packard

 

 

Accuracy: N/A

908A

 

 

 

 

2.11 SIGNAL

Range: 500 MHz, -22 dBm

Hewlett-Packard

 

GENERATOR

Accuracy: N/A

8663A

 

 

 

 

2.12 ATTENUATOR

Range: 20 dB at 50 MHz

Hewlett-Packard

 

 

Accuracy: N/A

8491B OPT 020

 

 

 

 

2.13 OSCILLOSCOPE

Range: 20 m to 20 V p-p, 1 kHz

Tektronix

 

CALIBRATOR

squarewave

9500B/3200AF

 

 

Accuracy: 0.18% of applied for

 

 

 

20 m, 0.2, 2 and 20 V p-p;

 

 

 

0.15% of applied for 50 m, 0.1,

 

 

 

0.5, 1, 5 and 10 V p-p

 

 

 

Range: 200 n to 5 s

 

 

 

Accuracy: 0.25% of frequency

 

 

 

setting

 

 

 

Range: 1 kHz to 5 MHz leveled sine wave

 

 

at 4 V p-p

 

 

 

Accuracy: 5% of applied

 

 

2.14 FEEDTHROUGH

Range: 50 , 1 kHz to 5 MHz

Tektronix

 

TERMINATION

Accuracy: 2.0% of nominal

011-0049-01

 

 

 

 

*A TAR of 2.7:1 is achieved by the RSS (Root Sum Square) value of the Power Meter and Power Sensor (2.4) for the TI REF OUT Calibration.

3 PRELIMINARY OPERATIONS:

3.1 Review and become familiar with the entire procedure before beginning the Calibration Procedure.

WARNING

Unless otherwise designated, and prior to beginning the Calibration Process, ensure that all test equipment voltage and/or current outputs are set to zero (0) or turned off, where applicable. Ensure that all equipment switches are set to the proper position before making connections or applying power. If not strictly observed, could result in injury to, or death of, personnel or long term health hazards.

7

T.O. 33K3-4-3299-1

3.2Connect TI to appropriate power source. Set TI MAIN POWER CUTOFF switch (rear panel) to . Set TI POWER APPLIED/ON switch to ON.

3.3TI will initiate a Power-up Self Test. A normal functioning TI will beep twice and display the IFR logo briefly before entering the Center Frequency Operation Screen. If an error message is displayed or the TI fails to beep, this indicates an abnormal condition. If this occurs, set TI POWER APPLIED/ON switch to APPLIED, wait about 30 seconds, then switch to ON. If the error re-occurs, refer to appropriate Commercial Data for corrective action.

3.4Allow the TI a 15 minute warm-up period.

3.5Connect all test equipment to appropriate power source. Set all POWER switches to ON or STBY and allow a warm-up period as required by the manufacturer.

3.6This procedure identifies the Hardkeys in bold and the Softkeys (Menu keys) in Italics. Values shall be entered by use of the DATA key pad. For example: Press TI AMPLITUDE BW, RBW and set to 30 kHz. The AMPLITUDE BW key is a Hardkey, the RBW is a Softkey (Menu key), the 3 and 0 are entries using the DATA key pad, then push the kHz key which is also a part of the DATA keys.

3.7Perform TI Self Test as follows:

3.7.1Press TI MEMORY CONFIG, then DIAG.

3.7.2Using the TI , , and keys and the TI Control Knob, select the first test listed on the TI screen.

3.7.3Toggle TI SELECT until this appropriate test is listed as TEST on the TI screen.

3.7.4Repeat steps 3.7.2 and 3.7.3 for each remaining test listed on the TI screen.

3.7.5Press TI TEST.

3.7.6Once the TI is finished, verify that each test status indicates PASSED on the TI screen.

3.7.7Press TI RETURN, then RETURN again.

3.8 Make copies of Tables 11 through 20 and 23 through 25 for use as Calibration Worksheets.

4 CALIBRATION PROCESS:

NOTE

Unless otherwise specified, verify the results of each test and take corrective action whenever the test requirement is not met, before proceeding.

4.1 REF OUT CALIBRATION:

4.1.1 Connect Frequency Standard 10 MHz REF OUT to the Electronic Counter EXT FREQ STD INPUT (1-10 MHz). Set Electronic Counter INT STD/EXT STD switch to EXT STD.

8

T.O. 33K3-4-3299-1

4.1.2 Connect TI REF OUT to the Electronic Counter CHANNEL A input. Set Electronic Counter 50 /1M switch to 50 .

NOTE

The values in the following step are derived from multiplication of the Aging Rate to determine the offset at one year. Use these calculated one year values regardless of the length of the calibration interval for this TI in

T.O. 33K-1-100-1/2. The longest aging rate specification not to exceed 1 year has been used to calculate the limits.

4.1.3Adjust Electronic Counter controls as required for a stable display indication and then push RESET. Verify Electronic Counter indication is 99 999 700 to 100 000 300 Hz 1 count of LSD if within the first year of service and 99 999 900 to 100 000 100 Hz 1 count of LSD thereafter for the STD TI or 99 999 900 to 100 000 100 Hz

1 count of LSD if within the first year of service 99 999 950 to 100 000 050 Hz 1 count of LSD thereafter for the AN930A OPT 09 TI.

4.1.4Set Electronic Counter INT STD/EXT STD switch to INT STD and disconnect test setup.

4.1.5To ensure reliability of the TI, the following action will be taken: If TI passed the above steps, perform the applicable adjustment steps in Appendix A, and enter NO ADJUSTMENT ACTION into the Maintenance Data Collection System. If TI failed, perform the applicable steps listed in Appendix A and enter appropriate ADJUSTMENT ACTION into the Maintenance Data Collection System.

4.1.6Connect the TI REF OUT to the TI ANALYZER INPUT connector.

4.1.7Set the TI controls as follows:

FREQUENCY CENTER

100 MHz

AMPLITUDE SCALE

1 dB

FREQUENCY SPAN

200 kHz

4.1.8Using the TI Control Knob, adjust the TI gain to place the peak of the signal at a convenient reference line.

4.1.9Disconnect the cable from the TI REF OUT connector and connect to the Synthesized Sweeper RF OUT connector using the same cable.

4.1.10Set the Synthesized Sweeper to 100 MHz and adjust the output controls to place the peak of the signal at the same reference set in step 4.1.8. Do not change the Synthesized Sweeper output controls for the remainder of this para.

4.1.11Disconnect the cable from the Synthesized Sweeper.

4.1.12Standardize the Power Meter and Power Sensor (2.4).

4.1.13Connect the Power Sensor (2.4) to the Synthesized Sweeper.

4.1.14Verify the Power Meter indicates within -30.5 to -29.5 dBm.

4.1.15Set the Synthesized Sweeper for minimum output and disconnect test setup.

9

T.O. 33K3-4-3299-1

4.2 FREQUENCY READOUT ACCURACY CALIBRATION:

4.2.1 Connect equipment as shown in Figure 1.

 

 

TI

 

 

 

10 MHz REF

SYNTHESIZED

 

 

 

EXT REF

OUT

SWEEPER

 

 

 

ANALYZER

 

 

 

RF OUTPUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INPUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 1.

