HyQuest Solutions PVD100 Operating Manual

Page 1
HyQuest Solutions Pty Ltd
1
QUALITY
SYSTEM
ISO: 9001
CERTIFIED
PVD100
Current Meter Counter
HYQUEST SOLUTIONS PTY LTD
48-50 Scrivener St, Warwick Farm, NSW 2170, AUSTRALIA
Phone: +61 2 9601 2022 Fax: +61 2 9602 6971
Email: sales@hyquestsolutions.com.au
Web: www.hyquestsolutions.com.au
Page 2
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
2 Issue 1.00 4 Apr, 2007
TABLE OF CONTENTS
1. Introduction........................................................................................ 3
2.1 Overview ............................................................................................................... 4
3. Installation .......................................................................................... 5
3.1 Hardware Connections ....................................................................................... 5
4. Operation ............................................................................................ 6
4.1 Description of Controls ....................................................................................... 6
4.1.1 Pushbuttons .......................................................................................................... 6
4.1.2 Indicator LEDs .................................................................................................... 7
4.1.3 Factory Default Settings ...................................................................................... 7
4.2 The Main Display ................................................................................................ 8
4.3 The Menus ............................................................................................................ 11
4.3.1 Special Operating Functions .............................................................................. 11
4.3.2 Velocity / Counts .................................................................................................. 11
4.3.3 Feet / Meters ......................................................................................................... 11
4.3.4 Magnetic Head / Cat whisker ............................................................................. 12
4.3.5 Normal / Slow ....................................................................................................... 12
4.3.6 Buzzer On/Off ...................................................................................................... 12
4.3.7 Measurement Time .............................................................................................. 13
4.3.8 Bluetooth® Communications ............................................................................... 13
4.3.9 Calibrate ............................................................................................................... 14
4.3.10Comms Test .......................................................................................................... 15
4.3.11Dwell Measurement ............................................................................................. 16
4.3.12Spin Test ............................................................................................................... 17
4.3.13 Battery Condition ................................................................................................. 18
4.3.14Meter Selection .................................................................................................... 19
4.3.15Software Version ................................................................................................. 20
4.4 Serial Communications Interface ...................................................................... 21
4.4.1 Data Conversion Example .................................................................................. 23
4.4.2 The User Serial Interface Program .................................................................... 25
4.4.3 Quick-Step Procedure for the User Serial Interface Program ........................ 29
4.4.4 Operating the Special Functions Using Serial Communications .................... 31
4.5 Changing the Batteries ........................................................................................ 32
4.6 Maintenance and Repair Procedure .................................................................. 32
5. Specifications ...................................................................................... 33
Appendix A Capturing a Spin Test ............................................................... 35
Appendix B Measuring Meter Dwell ............................................................ 40
Appendix C Serial Port Echo Command ..................................................... 41
Appendix D Example Calibration Certificate .............................................. 42
Page 3
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
3 Issue 1.00 4 Apr, 2007
1. Introduction
(The functions described in this manual apply to PVD100 software Version 2.5)
The HyQuest Solutions Point Velocity Display (PVD100) is a small electronic device that
can be used to measure flow velocity from most any mechanically rotating current meter. A
superior feature of the PVD100 is its ability to "clean" the signal and interface directly to a
PDA, PocketPC, or other computing device and provide serial data that can be used by the external device to compute discharge. Meters using a cat whisker head and ball and wire
contact to perform a mechanical switch closure can be used with the PVD100 to produce a clean, noise-free signal. To measure and compute velocity, the PVD100 provides a User Serial Interface Program (USIP). This USIP is used to enter the rating of the current meter
and can support up to two meters with each meter having up to three rating equations with three rotations/sec ranges.
