IDEAL INDUSTRIES 61-764, 61-766, 61-768 Technical Manual

Page 1
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-764
61-766 61-768
The Service Information provides the following information:
• Precautions and safety information
• Performance test procedure
• Calibration and calibration adjustment procedure
• Basic maintenance (replacing the battery)
Form number: TM61764-6-8 Revision: 5. Date: Nov 2007
Form number TM61764-6-8 Rev 5 November 2007
Page 2
TABLE OF CONTENTS
Title
Page Introduction 1 Precautions and Safety Information 1 Symbol Table 1 TightSight™ Display Notes 1 High Voltage Warning (HI-V) 2 Safety Information 2 Certifications and Compliance 3 Specifications 3 General Specification 3 Voltage Specifications 4 Current Specifications 4 Resistance Specifications 4 Frequency Specification 4 Capacitance Specifications 4 Continuity Specifications 4 Performance Verification 5/6 Calibration 7/8 Calibration Reference Points 9/10 Replacing the Battery 11
Form number TM61764-6-8 Rev 5 November 2007
Page 3
Introduction
Warning
To avoid shock or injury, do not perform the verification
tests or calibration procedures described in this manual
unless you are qualified to do so.
The information provided in this document is for the use
of qualified personnel only.
Caution
The 61-760 series contains parts that can be damaged by static
discharge. Follow the standard practices for handling static sensitive
devices.
For additional information about IDEAL INDUSTRIES, INC. and its products,
and services, visit IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
Precautions and Safety Information
Use the meter only as described in the Users Manual. If you do not do so, the protection provided by the meter may be impaired. Read the “Safety Information” page before servicing this product. In this manual, a Warning identifies conditions and actions that pose hazard (s) to the user; a Caution identifies conditions and actions that may damage the meter or the test instruments.
The Symbols
The symbols used on the meter and in this manual are explained in Table A.
Table A Symbols
Symbol Description Symbol Description
High Voltage Indicator >30 V indicator is on
Continuity indicator
CAT III CAT IV
Battery Cautionary or important
information in manual Danger- Risk of electrical shock Double Insulation- Protection Class II IEC Over-voltage Category III IEC Over-voltage Category IV
HI-V
TightSight™ Display Notes:
Only AC/DC amps units of measure are displayed in the TightSight™ display since primary use is for viewing current measurements in tight locations. The display will show n umerical values only for other functions. The main display is to be used to view units of measure for all other functions.
Page 1
Form number TM61764-6-8 Rev 5 November 2007
Page 4
Page 2
High Voltage Warning (HI-V):
The meter beeps and lights an LED when >30V AC/DC voltage is present through test leads of the meter. This enhanced safety feature alerts the user that dangerous voltage is present across the leads even if the meter is set on an incorrect function or range. Notes: This feature does not work through the clamp head as the clamp is intended to only measure current. Audible indication can be turned off by sliding the switch in battery compartment.
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use the product only as specified.
CAUTION.
These statements identify conditions or practices that could result in damage to the equipment or other property.
WARNING.
These statements identify conditions or practices that could result in personal injury or loss of life.
Specific precautions
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without the covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or damp conditions.
Form number TM61764-6-8 Rev 5 November 2007
Page 5
Certifications and Compliances
Safety
Designed to EN 61010-1, EN 61010-2-032, UL 61010B-1,
UL 61010B-2-032 specifications
1000V DC Category III
Input rating
600V DC Category IV
750V AC Category III
600V AC Category IV
Over voltage category CAT IV: Outside and service entrance.
CAT III: Distribution level mains, fixed installation. C AT II: Local level mains, appliances, and port able equipment.
CAT I: Signal level, special equipment or parts of equipment, telecommunication, electronics.
