IEI Technology POCm-W22C-ULT3, POCm-W24C-ULT3 Installation Manual

Page 1
Medical Panel PC with 6th Gen. Intel® Core™ / Celeron® CPU ,
4 GB DDR4 RAM, Wi-Fi 802.11a/b/g/n/ac, P-CAP Touchscreen,
Three Battery Bays, 5-Megapixel Camera and Microphone
POCm-W22C-ULT3 POCm-W24C-ULT3
Quick Installation Guide
Version 1.01
©2018 Copyright by IEI Integration Corp.
All rights reserved.
Page 2
1
Revision
Date
Version
Changes
September 11, 2018 1.01 Minor update August 28, 2018 1.00 Initial release
Intended Use
The POCm-W22/24C-ULT3 medi cal panel PC is intended to be used to display general purpose medical im ages. The device shall not be used for diagnosis purpose or life supporting system. Equipment connected to analog or digital interfaces must comply with the respective IEC Standards (e.g. IEC 60950 for data processing equipment and IEC 60601-1 for medical equipment). Furthermore all configurations shall comply with the current version of the standard for S Y STEMS IEC 60601-1-1. Everybod y who connects additional equipment to the signal input part or signal output part configure a medical system, and is ther efore responsible that the system complies with current version of the requirement s of the system standard IEC 60601-1-1. If in doubt, consult the technical service department or your local repr esentative.
Manual Conventions
WARNING
Warnings appear where overlooked details may cause damage to the equipment or result in personal injury. Warnings should be taken seriously.
CAUTION
Cautionary messages shou ld be heeded to help reduce the chance of
losing data or damaging the product.
NOTE
These messages inform the reader of essential but non-critical information. These messages should be read carefully as any directions or instructions contained therein can help avoid making
mistakes.
OPERATING INSTRUCTION
Follow operating instructions or consult instructi ons for use.
IEC 60417-5009: Stand-by
IEC 60417-5031: Direct current
Page 3
2
Overview
The POCm-W22/24C-ULT3 is a 6th generation Intel® Core™ i5-6300U processor powered medical-grade panel PC with a rich variety of functions and peripherals. All POCm-W22/24C-ULT3 models are designed for easy and simplified integration into point-of-care (POC) applications. The system comes with 4 GB of preinstalled DDR4 memory and supports a maximum of 32 GB ensuring smooth data throughputs with reduced bottlenecks and fast system access.
Two RS-232/422/485 serial ports, four USB 3.0 ports and two USB 2. 0 ports provide simplified connectivit y to a variety of external peripheral devices. Wi-Fi 802.11a/b/g/n/ ac high speed wireless and two RJ-45 GbE connectors allow for smooth connection of the system to an ext ernal LAN. Three hot-swappable battery bays all ow installation of three batterie s to provide continuous power for 8~16 hou rs.
Page 4
3
Front Panel
Bottom Panel
Page 5
4
Side Panel
Rear Panel
Page 6
5
System Specifications
POCm-W22C-ULT3
POCm-W24C-ULT3
LCD and Touchscreen
LCD Size 21.5" (16:9) 23.8" (16:9)
Max. Resolution
1920 (W) x 1080 (H)
1920 (W) x 1080 (H)
Brightness (cd/m2)
250
250
Contrast Ratio
1000:1
1000:1
LCD Color
16.7M (RGB 6-bit)
16.7M (RGB 6-bit + Hi-FRC)
Pixel Pitch (mm)
0.24825 (H) x 0. 24825 (V)
0.2745 (H) x 0. 2745 (V)
Viewing Angle (H-V)
170°/160°
178°/178°
Backlight MTBF
30,000 hrs (LED backlight)
30,000 hrs (LED backlight)
Touchscreen
Projected capacitive type wi th USB interface
Multi-touch
10-point touch
Touch Controller
EETI
Surface Hardness
6H
System
CPU
Intel® Celeron® 3855U / Intel® Core™ i5-6300U /