4.2.2Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.2.3Set the TI controls as follows:

MEMORY CONFIG

3

ENTER

9

ENTER

or

ENTER

4.2.4 Set the TI controls as follows:

FREQUENCY CENTER

FREQUENCY SPAN

AMPLITUDE BW

BW

COUPLE

To toggle the reference to EXT

First value listed in the Frequency column of Table 2

First value listed in the Span/Div column of Table 2

First value listed in the Bandwidth column of Table 2

UNCOUPLE (COUPLE not highlighted)

10

T.O. 33K3-4-3299-1

4.2.5 Press the Synthesized Sweeper INSTR PRESET and set the controls as follows:

CW

First value listed in the

 

Frequency column of Table 2

POWER LEVEL

-30 dBm

4.2.6Press the TI MODE MARKER, PEAK*, then <PEAK or PEAK> as necessary to place the active marker at the peak of the displayed signal.

4.2.7Verify the TI M 1 frequency indication is within the values listed in the Limits column of Table 2.

4.2.8Repeat steps 4.2.4 through 4.2.7 for the remaining values listed in Table 2.

Table 2.

Frequency (GHz)

Span/Div (MHz)

Bandwidth (Hz)

Limits (GHz)

 

 

 

 

 

 

1.5

20

5 M

1.491

5 to 1.508 5

1.5

1

300 k

1.499

55 to 1.500 45

1.5

0.5

30 k

1.499

83 to 1.500 17

4

20

5 M

3.991

5 to 4.008 5

4

1

300 k

3.999

55 to 4.000 45

4

0.5

30 k

3.999

83 to 4.000 17

9

20

5 M

8.991

5 to 9.008 5

9

1

300 k

8.999

55 to 9.000 45

9

0.5

30 k

8.999

83 to 9.000 17

16

20

5 M

15.991

5 to 16.008 5

16

1

300 k

15.999

55 to 16.000 45

16

0.5

30 k

15.999

83 to 16.000 17

21

20

5 M

20.991

5 to 21.008 5

21

1

300 k

20.999

55 to 21.000 45

21

0.5

30 k

20.999

83 to 21.000 17

4.2.9 Set the Synthesized Sweeper for minimum output and leave equipment connected.

11

T.O. 33K3-4-3299-1

4.3 FREQUENCY SPAN WIDTH CALIBRATION:

4.3.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.3.2Set the TI controls as follows:

FREQUENCY CENTER

First value listed in the Center

 

Freq column of Table 3

FREQUENCY SPAN

First value listed in the

 

Span/Div column of Table 3

4.3.3Set the Synthesized Sweeper frequency to the first value listed the Frequency 1 column of Table 3 at -30 dBm.

4.3.4Press TI FREQUENCY CENTER, then adjust TI Control Knob so the displayed signal is on the TI 2nd vertical graticule line (1 graticule from left-most).

4.3.5Set the TI controls as follows:

MODE MARKER

 

PEAK*

 

<PEAK or PEAK>

As necessary to place the active

 

marker at the peak of the displayed

 

signal.

4.3.6Set the Synthesized Sweeper frequency to the first value listed the Frequency 2 column of Table 3.

4.3.7Press TI RETURN, MKR 2, and use the TI Control Knob to superimpose the active marker over the displayed signal.

4.3.8Press TI DELTA and verify the TI marker frequency indication is within the values listed in the Limits column of Table 3.

Table 3.

Center

 

 

 

 

Freq (Hz)

Frequency 1 (Hz)

Frequency 2 (Hz)

Span/Div (Hz)

Limits (Hz)

 

 

 

 

 

10 G

2 G

18 G

2 G

15.0 to 17.0 G

 

6 G

14 G

1 G

7.5 to 8.5 G

 

8 G

12 G

500 M

3.75 to 4.25 G

 

9.2 G

10.8 G

200 M

1.5 to 1.7 G

12

 

 

 

 

T.O. 33K3-4-3299-1

 

 

Table 3. (Cont.)

 

 

 

 

 

 

 

Center

 

 

 

 

Freq (Hz)

Frequency 1 (Hz)

Frequency 2 (Hz)

Span/Div (Hz)

Limits (Hz)

 

 

 

 

 

1 G

600 M

1.4 G

100 M

750 to 850 M

 

800 M

1.2 G

50 M

375 to 425 M

 

920 M

1.08 G

20 M

150 to 170 M

100 M

60 M

140 M

10 M

75 to 85 M

 

80 M

120 M

5 M

37.5 to 42.5 M

 

92 M

108 M

2 M

15.0 to 17.0 M

10 M *

6 M

14 M

1 M

7.5 to 8.5 M

 

8 M

12 M

500 k

3.75 to 4.25 M

 

9.2 M

10.8 M

200 k

1.5 to 1.7 M

1 M

600 k

1.4 M

100 k

750 to 850 k

 

800 k

1.2 M

50 k

375 to 425 k

 

920 k

1.08 M

20 k

150 to 170 k

100 k

60 k

140 k

10 k

75 to 85 k

 

80 k

120 k

5 k

37.5 to 42.5 k

 

92 k

108 k

2 k

15 to 17 k

10 k

6 k

14 k

1 k

7.5 to 8.5 k

 

8 k

12 k

500

3.75 to 4.25 k

 

9.2 k

10.8 k

200

1.5 to 1.7 k

1 k

600

1.4 k

100

750 to 850

 

800

1.2 k

50

375 to 425

 

920

1.08 k

20

150 to 170

 

960

1.04 k

10

75 to 85

*Set the Synthesized Sweeper for minimum output. Disconnect the Synthesized Sweeper from the TI and replace with the Frequency Synthesizer with the timebases connected as in Figure 1.

13

T.O. 33K3-4-3299-1

4.3.9Press TI MKR 1 and MKR 2 as necessary to turn off the TI markers.

4.3.10Repeat steps 4.3.2 through 4.3.8 for the remaining values listed in Table 3.

4.3.11Set the Frequency Synthesizer for minimum output and disconnect test setup.

4.4 FREQUENCY COUNTER CALIBRATION:

4.4.1Connect equipment as shown in Figure 1.

4.4.2Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.4.3Set the TI controls as follows:

FREQUENCY SPAN

ZERO

AMPLITUDE ATTEN

0 dB, INT

4.4.4Set the Synthesized Sweeper for 100 MHz at -30 dBm.

4.4.5Press TI MODE AUX.

4.4.6Press TI CTR*.

4.4.7Set the TI counter resolution to the first value listed in the Resolution column of Table 4.

4.4.8Verify the TI CTR indication is within the first values listed in the Limits column of Table 4.

Table 4.

Resolution (Hz)

Limits (MHz)

 

 

 

1 k

99.999 to 100.001

100

99.999

9 to 100.000 1

10

99.999

99 to 100.000 01

1

99.999

999 to 100.000 001

4.4.9Repeat steps 4.4.7 and 4.4.8 for the remaining values listed in Table 4.

4.4.10Press TI AMPLITUDE BW, BW, then 5 MHz.

4.4.11Decrease the Synthesized Sweeper output controls until the TI CTR indication changes either -5 or +5 Hz.

4.4.12Verify the Synthesized Sweeper indicates -65 dBm.

4.4.13Press TI AMPLITUDE BW, BW, then 3 kHz.

4.4.14Press TI AMPLITUDE GAIN, then 10 dB.

