Centra-CL3 series units are high-speed, multipurpose
centrifuges, used
applications.
The CL3 Series
(Centra-CL3) and refrigerated (Centra-CL3R). Sections
this manual that apply
designated Centra-CL3R
Both units accommodate a wide variety
fixed angle, swinging bucket, and fixed horizontal designs.
They can process tubes, bottles, microplates, microcapillary
tubes, cytological slide carriers, and microsample tubes.
The units can centrifuge up to one liter
operation.
Each centrifuge has an easy to use front panel that provides
two modes
of
operation: Manual and Programmed.
in
medical, industrial, and scientific
is
available in two versions: ventilated
to
the refrigerated version will be
or
refrigerated only.
of
rotors, including
of
fluid in a single
of
f
l
Manual mode
speed/force, and time values for individual runs.
Program mode allows you to define and save a maximum
ninety-nine specific sets
reuse.
Centra-CL3 Series Operation Manual
is
used for entering temperature (CL3R only),
of
run parameters, to recall and
1
of
TABLE
OF
CONTENTS
I
t
.Centra-CL3 Series Operation Manual
.Chapter 1: Introduction
1.
1 General Product Description ............................................................................... 1
1.2 About this Manual .............................................................................................. 2
1.3
Warnings, Cautions,
Chapter 2: Installation
2.1
Receiving the Unit .............................................................................................. 5
2.2 Site Preparation .................................................................................................. 5
2.3 Power Configuration ........................................................................................... 7
2.4 Moving the Unit .................................................................................................. 8
2.5 The Front Panel .................................................................................................. 8
. Chapter 3: Operation
3.
1 Rotor and Accessories ......................................................................................
3
.2
Starting and Stopping a Run .............................................................................
The CL3 Series features a maintenance-free, brushless
motor, and an easy-to-use front panel, which provides three
versatile timing modes: automatic timed run, momentary
spin, and continuous operation (hold mode). Acceleration
and brake rates may be controlled,
for fast separations or slow for delicate samples. Repeat
runs, with the same speed and time settings, may be
achieved at the touch
A fail-safe cover interlock insures that
before a run can begin, and keeps the cover closed, until the
rotor has reached a safe low speed (below 100 rpm), even in
the event
The rugged steel cabinet and rigid construction provide
quiet operation and long term reliability.
1.2 About This Manual
of
a power failure.
of a key.
to
optimize runs: rapid
the
cover is closed,
!
;
I
The Operations Manual contains all
needed to install, operate, and maintain a CL3 Series
centrifuge. Refrigerated and ventilated models operate
similarly, and any differences are highlighted and noted,
throughout this manual. This manual, also, contains speed
and force, derating, chemical resistance, and
decontamination tables. The last chapter lists the units'
specifications.
This manual
to operation information, it contains a few basic
troubleshooting techniques, and a chapter on maintenance.
This Operation Manual
centrifuge units.
is
written for centrifuge operators. In addition
is
not a guide for servicing
of
the
information
Centra-CL3 Series Operation Manual
2
1.3 Warnings, Cautions, and Notes
The tenns warning, caution, and
in
meanings
Warning
• A
that could result
Caution
• A
this manual.
advises against certain actions
in
personal injury.
advises against actions
note
have specific
or
or
situations
situations that
could damage equipment, produce inaccurate data, or
invalidate a procedure.
• A Note provides useful infonnation regarding an
operation, function,
or
procedure.
Centra-CL3 Series Operation Manual
3
- !
t
-
2.1
Receiving the Unit
International Equipment Company (IEC) ships the
centrifuge
Follow the unpacking instructions on the carton.
complete and return the postage-paid warranty card.
2 Installation
in
a carton that protects it from shipping hazards.
Be
sure
to
2.2
·
Site Preparation
The unit normally resides on a bench top. The Centra-CL3
in
(ventilated model) can be placed
than 0°C), for processing temperature-sensitive samples.