General Specifications
Characteristics Description Display 3¾ Digit LCD display Display Count 6600 count, maximum reading 6600 Over range Indication “OL” is displayed Sampling Rate 2.0 time/second Operating Relative Humidity Storage Environment:
0°C to 50°C (32°F to 122°F) 0 ~ 70% RH
-20°C to 60°C (-4°F to 140°F) at <80% relative humidity Power source: 9V Battery (NEDA 1604) Battery Life: 200 hours typical (alkaline) {61-764}
200 hours typical (alkaline) {61-766} 150 hours typical (alkaline) {61-768}
Low Battery Indicator:
symbol indicates low battery voltage Auto power off Approximately 30 minutes Dimensions 8.7” H X 3.1 ” W X 1.8 ” D
222mm H X 80mm W X 45.5mm D Maximum Cable Size ACA 1¼” (36mm) Weight: Approximately 12.4 oz. or 350g including battery
RANGES and ACCURACY SPECIFICATION Accuracy: Accuracy specifications at 23°C ±5°C (73.4°F ±9°F) less than 75% RH. Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F) Electrical Specification: Accuracy are ±(reading plus number of digits) at 23°C ±5°C (73.4°F ±9°F) <75% RH
Page 3
Form number TM61764-6-8 Rev 5 November 2007
Page 6
Page 4 61-764
Function / Range
Ranges Accuracy
660.0mV, 50Hz - 60Hz 1.7% + 8digits
AC Voltage
6.600V/66.00V/660.0V, 50Hz - 100Hz 1.2% + 8 digits
6.600V/66.00V/660.0V, 100Hz - 400Hz 1.5% + 8 digits 750V, 50Hz - 400Hz 1.5% + 8 digits
DC Voltage
AC Current
Resistance
Continuity
660.0mV 1.0% + 2 digits
6.600V/66.00V/660.0V/1000V 0.5% + 2 digits
66.00A/660.0A, 50Hz - 60Hz 1.7% + 8 digits
66.00A/660.0A, 60Hz - 400Hz 3.0% + 8 digits
660.0Ω/6.600KΩ/66.00KΩ/400.0KΩ
4.000MΩ
10.00MΩ <40Ω on
Continuity
1.0% + 4 digits
5.0% + 4 digits
12.0% + 5 digits Not Specified
61-766
Function / Range
Ranges Accuracy
660.0mV, 50Hz - 60Hz 1.7% + 8 digits
AC Voltage (True RMS)
6.600V/66.00V/660.0V, 50Hz - 100Hz 1.2% + 8 digits
6.600V/66.00V/660.0V, 100Hz - 400Hz 1.5% + 8 digits 750V, 50Hz - 400Hz 1.5% + 8 digits
DC Voltage
660.0mV 1.0% + 2 digits
6.600V/66.00V/660.0V/1000V 0.5% + 2 digits
66.00A/660.0A, 50Hz - 60Hz 1.7% + 10 digits AC Current
(True RMS)
66.00A/660.0A, 60Hz - 400Hz 3.0% + 10 digits Capacitance 6.600µ/66.00µ/660.0µ F 3.0% + 10 digits Frequency 20Hz ~ 400Hz (Auto-ranging) 0.1% + 3 digits
1.0% + 4 digits
5.0% + 4 digits
12.0% + 5 digits Not specified
Resistance
Continuity
660.0Ω/6.600KΩ/66.00KΩ/400.0KΩ
4.000MΩ
10.00MΩ
<40Ω on
Continuity
61-768
Function / Range
Ranges Accuracy
660.0mV, 50Hz - 60Hz 1.7% + 8 digits AC Voltage
(True RMS)
6.600V/66.00V/660.0V, 50Hz - 100Hz 1.2% + 8 digits
6.600V/66.00V/660.0V, 100Hz - 400Hz 1.5% + 8 digits
750V, 50Hz - 400Hz 1.5% + 8 digits
DC Voltage
660.0mV 1.0% + 2 digits
6.600V/66.00V/660.0V/1000V 0.5% + 2 digits
660.0A, 50Hz - 60Hz 2.0% + 10 digits AC Current (True RMS)
660.0A, 60Hz - 400Hz 3.0% + 10 digits DC Current 660.0A 2.0% + 5 digits
1.0% + 4 digits
5.0% + 4 digits
12.0% + 5 digits Not specified
Resistance
Continuity
660.0Ω/6.600KΩ/66.00KΩ/400.0KΩ
4.000MΩ
10.00MΩ
<40Ω on
Continuity
Form number TM61764-6-8 Rev 5 November 2007
Page 7
Page 5
AC Converter: 61-764 - Average responding, RMS Calibrated to Sine Wave 61-766, 61-768 – True RMS sensing Overload Protection:
AC and DC Voltage: Not to exceed 1000V DC or 750VAC RMS
AC Current: Not to exceed 660A AC DC Current: Not to exceed 660A DC
Resistance: Not to exceed 600V DC or VAC RMS Capacitance, Frequency, Continuity: Not to exceed 600V DC or VAC RMS
PERFORMANCE VERIFICATIONS
Perform the following analysis; if the meter conforms to the limits listed in Table 1 through 7 the meter is functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure must be performed.
