All products manufactured by ICP DAS are warranted against defective materials for a
period of one year from the date of delivery to the original purchaser.
Warning
ICP DAS assumes no liability for damages consequent to the use of this product.
ICP DAS reserves the right to change this manual at any time without notice. The
information furnished by ICP DAS is believed to be accurate and reliable. However, no
responsibility is assumed by ICP DAS for its use, nor for any infringements of patents or
other rights of third parties resulting from its use.
Revision History ........................................................................................................................ 221
Page 5
1. Introduction
The CL-200 series of data logger devices can be used to record PM2.5, CO, CO2, H2S,
NH3, HCHO, TVOC, temperature, humidity and dew point information, including the
date and time stamps, and are able to store up to 450,000 downloadable records.
Real-time data can be accessed from the CL-200 data logger from anywhere and at any
time using the free Windows software, the iOS App, or the Android App, as long as they
are connected to the same local network as the data logger.
The CL-200 series module contains RS-485, Ethernet and PoE communication
interfaces, the most common communication interfaces in industrial network. With
additional Wi-Fi interface, the CL-200-WF series provides a WLAN connection which
makes an easy way to incorporate wireless connectivity into monitoring and control
systems.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 5 -
Page 6
Characteristics
PM2.5 measurement range: 0 to 400 ug/m3
CO measurement range: 0 to 1000 ppm
CO2 measurement range: 0 to 9999 ppm
NH3 measurement range: 0 to 100 ppm
H2S measurement range: 0 to 100 ppm
HCHO measurement range: 0 to 2000 ppb
TVOC measurement range: 0 to 60000 ppb
Non‐dispersive Infrared (NDIR) sensor with Automatic Baseline Correction
algorithm for CO2 measurement
HCHO : Electrochemical Sensor
TVCO : Metal-Oxide Sensor
Able to store up to 450,000 records
Remote control with a standard web-browser
iAir App for iOS or Android mobile devices to monitor on-line data
Supports the DCON, Modbus RTU, Modbus TCP protocols
Supports the MQTT Protocol for Ethernet Interface
One relay output for turning on/off alarm light/buzzer or IAQ control devices
Includes RS-485/Ethernet communication interfaces
Includes Wi-Fi Interfaces (CL-200-WF Series)
Compatible with IEEE802.11b/g/n standards (CL-200-WF Series) Support infrastructure and limit-AP modes for wireless networks(CL-200-WF Series)
Wide operating temperature range of 0 to 50°C
RoHS compliant with no Halogen
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 6 -
Page 7
Features
NDIR Sensor
NDIR (Non-Dispersion Infrared) is based on one of the natural properties of CO2
molecules: CO2 molecules absorb light at a specific wavelength of 4.26 µm. This
wavelength is in the infrared (IR) range. High concentrations of CO2 molecules
absorb more light than low concentrations. NDIR sensor can detect fast and
accurately in a wide range of CO2 concentration.
Built-in Web Server
With the built-in Web server, users can easily log in to the CL-200 module via a
standard web browser to monitor the data and configure the settings without install
any software in the terminal. Please note that the web server is only available to the
Ethernet interface.
Get Real-time Data Anywhere and Anytime
iAir App for iOS or Android Phones or Tablets is free and easy to install, it can
obtain the real-time data from CL-200 modules over a Wi-Fi network anytime and
anywhere. The iAir App can link to the CL-200 modules by specifying IP addresses
or by searching all the modules connected to the same Ethernet segment.
Data Logging Software
The DL-300 Utility can be used to configure the modules, monitor real-time data
and show the run chart, log alarm events, group CL-200 modules so that the status
of distribution groups can be
viewed and managed. The
utility also allows the log data
to be downloaded and
exported to a .CSV file that
can then be imported into
any industry-standard
software or spread sheet for
analysis.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 7 -
Page 8
Easy integration with SCADA software
Modbus is one of the most popular protocols used in the industrial world.
Supporting traditional serial protocols of RS-485 and Ethernet protocols allow the
CL-200 series well-integrated into the HMI/SCADA systems.
Alarm
CL-200 series allows users to set high alarm level for PM2.5/CO/CO2/H2S/NH3/
HCHO/TVOC/Temperature/Humidity/Dew Point and low alarm level for
Temperature /Humidity /Dew Point, and to enable/disable the alarm functions. An
Alarm LED indicator on the front of the CL-200 module will flash when an alarm
event is activated, and a relay output related to all alarm events can be use to tap
an alarm light/sound or control the IAQ devices such as ventilators, air cleaners,
and filters. Beep alarm is available when the CO/CO2/H2S/NH3/ HCHO/TVOC high
level alarm occurs.
Automatic Baseline Correction
The built-in ABC algorithm makes the CO2 sensor on the CL-202, CL-212, CL-203,
CL-213, maintenance-free. In most indoor applications, the carbon dioxide level
drops to nearly outside air - 400 ppm, and then the ABC algorithm constantly keeps
track of the lowest reading and slowly corrects it as the expected fresh air value of
400 ppm. The ABC algorithm can not apply for the places where are no periods that
the CO2 concentration drops to background level such as greenhouses, hospitals,
24-hour operation factories or stories. The ABC function needs be disabled where
the spaces the CO2 concentration may be elevated at all times.
Easy Wiring
Support for RS-485, Ethernet and Power over Ethernet (PoE) interfaces for users
to choose the appropriate one to meet the field requirements.
Power over Ethernet (PoE)
The CL-200 series features true IEEE802.3af-compliant (classification, Class 1)
PoE technology that allows both power and data to be carried over a single
Ethernet cable. PoE provides a unified power system, as well as backup provisions
for critical building functions, without any additional cables, outlets or connections.
It can reduce the power supply wiring and maintenance costs, and improve system
scalability.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 8 -
Page 9
Support for MQTT protocol
MQTT is a protocol designed for the
efficient exchange of real-time data
with sensor and mobile devices. It
runs over TCP/IP and is in widest
use on the "machine-to-machine"
(M2M) and "Internet of Things"
applications today
Smartphone
MQTT
Broker
CL-203-E
CL-212-E
Mobile Tablet
Subscriber
Compatible with IEEE 802.11b/g/n standards
The CL-200-WF modules are complied with IEEE 802.11b/g/n standard from
2.4~2.5 GHz. It can be used to provide up to 11 Mbps for IEEE 802.11b and 54
Mbps for IEEE 802.11g to connect to your wireless LAN.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 9 -
Page 10
2. Hardware
Model
CL-201-E
CL-201-BLE
CL-201-WF
CO Measurement
Range
0 to 1000 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
±5% of measured value
Response Time
30 seconds
Warm-up Time
60 seconds
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
CO Alarm
Yes
CO2 Alarm
-
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
2.1 Specifications
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 10 -
Page 11
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
1.7 W
1.8 W
1.9 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-202-E
CL-202-BLE
CL-202-WF
CO2 Measurement
Range
0 ~ 9999 ppm
Resolution
1 ppm
Accuracy
±40 ppm ±3%
Response Time
120 seconds
Warm-up Time
5 minutes
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 11 -
Page 12
System
CO Alarm
-
CO2 Alarm
Yes
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
1.8 W
1.9 W
2.0 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 12 -
Page 13
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 13 -
Model
CL-203-E
CL-203-BLE
CL-203-WF
CO Measurement
Range
0 to 1000 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
±5% of measured value
Response Time
30 seconds
Warm-up Time
60 seconds
CO2 Measurement
Range
0 ~ 9999 ppm
Resolution
1 ppm
Accuracy
±40 ppm ±3%
Response Time
120 seconds
Warm-up Time
5 minutes
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
CO Alarm
Yes
CO2 Alarm
Yes
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Page 14
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
1.8 W
1.9 W
2.0 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-204-E
CL-204-BLE
CL-204-WF
HCHO Measurement
Range
0 to 2000 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
-
Response Time
≤60 seconds
Warm-up Time
180 seconds
TVOC Measurement
Range
0 to 60000 ppb (MEMS Metal Oxide)
Resolution
1 ppb
Accuracy
± 15%
Response Time
60 seconds
Warm-up Time
180 seconds
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 14 -
Page 15
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
HCHO Alarm
Yes
TVOC Alarm
Yes
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
2.0 W
2.1 W
2.2 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 15 -
Page 16
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 16 -
Model
CL-205-E
CL-205-BLE
CL-205-WF
NH3 Measurement
Range
0 to 100 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
± 5% of measured value
Response Time
< 120 Sec
Warm-up Time
180 Sec
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
NH3 Alarm
Yes
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Page 17
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
1.9 W
2.2 W
2.2 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-206-E
CL-206-BLE
CL-206-WF
H2S Measurement
Range
0 to 100 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
± 5% of measured value
Response Time
< 120 Sec
Warm-up Time
180 Sec
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
H2S Alarm
Yes
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 17 -
Page 18
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
1.9 W
2.2 W
2.2 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-207-E
CL-207-BLE
CL-207-WF
HCHO Measurement
Range
0 to 2000 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
-
Response Time
≤60 seconds
Warm-up Time
180 seconds
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 18 -
Page 19
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
HCHO Alarm
Yes
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
2.0 W
2.1 W
2.1 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 19 -
Page 20
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 20 -
Model
CL-208-E
CL-208-BLE
CL-208-WF
TVOC Measurement
Range
0 to 60000 ppb (MEMS Metal Oxide)
Resolution
1 ppb
Accuracy
± 15%
Response Time
60 seconds
Warm-up Time
180 seconds
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
HCHO Alarm
Yes
TVOC Alarm
Yes
PM2.5 Alarm
-
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Page 21
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
1.9 W
2.0 W
2.0 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-210-E
CL-210-BLE
CL-210-WF
PM2.5 Measurement
Range
0 to 400 ug/m3
Resolution
1ug/m3
Response Time
<=1 min.
