The information contained in this is subject to change without notice. IBP Instruments GmbH, its distributors and
subsidiaries take no responsibility for any errors or omissions in this document. The contained software is being
delivered on the basis of a general licence contract or in single license. Use or reproduction of the software is allowed
only in agreement with the contractual arrangements. Whoever transfers this software and/or this manual on magnetic
tape, diskette or any other media, except for the purpose of own use, without written authorization of the IBP
Instruments GmbH, is liable to prosecution.
Copyright (C) 1997, 2002 IBP Instruments GmbH.
All rights reserved
Publishers: IBP Instruments GmbH IBP Medical Inc
Sutelstr. 7a 3853 E Edna Ave
D 30659 Hannover Phoenix, AZ 85032
Germany USA
Manual author: Dipl. Ing. Werner Pfingstmann Christopher Weed
Rev 1 August 2002 Temperature Coefficient for B.Braun now 2.10%/°C
Rev 2 September 2002 Temperature Coefficient for Gambro now 2,07%/°C
Specification: Temp. Accuracy 25 … 40°C ± 0.07 °C otherwise ± 0.1 °C
This user guide is only valid for software releases 3.6x (x = 0..9)
Conformity Declaration in accordance with Medical Product Law
Flow Measuring extended
Window Entrance Test added
Safety Instructions
For your own safety, and the safety of your patients, follow these important safety
instructions as well as other safety instructions noted throughout this User Guide.
• Please read the User Guide in its entirety before using the HDM99.
• Keep away the device from unauthorized persons.
• Never use the HDM99 on a dialysis machine to which a patient is connected.
• Never use the HDM99 on a dialysis machine with a connected battery charger.
• Never use the HDM99 in place of the hemodialysis machine’s primary sensors.
• Operate the instrument only in a dry environment, and do not touch it with damp
hands.
• Ensure that no fluids intrude into the interior of the device or into the sockets at
the front.
• Apply a clean Transducer Protector to the nozzle for the pressure measurement.
• Verify accurate function of the meter before taking measurements or whenever
inaccurate readings are suspected.
• Avoid a discharge of static electricity over the sockets. It could lead to the
destruction of your instrument. Before touching the sockets and lines that are
connected with them, dissipate any static electric charge that may be present in
your body.
• Only calibrate the HDM99 if you have understood the consequences to their full
extent. Use only the recommended standards later in this manual.
• Do not use abrasive cleaning agents and/or full strength bleach or acid to clean
HDM99 or the electrodes as this will cause damages.
• Voltages above 40 V can be dangerous. Take extreme care while working with
higher voltages.
• To avoid electrical shocks, and/or damage to your measuring system, do not apply
excess voltages above 40 V to the instrument.
• Do not open the device.
• To avoid current loops use the RS232-Interface only with notebooks which are not
connected with the battery charger.
• Federal law restricts the use of this device to sale by or on the order of a
The HDM99 is a self-contained hemodialysis test instrument designed for the measurement of
Conductivity
Temperature
Pressure
pH
Flow
Voltage
Frequency
Period time
Pulses..
This innovative device has been designed to be user-friendly, reliable, and rugged enough for
the demanding needs of hemodialysis technicians. In addition, the HDM99 may also be used
for environmental measurement and other non-medical applications.
The measured values are displayed numerically on the instrument’s large screen. The instrument
also may be connected to a computer, and using the supplied software, displayed, stored and
analyzed.
Please take a few moments upon initial receipt of your shipment to ensure that all the items
listed below have been included. In the event of a discrepancy, contact your supplier
immediately. Be sure to read this User Guide in its entirety before first using the product.
Applications for use
The HDM99 may be used by hemodialysis personnel to test the conductivity, temperature,
pressure, pH and flow of the dialysate solution used with hemodialysis delivering systems.
The HDM99 may also be used to test the conductivity/temperature and pH of acid
and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications.
The HDM99 may also be used to test the voltage and alternating signals in hemodialysis
delivering systems.
Introduction 4
EC Declaration of Conformity
according to the Council Directive 93/42/EEC
concerning medical devices
We
IBP Instruments GmbH
Sutelstraße 7A
30659 Hannover
declare under our sole responsibility that the product:
HDM99, Art.-Nr.: 31.0006.
meet the provisions of the Council Directive 93/42/EEC concerning medical devices which
apply to them.
CE – certified by : Medcert GmbH Hamburg, Germany Nr.: 0482
Date of issue: 09.08.02
Authorized person : Dipl. Ing. Werner Pfingstmann (Managing director)
5 Introduction
Delivery Content
Introduction 6
Pos. Standard Option
1 HDM99 x
2 Conductivity/Temperature Probe (CTP) x
3 Flow Through Adapter for CTP x
4 Battery Charger with country specific adapter x
5 Tube inc. Male Connector for Pressure Measurement x
6 13,63 or 14.00 mS/cm Conductivity Standard Solution x
7 User Manual x
8 Interface Lead Voltage x
9 Interface Lead Frequency/Period Time/Pulses x
10 Test prod x
11 Case x
12 Flow Probe x
13 pH Probe with Reference Solution x
14 HDMView Software and RS232 Interface Cable x
7 Introduction
Warranty
IBP Instruments GmbH warrants that it will repair or replace, at its option, any defective or
malfunctioning part without charge for the terms listed below. Parts used for repair or
replacement are warranted for the remaining warranty period only.