Intel® Core™ i7-6600U
Memory
Two 260-pin 2133/1866 MHz dual-channel non-ECC unbuffered DDR4 SO-DIMMs sup ported (system max. 32 GB)
Preinstalled with 4 GB memory
GbE Controller
LAN1: Intel I211 Ethernet controller
LAN2: Intel® I219 Ethernet controller
I/O Ports
1 x DC input jack
2 x HDMI output connector
2 x GbE LAN (RJ-45 connector)
2 x RS-232/422/485 serial port (DB-9 connector)
4 x USB 3.0 connectors
2 x USB 2.0 connectors (side panel)
1 x Audio out (side panel) 1 x Mic-in (side panel)
1 x Digital microphone
Storage
One 2.5” SATA 6Gb/s HDD bay
Audio
Two 3 W speakers
Webcam & Microphone
5-megapixel CMOS front-facing camera wi th auto focus and
digital microphone
Expansion Interface
2 x M.2 M-key 2242/2260/2280 slot (PCIe + SATA) with RAID 1 x M.2 A-/E-key 2242 slot (PCIe + USB)
1 x Full-size/Half-size PCIe Min i (PCIe + USB)
Page 7
6
TPM
TPM 2.0 (optional)
Other Features
1 x Power on/off
1 x Brightness up
Function Keys
1 x Brightness down
1 x Volume up
1 x Volume down
1 x Touch lock (clean mode) or LCD on/off
LED Indicators
3 x Battery indicator
1 x Power indicator
Cooling Method
Fanless
Connectivity
Wi-Fi and Bluetooth
802.11a/b/g/n/ac dual band, Bluetooth v4.2
(M.2 2230 module, Intel® 8265)
LAN
Two GbE LAN connectors
Physical
Construction Material
PC+ABS plastic with anti-bacterial material
Mounting
Wall and stand mounting
VESA 75 mm x 75 mm or 100 mm x 100 mm
Dimensions (W x H x D)
543 x 350 x 71 (mm)
594.6 x 379.6 x 71 (mm)
Net Weight
7.07 kg
8.18 kg
Environment
Storage/Transportation
Temperature
-20ºC ~ 60ºC
Humidity
10% ~ 90% (non-condensing)
Pressure
700 hPa ~ 1060 hPa
Operating
Temperature
0ºC ~ 40ºC
Humidity
10% ~ 90% (non-condensing)
Pressure
700 hPa ~ 1060 hPa
Vibration
1G
Shock
Operating Shock: 5G peak acceleration (11ms duration)
Non-Operating Shock: 10G peak acceleration (11ms duration)
IP Level
IP 65 compliant front panel
Power
Power Input
19 V DC
Power Adapter
150 W FSP FSP150M-ABA medical-grade power adapter
(P/N: 63040-010150-400-RS)
Input: 100 V AC ~ 240 V AC, 50 Hz ~ 60 Hz, 2A ~ 0.85A
Output: 19 V 7.89 A
Battery
3 x Hot-swappable battery bay
Page 8
7
Installation Precautions
When installing the medical panel PC, please follow the precautions liste d below:
Manufacturer authorization: Do not modify this equipment
without authorization of manufacturer.
Certified Engineers: Only certified engineers sho uld install and
modify the hardware settings.
Power turned off: When installing the medical panel PC, make
sure the power is off. Failing to turn off the power may cause severe injury to the body and/or damage to the system.
Anti-static Discharge: If a user open the rear pa nel of the m edical
panel PC, to configure the jumpers or plug in added peripheral devices, ground themselves first and wear an anti-static wristband.
WARNING:
Danger of explosion if battery is incorrectly replaced. Replace only
with the same or equivalent t ype recommended by the manufacturer.
Disposal of a battery into fire or a hot oven, or mechanically crushing
or cutting of a battery, that can result in an explosion.
Leaving a battery in an extremely high temperature surrounding
environment that can result in an expl osion or the leakage of flammable liquid or gas.
A battery subjected to extremely low air pressure that may result in
an explosion or the leakage of flamm able liquid or gas.
Page 9
8
Mounting the System
WARNING:
Use suitable mounting apparatu s and be sure to secure the screws of the mounting apparatus tightly to avoid risk of injury.
