ICU Medical 20678-04-01 Users Manual

Plum A+® Infusion Systems
Because the Dose falls outside the soft limits, the infuser beeps three times (Invalid Key Warning), and generates the error message:
“Dose > 8. Override?”
13 Select
[Yes].
Program Dose Calc
A
DOPamine
Conc Weight
Dose VTBI Duration Rate
Dose > 8. Override?
Yes
NOTE: A SOFT limit override will be recorded in the History Events Log.
This will bring up the PROGRAM CONFIRMATION screen.
14 If Dose, Drug, Conc., Weight &
Rate appear correct, select 
Program Dose Calc
A
Delivery will be:
Dose
DOPamine
Drug Conc Weight Rate
Confirm Program?
[Yes] to begin program.
NOTE: If the selected drug has been pre-programmed with dosing units and concentration, the concentration drug amount or diluent volume cannot be changed whether delivering or stopped. The only way to change to a different concentration is to clear the program.
Yes No
40070mgkg250 mL
10
mcg/kg/min
250
mL
09: 30
hr : min
26.
3
mL/hr
MedSurg
10
mcg/kg/min
400
mg
250 mL
70
kg
26
mL/hr
.3
No
However, if selected drug has not been pre-programmed (or still operating with the DDL), the concentration can be changed for a program.
430-95597-001 (Rev. 06/05) 5- 7
For Systems With Hospira MedNet
TM
Software
The Main Delivery screen appears displaying the information just entered.
NOTE: The Program Type and Drug are displayed as well as the Dose Rate and Dose Units.
PUMPING STOPPED
A
DOPamine
Dose Calculation
26.
mcg/kg/min
Rate
3
mL/hr Vol Inf
00
mL
10
Dose
MedSurg
B
A
0
Options/
Vol Inf
B
5- 8 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
Dose Calc (mg/min on B)
DOSE CALCULATION- Allows programming Dose
Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep.
Clear ALL settings when changing patients or multiple programming parameters.
1 Select
[B] at the Main
Delivery Screen.
The Line B Programming screen appears.
2 Select
[Drug List].
PUMPING STOPPED
A
DOPamine
Dose Calculation
26.
mcg/kg/min
MedSurg
Mode
Rate
VTBI
Duration
Change using Change Mode
Program
Options
Rate
3
mL/hr Vol Inf
00
mL
10
Dose
MedSurg
B
A
PROGRAM
Piggyback
0 0
00:00
Change
Drug
Mode
List
0
Options/
Vol Inf
mL/hr
mL
hr : min
Cancel/
Back
B
B
430-95597-001 (Rev. 06/05) 5- 9
For Systems With Hospira MedNet
TM
Software
The Drug List Screen appears, in mainly alphabetical order.
NOTE: Default entry is “No Drug Selected”.
3 Use to highlight desired
drug name.
When list does not contain desired generic drug name or group name select
[Page Down] (or
[Page Up]) until it shows,
then use to highlight choice.
MedSurg
No Drug Selected
Albumin 5% Aldesleukin Alfentanil Alteplase Alteplase Amikacin Amiodarone
Select, then Enter
PageUpPage
Down
MedSurg
Labetalol Lepirudin Lidocaine
Lidocaine
Lidocaine Liposyn 10% Liposyn 20% Milrinone
Select, then Enter
PageUpPage
Down
25 grams/500 mL
2.6 mg/50 mL
10 mg/250 mL
50 mg/50 mL
100 mg/100 mL
100 mL
150 mg/100 mL
Cancel/
Enter
Back
200 mg/200 mL
100 mg/250 mL 1 grams/250 mL
2 grams/500 mL
4 grams/500 mL
100 mL 500 mL
200 mcg/100 mL
Cancel/
Enter
Back
B
B
4 Select
[Enter] to indicate selection and advance to
the Program Selection screen.
The Select Program screen for Line B appears.
5 Select
[Choose] to continue.
