IFM RD52 Installation Instructions Manual

Page 1
Installation Instructions
Universal process adapter RD52
E33340
704966/01 02/2011
UK
Page 2
2
B
A
► Grease thread and sealing (B) of the
coupling nut (C) with paste (A). The paste must be suitable and ap­proved for the application and compat­ible with the elastomers used.
► Slip the coupling nut onto the sensor (D)
from the bottom.
F
E
A
► Grease the sensor thread with the paste
(A) (see above ).
► Place the sealing ring (E) in the groove
(F) of the sensor. The adapter is supplied with an EPDM O-ring. More sealing rings are available as accessories:
• FKM O-ring, order no. E30123.
• PEEK sealing ring, order no. E30124. Use a new adapter when changing from a PEEK seal to O-ring sealing.
Page 3
3
UK
G
► Screw the sensor into the adapter (G)
until it is hand-tight. When doing so, avoid mechanical effects on the sealing areas, the hygienic area and the pres­sure measuring cell..
► Make sure that the O-ring (H) is correctly
positioned.
I
► Clamp the sensor + adapter into a
lamping device (I). Tighten the clamping device only slightly so that the adapter does not warp.
► Tighten the sensor using a spanner until
you can feel the end stop (this corre­sponds to a maximum tightening torque of 35 Nm). Note: Further tightening may affect the sealing effect.
► Insert the unit in a universal process connection RD52. ► Tighten the coupling nut.