NOTE:
When mounting the POCm-W22/24C-ULT3 flat panel PC, it is bet ter to have more than one person to help with the installation to make sure the POCm-W22/24C-ULT3 does not fall down and get dam aged.
The POCm-W22/24C-ULT3 is VESA (Video Electronics Standards Association) compliant and can be m ounted on a mounting device wit h a 75 mm or 100 mm interface pad. The POCm-W22/24C-ULT3 VESA mount retention screw holes are shown below. Refer to the installation guide that came with the mounting device to mount the POCm-W22/24C-ULT3.
Page 10
9
Powering On the System
WARNING:
To avoid risk of electric shock, this equipment must only be connected to supply mains with protective earth.
CAUTION:
The FSP FSP150M-ABA power adapter came with the POCm-W22/24C-ULT3 is a forming part of the medic al device.
Step 1: Connect the power cord to the powe r adapter. Connect the other
end of the power cord to a power source.
Step 2: Connect the power adapter to the power connector of the
POCm-W22/24C-ULT3.
Step 3: Locate the power button on the right panel. Step 4: Short press the power button to t urn on the POCm-W22/24C-ULT3.
Shutdown Procedure
Turn off the power and disconnect the power cord.
To prevent the risk of electric shock, make sure power cord is unplugged from wall socket. To fully disengage the power to the unit,
please disconnect the power cord from the AC outlet. Refer servicin g to qualified service personnel. The AC out l et shall be readily available and accessible.
Page 11
10
Troubleshooting
If the following situations happen, contact your distributor, sales representatives or IEI customer serv ice center for technical support.
The HDD is installed correctly, but the POCm-W22/24C-ULT3 is
unable to boot with AC power input after pressing the power button.
Unable to shut down the POCm-W22/24C-ULT3 normally Please have the following informatio n prepared prior to reporting the abnormal situations:
Product name and S/N
OS, BIOS version and applications
A complete descriptio n of the ab normal sit uation (with photos or vid eo
if available)
System Maintenance
If the components of the POCm-W22/24C-ULT3 fail they must be replaced. Please contact the system reseller o r vendor to purchase the replacement parts.
Maintenance and Cleaning
Prior to cleaning any part or component of the POCm-W22/24C-ULT3, please read the details below.
To clean the POCm-W22/24C-ULT3,
o remove dirt with a lightly moistened cloth. Then wipe the external
chassis with a soft dry cloth.
o use 75% ethanol alcohol to clean the ex ternal chassis.
Cleaning frequency: follow the cleaning method guidel i nes of the
hospital.
Except for the LCD panel, never spray or squirt liquids directly onto
any other components.
The interior of the POCm-W22/24C-ULT3 does not require cle aning.
Keep fluids away from the POCm-W22/24C-ULT3 interior.
Never drop any objects or liquids through the opening s of the
POCm-W22/24C-ULT3.
Page 12
11
Accessories
The medical panel PC is shipped with t he following components:
Quantity Item Image
1
POCm-W22/24C-ULT3
medical panel
PC
1
Power cord (P/N: 32702-000200-100-RS)
1
150 W FSP FSP150M-ABA medical-grade power adapter (P/N: 63040-010150-400-RS)
4
Round-head screw (M3*3) for HDD installation (P/N: 44003-030032-RS)
1 Quick Installation Guide
Optional Items
The following components may be separately purchased:
Item and Part Number Image
Li-ion battery pack, 7800 mAh (P/N: 31603-000069-RS)
Cable cover (P/N: 43106-0272Q7-00-RS)
20-pin Infineon TPM 2.0 module, software management tool, firmware v5.5 (P/N: TPM-IN02-R20)
Page 13
12
DECLARATION OF CONFORMITY
This equipment is in conformit y with the following EU directives:
EMC Directive (2004/108/EC, 2014/30/EU)
Low-Voltage Directive (2006/95/EC, 2014/35/EU)
RoHS II Directive (2011/65/EU, 2015/863/EU)
Medical Device Directive 93/42/EE C: EN 60601-1
If the user modifies and/or install other dev ices in the equipment, the CE conformity declaration may no l onger apply. If this equipment has telecommunications functionality, it also complies with the requirements of the Radio E quipment Directive 2014/53/EU. IEI Integration Corp declares that t his equipme nt i s in complia nce wit h the essential requirements and ot her relevant provisions of Directive 2014/53/EU.