The PVD100 operates with two AA batteries and can mount directly onto a top-setting
wading rod. It offers connectivity between the current meter and a hand-held computer using
either a connecting lead or an optional Bluetooth
®
wireless interface. Commands can be sent
to the PVD100 to perform a measurement and make operating selections. The PVD100 can
operate with current meter velocities from less than 0.015 m/sec to more than 6.1 m/sec for a
Price AA meter. The PVD100 also provides a slow-speed mode to improve measurements with extremely slow rotating current meters [velocities less than 0.076 m/sec]. The PVD100
does not need to be connected to an external device to operate. The visual indicators, LCD
and internal buzzer of the PVD100 can be used to make a conventional current meter
measurement and compute water velocity.
As you continue to read this manual, many features and operating capabilities of the PVD100 will be revealed and described. The PVD100 is extremely easy to use and yet versatile
enough to offer selections to operate in a wide variety of scenarios and field settings. Its lightweight design allows the unit to connect directly onto the phone jack connection of a top­setting wading rod giving the user a nearly ‘hands free’ feel of operation. A single press of a
button can begin a measurement. With its optional Bluetooth
®
capability, true ‘wireless’
connectivity can be made.
The software functions described in these instructions are sufficiently detailed for developers
to integrate the PVD100 with other hardware and software devices giving users the ability to
develop their own applications.
As an added benefit, whenever the PVD100 is inactive, the PVD100 will automatically turn itself Off after 5 minutes to help conserve the batteries. When the PVD100 turns Off, all important settings are saved in
memory, eliminating the burden of re-setting up the device when the unit is turned back
On. To turn it On, simply press the ON/OFF
button and all previous settings are restored ready for the user to begin a measurement.
Page 4
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
4 Issue 1.00 4 Apr, 2007
2. Product Overview
2.1 Overview
The PVD100 can measure and display velocity. Velocity can be either displayed in
meters/sec (m/sec) or feet/second (ft/sec) and is chosen by the user in a menu. When a
measurement is being made, the PVD100 counts contact closures, measures elapse time, and
computes velocity to a precision of 0.001 m/sec or 0.01 ft/sec. The velocity is computed based on a selection of a meter rating again chosen from a menu. During a measurement, the
PVD100 sends through the serial port the number of accumulating meter contact closures and
elapsed time at 1-second intervals. At the end of the measuring period, typically set for 40 seconds, a final count of the total number of contact closures and total measurement time is transmitted. From this data, a connected hand-held computer with an appropriate applications program can compute velocity for a discharge computation.
One of the most important operating features of the PVD100 is the ability to signal process
each contact closure. Contact closures from any type of meter are inherently noisy causing
miscounting or double counting. The PVD100 samples the input signal from the connected
meter and examines the signal to determine if a contact closure has occurred and if successive
signals are of the same characteristics. The PVD100 then determines contact closure
conditions and either counts the signal closure or eliminates the condition as a noise pulse and does not count the signal. For cat whisker contacts, this becomes extremely important since cat whisker contacts offer the poorest quality for marking and counting rotations.
The PVD100 offers (2) modes of operation, i.e. NORMAL and SLOW speed mode.
NORMAL is selected for meter rotations >0.076 m/sec while SLOW mode is for velocities below 0.076 m/sec. To improve noise rejection, the sample rate of the input signal at the SLOW speed mode (lower velocities) is 1/10 of the sampling rate that occurs during NORMAL mode. This slower sampling rate improves performance for back rotations and corroded whisker contacts at extremely low meter cup rotations. It is important to know that
the PVD100 offers operational overlap at this 0.076 m/sec velocity selection in either mode
and therefore it is not necessary for the operator to know precisely the velocity of water for selecting either mode.
Another important feature of the PVD100 is the ability to detect faulty connections from the meter through the wading rod to the connected PVD100. During power up of the PVD100, a
calibration of the input signal is performed. If a connection short is detected or other
condition such as faulty input electronics, the PVD100 signals the operator of the condition
(by flashing the COUNT indicator) until the faulty condition is removed. Open circuit fault conditions are observed simply as a failure to detect a contact closure. A very important
feature of the PVD100 is the ability to detect a signal short during a measurement. If such a
condition is detected, the ‘*’ symbol is displayed until the fault condition is removed. Importantly is the appearance of the ‘*’ symbol after the measurement to show the operator that a fault condition occurred during the measurement and that the measurement should be repeated. The measurement data that is transmitted through the serial port is also flagged to indicate a faulty measurement.
