IBM M-02-212 User Manual

BROOKFIELD R/S RHEOMETER
Operating Instructions
Manual No. M/02-212
Brookfield Engineering Labs., Inc. Page 1 M/02-212
Version 1.10 E, valid for firmware version 4.02 All rights reserved, the translation included.
Any part of this Operation Manual cannot be reproduced in any form (as printed matter, photocopies, microfilms or in any other form) or processed, copied or disseminated using electronic system without written consent of Brookfield Viscometers Ltd.
Brookfield Viscometers Ltd bears no responsibility for any technical or print technical errors or incomplete information in this Operation Manual.
We reserve the right to change this Operation Manual without previous announcement.
The reproduction of users’ names, trade names, trade marks and so on in this Operation Manual does not create the necessary prerequisites for their use by anybody; frequently it is a question of protected by law registered trade marks, even if they are not marked as such.
Copyright 1998 by
Brookfield Viscometers Ltd 1 Whitehall Estate Flex Meadow Pinnacles West Harlow, Essex U.K. CM19 5TJ
MS-WINDOWS - Copyright by Microsoft
IBM, IBM-PC - Registered trade marks of International Business Machines Corporation
All rights reserved
Revision 10/9
Brookfield Engineering Labs., Inc. Page 2 M/02-212
CONTENTS
I. GENERAL DESCRIPTION 5
I.1 Use of the Rheometer 5 I.2 Measuring Principle 5
II.
SYSTEM CONFIGURATION 6
II.1 R/S Rheometer 6 II.2 Measuring Devices 8 II.3 Computer System 8
III. INSTRUMENT INSTALLATION 9
III.1 Stand Mounting 9 III.2 Electrical Connections 10 III.2.1 Temperature Sensor Pt100 10 III.2.2 AC-Adaptor 11 III.2.3 Printer Connection 11 III.2.4 Computer Connection 11 III.3 Mounting of temperature controlled measuring devices 12 III.3.1 Mounting of temp control device FTK-CC with and without cooling device KE 12 III.3.2 Mounting of the measuring device ME-CP/PP 14 III.3.3 Mounting of the cooling device KE 17
IV. ENVIRONMENT, HANDLING, CLEANING AND MAINTENANCE 18
IV.1 Operating environment, Storage 18 IV.2 Handling 18 IV.3 Cleaning 18 IV.4 Maintenance 19
V. MEASURING SYSTEMS 20
V.1 Measurement directly in the substance 20 V.2 Measurement with substance in measuring system 21 V.3 Measurement with temperature control device FTK-CC 22 V.4 Measurement with cone/plate and plate/plate measuring systems 23
VI. OPERATION AND MENUE SYSTEM
VI.1 Keyboard 26 VI.2 Menu System of R/S Rheometer 28 VI.3 Selecting from Lists 29 VI.4 Input of numerical values and alphanumeric texts 30 VI.5 Menue entries (MAIN menue) 32 VI.5.1 MAIN-Menue → Run Single 32 VI.5.2 MAIN-Menue → Run Program 34 VI.5.3 MAIN menue → Remote 36 VI.5.4 MAIN menue → Utilities 36 VI.5.5 MAIN menue → Configuration 36 VI.6 Menue entries in the Utilities-Menue 37 VI.6.1Utilities → Zero Calibration 37 VI.6.2 Utilities → Edit Programs 37 VI.6.3 Utilities → Print Programs 41 VI.6.4 Utilities → Measuring systems 41 VI.6.5 Utilities → Print memory 42 VI.6.6 Utilities → Clear memory 43 VI.6.7 Utilities → Measure temperature 43
Brookfield Engineering Labs., Inc. Page 3 M/02-212
25
VI.7 Menu entries of the CONFIGURATION menu 44 VI.7.1 Configuration → Set output mode 44 VI.7.2 Configuration → MeasCount mode 45 VI.7.3 Configuration → Reset meascount 45 VI.7.4 Configuration → Set Time/Date 45 VI.7.5 Configuration → Set RS232 Parameter 46 VI.7.6 Configuration → Language 47 VI.7.7 Configuration → Service, Service 2 47 VI.8 Serial Data transfer via Interface RS232 47
VII. MEASUREMENTS 50
VII.1 Measuring in manual mode 50 VII.2 Measuring in Remote-mode 51
VIII. TECHNICAL DATA 52
IX. GUARANTEE 54
APPENDIX
A1 Data sheets of standard measuring systems 55
A2 Error messages 58 A3 Pin-layout of the serial data cable 61 A4 Requirements to the mains connecting cables 62 A5 Language cross reference 63
Brookfield Engineering Labs., Inc. Page 4 M/02-212
I. General Description
In this chapter you get general information concerning use and measurement principle of your R/S Rheometer.
I.1 Use of the Rheometer
The R/S Rheometer system provides possibility of Newtonian fluids viscosity measurements, recording of flow curves as well as determination of viscosity functions of non-Newtonian fluids in steady shear flows. Furthermore, it assures carrying out of measurements in non-stationary shear flows and the measurements of viscoelastic substances values in creep processes. The R/S Rheometer is used in quality control, production control and researches
I.2 Measuring Principle
The R/S Rheometer is a rotational-shear rate /shear stress-rheometer. Concentric cylinders, measuring cones and plates are available as measuring system. The measuring­sample is positioned in measuring gap between the stationary measuring cup and the rotating measuring bob (Searle-principle), respectively between the rotating cone or plate and the stationary lower plate (cone/plate-, cone/cone-measuring-system). The measuring drive developed for this instrument operates with a high dynamic precision-drive­system with optical encoder without gearing and without mechanical force transducers, so torque is measured without deflection. The R/S Rheometer measuring drive is applicable for carrying out rotational measurements with pre-set of speed (shear rate) and pre-set of torque (shear stress). In case of shear rate pre-set the torque acting on rotating measuring bob/cone/plate by measuring substance flow resistance is measured (shear stress is measured). In case of shear stress pre-set resulting speed of rotating measuring bob/cone/plate is measured (shear rate). Shear stress measurements can be carried out with R/S Rheometer in order to determine flow behavior of plastic substances and enables exact measurement of yield point without shearing the measuring substance.
The types of measurements
- rotation measurement with controlled shear rate (CSR)
- rotation measurement with controlled shear stress (CSS)
can be carried out in manual (without PC support) or in connection with a computer system with the software RHEO 2000.
The creep measurements requires computer system with software RHEO 2000.
Brookfield Engineering Labs., Inc. Page 5 M/02-212
II. System Configuration
In this chapter you get information concerning system configuration, temperature control device, optional computer system and data- and signal flow of your Rheometer system.
The Rheometer system R/S Rheometer consists of:
- electronic unit with measuring drive integrated in one housing
- stand with working surface
- AC-adaptor
as accessories are supplied:
printer
coaxial cylinder measuring system (see Appendix A)
temperature measuring sensor Pt100
standard measuring device ME-CP/PP for cone/plate and plate/plate measuring systems
cone/plate- and plate/plate measuring systems
temperature control device “FTK-CC“
cooling device “KE“
fluids circulation thermostat
computer system
software RHEO 2000
bold printed accessories are necessary for minimal configuration
II.1 R/S Rheometer
The Rheometer electronics with measuring drive are built-in in one housing. The R/S Rheometer system can serve due to consequent use of microelectronics with special capacities:
digital control of rotational speed and torque automatic matching of control parameters during measurement direct indication of measured and calculated values of speed, shear rate, torque, shear stress,
viscosity, temperature, time etc.
internal measuring values storage measuring values output to a printer (parallel centronics) user support by dialog mode at LCD-display with clear text output built-in system interface with serial standard interface (RS-232-C) for the direct connection of a
computer or other serial data-loggers.
printing and serial data-transmission while measuring possible
The R/S Rheometer is operated manually using foil keyboard at the front panel or under computer control. The power supply of the R/S Rheometer with direct current is carried out by AC-adaptor.
Brookfield Engineering Labs., Inc. Page 6 M/02-212
R/S Rheometer main instrument (minimum configuration)
BROOKFIELD
- Run Single
- Run Program
- Remote
6
2
Main-Menue
Power
S T
O K
3
1
14
4
5
7
13
10
12
8
9
11
1 R/S Rheometer 2 Measuring bob coupling 3 Mounting flange 4 Pt100-clamp fixture (accessory) 5 Pt100 (accessory) 6 Standard measuring bob (accessory) 7 Standard measuring cup (accessory) 8 Measuring cup bottom, or thread protection (accessory) 9 Measuring cup screw fitting 10 Stand 11 AC-adaptor 12 Direct current coupling 13 Mains connection cable 14 Data transmission cable (accessory
Brookfield Engineering Labs., Inc. Page 7 M/02-212
II.2 Measuring Devices
Measuring devices are not included in the main delivery volume of R/S Rheometer and must be ordered in accordance with your measuring requirements.
As measuring devices are optional supplied:
coaxial standard measuring systems for R/S Rheometer (see Appendix A) with and without built-in temperature sensor Pt100
temperature control device “FTK-CC“ for use of cylinder measuring system in temperature range -10°C ... +90°C
cooling device “KE“ for use of cylinder measuring system in temperature range -20°C ... +180°C
measuring device “ME-CP/PP“ for use of cone/plate or plate/plate measuring systems (see
Appendix A) in temperature range -10°C ... +90°C
II.3 Computer System
The computer system is optional and provides automatic measuring, graphical representation of measuring results, printing of report as well as analysis of measuring data and quality control.
The computer system consists of:
IBM-compatible PC with the following minimal system requirements:
- CPU 486 DX2 66 MHz
- 8 MB RAM (main memory)
- 10 MB free hard disk drive capacity
- operating system Microsoft Windows version 3.x, Microsoft Windows 95 or Microsoft Windows 98
- mouse and keyboard
- VGA-graphic card and monitor
- 1 free serial interface RS-232
Printer
user software program package RHEO 2000
BROOKFIELD
Main-Menue
- Run Single
R/S Rheometer
- Run Program
- Remote
Power
ST
OK
Centronics
Printer when using R/S Rheometer without Computer
RS232 (Data transmission)
Printer when using R/S Rheometer
PC
Software RHEO 2000
remo te controlled
LPT1
Fig. 1.: Computer system for R/S Rheometer
Brookfield Engineering Labs., Inc. Page 8 M/02-212
III. Instrument Installation
In this chapter you will know how to prepare your R/S Rheometer for the first measurements.
You will know particularly,
- how to mount the R/S Rheometer,
- details about the electric connections
- installation of accessories such as temperature control device, cooling device, measuring system
- hose connections.
III.1 Stand Mounting
The stand consists of:
- the stand base plate
- the stand column with Rheometer
- the shelf
Hex-nut wrench size 6 and bolt DIN 912 M8x40 are supplied for mounting.
1. The stand column is joined with the stand base plate by the supplied bolt. It is necessary to check during this operation that the parallel pin at the down side of the stand column fits into the corresponding drilled hole at the base plate.
2. Lay the shelf on the stand base plate.
Brookfie ld
Main-Menue
- Run Single
- Run Program
- Remote
ST
OK
Rheometer
Pt100-temperature-sensor
Cylindric measuring system
Shelf
AC-adaptor
Fig. 2.: R/S Rheometer (minimum configuration)
Hints for height adjustment of the stand:
Locking button
In order to change the height adjustment of the stand you must release the clamp handle and move the stand to the desired height. Caution:
Hold tight the R/S Rheometer as you release the clamp handle!
You can press the locking button to change the clamp handle position
Clamp handle
without screw/unscrew the thread. After adjustment of the height screw tight the clamp handle thread again with hand
Brookfield Engineering Labs., Inc. Page 9 M/02-212
III.2 Electrical Connections
Connections for the electrical components of the R/S Rheometer are located at the instrument back side:
Socket “Centronics“ parallel Printer interface
Socket “RS232“ serial communication port
Socket “Pt100“ for temperature sensor
Socket “DC“ for AC­adaptor connection
Power
Centronics
RS232
DC
Pt 100
Mains switch “POWER“
Any cables from and to R/S Rheometer can be connected or disconnected only when instrument is switched off!