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
CO Alarm
-
CO2 Alarm
-
PM2.5 Alarm
Yes
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 21 -
Page 22
Communication
Protocol
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
+12 to +48 VDC
Powered via PoE
IEEE 802.3af, Class 1 (require a PoE switch or injector)
Power Consumption
1.7 W
1.8 W
1.8 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-211
CL-211-E
CL-211-BLE
CL-211-WF
PM2.5 Measurement
Range
0 to 400 ug/m3
Resolution
1ug/m3
Response Time
<=1 min.
CO Measurement
Range
0 to 1000 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
±5% of measured value
Response Time
30 seconds
Warm-up Time
60 seconds
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 22 -
Page 23
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 23 -
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
CO Alarm
Yes
CO2 Alarm
-
PM2.5 Alarm
Yes
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus
RTU
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
Block
+10 to +30
VDC
+12 to +48 VDC
Powered via PoE
-
IEEE 802.3af, Class 1 (require a PoE switch or
injector)
Power Consumption
1.2 W
2.0 W
2.0 W
2.1 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Page 24
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-212
CL-212-E
CL-212-BLE
CL-212-WF
PM2.5 Measurement
Range
0 to 400 ug/m3
Resolution
1ug/m3
Response Time
<=1 min.
CO2 Measurement
Range
0 ~ 9999 ppm
Resolution
1 ppm
Accuracy
±40 ppm ±3%
Response Time
120 seconds
Warm-up Time
5 minutes
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
CO Alarm
-
CO2 Alarm
Yes
PM2.5 Alarm
Yes
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 24 -
Page 25
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus
RTU
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
Block
+10 to +30
VDC
+12 to +48 VDC
Powered via PoE
-
IEEE 802.3af, Class 1 (require a PoE switch or
injector)
Power Consumption
1.2 W
2.1 W
2.2 W
2.2 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
Model
CL-213
CL-213-E
CL-213-BLE
CL-213-WF
PM2.5 Measurement
Range
0 to 400 ug/m3
Resolution
1ug/m3
Response Time
<=1 min.
CO Measurement
Range
0 to 1000 ppm (Electrochemical)
Resolution
1 ppm
Accuracy
±5% of measured value
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 25 -
Page 26
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 26 -
Response Time
30 seconds
Warm-up Time
60 seconds
CO2 Measurement
Range
0 ~ 9999 ppm
Resolution
1 ppm
Accuracy
±40 ppm ±3%
Response Time
120 seconds
Warm-up Time
5 minutes
Temperature Measurement
Range
-10 to +50°C
Resolution
0.1°C
Accuracy
±0.6°C
Relative Humidity Measurement
Range
0 to 100% RH, Non-condensing
Resolution
0.1% RH, Non-condensing
Accuracy
±5% RH, Non-condensing
Dew Point
Range
Calculated using temperature and relative humidity
Resolution
0.1°C
System
CO Alarm
Yes
CO2 Alarm
Yes
PM2.5 Alarm
Yes
Real Time Clock
Yes
Data Logger
Yes, 450,000 Records
Alarm Relay Output
Form A×1, SPST. 30 VDC @ 16 A or 250 VAC @ 16 A
Communication
Protocol
DCON, Modbus
RTU
DCON, Modbus RTU, Modbus TCP, MQTT
Wire Interface
Yes, RS-485 X 1
Yes, RS-485 X 1 and Ethernet/PoE X1
Wireless Interface
-
Bluetooth
Wi-Fi
Standard Supported
-
BT 4.0
IEEE 802.11 b/g/n
Wireless Mode
-
Slave
Infrastructure/
Limited AP
Wireless Security
-
AES 128
WEP, WPA ,WPA2
Page 27
Transmission Range
-
20 m (LOS)
50 m (LOS)
Electrical
Powered via Terminal
Block
+10 to +30
VDC
+12 to +48 VDC
Powered via PoE
-
IEEE 802.3af, Class 1 (require a PoE switch or
injector)
Power Consumption
1.2 W
2.1 W
2.2 W
2.2 W
Mechanical
Dimensions (D x H)
Ø 150 mm x 53 mm
Installation
Ceiling Mounting or Wall Mounting
Environment
Operating
0 to +50°C
Storage Temperature
-30 to +75°C
Humidity
10 to 90% RH, Non-condensing
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 27 -
Page 28
2.2 Appearance
DIP Switch
Rotary Switch
PoE/Non-PoE
Ethernet Port
Connectors:
Power
RS-485
Alarm Relay Output
LED Indicators:
Power
Alarm
Internal Wi-Fi Antenna
(For CL-200-WF Only)
LED Indicators
PWR: Red for normal operation.
Alarm: Green for alarm condition.
DIP Switch
The functions are printed on the right beside the SW1 DIP switch.
PoE/ non-PoE Ethernet port
The Ethernet port can be used to connect to a PoE switch or a non-PoE switch.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 28 -
Page 29
Connector for Power/ Frame Ground / RS-485/ Alarm Relay Output
The Push-in connector can easily connect and
detach solid wires or wires with wire-end
ferrules without using tools. Just push in the
solid wire to lock it and press the white button to
release the wire.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 29 -
Page 30
2.3 Dimensions (unit: mm)
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 30 -
Page 31
2.4 Cabling for Power and Network
Note
Do not install the CL-200 module near a vent, a ventilation fan or a door where the
air flows faster.
Avoid installing in locations where the temperature is below 0°C or above 50°C.
Avoid installing in locations near a strong electromagnetic field.
For connecting with a PC or a Android device
The CL-200 logger can connect to a PoE network without a power source or connect to
a non-PoE network. When using the Search function in iAir App on Android or iOS
mobile devices, mobile devices need to connect to the same subnet that the CL-200
connected to over Wi-Fi. Similarly to using the Search function in DL-300 Utility running
on Windows, the module and the host PC need to connect on the same subnet, too.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 31 -
Page 32
The iAir App and DL-300 Utility search the logger by broadcast, therefore only the
RS-485 Network
CL-200-E
Power Supply
USB/RS-485Converter
I-7561
CL-200-E
devices on the same subnet can be searched out. It means that the host PC, Android
devices and the logger must have the same broadcast address. The broadcast address
for an IPv4 device can be obtained by performing a bitwise OR operation between the
bit complement of the subnet mask and the IP address for a device. In other words, take
the device's IP address, and set to '1' any bit positions which hold a '0' in the subnet
mask.
For example, in an entire IPv4 subnet, the host PC or the Android device uses the
private IP address space 172.16.0.0/12 and subnet mask address 255.240.0.0, the
broadcast address is 172.16.0.0 | 0.15.255.255 = 172.31.255.255. Only the loggers
which have the same broadcast address could be searched out in the iAir App or
DL-200 Utility. Please contact with your network administrator to make sure the CL-200
logger is connected to the same sub-network that your Android devices or PC is
connected to.
For connecting with PC via RS-485 network
The CL-200 logger can connect to the PC through a RS-485 network with power input
requirement of +12 ~ +48 VDC.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 32 -
Page 33
For connecting with PC via Wi-Fi
The CL-200-WF logger can connect to the PC through Wi-Fi with power input
requirement of +12 ~ +48 VDC.
The CL-200-WF device can be configured as station mode, such that the PC/Laptop
can be connected through Wi-Fi AP.