PROGRAM
Standard Program
Loading Dose
Multistep
Select, then Choose
B
Lidocaine
Choose Back
5- 10 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
If selected drug has been programmed with established dosing units and concentration, skip to step 8.
The Program Dose Calc screen appears.
mL/hr
NOTE: Cursor defaults to package insert (or hospital) recommended dose units associated with the selected drug.
6 Select
[Choose] to continue.
mcg/kg/min mcg/kg/hr mcg/min mcg/hr mg/kg/hr
Select, then Choose
Page
Up
Page
Down
Or, use to change the units to use in the delivery.
If desired, select  [Page Down] to see more units of measure.
The Container Concentration
Program Dose Calc
Selection screen appears whenever the selected Dose Units are gram-based (e.g., grams, mg, mcg, or ng)
Drug Conc in Container
mcg
mg
grams
otherwise this screen is skipped.
Select, then Choose
Choose Back
Lidocaine
mg/min
mg/hr grams/hr ng/kg/min mmol/min mmol/hr mEq/hr
Choose Back
Lidocaine
BProgram Dose Calc
B
NOTE: Cursor defaults to package insert recommended concentration units associated with the selected drug.
7 Select
[Choose] to continue. Or, use to change
the Container Concentration units to use in the delivery.
430-95597-001 (Rev. 06/05) 5- 11
For Systems With Hospira MedNet
TM
Software
The next page of the Program Dose Calc screen appears to provide for entry of delivery program parameters.
8 Use keypad to enter desired
values. Once entered, use
Conc
Mode Piggyback
Dose VTBI Duration Rate
Enter value using keypad
Program Options
500
2
grams mL
00: 00
0 0
0
mg/min mL hr : min mL/hr
to highlight the next field.
NOTE: If selected drug has established Rule Sets (e.g., non­blank concentration), the Mode field will be highlighted upon entry to this screen.
9 To change the Piggyback Delivery Mode default,
scroll down to Mode using , then select 
[Change Mode].
NOTE: Delivery Mode Option is available on Line B, its default is Piggyback, and the Weight field is gone.
Lidocaine
Cancel/
Back
BProgram Dose Calc
10 Continue entering desired
values using the keypad. Use
to advance to the next field.
NOTE: When entering the RATE, with a VTBI value entered, the DOSE and DURATION fields are
Conc 2
Mode Concurrent
Dose VTBI Duration Rate
Enter value using keypad
Program Options
500
grams mL
500
16 : 40
2
30
mg/min mL hr : min mL/hr
Lidocaine
Cancel/
Back
BProgram Dose Calc
automatically updated. To understand the Automatic Calculation feature,
see Examples of Automatic Calculation on page 6-10.
5- 12 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
NOTE: The Duration field temporarily changes to dashes when the computed value is larger than what can be displayed.
11 Continue entering the RATE value, if desired.
NOTE: As Rate is entered, the Duration and Dose are
automatically adjusted as individual numbers are entered.
NOTE: If the Duration value is changed, the Rate and Dose are automatically adjusted.
NOTE: At this point, programming is complete unless
[Program Options] are desired. See Section 6, Additional
Features, for more information.
CAUTION: V
12 To start the delivery, press
This will bring up the PROGRAM CONFIRMATION screen.
13 If Dose, Drug, Conc., Weight, &
ERIFY ALL VALUES BEFORE STARTING INFUSION.
Program Dose Calc
.
Delivery will be:
Dose
Lidocaine
Drug Conc Weight Rate
Confirm Program?
Yes No
70
30
Rate appear correct, select 
[Yes] to begin program.
NOTE: If the selected drug has been pre-programmed with dosing units and concentration, the concentration drug amount or diluent volume cannot be changed whether delivering or stopped. The only way to change to a different concentration is to clear the program.
However, if selected drug has not been pre-programmed (or still operating with the DDL), the concentration can be changed for a program.
2
2
mg/min
grams kg mL/hr
B
MedSurg
500 mL
430-95597-001 (Rev. 06/05) 5- 13
For Systems With Hospira MedNet
Upon , the Main Delivery Screen shows delivery status
for both A and B lines.
