ICP DAS Co., LTD PM-3133 Quick Start V1.20 Page 2
3.1.
Dimension: 127mm(length)× 33mm(wide)× 105mm(high)
Products come with external split type clip-on CT’s. Disconnect the CT’s or use other CT’s is highly
prohibited.
Please read this operation manual carefully before using.
Please re-confirm the measure position.
Reconfirm the RST (ABC) phase sequence of the power system.
PM-3133 series can be installed as rail mounting mode or embedded, no need to drill a hole or screw to fix it
(rail mounting width can up to the length of 35 mm).
Meter auxiliary power for PM-3133 series is DC +12V ~+48V.
3.2. V oltage Input
1. PM-3133 series: Input Voltage up to 500V.
For any higher Input Voltage large than 500V, please add the PT (power transformer), and Change PT RATIO
setup.
2. Confirm the RST (ABC) phase sequence.
3.3. Current Input
1. The external CT’s are fragile, please handle with care.
2. The current input of PM-3133 series is in mA range. Only the ex-factory attached CT’s can be used. The other
CT’s, for example, from panel will damage the instrument due to its large current (around 5A).
3. The current input of PM-3133-xxxP series is in mV range. Only the ex-factory attached CT’s can be used. The
other CT’s, for example, from panel will damage the instrument due to its large current (around 5A or mA).
4. When more than one smart meter (PM-3133 series) are installed, please do not disconnect the CT with its
original meter and mix use with each other. Since each set of smart meter (PM-3133 series) and its attached
split type clip-on CT are calibrated set by set. The mix use may cause wrong measurements.
5. To install CT’s correctly, please ensure the CT lines sequences is right before clip the CT’s onto the power
cable of the monitoring equipment. (Detail will be found in next section)
6. When measuring the current, the secondary circuit of a CT should never be opened
when a load is passing through its primary. Make sure you always open the CT clip to
detach the CT before removing the terminal lines. Otherwise, it will cause severe injury.
7. Please handle with extra care, especially when the operation space of CT’s is limited.
8. The current direction must follo w K -L marked on CT’s.
9. Please select the right size CT’s for different size of monitoring equipment cables:
power cable diameter <Φ10 use 60A CT,Φ10~Φ16 use 100A CT,Φ16~Φ24 use 200A CT
,Φ24~Φ36 use 300A or 400A CT。
10. The maximum current value cannot exceed the CT rating.