The CL-200-WF device can be configured as AP mode, such that the PC/Laptop can be
connected through Wi-Fi directly. Only one device is allowed to be connected to the
DL-300-WF module in AP mode.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 33 -
Page 34
2.5 Hardware Mounting
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 34 -
Page 35
(Requires RJ45 Cable, Male-Female, 30cm (90º))
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 35 -
Page 36
3. Configuration via Web Browser
CL-200 logger has a built-in web server that provides simple web pages for remote
monitoring real-time data and configuring the logger with a standard browser. For
opening the web page in CL-200, the factory default IP address (192.168.255.1),
Subnet Mask (255.255.0.0) and Gateway (192.168.0.1) need be set to available
IP/Subnet Mask/Gateway addresses in your Ethernet environment. The Ethernet
configuration can be set by entering the Settings menu from the web pages. Please
note that the web configuration is only available to the Ethernet interface.
3.1 Search the CL-200 logger
eSearch is designed to search out the CL-200 logger connected on the same Ethernet
network, it supports for Linux and Windows and is needless to install. Please note that
the eSearch Utility cannot be used to configure CL-200-WF device via the Wi-Fi
interface. In this case, use the Wi-Fi IIOT Utility instead.
The eSearch can be downloaded from
http://ftp.icpdas.com/pub/cd/iiot/utility/
Before running eSearch, turn off firewall on computer, and connect the computer and
CL-200 logger to Ethernet network.
1. Launch eSearch, click the Search Servers button to search the CL-200 modules
connected to the network, the modules searched out will be listed as below.
2. Double click the module name searched in the list.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 36 -
Page 37
3. Set available IP Address, Sub-net Mask, Gateway (designated by your network
1
2
administrator) and alias and click the OK button. The Alias for easy to identify each
item will be shown at the bottom-left corner of the CL-200 screen.
3.2 Logging into the CL-200
1. Enter the IP address for your CL-200 in the address bar of a web browser. (sec. 3.1).
2. Type the Login password, and click the Submit button.
(The default Login password is Admin, case sensitive.)
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 37 -
Page 38
3.3 Home
The first page displayed is Home, it shows the based configuration of the CL-200
module and the real-time data as below:
In the Sensor Readings field is the real-time data of PM2.5/CO/CO2/NH3/H2S/HCHO
/TVOC concentration, temperature, humidity and dew point, the minimum value (Low
Latched) and maximum value (High Latched) logged. Clicking on the Clear Low Latched button and the Clear High Latched button can reset the latched data to
current value and latch new minimum or maximum value.
The Alarm table displays the settings of alarm mode, high alarm limit for PM2.5/CO/CO
/NH3/H2S/HCHO/TVOC concentration, temperature, humidity and dew point, low alarm
limit for temperature, humidity and dew point, and the alarm status for each. Clicking on
2
the Clear Latched Alarm button can clear the activated alarm status.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 38 -
Page 39
The Digital Output table shows the status of the relay output and the control button Set Digital Output to change the relay output status. The control function is invalid when
any of the alarm modes is not disabled. If one of the alarm modes is enabled, the relay
is linked to the alarm status for tapping audible/visual alarm.
At the end of the page are the data, time and device online time since powered on.
3.4 Network
The networks parameters are set on this page including DHCP enabled/disabled,
IP/Subnet Mask/Gateway addresses, the port number and the NetID for Modbus TCP
communication. Remember to click on the Update Settings button to update new
parameters.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 39 -
Page 40
Item
Description
Default
System
Timeout
(Network
Watchdog)
Sets the timeout for rebooting a CL-200 logger when it is
abnormal or failure to communicate.
Range: 30 ~ 65535 (unit: second)
0 = Disable
0
(Disable)
TCP Timeout
Sets the timeout for disconnecting a TCP connection
when a CL-200 does not receive data coming from the
Ethernet port.
Range: 5 ~ 65535 (unit: second)
0 = Disable
180
Web
Auto-logout
Sets the timeout for logout the web server in a logger
when there is no any operation from the web browser
interface.
Range: 1 ~ 65535 (unit: minute)
0 = Disable
10
Alias Name
Sets an alias name for easy to identify a CL-200. The
maximum length is 18 characters.
EtherIO
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 40 -
Page 41
Item
Factory Default
IP address type
Static IP
Static IP
192.168.255.1
Default gateway
192.168.0.1
Subnet Mask
255.255.0.0
MAC address
Factory MAC address
Modbus TCP port
502
Modbus TCP NetID
1
Modbus TCP NetID
Enabled
System Timeout
0 (disabled)
TCP Timeout
180 seconds
Web auto logout
10 minutes
Alias name
EtherIO
Accessible IP
Disabled
The Reboot button is used to reboot the CL-200. After pressing the button, a user
needs to login the CL-200 logger again to using the web interface.
The Restore Defaults button can be used to restore the following settings to factory
default values.
The Update button is used to update firmware. Please Refer to firmware update manual
first.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 41 -
Page 42
3.5 MQTT
MQTT stands for MQ Telemetry Transport, it is a publish/subscribe, extremely simple
and lightweight messaging protocol, designed for constrained devices and
low-bandwidth, high-latency or unreliable networks.
The Publish-Subscribe messaging pattern requires a message broker. The broker is
responsible for distributing messages to interested clients based on the topic of a
message. Now the MQTT Version 3.1.1 becomes an OASIS standard, it is an ideal
protocol for communicating with connected devices in the emerging
"machine-to-machine" (M2M) and "Internet of Things" applications, and for mobile
applications where bandwidth and battery power are at a premium.
Input the IP address and port number for the MQTT broker and click on the Update Settings button to save the parameters.
The MQTT Last Will and Testament (LWT) feature is used to notify other clients about
an ungracefully disconnected client. A CL-200-E can register an offline message (LWT)
to the broker. The LWT message will be deliver to all clients who subscribe to the offline
topic if the CL-200-E disconnects unexpectedly.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 42 -
Page 43
- Cycle: sets the time period for update the publish messages in millisecond.
Temperature (°F)/ Dew Point (°C)/ Dew Point (°F) Sub Topic Name: sets the sub topic
name for each item.
A MQTT client subscribes the messages form a MQTT broker by specifying the topic
name as
Module Topic Name + Sub Topic Name
For example, to subscribe the CO2 level in this case, a MQTT client subscribes the
topic name from a MQTT broker as
EtherIO/CO2
If a MQTT control message is published to topic name: “Module Topic Name + DO0 Sub
Topic Name ” for a CL-200 logger, the logger will follow the MQTT message described
to set the Relay Output.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 43 -
Page 44
3.6 I/O Settings
Users can change the temperature unit to Fahrenheit or Celsius in this field.
To Enable/Disable the CO2 Automatic Baseline Correction function.
Q & A
Q: What is ABC (Automatic Baseline Correction)?
A: ABC stands for the Automatic Baseline Correction which is used to adjust a shifted
baseline to the carbon dioxide level in fresh air. In case of normal indoor application,
the carbon dioxide level drops to nearly outside air where there are no human, green
plants or anything to elevate the carbon dioxide levels on weekday evenings or
weekends, the ABC algorithm constantly keeps track of the lowest reading and
slowly corrects it as the expected value in fresh air typically around 400 ppm.
Q: Why I need to enable the ABC?
A: When the CO2 concentration detected in a period time of unoccupied space is
greater than the base value of 400ppm, enable the ABC function to adjust the
baseline. Be careful that the ABC will not work if a space is constantly occupied such
as a hospital, 24-hr factory, 24-hr store, green house or other applications where
CO2 levels may be elevated at all times.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 44 -
Page 45
All the settings take effect after clicking the Update Settings button.
Item
Description
Default
Alarm Mode
- Disabled:
Disables alarm function.
- Momentary:
If a measurement value higher than the High Alarm Limit
or lower than the Low Alarm Limit, the alarm occurs until
the measurement value is within a range from Low Alarm
Limit to High Alarm Limit. (For CO/CO2 level, until the
measurement value is lower than the High Alarm Limit.)
The Alarm LED turns red, and the relay turns to on for
every alarm event, and a sound alarm beeps as the
setting in Beep on Alarm Time for CO/CO2 high limit
alarm events during the alarm stage.
- Latched:
If a measurement value higher than the High Alarm Limit
or lower than the Low Alarm Limit, the alarm occurs. The
Alarm LED turns red, the relay turns to on for every alarm
event, and a sound alarm beeps as the setting in Beep on Alarm Time for CO/CO2 high limit alarm events.
Even though the alarm event is not presented, the alarm
status is latched; the Alarm LED keeps red, and the relay
keeps on and the sound alarm keeps beeping if it is set to
beeping continuously.
Disabled
Low Alarm
Limit
Sets the Low alarm limit conditions for Relative Humidity/
Temperature/ Dew Point.
High Alarm
Sets the High alarm limit conditions for CO/CO2/Relative
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 45 -
Page 46
Limit
Humidity/ Temperature/ Dew Point.