III.2.1 Temperature Sensor Pt100
The connecting cable of the temperature sensor Pt100 is inserted into connector “Pt100“ at the back side of the R/S Rheometer. The Pt100 used is dependent upon measuring system device.
If you use standard cylindric measuring cups (MB-CC45...CC8) for measurements, put the sensor Pt100 into Pt100 clamp fixture and fix it parallel to the measuring system at mounting flange of the R/S Rheometer with the help of knurled-head screw.
Fig. 3: Pt100 connection and mounting
Insert the plug of Pt100 temperature sensor into the connector “Pt100” at the back side of the Rheometer.
As the viscosity is a function of the temperature, the temperature should be measured preferly in the measuring substance. For this purpose, the standard measuring systems CC45...CC8 DIN can be equipped with a Pt100 in measuring cup bottom (only in connection with FTK-CC). If you use a measuring system with built-in Pt100, insert the cable “VK-MB“ there. With this option you can measure the substance temperature directly.
Brookfield Engineering Labs., Inc. Page 10 M/02-212
III.2.2 AC-Adaptor
The AC adaptor supplies R/S Rheometer with power.
Only the AC-adaptor delivered by BROOKFIELD with your Rheometer may be used for power supply of R/S Rheometer.
The AC-adaptor may be inserted to the socket with corresponding grounding only. Connect the AC adaptor only using the plug in a proper way grounded to avoid electric shocks or damage of the system components. Pay attention also to “Requirements to the mains cable” (see Appendix).
Connecting the AC-adaptor:
- be sure that R/S Rheometer is switched off (mains switch “POWER” at the back side of the instrument);
- insert the instrument connector of the mains cable into AC adaptor;
- insert the plug of DC cable into the connector “DC” at the back side of R/S Rheometer;
- insert the mains plug of the mains cable into a grounded socket;
- turn on the R/S Rheometer again.
The AC-adaptor should not remain connected to the mains socket during long time with disconnected plug of the DC cable from the connector “DC” at the back side.
Before disconnecting R/S Rheometer from mains supply be sure that the Rheometer is switched off.
III.2.3 Printer Connection
Printer can be connected directly to the interface connector of the R/S Rheometer when measuring without PC support. You must preset “Printer” as output device to print the measuring values during measurement (see Chapter 6). You should use obligatory a printer with parallel “CENTRONICS” interface.
- Turn off the R/S Rheometer with the mains switch “POWER” at the back side.
- Insert the printer connecting cable into the connector “CENTRONICS” at the back side of rheometer.
- Turn on R/S Rheometer again.
Standard printer cable (Printer PC) can be used as printer connecting cable. This cable is supplied with the printer in most cases.
To print data-values from R/S Rheometer all printers can be used which are printing text in ASCII-mode.
III.2.4 Computer Connection
If the R/S Rheometer has to be used in “REMOTE” mode with PC support (Rheo-program package) or with serial data-terminal for data logging it is necessary to connect the data link with the plug “RS232” at the instrument back side.
- Turn off the R/S Rheometer with the mains switch “POWER” at the back side.
- Turn off your computer system.
- Insert the data link into the plug “RS232” at the back side of the rheometer.
- Connect the other end of the data link with a free serial interface RS-232 (f.e. “COM2”) of your computer.
- Turn on the R/S Rheometer and your computer system again.
Use data link cable of BROOKFIELD delivery only!
Brookfield Engineering Labs., Inc. Page 11 M/02-212
j
Please read information concerning installation of the computer system in this Operation Manual.
III.3 Mounting of temperature controlled measuring devices
In this chapter you get information concerning mounting and attachment of the following accessories components:
FTK-CC temperature control device for use of cylinder measuring system in the
temperature range -10°C...+90°C (liquid’s temperature control)
KE cooling device in conjunction with FTK-CC temperature control device or ME-CP/PP
measuring device to rise the temperature range up to -20°C...+180°C
ME-CP/PP measuring device for use of cone/plate and plate/plate measuring system in the
temperature range -10°C...+90°C (liquid’s temperature control)
III.3.1 Mounting of temperature control device FTK-CC with and
without cooling device KE
As temperature control device is available optionally:
- FTK-CC
for use of cylinder measuring system in the temperature range -10°C...+90°C (liquid’s temperature control).
The temperature control device “FTK-CC” should be used only in the temperature range -10°C ...+90°C. The cooling device “KE” is necessary additionally for temperature range of ­20°C...+180°C!
oint for mounting flange
thread
hose connection of flow output
Pt 100
thermostating chamber
hose connection of flow input
measuring cup screw fitting
Fig.: 4.: Temperature control device FTK-CC
Mounting of the temperature control device “FTK-CC”
- Turn off the R/S Rheometer with the mains switch “POWER” at the instrument back side.
- When using cooling device KE mount cooling device first (see chapter 3.3.3)
- Set the “FTK-CC” from below on the mounting flange of R/S Rheometer and tighten the thread.
- Fix the hoses of liquid circulation thermostat at the “ FTK-CC ” as described below.
- Connect the FTK-CC´s built-in Pt100 cable with the connector “PT100” at the instrument back side of the R/S Rheometer.
Brookfield Engineering Labs., Inc. Page 12 M/02-212
BROOKFIELD
Main-Menue
- Run Single
- Run Program
- Remote
ST
OK
Cooling device KE
Temperature control device FTK-CC
Built in Pt100
Temperature control liquid from / to thermostatic device
Connecting cable Pt100
Fig. 5.: Operation with FTK-CC and cooling device KE
Thermostat connection to temperature control device FTK-CC
Hoses’ connections are necessary to join a liquid circulation thermostat by operation of the R/S Rheometer with liquid circulation temperature control device “FTK-CC”. The hoses of liquid circulation thermostat are connected with the help of quick fitting couplings with the temperature control device “FTK-CC” (below: inlet; above: outlet). For this purpose it is necessary to move back the coupling’s bush, insert the hose and release the coupling. It fixes the hose (without screwing or rotation) by groove. Check by pulling gently whether the hoses’ connections are securely fixed.
The temperature control device “FTK-CC” without cooling device KE must only be used in the temperature range -10°C ...+90°C.
The temperature control device “FTK-CC” can be used in the temperature range -20°C...+180°C in case of using optionally obtainable cooling device “KE”.
It is recommended urgently to preset at liquid circulation thermostat the upper temperature limitation: when using water at 90°C and when using oil at 180°C (only with cooling device KE).
As thermostatic liquids are usually used:
-10 °C to +90 °C water (deionized) - glycol-mixture
Brookfield Engineering Labs., Inc. Page 13 M/02-212
-20 °C to +250 °C thermostat oil
Thermostatic fluids can be ordered from BROOKFIELD.
During measurements in the temperature range below -10°C and above +90°C the temperature control device “FTK-CC” may be put into operation only if the cooling liquid flows through the “KE” in order to prevent overheat of the measuring sensor.
III.3.2 Mounting of the measuring device ME-CP/PP
- Turn off the R/S Rheometer with the mains switch “POWER” at the instrument back side.
- When using cooling device KE mount cooling device first (see chapter 3.3.3)
- Set the “ME-CP/PP” from below on the mounting flange of measuring cup and tighten the thread. Check that before tightening the guide pin of the measuring device must be in the slot of the R/S Rheometer mounting flange!
- Fix the hoses of liquid circulation thermostat at the “ME-CP/PP” (see below).
- Insert the cable “VK-MB” supplied to the “ME-CP/PP” in the built-in Pt100 and in the connector “PT 100” at the instrument back side of the R/S Rheometer.
- For mounting the cone/plate or plate/plate measuring systems to measuring device refer to chapter
5.
Thermostat connection to temperature control device FTK-CC
Hoses’ connections are necessary to join a liquid circulation thermostat by operation of the R/S Rheometer with liquid circulation measuring device “ME-CP/PP”. The hoses of liquid circulation thermostat are connected with the help of quick fitting couplings with the measuring device “ME-CP/PP” (below: inlet; above: outlet). For this purpose it is necessary to move back the coupling’s bush, insert the hose and release the coupling. It fixes the hose (without screwing or rotation) by groove. Check by pulling gently whether the hoses’ connections are securely fixed.
The measuring device “ME-CP/PP” without cooling device KE must only be used in the temperature range -10°C ...+90°C.
The measuring device “ME-CP/PP” can be used in the temperature range -20°C ... +180°C in case of using optionally obtainable cooling device “KE”.
It is recommended urgently to preset at liquid circulation thermostat the upper temperature limitation: when using water at 90°C and when using oil at 180°C (only with cooling device KE).
As thermostatic liquids are usually used:
-10 °C ... +90 °C Water (deionized) - glycol-mixture
-20 °C ... +180 °C Thermostat oil
The proper thermostatic liquids can be ordered from BROOKFIELD.
Brookfield Engineering Labs., Inc. Page 14 M/02-212
During measurements in the temperature range below -10°C and above +90°C the measuring device “ME-CP/PP” may be put into operation only if the cooling liquid flows through the “KE” in order to prevent overheat of the measuring sensor.
Brookfield Engineering Labs., Inc. Page 15 M/02-212
Measuring device for cone/plate and plate/plate measuring systems ME CC/CP
Brookfield Engineering Labs., Inc. Page 16 M/02-212
III.3.3 Mounting of the cooling device KE
The optionally supplied cooling device “KE” must be used in case of the R/S Rheometer operation with the liquid temperature control device “FTK-CC” or measuring device “ME-CP/PP“ in the temperature range below -10°C and above +90°C. When using cooling device KE temperature range of temperature control device “FTK-CC“ and measuring device “ME-CP/PP“ is expanded to -20°C... +180°C. The cooling liquid flows through the cooling channel of the cooling device KE and prevents the heat transfer from the liquid’s temperated chamber of temperature control device “FTK-CC” or measuring device “ME-CP/PP“ to the R/S Rheometer.
Mounting
- Turn off the R/S Rheometer with the mains switch “POWER” at the instrument back side.
- Lift the “KE” from below to the R/S Rheometer and tighten the thread.
- Fix the hoses of the cooling loop at the “KE” (see below “Cooling water connection to cooling device KE”).
Fig. 6: cooling device KE
Cooling water connection to cooling device KE
The hoses of cooling circulation loop are connected with by quick fitting couplings the cooling device “KE”. For this purpose it is necessary to move back the coupling’s bush, insert the hose and release the coupling. It fixes the hose (without screwing or rotation) by groove. Check by easy pull if the hoses’ connections are firm.
During measurements in the temperature range below -10°C and above +90°C the “FTK-CC” / “ME- CP/PP“ may be put into operation only if the cooling liquid flows through the “KE” in order to prevent overheat of the measuring sensor.
Brookfield Engineering Labs., Inc. Page 17 M/02-212
IV. Environment, handling, cleaning and maintenance
In this chapter you get information concerning environment, handling, cleaning and maintenance of your R/S Rheometer.
IV.1 Operating environment, Storage
Find comfortable, convenient working place where you install your R/S Rheometer. It must be enough room for installation of the stand, measuring system, measuring substances and peripheral devices (for example printer, computer and thermostat).
You need a grounded socket for operation of the R/S Rheometer. You need also a socket for connection of the peripheral devices.
Your operating environment and the place for storage of the R/S Rheometer should not be extremely hot, extremely cold or moist.
Also avoid places where temperature and air humidity fluctuate strongly. Moreover check that the R/S Rheometer is not subjected to
dirt or dust,
direct sun radiation,
objects that emit strong heat (e.g. heating radiators),
objects with strong electromagnetic field (e.g. loudspeakers, motors etc.),
liquids or corrosive chemicals.
IV.2 Handling
The R/S Rheometer is designed in such way that even by easy kicks or vibrations its’ performance is failure-free. However, pay attention to absence of strong kicks or falls.
Never lift your R/S Rheometer at the measuring bob coupling or a built-in measuring bob and avoid everything that could influence free and concentric rotation of the measuring bob coupling (e.g. a shock).
If the measuring torque is exceeded, e.g. by shut-down of the measuring drive at full load or due to viscosity raise in solidification process, the safety device in electronic operates and prevents a failure. The R/S Rheometer could be under the load unlimited time also in such extreme cases.