14

T.O. 33K3-4-3299-1

4.4.15Decrease the Synthesized Sweeper output controls until the TI CTR indication changes either -5 or +5 Hz.

4.4.16Verify the Synthesized Sweeper indicates -85 dBm.

4.4.17Set the Synthesized Sweeper for minimum output and disconnect test setup.

4.4.18Set the TI controls as follows:

AUX MODE

CTR*

1 Hz

MEMORY CONFIG

3

ENTER

9

ENTER

or

ENTER

RETURN

MEMORY CONFIG 4.5 AVERAGE NOISE LEVEL CALIBRATION:

4.5.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.5.2Connect the Termination to TI ANALYZER INPUT connector.

4.5.3Set the TI controls as follows:

FREQUENCY CENTER

FREQUENCY SPAN

AMPLITUDE GAIN

AMPLITUDE BW

COUPLE

VIDEO

To toggle the reference to INT

First value listed in the Applied column of Table 5

1 kHz

50 dB

300 Hz

UNCOUPLE (COUPLE not highlighted)

10 Hz

15

T.O. 33K3-4-3299-1

4.5.4Press the TI MODE MARKER, then MKR 1. Using the DATA ENTRY keys, set the TI marker 1 to the first value listed in the Applied column of Table 5.

4.5.5Verify the TI MRK 1 amplitude is within the value listed in the Limits column of Table 5.

4.5.6Press TI FREQUENCY CENTER and set to the next value listed in the Applied column of Table 5.

4.5.7Repeat steps 4.5.4 through 4.5.6 for the remaining values listed in Table 5.

Table 5.

 

 

 

Limits (dBm)

Applied (Hz)

AN930

AN930A

 

 

 

 

9 k

 

-70

-95

60 k

-70

-95

100 k

-70

-95

101 k

-100

-115

9.9

M

-100

-115

10 M

-120

-115

11 M

-120

-115

2.9

G

-120

-115

3.0

G

-110

-110

7.5

G

-110

-110

12 G

-110

-110

13 G

-105

-105

17 G

-105

-105

22 G

-105

-105

4.5.8For TI AN930A, proceed to step 4.5.9. Otherwise, proceed to step 4.5.10

4.5.9Set the TI FREQUENCY SPAN to 20 Hz and AMPLITUDE BW to 3 Hz.

4.5.10Set the TI FREQUENCY CENTER to the first value listed in the Applied column of Table 6.

4.5.11Repeat steps 4.5.4 through 4.5.7 using Table 6.

16

T.O. 33K3-4-3299-1

 

Table 6.

 

 

Applied (Hz)

Limits (dBm)

 

 

9 k

-115

60 k

-115

100 k

-115

101 k

-135

9.9 M

-135

10 M

-135

11 M

-135

2.9 G

-135

3.0 G

-130

7.5 G

-130

12 G

-130

13 G

-125

17 G

-125

22 G

-125

4.5.12 Disconnect test setup.

4.6 DISPLAY LINEARITY CALIBRATION:

4.6.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.6.2Connect the Synthesized Leveled Generator OUTPUT 50 Ω to the TI ANALYZER INPUT connector though the Attenuator.

4.6.3Set the TI controls as follows:

FREQUENCY CENTER

50 MHz

FREQUENCY SPAN

1 MHz

AMPLITUDE SCALE

5 dB

AMPLITUDE BW

300 kHz

COUPLE

UNCOUPLE (COUPLE not

 

highlighted)

17

T.O. 33K3-4-3299-1

VIDEO

100 Hz

AMPLITUDE GAIN

0 dB, INT

AMPLITUDE ATTEN

10 dB, INT

4.6.4Set the Synthesized Leveled Generator to 50 MHz at 0 dBm.

4.6.5Set the TI controls as follows:

MODE MARKER

MKR 1

4.6.6Press TI AMPLITUDE GAIN and adjust the TI Control Knob for a reference at the top graticule line of the TI screen.

4.6.7Adjust the TI Control Knob for maximum amplitude as indicated by the TI M 1 amplitude readout.

4.6.8Press TI MODE MARKER, REF*, then +REF.

4.6.9Press the Synthesized Leveled Generator AMPL INCR key and set to 1 dB. Press the AMPLITUDE key, then

press the step key to set the Synthesized Leveled Generator amplitude to the next value listed in Applied column of Table 7.

4.6.10Verify the TI M 1 marker amplitude indicates within the values listed in the Limits column of Table 7.

 

Table 7.

 

 

Applied (dB)

Limits (dB)

0

Reference

-2

-2.3 to -1.7

-4

-4.6 to -3.4

-6

-6.9 to -5.1

-8

-9.2 to -6.8

-10

-11.5 to -8.5

-11

-12.5 to -9.5

-12

-13.5 to -10.5

-13

-14.5 to -11.5

-14

-15.5 to -12.5

-15

-16.5 to -13.5

18

T.O. 33K3-4-3299-1

Table 7.

(Cont.)

 

 

Applied (dB)

Limits (dB)

 

 

-16

-17.5 to -14.5

-17

-18.5 to -15.5

-18

-19.5 to -16.5

-19

-20.5 to -17.5

-20

-21.5 to -18.5

-21

-22.5 to -19.5

-22

-23.5 to -20.5

-23

-24.5 to -21.5

-24

-25.5 to -22.5

-25

-26.5 to -23.5

-26

-27.5 to -24.5

-27

-28.5 to -25.5

-28

-29.5 to -26.5

-29

-30.5 to -27.5

-30

-31.5 to -28.5

-31

-32.5 to -29.5

-32

-33.5 to -30.5

-33

-34.5 to -31.5

-34

-35.5 to -32.5

-35

-36.5 to -33.5

-36

-37.5 to -34.5

-37

-38.5 to -35.5

-38

-39.5 to -36.5

-39

-40.5 to -37.5

-40

-41.5 to -38.5

19

T.O. 33K3-4-3299-1

4.6.11Repeat steps 4.6.9 and 4.6.10 for the remaining values listed in Table 7.

4.6.12Press TI MODE MARKER, then MKR 1 to turn off the TI markers.

4.6.13Press TI AMPLITUDE SCALE, then 1 dB.

4.6.14Repeat steps 4.6.4 through 4.6.8.

4.6.15Press the Synthesized Leveled Generator AMPL INCR key and set to 1 dB. Press the AMPLITUDE key,

then press the step key to set the Synthesized Leveled Generator amplitude to the next value listed in Applied column of Table 8.

4.6.16Verify the TI M 1 marker amplitude indicates within the values listed in the Limits column of Table 8.

 

Table 8.

 

 

Applied (dB)

Limits (dB)

0

Reference

-1

-1.5 to -0.5

-2

-2.5 to -1.5

-3

-3.5 to -2.5

-4

-4.5 to -3.5

-5

-5.5 to -4.5

-6

-6.5 to -5.5

-7

-7.5 to -6.5

-8

-8.5 to -7.5

4.6.17Repeat steps 4.6.15 and 4.6.16 for the remaining values listed in Table 8.

4.6.18Set the TI controls as follows:

AMPLITUDE SCALE

 

LIN

 

UNITS *

 

mV

 

RETURN

 

AMPLITUDE GAIN

-0.1 dB, INT

AMPLITUDE ATTEN

10 dB, INT

20

T.O. 33K3-4-3299-1

4.6.19Set the Synthesized Leveled Generator to 50 MHz at 0 dBm.