Beep On
Alarm
Enable/disable beep on alarm for PM2.5 /CO /CO2
/Temp /RH /Dew point
Beep On
Alarm Time
Sets the time for beeping alarm.
Range: 1 ~ 250 (unit: second)
0 = disable the beeping alarm
251 = continue the beeping alarm without stop
30
Set the Power On Value and Safe Value for the relay output, and the Host Watchdog Timeout timer for RS-485 communication; if a host does not send a command over the
setting time, the Host Watchdog timeout occurs and the relay outputs the status set for
Safe value. The settings for Power On Value and Safe Value are unavailable when any
one setting in the Alarm Mode is enabled.
All the settings take effect after clicking the Update Settings button.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 46 -
Page 47
In this table it shows the settings for data logger.
Item
Description
Default
Status
- Running: the data logger is running
- Stopped: the data logger is stopped
Change
Logging
Sets the mode for data logger
- Stop: stops the data logger
- Run: continues logging data
- Period: logs data in the specified period time
Stop
All the settings take effect after clicking the Update Settings button.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 47 -
Page 48
Overwrite
on Full
Sets whether to overwrite old data by new ones when the
memory for data storage is full. (Over the upper limit of
450,000.)
- No: discards the new data (default)
- Yes: overwrites the old data by new ones
No
Sampling
Interval
Sets the time interval for logging data. It is valid for both Run
mode and Period mode.
- Sampling Interval – Hour: sets the hour for log interval
- Sampling Interval – Minute: set the minute for log interval
- Sampling Interval – Second: sets the second for log
interval
10 (s)
Period
Start
Sets the start time for Period mode.
Period End
Sets the stop time for Period mode
Click the Reset Data Logger button to clear the data in data storage memory.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 48 -
Page 49
3.7 Filter IP
For limiting the devices to access the CL-200 logger, users can specifies particular
devices by setting their IP addresses on this page. When the addresses are 0.0.0.0
from IP1 to IP5, all the devices can access the logger. Once any of the 5 IP address
columns is set, only the device with which IP is saved in the list can assess the logger.
Filter Settings
1. Select the radio button for Add ___.___.___.___ To The List and type the IP
address for the accessible device in the following text box.
2. Click on the Submit button to the setting effect without restarting.
If the IP setting needs be saved for using after repowered, check the checkbox for
Save to Flash before clicking the Submit button.
Delete IP setting
Select the radio button for Delete IP# to delete a specified IP or the radio button for
Delete All to delete all the IP, check the checkbox for Save to Flash and then click
the Submit button to take the delete operation effect.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 49 -
Page 50
3.8 Change Password
On this page users can change the passwords for login the logger and locking the touch
screen. The factory default for the CL-200 touch screen has no password protection.
After setting the password for touch screen, each time whoever wants to change to
settings from the touch screed, the password will be requested.
Change Web Password
The password for logging into the web page is Admin and can be changed in the
Change Web Password field. The password can be alphabetic characters or
numbers and up to 12 characters (case sensitive).
To change the password, uses need enter the Current password, New password,
and Confirm new password columns and click the Submit button for Change Web
Password to take the setting effect.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 50 -
Page 51
3.9 Logout
Click the Logout on any page to logout the CL-200.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 51 -
Page 52
3.10 Wi-Fi
Item
Description
Connection Status
The Wi-Fi connection status of the CL-200-WF device.
Signal Strength
The Wi-Fi signal strength of the CL-200-WF device in station mode. It
can be High, Medium, Low, or Not Connected.
MAC Address
The MAC address of the Wi-Fi interface of the CL-200-WF device.
IP Address
The IP address of the Wi-Fi interface of the CL-200-WF device.
Update Wi-Fi Status
Click this button to update the Wi-Fi status of the CL-200-WF device.
For CL-200-WF module, the Wi-Fi related parameters can be set via the Wi-Fi page.
This page including Wi-Fi Status and Wi-Fi Settings, each of which will be described in
more detail below.
Wi-Fi Status
The following table provides an overview of the parameters contained in the Wi-Fi
Status section:
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 52 -
Page 53
Wi-Fi Settings
Item
Description
Mode
This parameter is used to specify the Wi-Fi mode of the CL-200-WF device.
It can be station or AP. For AP mode, only one device can be connected.
Wireless Security
This parameter is used to specify which security protocol is used to secure
wireless computer network. It can be open, WEP, or WPA/WPA2. It is
recommended to use WPA/WPA2 if possible.
DHCP Server (AP
Mode)
This parameter is used to specify whether to turn on the DHCP server
function. It is only available to the AP mode.
Wi-Fi Channel (AP
Mode)
This parameter is used to specify which channel is used for Wi-Fi
transmission. It can be 1 to 11. It is only available to the AP mode.
IP Address Type
(Station Mode)
This parameter is only available to the station mode and it can be Static IP
or DHCP. If DHCP is supported by the AP you would like to connect, then
DHCP should be selected. Otherwise, select Static IP and the following
three parameters IP Address, Subnet Mask and Gateway should be set, too.
The column of Current shows the current Wi-Fi settings. You can change the settings
by changing the column of New. The following table provides an overview of the
parameters contained in the Wi-Fi Settings section:
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 53 -
Page 54
IP Address
Each CL-200-WF device connected to the Wi-Fi network must have its own
unique IP address. This parameter is used to assign a specific IP address.
Subnet Mask
This parameter is used to assign the subnet mask for the CL-200-WF
device. The subnet mask indicates which portion of the IP address is used to
identify the local network or subnet.
Gateway
This parameter is used to assign the IP address of the gateway to be used by
the CL-200-WF device. A gateway (or router) is a device that is used to
connect an individual network to one or more additional networks.
SSID
This parameter is used to specify the Service Set Identifier. For station
mode, specify the SSID of the AP you would like to connect. For AP
mode, the SSID will be used by the device to be connected.
Modbus TCP Port
This parameter is used to set the local port of the Wi-Fi interface to be used
by the Modbus slave device. The default value is 502.
Update Settings
Click this button to save the revised settings to the CL-200-WF device.
Factory Default Wi-Fi Settings
Mode
AP
Wireless Security
WPA/WPA2, "00000000"
DHCP Server (AP Mode)
DHCP Server on, start IP: 192.168.255.2
Wi-Fi Channel (AP Mode)
11
IP Address
192.168.255.1
Gateway Address
192.168.255.254
Subnet Mask
255.255.0.0
SSID
CL-200-WF
Modbus TCP Port
502
The following table provides an overview of the factory default Wi-Fi settings:
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 54 -
Page 55
Set COM port
Start Search
Stop Search
4. Configuration via RS-485
The factory default settings for RS-485 communication
If there are multiple CL-200 loggers connected to the same RS-485 network, each
logger needs be set with a unique RS-485 address. More than one module having
the same address will cause communication failure
Testing RS-485 Communication
1. Download the DCON Utility Pro from
http://ftp.icpdas.com/pub/cd/iiot/utility/
2. Launch the DCON_Utility_Pro.exe.
CL-200 Series User ManualVersion 1.2.0 Apr. 2019 - 55 -
3. Click the icon to configure the COM port.
Page 56
4. Select the COM Port number used to connect the CL-200 logger.
5. The Baud Rate is factory default to 9600 bps.
6. Select the Protocol tab and check the protocol that set in the logger.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 56 -
Page 57
7. Select the Format tab and check the parity that set in the logger.
8. Click the Start Search icon.
9. The CL-200 logger searched out will be listed as below.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 57 -
Page 58
10. Click the module name to configure the logger.
Note
The Protocol/Baud Rate/Parity/Checksum items marked with ”(INIT*)” means that
when any of those items needs be modified, the pin 4.INIT needs to be set in ON
position and power cycle the logger, then the item can be modified. After complete
setting, set the pin 4.INIT back to OFF position and power cycle the logger again to
take the setting effect.
AI tab
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 58 -
Page 59
Degree of offset
Clear High/Low
Alarm Latch
PM2.5/CO
2
level
Humidity、Temperature and
Dew point temperature
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 59 -
Page 60
Alarm tab
Set Alarm Configurations
Alarm Mode
Low Limit
High Limit
Alarm Buzzer On
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 60 -
Page 61
When the buttons are
clicked, the DO status
will be set as the
Power On value or
Safe value.
Select one of the radio button and the
checkbox next to DO will display the
setting for selected item.
The checkbox can be used to control and display the DO status.
Check/Uncheck the checkbox can control the relay output.
Checked/Unchecked symbol also means the DO status
DO tab
On this DO tab, users can control the relay to output ON or OFF status, and set the
power on value and safe value for the relay output.