IV.3 Cleaning
The paint of the R/S Rheometer resists attack by usual solvents and weak acids. Use dry, clean, soft and nap-free cloth to clean the housing. Use neutral detergent liquids at a soft cloth in case of severe contamination of the housing.
Do not use any chemical products such as strong solvents or strong acids to clean the housing, especially the operating fields with the foil keyboard.
Brookfield Engineering Labs., Inc. Page 18 M/02-212
Check that any liquid does not penetrate into the housing (e.g. through the instrument connecting plugs) and into the bearings of the measuring drive. It results in the instrument damage!
IV.4 Maintenance
The R/S Rheometer is designed for long-term operation.
Nevertheless we recommend a regular maintenance in one-year cycle by a service engineer of BROOKFIELD or of corresponding representation.
Works on control electronics, all accessories, measuring drive as well as with the AC-adaptor and all electric circuits and connections may only be carried out by authorized service personnel trained by us.
Check of the measurements’ accuracy by the customer is possible any time. We recommend the measurement with the standard oils (normal oils).
Type: standard oil 2000 A, viscosity approx. 1.9 Pas
The measurement is carried out preferably with the temperature control device “FTK-CC” with measuring system CC25 DIN with Pt100 built-in in the bottom of the measuring cup. The standard oil is thermostatted for at least 20 minutes at 20°C ± 0.05°C.
Select the appropriate measuring system.
Carry out measurements at the following preset values of shear rates:
10 s-1, 25 s-1, 50 s-1 and 100 s-1.
Print viscosities or read off viscosities from display.
In case of the instrument failure (or severe deviation from the preset value), please, apply to the service department of BROOKFIELD.
Brookfield Engineering Labs., Inc. Page 19 M/02-212
V. Measuring Systems
In this chapter you get information concerning use of the measuring system.
At present the following measuring system types are supplied:
a) Standard measuring systems MS-CC48...CC8 DIN/RC and DG-DIN/RC for measurement without
temperature control device FTK-CC consists of:
- measuring cup for R/S Rheometer (MB-CC48...MB-CC8/RC and MB-DG/RC) depending on measuring-system
- measuring cup bottom
- measuring bob (MK-CC48...CC8/RC and MK-DG/RC) depending on measuring system
- thread protection
-
b) Standard measuring systems MS-CC48...CC8 DIN/FTK and DG-DIN/FTK for measurement with
temperature control device FTK-CC consists of:
- measuring cup for R/S Rheometer (MB-CC48 ...CC8 DIN/FTK u. MB-DG DIN/FTK) depending on measuring system
- bottom screw
- measuring bobs (MK-CC48...CC8 DIN and MK-DG DIN) depending on measuring system
- optionally Pt100 temperature sensor in measuring cup bottom or in temperature control device
c) One-way measuring system EWS-CC48...CC8 DIN/FTK for measurement with temperature control
device FTK-CC consists of:
- measuring cup mounting MBA-CC48...CC8 DIN/FTK depending on measuring system (holder for disposable one-way cups)
- one-way measuring cup (aluminum) EMB-CC48...CC8 DIN/FTK depending on measuring system
- measuring cup ejector
- measuring bob (MK-CC48...CC8 DIN)
d) Cone/plate measuring system C25-1...C50-2 for measurement with measuring device ME-CP/PP
consists of:
- measuring cone (C25-1...C50-2) depending on measuring system
e) Plate/plate measuring system P25, P50 for measurement with measuring device SM-KP consists of:
- measuring plate (P25, P50) depending on measuring system
Please select a suitable measuring system for the desired measuring range to carry out measurements. (for details of measuring systems ranges refer to Appendix A)
V.1 Measurement directly in the substance
If necessary remove in case of measuring systems CC48...CC8-DIN the bottom from measuring cup and screw on the thread protection.
Attention when using DG-DIN!
When using the measuring system DG-DIN remove very carefully the O-ring seal between the measuring cup case and inner part. To do this open the measuring system below and take out the inner part. Take care not to damage or stretch the O-ring seal during removal! Insert the inner part again and screw the measuring system.
Lift the coupling sleeve of the measuring bob coupling (ring visible).
Brookfield Engineering Labs., Inc. Page 20 M/02-212
Insert carefully the measuring bob into the measuring bob coupling. Pay attention to insert the measuring bob shaft into the measuring bob coupling without impact.
Move down the coupling sleeve of the measuring bob coupling (ring covered). Fasten the measuring cup at the measuring cup mounting flange with the help of measuring cup thread.
By moving down the stand, immerse the measuring cup into the substance up to the ring mark or up to the point where diameter of measuring cup increases.
Fig. 7.: Depth of immersion
Pay attention to the fact that any substance or solvent does not ingress into measuring bob coupling, measuring drive or electronics.
Now the measurement can be carried out.
After the end of measurement unscrew the measuring cup thread and remove measuring cup. Then open the measuring bob coupling and remove the measuring bob.
Clean measuring cup and measuring bob carefully. Do not use hard objects, always prevent scratches!
Store measuring bobs on a soft pad. \
V.2 Measurement with substance in measuring system
Fill the measuring cup with substance (sample filling quantity see Appendix A1 “Data sheets of the standard measuring system”). Avoid air bubbles when filling in the substance as they result in non-reproducibility of the measured values.
Place the measuring bob in measuring cup. Lift the coupling sleeve of the measuring bob coupling (ring visible). Insert the complete measuring system from below into the measuring cup mounting flange and screw tight using the measuring cup thread. Insert the measuring bob shaft into the measuring bob coupling without impact. Now insert the measuring bob shaft completely into the measuring bob coupling and move down the coupling sleeve of the measuring bob coupling (ring is covered).
In this case of measurement the substance temperature control can be carried out by immersing the measuring system into a thermostat (delivery as special accessory). It is necessary to pay attention to the immersion depth (Chapter 5.1).
Brookfield Engineering Labs., Inc. Page 21 M/02-212
Temperature range max.: 0°C ... +90°C.
The cooling device “KE” is required to expand the temperature range to
-20°C ...+180°C
The temperature sensor Pt100 may be immersed in medium only at 2/3 of the metal rod length! Cable must be always outside the medium.
If you want to measure with temperature control, the thermostat must be pre-set at the desired temperature and you should wait till the necessary temperature of the substance is reached (e.g. temperature measurements through the Pt100 menue point “Utilities Meas.-Temperature”, see Chapter 6).
Now the measurement can be carried out.
To remove the measuring system after the measurement first open the measuring bob coupling and then unscrew the measuring cup thread.
Attention: If you have carried out measurements at higher or very low temperatures please take care: Some accessible surfaces can become very hot or cold. Before removing measuring system please wait till the system-surface-temperature equalized to nearly room temperature and allows you to touch the system without danger.
When unscrewing the measuring cup, you must hold the measuring system tightly with one hand! Due to its own weight measuring system would fall downwards.
Clean measuring cup and measuring bob carefully, do not use hard objects, always prevent scratches! Store measuring bobs on a soft pad.
V.3 Measurement with temperature control device FTK-CC
In case of one-way measuring systems insert the one-way measuring cup into the measuring cup mounting. Fill the measuring cup with substance (sample filling quantity see Appendix A1 “Data sheets of the standard measuring system”).
Avoid air bubbles when filling in the substance as they result in non-reproducibility of the measured values.
Place the measuring bob in measuring cup. Lift the coupling sleeve of the measuring bob coupling (ring visible). Insert the complete measuring system from below into the temperature control device and screw tight using the measuring cup screw. Insert the measuring bob shaft into the measuring bob coupling without impact. Now insert the measuring bob shaft completely into the measuring bob coupling and move down the coupling sleeve of the measuring bob coupling (ring is covered).
The thermostat must now be pre-set at the desired temperature and you should wait till the necessary temperature of the substance is reached (e.g. temperature measurements through the Pt100 menue point “Utilities Meas.-Temperature”, see Chapter 6).
Now the measurement can be carried out.
Brookfield Engineering Labs., Inc. Page 22 M/02-212
To remove the measuring system after the measurement first open the measuring bob coupling and then unscrew the measuring cup screw of temperature control device.
Attention: If you have carried out measurements at higher or very low temperatures please take care: Some accessible surfaces can become very hot or cold. Before removing measuring system please wait till the system-surface-temperature equalized to nearly room temperature and allows you to touch the system without danger.
When unscrewing the measuring cup, you must hold the measuring system tightly with one hand! Due to its own weight measuring system would fall downwards.
In case of one-way measuring cups eject the one-way measuring cup with the measuring cup ejector.
Clean the measuring bob carefully, do not use hard objects, always prevent scratches! Store measuring bobs on a soft pad.
V.4 Measurement with cone/plate and plate/plate measuring systems
Insert and adjustment of the measuring cone or measuring plate
To install the measuring bob the bayonet lock must be released (position “opened”) and the lower measuring plate must be displaced down. Caution: during release of the bayonet lock the measuring device has to be hold tight at height adjustment ring. Lift the coupling sleeve of the measuring bob coupling (ring visible). Insert the measuring bob (cone or plate) from below into measuring coupling. Move down the coupling sleeve of the measuring bob coupling (ring covered).
Loosen the inner hexagon screw at measuring bob shaft.
Press measuring plate upward again and close bayonet lock (position “closed”).
Loosen scale clamp screw at height adjustment ring.
0-/red-position
0
Scale fixing screw
Height adjustment ring
Turn left lower
Turn right lift
The upper marker (“0”) and the lower red marker (red mark) are brought into coincidence rotating the height adjustment ring (=0-/red position, see picture).
Brookfield Engineering Labs., Inc. Page 23 M/02-212
In order to avoid thread backlash, the height adjustment ring must be now obligatory turn through to “0.4” (turn to the right = lift) and then again return back to the 0/red position.
Now screw on the inner hexagon screw at measuring bob shaft finger-tight ; this fixes the measuring bob in this position.
After rotating the scale mark from 0-/red position to the 0/0 position the cone/plate measuring device is ready for measurements. To protect lower plate against accidental turning please fasten scale fixing screw finger-tight .
Open the bayonet lock and lower the lower plate. Apply the desired measuring substance to the lower plate. Lift up carefully the lower plate again and fix the bayonet lock.
Caution: Excess of medium at plate edge must be removed (error influence!)
The thermostat must now be pre-set at the desired temperature and you should wait till the necessary temperature of the substance is reached (e.g. temperature measurements through the Pt100 menue point “Utilities Meas.-Temperature”, see Chapter 6).
Brookfield Engineering Labs., Inc. Page 24 M/02-212
VI. Operation and Menue System
In the following chapter you will get overview on operations, menue system of R/S Rheometer and measurements, both in manual and PC controlled modes.
The menue-system of R/S Rheometer (firmware-version 4.02) is a dual-language system. Following languages are software selectable:
English
German
Desired language can be selected via menue item ConfigurationLanguage (if English language is active) or KonfigurationSprache (if German language is active). More details on language selection see chapter 6.7.6.
The following chapter will describe instrument handling with ENGLISH language active.
After switching on the R/S Rheometer, LCD indicates the following information for few seconds:
* Name of the Rheometer; * Software firmware-version installed in the instrument (Brand name); * Serial number of the instrument (xxxxxx); * Date and time;
Example:
After about 5 sec the instrument checks voltages of the Power Supply Unit. Displayed voltages are to lie in the range from 14.9 to 16 V. Otherwise, some technical fault has occurred ( Repairs).
Beispiel:
This message is displayed for about 3 seconds. Then the MAIN menue is displayed. Before you will get overview of the menue system of R/S Rheometer (Chapter 6.4), at this place it is worth to make a brief summary on the rheometer keyboard and corresponding rheometer functions.
R/S Rheometer
Ver.: 4.02 #xxxxxx
07.05.98 15.12 © Brookfield
Voltage check: VCC:15.25 V
Brookfield Engineering Labs., Inc. Page 25 M/02-212
K
t
VI.1 Keyboard
START/STOP
- back to parent menue
- break
- st art of measur ement
- break of measurement
- menue: previ ous menue-ent ry
- input: increment value/ l et t er
BROOKFI ELD
MAIN- men ue
- Run Single
- Run Program
- Remote >
R/S Rheometer
- menue next menue entry
- input: decr ement val ue/letter
- one digit to the left
ST
O
OK
- enter
- select
- one digit to the righ
All user inputs are made with help of 6 keys located below LCD-Display. Some of the keys are of multiple use, i.e. their function depends on actual executed operation. The following table shows detailed overview on keyboard functions.