Page 5
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
5 Issue 1.00 4 Apr, 2007
3. Installation
3.1 Hardware Connections
Current Meter Input Connector
The PVD100 provides two banana plug
connections to the current meter.
The DB9 female connector on the PVD100,
provides the RS232 interface and the signal driver output.
RS232 Interface – This interface is designed to directly connect to a handheld computer with
an RS232C interface. Use the Comms Test function to test this interface.
Pin PVD100 Signal Description Direction
1 N.C. 2 Tx Data (RS232 data output) 3 Rx Data (RS232 data input) 4 N.C. 5 Data Ground 0V 6 N.C. 7 N.C. 8 N.C. 9 Signal Driver (O/C output)
Signal Driver Interface—Pin 9 provides an interface to other instruments that may use a
current meter to perform velocity or flow measurements. Examples include the JBS Instruments’ AquaCalc and Sutron’s DMX Discharge Measurement Recorders. The signal provided through this connection simulates a current meter contact closure. Whenever a
contact closure is detected from the current meter, the PVD100 outputs through this
connector, a fixed pulse-width driving signal that represents the contact closure. The signal driver is an open-drain Field Effect Transistor (FET) that will act like a switch. The signal is
processed by the PVD100. To verify operation of this signal with other equipment, use the Comms Test function which simulates a contact closure and operates this output.
CAUTION : Do not connect any type of DC voltage directly to pin 1 without a limiting resistor as this will permanently damage the driver circuit.
Page 6
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
6 Issue 1.00 4 Apr, 2007
4. Operation
4.1 Description of Controls
4.1.1 Pushbuttons
The PVD100 has three front panel buttons that are used to turn power On and Off, view
menus, and choose operating settings. The buttons are labeled as follows:
ON/OFF – This button is the far left button on the front panel and is used to turn power
On and Off on the PVD100. It is also used to quit from operating functions
or exit from selections.
To turn On the PVD100, depress and release the ON/OFF button. The PVD100 will turn On and go through a short initialization restoring settings and selections. The PVD100 will then perform a self-calibration. Some
settings and selections are stored permanently in memory even when the batteries are changed.
To turn Off the PVD100, press and hold the ON/OFF button for more than 3 seconds. The PVD100 will turn off all indicators and the LCD, and then
permanently store all user selectable settings. These settings will be
automatically restored the next time the PVD100 is powered on.
The PVD100 will automatically shut off after (5) minutes of no activity
(i.e. no meter contact closures, no serial port activity, no button presses and Bluetooth switched off or in Discover mode.)
SELECT – This center located button is used to perform selections within the different
menus, and for special operating functions, this button is used to begin or start an operation.
FUNCTION – This button, located on the far right of the front panel, is used to go into a
selection menu. If a measurement is not in progress, depressing this button will bring up a series of menus for choosing operating selections and functions.
When the PVD100 is first turned on, previously selected settings are restored and the battery
is checked. When the battery-check message appears on the LCD, the operator is required to acknowledge the message by depressing any button (i.e. ON/OFF, SELECT, or FUNCTION).
After acknowledging the battery condition message, the PVD100 performs a self-calibration.
This calibration automatically adjusts the internal electronics for the conductivity of the water
and check for any meter fault conditions. After this calibration, the PVD100 shows the Main
display.
NOTE : When all 3 buttons on the PVD100 are held down simultaneously, the electronics performs a hardware reset, which is the same as removing and reinstalling the batteries!
Page 7
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
7 Issue 1.00 4 Apr, 2007
4.1.2 Indicator LEDs
The PVD100 provides two LED indicators to show the unit is operating and to indicate
current meter contact closures. Both indicators also show cable faults and other operating
conditions of the PVD100. The following is a brief description for each indicator.