4.6.20Press TI MODE MARKER, then MKR 1. Adjust the TI Control Knob for maximum amplitude as indicated by the TI M 1 amplitude readout.

4.6.21Set the Synthesized Leveled Generator amplitude controls for a TI M 1 indication as close as to possible 22.4 mV.

4.6.22Press the Synthesized Leveled Generator AMPL INCR key and set to 6 dB. Press the AMPLITUDE key,

then press the step key to set the Synthesized Leveled Generator amplitude to the next value listed in Applied column of Table 9.

4.6.23Verify the TI M 1 amplitude indicates within the values listed in the Limits column of Table 9.

 

Table 9.

 

 

Applied (Nominal dBm)

Limits (mV)

 

 

0

22.4 (ref)

-6

10.8 to 11.7

-12

5.2 to 6.1

-18

2.4 to 3.3

4.6.24Repeat step 4.6.22 and 4.6.23 for the remaining values listed in Table 9.

4.6.25Set the TI controls as follows:

 

FREQUENCY SPAN

10 MHz

 

AMPLITUDE BW

10 MHz

 

AMPLITUDE GAIN

9.9 dB, INT

 

AMPLITUDE ATTEN

20 dB, INT

4.6.26

Repeat steps 4.6.19 through 4.6.24 utilizing Table 10.

 

 

Table 10.

 

 

 

 

 

Applied (Nominal dBm)

Limits (mV)

 

 

 

 

0

22.4 (ref)

 

-6

9.0 to 13.4

 

-12

3.4 to 7.9

 

-18

0.6 to 5.1

4.6.27

Set the Synthesized Leveled Generator for minimum output.

 

21

T.O. 33K3-4-3299-1

4.7 IF GAIN CALIBRATION:

4.7.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.7.2Set the TI controls as follows:

FREQUENCY CENTER

50 MHz

FREQUENCY SPAN

1 kHz

AMPLITUDE BW

1 kHz

COUPLE

UNCOUPLE (COUPLE not

 

highlighted)

VIDEO

100 Hz

AMPLITUDE SCALE

1 dB

AMPLITUDE GAIN

-10 dB, INT

4.7.3Set the Synthesized Leveled Generator to 50 MHz at -1.0 dBm.

4.7.4Set the TI controls as follows:

MODE MARKER

MKR 1

4.7.5Adjust the TI Control Knob for maximum amplitude as indicated by the TI M 1 amplitude indication. Record the TI M 1 indication in the TI M 1 column of Table 11 as ref.

4.7.6Press the Synthesized Leveled Generator AMPL INCR key and set to 10 dB.

4.7.7Press the Synthesized Leveled Generator AMPLITUDE key, then press the step key to set the Synthesized Leveled Generator amplitude to the next nominal value listed in Applied column of Table 11.

4.7.8Press TI AMPLITUDE GAIN and set to the first value listed on the TI Gain column of Table 11.

4.7.9Press TI MODE MARKER, PEAK *, then <PEAK or PEAK> as necessary to place the TI M 1 marker to the peak of the displayed signal. Record the TI M 1 amplitude indication in the TI M 1 column of Table 11 for the appropriate applied value.

Table 11.

Applied (dBm)

TI Gain (dB)

TI M 1 (dB)

Limits (dB)

 

 

 

 

-1

-10

_________ (ref)

Reference

-11

0

_________

-0.25 to +0.25

-21

+10

_________

-0.25 to +0.25

22

 

 

 

T.O. 33K3-4-3299-1

 

Table 11.

(Cont.)

 

 

 

 

 

Applied (dBm)

TI Gain (dB)

TI M 1 (dB)

Limits (dB)

 

 

 

 

-31

+20

_________

-0.25 to +0.25

-41

+30

_________

-0.50 to +0.50

-51

+40

_________

-0.50 to +0.50

-61

+50

_________

-0.50 to +0.50

-71

+60

_________

-0.50 to +0.50

-76 *

+65

_________

-0.50 to +0.50

* This setting must be manually entered into the Synthesized Leveled Generator.

4.7.10Repeat steps 4.7.7 through 4.7.9 for the remaining values listed in Table 11.

4.7.11From the recorded values in Table 11, calculate the TI IF Gain as follows:

IF Gain = (Current Value - Previous Value) + 10 dB (for last step,

IF Gain = (Current Value - Previous Value) + 5 dB)

4.7.12Verify the results of step 4.7.11 are within the values listed in the Limits column of Table 11 for the appropriate TI Gain setting.

4.7.13Set the TI controls as follows:

FREQUENCY SPAN

10 MHz

AMPLITUDE BW

10 MHz

MODE MARKER

MKR 1, as necessary to turn

 

off TI markers

AMPLITUDE GAIN

0 dB, INT

AMPLITUDE SCALE

 

LIN

 

UNITS *

 

dB

 

4.7.14Set the Synthesized Leveled Generator to 50 MHz at -13.0 dBm.

4.7.15Repeat steps 4.7.4 through 4.7.12 utilizing Table 12.

23

T.O. 33K3-4-3299-1

Table 12.

Applied

 

 

 

Limits (dB)

(dBm)

TI Gain (dB)

TI M 1 (dB)

AN930

AN930A

 

 

 

 

 

 

-13

0

 

(ref)

Reference

Reference

-23

+10

________

 

-3.0 to +3.0

-2.0 to +2.0

-33

+20

________

 

-3.0 to +3.0

-2.0 to +2.0

4.7.16 Set the Synthesized Leveled Generator for minimum output.

Disconnect the test setup.

4.8 ATTENUATOR CALIBRATION:

4.8.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.8.2Connect the Synthesized Leveled Generator OUTPUT 50 Ω to the TI ANALYZER INPUT connector. Set the Synthesized Leveled Generator to -60 dBm at 50 MHz.

4.8.3Set the TI controls as follows:

FREQUENCY CENTER

50 MHz

FREQUENCY SPAN

20 kHz

AMPLITUDE SCALE

1 dB

AMPLITUDE ATTEN

0 dB, INT

AMPLITUDE GAIN

28 dB, INT

MODE MARKER

 

MKR 1

 

4.8.4Adjust the TI Control Knob for maximum amplitude as indicated by the TI M 1 amplitude indication. Record the TI M 1 indication in the TI M 1 column of Table 13 as ref.

4.8.5Press the Synthesized Leveled Generator AMPL INCR key and set to 10 dB.

4.8.6Press TI AMPLITUDE ATTEN and set to the first value listed on the TI Attenuator column of Table 13.

4.8.7Press the Synthesized Leveled Generator AMPLITUDE key, then press the step key to set the Synthesized Leveled Generator amplitude to the next nominal value listed in Applied column of Table 13.

4.8.8Press TI MODE MARKER, PEAK *, then <PEAK or PEAK> as necessary to place the TI M 1 marker to the peak of the displayed signal. Record the TI M 1 amplitude indication in the TI M 1 column of Table 13 for the appropriate applied value.

24

 

 

 

T.O. 33K3-4-3299-1

 

 

Table 13.