When any one of the high/low limit alarm for CO/CO2 concentration, temperature,
humidity and dew point is enabled, the functions on this tab are all disabled as
below.
If all the alarm events are disabled, the functions are available as below:
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 61 -
Page 62
Host Watchdog
Host Watchdog is used to monitor the RS-485 communication status; if the host
(PC) does not send command “~**” in the time period of WDT Timeout setting, the
enabled Host Watchdog will announce the timeout error and turn the relay output
to Safe value to avoid an unsafe act. Users can not control the relay until the
command “~AA1” is sent to clear the WDT timeout status.
Note
On this tab:
1.Set the time period for WDT timeout, check the checkbox next to Enable WDT
and click the Set WDT button to enable the Host watchdog.
2.Check the checkbox next to Send Host OK to send the “~**” command.
3.Uncheck the checkbox next to Send Host OK to stop sending ~** command, the
Host watchdog timeout will occur and relay will turn to Safe value.
4.Click the Reset WDT button to clear the Host watchdog timeout status.
5.Uncheck the checkbox next to Enable WDT and click the Set WDT button to
disable the Host watchdog.
The relay will not turn to Safe value when any one of the alarm for PM2.5/CO/CO
concentration, temperature, humidity and dew point is enabled. If any one alarm is
2
enabled, the relay will be linked to the Alarm status. In case an Alarm occurs, the
relay turns ON, it can be used to turn on the user’s alarm light or beeping alarm or
other device.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 62 -
Page 63
Logger Configuration
Set the logger configuration on this TAB.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 63 -
Page 64
Check/Uncheck the item to Enable/Disable
ABC function(For CL-202-E, CL-212-E,
CL-203-E, CL-213-E only)
Click the Apply button to
save settings.
Buzzer On Alarm Time setting.
System Tab
Click the Edit button to enable settings on this tab.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 64 -
Page 65
INIT
In case of the following situations, users have to set the pin 4.INIT on SW1 in the ON
position and power-cycle the CL-200 module:
- Change protocol from PC
- Change DCON configuration such as baudrate, parity and checksum
- Communication failure with a CL-200 module.
When a CL-200 module is powered-on with the pin 4.INIT in ON position, the protocol is
DCON, address is 0, Baud Rate is 9600 bps, Parity is set to N/8/1 and Checksum is
disabled.
After configuring the communication parameters, click the Set Module Configurations
button, set the INIT to OFF position and power-cycle the CL-200 to take the settings
effect.
Note
The INIT switch does not need to be set in the ON position when changing the
address, baudrate and parity for ModbusRTU communication; users only have to
power-cycle the module after complete configuration.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 65 -
Page 66
5. Configuration via Wi-Fi
The factory default settings for Wi-Fi communication of the CL-200-WF are as
follows.
Mode: AP
Wireless Security: WPA/WPA2, "00000000"
DHCP Server (AP Mode): DHCP Server on, start IP: 192.168.255.2
Wi-Fi Channel (AP Mode): 11
IP Address: 192.168.255.1
Gateway Address: 192.168.255.254
Subnet Mask: 255.255.0.0
SSID: CL-213-WF
Modbus TCP Port: 502
The Wi-Fi IIOT Utility is provided to configure and test the CL-200-WF module through
the Wi-Fi interface.
Install Wi-Fi IIOT Utility
The installation file location of the Wi-Fi IIOT Utility is at:
http://ftp.icpdas.com/pub/cd/iiot/utility/
Search and Find the Module
Click on the search button to find the modules via the Wi-Fi interface.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 66 -
Page 67
Select the Wi-Fi network interface and click on the OK button.
Configure and Test the Module
When the module is found, click on the module name to enter the configuration
form.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 67 -
Page 68
In the Configuration form, you can change the Wi-Fi related settings. Click on the Set
Item
Description
WiFi Mode
This parameter is used to specify the Wi-Fi mode of the SL device. It can be
Station or AP. For AP mode, only one device can be connected.
SSID
This parameter is used to specify the Service Set Identifier. For station
mode, specify the SSID of the AP you would like to connect. For AP
mode, the SSID will be used by the device to be connected.
Encryption
This parameter is used to specify which security protocol is used to secure
wireless computer network. It can be open, WEP, or WPA. It is
recommended to use WPA if possible.
Modbus TCP Port
This parameter is used to set the local port of the Wi-Fi interface to be used
by the Modbus slave device. The default value is 502.
DHCP Server
This parameter is used to specify whether to turn on the DHCP server
function. It is only available to the AP mode.
IP Address Type
This parameter is only available to the station mode and it can be Static or
DHCP. If DHCP is supported by the AP you would like to connect, then
DHCP should be selected. Otherwise, select Static and the following three
parameters Static IP, Subnet Mask and Gateway should be set, too.
Module Configurations button to save the changes to the module.
The followings show the detailed description of each setting.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 68 -
Page 69
WiFi Channel
This parameter is used to specify which channel is used for Wi-Fi
transmission. It can be 1 to 11. It is only available to the AP mode.
Static IP
Each SL device connected to the Wi-Fi network must have its own unique
IP address. This parameter is used to assign a specific IP address.
Subnet Mask
This parameter is used to assign the subnet mask for the SL device. The
subnet mask indicates which portion of the IP address is used to identify the
local network or subnet.
Gateway
This parameter is used to assign the IP address of the gateway to be used by
the SL device. A gateway (or router) is a device that is used to connect an
individual network to one or more additional networks.
In the AI form, you can read the sensor readings such as PM2.5, CO, CO2, humidity,
temperature and dew point temperature.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 69 -
Page 70
In the Alarm form, you can change the alarm related settings. Click on the Set Alarm
Configurations button to save the changes to the module.
In the DO form, you can change the digital output status and the power on and safe
digital output settings.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 70 -
Page 71
In the Data Logger form, you can change the data logger related settings. Click on the
Apply button to save the changes to the module.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 71 -
Page 72
6. Monitoring via Mobile Devices
The iAir App can be used to monitor real-time data of PM2.5, CO/CO2 level, NH3, H2S,
HCHO, TVOC, temperature and humidity anywhere and anytime without any
complicated configuration. The CL-200 modules and your mobile devices such as smart
phones or tablets need be addressed on the same network, and then you can get the
real-time data from CL-200 loggers by entering a specific IP address, or by performing
an automatic search for available devices.
If a CL-200 can’t be searched in the iAir App, please contact with the network
administrator to make sure the module and your mobile devices are addressed on the
same sub-network. It means that they have the same broadcast address.
The iAir app is available to free download in Google Play and App Store. Search “iAir” in
or search “iAir”, “ICPDAS” in App Store and tap on install.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 72 -
Page 73
7. Utility to Get/Manage Data Log
DL-300 Utility is a convenient, easy-to-use management utility running on Windows
platform that allows users to monitor the real-time data and trend chart from CL-200
modules on the Ethernet, it can group the CL-200 modules for group view management,
log alarm events with timestamp, download the logged data from a CL-200 logger and
export the data to *.csv files for performing statistical analysis in Excel.
The DL-300 Utility can be obtained from:
http://ftp.icpdas.com/pub/cd/iiot/utility/
1. Run the DL-300_utility_setup_yyyymmdd.exe, the default install location is
C:\ICPDAS\DL300_Utility\DL-300 Utility
2. Open the DL-300 Utility by double
clicking on the DL-300 Utility shortcut
on desktop.
3. Search out a CL-200 module on the Ethernet and set the configuration.
3-1. Select the Device Settings on the Settings menu.
3-2. Click the Search New Device button to search the CL-200 modules connected
on the same Ethernet network.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 73 -
Page 74
3-3. Check the checkbox next to a module and click the Add button to add the
1
2
3
4
module in the utility.
3-4. Highlight a module and click the Edit Device button to configure the module.
3-5. Set the configuration, andclick on theOKbutton.
Note
Consult your network administrator before making changes to IPAddress/ Mask Address/ Gateway
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 74 -
Page 75
5
4. Get real-time data, trend chart and alarm event.
4-1. Click the List icon to obtain the real-time data. It also lists the connect status,
group information and IP address for every CL-200 logger.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 75 -
Page 76
4-2. Click the Trend icon to display the trend chart. Users can select the radio button
for CO/CO2 level, Temperature or Humidity to access the trend chart for those
real-time data, check the checkbox next to each CL-200 logger to display its
trend chart or uncheck it to cancel display. Drag and drop the trend chart can
move it to see the data not be displayed in the chart.
4-3. Click the Alarm icon to review the alarm events.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 76 -
Page 77
4-4. Modify the event condition.
1
2
Select the System Settings on the Settings menu.