NOTE: See Delivery Rate Range in Section 9. The ratio of concurrent
rates has recommendations specified on page 1-7.
TM
Software
STOPPED PUMPING
A
DOPamine
Dose Calculation
26.
3
10.
mcg/kg/min
A
Rate
mL/hr
Vol Inf
mL
Dose
MedSurg
Lidocaine
Dose Calculation
30
1
2
mg/min
Options/
B
Vol Inf
B
5- 14 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
Loading Dose
LOADING DOSE- Allows programming of an initial
infusion rate for a specific volume, followed automatically by a maintenance rate from the same container (e.g., a fluid challenge). If Dose Calc is used, the Loading Dose and Maintenance Dose are in the same unit of measure, over the same period of time (mcg/min), from the same container.
Clear ALL settings when changing patients or multiple programming parameters.
The Area Selection screen will appear if your system has been programmed with the MedNet accessory.
Use to select CCA and
MM/DD/YY XXYYZZ- 123456
ICU
MedSurg OR Cardio Urology Psych Recov L&D 5 West
Select, then Enter
bring you to the Main Delivery Screen.
NOTE: To change CCA while infusing, see Changing CCA While Infusing on page 6-12.
1 Select
[A] (or  [B]) at the
STOPPED STOPPED
A
Main Delivery screen.
Back
Prime
Area Selection
Any Hospital, USA
Renal Neo ICU OB/GYN Oncol Ad ICU ER 7 South 3 North 4 North
Enter
B
Rate
0
mL/hr
Vol Inf
00
mL
MedSurg
A
0
Options/
B
Vol Inf
430-95597-001 (Rev. 06/05) 5- 15
For Systems With Hospira MedNet
TM
Software
The selected Line (A or B) Programming screen appears.
2 Select
[Drug List].
The Drug List Screen appears, in mainly alphabetical order.
NOTE: Default entry is “No Drug Selected”. This process will guide you through a “fluid challenge”, so “No Drug Selected” will be entered.
3 Select
[Enter] to indicate
selection and advance to the Program selection screen.
The Select Program screen for Line A (or B) appears.
A
PROGRAM
Rate
VTBI
Duration
Enter Value using keypad
Program Options
A
No Drug Selected
Albumin 5% Aldesleukin Alfentanil Alteplase Alteplase Amikacin Amiodarone
Select, then Enter
PageUpPage
A
00:00
MedSurg
25 grams/500 mL
Down
PROGRAM
Drug
100 mg/100 mL
150 mg/100 mL
Enter
MedSurg
mL/hr
0
mL
0
hr : min
Cancel/
Back
List
2.6 mg/50 mL
10 mg/250 mL
50 mg/50 mL
100 mL
Cancel/
Back
NOTE: This example shows line A being programmed.
Standard Program
Loading Dose
Multistep
Select, then Choose
Choose Back
5- 16 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
4 Use to scroll down to
Loading Dose.
5 Select
The next page of the Program Loading Dose screen appears.
6 Select
Or, use to change the units to use in the delivery.
If desired, select  [Page
Down] to see more units of measure.
The Loading Dose Programming screen appears.
[Choose] to continue.
[Choose] to continue.
A
Select, then Choose
A
Select, then Choose
A
PROGRAM
Standard Program
Loading Dose
Multistep
Choose Back
Program Loading Dose
mL/hr
mcg/kg/min mcg/kg/hr mcg/min mcg/hr mg/kg/hr
Page
Page Down
Choose Back
Up
Program Loading Dose
mg/min mg/hr grams/hr ng/kg/min mmol/min mmol/hr mEq/hr
Rate
VTBI
Duration
mL
7 The default field is Rate. Since
mL/hr
0
1
0
2
hr : min
0
00 : 00
0
00 : 00
a “Fluid Challenge” is being performed in this example,
press to advance to the
Enter value using keypad
Program
Options
Cancel/
Back
VTBI field.
430-95597-001 (Rev. 06/05) 5- 17
For Systems With Hospira MedNet
TM
Software
8 Enter value using keypad.
9 Use to highlight the next
field.
