Maintenance and Service ....................................................16
Disposal and Warranty ........................................................17
3
Introduction
The IDEAL 61-637 and 61-627 Single Range Non-Contact Voltage Testers
automatically detect and indicate the presence of voltage within specified ranges.
The 61-637 detects voltage within a range of 24-600V AC and the 61-627 detects
voltage within a range of 50-600V AC. Both testers provide non-contact detection
of voltage through an antenna enclosed in plastic at the tip of the tester which
senses the presence of the electromagnetic field present around live conductors.
The 61-637 has a flash light that works independently of the testing function and is
activated by an independent On/Off button.
Arc Flash and Shock Hazard, Proper PPE Required. Follow all safety procedures,
wear proper PPE in accordance to NFPA 70E. Read and fully understand the
instruction manual prior to using this product. Failure to comply can result in
serious injury or death.
4
Safety Information
possible death or serious injury if the hazard is realized.
Caution - Identifies conditions and actions that could result in tester dam-
age, equipment under test damage or data loss if the hazard is realized.
Warning - Identifies conditions and actions that could result in
WARNING
Arc Flash and Shock Hazard, Proper PPE Required. Follow all safety procedures,
wear proper PPE in accordance to NFPA 70E and follow the guidelines below and
the instructions in this manual when operating the tester. Failure to comply can
result in serious injury or death.
•Choking Hazard, Small Parts. Keep Away from Children Sharp Objects
Hazard, This is not a toy. It is not for use or play by children. Keep
Away from Children. Failure to do so can result in serious injury.
• Only experienced or technically competent consumers should use this
equipment. When in doubt, call an experienced electrician to make any and
all necessary repairs or installations. At all times, perform any necessary
work on a de-energized circuit that has had its circuit breaker turned off and
has been locked out.
• Use the Tester only as specified in this manual or protection provided by the
Tester can be compromised.
• Before using the Tester, visually inspect it to ensure the case or tip are not
cracked and the battery cap is securely in place. Do not use if the Tester
appears damaged.
• When using the tester, keep fingers behind the tactile barrier at the base of
the tester. (see #4 on page 9)
• This Tester is intended for use by qualified electricians and competent
consumers. Follow NFPA 70E Standards for Electrical Safety when using
this Tester.
• Do not use without the batteries correctly in place and the slide on battery
cover secured.
• Do not use Tester if it operates incorrectly as protection may be
compromised. When in doubt, have the Tester serviced.
5
WARNING
Arc Flash and Shock Hazard, Proper PPE Required. Follow all safety procedures,
wear proper PPE in accordance to NFPA 70E and follow the guidelines below and
the instructions in this manual when operating the tester. Failure to comply can
result in serious injury or death.
• Have the Tester serviced only by qualified service personnel.
• Do not use the Tester around explosive gas, dust, or vapor, or during electrical
storms, or in wet environments.
• Do not apply more than the rated voltage, as marked on the Tester. Observe
CAT Rating of Tester when in use.
• To avoid false readings that can lead to electrical shock and injury, replace the
batteries as soon as the low battery indication appears.
• Voltages exceeding 30VAC or 60VDC pose a shock hazard so use caution.
• Do not work alone so that assistance can be rendered in an emergency.
Use extreme caution when working around bare conductors or bus bars.
•
Contact with the conductor could result in electric shock.
• Adhere to local and national safety codes. Individual protective equipment
must be used to prevent shock and arc blast injury where hazardous live
conductors are exposed.
• Cancer and Reproductive Harm
CAUTION
Tester damage can occur if the following guidelines are not adhered to.
• Use the proper settings and voltage ratings for the measurement application.
• Clean the case and accessories with a damp cloth and mild detergents only.
Do not use abrasives or solvents. Make sure the tester is completely dry
before use.
• Not intended for insertion into electrical outlet slots.
• Do not apply excessive side or torsional loading to the sensing tip as
damage may occur.
6
Symbols & Descriptions
SYMBOLDESCRIPTION
Arc Flash and Shock Hazard
Shock Hazard
Warning or Caution
Choking Hazard
AC (Alternating Current)
Earth Ground
CAT III
CAT IV
IEC Measurement Category III
CAT III has protection against transients in equipment in fixedequipment installations such as distribution panels feeders,
and short branch circuits. Also included are lighting systems in
larger buildings.
IEC Measurement Category IV
CAT IV has protection against transients from the primary
supply level such as a Meter or overhead or underground utility
service.
Voltage AC
V
Double Insulation
Do not dispose of this product as unsorted municipal waste.
It must be properly disposed of in accordance with local
regulations.
Conforms to applicable North American Safety Standards
Conforms to applicable Australian Safety Standards
Conforms to European Directives
NOTE: Tester must be used within the designated CAT Rating.