Set the CO/CO2 Alert Value, CO/CO2 Alarm Value (If it is supported in the
logger), Temperature Alert Value and Temperature Alarm Value for trigger
events. Check the checkbox next to The Record Time Everyday can schedule
auto generate report everyday at the time set in the dropdown menu. Click on
the OK button to complete the settings.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 77 -
Page 78
5. Download data in a CL-200 logger and export the data
2
3
1
5.1. Select Update Database on the Settings menu
5.2. Click the Start button to download the data in CL-200 modules.
5.3. Click the close icon to exit the download procedure when all data are
downloaded.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 78 -
Page 79
5.4. Click the Query icon.
4
5
5.5. Highlight the desired module, set the Start Time and End Time, and then click
the Search button. The data in the time period will be listed as below.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 79 -
Page 80
5.6. Click the Export button to export the searched data in *.csv files for performing
6
statistical analysis in Excel.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 80 -
Page 81
6. Group the devices by location or users
2
1
3
6.1. Select Group Settings on the Settings menu.
6.2 Click the New button, enter the group name and click the OK button in the
pop-up box, and then click the OK button in the Group Settings box.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 81 -
Page 82
6-3. Select Device Settings on the Settings menu; highlight the desired device
1 2 3
and click the Edit Device button, select the group name for the module and
click the OK button in the pop-up Device Property box to complete the setting.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 82 -
Page 83
6-4. Monitor the group data by clicking the Group icon and then highlighting the
group name.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 83 -
Page 84
8. FAQ
Q1: What is ABC (Automatic Baseline Correction)?
A1: ABC stands for the Automatic Baseline Correction which is used to adjust a shifted
baseline to the carbon dioxide level in fresh air. In case of normal indoor application,
the carbon dioxide level drops to nearly outside air where there are no human,
green plants or anything to elevate the carbon dioxide levels on weekday evenings
or weekends, the ABC algorithm constantly keeps track of the lowest reading and
slowly corrects it as the expected value in fresh air typically around 400 ppm.
Q2: Why I need to enable the ABC?
A2: When the CO2 concentration detected in a period time of unoccupied space is
greater than the base value of 400ppm, enable the ABC function to adjust the
baseline. Be careful that the ABC will not work if a space is constantly occupied
such as a hospital, 24-hr factory, 24-hr store, green house or other applications
where CO2 levels may be elevated at all times.
Q3: Does the CL-202/CL-212/CL-203/CL-213 enable the ABC as the factory default
setting?
A3: No, the ABC is default disabled in a CL-202/CL-212/CL-203/CL-213 logger to
prevent the baseline from being adjusted to an incorrect value in case of using in a
constantly occupied space.
Q4: What to do when the ABC is no work?
A4: When the ABC is no work regarding baseline correction, the
CL-202/CL-212/CL-203/CL-213 needs be returned to ICP DAS.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 84 -
Page 85
Q5: How to set the Accessible IP?
A5: Enter the IP address for your logger in the address bar of a web browser and go to
the Accessible IP Settings page, select the radio button next to
Add ___.___.___.___ To The List and key in the IP for a device which is allowed to
access the CL-200, and then click the submit button.
Check the checkbox next to the Save to Flash before clicking the submit button to
save the IP setting and use after repowering. Once any of those in the list is set,
only the device for which the IP address is saved in the list can assess the CL-200.
Q6: How to delete the Accessible IP settings?
A6: Enter the IP address for your logger in the address bar of a web browser and go to
the Accessible IP Settings page, select the radio button next to Delete IP# to delete
a IP by the IP number or select the radio button next tot Delete All and then click the
submit button.
Check the checkbox next to the Save to Flash before clicking the submit button to
save the IP setting and use after repowering.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 85 -
Page 86
Q7: How to clear the data logged in a CL-200 module?
A7: Enter the IP address for the module in the address bar of a web browser and go to
the I/O Settings page, click the Reset Data Logger button at the bottom of the page.
CL-200 Series User Manual Version 1.2.0 Apr. 2019 - 86 -
Page 87
Appendix A: DCON Command Sets
Command
Description
$AAF
read firmware version
$AAI
read INIT status
response:
!AA0 -> INIT short to GND
!AA1 -> else
Set Modbus RTU/DCON protocol
N-> 0: DCON, 1: Modbus RTU
$AA2
read configuration
$AA5
read reset status
!AA1 first after power on, !AA0 others
#AA
Read All Analog Inputs
response
> (CO in 1 ppm) (relative humidity in 0.01%)(temperature in
0.01°C)(temperature in 0.01°F) (dew point temperature in
0.01°C)(dew point temperature in 0.01°F)
#AAN
Read Channel Analog Input
N = 0 for CO in 1 ppm, 1 for relative humidity in 0.01%, 2 for
temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew point
temperature in 0.01°C, 5 for dew point temperature in 0.01°F
%AANNTTCCFF
set configuration, NN: new address, TT = 00, CC: new baud rate
FF: data format
@AABA
Read beep on alarm time
response
!AAHH, HH in hex, 0: disabled, 1 ~ 250: beep on alarm
time in seconds, 251: beep on alarm continuously
A-1. CL-201-E DCON Command Sets
A-1. CL-201-E DCON Command Sets - 87 -
Page 88
Command
Description
@AABAHH
Set beep on alarm time
HH in hex, 0: disabled, 1 ~ 250: beep on alarm time in seconds,
251: beep on alarm continuously
@AABE
Read enable/disable beep on alarm
response
!AAHH, HH in hex, bit 0 for channel 0, bit 1 for channel 1,
etc, for each bit, 0: disabled, 1: enabled
@AABEHH
Enable/disable beep on alarm
HH in hex, , bit 0 for channel 0, bit 1 for channel 1, etc, for each bit,
0: disabled, 1: enabled
@AACH
Clear all high latched analog inputs to the current values
@AACHN
Clear channel high latched analog input to the current value, N = 0
for CO, 1 for relative humidity, 2 for temperature in 0.01°C, 3 for
temperature in 0.01°F, 4 for dew point temperature in 0.01°C, 5 for
dew point temperature in 0.01°F
@AACHCN
Clear high latched alarm of a channel, N = 0 for CO, 1 for relative
humidity, 2 for temperature in 0.01°C, 3 for temperature in 0.01°F, 4
for dew point temperature in 0.01°C, 5 for dew point temperature in
0.01°F
@AACL
Clear all low latched analog inputs to the current values
@AACLN
Clear channel low latched analog input to the current value, N = 0
for CO, 1 for relative humidity, 2 for temperature in 0.01°C, 3 for
temperature in 0.01°F, 4 for dew point temperature in 0.01°C, 5 for
dew point temperature in 0.01°F
@AACLCN
Clear low latched alarm of a channel, N = 1 for relative humidity, 2
for temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew
point temperature in 0.01°C, 5 for dew point temperature in 0.01°F
@AADACN
Disable AI alarm of a channel, N = 0 for CO2, 1 for relative humidity,
2 for temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew
point temperature in 0.01°C, 5 for dew point temperature in 0.01°F
@AADI
read DO
response
!AA00O00
@AADLB
Read the beginning of the period setting of the data logger for
period logging mode
response
!AAyyyymmddhhmmss,
A-1. CL-201-E DCON Command Sets - 88 -
Page 89
Command
Description
@AADLByyyymmd
dhhmmss
Set the beginning of the period setting of the data logger for period
logging mode
yyyy: year, 2000 to 2199, mm: month, 01 to 12, dd: date, 01 to 31,
hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second, 0 to 59
@AADLC
Read the data logger command
response
!AAh, 0: stop, 1: run, 2: run in period mode
@AADLCh
Set the data logger command, h->0: stop, 1: run, 2: run in period
mode
@AADLE