Brookfield Engineering Labs., Inc. Page 26 M/02-212
Key-function dependent to executed operation
key operation Function of key example
Menue goto previous menue-entry „Utilities“ → „Remote“ (above active one) Value-input increment „8 „A Selection from List list entry above active Select Meas.system entry (previous) „CC45“ → „CC48“
Menue goto next menue-entry „Remote“ „Utilities“ (below active one) Value-input decrement „5 „F Selection from List list entry below active Select Meas.system entry (next) „CC48“ → „CC45“
„9“
„B“
„4“ “ „G“
Menue goto previous menue-entry „Utilities“-> „Remote“ (above active one) Value-input one digit to the right „1
00.00“->„100.00“ „Test“->„Test“ Selection from List list entry above active Select Meas.system entry (previous) „CC45“ -> „CC48“
Menue goto next menue-entry „Remote“->Utilities“ (below active one) Input one digit to the left „1
00.00“->„100.00“ „Test“->„Test“ Selection from List list entry below active Select Meas.system entry (next) „CC45“ -> „CC48“
ST
Menue return to parent-menue „UTILITY“->„MAIN“
Value-input Input break (only if possible) Selection from List Selection break Select Meas.system (only if possible) ->back to menue Messung START and BREAK break while measuring of measurement Messung Remote active Break of measurement Back to MAIN-menue
OK
Menue execute Menue item (open sub-menue) Value-input End of input / enter value
Selection from List select active item
A novice might perceive this keyboard layout excessively complicated, however, even after your first experiments, you will surely come to the conclusion, that you do not need the layout Table for reference and multi-function keys ensure major simplification of the operations.
Brookfield Engineering Labs., Inc. Page 27 M/02-212
g
y
VI.2 Menu System of R/S Rheometer
Main-
This page demonstrates organization of R/S Rheometer menue system.
Run Single
Run
Remote
Utilities
Configuration
Measurement at constant shear rate or shear stress
Measurement with pre-defined program
Measurement under PC control (remote control)
Zero
Edit
rams
Pro
Print
Measuring-
stems
S
Print
Clear
Measure
Set
-
Set meas-
-
Reset
Set
Set RS232-
Language
Service
Zero-point calibration
Define/edit program
Print programs from memory
Define / edit of measuring systems
Print measuring data from memory
Clear measuring data in memory
Permanent measuring of temperature
Selection of output mode for measuring data (via printer /
Selection of measuring counter mode
Reset measuring counter
Setting of time and date
Parameter input for serial communication port (RS232)
Language selection of user­language
Only for service personnel!
Brookfield Engineering Labs., Inc. Page 28 M/02-212
Menu handling
Because LCD display of R/S Rheometer cannot show simultaneously all menue items, only part of the menue (three entries) is always displayed. Two arrows “>“ at the right side of display inform you, if there are still entries in the menue, either above (the arrow in the 2nd line), or below (the arrow in the 4th line).
Currently active (but still not selected) entry is marked by blinking black field (cursor) at the left part of LCD.
Example:
This entry is pre­selected and can be selected by pressing the “OK“-key
Using and keys you can move cursor upwards and downwards in the menue until desired menue item is reached. Note: If there are more menue entries in the menue and you are reaching the end of the display, next part of menue is opened automatically (scrolling).
You can “start” the pre-set menue entry or open with key.
The major menue (getting active ca. 10 seconds after switching on the instrument) is the MAIN menue.
In the case, you are in a submenue and wish to return to upper menue, press key.
Example:
Press ,you return back to MAIN-menue.
If you push key now, the “Run Single“ function is started.
Before we come to detailed description of this functions like “Run Single“ (see Chapter 6.5), it is worth to explain, how to input numbers and alphanumeric texts as well as select from a list, because majority of the Rheometer functions require such actions from the user before execution of a desired function.
ST
OK
MAIN menue
- Run Single
- Run Program
- Remote >
OK
CONFIGURATION-,emie
- Set time/date >
- Set RS232 Param.
- Language
Main-menue
- Run Single
- Run Program
- Remote >
This menue contains more items below.
ST
Menu has more entries above.
VI.3 Selecting from Lists
If the Rheometer wants you to select an entry of a list, there are the same rules and functions of LCD and keyboard as in menue operations.
Selection from a list is requested, for example at:
Selection of a measuring system for measurements in “Run Single” or “Edit Program”;
Selection of pre-set values for measurements in “Run Single” or “Edit Program”;
Brookfield Engineering Labs., Inc. Page 29 M/02-212
Selection of a program or a measuring system you want to edit in “Edit Program” and “Meas. systems”;
Answering a request “YES” “NO”.
Selection of a program to start in “Run Program”
Keys and move cursor (= pre-selected entry) in the list upwards and downwards. Key selects the pre-set entry from the list and key interrupts selection from the list (only if
OK
possible).
As well as in the menue, an arrow in the right part of LCD display shows if there are still more elements of the list above or below the marked element.
Example for selection from a list: Selection of a measuring system.
Select Measuring system:
1) CC48 >
2) DG
3) CC45 >
VI.4 Input of numerical values and alphanumeric texts
Majority of inputs, requested from user, are input of numbers. Pre-set values (e.g. start and end values of a slope), times, number of measurement points, factors, time, date, etc. are set as numbers with or without decimal digits.
If LCD display shows the decimal point “.” in a number to be entered, input of floating point number is requested. However, number of digits after decimal point is fixed to number of displayed decimal digits, the decimal point cannot be moved.
Cursor (= the digit to be changed) is here a bar under the digit.
In the following example, the set value for Single Program measurement is to be adjusted by Digit Input. We want to enter the value 290.00 s-1 for Shear Rate.
Input Values: Val.[1/s]: 0100.00 Nr. of MP: 010 Time[s]: 0100
The cursor is located under “1” in the value of shear rate to be set (currently 100.00 s
Using and keys one can change “1” (increment or decrement).
We push 1x and get:
Input Values: Val.[1/s]: 0200.00 Nr. of MP: 010 Time[s]: 0100
Digit to be entered
-1
).
Brookfield Engineering Labs., Inc. Page 30 M/02-212
Using and keys you can move cursor leftwards (rightwards). To process the next digit we push now key and get:
Input Values: Val.[1/s]: 0200.00 Nr. of MP: 010 Time[s]: 0100
Now we can process the next digit. In order to insert here “9” it’s advisable to decrement once current “0”, rather than to increment it nine times. So, we push key and get:
Input Values: Val.[1/s]: 0290.00 Nr. of MP: 010 Time[s]: 0100
Note:
In the case, you push and hold one of the keys or the cursor-underlined digit will first
increment or decrement for +/- 1 only. However, after short period of time, the digit is automatically changed. This corresponds to the Repeat Function of the computer keyboard. In our example you could also push and hold key until “9” will be displayed.
Now, when the desired number is displayed we can enter it by pressing key. In our example cursor now jumps to the first digit of the next value of user input (here Number of Measpoints).
Input Values: Val.[1/s]: 0290.00 Nr. of MP: 010 Time[s]: 0100
OK
Alphanumeric inputs
(= text with or without numeric digits) are identical to digit inputs. However, in this case, currently selected digit could not only be changed from “0” to “9”, but also to “A” to “Z” and to blank symbol “ “.
Selectable values for alphanumeric input digit in order of increase:
“ ABCDEFGHIJKLMNOPQRSTUVWXYZ1234567890”
If for example, your current digit is “B” and you want to input “7”, you have to hold key until desired “7” will appear at the display.
Alphanumeric inputs are requested for example to give a program, measuring system or measurement a name.
Brookfield Engineering Labs., Inc. Page 31 M/02-212
VI.5 Menue entries (MAIN menue)
Menue entries (see Menue System in Chapter 6.2) may either contain submenues (e.g. “Utilities” or “Configuration”) or directly executing of one of the Rheometer’s functions.
At Tree Chart in Chapter 6.2 all gray fields without further right branches serve to call functions. Those ones with right branches are submenues. Let us remind keyboard layout from Chapter 6.2. Using and keys we move cursor (black rectangle) upwards and downwards, and with key we start (activate) function (if the current
OK selection is a submenue, we open it, otherwise the function of the Rheometer starts). Functions of R/S Rheometer are described in detail elsewhere in this Chapter.
VI.5.1 MAIN-Menue → Run Single
This function carries out a measurement of shear stress or shear rate at constant preset values. The physical units of preset values may be selected from:
- shear rate D [s-1]
- speed n [rpm]
- shear stress Tau [Pa]
- torque M[‰] [1000 ‰ correspond ca. 50 mNm]
Before you start the measurement itself you have to input the following values:
Selection of the measuring system used at the instrument (selection from the list, operation ­see Chapter 6.3)
Selection of physical units of preset value (see above, selection from the list, operation, see Chapter 6.3). Here you have to decide, if you want to carry out a shear rate measurement (shear rate, speed) or a shear stress measurement (shear stress, torque).
Input of preset values (input of numbers, operation, see Chapter 6.4) D [s-1] The range of the value depends on the measuring system
n [rpm] 0.7... 800 rpm (U/min)
Tau [Pa] The range of the value depends on the measuring system
M[‰] 0 ... 999 ‰
Input of desired number of measured points (input of numbers, operation, see Chapter 6.4)
Input of the measurement duration in seconds. Here it is necessary to take into account, that minimal duration of the measurement depends on kind of the measurement and on number of desirable measurement points. For minimal duration between 2 measurement points it is valid
Shear stress measurement t
Shear rate measurement t
Note: The more is the time period between 2 measured values, the more is the precision of physical values determined!
(input of numbers, operation, see Chapter 6.4)
Input of ID (identification text of the measurement). Here you can input textual notes of the measurement (max. 15 characters). (Alphanumeric input, operation - see Chapter 6.4)
If you have passed these points, before starting a measurement, the Rheometer will indicate, where the measure points will be written to.
MP
MP
>= 1 s
> =4 s
Brookfield Engineering Labs., Inc. Page 32 M/02-212
p
Output of MPs to:
-No output device
-Memory <^> Menu <ST> Start
At this point you either can start a measurement by pressing key, or return back to Main menue
ST
using key.
In this example, we see that the measurement data will be saved in Memory only. No other output device is defined. Output devices are printer or serial interface RS232 of the Rheometer. Selection of these devices is described in Chapter ”Configuration -> Set output mode”. In case, the Memory is full with measurement data, you have to cancel function Run Single, print out the measurement data from the memory or send data to a PC via serial interface (see Chapter “Utilities Print Memory”) and then delete measurement data from the Memory (see Chapter “Utilities Delete Memory”). However, you can start a measurement even with full Memory, in this case measurement results will be shown only at LCD display and will be lost after completion of the measurement (because no output device archives measured data).
If you start the measurement, the instrument shows:
Prog. Running... Wait for 1.MP
#1
This message will be displayed until the first measurement point is calculated.
First Measure point is displayed like this:
Torque
Actual value of preset
arameter
Duration
Temperature
100.5 ‰ 000.90 1/s Eta: 1.728 Pas Tau: 1.572 Pa 04s 20.7 °C #01
Measurement result
Step (in Single meas. always 01)
If torque is less than 10 ‰
measurement results are often of low precision. In this case, preset values
are to be changed in such a manner, that the torque is higher than 10 ‰.
If the display field for the torque indicates: “M low!”, these values are outside the range of resolution of the Rheometer. If the temperature is not to displayed, the measuring sensor is not connected. At printout, the temperature is printed as 1000.0°C!
The display is updated on every new measure point. Current measurement can be canceled at any time by pushing key.
Brookfield Engineering Labs., Inc. Page 33 M/02-212
ST
A
)
After measurement, or after a break, the display field for Step indicates “END” or “BREAK”. The Display toggles in intervals of about 4 seconds between last displayed measure point and a measuring end information like this: angezeigt:
Break reason or end
Type / name of executed progr.