COUNT (Orange LED)—This indicator will flash whenever a contact closure from the meter
has been detected. It also will indicate when special functions such as the “Meter Dwell Measurement” are being performed or the User Serial Interface program has been invoked.
In the case of a detected fault, the indicator will turn On and stay On until the fault is cleared.
OPERATING STATUS (Green/Red/Blue LED)—This indicator remains illuminated whenever the PVD100 is turned On and showing the operating status mode. The LED has
three colors showing the type of meter (Magnetic Head or Cat Whisker Head) selected as well as Bluetooth
®
operational status. Green indicates the Magnetic Head is selected and red
indicates the Cat Whisker Head is selected. The operating default (power-up setting) is Magnetic Head. The indicator also shows the operating mode. If the indicator is not
blinking, then the operating mode is ‘Normal’. If the indicator is blinking rapidly, then the operating mode is ‘Slow’.
4.1.3 Factory Default Settings
The PVD100 can be set with Factory Settings (or defaults) and these settings allow the PVD100 to operate with preset conditions. These settings can be changed. To set the PVD100 with Factory Settings, depress and hold the FUNCTION button, then turn On the PVD100. When the ‘Factory Settings’ message appears on the display, the FUNCTION
button can be released.
The following is a list of the Factory Settings:
- Magnetic Head/Cat Whisker Head selection set to MAG, Magnetic Head meter
- STATUS is set to Normal speed
- Measurement Time is set to 40-seconds
- The Buzzer is turned Off
- The following are the rating equation defaults:
o A: S/N 1000-00 (Meter ‘A’ Serial Number) o 1 Rating (Number of ratings or equations) o 1: 2.2048(n)+0.0178 (Rating equation, n= rotations/sec) o B: S/N 2000-00 (Meter ‘B’ Serial Number) o 1 Rating (Number of ratings or equations) o 1: 0.9604(n)+0.0312 (Rating equation, n=rotations/sec)
- Meter ‘A’ selected for velocity computation
- Feet – Meters/second display set to ‘Feet’/sec
- All other settings are initialized to ‘0’ or turned Off
Page 8
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
8 Issue 1.00 4 Apr, 2007
M XXX 40
N XXX.XX *
M XXX 40
N XXX.XX *
M XXX 40
N XXX.XX *
M XXX 40
N XXX.XX *
4.2 The Main Display
The following describes the Main display presentation on the LCD. Note: This display presentation is shown when the Menu selection VELOCITY/COUNTS is set to COUNTS.
This display position shows the selection of the meter contact
switch type. The selection can be M for a Magnetic Head contact or C for a Cat Whisker contact switch.
This display position shows the selection of the operating
speed. The operating speed (or mode) selection can be N for Normal Speed (for velocities >0.076 m/sec) or S for Slow
Speed (velocities <0.076 m/sec).
This display position shows the total accumulating meter contact counts during a measurement. Maximum number that can be shown is 999. Whenever meter contact counts exceed
999, the PVD100 will continue to count but the most
significant number position will begin to show special characters such as ‘:’ or ‘;’, etc. These characters indicate to the operator that a rollover count of ‘999’ has occurred.
This display position shows the total elapsed time of the measurement. During a measurement, the display shows elapsed time in seconds. When the measurement is completed, the display shows the total elapsed time to a tenth of second. If any measurement exceeds 300 seconds of elapsed time, the
display will show the final elapsed time only in seconds when the measurement is completed. A ‘^’ character is shown in front of the accumulating elapsed time during and after completion of the measurement to indicate elapsed time has exceeded
300 seconds. The maximum displayed number is 999, however the PVD100 will ‘roll-over’
to ‘0’ and continue to count until a meter contact closure is detected. To terminate the measurement upon the next meter contact closure, depress the ‘SELECT’ button. To terminate the measurement at any time and stop the count, depress and release the ‘ON/OFF button. After the measurement, to clear the total elapsed time, depress and release the ON/OFF button while in the Main display.