Read the ending of the period setting of the data logger for period
logging mode
response
!AAyyyymmddhhmmss
@AADLEyyyymmd
dhhmmss
Set the ending of the period setting of the data logger for period
logging mode
yyyy: year, 2000 to 2199, mm: month, 01 to 12, dd: date, 01 to 31,
hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second, 0 to 59
@AADLN
Read number of log records in the data logger
response
!AAhhhhhhhh, hhhhhhhh in hex format
@AADLO
Read the overwriting mode when data logger is full
response
!AAh, 0: stop logging when full, 1: overwrite
@AADLOh
Set the overwriting mode when data logger is full
h->0: stop logging when full, 1: overwrite
@AADLP
Read the samplig period setting of the data logger
response
!AAhhmmss, hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second,
0 to 59
@AADLPhhmmss
Set the samplig period setting of the data logger
@AADLS
Read logging status of the data logger
response
!AAhh, hh in hex format, 00: stopped, 01: running, others: error
@AADO0V
set DO, V-> 0: off, 1: on
A-1. CL-201-E DCON Command Sets - 89 -
Page 90
Command
Description
@AAEATCN
Enable AI alarm of a channel, N = 0 for CO, 1 for relative humidity, 2
for temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew
point temperature in 0.01°C, 5 for dew point temperature in 0.01°F
T->M: momentary alarm, L: latched alarm
@AAHI(data)CN
Set high alarm limit of an AI channel, N = 0 for CO in 1ppm, 1 for
relative humidity in 0.01%, 2 for temperature in 0.01°C, 3 for
temperature in 0.01°F, 4 for dew point temperature in 0.01°C, 5 for
dew point temperature in 0.01°F
@AAHO
Read humidity offset
@AAHO(data)
Set humidity offset, data in format of -100.00 ~ +100.00
@AALO(data)CN
Set low alarm limit of an AI channel, N = 1 for relative humidity in
0.01%, 2 for temperature in 0.01°C, 3 for temperature in 0.01°F, 4
for dew point temperature in 0.01°C, 5 for dew point temperature in
0.01°F
@AARACN
Read AI alarm enabled/disabled status of a channel
response
!AAN, N->0: disabled, 1: momentary, 2: latched
@AARAO
Read AI alarm status
response
!AAHHLL
@AARH
Read all high latched values of analog input channels
@AARHN
Read channel high latched value of analog input
@AARHCN
Read high alarm limit of an AI channel
@AARL
Read all low latched values of analog input channels
@AARLN
Read channel low latched value of analog input
@AARLCN
Read low alarm limit of an AI channel
@AART
Read RTC data
response
!AAYYYYMMDDHHMMSS
@AARTYYYYMM
DDHHMMSS
Set RTC data
@AATO
Read temperature offset in 0.01°C
@AATO(data)
Set temperature offset in 0.01°C, -100.00 ~ +100.00
~**
clear host watchdog timeout counter
~AA0
read host watchdog status
~AA1
clear host watchdog timeout status
~AA2
read host watchdog enable/disable status and timeout value
A-1. CL-201-E DCON Command Sets - 90 -
Page 91
Command
Description
~AA3ETT
enable/disable host watchdog and set timeout value
E-> 0: disable host watchdog, 1: enable host watchdog
TT: host watchdog timeout in 0.1s in hex format
~AA4
read DO power on and safe value
~AA50P0S
set DO power on and safe value
P-> 0: power on value off, 1: power on value on
S-> 0: safe value off, 1: safe value on
~AARD
read response delay time in ms in hex format
~AARDVV
set response delay time in ms, VV in hex format, 00 - 1E
A-1. CL-201-E DCON Command Sets - 91 -
Page 92
A-2. CL-202-E DCON Command Sets
Command
Description
$AAF
read firmware version
$AAI
read INIT status
response:
!AA0 -> INIT short to GND
!AA1 -> else
Set Modbus RTU/DCON protocol
N-> 0: DCON, 1: Modbus RTU
$AA2
read configuration
$AA5
read reset status
!AA1 first after power on, !AA0 others
#AA
Read All Analog Inputs
response
> (CO2 in 1 ppm) (relative humidity in 0.01%)(temperature in
0.01°C)(temperature in 0.01°F) (dew point temperature in
0.01°C)(dew point temperature in 0.01°F)
#AAN
Read Channel Analog Input
N = 0 for CO2 in 1 ppm, 1 for relative humidity in 0.01%, 2 for
temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew point
temperature in 0.01°C, 5 for dew point temperature in 0.01°F
%AANNTTCCFF
set configuration, NN: new address, TT = 00, CC: new baud rate
FF: data format
@AAABC
Read status of the automatic baseline correction
response
!AAN, N=0: disabled, 1: enabled
@AAABCN
Set the automatic baseline correction
N->0: disabled, 1: enabled
@AABA
Read beep on alarm time
response
!AAHH, HH in hex, 0: disabled, 1 ~ 250: beep on alarm
time in seconds, 251: beep on alarm continuously
A-2. CL-202-E DCON Command Sets - 92 -
Page 93
Command
Description
@AABAHH
Set beep on alarm time
HH in hex, 0: disabled, 1 ~ 250: beep on alarm time in seconds,
251: beep on alarm continuously
@AABE
Read enable/disable beep on alarm
response
!AAHH, HH in hex, bit 0 for channel 0, bit 1 for channel 1,
etc, for each bit, 0: disabled, 1: enabled
@AABEHH
Enable/disable beep on alarm
HH in hex, , bit 0 for channel 0, bit 1 for channel 1, etc, for each bit,
0: disabled, 1: enabled
@AACH
Clear all high latched analog inputs to the current values
@AACHN
Clear channel high latched analog input to the current value, N = 0
for CO2, 1 for relative humidity, 2 for temperature in 0.01°C, 3 for
temperature in 0.01°F, 4 for dew point temperature in 0.01°C, 5 for
dew point temperature in 0.01°F
@AACHCN
Clear high latched alarm of a channel, N = 0 for CO2, 1 for relative
humidity, 2 for temperature in 0.01°C, 3 for temperature in 0.01°F, 4
for dew point temperature in 0.01°C, 5 for dew point temperature in
0.01°F
@AACL
Clear all low latched analog inputs to the current values
@AACLN
Clear channel low latched analog input to the current value, N = 0
for CO2, 1 for relative humidity, 2 for temperature in 0.01°C, 3 for
temperature in 0.01°F, 4 for dew point temperature in 0.01°C, 5 for
dew point temperature in 0.01°F
@AACLCN
Clear low latched alarm of a channel, N = 1 for relative humidity, 2
for temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew
point temperature in 0.01°C, 5 for dew point temperature in 0.01°F
@AADACN
Disable AI alarm of a channel, N = 0 for CO2, 1 for relative humidity,
2 for temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew
point temperature in 0.01°C, 5 for dew point temperature in 0.01°F
@AADI
read DO
response
!AA00O00
@AADLB
Read the beginning of the period setting of the data logger for
period logging mode
response
!AAyyyymmddhhmmss,
A-2. CL-202-E DCON Command Sets - 93 -
Page 94
Command
Description
@AADLByyyymmd
dhhmmss
Set the beginning of the period setting of the data logger for period
logging mode
yyyy: year, 2000 to 2199, mm: month, 01 to 12, dd: date, 01 to 31,
hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second, 0 to 59
@AADLC
Read the data logger command
response
!AAh, 0: stop, 1: run, 2: run in period mode
@AADLCh
Set the data logger command, h->0: stop, 1: run, 2: run in period
mode
@AADLE
Read the ending of the period setting of the data logger for period
logging mode
response
!AAyyyymmddhhmmss
@AADLEyyyymmd
dhhmmss
Set the ending of the period setting of the data logger for period
logging mode
yyyy: year, 2000 to 2199, mm: month, 01 to 12, dd: date, 01 to 31,
hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second, 0 to 59
@AADLN
Read number of log records in the data logger
response
!AAhhhhhhhh, hhhhhhhh in hex format
@AADLO
Read the overwriting mode when data logger is full
response
!AAh, 0: stop logging when full, 1: overwrite
@AADLOh
Set the overwriting mode when data logger is full
h->0: stop logging when full, 1: overwrite
@AADLP
Read the samplig period setting of the data logger
response
!AAhhmmss, hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second,
0 to 59
@AADLPhhmmss
Set the samplig period setting of the data logger
@AADLS
Read logging status of the data logger
response
!AAhh, hh in hex format, 00: stopped, 01: running, others: error
@AADO0V
set DO, V-> 0: off, 1: on
A-2. CL-202-E DCON Command Sets - 94 -
Page 95
Command
Description
@AAEATCN
Enable AI alarm of a channel, N = 0 for CO2, 1 for relative humidity,
2 for temperature in 0.01°C, 3 for temperature in 0.01°F, 4 for dew
point temperature in 0.01°C, 5 for dew point temperature in 0.01°F
T->M: momentary alarm, L: latched alarm