7
Operation
Identification and Description of Operating
Functions for the 61-637 Single Range 24
to 600V AC NCVT w/ Flashlight and 61-627
Single Range 50 to 600V AC NCVT:
The tester can be operated with a visual only indication of voltage when working in
noise sensitive/noise restricted areas. With the tester On, press the Speaker button
for less than 1s. The tester will beep once to indicate the audible warning is muted
and only a visual indication will be provided. To restore the audible warning, press
the speaker button for less than 1s. The tester will beep twice to indicate the audible
warning is active and both a visual and audible indication will be provided. The
tester is set to Audible Notification On as the default setting.
Flashlight (61-637 Only)
An LED Flashlight in the tip operates independently of the tester functions and can
be used to illuminate the work area. Press the Flashlight Button to turn it On/Off.
The tester is set to Flashlight Off. The Flashlight will Auto Power Off after 5 minutes
of no voltage detection. Further voltage detection will reset the APO function to 5
minutes again.
Note: Use of the flashlight will decrease battery life.
Auto Power Off
After 10 minutes of inactivity (no button pressed and no signals detected), the
tester will emit one long beep (~0.5s) and then shut off. Note: if the flashlight is on,
it will be also shut off.
Low Battery Indication
If battery voltage is less than around 2.4V, the green led in the tip will flash 3 times
and the buzzer will send out one long beep, then turn off the device.
10
Tester Operation
Turning The Tester On
Press the Power On/Off button for less than 1s to turn the tester On. The tester
will beep twice, and will light a steady Green LED in the tip of the tester to indicate
Power On.
Steady Green LED in the tip
indicates power-on
Turning The Tester Off
Press the Power On/Off button for more than 1s to turn the tester Off. Listen for a
long steady beep (~.5s) and watch for LED in the tip to turn off. The tester is now
deactivated and is not operational.
11
Checking for The Presence of AC Voltage
Prior to use, test on known live circuit to verify tester functionality. Place tip of the
tester near an AC voltage source and refer to the NCV Sensing Indications Table for
tester indication during operation.
Flashing Red LED in the tip and
intermittent beeping indicate
voltage is present. (See NCV
Sensing Indications table for
indications associated with each
range and mode setting)
While the NCV is a helpful function, it is ALWAYS RECOMMENDED that the
operator verify that any electrical conductor is completely deenergized and that
no voltage is present by measuring for voltage AND CONFIRMING THAT NO
VOLTAGE IS PRESENT and that all applicable PPE and lock out tag out procedures
be followed before attempting any work on ANY electrical distribution system.
Voltages with frequencies higher than 60Hz or electrostatic charges may also be
detected by the NCV sensing antenna. Due to the variability of designs across
device manufactures the 61-627 and 61-637 may not detect the presence of voltage
in Tamper Resistant Outlets.
12
61-637 NCV Sensing Indications Table
NCV
Indications
Mode
18-25V26-34V35-45V46-54V55-600V
Ranges (V AC)
Red
Normal Mode
(24-600V AC)
Audible
Notification
Disabled
(24-600V AC)
Note: Indications above may be subject to variability caused by humidity, insulation
thickness, distance, and other factors.
Flashing
Light in Tip
and beeps
@ 2Hz
Red
Flashing
Light in Tip
@ 2Hz, No
beeps
Red
Flashing
Light in
Tip and
beeps
from 2Hz
to 4Hz
Red
Flashing
Light in
Tip from
2Hz to
4Hz, No
beeps
Red
Flashing
Light in Tip
and beeps
@ 4Hz
Red
Flashing
Light in Tip
@ 4Hz, No
beeps
Red
Flashing
Light in Tip
and beeps
from 4Hz
to 12 Hz
Red
Flashing
Light in Tip
from 4Hz
to 12Hz,
No beeps
Red Flashing
Light in Tip
and beeps
@ 12Hz
Red Flashing
Light in Tip
@ 12Hz, No
beeps
61-627 NCV Sensing Indications Table
NCV
Indications
Mode
Normal Mode
(50-600V AC)
Note: Indications above may be subject to variability caused by humidity, insulation
thickness, distance, and other factors.
Ranges (V AC)
50-600V AC
Red Flashing Light in Tip and beeps @ 12Hz
Functions Operation Table
IndicationResponseDefault FunctionOperation
On OffMeter OnGREEN LEDPush Button
Audible
(61-637 Only)
Flashlight
(61-637 Only)
BeepsOnPush Button
Flashlight OnOffPush Button
13
Functions Indication Table
FunctionDescription
Power OnSteady Greenlight in Tip,
Power OffNo Light, Long Beep
Tester Auto Power OffGreen Light Flashes, 1 Beep
Audible OFF
(61-637 Only)
Audible ON
(61-637 Only)
Flashlight Button
(61-637 Only)
2 Short Beeps
One Beep
Two Beeps
Flashlight On or Off
Electrical Specifications
FunctionRange
AC Voltage Sensing 61-63724V AC to 600V AC
AC Voltage Sensing 61-62750V AC to 600V AC
Frequency50Hz or 60Hz AC
Environmental Specifications
Operating Temperature:32ºF to 104ºF (0ºC to 40ºC) (<80%RH)
Operating Altitude:
Storage Temperature:
Intended for indoor use.
