AC Voltage (V~)
#61-340 
#61-342
Digital Multimeter 
Instruction Manual
Read First: Safety Information
Understand and follow operating instructions 
carefully. If this tester is not used in a manner 
specified by IDEAL, protection provided by the 
product may be impaired.
WARNINGS
To avoid possible electric shock, personal injury or 
death follow these instructions: 
• Do not use if meter appears damaged. 
• Visually inspect the meter to ensure case is not 
cracked and back case is securely in place.
• Inspect and replace leads if insulation is damaged, metal is exposed, or probes are cracked. 
Pay particular attention to the insulation surrounding the connector.
• Do not use meter if it operates abnormally as 
protection maybe impaired. 
• Do not use during electrical storms or in wet 
weather.
• Do not use around explosive gas, dust, or vapor.
• Do not apply more than the rated voltage to the 
meter. 
• Do not use without the battery and the back case 
properly installed.
• Remove the test leads from the meter before 
removing battery cap.
• Do not attempt to repair this unit as it has no 
user-serviceable parts.
• Disconnect power and discharge capacitors 
before testing resistance, continuity, diodes, 
capacitance or temperature.
• Replace battery as soon as low battery indicator  
appears to avoid false readings.
• Use the proper terminals, function and range for 
your measurements.
• Comply with local and national safety 
requirements, including the use of appropriate 
personal protective equipment.
CAUTION:
To protect yourself, think "Safety First":
• Voltages exceeding 30VAC or 60VDC pose a 
shock hazard so use caution.
• Use appropriate personal protective equipment 
such as safety glasses, face shields, insulating 
gloves, insulating boots, and/or insulating mats.
• Before each use:
- Perform a continuity test by touching the 
test leads together to verify the functionality 
of the battery and test leads.
- Use the 3 Point Safety Method. (1) Verify 
meter operation by measuring a known 
voltage. (2) Apply meter to circuit under 
test. (3) Return to the known live voltage 
again to ensure proper operation.
• Never ground yourself when taking electrical 
measurements.
• Connect the black common lead to ground or 
neutral before applying the red test lead to 
potential voltage. Disconnect the red test lead 
from the voltage first.
• Always work with a partner.
• When using the probes, keep fingers as far 
behind the probe tips as possible.
Ranges & Accuracies: 
AC Converter: 61-340 model is averaging
sensing, rms calibrated; 61-342 model is true rms 
sensing. 
Accuracy: Accuracy is specified as +/-(a 
percentage of the reading + a fixed amount) at 
23°C±5°C (73.4°F ± 9°F), less than 75% relative 
humidity. 
Temperature Coefficient: 0.1 times the 
applicable accuracy specification from 32°F to 
64°F and 82°F to 122°F (0°C to 18°C ; 28°C to 
50°C).
Ranges & Accuracies
DC Voltage (V )
Toggle for 
Frequency 
(>20% of 
range)
Function Range & Resolution Accuracy Overload Protection
DC Voltage
400.0m/4.000/40.00/400.0 V ±(0.5%+5)
600.0V ±(1.0%+3)
AC Voltage
400.0m/4.000/40.00/400.0/600 V ±(1.5%+5)
(40~400Hz)
DC Current
400.0/4000 µA ; 40.00/400.0 mA ±(0.5%+5) 500mA/600V Fast Fuse
4.000/10.00 A ±(1.5%+5) 10A/1000V Fast Fuse
AC Current
400.0/4000 µA ±(1.5%+5) 
500mA/600V Fast Fuse
(40~400Hz)
40.00/400.0 mA ±(2.0%+5) 
4.000/10.00 A ±(2.5%+5)  10A/1000V Fast Fuse
400.0Ω ±(1.2%+5) 
Resistance
4.000k/40.00k/400.0k Ω ±(1.0%+2)
600V DC/AC rms
4.000MΩ ±(1.2%+2)
40.00MΩ ±(2.0%+5)
40.00nF ±(3.0%+10)
Capacitance* 400.0n/4.000µ/40.00µF ±(3.0%+5)  600V DC/AC rms
400.0µ/4000µ F ±(20%+5) 
Frequency 10.00/100.0/1.000k/10.00k/100.0k/1.000M/10.00M Hz
±(0.1%+3) 
600V DC/AC rms
Sensitivity: <1MHz: 0.7Vrms ; >1MHz: 5Vrms 
Duty Cycle 0.1 - 99.9% ±(2.5%+5)  (<10kHz)
Diode Check Test current : (1±0.6) mA and then open circuit voltage is 2.5VDC typical. 600V DC/AC rms
Continuity The beeper turns on <25Ω and turns off at >120Ω. 600V DC/AC rms
Temperature**
-58~1500°F
±(3.0%+5)
—
-50~800°C
±(3.0%+3)
*Accuracy not available for <10nF capacitance.  **Accuracy is stated for meter only. Thermocouple accuracy adds another ±2.5% to reading.
Input Impedance : 10MΩ for VAC &VDC.       CF > 2, add ±1.0% to accuracy.
900VDC
750VAC rms