@AAHI(data)CN
Set high alarm limit of an AI channel, N = 0 for CO2 in 1ppm, 1 for
relative humidity in 0.01%, 2 for temperature in 0.01°C, 3 for
temperature in 0.01°F, 4 for dew point temperature in 0.01°C, 5 for
dew point temperature in 0.01°F
@AAHO
Read humidity offset
@AAHO(data)
Set humidity offset, data in format of -100.00 ~ +100.00
@AALO(data)CN
Set low alarm limit of an AI channel, N = 1 for relative humidity in
0.01%, 2 for temperature in 0.01°C, 3 for temperature in 0.01°F, 4
for dew point temperature in 0.01°C, 5 for dew point temperature in
0.01°F
@AARACN
Read AI alarm enabled/disabled status of a channel
response
!AAN, N->0: disabled, 1: momentary, 2: latched
@AARAO
Read AI alarm status
response
!AAHHLL
@AARH
Read all high latched values of analog input channels
@AARHN
Read channel high latched value of analog input
@AARHCN
Read high alarm limit of an AI channel
@AARL
Read all low latched values of analog input channels
@AARLN
Read channel low latched value of analog input
@AARLCN
Read low alarm limit of an AI channel
@AART
Read RTC data
response
!AAYYYYMMDDHHMMSS
@AARTYYYYMM
DDHHMMSS
Set RTC data
@AATO
Read temperature offset in 0.01°C
@AATO(data)
Set temperature offset in 0.01°C, -100.00 ~ +100.00
~**
clear host watchdog timeout counter
~AA0
read host watchdog status
~AA1
clear host watchdog timeout status
~AA2
read host watchdog enable/disable status and timeout value
A-2. CL-202-E DCON Command Sets - 95 -
Page 96
Command
Description
~AA3ETT
enable/disable host watchdog and set timeout value
E-> 0: disable host watchdog, 1: enable host watchdog
TT: host watchdog timeout in 0.1s in hex format
~AA4
read DO power on and safe value
~AA50P0S
set DO power on and safe value
P-> 0: power on value off, 1: power on value on
S-> 0: safe value off, 1: safe value on
~AARD
read response delay time in ms in hex format
~AARDVV
set response delay time in ms, VV in hex format, 00 - 1E
A-2. CL-202-E DCON Command Sets - 96 -
Page 97
A-3. CL-203-E DCON Command Sets
Command
Description
$AAF
read firmware version
$AAI
read INIT status
response:
!AA0 -> INIT short to GND
!AA1 -> else
Set Modbus RTU/DCON protocol
N-> 0: DCON, 1: Modbus RTU
$AA2
read configuration
$AA5
read reset status
!AA1 first after power on, !AA0 others
#AA
Read All Analog Inputs
response
>(CO in 1 ppm) (CO2 in 1 ppm) (relative humidity in
0.01%)(temperature in 0.01°C)(temperature in 0.01°F) (dew
point temperature in 0.01°C)(dew point temperature in
0.01°F)
#AAN
Read Channel Analog Input
N = 0 for CO in 1 ppm, 1 for CO2 in 1 ppm, 2 for relative humidity in
0.01%, 3 for temperature in 0.01°C, 4 for temperature in 0.01°F, 5
for dew point temperature in 0.01°C, 6 for dew point temperature in
0.01°F
%AANNTTCCFF
set configuration, NN: new address, TT = 00, CC: new baud rate
FF: data format
@AAABC
Read status of the automatic baseline correction
response
!AAN, N=0: disabled, 1: enabled
@AAABCN
Set the automatic baseline correction
N->0: disabled, 1: enabled
A-3. CL-203-E DCON Command Sets - 97 -
Page 98
Command
Description
@AABA
Read beep on alarm time
response
!AAHH, HH in hex, 0: disabled, 1 ~ 250: beep on alarm
time in seconds, 251: beep on alarm continuously
@AABAHH
Set beep on alarm time
HH in hex, 0: disabled, 1 ~ 250: beep on alarm time in seconds,
251: beep on alarm continuously
@AABE
Read enable/disable beep on alarm
response
!AAHH, HH in hex, bit 0 for channel 0, bit 1 for channel 1,
etc, for each bit, 0: disabled, 1: enabled
@AABEHH
Enable/disable beep on alarm
HH in hex, , bit 0 for channel 0, bit 1 for channel 1, etc, for each bit,
0: disabled, 1: enabled
@AACH
Clear all high latched analog inputs to the current values
@AACHN
Clear channel high latched analog input to the current value, N = 0
for CO, 1 for CO2, 2 for relative humidity, 3 for temperature in
0.01°C, 4 for temperature in 0.01°F, 5 for dew point temperature in
0.01°C, 6 for dew point temperature in 0.01°F
@AACHCN
Clear high latched alarm of a channel, N = 0 for CO, 1 for CO2, 2 for
relative humidity, 3 for temperature in 0.01°C, 4 for temperature in
0.01°F, 5 for dew point temperature in 0.01°C, 6 for dew point
temperature in 0.01°F
@AACL
Clear all low latched analog inputs to the current values
@AACLN
Clear channel low latched analog input to the current value, N = 0
for CO, 1 for CO2, 2 for relative humidity, 3 for temperature in
0.01°C, 4 for temperature in 0.01°F, 5 for dew point temperature in
0.01°C, 6 for dew point temperature in 0.01°F
@AACLCN
Clear low latched alarm of a channel, N = 2 for relative humidity, 3
for temperature in 0.01°C, 4 for temperature in 0.01°F, 5 for dew
point temperature in 0.01°C, 6 for dew point temperature in 0.01°F
@AADACN
Disable AI alarm of a channel, N = 0 for CO, 1 for CO2, 2 for relative
humidity, 3 for temperature in 0.01°C, 4 for temperature in 0.01°F, 5
for dew point temperature in 0.01°C, 6 for dew point temperature in
0.01°F
A-3. CL-203-E DCON Command Sets - 98 -
Page 99
Command
Description
@AADI
read DO
response
!AA00O00
@AADLB
Read the beginning of the period setting of the data logger for
period logging mode
response
!AAyyyymmddhhmmss,
@AADLByyyymmd
dhhmmss
Set the beginning of the period setting of the data logger for period
logging mode
yyyy: year, 2000 to 2199, mm: month, 01 to 12, dd: date, 01 to 31,
hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second, 0 to 59
@AADLC
Read the data logger command
response
!AAh, 0: stop, 1: run, 2: run in period mode
@AADLCh
Set the data logger command, h->0: stop, 1: run, 2: run in period
mode
@AADLE
Read the ending of the period setting of the data logger for period
logging mode
response
!AAyyyymmddhhmmss
@AADLEyyyymmd
dhhmmss
Set the ending of the period setting of the data logger for period
logging mode
yyyy: year, 2000 to 2199, mm: month, 01 to 12, dd: date, 01 to 31,
hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second, 0 to 59
@AADLN
Read number of log records in the data logger
response
!AAhhhhhhhh, hhhhhhhh in hex format
@AADLO
Read the overwriting mode when data logger is full
response
!AAh, 0: stop logging when full, 1: overwrite
@AADLOh
Set the overwriting mode when data logger is full
h->0: stop logging when full, 1: overwrite
@AADLP
Read the samplig period setting of the data logger
response
!AAhhmmss, hh: hour, 0 to 24, mm: minute, 0 to 59, ss: second,
0 to 59
A-3. CL-203-E DCON Command Sets - 99 -
Page 100
Command
Description
@AADLPhhmmss
Set the samplig period setting of the data logger
@AADLS
Read logging status of the data logger
response
!AAhh, hh in hex format, 00: stopped, 01: running, others: error
@AADO0V
set DO, V-> 0: off, 1: on
@AAEATCN
Enable AI alarm of a channel, N = 0 for CO, 1 for CO2, 2 for relative
humidity, 3 for temperature in 0.01°C, 4 for temperature in 0.01°F, 5
for dew point temperature in 0.01°C, 6 for dew point temperature in
0.01°F
T->M: momentary alarm, L: latched alarm
@AAHI(data)CN
Set high alarm limit of an AI channel, N = 0 for CO in 1ppm, 1 for
CO2 in 1ppm, 2 for relative humidity in 0.01%, 3 for temperature in
0.01°C, 4 for temperature in 0.01°F, 5 for dew point temperature in
0.01°C, 6 for dew point temperature in 0.01°F
@AAHO
Read humidity offset
@AAHO(data)
Set humidity offset, data in format of -100.00 ~ +100.00
@AALO(data)CN
Set low alarm limit of an AI channel, N = 2 for relative humidity in
0.01%, 3 for temperature in 0.01°C, 4 for temperature in 0.01°F, 5
for dew point temperature in 0.01°C, 6 for dew point temperature in
0.01°F
@AARACN
Read AI alarm enabled/disabled status of a channel
response
!AAN, N->0: disabled, 1: momentary, 2: latched
@AARAO
Read AI alarm status
response
!AAHHLL
@AARH
Read all high latched values of analog input channels
@AARHN
Read channel high latched value of analog input
@AARHCN
Read high alarm limit of an AI channel
@AARL
Read all low latched values of analog input channels
@AARLN
Read channel low latched value of analog input
@AARLCN
Read low alarm limit of an AI channel
@AART
Read RTC data
response
!AAYYYYMMDDHHMMSS
@AARTYYYYMM
DDHHMMSS
Set RTC data
A-3. CL-203-E DCON Command Sets - 100 -
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.