The user must deliver, at its own expense, the product to IBP Instruments Inc., USA, from any
country in America or IBP Instruments GmbH, Germany from all other countries.
Parts Warranty Terms* Conditions
HDM99 60 months Annual calibration through IBP**
Conductivity/Temperature Probe 24 months Annual calibration through IBP**
HDM99 36 months No
Conductivity/Temperature Probe 12 months No
* from date original purchase of original purchaser
** Calibration service must be performed by IBP Instruments every year
(maximal plus 2 months of the first calibration) without interruption
The warranty does not cover:
• Disposable parts as battery or pH electrode
• Annually calibration
• Cell cleaning
• Defects caused by:
1. Modification, alteration, repair or service of the product by anyone
other than IBP Instruments or an authorized service center
2. misuse due to negligence or accident
3. operation or maintenance of the product in a manner contrary
to IBP instructions
Any express warranty not provided herein, and any remedy for breach of contract that but for
this provision might arise by implication or operation of law, is hereby excluded and disclaimed.
The implied warranties of merchantability and of fitness for any particular purpose are expressly
limited to the terms mentioned above. Some states do not allow limitations on the duration of an
implied warranty, so the above limitation may not apply to you.
Under no circumstances shall IBP Instruments GmbH be liable to the original purchaser or to
any other person for any special or consequential damages, whether arising out of breach of
warranty, breach of contract, or otherwise. Some states do not allow the exclusion or limitation
of special or consequential damages, so the above exclusion or limitation may not apply to you.
Introduction 8
This warranty gives you specific legal rights, and you may also have other rights that vary from
state to state.
For further warranty information, contact IBP Instruments GmbH.
9 Introduction
t
r
Product Overview
Voltage Connecto
Flow/Alternating Signal Connector
Conductivity/Temperature Connector
Measuring Channels
Pressure Inle
pH Connector
Only use accessories provided with the HDM99. Use of other devices may damage the
instrument and will void the warranty.
Conductivity/Temperature Transducer
The Conductivity/Temperature probe is a quadropole design for greater accuracy and longevity
than other designs. Take care to properly align the pins in the connector when attaching the
transducer.
Pressure
To avoid damage or inaccurate measurement, ensure that no fluids enter the instrument. We
recommend the use of a Transducer Protector for this purpose.
pH
The input connector is designed exclusively for pH electrodes. Use of other devices may
destroy the instrument.
Introduction 10
Flow
The connection is to be found at the right side of the device.
(8 pol. connector)
Frequency, Periodicity and Events
The connection on the right side of the device is also used for the measuring of the flow. The
incoming voltage range is +5...+24 V. The form of the signal should be of a rectangular form
and has to be positive.
Voltage
The sockets are intended only for AC and DC voltages up to ± 40 V. They are protected up
to ± 80 V.
Battery Charger
The HDM99 is equipped with an internal NiMH battery for extended run times when not
connected to the external charger. The charger provides a 12V/500mA charge to the battery. A
message appears on the display when the battery requires charging. For further details, consult
the Maintenance section of this User Guide.
11 Introduction
Use of the Flow Through Adapter
The Conductivity/Temperature Transducer may be used in
two ways.
In the Flow-Through Mode (illustrated at left) the user may
measure pressure, temperature and conductivity with the
included Flow Through Adapter. The preferred position of
the Hansen connector is vertical, the axial supply is show at
the bottom of the illustration. In the Flow-Through Mode the
axial supply is used as input and the radial supply as output.
The pressure probe in the HDM99 is fastened to the radial
connection above the input. Before any measurements you
must shake the connector to assure that any air within the
system escapes.
In the Dip Mode temperature and conductivity can be
measured. For this mode open the screw cap at the upper end
of the connector and extract the combined electrode. You
must not remove the tube covering the probe carrier. Dip the
electrode into the media up to a level above the holes in the
covering tube. By moving the electrode you assure that the
trapped air can escape and the temperature is well balanced.
NOTE: Do not touch the surface of the electrode with
your fingers.
Introduction 12
Switching on the unit
The unit is switched on and off with the keys marked I for on and O for off. The unit also
switches itself off automatically when the switch-off time after the last key activation is over
or the battery is fully discharged.
Main Menu
After switching on the device you receive an information menu for approx. two seconds. After
that the system goes back to the last setting. In some cases you will receive the following main
menu.