When you remove the centrifuge from a cold environment,
do not operate for a minimum
condensation will evaporate.
Note: When used
noise may become evident. The bearing lubricant thickens
at low temperatures.
thinned and distributed more evenly. Once this occurs, any
noise should subside.
in
a cold room environment, some bearing
As
the centrifuge speeds up, it
a cold room (no colder
of
two hours, so that any
is
Centra-CL3 Series Operation Manual
s
The following table lists the physical dimensions for
Centra-CL3 and Centra-CL3R:
the
CL3R
Sample
Cover
Cover
Width
Depth
A clearance
side
centrifuge on a clean, dry surface,
suction feet at the
area beneath the unit free
The resting surface must
free operation.
improperly loaded centrifuge may vibrate
The CL3 Series uses
appropriate for use throughout the world. Please check
catalog number
ensure that the machine you have is the proper power
configuration. For best results, the refrigerated centrifuge,
Centra-CL3R, should be used
Variations in line voltage
speed and other characteristics. Less than nominal line
voltage may prevent the centrifuge from reaching maximum
published specifications
Power line voltage, at some locations, may sag, when
refrigeration system turns on.
The unit requires a grounded power supply
If
outlet).
outlets, arrange for a proper grounding. The power
plugs in on the left rear side
your facility does not have grounded
AC
power in different configurations,
of
the model that you have purchased,
on
a dedicated line.
or
frequency affect the
of
speed and/or temperature.
(3
prong
of
the unit.
the
to
unit's
the
power
cord
~=~o~=:i:ili~~~t~~~e
power plug tbatdoes not
have
~
gr~Undilig
pin.
··
·.···
· ·
Circuit Breaker
The
power
highest
interchanged
demand.
createa
The system provides
emergency situations, such as power surges, that could
damage the unit.
If
the circuit breaker trips:
1.
Unplug the unit.
2.
Press the white button, on the left side
3.
Plug the unit back
cord
current
with
Exchange
fire
hazard.
provided with the tiJlltiscorrectly rated
demand.
cords
of
This
poweirord.shouldJ1ot
from
equipinentwitb.Jower current
power
cords
befiVeell
an
automatic circuit breaker, for
in.
equipment•may
of
the unit.
be
forthe
Centra-CL3 Series Operation Manual
7
2.4 Moving the Unit
Suction cups, at the bottom
the work surface. Keeping the unit stationary is a safety
feature.
To move the unit to a new location:
1.
Check that the new site meets the criteria
2.2., before moving the unit.
2.
Position a flat object, such as a tongue depressor, near a
suction cup at the bottom
3.
Lift up
enough to break the vacuum seal.
4.
When
work surface.
an
edge
of
all
four cups are disengaged, lift the unit from the
Caution: When the unit
the suction cups adhere correctly to the work surface.
use
control panel and refiigeration system (Refrigerated model
only). The
Stop a run with the
display chamber temperature whenever they are plugged in,
but will not cool down
light indicates that the centrifuge is plugged
key:
The On/Off key must be activated,
of
the front panel. This key applies power
On/Off key is inoperative during actual runs.
STOP key. Refrigerated models
if
the unit
is
off
to
enable
to
the
The red STOP
in.
The control panel contains numeric displays for
(Speed/Force), Time, and Temperature (Refrigerated only).
These displays have
display) and Set (dim display).
In Actual mode (bright display), they indicate current run
conditions, such as:
• rotor speed
• elapsed time of,
• actual temperature (Refrigerated only).