10 Enter Duration using keypad.
The Rate value is automatically calculated. To understand the Automatic Calculation feature,
see Examples of Automatic Calculation on page 6-10.
Program Loading Dose
A
Rate
VTBI
mL/hr
mL
0
1 2
Enter value using keypad
Program
Options
Program Loading Dose
A
1 2
Enter value using keypad
Program
Options
Rate
mL/hr
500
250
0
0
VTBI
mL
250
0
0
Duration
hr : min
00 : 00 00 : 00
Cancel/
Duration
hr : min
00 : 30 00 : 00
Cancel/
Back
Back
11 After the values for loading
Program Loading Dose
A
dose Step-1 are entered, use
Rate
VTBI
to highlight the
mL/hr
500
1
75
2
mL
250
Duration
hr : min
00 : 30
0
00 : 00
maintenance dose, Step-2.
NOTE: The Loading Dose program has two steps. The first is the loading dose followed by the maintenance dose. These words will appear on the main screen as each step is delivered.
Enter value using keypad
Program
Options
Cancel/
Back
5- 18 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
12 Enter the set of desired values
for Step-2 of the program.
13 Select
select a Delayed Start or Standby.
NOTE: At this point, programming is complete unless
Options] are desired. See Section 6, Additional Features, for more information.
NOTE: Nurse Callback enables an alarm when a change in the delivery on a line occurs such as completing Step-1 and starting Step-2. Or on line B, a piggyback delivery is completed and line A is starting its delivery.
[Program Options] to
[Program
Program Loading Dose
A
Rate
VTBI
mL 250
750
Enter
Duration
hr : min
00 : 30 10 : 00
mL/hr
500
1
75
2
Enter value using keypad
Program
Options
A
Program Options
Loading Dose
Delay Start for 00:00 hr:min
Callback No
Enter Value using keypad
Standby
Cancel/
Back
Cancel/
Back
14 Enter a delay time (less than
24 hours) using the keypad.
A
Program Options
Loading Dose
Delay Start for 00:05 hr:min
NOTE: Not available if a delivery is taking place on the line.
NOTE: Nurse Callback cannot be changed while a delivery is taking
Callback Yes
Change using Yes/No
Enter
Yes/No
Cancel/
Back
place on the line.
15 Select
[Enter] to accept the changes and return to
the Program Loading Dose screen.
430-95597-001 (Rev. 06/05) 5- 19
For Systems With Hospira MedNet
TM
NOTE:  [Cancel/Back] returns the fields to their original value
and displays the previous screen.
16 To start the delivery, press .
Software
This will bring up the PROGRAM CONFIRMATION screen.
CAUTION: VERIFY ALL VALUES
BEFORE STARTING INFUSION.
17 If Drug and Rate appear
correct, select  [Yes] to begin program.
The Main Delivery Screen appears.
NOTE: The words “DELAYED” and “Loading Dose” appear in the upper­left corner. These indicate the options which have been selected.
After the programmed delay period has expired, the Plum A+ starts “Pumping”. This is indicated where “Delayed” previously was displayed.
Program Loading Dose
A
Delivery will be:
Drug Rate VTBI
A
!
A
No Drug Selected
500 250
Confirm Program?
Yes No
DELAYED STOPPED
Loading Dose
500
Loading Dose
500
Rate
mL/hr
Vol Inf
00
mL
MedSurg
A
PUMPING STOPPED
Rate
mL/hr
Vol Inf
00
mL
MedSurg
mL/hr mL
B
MedSurg
!
0
Options/
Vol Inf
0
B
B
A
Vol Inf
Options/
B
5- 20 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
After the Step-1 has completed, the Main Delivery Screen changes to the second step.
NOTE: A flashing “Callback Line A” message and the audio alarm sound indicates activation of that option.
PUMPING
A
Maintenance Dose
75
250 0
MedSurg
N105 Callback Line A
A
18 The user should acknowledge
the Callback message and stop the audio alarm by
pressing .