Activation of the function keys leads to
display of the info-menu
display of the measuring menu
display of the calibration menu
display of the installation menu
13 Introduction
Info
From the root menu via info you receive the following information:
There are device specific data like product name, software version and version date as well as
the serial number of your HDM99.
Introduction 14
Installation
Language
By pressing the function key beside the menu point language you will reach the language
choice. With the keys beside the arrows you can select the desired language for your work with
HDM99. The selection is shown inverted. With activation of ok you will return to the
installation menu.
After altering the language the entire menu leadership and display appears in the chosen
language.
Switch-off time
The switch-off time is the time after the last key activation. At the end of this time the device
switches off automatically.
The activation of the function switch-off leads to a selection of times which can be selected with
the arrow-keys. The chosen time is shown inverted and taken over with ok as new switch-off
time in the installation menu. If >--< is chosen, the automatic power shutoff is cleared, the
device remains active until it is switched off manually with the key >OFF<.
15 Introduction
Adjust-reset
With the adjust-reset the calibration values for all measuring channels can be set to basic values.
These basic values are the calibration values of the company conducted calibration. Operate the
function key beside this menu point. You will have to enter a code via the numeric keyboard.
This code is 1704. When you confirm your input with >enter< you will receive the following
images:
With the function keys beside the individual measuring channels you can decide on which
calibration values this function should be applied. The calibration values will be reset to basic
values.
If for individual parameters the function adjust reset was used, then a new verification is
necessary for these parameter.
The reset of the flow measurement only resets the parameters of the flow sensors 1....7. The
possibly existing calibration values of own sensors remain unchanged.
Introduction 16
System Error Message
When the HDM99 is switched on, an internal check is initiated. If the check shows no errors the
HDM99 is started using the menu which was last in use. If the check reveals an error, following
error message (window) is displayed:
Example: AD-Converter Error
An" OK" error display shows that the internal test routine has been completed with no errors
found. An "ERR" in the display shows that a problem has occurred during the check. When this
error report is displayed, please contact IBP.
#1 MEM – Memory Error
This error is displayed when an internal memory check shows that a problem has occurred. The
unit is need of repair.
#2 ADC - AD-Converter Error
This error display is shown when a defect is fount in the Analog/Digital-Converter. The unit is
need of repair.
17 Introduction
#3 ADJ – Cecksum Error
The calibration values are saved to memory together with an internal check sum.
This check sum is used each time the system is started to verify that all memory data remains
unchanged.
The aforementioned text is displayed when an error occurs during system start-up; the unit
cannot function while the error is displayed and a renewed calibration is mandatory.
#4 SYS - Watchdog Error
This message is displayed when the unit is restarted after a system crash. To delete the message,
switch the unit off and then on again. If this error occurs at regular intervals, please contact IBP
for support.
MC, ECT – Additional Information
The entry on the display following the MC and ECT are IBP error messages facilitating problem
analysis. When contacting IBP to report an error, please report the messages describing the
system error.
Measurement 18
p
Measurement
From the main menu over measuring you will receive the following menu:
Examples for measurement displays
F1
F2
F3
F4
F5
F6
F1
F2
F3
F4
F5
F6
Activation of the function keys
F5 leads to the menu for the
arameter setting
F6 leads back to the main menu
The handling of all menus
corresponds to the already
explained operating instructions.
The individual menus and submenus are presented and
explained below.
19 Measurement
Temperature measurement
Time range
By activation of the function key param. you will receive a menu which allows to change the
time range, in which the course graphic will run through horizontally once. Long term
measurements up to twelve hours are possible. Since you do not have all possible times
presented simultaneously, a black beam on the right side shows that above, or - like in the
example - further options hide themselves below.
The selected time range is shown inverted. To change the time range, operate the function keys
beside the arrows upward or downward. When you have selected the suitable time range for
your application, return to the temperature measurement with the activation of ok. The length of
the horizontal axis of the graph now corresponds to the chosen time range.
Measurement 20
Conductivity measurement
After activation of the function key param. you will get to the menu which allows you to change
the time range (horizontal axis of the graph) or the y-range (vertical axis) and also the
temperature coefficient. By activation of the function key for the range to be changed you see
the selection menus.
At the time range long term measurements up to twelve hours are possible. Since you do not
have all possible times presented simultaneously, a black beam on the right side shows that
above, or - like in the example -, further options hide themselves below.
The selected time range is shown inverted. To change the time range, operate the function keys
beside the arrows upward or downward. When you have selected the suitable time range for
your application, return to the temperature measurement with the activation of ok. The length of
the horizontal axis of the graph now corresponds to the chosen time range.
After the choice of the field y-range four fixed measuring ranges can be selected in the already
explained way with the function keys beside the arrows. If you choose > auto <, the HDM99
automatically adjusts the measuring range so that an optimal graphic representation of the
running measurement is assured.
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.