6500 ft (2000 m)
14ºF to 122ºF (-10ºC to 50ºC) (<80%RH)
14
Mechanical Specifications
Dimensions: (L x W x H) 6.0 in. x 1.0 in. x 1.1 in.
Weight:0.13 LBS (0.06 KG)
Power Source:(2) 1.5V AAA Batteries
Flashlight:
(61-637)
(154 mm x 26 mm x 29 mm)
>500LUX @ 4 in
EMC/EMI
CISPR 22 3rd Edition. Class B Limits.
EN 55032
CISPR 32
CISPR 11
FCC 15. 107 with reference to Section 15.109 (g).
ICES-003
EN 61326-2-2 Sec 6.4.2.101
USA (FCC)
47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Safety
Complies with the following:
IEC 61010-1, Edition 3 (2010-06) +AMD1 (2016-12)
ANSI/UL/IEC EN 61010-1, Edition 3 (2016-04-29)
CSA-C22.2 No. 61010-1-12 Edition 3 (2019-07-19)
IEC 60529
Overvoltage CAT IV 600V.
Any voltages exceeding the defined maximum voltage measurement categories
described above are outside the normal use of the equipment and protection cannot
be guaranteed.
Pollution Degree Class 2
15
Maintenance and Service
Tester Inspection
Before using the Tester, visually inspect it to ensure the case or tip are not cracked
and the battery cover is securely locked in place. Do not use if the Tester appears
damaged.
Battery Inspection/Replacement
Maintenance and Storage
When not in use, turn the meter off. If the tester will not be used
for more than a month remove the batteries for storage. Clean the
case with a damp cloth and mild detergent. Do not use abrasives or
solvents. If the meter is exposed to water ensure the tester is
completely dry before use. Do not expose to chemicals or solvents.
Service and Replacement Parts
This unit has no user-serviceable parts.
16
Disposal of Waste, Electrical & Electronic Equipment
In order to preserve, protect and improve the quality of the environment, protect
human health and utilize natural resources prudently and rationally, the user should
return unserviceable product to relevant facilities in accordance with statutory
regulations. The crossed-out wheeled bin indicates the product needs to be
disposed separately and not as municipal waste.
Do not dispose of this product as unsorted municipal waste. It must be properly
disposed of in accordance with local regulations.
Disposal of Used Batteries/Accumulators
The user is legally obliged to return used batteries and accumulators. Disposing
used batteries in household waste is prohibited! Batteries/accumulators containing
hazardous substances are marked with the crossed-out wheeled bin. The symbol
indicates that the product is forbidden to be disposed via domestic refuse. The
chemical symbols for the respective hazardous substances are Cd = Cadmium, Hg
= Mercury, Pb = Lead.
You can return used batteries/accumulators free of charge to any collecting point
your local authority, our stores, or where batteries/accumulators are sold.
Consequently, you must comply with your legal obligations and contribute to
environmental protection.
TWO YEAR LIMITED WARRANTY
This tester is warranted to the original purchaser against defects in material and
workmanship for a period of two (2) years from date of purchase. With proof of
purchase from an authorized IDEAL distributor, a defective tester will be repaired or
replaced with the same product or a functionally equivalent product, at the option of
IDEAL INDUSTRIES, INC. during the warranty period, subject to verification
of the defect or malfunction. Warranty does not cover consumables such as fuses,
batteries, and excludes defects caused by leakage from batteries, abuse,
mishandling, dropping, ordinary wear and tear, misuse, neglect, unauthorized
repair, improper use, alterations, accidents or any causes beyond IDEAL’s
reasonable control. Consequential or incidental damages are not recoverable under
this warranty. Some states do not allow the exclusion or limitation of incidental or
consequential damages, so the above limitation or exclusion may not apply to you.
This LIMITED WARRANTY gives you specific legal rights, which vary from state to
state. This warranty constitutes the sole and exclusive remedy of the purchaser and
the exclusive liability of IDEAL, and is in lieu of any and all other warranties, and
expressly disclaims all other warranties, implied, or statutory as to merchantability,
fitness for purpose sold, description, quality productiveness, or any other matter.
No agent, distributor or other supplier has the authority to modify or amend this
warranty or make other representations or warranties other than those contained in
this warranty without express written authorization from IDEAL. For warranty
service, call IDEAL customer service at 1-800-435-0705.
Made in China.
of
17
Loading...
+ 37 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.