Program end Single mode Total time: 100s Total #MP: 10
Duration until break or end
Number of measure points measured
By pressing key you return to the MAIN-Menue.
OK
Note: The last selected program parameters remain nonvolatile in the instrument memory also after switching off the Rheometer.
VI.5.2 MAIN-Menue → Run Program
This function starts measurements, which were previously defined as a program. Contrary to “Run Single”, here you can also set values as a linear function of time, for example D[s to use function Run Program always when you want to carry out repeating measurement series with the same preset values and with the same measuring system. The second option is used when a user can carry out measurement without large previous information of rheology. Actually, not every person, starting a measurement, has unlimited possibilities to determine correct preset values and measurement system for a medium under study (optimal use of measurement range, irreversible rheological changes of the substance studied, etc.). 4 standard programs can be defined. How to define or to edit pragram is described in Chapter “Utilities Edit Program”.
Function Run Program starts one of the programs defined in memory. If there is no program available, the following error message is displayed:
No valid program! Enter program first!
In such a case, you first have to define a program (“Utilities Edit Program”). Using key you return back to the Main menue.
However, if programs are available, the list for selection of programs, stored in the memory, appears. Here you must select the program to start (selection from the list, operation see Chapter 6.3).
Select program
1) Prog xyz
2) Test
3) Prog Öl 2000 >
Press key to select the preselected program (here the program named “Test”), this operation will be
OK
followed by ID input (identification text of the measurement).
You can give your measurement a name (max. 15 characters) (alphanumerical input, operation see Chapter 6.4).
Measurement identification
Input measuring id: ID- number: 125 ID: Ö 2000 040596
utomatic counter of measurement series MEASCOUNT (automatic increment at every measurement
-1
] = f(t). It is advisable
OK
Brookfield Engineering Labs., Inc. Page 34 M/02-212
If you have passed these points, before starting a measurement, the Rheometer will indicate, where the measure points will be written to.
Output of MPs to:
-Printer
-Memory <^> Menue <ST> Start
At this point you either can start a measurement by pressing key, or return back to Main menue
ST
using key.
In this example we see, that measurement points will be saved in the memory and will be printed at the printer, so be sure that printer is connected to the Rheometer and ready for operation (=output device). Output devices are printer or serial interface RS232 of the Rheometer or no device (no output device). Their preliminary selection is described in Chapter “Configuration Set Output Mode”. In case, the Memory is full with measurement data, you have to cancel function Run Single, print out the measurement data from the memory or send data to a PC via serial interface (see Chapter “Utilities Print Memory”) and then delete measurement data from the Memory (see Chapter “Utilities Delete Memory”). However, you can start a measurement even with full Memory, in this case measurement results will be shown only at LCD display and will be lost after completion of the measurement (because no output device archives measured data).
If you start the measurement, the instrument shows:
Prog. running... Wait for 1.MP #1
This message will be displayed until the first measurement point is calculated.
First Measure point is displayed like this:
Torque
Actual value of preset parameter
Duration
Temperature
100.5 ‰ 000.90 1/s Eta: 1.728 Pas Tau: 1.572 Pa 05s 20.7°C #01
Measurment result
Step (in Run Single always 01)
If torque is less than 10 ‰
measurement results are often of low precision. In this case, preset values
are to be changed in such a manner, that the torque is higher than 10 ‰.
If the display field for the torque indicates: “M low!”, these values are outside the range of resolution of the Rheometer. If the temperature is not to displayed, the measuring sensor is not connected. At printout, the temperature is printed as 1000.0°C!
The display is updated on every new measure point. Current measurement can be canceled at any time by pushing key.
ST
Brookfield Engineering Labs., Inc. Page 35 M/02-212
After measurement, or after a break, the display field for Step indicates “END” or “BREAK”. The Display toggles in intervals of about 4 seconds between last displayed measure point and a measuring end information like this:
Break reason or end
Type / name of executed progr.
By pressing key you return to the MAIN-Menue.
OK
Program end Test Total time: 200s Total #MP: 40
Duration until break or end
Number of measuring points measured
VI.5.3 MAIN menue → Remote
Function Remote serves measurement under PC control. In this operation mode all functions of the R/S Rheometer are controlled by a PC. In order to carry out PC-controlled measurements, you need program package RHEO 2000 for R/S Rheometer. This software operates under operational systems Microsoft “Windows 3.x and Windows 95 or 98. More detailed information of RHEO 2000 software you will get from your supplier.
After selection of menue option “Remote”, the Rheometer shows following display:
Remote
Wait for RS232...
<ST>=stop
Now the Rheometer waits for communication with a PC. The data between a PC and the Rheometer is transferred via serial interface RS232 of the R/S Rheometer. If you have RHEO program package, you can start now the MEASURE/ANALYSIS program on PC.
At any time you can cancel REMOTE operation with key. But caution: current measurements are also canceled by key in REMOTE operation.
After completion of communication with the PC, LCD display shows: “Remote done ...” and you can again return back to the Main menue with key.
During measurement with PC in REMOTE operation LCD display of the Rheometer indicates some of preset values. These values, however, are not relevant for the user and serve for information in case of occurring problems.
ST
ST
ST
VI.5.4 MAIN menue → Utilities
Entry to open “Utilities” submenue - see Chapter 6.6.
VI.5.5 MAIN menue → Configuration
Entry to open “Configuration” submenue - see Chapter 6.7.
Brookfield Engineering Labs., Inc. Page 36 M/02-212
VI.6 Menue entries in the Utilities-Menue
Utilities menue, as its title presumes, contains several useful functions corresponding to measurements.
The entries:
Zero Calibration
Edit Programs
Print Programs
Measuring Systems
Print Memory
Delete Memory
Measuring Systems
Chapters below explain these functions more detailed.
Executes zero-calibration of the Rheometer input and editing of programs (which can be started with Run Program) prints all parameters of all programs in memory at the printer input of measuring system’s parameters or generation of new measuring
systems output of measurement data from memory to the printer or to serial interface clears measurement data in memory Measures temperature without rheological measurements
VI.6.1 Utilities → Zero Calibration
The function “Zero calibration” serves to calibrate the Rheometer zero point. This function continues for approximately 10 minutes and should be done once a week.
Before starting this function, ensure that R/S Rheometer was warmed-up for not less than 10 minutes and that no measuring system is connected to RC20´s measuring bob coupling.
The requirement, that a measuring system is removed, is confirmed after checking once more by
OK
OK
ST
ST
pressing key (= start of zero point calibration) or pressing key to return back to Utilities menue.
Zero point calibration is performed automatically. The progress of calibration is shown as executed step of total steps to be executed.
In the case of an error message occurs while calibrating you should start zero point calibration again. In the case, the error message is displayed again, there is a critical system fault (=>Repairs).
After a successful calibration, the message on its completion is displayed and values of the zero point are saved internally. These values are saved till the next calibration.
Push key to save values or any other key to return back to Utilities menue without saving.
Note: At any time you can cancel zero point calibration by key. Zero point values, determined up to this moment will be ignored.
VI.6.2 Utilities Edit Programs
This function is for input of new or for edition of already existing programs. The programs can be started after successful installation via the menue entry “Run Program” in the main menue.
The following values are preset in a program:
Brookfield Engineering Labs., Inc. Page 37 M/02-212
measuring system to be used in the measurement
Number of steps
start and end value of each step
Number of measurement points for each step
Duration of each step
preset value as a function of time:
D[1/s]
100
80
60
40
20
0
0 60 120 180
t[s]
D[1/s]=f(t)
As an example, we would like to consider a standard measurement.
1. Step: shear rate increases within 60 sec from 10.00 to 100.00 s
2. Step: shear rate remains at the level of 100.00 s
-1
for 60 sec.
3. Step: shear rate decreases again within 60 sec from 100.00 to 10.00 s
-1
.
-1
.
This measurement consists of 3 steps.
This example might be used to consider program input.
However, besides the shear rate D [s
-1
] = f(t), preset value can also be defined as:
Speed n[rpm]=f(t) Torque M[‰]=f(t) Shear stress Tau[Pa]=f(t)
Slopes and plain parts of the preset values are linear, measurement point is also defined linear, i.e. there is no possibility to define logarithmic measurements. For logarithmic measurement you need RHEO program package.
Preset value substeps in a defined step according to the following equation:
Preset Value = End Value - Start Value/(Number of Measuring points - 1)
where Preset Value is the difference between preset value of measurement point n and measurement point n+1.
The first preset value (=the first measurement point) is always the start value of the slope, the last measurement point is determined by the end value. As for our example, in order to reach the values D= 10, 20, 30, ... , 100 [s
-1
] we need Preset Value of 10 [s-1]. So we need 10 measurement points.
Sample: 100-10/(10-1)=10.
Number of Measurement Points = (End Value - Start Value) / Preset Value)+1
(for slopes only)
Let’s return back to the example.
After selection of the menue point “Edit Program” you will be requested to select a program from the list of all programs available, which you want to generate or to edit.
Brookfield Engineering Labs., Inc. Page 38 M/02-212
Select PROG to edit
1) Prog xyz
2) Test
3) NEW >
Available programs contain a name
All free (non-defined) programs are marked as NEW
All free memory parts are marked as NEW. In order to avoid overwriting of already available programs, please, select a NEW program to be edited.
After selection of the program with key you will be requested to select a measuring system.
OK
Attention: the program has to be executed later with the same measuring system, otherwise senseless measurement results will be calculated.
Select Meas. system:
3) CC45 >
4) CC25
5) CC14 >
After selection of the measurement system, input of number of steps requested (=number of slopes and planes of the function).
Nr. of steps : 01
Input number of steps 1...10 Our example: 03
Number of steps can be from 1 to maximum 10 steps. In our example we need 3 steps. So, number “01” must be changed to “03”. (Input of numbers - see Chapter 6.4). If you try to input numbers <1 or >10, an error message “Range-Error” will appear. After successful input of number of steps, selection of kind of measurement is carried out:
Select Input-Mode:
-D[1/s]
-n[rpm]
-Tau[Pa] >
Possible types of measurement Range of parameters preset
Shear rate D [s-1] Depending on measuring system Speed n [rpm] 0.7 ... 800 rpm Shear stress Tau [Pa] Depending on measuring system
Torque M [‰] 0 ... 999 promille
So selectent (in our example) “D[1/s]” and press key. Now, you have to define each step.
OK
Brookfield Engineering Labs., Inc. Page 39 M/02-212
For every step, there are the following inputs:
Input Step Nr 1 Start[1/s]: 0010.00 End[1/s]: 0100.00 Nr of MP: 01
0
Start Value
End Value
Number of MP (measur. points)
(input of numbers: see Chapter 6.4)
Minimum- and maximum-values of start values and end values depend on selected measuring system for shear rate (D[s-1]) and shear stress (Tau[Pa]). At input the Rheometer checks start and end values (if these values lie within due range) and indicates error massage “Range-Error” if out of range:
Range-Error
Range-Error Start-Value: min: 0.90 max: 1032.80
This message will be indicated, for example, at input of the start value D[s CC25, which lies outside the range of 0.9 ... 1032.8 s
Name of Value out of range
Allowed mimum-value
Allowed maximum-value
-1
.
-1
] with measuring system
However, in the case, start value, end value and number of measurement points were input correctly, the request for input of step duration will be indicated:
Input Step Nr. 1 time [s]: 0060
Number of currently editing step
Step duration
Minimum step duration:
D [s-1] tmin = number of measurement points * 4 s
n [rpm] tmin = number of measurement points * 4 s
Tau [Pa] tmin = number of measurement points * 1 s
M [‰] tmin = number of measurement points * 1 s
Maximum step duration allowed is: 3600 s
The duration value is automatically checked for the correct range (dependent on number of meas. points). At wrong input, the “Range-Error” message will be indicated, accompanied by allowed values.
Attention: The more time between two measuring points, the higher precision of determination of physical parameters!
This procedure of input Start-, End-Value, Number of MP, time is repeated for each step.