FCC WARNING
This equipment complies with part 18 of the FCC Rules.
This equipment generates, use s and can radiate radio frequency energy and, if not installed and used in acco rdance with the instructions, m ay cause harmful interference to radio co mmunicat ions. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to corre ct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the e quipment and receiver.
Connect the equipment into an outlet on a circuit dif ferent from t hat
to which the receiver is connected.
Page 14
13
ROHS STATEMENT
The label on the product indicates this product conforms to European (EU) Restriction of Hazardous Sub st ances (RoHS) that set maximum concentration limits on hazardous materials used in electrical and electronic equipment.
CHINA RO H S
The label on the product indicates the estimated “Environmentally Friendly Use Period” (EFUP). Thi s is an estimate of the number of years that these substances would “not l eak out or undergo abrupt change.” This product may contain repla ceable sub-assemblies/components which have a shorter EFUP such as batteries a nd lamps. These components will
be separately marked.
Page 15
14
Safety Precautions
WARNING:
The precautions outlined below should be strictly followed. Failure to follow these precautions may result i n permanent damage to the
POCm-W22/24C-ULT3.
General Safety Precautions
Please ensure the following safety precautions are adhered to at all times.
To prevent the risk of electric shock, make sure power cord is
unplugged from wall socket. To fully disengage the power to the unit, please disconnect the power cord from the ac outlet. Refer servicing to qualified service personnel. The AC outlet shall be readily available and accessible.
Users must not allow SIP/SOPs and the patient to come int o contact
at the same time.
Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked “Hospital Only” or “Hospital Grade”.
Follow the electrostatic precautions outlined below whenever the
POCm-W22/24C-ULT3 is opened.
Make sure the power is turned off and the po wer cord is di sconnect ed
whenever the POCm-W22/24C-ULT3 is being installed, m oved or modified.
Do not apply voltage levels that exceed the specified voltage range.
Doing so may cause fire and/or an electri cal shock. Use a powe r cord that matches the voltage of the power outlet, which has been approved and complies with the safety standard of your particular country.
Electric shocks can occur if the POCm-W22/24C-ULT3 chassis is
opened when the POCm-W22/24C-ULT3 is running. To avoid risk of electric shock, this equipment must onl y be connected to a supply mains with protective earth.
DO NOT LEA VE THIS EQUIPMENT IN AN UNCONTROLLED
ENVIRONMENT WHERE THE STORAGE TEMPERATURE IS BELOW -20° C (-4°F) OR ABOVE 60° C (140° F). IT MAY DAMAGE THE EQUIPMENT.
Do not drop or insert any objects into the ventilation openi ngs of the
POCm-W22/24C-ULT3.
If considerable amounts of dust, water, or fluids enter the
POCm-W22/24C-ULT3, turn off the power supply immediat ely, unplug the power cord, and contact the POCm-W22/24C-ULT3 vendor .
Page 16
15
DO NOT:
o Drop the POCm-W22/24C-ULT3 against a hard surface. o Strike or exert excessive force onto the LCD panel. o Touch any of the LCD panels with a sharp object o In a site where the ambient temperature exceeds the rated
temperature
Product Disposal
CAUTION:
Risk of explosion if battery is replaced by an incorrect ty pe. Only certified engineers should replace the on-board battery. Dispose of used batteries acc ordi ng to i n stru ction s a nd l ocal r egul a tions.
Outside the European Union - If you wish to di spose of used electri cal
and electronic products outside the European Union, please contact your local authority so as to comply with the correct disposal method.
Within the European Union–The device that produces less waste and
is easier to recycle is classifi ed as electronic device in terms of the European Directive 2012/19/E U (WE E E), and must not be disposed of as domestic garbage.