Page 9
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
9 Issue 1.00 4 Apr, 2007
M XXX 40
N XXX.XX *
M XXX 40
N XXX.XX *
This display position shows the selection of the measuring interval. The measuring interval is chosen within a selection menu. The measuring interval is selected in 10-second increments to a maximum of 90 seconds. A measuring interval selection of ‘0’ means a ‘continuous’ measurement with no
time limit. With this time interval selection, the operator must depress the ‘SELECT’ button to stop the measurement at the next meter contact closure or the ‘ON/OFF’ button to simply stop and end the measurement.
This display position shows the indication of a current meter fault. When the indicator appears, a meter fault has been detected. When the fault is cleared, the indicating character will disappear. If a fault has been detected during the measurement, the symbol will appear at the end of the measurement and will remain shown on the display until a new measurement is made.
In Summary (VELOCITY/COUNTS selected for COUNTS)
M XXX 40
N XXX.XX *
M = Mag Head C = Cats Whiske
r
Contact Counts
Measuring Interval in Secs 0 = Continuous
N = Normal Speed S = Slow S
peed
Elapsed Time in Secs
Blank = OK * = Fault Detected
Page 10
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
10 Issue 1.00 4 Apr, 2007
The following describes the Main display presentation on the LCD when the Menu selection VELOCITY/COUNTS is set to VELOCITY.
This display position shows the units as either ‘m’ for meters/second or ‘f’ for feet/second. The selection is performed in the FEET/METERS menu selection.
This display position shows velocity. Velocity precision (number of decimal places) is presented based on the selection of either feet/second or meters/second. If the FEET/METERS selection menu is set for FEET, the display will show velocity to two decimal places, while if the selection is set to METERS, the display will show velocity to three decimal places.
(Trick: if you want to get more resolution in the display of feet/second, simply select meters/second
knowing that the rating equation is entered as feet/second. When ratings are entered, they are dimensionless; hence the calculation is performed without units. You must know that the rating is entered as feet/second even though the display shows meters/second.)
In Summary (VELOCITY/COUNTS selected for VELOCITY)
X. XX f 40
N XXX.XX *
Units: f = feet/sec
m= meters/sec
Velocity:
-in meters: X.XXX m
-in feet: X.XX f Measuring Interval in Secs
0 = Continuous
N = Normal Speed S = Slow S
peed
Elapsed Time in Seconds
Blank = OK * = Fault Detected
X.XXf 40
N XXX.XX *
X.XXXm40
N XXX.XX *
Page 11
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
11 Issue 1.00 4 Apr, 2007
FEET
METERS
4.3 The Menus
To review or choose operating conditions and settings, a series of menus provide for each selection or function. To enter a menu, depress and release the FUNCTION button. Each time the FUNCTION button is depressed, a new menu is presented. To make a selection within a menu, depress the SELECT button. For most menus, an arrow character, ‘ ‘, points to or shows the present selection within that menu. To exit from a menu at any time, depress and release the ON/OFF button. To continue through the menus, depress and release the FUNCTION button. Menus can only be viewed when a measurement is not being performed.
4.3.1 Special Operating Functions
Within the PVD100 are special operating functions that offer tools for adjusting the current
meter and measuring its performance. These special functions are described in order of
appearance within the menu list. Refer to Section 4.4.4 below for the details of operating
these functions using the serial communications interface.
4.3.2 Velocity / Counts
This is the first menu. The menu allows selection of the main display to show velocity; or counts and meter type selection. Use the FUNCTION button to select this menu, then press SELECT to toggle between Velocity (spelled VELOCTY in menu) or Counts. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item.
4.3.3 Feet / Meters
Use the FUNCTION button to select this menu, then press SELECT to toggle between Feet or Meters. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item.