Finally the Program must be named and saved:
Edit PROGRAM name: TEST
Input name of program
(for input of alpha-numerical texts see Chapter 6.4)
Brookfield Engineering Labs., Inc. Page 40 M/02-212
Saving of the Program:
(2) TEST
Press <OK> to save
If you push key now, the Program will be saved. If you push any other key, the entries are abandoned and previous parameters remain (the current ones, before the edit).
OK
VI.6.3 Utilities → Print Programs
This function prints the parameters of defined programs in memory at the printer. Be sure that printer is connected to the Rheometer and ready for operation.
In the case, the printer is not ready for operation at launching “Print Programs” function, the error message will be indicated:
ERROR #1 Printer not ready!
<OK> cont. <ST> stop
In case, this error message appears, please ensure, that the printer is ready and its sheet feeder contains paper. Press key to try printing again or push key to return back to the Menue.
OK
ST
VI.6.4 Utilities → Measuring systems
With this function you can create new measuring system or change already defined measuring systems.
Measuring systems are represented by following values. This values might be edited:
Name of measuring system
Shear rate factor Kγ [min/s] (k_gamma)
Shear stress factor τ‰ [Pa] (tau_prom)
Only authorized personnel is permitted to change constants of the measuring system!
After starting the function you will be prompted to select the measuring system you want to edit. Select the list item “NEW” to generate a new measuring system.
Select Meas. system:
3) CC45 >
4) CC25
5) CC14 >
Brookfield Engineering Labs., Inc. Page 41 M/02-212
After selection of the measuring system, you will be requested to input sequentially: its name, tau_prom and k_gamma.
Enter Meas. Sys. #4 Name: CP25 tau_prom: 01.1418 k_gamma: 01.2910
The measuring system name is a alpha-numerical entry, factors tau_ prom and k_gamma are input as numerical entries (see Chapter 6.4).
When done, a security request follows, if you really want to save new parameters.
PRESS <OK> to save Name: CC25 tau_prom: 01.1418 k_gamma: 01.2990
Using key you can save the changes, otherwise, using key you can return to the Menue, without saving.
OK
ST
VI.6.5 Utilities → Print memory
This function serves output of measuring values from internal memory to the printer, connected to the Rheometer, or to the serial interface RS232 of the Rheometer. First, you will be requested to select the output device, where you want to transfer the measuring values to.
Sel. Output-device:
- Printer
- RS232
a) Output to printer: The printer must be connected to the Rheometer and be ready for operation.
b) Output to the serial interface RS232:
Data receiving device (mostly a PC) must be adjusted to same data transfer parameters as R/S Rheometer and be ready for receiving data, because the data transfer of the Rheometer to the receiving device is carried out without handshaking. If the receiving side is not ready or is not adequately tuned, the data will be transmitted “into emptiness”. (See also Chapter 6.8 “Serial data transfer”).
Output to printer
Output to RS232 serial interface
Brookfield Engineering Labs., Inc. Page 42 M/02-212
If you have chosen the output device, you have to select the measuring series to transfer. All available data is displayed in a list.
Select Program:
- 1;17.10;11:20
- 2;17.10;12:30
- 3;17.10;13:40
After selection of a measurement series, the data of the series will be printed as Table or transferred via the serial interface. After completion of data transfer the Rheometer returns to Utilities Menue.
Meas. series counter MEASCOUNT
Date of measurement
Time of measurement
VI.6.6 Utilities → Clear memory
This function deletes all measured data stored in the memory of the R/S-CPS Rheometer.
Before deleting, please ensure that the measurement data have been archived on printer or PC. (see Print Memory, the last chapter). Before the final deleting from the memory the following security request is indicated:
Clear Memory?
- NO
- YES
If you select “YES” entry, the memory of measuring results will be erased, otherwise you will return back to the Menue without deleting.
VI.6.7 Utilities → Measure temperature
This function serves measurement of temperature with the temperature sensor, connected to the Rheometer.
After starting this function, permanent temperature measurement is performed:
Temperature:
20.5 °C
<OK> return
Pressing key you return back to Utilities-Menue .
Brookfield Engineering Labs., Inc. Page 43 M/02-212
OK
A
A
VI.7 Menu entries of the CONFIGURATION menu
The Entries:
Set output-mode sets output-mode of measuring data at the time of measurement into the
printer or into the serial interface.
Set meascount-mode sets the counter mode (if the MEASCOUNT counter of measuring series
will be daily cleared or not).
Reset meascount reduces the MEASCOUNT counter of measuring series by 1.
Set time/date Input of date and time
Set RS232 Param. Sets data transfer parameters for the serial interface RS232
Language Selects user language of Rheometer (English/German)
Service, Service2 for service personnel only
In the following chapters these functions are explained in more details.
VI.7.1 Configuration → Set output mode
This function serves definition of output device (printer, serial interface) which will receive measurement results while measuring. These settings do not depend on measuring values, being saved in the inner memory of measurement data, i.e. when you set the output device, measuring points are saved nevertheless in the inner memory of measurement results (if there is available memory). Selection of an output device is the additional output of measurement values to the printer or the serial interface RS232 at the time of a measurement.
At starting this function you may choose between possible output devices:
Select output device
- No output device
- Printer
- RS232
Internal memory only
dditional output to printer
dditional output to
serial interface
If you have selected an output device, your selection is saved in the instrument by pressing key.
OK
This selection remains stored even after switching off the instrument!
Currently set output device is indicated before start of every measurement (see also “Main Run Single” and “Main Run Program”).
Output of MPs to
- Printer
- Memory <ST>Start UP>Menue
Selected output device
After selecting an output device you have to make sure before the start of every measurement that:
a) for output to a printer, the printer is connected to the Rheometer and is ready for operation. If the
printer is not ready, you cannot start the measurement.
Brookfield Engineering Labs., Inc. Page 44 M/02-212
b) for output to the serial interface (RS232), the receiving device (normally a PC) is set to the data
transfer parameters of the Rheometer (see chapter 6.7.5 “RS232 Parameters”) and ready to receive data. If the receiving side is not ready, the data will not be transmitted or will be transmitted to Null. (See also chapter 6.8 “Serial data transfer”).
VI.7.2 Configuration → MeasCount mode
Measurement series counter MEASCOUNT is a counter, increasing its count by 1 when starting a measurement. Measurement series counter serves, as well as ID of the measurement for identification of a measurement series. Counters’ count rises until automatic of manual reset (“Reset meascount”) will be performed.
Using the function “Set meascount-mode” you set the counter mode (if the measurement series counter will be daily reset or not).
Select meascount mode:
- No reset
- Reset every day
After choosing one of two reset modes you again come to the Configuration-Menue.
Reset only via function „Reset meascount“
Automatic resetting every new day
VI.7.3 Configuration → Reset meascount
This function resets the measurement series counter (regardless a MEASCOUNT mode set) = manual reset of the measurement series counter.
No further selections or settings are required.
VI.7.4 Configuration → Set Time/Date
This function allows the user to set the internal clock and internal calendar.
Setting of time is carried out in 24 hour format.
Set Time Hour: 17 Minutes: 15
Setting of date is carried out as dd.mm.yy. :
Set Date Day: 17 Month: 03 Year: 01
Hours 00 to 23
Minutes 00 to 59
Day 01 to 31
Month 01 to 12
Year 00 to 99
Brookfield Engineering Labs., Inc. Page 45 M/02-212
After setting time and date you will be requested if you want to set clock for new time.
New Time/Date 17:15 17.03.01
Press <OK> to save
Using key, you can set new time, or using key you can return back to the menue without
OK
ST
saving.
Note: New time is input into the clock only after pressing key. When pressing the key, time starts
OK
running with seconds=00.
VI.7.5 Configuration → Set RS232 Parameter
This function allows changing communication parameters of the serial interface. The serial interface operates without handshake. Data are transferred as ASCII-texts.
Parameters to set: Baud rate [Baud]: 110, 150, 300, 600, 1200, 2400, 4800, 9600, 19200
Parity: n(=no parity), e(=even), o(=odd)
Stopbits [Bit]: 1 or 2
Databits [Bit]: 7 or 8
The standard setting is:
Baud rate = 4800 Parity = n Stopbits = 1 Databits = 8
Caution: If you changed these settings, you have also to set new transfer parameters of the receiving device!
After starting the function “Set RS232 Param.” you will be requested to input the baudrate:
OK
With key you start change of parameters.
RS232: 4800,8,n,1
Change RS232-Par.? <OK>YES <ST>NO
You have to select following parameters:
Baudrate [Baud] : 10, 150, 300, 600, 1200, 2400, 4800, 9600, 19200
Datenbits [Bit]: or 8
Parity: (=none), e(=even), o (=odd)
Brookfield Engineering Labs., Inc. Page 46 M/02-212
Stopbits [Bit]: 1 or 2
Choose the requested parameter from the list of available values and select it by pressing key.
Example Baudrate:
Baudrate:
-4800 _9600
-19200
After selecting all parameters, new setting of communication parameters is displayed:
RS232: 19200,8,n,1
<OK>=storing
with key you transfer this parameters to nonvolatile memory of R/S Rheometer.
OK
These changed transfer settings will be saved into memory and will remain in the memory also after switching off the instrument.
OK
VI.7.6 Configuration → Language
This function selects the user-language of R/S Rheometer. Languages available are:
English
German
Choose desired language and press .
OK
VI.7.7 Configuration Service, Service 2
This function is only for service-personnel. Password protected!
VI.8 Serial Data transfer via Interface RS232
Serial data transfer has to be used only in the case, when a user has basics of knowledge on data processing and is capable to change data transfer parameters of sending and receiving sides, by him/herself.
Successful data transfer can be achieved only in the case, when a transmitter (=R/S Rheometer) and the receiver (e.g. a PC) comply to the following:
1. Instruments are connected with supplied interface cable. Caution: While making the connection, both instruments must be switched off! (Pin layout: see Appendix to this documentation).
2. The transmitter (R/S Rheometer) and the receiver are configurated to identical data transfer parameters (for R/S Rheometer see Chapter 6.7.5. “Set RS232 Parameter”).
Brookfield Engineering Labs., Inc. Page 47 M/02-212
3. The receiver is ready for reception and can (due to sufficient PC and memory capacity) receive or save received data at selected baudrate.
It is worth to demonstrate reception with the help of Terminal program under Microsoft Windows.
1. Switch off the Rheometer and PC.
2. Connect the Rheometer (connector RS232 at the instrument back side) and one of free serial interface (e.g. COM2) of the PC. Here use only the data link (supplied by BROOKFIELD as an accessory).
3. Switch on the PC and the Rheometer.
4. Set data transfer parameters at the Rheometer (see Chapter 6.7.5). Here: Standard setting 4800 Baud, Parity n, Stopbit 1, Databit 8
5. Select RS232 interface as the output device at the Rheometer (see Chapter 6.7.1).
6. Start WINDOWS operating system at your PC.
7. Open “Accessories” program group.
8. Start “Terminal” program.
9. Choose menue option “Settings” “Data transfer”
First the window appears, where you can input data transfer parameters:
- select under “Connection” the COM-port, connected to the Rheometer.
- set “Baudrate” to “4800”
- set “Databits” to “8”
- set “Stopbits” to “1”
- set “Parity” to “no parity”
- set “Protocol” to “no”
- deactivate “Parity check” checkbox (if crossed)
- deactivate “Carrier signal detection” checkbox (if crossed)
- finish input by “OK”
10. Select the Menue option “Settings” “Terminal Settings”
The window appears, where the terminal functions are set:
- scroll and select “Germany” option
- activate “IBM in ANSI” (if not crossed yet)
11. Now you can save parameters of Terminal program, changed up to this time, under the Menue option “File” “Save” and set the file name, e.g. “RC20.TRM”.
(At the next start of the Terminal program, you can open this file with “File” “Open” and load data transfer parameters for the R/S Rheometer.)
12. Now select the Menue option “Transfer” “Text file reception”.
- input the file name, under which you want to save the data sent from the Rheometer (e.g. “TEST.TXT”).