 

 

 

 

 

TI Attenuator (dB)

Applied (dBm)

TI M 1 (dB)

Limits (dB)

 

 

 

 

0

-60

________ (ref)

Reference

10

-50

________

-0.5 to +0.5

20

-40

________

-0.5 to +0.5

30

-30

________

-0.6 to +0.6

40

-20

________

-0.8 to +0.8

50

-10

________

-1.0 to +1.0

60

0

________

-1.6 to +1.6

4.8.9Repeat steps 4.8.6 through 4.8.8 for the remaining values listed in Table 13.

4.8.10From the values recorded in Table 13, calculate the TI Attenuator Error as follows:

Attenuator Error = (Current Value - Previous Value) - 10 dB

4.8.11Verify the results of step 4.8.10 are within the values listed in the Limits column of Table 13 for the appropriate TI Attenuator setting.

4.8.12Set the Synthesized Level Generator for minimum output and disconnect test setup.

4.9 FREQUENCY RESPONSE CALIBRATION:

4.9.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.9.2Standardize and zero the Power Meter and the Power Sensor (2.5).

4.9.3Connect the equipment as shown in Figure 2 with the Power Sensor (2.5) connected to the Power Splitter.

4.9.4Set the Synthesized Leveled Generator controls as follows:

FREQUENCY

50 MHz

AMPLITUDE

-18.0 dBm

25

T.O. 33K3-4-3299-1

 

 

SYNTHESIZED

 

 

POWER METER

LEVEL

 

TI

GENERATOR

 

SENSOR

 

OUTPUT 50

 

ANALYZER INPUT

 

 

 

 

 

 

 

 

POWER

SPLITTER

POWER SENSOR

(2.5)

TERMINATION

Figure 2.

NOTE

Throughout this para, ensure the Power Meter Cal Factor is set to the appropriate value as charted on the Power Sensor (2.5).

4.9.5 Set the TI controls as follows:

 

FREQUENCY CENTER

50 MHz

FREQUENCY SPAN

20 kHz

AMPLITUDE SCALE

1 dB

AMPLITUDE ATTEN

10 dB, INT

AMPLITUDE GAIN

0 dB, INT

MODE MARKER

 

MKR 1

 

4.9.6Adjust the TI Control Knob for maximum amplitude as indicated by the TI M 1 amplitude readout.

4.9.7Set Synthesized Leveled Generator amplitude controls for a TI M 1 amplitude indication of

-24 dBm ±0.1 dB.

4.9.8Set the TI REF level to -20 dBm.

4.9.9Press the Power Meter MODE dB [REF] key.

4.9.10Disconnect the Power Sensor (2.5) from the Power Splitter and connect the Termination to the unused port of the Power Splitter.

4.9.11Set the Synthesized Leveled Generator frequency to the first value listed in the Applied column of

Table 14.

26

T.O. 33K3-4-3299-1

4.9.12Press TI FREQUENCY CENTER and set to the first value listed in the Applied column of Table 14.

4.9.13Press TI MODE MARKER, then using the TI Control Knob, set the TI M 1 cursor on top of the displayed

signal.

4.9.14Algebraically subtract -24 dB to the TI M 1 amplitude indication and record results in a copy of Table 14.

4.9.15Repeat steps 4.9.11 through 4.9.14 for the remaining values listed in Table 14.

 

Table 14.

 

 

Applied (Hz)

TI M 1 (dB)

 

 

9 k

______

20 k

______

50 k

______

75 k

______

100 k

______

1 M

______

5 M

______

10 M

______

20 M

______

4.9.16Set the Synthesized Leveled Generator for minimum output and disconnect Synthesized Leveled Generator from test setup.

4.9.17Connect equipment as shown in Figure 2 with Power Sensor (2.5) connected to Power Splitter, except replace the Synthesized Leveled Generator with the Synthesized Sweeper.

4.9.18Set the Synthesized Sweeper as follows:

CW

50 MHz

POWER LEVEL

-18.0 dBm

4.9.19 Set the Synthesized Sweeper amplitude controls as close as possible for a 0.0 dB indication on the Power Meter.

27

T.O. 33K3-4-3299-1

4.9.20Set the Synthesized Sweeper frequency to the first value listed in the Applied column of Table 15.

4.9.21Press the TI FREQUENCY CENTER and set to the first value listed in the Applied column of Table 15.

4.9.22Press TI MODE MARKER, then using the TI Control Knob, set the TI MKR 1 cursor on top of the displayed signal.

4.9.23Adjust Synthesized Sweeper amplitude controls for a TI M 1 amplitude indication of -24 dBm ±0.1 dB.

4.9.24Record the Power Meter indication as the opposite polarity value in the Power Meter column in a copy of Table 15.

4.9.25Repeat steps 4.9.20 through 4.9.24 for the remaining applied values listed in Table 15.

 

Table 15.

 

 

Applied (MHz)

Power Meter (dB)

 

 

75

_______

175

_______

275

_______

375

_______

475

_______

575

_______

675

_______

775

_______

875

_______

975

_______

1075

_______

1175

_______

1275

_______

1375

_______

1500

_______

1525

_______

1625

_______

1725

_______

28

 

T.O. 33K3-4-3299-1

Table 15.

(Cont.)

 

 

Applied (MHz)

Power Meter (dB)

 

 

1925

______

2125

______

2325

______

2525

______

2725

______

2900

______

4.9.26Set the Synthesized Sweeper frequency to the first value listed in the Applied column of Table 16.

4.9.27Press the TI FREQUENCY CENTER and set to the first value listed in the Applied column of Table 16.

4.9.28Press TI MODE MARKER, then using the TI Control Knob, set the TI MKR 1 cursor on top of the displayed signal.

4.9.29Press TI AMPLITUDE GAIN, then PS AUTO.

4.9.30Adjust Synthesized Sweeper amplitude controls for a TI Marker amplitude indication of -24 dBm ±0.1 dB.

4.9.31Record the Power Meter indication as the opposite polarity value in the Power Meter column in a copy of Table 16.

4.9.32Repeat steps 4.9.26 through 4.9.31 for the remaining applied values listed in Table 16.

 

Table 16.

 

 

Applied (GHz)

Power Meter (dB)

 

 

3.05

______

3.1

______

3.3

______

3.5

______

3.7

______

3.9

______

4.1

______

29

T.O. 33K3-4-3299-1

Table 16.

(Cont.)

 

 

Applied (GHz)

Power Meter (dB)

 

 

4.3

_______

4.5

_______

4.7

_______

4.9

_______

5.1

_______

5.3

_______

5.5

_______

5.7

_______

5.9

_______

6.1

_______

6.3

_______

6.5

_______

6.7

_______

6.75

_______

6.8

_______

7.0

_______

7.4

_______

7.8

_______

8.2

_______

8.6

_______

9.0

_______

9.4

_______

9.8

_______

10.2

_______

11.0

_______

30

 

T.O. 33K3-4-3299-1

Table 16.

(Cont.)

 

 

Applied (GHz)

Power Meter (dB)

 

 

11.4

______

11.8

______

12.0

______

4.9.33 Repeat steps 4.9.26 through 4.9.31 for the remaining applied values listed in Table 17.

 

Table 17.

 

 

Applied (GHz)

Power Meter (dB)

 

 

12.6

______

12.8

______

13.2

______

13.25

______

13.3

______

13.7

______

14.1

______

14.5

______

14.9

______

15.3

______

15.7

______

16.1

______

16.5

______

16.9

______

17.3

______

17.7

______

18.0

______

31

T.O. 33K3-4-3299-1

4.9.34 Repeat steps 4.9.26 through 4.9.31 for the remaining applied values listed in Table 18.

 

Table 18.