Performance Verification Preparation
1. Turn on the calibrator, allow calibrator to warm up. Temperature stabilization should be reached after 30 minutes.
2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5V DC. If the battery measures under 7.5V DC, replace the battery (see Battery Replacement page 11) before beginning the performance test.
3. Input the values listed in Table 1 through 7.
Table 1 AC Voltage Test
Function /Range
Input
Low Limit
High Limit
Model Number
V AC 660mV 350mV AC @ 50Hz 343.2 356.8 61-764, 61-766, 61-768 V AC 6.6V 3.5V AC @ 50Hz 3.450 3.550 61-764, 61-766, 71-768 V AC 6.6V 3.5V AC @ 400Hz 3.439 3.561 61-764, 61-766, 71-768 V AC 66V 35V AC @ 50Hz 34.50 35.50 61-764, 61-766, 61-768 V AC 66V 35V AC @ 400Hz 34.39 35.61 61-764, 61-766, 61-768 V AC 660V 350V AC @ 50Hz 345.0 355.0 61-764, 61-766, 61-768 V AC 660V 350V AC @ 400Hz 343.9 356.1 61-764, 61-766, 61-768 V AC 750V 700V AC @ 50Hz 681 719 61-764, 61-766, 61-768 V AC 750V 700V AC @ 400Hz 681 719 61-764, 61-766, 61-768
Table 2 DC Voltage Test
Function /Range
Input
Low Limit
High Limit
Model Number
V DC 660mV 350mV 346.3 353.7 61-764, 61-766, 61-768 V DC 6.6V 3.5V 3.480 3.520 61-764, 61-766, 61-768 V DC 66V 35V 34.80 35.20 61-764, 61-766, 61-768 V DC 660V 350V 348.0 352.0 61-764, 61-766, 61-768 V DC 1000V 900V 893 907 61-764, 61-766, 61-768
Form number TM61764-6-8 Rev 5 November 2007
Page 8
Page 6
Table 3 AC Current Test
Function /Range
Input
Low Limit
High Limit
Model Number
A AC 66A 10A AC @ 50Hz 9.75 10.25 61-764 A AC 66A 10A AC @ 50Hz 9.73 10.27 61-766 A AC 66A 10A AC @ 400Hz 9.62 10.38 61-764 A AC 66A 10A AC @ 400Hz 9.60 10.40 61-766 A AC 660A 500A AC @ 50Hz 490.7 509.3 61-764 A AC 660A 500A AC @ 50Hz 490.5 509.5 61-766 A AC 660A 500A AC @ 50Hz 489.0 511.0 61-768 A AC 660A 500A AC @ 400Hz 484.2 515.8 61-764 A AC 660A 500A AC @ 400Hz 484.0 516.0 61-766, 61-768
Table 4 DC Current Test
Function /Range
Input
Low Limit
High Limit
Model Number
A DC 660A 100A DC 97.5 102.5 61-768 A DC 660A 500A DC 489.5 510.5 61-768
Table 5 Resistance Test
Function /Range Input
Ω 660 100Ω Ω 6.6K 1KΩ Ω 66K 10KΩ Ω 400K 100KΩ Ω 4M 1MΩ Ω 10M 10MΩ
Low Limit
98.6 101.4 61-764, 61-766, 61-768 .986 1.014 61-764, 61-766, 61-768
9.86 10.14 61-764, 61-766, 61-768
98.6 101.4 61-764, 61-766, 61-768 .946 1.054 61-764, 61-766, 61-768
8.75 112.5 61-764, 61-766, 61-768
High Limit
Model Number
Table 6 Capacitance and Frequency Test
Function /Range Input
Hz
50Hz @ 5A 49.6 50.4 61-766
Low Limit
High Limit
Model Number
{through clamp} Hz {auto}
200Hz @ 10A 199.5 200.5 61-766 {through clamp} MFD 6.6µF 1µF .960 1.040 61-766 MFD 66µF 10µF 9.60 10.40 61-766 MFD 660µF 100µF 96.0 104.0 61-766
Table 7 Continuity Check
Function /Range
Continuity
Test Value
20Ω beep on 40Ω beep off
Low limits
High Limit
19.4 20.6
39.2 40.8
Model Number
61-764, 61-766, 61-768
Form number TM61764-6-8 Rev 5 November 2007
Page 9
Page 7
CALIBRATION
Calibration Preparation
1. Turn on the calibrator, allow calibrator to warm up. Perform calibration at 23±2°C (73.4°F ±3.5°F) at relative humidity of < 70%. Temperature stabilization should be
reached after 30 minutes.