- In the status line of the Terminal program, the message appears now “receiving: TEST.TXT”
13. The Terminal program is now ready to receive data (for more detailed instructions or troubleshooting on the Terminal program - please read your Windows-documentation).
Brookfield Engineering Labs., Inc. Page 48 M/02-212
14. Start measurement at the Rheometer.
Measurement points must now be seen as text at the PC display after a short period of time.
15. After completed data transfer of one or several measurements into the selected text file, break the data transfer and end the Terminal Program.
If it is necessary to select some other receiver for serial data transfer (not a PC), pin layout of this receiver must be checked before the connection. You can find pin layout of the serial interface plug and pin layout of the data link in the Appendix to this documentation.
Brookfield Engineering Labs., Inc. Page 49 M/02-212
VII. Measurements
VII.1 Measuring in manual mode
To carry out measurements in manual mode you can act in correspondence with the following brief instructions:
Measurement control:
Fasten the R/S Rheometer at stand (see Chapter 3).
Connect the AC adaptor (see Chapter 3).
Connect, if necessary, the printer (see Chapter 3).
Fill and mount the standard measuring system (see Chapter 5).
Point to the fact that any substance or solvent do not ingress into measuring bob coupling, measuring drive or electronics.
Wait for temperature control till the attainment of the desired temperature in the measuring
substance.
Start a program or a single measurement (see Chapter 6).
After ending of the measurement and switching off of the temperature control wait till the
cooling/warming of the medium to room temperature.
Remove and clean the measuring system.
Measurement with “FTK-CC”
Fasten the R/S Rheometer at stand (see Chapter 3).
Connect the AC adaptor (see Chapter 3).
Connect, if necessary, the printer (see Chapter 3).
Mount the temperature control device “FTK-CC” and connect, if desired, the built-in Pt100 (see
Chapter 3).
Fill and mount the standard measuring system (see Chapter 5).
Point to the fact that any substance or solvent do not ingress into measuring bob coupling, measuring drive or electronics.
Check the presence of temperature control medium and cooling liquid (if cooling device is
used).
Wait for temperature control till the attainment of the desired temperature in the measuring
substance.
Start a program or a single measurement (see Chapter 6).
After ending of the measurement and switching off of the temperature control wait till the
cooling/warming of the medium to room temperature.
Remove and clean the measuring system.
Measurement with cone/plate measuring device “ME-CP/PP”
Fasten the R/S Rheometer at stand (see Chapter 3).
Connect the AC adaptor (see Chapter 3).
Connect, if necessary, the printer (see Chapter 3).
Mount the cone/plate measuring device “ME-CP/PP” and connect the built-in Pt100 (see
Chapter 3).
Mount, adjust and fill the standard measuring system (see Chapter 5).
Brookfield Engineering Labs., Inc. Page 50 M/02-212
Point to the fact that any substance or solvent do not ingress into measuring bob coupling, measuring drive or electronics.
Check the presence of the temperature control liquid.
Wait for temperature control till the attainment of the desired temperature in the measuring
substance.
Start a program or a single measurement (see Chapter 6).
After ending of the measurement and switching off of the temperature control wait till the
cooling/warming of the medium to room temperature.
Remove and clean the measuring system.
VII.2 Measuring in Remote-mode
Measurement in remote control mode:
Fasten the R/S Rheometer at stand (see Chapter 3).
Connect the AC-adaptor (see Chapter 3).
Connect the data link cable between Rheometer and PC.
Switch on the R/S Rheometer and select the menu point “Remote”.
Switch on computer and all peripheral devices.
Start the RHEO 2000 Software.
Load a program in the RHEO 2000 Software
Fill and mount your measuring system standard measuring system (see Chapter 5).
Point to the fact that any substance or solvent do not ingress into measuring bob coupling, measuring drive or electronics.
Check the presence of the temperature control medium and cooling liquid (if cooling device is
used).
Wait for temperature control till the attainment of the desired temperature in the measuring
substance.
Start the program.
After ending of the measurement and switching off of the temperature control wait till the
cooling/warming of the medium to room temperature.
Remove and clean the measuring system.
Brookfield Engineering Labs., Inc. Page 51 M/02-212
VIII. Technical data
R/S Rheometer
Dimensions Weight Nominal operating voltage
Power consumption (average) Power consumption (maximum)
Ambience conditions Temperature
in operation out of operation
Relative humidity (not condensable) in operation out of operation
Accuracy
Torque range Mains operation Torque resolution
Speed range Angle resolution
Temperature range
Shear rate range
Shear stress range
Viscosity range
The given range is a standard value (not maximum value)
AC-Adaptor
Dimensions Weight Power supply
Mains voltage secondary voltages (Output) output-current output power
Frequency range of mains voltage Environment conditions
Temperature in operation out of operation Relative humidity (not condensable) in operation out of operation
480 mm x 300 mm x 290 mm
8 kg
+/- 15 V, 5 V 12 W 22 W
10° to 40°C 10° to 45°C
20% to 80% 10% to 90%
± 1.0 % of maximum range value ± 1 digit
0.05 to 50 mNm
0.01 mNm
0.7 min
-1
to 800 min-1
0.8 mrad
Depending on temperature control device used
-20°C to +180°C
Depending on measuring system used
0.9 s-1 to 4 × 103 s-1
Depending on measuring system used
0.7 Pa to 3.4 × 104 Pa
Depending on measuring system used
1 × 10-3 Pas to 3 × 103 Pas
160 mm x 85 mm x 35 mm
0.5 kg
from 100 up to 240 V AC 5 V, +/- 15 V DC 2 A, 0.9 / -0.2 A 20 W
50 bis 60 Hz
+10°C to +40°C +10°C to +45°C
20% to 80% 10% to 90%
Brookfield Engineering Labs., Inc. Page 52 M/02-212
FTK-CC
Dimensions (Width x Height x Depth) Weight Temperature range
standard range with cooling device
94 mm x 55 mm x 170 mm
600 g
-10°C to +90°C
-20°C to +180°C
Measured or evaluated values The preset values and measured values are listed below as well as all evaluated values:
Value Symbol Physical Unit
Speed n [min-1]
Torque (relative)
(1000 ‰ =^ 50 mNm)
M [1]
Temperature T [°C]
Time t [s]
Shear rate
Shear stress
Dynamic viscosity
γ⋅
τ
η
[s-1]
[Pa]
[Pas]
Brookfield Engineering Labs., Inc. Page 53 M/02-212
IX. Guarantee
Brookfield Engineering Laboratories guarantees the faultless functioning of this instrument insofar as it is Brookfield Viscometers Ltd guarantees the faultless functioning of this instrument as far as it is used, maintained, connected and handled in accordance with this Operation Manual.
The guarantee period is 1 year from the date of delivery. Place of the guarantee fulfillment is Brookfield Viscometers Ltd, Harlow, U.K.
All claims of the Customer concerning guarantee and damages forfeited if he has improperly handled, worked on, opened or changed the delivery goods or gave it to a third party without our prior approval.
The total liability of Brookfield Viscometers Ltd and your exclusive claim is on Brookfield’s choice
either
(a) the refund of the price paid
or
in repair or replacement of the instrument, when the instrument does not meet the requirements stated in the limited guarantee of Brookfield Viscometers Ltd. This instrument should then be returned with a copy of receipt-formular to Brookfield Viscometers Ltd, Harlow, U.K.
This limited guarantee is not valid if instrument’s failure is result of accident, misuse or incorrect application. For a replacement delivery Brookfield gives guarantee only for the rest of the original guarantee period or for 30 days, the more long time period is decisive.
Brookfield Viscometers Ltd suspends any further guarantee concerning the instrument and related manuals and written materials.
Neither Brookfield Viscometers Ltd nor the Suppliers of Brookfield are responsible for any damages (lost profit, production interruption, losses of trade information or data or any other financial losses) that arise due to usage of this Brookfield product or from the inability of use of this Brookfield product, even when Brookfield Viscometers Ltd was informed of possibility of such damages.
In any case the liability of Brookfield Viscometers Ltd is limited to the amount which was paid for this product. This exception is not valid for damages which were caused due to intention or negligence from the side of Brookfield. Also the claims remain valid that are based on the laws concerning products quality responsibility.
Furthermore the guarantee conditions of “General Delivery Conditions of supply and delivery for products and services of the Electric Industry” are valid.
Brookfield Engineering Labs., Inc. Page 54 M/02-212
APPENDIX
A1 Data sheets of standard measuring systems
Table Standard cylinder measuring systems according to DIN 53019 / ISO 3219 (consists of measuring bob and measuring cup) For R/S Rheometer three types of measuring bobs are available:
MB-CC48 ... CC8 DIN/RC
MB-CC48 ... CC8 DIN/FTK
MBA-CC48...CC8 DIN/FTK with EMB-CC48...CC8
Measuring system CC48 CC45 CC25 CC14 CC8
Shear rate range
-1
]
[s
Shear stress range [Pa]
Viscosity range [Pas]
Filling volume [ml]
Shear rate factor K [min/s]
γ
0 ... 4,114
0 ... 206
0.005 ... 32
0 ... 1,032
0 ... 195
0.020 ... 15
0 ... 1,032
0 ... 1141
0.118 ... 100
0 ... 1,032
0 ... 6501
0.672 ... 500
0 ... 1,032
0 ... 34,844
3.60 ... 3,000
70 100 17 3 0.5
5.142 1.291 1.291 1.291 1.291
α
Measuring geometrie DIN 53019
Shear stress factor τ [Pa]
Radius of measuring bob Ri [mm]
Radius of measuring cup Ra [mm]
Radius of shaft Rs [mm]
0.1900 0.1958 1.1418 6.501 34.844
23.9 22.5 12.5 7 4
24.4 24.4 13.56 7.59 4.34
3.5 3.5 3.5 2.1 1.2
Angle of measuring bob cone α
120 120 120 120 120
[°]
Distance between lower edge of meas. bob and
35 35 15.5 13 12
meas. cup bottom L' [mm]
Immersion of measuring shaft L'' [mm]
Length of meas. bob L [mm]
Ratio of Radii δ = Error!
Resistance coefficient cL
22.5 22.5 12.5 7 4
67.5 67.5 37.5 21 12
1.0209 1.0844 1.0848 1.0843 1.0850
1.1 1.1 1.1
1.1
The given ranges are standard values (not maximum values).
1.1
Brookfield Engineering Labs., Inc. Page 55 M/02-212
Table Double Gap cylinder measuring system according to DIN 53453 (consists of measuring bob and measuring cup)
Measuring system DG DIN
Shear rate range[s-1]
0 ... 4,031
Shear stress range [Pa] 0 ... 67
Viscosity range [Pas] 0.001 ... 1.30
Filling volume [ml] 22.5
R
4
R
3
Shear rate factor K
Shear stress factor τ
Inner radius of measuring bob R2 [mm]
Outer radius of measuring bob R
Inner radius of measuring cup R1 [mm]
Outer radius of measuring cup R4 [mm]
[min/s]
γ
[Pa]
[mm]
3
5.039
0.067
22.75
23.5
22.25
24
Length of measuring bob L [mm] 111
Ratio of radii δ = Error! = Error!
Resistance coefficient cL
1.021
1
The given ranges are standard values (not maximum values).