In Set mode (dim display), the display indicates the desired
settings for the run. Set mode
• whenever you use the up and down arrows
• briefly, at the start
• briefly, after the unit
The display
The display
numeric displays can, also, display warning
messages (see Section 3 .5). Descriptions
appear on the following pages.
is
is
two
states
or
force
or
time remaining
of
is
bright, when it shows Actual run conditions.
dim, when it shows Set parameters.
or
modes: Actual (bright
is
operative:
a run
switched ON
in,
the run
RPM/RCF
or
error
of
the displays
The
Centra-CL3 Series Operation Manual
9
0
Speed/Force display: The number in the speed/force
rotor
speed
display (above this symbol) represents the
RPM
or
force in RCF. Press this symbol
RPM
and RCF. When
indicates revolutions per minute. When
the display indicates relative centrifugal force.
to
arrow keys
shows speed within 50 RPM. It never requires calibration.
Select speed in increments
through 8,500 RPM. Select RCF in increments from
1 - 1,000 xg by 50 xg and above 1,000 xg by 100 xg. The
numeric display can, also, display warning
messages (see Section 3.5).
change the set speed
RPM
is selected, the display
of
100 RPM, from 1,000
or
to
toggle between
RCF
is selected,
force.
or
The
error
Use
in
the
display
Time display: The number
symbol) indicates time. Below ten minutes, time is
displayed as minutes:seconds. Above ten minutes,
minutes are displayed.
Time is set
• 1 second increments, from 1 15
•
• 1 minute increments, from 5 - 99 minutes.
In normal timed mode, the system counts
point. In time
counts up.
In At-Start mode, the timer begins counting
run. In At-Speed mode, the timer begins counting when the
rotor
Temperature display: The number
this symbol) represents temperature
from - 9 °C
in:
second increments, from 1 - 5 minutes.
Hold
or
reaches
95%
of
through+
in
the display (above this
59
seconds.
down
momentary spin modes,
at
the set speed.
in
the display (above
in
degrees Celsius,
40 °C (Refrigerated only).
just
from the set
the
system
the
start
of
a
Centra-CL3 Series Operation Manual
10
:I
Program
key: This key saves the currently displayed
desired settings as stored programs I through 99 (see
Section 3.3). The numeric display (above this symbol)
shows the stored program number and mode
of
operation
(see Section 3 .2).
Rotor/Radius
indicates the selected rotor number
Key: The display (above this symbol)
or
the
rotor radius,
in
centimeters. The rotor/radius key toggles between the two
displays. The applicable IEC rotor numbers are supplied
in
the memory, along with their maximum radii, in
centimeters. To select a rotor number, toggle
to
ROTOR,
and press an arrow key under the rotor/radius display. To
change the radius, toggle to RADIUS, and press an arrow
Note
key under the rotor/radius display.
that the radius
cannot be changed to a radius larger than the maximum
radius,
or
less than the minimum radius, for the selected
rotor.
Use the arrow keys to view or change the Set parameters
for Speed/Force, Time, Temperature (Refrigerated only)
Rotor/Radius,
or
Program. The first time the key is
pressed, the numeric display switches from Actual readings
If
to Set parameters, without changing them.
key a second time, the selected parameter increases
decreases once for each depression.
will
down, the setting
keep changing, until you release the
If
you hold the key
you press the
or
key.
The longer you hold the key, the more rapidly the setting
changes. Hold a key down, to approach a desired setting.
Then, press the up or down key, repeatedly, to select the
exact setting. When you release the arrow keys for
3 seconds, the display returns to the Actual readings.
Centra-CL3 Series Operation Manual
11
l
Acceleration and Braking
Gentle acceleration and braking can be selected, when
centrifuging delicate samples. The gentle settings avoid
mixing
This key controls rotor acceleration up to 400 RPM.
yellow light over the rabbit is lit,
selected.
acceleration
3 5 seconds to achieve 400 RPM, depending on the rotor
and its contents. After 400 RPM,
applied, until the set speed is reached.
This key controls rotor braking.
rabbit
the turtle
rotor
out,
operating speed to a stop.