2. Disco nnect the test leads and turn the range switch to “OFF”.
3. Remove the screws holding the battery cover, one at the jaw, and the screw for the TightSight™
cover.
4. Remove the case bottom using care not to damage the leads of battery snap and spring
to the continuity beeper. (Beeper is attached to the bottom case cover.)
5. Using a calibrated meter ensure the battery measures a minimum of 7.5V DC.
If the battery measures under 7.5V DC, replace the battery (see Battery Replacement page 11).
Calibration Procedure
It is recommended that all IDEAL meters undergo the following calibration procedure on an annual basis. The class of calibrator or equipment should have an accuracy that exceeds, by an expectable ratio the accuracy of this instrument.
V DC Calibration: 61-764 (Refer to Figure 1A), 61-766 (Refer to Figure 2A), 61-768 (Refer to Figure 3A)
1. Set the function / range to 6.6V DC.
2. Connect th e calibrator to the V and COM inputs on the meter.
3. Output 3.900V DC.
Adjust VR1 (VR 1KΩ) until unit display reads 3.900. (61-764, 61-766)
Adjust VR2 (VR 1KΩ) until unit display reads 3.900. (61-768)
4. De-energize source and remove test leads
V AC Zero Calibration: 61-766 (Refer to Figure2A), 61-768 (Refer to Figure 3A)
1. Set the function /range to 750V AC.
2. Short the V and COM input on the meter.
3. Adjust VR3 (VR 220kΩ) until display reads 000.
4. De-energize source and remove test leads.
V AC Calibration: 61-766 (Refer to Figure 2A), 61-768 (Refer to Figure 3A)
1. Set the function/range to the 6.6V AC.
2. Connect the calibrator to th e V and COM inputs on the meter.
3. Output 3.900VAC/60Hz.
Adjust VR2 (VR 1KΩ) until unit display reads 3.895 ± 1 digit. (61-766) Adjust VR1 (VR 1KΩ) until unit display reads 3.895 ± 1 digit. (61-768)
4. De-energize source and remove test leads.
Form number TM61764-6-8 Rev 5 November 2007
Page 10
Page 8
Calibration Procedure (cont’d)
A AC Calibration: 61-764 (Refer to Figure 1B), 61-766 (Refer to Figure 2B) (Adjustments made under front panel label.)
1. Set the function / range to the 660A AC.
2. Set output of the AC calibrator for 10.00A/60Hz +/- 0.01% and connect it to Coil = 10N = 100.0A
AC.
3. Clamp the jaws to the coil = 10N.
4. Adjust VR2 (VR 1KΩ) for a display reading of 100.0 ±1 digit. (61-764)
Adjust VR4 (VR 2KΩ) for a display reading of 100.0 ±1 digit. (61-766)
5. De-energize source and remove test leads.
A DC Zero Calibration 61-768 (Refer to Figure 3B)
1. Set the function/range to A DC.
2. Short the V and COM inputs on the meter.
3. Adjust VR5 (VR 5KΩ) until display reads 000.
4. De-energize source and remove test leads.
A DC Calibration 61-768 (Refer to Figure 3B)
1. Set the function / range to the A DC.
2. Set output of the DC calibrator for +10.00A ± 0.01% and connect it to Coil = 10N = +100.0A DC.
3. Clamp the j aws to the coil = 10N.
4. Adjust VR4 (VR 3.3KΩ) for a display reading of 100.0 ±1 digit.
5. De-energize source and remove test leads.
Calibration of the 61-760 series is complete. Remove all leads from the calibrator and equipment. Return unit to proper operating condition.
Form number TM61764-6-8 Rev 5 November 2007
Page 11
Page 9
61-764 (Figure 1A) 61-764 (Figure 1B)
61- 766 (Figure 2A) 61-766 (Figure 2B)
Form number TM61764-6-8 Rev 5 November 2007
Page 12
Page 10
61-768 (Figure 3A) 61-768 (Figure 3B)
Form number TM61764-6-8 Rev 5 November 2007
Page 13
Battery Replacement (Refer to Figure 4)
1. Disconnect the test leads from any circuit under test and t urn off m eter.
2. Use a Philips head screwdriver to remove the screws on battery cover.
3. Remove battery from the battery compartment.
4. Install new 9V battery (NEDA #1604). An alkaline type is recommended.
5. Install new battery into compartment using care to install to proper polarity.
6. Reinstall battery cover.
Page 11
Figure 4
Form number TM61764-6-8 Rev 5 November 2007