 

 

Applied (GHz)

Power Meter (dB)

 

 

18.5

_______

18.9

_______

19.3

_______

19.7

_______

20.1

_______

20.5

_______

20.9

_______

21.3

_______

21.7

_______

22.0

_______

4.9.35Set the Synthesized Sweeper for minimum output and disconnect test setup.

4.9.36Verify the TI Frequency Response as follows:

4.9.36.1From the values recorded in Tables 14 and 15, algebraically subtract the most negative from the most positive. Verify the result is 3.0 dB.

4.9.36.2From the values recorded in Table 16, algebraically subtract the most negative from the most positive. Verify the result is 4.0 dB.

4.9.36.3From the values recorded in Table 17, algebraically subtract the most negative from the most positive. Verify the result is 6.0 dB.

4.9.36.4From the values recorded in Table 18, algebraically subtract the most negative from the most positive. Verify the result is 8.0 dB.

4.10 RESOLUTION BANDWIDTH AND SELECTIVITY CALIBRATION:

4.10.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.10.2Connect equipment as shown in Figure 3.

32

T.O. 33K3-4-3299-1

 

 

 

 

 

 

 

 

SYNTHESIZED

 

 

TI

 

 

 

REF

LEVELED

 

 

 

EXT REF

OUT

GENERATOR

 

 

 

ANALYZER

 

 

 

50 OUTPUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INPUT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 3.

4.10.3 Set the TI controls as follows:

MEMORY CONFIG

3

ENTER

9

ENTER

or

RETURN

MEMORY CONFIG

FREQUENCY CENTER

AMPLITUDE SCALE

AMPLITUDE GAIN

4.10.4Set the Synthesized Leveled Generator for 50 MHz at -20 dBm.

4.10.5Set the TI controls as follows:

FREQUENCY SPAN

AMPLITUDE BW

COUPLE

VIDEO

to toggle the reference to EXT

50 MHz

1 dB

-10 dB, INT

First value listed in the TI Span/Div column of Table 19

First value listed in the TI BW column of Table 19

UNCOUPLE (COUPLE not highlighted)

100 Hz

4.10.6 Adjust the Synthesized Leveled Generator amplitude controls to place the displayed signal on the TI top graticule line.

33

T.O. 33K3-4-3299-1

4.10.7Press TI MODE MARKER and using the TI Control Knob, adjust for a TI M 1 indication of -23.00 0.1 dB

if measuring a 3 dB RBW or -26.00 0.1 dB if measuring a 6 dB RBW. The marker should be on the left-hand skirt of the signal.

4.10.8Press TI MKR 2 and using the TI Control Knob, adjust the active cursor past -20.00 dB and back down

to -23.00 0.1 dB if measuring a 3 dB RBW or -26.00 0.1 dB if measuring a 6 dB RBW. The marker should be on the right-hand skirt of the signal.

4.10.9Press TI DELTA.

4.10.10Verify the TI frequency indication is within the values listed in the Limits column of Table 19.Record the value in the appropriate Actual 3 dB BW or Actual 6 dB BW column of Table 19.

4.10.11Press TI MKR 1 and MKR 2 as necessary to turn off the TI markers.

Table 19.

TI

 

 

 

 

Span/Div

TI BW

Actual 3 dB

Actual 6 dB

 

(Hz)

(Hz)

BW (Hz)

BW (Hz)

Limits (Hz)

 

 

 

 

 

2 M

5 M

________

N/A

3.5 to 6.5 M

500 k

1 M *

________

N/A

700 k to 1.3 M

100 k

300 k

________

N/A

210 to 390 k

50 k

120 k *

N/A

________

108 to 132 k

10 k

30 k

________

N/A

24 to 36 k

5 k

9 k *

N/A

________

8.1 to 9.9 k

1 k

3 k

________

N/A

2.4 to 3.6 k

500

1 k *

________

N/A

0.8 to 1.2 k

300

300

________

N/A

240 to 360

201

200 *

N/A

________

180 to 220

* For AN930A OPT 08 Only.

4.10.12Repeat steps 4.10.5 through 4.10.11, as required, for the remaining values listed in Table 19.

4.10.13Set the TI AMPLITUDE SCALE to 10 dB.

4.10.14Set the TI AMPLITUDE BW to the first value listed in the TI BW column of Table 20.

4.10.15Set the TI FREQUENCY SPAN as necessary to view several div of width of the displayed signal at the 60 dB or 50 dB point. Ensure the TI UNCAL light is not lit.

4.10.16Adjust the Synthesized Leveled Generator amplitude controls to place the displayed signal on the TI top graticule line.

34

T.O. 33K3-4-3299-1

4.10.17Using the TI Control Knob, adjust for a TI M 1 indication of -80.00 0.1 dB if measuring a 60 dB point or -70.00 0.1 dB if measuring a 50 dB point. The marker should be on the left-hand skirt of the signal.

4.10.18Press TI MKR 2 and using the TI Control Knob, adjust the active cursor past -20.00 dB and back down

to -80.00 0.1 dB if measuring a 60 dB point or -70.00 0.1 dB if measuring a 50 dB point. The marker should be on the right-hand skirt of the signal

4.10.19Press TI DELTA.

4.10.20Record the TI frequency value in the appropriate Actual 60 dB BW or Actual 50 dB BW column of Table 20.

4.10.21Press TI MKR 1 and MKR 2 as necessary to turn off the TI markers.

Table 20.

TI BW (Hz)

Actual 60 dB BW (Hz)

Actual 50 dB BW (Hz)

 

 

 

 

5 M

______

 

N/A

1 M *

______

 

N/A

300 k

______

 

N/A

120 k *

 

 

N/A

30 k

______

 

N/A

9 k *

 

 

N/A

3 k

______

 

N/A

1 k *

______

 

N/A

300

______ (AN930)

______ (AN930A)

200 *

N/A

______

* For AN930A OPT 08 Only.

4.10.22Repeat steps 4.10.14 through 4.10.21 for the remaining values listed in Table 20.

4.10.23Divide each Actual 60 dB or 50 dB BW value recorded in Table 20 by the appropriate Actual 3 dB or Actual 6 dB BW value recorded in Table 19. Verify the resulting is within the value listed in the Limits column of Table 21 for the TI being certified.

35

T.O. 33K3-4-3299-1

Table 21.

TI

TI BW (Hz)

Ratio (dB)

Limits

 

 

 

 

AN930

5 M

60:3

<12:1

 

300 k

60:3

<12:1

 

30 k

60:3

<12:1

 

3 k

60:3

<12:1

 

300

60:3

<12:1

AN930A

5 M

60:3

<12:1

 

300 k

60:3

<12:1

 

30 k

60:3

<12:1

 

3 k

60:3

<12:1

 

300

50:3

<10:1

AN930A OPT 08

5 M

60:3

<12:1

 

1 M

60:3

<12:1

 

300 k

60:3

<12:1

 

120 k

60:6

<5:1

 

30 k

60:3

<12:1

 

9 k

60:6

<5:1

 

3 k

60:3

<12:1

 

1 k

60:3

<12:1

 

300

50:3

<10:1

 

200

50:6

<10:1

4.10.24 For TI AN930, proceed to step 10.25. Otherwise, proceed to step 4.10.34.

36

 

 

T.O. 33K3-4-3299-1

4.10.25

Set the TI controls as follows:

 

 

AMPLITUDE SCALE

 

 

LIN

 

 

UNITS *

 

 

dB

 

 

AMPLITUDE ATTEN

10 dB, INT

4.10.26

Set the TI controls as follows:

 

 

FREQUENCY SPAN

First value listed in the TI

 

 

Span/Div column of Table 22

 

AMPLITUDE BW

First value listed in the TI BW

 

 

column of Table 22

4.10.27Adjust the Synthesized Leveled Generator amplitude controls to place the displayed signal on the TI top graticule line.