This key starts a run. A run is governed by the Set
parameters (manual
light blinks, until the rotor reaches 95%
speed. The light stays on until the end
of
density gradients
If
the yellow light over the turtle
is
selected. Slow acceleration takes from
is
lit,
full
braking
is
lit,
slow braking is selected. (This means the
will
coast down from 600 RPM.)
all
braking is disabled; and the rotor
or
or
breakup
full
If
is
selected.
programmed). The associated green
of
pellets.
If
acceleration
is
full
acceleration
the yellow light over the
If
the yellow light over
If
both lights are
will
of
the set run
of
the run .
is
lit, slow
15
is
coast from
the
to
l
}
I
l
1
•
This key stops a run. (A run will, also, stop automatically
when the set time has elapsed
momentary mode.) The red light will flash, to indicate the
rotor
is
still slowing down (braking
run ends, the red light stays on, indicating that the rotor has
stopped.
This key unlocks the cover. This key
is
in
progress. Pressing it will not stop the run. The cover
will not unlock, until the rotor has slowed to below 100
rpm. The cover will open automatically, on ventilated units
only, at the end
of
a run.
or
the start key
or
coasting). When the
is
is
released, in
inoperative
if
a run
Centra-CL3
Series
Operation Manual
12
3 OPERATION
3.1
Rotor and Accessories
A balanced
unbalanced load produces vibration, and can damage the
unit. A 2 gram load imbalance, at a speed
imparts force equivalent to
Always ensure that the rotor is loaded symrn.etrically, with a
full
complement
tubes. Tube adapters should also be installed symmetrically.
IEC rotors are dynamically balanced at the factory. IEC
matches removable parts (trunnion rings, shields, buckets,
and carriers) to within I gram, and stamps the weight on
each piece. Check these markings, whenever you
interchange parts, to ensure that opposite parts are matched.
Ensure that the total weight
parts, loaded
within 1 gram. The position numbers, present on many
rotors and adapters, identify opposing tube positions.
To
obtain good dynamic balance, opposite loads must not
only be equal
of
gravity. Opposing containers must be alike
thickness, and distribution
especially important for large containers.
lo.ad
is
essential for all centrifuges.
9.
I kg (20 pounds) at rest.
of
accessories, and a
of
samples and removable
in
opposing positions, are equal
in
mass, but must, also, have the same centers
of
glass
full
or
plastic. This
An
of
4600 RPM,
(or paired) set
in
weight, to
in
of
shape,
is
Centra-CL3 Series Operation Manual
13
I
Tubes loaded into swinging bucket
symmetric, around the
ax.is
of
rotating the entire rotor 180°,
in
the same apparent positions (not mirror images). In
rotors
must be
rotation. Verify this by
by
hand.
The
loads should be
addition, the loads within each bucket must, also, be
symmetric around the bucket's pivot axis. Verify this
ensuring that each bucket
from the vertical, when the rotor is
balance within each bucket ensures that
tubes swing out to horizontal, when
is
loaded so that it does not tilt
at
rest. Maintaining
the
bucket and the
the
rotor reaches
by
operating speed, applying centrifugal force toward the
of
bottom
causes vibration and premature wear
the tubes. Failure to achieve full swing-out
of
the rotor and the
motor.
Samples oflike (similar) specific gravities may be processed
in
the same run, provided that the samples
of
the same type
are balanced around the rotor, as though they were the only
in
pairs
the rotor.
I
Rotor Balance
Load tubes
1.
Load two tubes at positions:
2. Load four tubes at positions:
3.
Load six tubes at positions:
4.
Loading
recommended.
in
the following manner:
9
and
18.
1,
5 and 10,
7, 3 and 16,
1,
9, 5 and 14, 18,
7, 9, 3 and 12, 18,
an
14
or
12
or
odd number
10
16
of
tubes is not
Caution:
Do
not exceed maximum rated speed for each
rotor/accessory combination. Maximum rated speeds can
in
be found
Section 4.2 - Speed
And
Force
Tables.
I
r -
Centra-CL3 Series Operation Manual
14
Loading...
+ 34 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.