EU-wide legislation, as implemented in e ach Member State, requires that waste electrical and ele ct ronic products carrying the mark (left) must be disposed of separately from normal
household waste. This includes monit ors and electrical accessories, such as signal cables or power cords. When you need to dispose of your display product s, pl ease follow the guidance of your local authority, or ask the shop where you purchased the product. The mark on electrical and electronic produ ct s only applies to the current European Union Member States. Please follow the national guidelin es for electrical and electronic product disposal.
Classification
Power by Class I power supply (IEI, POCm-W22/24C-ULT3) No Applied Part. No protection against the ingress of water: IPX0 Mode of operation: Continuous Operation
Page 17
16
The equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide: Not AP or APG Category.
EMC Test Summary
Guidance and manufacturer’s declaration – electromagnetic emiss ions
The model POCm-W22/24C-ULT3 is intended for use in the electromagnetic
environment specified below. The customer or the user of the model
POCm-W22/24C-ULT3 s hou ld assur e that it i s used in such an environ m ent.
Emissions test
Compliance
Electromagnetic environm ent – guidance
RF emissions
CISPR 11
The model POCm-W22/24C-ULT3
uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nea rby
electronic equipment.
RF emissions CISPR 11
The model POCm-W22/24C-ULT3 is
suitable for use in all establishments, including domestic establishm e nts and those directly connec ted to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissi ons IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances betw een portable and mobile RF communications equip ment and the model POCm-W22/24C-ULT3
The model POCm-W22/24C-ULT3 is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of the model POCm-W22/24C-ULT3 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model POCm-W22/24C-ULT3as recommended below, according to the maximum output power of the
communications equip ment.
Rated maximum
output power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80
MHz
d = 1,2
80 MHz to 800
MHz
d = 1,2
800 MHz to 2,5
GHz
d = 2,3
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
PPP
Page 18
17
For transmitters rated at a maximum output power not listed above, the
recommended separation distan ce d in metres (m) can be esti ma ted using the equation applicable to th e f requency of the transmitter, where P is t h e maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration – electromag netic immunity
The model POCm-W22/24C-ULT3 is intended for use in the electromagnetic
environment specified below. The customer or the user of the model
POCm-W22/24C-ULT3 s hou ld assur e that it i s used in such an environ m ent.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity
should be at least 30 %.
Electrical fas t
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output lines
±2 kV for power
supply lines
±1 kV for input/output lines
Mains power quality
should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
Mains power quality
should be that of a typical commercial or hospital environment.
interruptions
and voltage variations on power supply input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT) for 0,5 cycle
40 % UT (60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT)
<5 % UT
(>95 % dip in UT) for 0,5 cycle
40 % UT (60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT)
Mains power quality
should be that of a typical commercial or hospital environment. If the user of the model POCm-W22/24C-ULT3 requires continued operation during power mains interruptions, it is recommended that the
model
Page 19
18
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
POCm-W22/24C-ULT3
be powered from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should be at levels characteristic of a typical location in a typical commercial or
hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level .
Guidance and manufacturer’s declaration – electromag netic immunity
The model POCm-W22/24C-ULT3 is intended for use in the electromagnetic
environment specified below. The customer or the user of the model POCm-W22/24C-ULT3 s hou ld assur e that it i s used in such an environ men t.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2,5 GHz
Vrms
V/m
Portable and mobile RF
communications equipment should be used no closer to any part of the model POCm-W22/24C-ULT3, including cables, than the rec omme nded separation distance calculated from the equation applicable to the frequency of the tr ansmitter.
Recommended separation distance
d = 1,2
d = 1,2
80 MHz to 800 M Hz
d = 2,3
800 MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).
P
P
P
Page 20
19
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, su ch as b ase stations for radio (cellular/cordless) telepho ne s and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy . To assess the e lectromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the model POCm-W22/24C-ULT3 is used exceeds the applicable RF compliance level above, the model POCm-W22/24C-ULT3 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the model POCm-W22/24C-ULT3.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than V/m.
Contact Information
IEI Integration Corp.
No. 29, Zongxing Rd., Xizhi Dist., New Taipei City 221, Taiwan
TEL: +886-2-8691-6798
Fax: +886-2-6616-0028
www.ieiworld.com
sales@ieiworld.com.tw