4.10.28Press TI MODE MARKER, then MKR 1 and using the TI Control Knob, adjust for a TI M 1 indication of -23.00 0.1 dB. The marker should be on the left-hand skirt of the signal.

4.10.29Press TI MKR 2 and using the TI Control Knob, adjust the active cursor past -20.00 dB and back down

to -23.00 0.1 dB. The marker should be on the right-hand skirt of the signal.

4.10.30Press TI DELTA.

4.10.31Verify the TI frequency indication is within the values listed in the Limits column of Table 22.

4.10.32Press TI MKR 1 and MKR 2 as necessary to turn off the TI markers.

Table 22.

TI Span/Div (MHz)

TI BW (MHz)

Limits (MHz)

 

 

 

2

10

7 to 13

5

30

21 to 39

4.10.33Repeat steps 4.10.26 through 4.10.32 for the remaining values listed in Table 22.

4.10.34Set the Synthesized Leveled Generator for minimum output.

4.11 RESOLUTION BANDWIDTH SWITCHING ERROR CALIBRATION:

4.11.1 Press TI MEMORY SETUP, then INIT to restore TI initial settings.

37

T.O. 33K3-4-3299-1

4.11.2 Set the TI controls as follows:

FREQUENCY CENTER

50 MHz

AMPLITUDE SCALE

LIN UNITS * dB

4.11.3Set the Synthesized Leveled Generator for 50 MHz at -30 dBm.

4.11.4Set the TI controls as follows:

AMPLITUDE BW

First value listed in the TI BW

 

column of Table 23

COUPLE

UNCOUPLE (COUPLE not

 

highlighted)

VIDEO

100 Hz

FREQUENCY SPAN

First value listed in the TI

 

Span/Div column of Table 23

4.11.5Adjust the Synthesized Leveled Generator amplitude controls to place the displayed signal 2 division below the TI top graticule line.

4.11.6Press the TI MODE MARKER, PEAK *, then <PEAK or PEAK> as necessary to place the TI M 1 marker to the peak of the displayed signal.

4.11.7Record the TI M 1 amplitude indication.

4.11.8Set the TI AMPLITUDE BW the next value listed in the TI BW column of Table 23. Press the TI FREQUENCY SPAN key and set to the next value listed in the TI Span/Div column of Table 23.

4.11.9Press the TI MODE MARKER, PEAK *, then <PEAK or PEAK> as necessary to place the TI M 1 marker to the peak of the displayed signal.

4.11.10Record the M 1 amplitude indication.

4.11.11Subtract the value recorded in step 4.11.10 from the value recorded in step 4.11.7. Verify the results are within the values listed in the Limits column of Table 23.

4.11.12Repeat steps 4.11.8 through 4.11.11 for the remaining values listed in Table 23.

38

 

 

T.O. 33K3-4-3299-1

 

Table 23.

 

 

 

 

TI BW (Hz)

TI Span/Div (Hz)

Limits (dB)

 

 

 

30 k

10 k

Reference

30 M *

10 M

-2.0 to +2.0

10 M *

5 M

-2.0 to +2.0

5 M

2 M

-0.5 to +0.5

300 k

100 k

-0.5 to +0.5

3 k

1 k

-0.5 to +0.5

300

250

-1.0 to +1.0

100

50

-1.0 to +1.0

30

50

-1.0 to +1.0

10

20

-1.0 to +1.0

3

20

-1.0 to +1.0

* For AN930A Only.

4.11.13Set the Synthesized Leveled Generator for minimum output and disconnect test setup.

4.11.14Set the TI controls as follows:

MEMORY CONFIG

3

ENTER

9

ENTER

or

To toggle the reference to INT

RETURN

MEMORY CONFIG

39

T.O. 33K3-4-3299-1

4.12 SWEEP TIME CALIBRATION:

4.12.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.12.2Connect the Oscilloscope Calibrator through the Active Head and Attenuator to the TI ANALYZER INPUT connector.

4.12.3Set TI controls as follows:

FREQUENCY CENTER

First value listed in the TI

 

Frequency column of Table 26

FREQUENCY SPAN

ZERO

AMPLITUDE BW

300 kHz

AMPLITUDE SCALE

2 dB

MODE SWEEP

First value listed in the TI Sweep

 

Time column of Table 26

4.12.4Set Oscilloscope Calibrator for pulsed time marks into 50 at the first value listed in the Applied column of Table 26. Set Oscilloscope Calibrator OUTPUT ON/OFF to ON.

4.12.5Set TI controls as follows:

MODE SWEEP

SOURCE*

 

VIDEO

 

RETURN

 

TRIG*

 

TRIG LVL

4.12.6Using the TI Control Knob, adjust the TI Trigger Level for a stable display. It may be necessary to adjust TI Center Frequency to obtain suitable marker amplitude on the display.

4.12.7Press TI MODE SWEEP, MODE*, SNGL then ARM. Allow sufficient time for the TI sweep to finish.

4.12.8Press TI MODE MARKER, then MKR 1.

4.12.9Using the TI Control Knob, adjust the TI MKR 1 so that it is superimposed on the 2nd signal peak from the far left graticule line.

4.12.10Press TI MKR 2 and using the TI Control Knob, adjust the TI active marker so that it is superimposed on the 10th signal peak from the far left graticule line.

4.12.11Press TI DELTA.

4.12.12Verify the TI frequency indication is within the values listed in the Limits column of Table 24.

40

4.13 RESIDUAL RESPONSES CALIBRATION:

T.O. 33K3-4-3299-1

4.12.13Press TI MODE MARKER, then MKR 1 and MKR 2 as necessary to turn off the TI markers.

4.12.14Press TI MODE SWEEP, MODE*, then AUTO.

4.12.15Repeat steps 4.12.3 through 4.12.14 for the remaining values listed in Table 24.

Table 24.

Frequency (Hz)

Sweep Time (s)

Applied (s)

Limits (s)

 

 

 

 

5 M

200 n *

0.2

1.58 to 1.62

50 k

20

20

158 to 162

500

2 m

2 m

15.8 to 16.2 m

20

50 m

50 m

395 to 405 m **

10

100 m

0.1

790 to 810 m

10

1

1

7.90 to 8.10

10

10

10

79.0 to 81.0

*For AN930 Only.

**Due to TI marker resolution, limits may have to be verified using the TI display.

4.12.16Set the Oscilloscope Calibrator OUTPUT ON/OFF to OFF and disconnect test setup.

4.13.1

Press TI MEMORY SETUP, then INIT to restore TI initial settings.