Upon completion of delivery, the screen shows a flashing “Line A VTBI complete” message and the audible alarm sounds.
NOTE: The KVO rate will change to the rate of 1 mL/hr or less depending on delivery rate and “Pumping” mode changes to “KVO”. If Continue Rate option was selected, pumping continues at original rate.
KVO
A
Maintenance Dose
1
1000 0
MedSurg
N161 Line A VTBI complete
A
STOPPED
Rate
mL/hr
Vol Inf
mL
STOPPED
Rate
mL/hr
Vol Inf
mL
B
0
Options/
B
Vol Inf
B
0
Options/
B
Vol Inf
19 Press to stop alarm sound.
NOTE: Warning message continues to flash and audio alarm
will return after a two minute period of silence, unless delivery is stopped or VTBI of last line is changed.
430-95597-001 (Rev. 06/05) 5- 21
For Systems With Hospira MedNet
TM
Multistep Programming
MULTISTEP- Allows a sequential program to deliver
up to 10 steps; fluid volumes and delivery rates may be programmed for each step. The program may be entered based on Rate and Volume or Volume and Time. If Dose Calc is used, the delivery steps are in the same unit of measure, over the same period of time, from the same container.
Clear ALL settings when changing patients or multiple programming parameters.
Software
The Area Selection screen will appear if your system has been programmed with the MedNet accessory.
Use to select CCA and
MM/DD/YY XXYYZZ- 123456
ICU
MedSurg OR Cardio Urology Psych Recov L&D 5 West
Select, then Enter
bring you to the Main Delivery Screen.
NOTE: To change CCA while infusing, see Changing CCA While Infusing on page 6-12.
1 Select
[A] at the Main
STOPPED STOPPED
A
Delivery screen.
Back
Prime
Area Selection
Any Hospital, USA
Renal Neo ICU OB/GYN Oncol Ad ICU ER 7 South 3 North 4 North
Enter
B
Rate
0
mL/hr
Vol Inf
00
mL
MedSurg
A
0
Options/
B
Vol Inf
5- 22 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
The Line A Programming screen appears.
2 Select
[Drug List].
The Drug List Screen appears, in mainly alphabetical order.
NOTE: Default entry is “No Drug Selected”.
3 Use to highlight desired
drug name.
When list does not contain desired drug name or drug class name, select
[Page Down] (or  [Page
Up]) until it shows, then use
A
PROGRAM
Rate
VTBI
Duration
Enter Value using keypad
Program
Options
A
No Drug Selected
Albumin 5% Aldesleukin Alfentanil Alteplase Alteplase Amikacin Amiodarone
Select, then Enter
PageUpPage
A
Tirofiban Tobramycin Tobramycin Vancomycin VinCRIStine
VinCRIStine
Select, then Enter
PageUpPage
00:00
MedSurg
25 grams/500 mL
Down
MedSurg
Down
100 mg/100 mL
150 mg/100 mL
Enter
Enter
MedSurg
mL/hr
0
mL
0
hr : min
Cancel/
Drug
Back
List
2.6 mg/50 mL
10 mg/250 mL
50 mg/50 mL
100 mL
Cancel/
Back
5 mg/100 mL 40 mg/50 mL
80 mg/100 mL
500 mg/100mL
100 mL
50 mL
Cancel/
Back
to highlight choice.
4 Select
[Enter] to indicate selection and advance to
the Program Selection screen.
430-95597-001 (Rev. 06/05) 5- 23
For Systems With Hospira MedNet
TM
Software
The Select Program screen for Line A (or B) appears.
NOTE: This example shows line A being programmed.
5 Use to scroll down to
Multistep.
6 Select
If selected drug has been programmed with established dosing units and concentration, skip to step 9.
[Choose] to continue.
A
VinCRIStine
Select, then Choose
A
VinCRIStine
Select, then Choose
PROGRAM
Standard Program
Loading Dose
Multistep
PROGRAM
Standard Program
Loading Dose
Multistep
Choose Back
Choose Back
The Program Multistep screen appears.