R
2
R
1
Measuring geometry
DIN 53453
L
Brookfield Engineering Labs., Inc. Page 56 M/02-212
Table Cone/Plate measuring systems (DIN 53018) (consists of measuring cone and measuring device ME-CP/PP)
Measuring system C25-1 C25-2 C50-1 C50-2 C75-1 C75-2
Shear rate range
-1
]
[s
Shear stress range
0 ... 4,800 0 ... 2,400 0 ... 4,800 0 ... 2,400 0 ... 4,800 0 ... 2,400
0 ... 12,223 0 ... 12,223 0 ... 1,528 0 ... 1,528 0 ... 452 0 ... 452
[Pa]
Viscosity range
0.25 ... 1,629 0.5 ... 3,259 0.03 ... 203 0.06 ... 407 0.009 ... 60 0.02 ... 120
[Pas]
Filling volume [ml] 0.08 0.3 0.7 1.5 2.0 3.9
Shear rate factor K
6.00 3.00 6.00 3.00 6.00 3.00
γ
[min/s]
Shear stress factor
τ
[Pa]
Radius of measu-
12.223 12.223 1.528 1.528 0.4527 0.4527
12.5 12.5 25 25 37.5 37.5
ring cone R [mm]
Angle of measu-ring
1 2 1 2 1 2
cone α [°]
Cone truncation
50 50 50 50 50 50
[µm]
Table Plate/Plate measuring systems (DIN 53018)
(consists of measuring plate and measuring device ME-CP/PP)
The following data is valid for a measuring gap of 1 mm. For other measuring gaps please calculate
K
γ
with formula:
K
γ
=(2π/60) * (R/H) R=
radius
, H=gap
Measuring system P25 P50 P75
Shear rate range [s-1]
0 ... 1,047 0 ... 2,094 0 ... 3,141
Shear stress range [Pa] 0 ... 16,298 0 ... 2,037 0 ... 603
Viscosity range [Pas] 1.56 ... 9,960 0.097 ... 623 0.02 ... 122
Filling volume [ml] 0.5 2.0 4.5
Shear rate factor K
Shear stress factor τ
[min/s]
γ
[Pa]
1.309
16.2975
2.6175 3.927
2.037 0.6036
Radius of measuring plate R [mm] 12.5 25 37.5
Brookfield Engineering Labs., Inc. Page 57 M/02-212
A2 Error messages
The R/S Rheometer is rather error resistive, when errors occur they are determined and the user is informed on the LCD display. The most frequent error massages are explained in this Chapter.
Range Error
Cause: User’s error at input. The user has tried to input a value that is less than the allowed
minimum value or is more than the allowed maximum value.
Message:
Range error End value: min : 0.90 max : 1032.80
What to do: Input the value again. Be sure that new value is inside of the allowed value range.
Example of error input of a preset value
Allowed minimum value
Allowed maximum value
Name of value which is out of range
Printer Error
Cause: The printer is selected as output device and is not ready for printing.
a) Printer cable is not connected to the Rheometer b) Paper out, printer not on line, other printer errors
Message:
Error #1 Printer not ready!
<OK> cont. <ST> stop
What to do: Check if the printer cable is connected to the Rheometer.
Example: printer not ready
Try again
Cancel action
Check if paper is in the feed sheeder of the printer. Check if the “Ready” LED lights on printer. Remove the error cause at printer and push key at Rheometer. If you cannot find the error cause, push key in order to break the action that is carrying out now.
ST
OK
Brookfield Engineering Labs., Inc. Page 58 M/02-212
Zero Calibration Error
Cause: An impermissible high value was measured during carrying out of zero point calibration of
the Rheometer
Message:
What to do:
Error #3 Zero cal. error! Please retry cal. <OK> cont. <ST> stop
Push key. Check if the measuring system has been really removed from the Rheometer. Retry zero point calibration. If the error is presented again, that means that there is a fault in the measuring instrument
ST
BREAK: MOMMAX
Cause: Maximum torque has been exceeded during carrying out of a measurement.
a) You have selected inappropriate measuring range load for the measuring medium. b) Rotation bob distorted in the measuring cup.
Message:
BREAK: torque max! Single mode total MP: 10 total time: 60
What to do:
Brookfield Engineering Labs., Inc. Page 59 M/02-212
OK
Push key. Check if the measuring system is fixed at the Rheometer properly. If so, select for this measurement smaller shear rates or speed or use a measuring system with higher shear stress factor.
BREAK: SPEED MAX
Cause: Maximum speed exceeded during carrying out a measurement.
a) You have selected too high shear stress or too high torque value for the measuring substance. b) You have started torque- or shear stress measurement without substance in measuring system.
Message:
BREAK: Speed max! Single mode total MP: 10 total time: 60
What to do:
OK
Push key. Select smaller shear stress or torque values for this measurement.
Brookfield Engineering Labs., Inc. Page 60 M/02-212
A3 Pin-layout of the serial data cable
The Rheometer is provided with a serial interface with 25 pin Sub-D-connector (male) at the back side of the instrument. The serial interface is marked as RS232.
The signal level in accordance with RS232 is in the range between +12 V and -12 V.
Pin-connection of the serial data cable to the serial data transfer to a PC:
Rheometer side: 25-pin Sub-D-connector (female) View on soldered joints All other pin must not be connected!
PC Side 9-pin Sub-D-connector (female)
13
25
5
7
GND
9
3
RXD 2 TXD
6
1
1
14
Brookfield Engineering Labs., Inc. Page 61 M/02-212
A4 Requirements to the mains connecting cables
The AC-adaptor unit of the R/S Rheometer enables operation of the Rheometer with supply voltage in the range of 100 to 240 V AC with frequencies from 50 to 60 Hz.
The mains connecting cable that you have received with your R/S Rheometer in some cases may not meet with the requirements of the country in which you are using this Rheometer. You must obligatory use a mains connecting cable that meets the specific requirements of the relevant country.
The following information explains the requirements that are to taken into account when choosing the mains connecting cable.
General information
The connecting cable must be permitted (authorized) in the country where it is used.
The AC-adaptor plug connector (that is the connection with the AC adaptor itself, not the
mains plug) must suitable for the input socket type CEE22/IEC 320 (data sheet C14).
The mains cable must be at least 2 m and at the most 3 m long.
USA and Canada
- The mains connecting cable must correspond to the UL-permission and CSA Certificate.
- The following minimal requirements are valid for the cable:
No. 18 AWG
Type SV or similar
3-phase
- The cable must have a nominal load-capacity of at least 10 A.
- The mains plug of type NEMA 5-15P (15A, 125V) or NEMA 6-15P (15A, 250V) must have plug protection.
Other countries
- The connectors of the mains connecting cable must have approval certificate for the corresponding country in which this Rheometer is used. These places are:
- BSI (Great Britain) NEMKO (Norway)
- CEBEC (Belgium) OVE (Austria)
- DEMKO (Denmark) SEMKO (Sweden)
- EANSW (Australia) SETI (Finland)
- IMQ (Italy) SEV (Switzerland)
- KEMA (Netherlands) UTE (France)
- VDE (Germany)
- The cable must be three-phase HAR-cable of type HO5VV-F3 with a minimal phase diameter­section of 1.0 mm The mains connecting cable must be permitted for a current load of at least 10 A and, depending on country, a nominal voltage of 125 V or 250 V AC.
Japan
In Japan the following cable types and connections values are required:
- All parts of the cables (cable, socket and plug) must have brand name and registration number according to the Dentori law.
- The following minimal requirements are valid for the cable:
0.75 mm
Type VCT or VCTF
3-phase
- The cable must have nominal current load capacity of 7 A.
2
.
2
, 2-phase
Brookfield Engineering Labs., Inc. Page 62 M/02-212
- The mains plug must be 2-pin shockproof plug according to the Japanese Industrial
Standard C8303 (15A,125V).
A5 Language cross reference
English German English German
!M low! !M -> 0! BREAK ABBR. Break:speed max! Abbruch Drehz. max! Break:torque max! Abbruch Mom. max.! Change RS232 Param.? Aendern RS232-Par.? Clear memory? Speicher loeschen? continue weiter Edit Meas.-Sys. Eing. Mess-Sys. Edit programm first! Zuerst Eingabe Prog! Edit PROGRAM-Name Eing. Programmname End-Value: Endwert: ERROR FEHLER Input measuring-id Eingabe Mess-ID Input Step Nr. Eingabe Schritt# Input Values: Eingabe Werte: Meascount Messzaehler Meascount-mode Messzaehler-Modus Measuring-systems Messsysteme Memory cleared! Speicher geloescht! Memory Speicher No free memory! Kein Speicher frei! no output-device kein Ausgabegeraet no reset nicht ruecksetzen no valid data! keine guelt. Daten! No valid programm! Kein guelt. Progr.! NO NEIN not valid nicht gueltig Nr. of steps Anzahl Schritte Output of MPs to: Ausgabe Messwerte: Please do DAC-CAL! DAC-Kal. notwendig! Please do ZERO-CAL! ZERO-Kal. notwendig! please wait... bitte warten... Press any key... Taste druecken... Printer not ready! Drucker n. bereit! Printer Drucker Printer-Error! Drucker-Fehler!
Program end Programmende Program running... Programm laeuft... Range-Error Eingabe-Fehler Remove Meas.-System! Messystem entfernen! reset every day taeglich ruecks. Reset Meascount? Messzaehler=0? return zurueck rpm U/min RS232 RS232 run without mem. Start o.Speicher save Speichern Sel. output-device: Wahl Ausgabe-Geraet: Select Input-Mode: Auswahl Einheit: Select Meas.system: Auswahl Messystem: Select PROG to edit Welches PROG bearb.? Select PROG to run Auswahl Programm Select Program: Auswahl Programm: Single mode Einzelmessung Start-Value: Startwert: stop Stop Temperature Temperatur t i m e Z e i t Time/Date Uhrzeit/Datum Todays date Heutiges Datum: total #MP: Gesamt #MP: total time: Gesamtzeit: User-Break! Abbruch Benutzer! User-defined Programs Benutzerdef.Programme Voltage-Control Spannungsversorgung Wait for 1.MP Warte auf 1.MP Wait for RS232... Warte auf RS232... YES JA Zero-Cal. done... Abgleich fertig... Zero-Cal.running... Abgleich laeuft... Zero-calibration Nullpkt.abgleich
Brookfield Engineering Labs., Inc. Page 63 M/02-212
German menue system of RC20
HAUPT­Menue
Einzel-
messung
Programm-
Messung
Remote
Hilfsmittel
Measurement with constant shear rate or shear stress
Measurement with pre-defined programm
Measurement under PC-control (remote control)
Nullpkt.-
Abgleich
Programme
bearb.
Programme
drucken
Messyteme
Speicher drucken
Zero-point calibration
Define/edit programs
Print program-parameters to printer
Define/edit of measuring-systems
Print measuring data from memory
Konfiguration
Speicher loeschen
Temperatur-
messung
Ausgabe
Modus
Messzaehler-
Modus
Messzaehler
=0
Uhrzeit/ Datum
RS232-
Parameter
Sprache
Service,
Service 2
Clear measuring data in memory
Permanent measuring of temperature
Selection of output-mode for measuring-data (via printer/RS232)
Selection of measuring-counter mode
Reset measuring-counter
Setting of Time and Date
Setting of RS232-parameters
Selection of user-language
Only for service-personel !
Brookfield Engineering Labs., Inc. Page 64 M/02-212
Index
C
cleaning, 18 computer system, 6, 8, 11 cooling device, 12
ME-CP/PP, 23 ME-CP/PP measuring device, 6, 8, 12, 14, 17 memory, 33, 35 memory, clear, 43 memory, internal, 44 memory, print data, 42 Menü-System, 25
D
Date, setting, 45
E
electrical connections, 10 Error messages, 58
F
FTK-CC, 6, 8, 12, 22
G
General description, 5 guarantee, 54
I
Immersion-measuring, 20 input of numeric values, 30 Installation, 9 instrument handling, 18
K
KE, 12
L
language, 47
O
output mode, 44
P
power supply, 11 printer, 44 Printer, 11 Program, edit, 37 Program, print, 41 PT 100, 10 Pt 100-clamp fixture, 10 Pt100, 8
Q
quick fitting couplings, 13, 14
R
Remote, 51 REMOTE, 11 RS 232, 11 RS232 data cable, 61 RS232 parameter, 46, 47
S
serial data cable, 61 serial interface, 44 Stand, 9 System configuration, 6
M
mains cable, 11 mains connecting cable, 62 maintenance, 19 measurement counter, 45 measurement ID, 45 measurements, 50 measuring counter, 44 measuring system, 6, 12, 20 measuring system, create, 41 measuring system, edit, 41 measuring system, one-way, 20 Measurment Remote, 36 Measurment Single, 32
Brookfield Engineering Labs., Inc. Page 65 M/02-212
T
technical data, 52 temperature control, 22 temperature control device, 6, 8, 12 temperature, permanent measuring, 43 thermostat, 6 thermostatic liquids, 13, 14, 18 Time, setting, 45
U
user language, 47
Loading...