 

4.13.2 Connect the Termination to the TI ANALYZER INPUT connector.

 

4.13.3

Set the TI controls as follows:

 

 

FREQUENCY START

9 kHz

 

FREQUENCY STOP

100 kHz

 

AMPLITUDE BW

RBW and VIDEO, as necessary

 

SWEEP TIME

As necessary to obtain a calibrated

 

 

sweep

4.13.4

Verify any Residual Response is within the first value listed in the Limits column of Table 25.

41

T.O. 33K3-4-3299-1

Table 25.

Freq Start (Hz)

Freq Stop (Hz)

Limits (dBm)

 

 

 

9 k

100 k

-90

100 k

22 G

-100

4.13.5Increase the TI FREQUENCY START and FREQUENCY STOP controls to cover the remaining frequency range in Table 25 in about 200 MHz increments. Adjust the AMPLITUDE BW and SWEEP TIME as necessary to maintain a calibrated sweep. Repeat step 4.13.4 for the remaining values in Table 25.

4.13.6Disconnect test setup.

4.14 SCOPE MODE AMPLITUDE CALIBRATION:

4.14.1Press TI MEMORY SETUP, then INIT to restore TI initial settings.

4.14.2Connect the Oscilloscope Calibrator through the Active Head to the TI EXTERNAL INPUT connector.

4.14.3Press TI MODE AUX, EXT IN *, AC, then SCOPE.

4.14.4Press TI AMPLITUDE SCALE and using the TI control Knob, set to the first value listed in the TI Vertical Scale column of Table 26. Press TI MODE SWEEP and set to 500 s.

4.14.5Set the Oscilloscope Calibrator for a 1 kHz squarewave into 1 M at the first value listed in the Applied column of Table 26. Set the Oscilloscope Calibrator OUTPUT ON/OFF to ON.

4.14.6Verify the applied signal is displayed on the TI screen.

4.14.7Press TI MODE MARKER.

4.14.8Using the TI Control Knob, adjust the TI MKR 1 so the marker is at the positive peak of the displayed signal.

4.14.9Press TI MKR 2.

4.14.10Using the TI Control Knob, adjust the TI MKR 2 so the marker is at the negative peak of the displayed

signal.

4.14.11Press TI DELTA.

4.14.12Press TI MODE SWEEP, MODE *, SNGL, then ARM.

4.14.13Press TI ARM several times and verify the TI amplitude indicates within the values listed in the Limits column of Table 26.

4.14.14Set the Oscilloscope Calibrator OUTPUT ON/OFF to OFF.

4.14.15Press TI AMPLITUDE SCALE and using the TI control Knob, set to the next value listed in the TI Vertical Scale column of Table 26. Press TI MODE SWEEP, MODE *, SNGL, then AUTO.

42

T.O. 33K3-4-3299-1

4.14.16Set the Oscilloscope Calibrator for a 1 kHz squarewave into 1 M at the next value listed in the Applied column of Table 26. Set the Oscilloscope Calibrator OUTPUT ON/OFF to ON.

4.14.17Verify the applied signal is displayed on the TI screen.

4.14.18Press TI MODE MARKER, then MKR 1.

4.14.19Using the TI Control Knob, adjust the TI MKR 1 so the marker is at the positive peak of the displayed

signal.

4.14.20Press TI MKR 2.

4.14.21Using the TI Control Knob, adjust the TI MKR 2 so the marker is at the negative peak of the displayed

signal.

4.14.22Press TI MODE SWEEP, MODE *, SNGL, then ARM.

4.14.23Press TI ARM several times and verify the TI amplitude indicates within the values listed in the Limits column of Table 26.

4.14.24Set the Oscilloscope Calibrator OUTPUT ON/OFF to OFF.

Table 26.

TI Vertical Scale (V/div)

Applied (V)

Limits (V)

 

 

 

5 m

20 m

19.8 to 20.2

10 m

50 m

49.7 to 50.3

20 m

0.1

99.4 to 100.6

50 m

0.2

198.5 to 201.5

100 m

0.5

497 to 503

200 m

1

0.994 to 1.006

500 m

2

1.985 to 2.015

1

5

4.97 to 5.03

2

10

9.94 to 10.06

5

20

19.85 to 20.15

4.14.25Repeat steps 4.14.15 through 4.14.24 for the remaining values listed in Table 26.

4.14.26Disconnect test setup.

43

T.O. 33K3-4-3299-1

4.15 SCOPE MODE BANDWIDTH CALIBRATION:

4.15.1Press TI MODE SWEEP, then AUTO.

4.15.2Press TI AMPLITUDE SCALE and using the TI Control Knob, set the TI vertical scale to 1 V/div.

4.15.3Connect the Oscilloscope Calibrator through the Active Head and Feedthrough Termination to the TI EXTERNAL INPUT connector.

4.15.4Set the Oscilloscope Calibrator OUTPUT ON/OFF to ON and the output for a 1 kHz leveled sine wave and 4 div of vertical display.

4.15.5Increase the Oscilloscope Calibrator frequency until the TI display drops to 2.8 V p-p.

4.15.6Press TI MODE SWEEP and using the TI Control Knob, reduce the TI sweep time as necessary for the best signal presentation.

4.15.7Verify the 2.8 V p-p signal frequency is 5 MHz as indicated on the Oscilloscope Calibrator.

4.15.8Set the Oscilloscope Calibrator OUTPUT ON/OFF to OFF and disconnect test setup.

4.15.9Set all POWER switches to OFF or STBY. Disconnect and secure all equipment.

CALIBRATION PERFORMANCE TABLE

Not Required

44

T.O. 33K3-4-3299-1

APPENDIX A

A-1 TIME BASE ADJUSTMENT: [Room Temperature Crystal Oscillator (RTXO)]

A-1.1 Set TI POWER APPLIED/ON switch to APPLIED and disconnect power source. A-1.2 Remove the TI cover.

A-1.3 Connect TI to appropriate power source and set POWER APPLIED/ON switch to ON.

A-1.4

Allow a 15 minute warm-up period.

 

A-1.5

Set the TI controls as follows:

 

 

MEMORY CONFIG

 

 

3

 

 

ENTER

 

 

9

 

 

ENTER

 

 

or

To toggle the reference to INT

A-1.6 Connect Frequency Standard 10 MHz REF OUT to the Electronic Counter EXT FREQ STD INPUT (1-10 MHz). Set Electronic Counter INT STD/EXT STD switch to EXT STD.

A-1.7 Disconnect TI W25P1 cable from A4J5 connector.

A-1.8 Connect the Electronic Counter CHANNEL A input to the TI A4J5 connector. Set Electronic Counter 50 /1M switch to 50 .

A-1.9 Adjust the TI Reference Oscillator Adjustment (A27C1) for an Electronic Counter indication as close as possible to 100 MHz.

A-1.10 Allow TI 100 MHz Reference Oscillator a minimum of one (1) hour to stabilize. Repeat step A-1.9 as required.

A-1.11 Disconnect the test setup.

A-1.12 Set TI POWER APPLIED/ON switch to APPLIED and disconnect power source. Replace the TI cover. A-1.13 Connect TI to appropriate power source and set POWER APPLIED/ON switch to ON.

A-1.14 Continue with step 4.1.6.

A-1/(A-2 Blank)

Loading...