NOTE: Cursor defaults to package insert recommended dose units associated with the selected drug.
7 Use key to highlight the
Program Multistep
A
VinCRIStine
mL/hr
mcg/kg/min mcg/kg/hr mcg/min mcg/hr mg/kg/hr
Select, then Choose
Page
Page
Up
Down
mg/min
mg/hr
grams/hr
ng/kg/min
mmol/min
mmol/hr
mEq/hr
Choose Back
units to use in the delivery.
8 Select
[Choose] to continue.
If desired, select  [Page Down] to see more units of
5- 24 430-95597-001 (Rev. 06/05)
measure.
Plum A+® Infusion Systems
The Multistep Programming screen appears.
9 Rate is the default field. Enter
value using keypad.
10 Use to highlight the
Duration or VTBI field.
NOTE: The VTBI value is automatically computed when the Duration value is entered. If VTBI is entered, the Duration value is automatically computed.
Program Multistep
A
VinCRIStine
VTBI
Rate
mL/hr 1 2 3 4
Enter value using keypad
Program
Options
A
VinCRIStine
Rate
mL/hr 1 2 3 4
Enter value using keypad
Program
Options
mL
0
50
0
0
0
0
0
0
Program Multistep
VTBI
mL
50
50
0
0
0
0
0
0
More
Steps
More
Steps
Duration
hr : min
00 : 00 00 : 00 00 : 00 00 : 00
Cancel/
Duration
hr : min
01 : 00
00 : 00 00 : 00 00 : 00
Cancel/
Back
Back
Program Multistep
11 Use to drop down to Step-
2, Rate field.
12 Enter the set of desired values
for Step-2 of the program. Continue process for each additional Step.
A
VinCRIStine
VTBI
Rate
mL/hr 1 2 3 4
Enter value using keypad
Program
Options
100
mL
50
50
0 0
0
0
0
More
Steps
Duration
hr : min
01 : 00 00 : 00 00 : 00 00 : 00
Cancel/
Back
430-95597-001 (Rev. 06/05) 5- 25
For Systems With Hospira MedNet
TM
Software
NOTE: When Step-4 is programmed, using [SELECT] to advance will highlight another item on this screen.
13 When Step-4 is programmed,
select
[More Steps] to program up
to 10 Steps.
NOTE: Using [SELECT] will not advance you to the additional steps.
14 To set up Steps 5 through 10,
enter desired values as performed for Steps 1 through
4.
To see the previous steps, select  [Prev Steps].
Program Multistep
A
VinCRIStine
VTBI
Rate
mL/hr 1 2 3 4
Enter value using keypad
Program
Options
A
VinCRIStine
5 6 7 8 9
10
Enter value using keypad
Program
Options
mL
50
50
825
100
75
75
50
50
Steps
Program Multistep
VTBI
Rate
mL/hr
mL
0
0
0
0
0
0
0
0
0
0
0
0
Steps
More
Prev
Duration
hr : min
01 : 00 08 : 15 01 : 00
01 : 00
Cancel/
Duration
hr : min
00 : 00 00 : 00 00 : 00 00 : 00 00 : 00 00 : 00
Cancel/
Back
Back
15 To start the delivery, press
.
This will bring up the Confirmation screen.
CAUTION: VERIFY ALL VALUES
BEFORE STARTING INFUSION.
Program Multistep
A
Delivery will be:
VinCRIStine
Drug Rate VTBI
Yes No
5050mL/hr
Confirm Program?
MedSurg
mL
!
5- 26 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
The Main Delivery Screen displays Step-1 in the upper­left corner.
NOTE: In units other than mL/hr, the Dose rate and units also will be displayed.
Example of Multistep Screen Pumping on Step-2.
16 To change a program
parameter while pumping, select  [A].
NOTE: The “*” in a Step number field indicates the step is delivering or “d” indicates each completed step.
17 Enter a value using the
keypad.
PUMPING STOPPED
A
VinCRIStine
Step 1
Rate
50
mL/hr
Vol Inf
5
27.
PUMPING STOPPED
A
VinCRIStine
Step 2
100
652 0
A
VinCRIStine
d
3 4
Enter value using keypad
Program
Options
mL
B
A
Rate
mL/hr
Vol Inf
mL
B
A
Program Multistep
VTBI
Rate
mL/hr
100
mL
50
50
825
75
75
50
50
More
Steps
Options/
Options/
Duration
hr : min
01 : 00 08 : 15 01 : 00 01 : 00
Cancel/
B
0
0
Vol Inf
B
0
Vol Inf
Back
430-95597-001 (Rev. 06/05) 5- 27
For Systems With Hospira MedNet
TM
Software
If Rate value is changed, the Duration will automatically change.
NOTE: Only the current delivering step and future steps can be changed.
18 To start the delivery, press
.
This will bring up the Confirmation screen.
CAUTION: VERIFY ALL VALUES
BEFORE STARTING INFUSION.
NOTE: The Values are now updated
on the Main Delivery Screen.
The infuser reverts to KVO when last Step completes.
The last step can be restarted by entering a new VTBI value.
Program Multistep
A
VinCRIStine
Duration
VTBI
Rate
mL/hr
50
d
125
75
3
50
4
Enter value using keypad
Program
Options
Program Multistep
A
Delivery will be:
VinCRIStine
Drug Rate VTBI
Confirm Program?
Yes No
PUMPING STOPPED
A
VinCRIStine
Step 2
125
658 0
A
mL
825
125 825
Rate
mL/hr
Vol Inf
mL
50
75 50
More
Steps
mL/hr mL
hr : min
B
01 : 00 06 : 36 01 : 00 01 : 00
Cancel/
Back
MedSurg
0
Options/
Vol Inf
!
B
5- 28 430-95597-001 (Rev. 06/05)
Plum A+® Infusion Systems
SECTION 6
Additional Features
Simple Delivery Using Delayed Start
1 Before or after entering the
simple delivery program, select
[Program Options].
NOTE: To program a Delayed Start using an advanced program (i.e., Dose Calc) you must first select which line you want (A or B), next select
the Delayed Start first, it will be cleared upon entry into Select Program mode.
The Default screen for Program Options appears.
[Drug List], and then
[Program Options]. If you program
PROGRAM
MedSurg
Mode
Concurrent
Rate
VTBI
Duration
Change using Change Mode
Program
Change
Options
Mode
Delay Start for 00:00 hr:min
Callback No
Enter Value using keypad
Standby
250
50:00
Drug
List
Enter
5
mL/hr
mL
hr : min
Cancel/
Back
Cancel/
Back
B
BProgram Options
430-95597-001 (Rev. 06/05) 6- 1
For Systems With Hospira MedNet
TM
Software
2 Enter values for desired hours
and/or minutes.
3 Select
[Enter] to confirm
delayed start.
CAUTION: VERIFY ALL VALUES
BEFORE STARTING INFUSION.
4 Press to confirm
programming.
Line B is delayed for the time entered.
NOTE: Delayed line drip indicator is solid, not flashing. It will begin flashing when actually pumping.
Delay Start for 01:00 hr:min
Callback No
Enter Value using keypad
Enter
Standby
PUMPING DELAYED
A
Rate
125
!
mL/hr
Vol Inf
850 0
MedSurg
A
mL
B
Cancel/
Back
5
Options/
Vol Inf
BProgram Options
B
!
Piggyback with Nurse Callback
After programming in desired
1
values for both lines, select
[Program Options].
NOTE: To set a nurse callback, the infuser must be stopped. This function can be selected either before or after entering the Piggyback delivery program.
6- 2 430-95597-001 (Rev. 06/05)
PROGRAM
MedSurg
Mode
Piggyback
Rate
VTBI
Duration
Enter value using keypad
Program
Options
200
100
00:30
Drug
List
mL/hr
mL
hr : min
Cancel/
B
Back
Loading...
+ 54 hidden pages