VELOCTY
COUNTS
Page 12
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
12 Issue 1.00 4 Apr, 2007
MAGHEAD
CATWSKR
NORMAL
SLOW
ON-BUZR
OFF
4.3.4 Magnetic Head / Cat whisker
Allows selection between a Magnetic Head contact or a Cat Whisker contact switch. Use the FUNCTION button to select this menu, then press SELECT to toggle between Magnetic Head or Cat Whisker contact. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item.
One of the most important operating features of the PVD100 is the ability to signal process
each contact closure. Contact closures from any type of meter are inherently noisy causing
miscounting or double counting. The PVD100 samples the input signal from the connected
meter and examines the signal to determine if a contact closure has occurred and if successive
signals are of the same characteristics. The PVD100 then determines contact closure
conditions and either counts the signal closure or eliminates the condition as a noise pulse and does not count the signal. For cat whisker contacts, this becomes extremely important since cat whisker contacts offer the poorest quality for marking and counting rotations.
4.3.5 Normal / Slow
Use the FUNCTION button to select this menu, then press SELECT to toggle between Normal or Slow. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item.
The PVD100 offers (2) modes of operation, i.e. NORMAL
and SLOW speed mode. NORMAL is selected for meter rotations >0.076 m/sec while SLOW mode is for velocities below 0.076 m/sec. To improve noise rejection, the sample rate of the input signal at the SLOW speed mode (lower velocities) is 1/10 of the sampling rate that occurs during NORMAL mode. This slower sampling rate improves performance for back rotations and corroded whisker contacts at
extremely low meter cup rotations. It is important to know that the PVD100 offers
operational overlap at this 0.076 m/sec velocity selection in either mode and therefore it is not necessary for the operator to know precisely the velocity of water for selecting either mode. When in SLOW mode, the status LED flickers, and the sounder beeps longer whenever the whisker contacts are made.
4.3.6 Buzzer On/Off
Use the FUNCTION button to select this menu, then press
SELECT to toggle between Buzzer On or Off. Press ON/OFF
to return to the main display or press FUNCTION to step to the next menu item.
Page 13
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
13 Issue 1.00 4 Apr, 2007
TIME
40 SEC
OFF - BT
ON COM
BLUTOOTH
DISCOVER
4.3.7 Measurement Time
Use the FUNCTION button to select this menu, then press
SELECT to increment the time in 10 second steps between 00
and 90 seconds. At this instant the selection is saved, so either
press ON/OFF to return to the main display or press
FUNCTION to step to the next menu item.
This is the nominal measurement time used when taking a current meter reading. When the time is set to 00 seconds, the timer runs until stopped manually by pressing the SELECT button, which completes the measurement at the next contact closure.
4.3.8 Bluetooth
®
Communications
Bluetooth
®
provides for a wireless link to an external device such as a PocketPC (PDA) or
laptop that eliminates the serial RS-232 cable link connection. Bluetooth
®
is a wireless
standard (IEEE 802.14) that offers a transparent serial port protocol for communication.
Since the PVD100 can output data and be controlled using the RS-232 serial interface, the physical cable connection can be eliminated using Bluetooth®. When the PVD100 is turned
On, the PVD100 determines if Bluetooth® is installed. If available, Bluetooth
®
can be
enabled and ‘discovered’ by another Bluetooth
®
device to establish a connection. Set the
external device’s communication protocol in the following manner:
Baud Rate : 19200 Data bits : 8 Stop bits : 1 Parity : None
Flow control : None
Use the FUNCTION button to select this menu, then press
SELECT to toggle between Bluetooth
®
On and Off, then
press ON/OFF to enable Bluetooth
®
.
The following display presentation is shown when setting up
and enabling Bluetooth
®
.
When this display is presented, the STATUS indicator will
be blue and flashing at a rate of two quick flashes every two seconds. When another Bluetooth
®
compatible device is in
proximity of the PVD100, and the devices application requests a connection, each automatically establishes the Bluetooth
®
serial interface protocol. When the connection has
been established, the following display is shown on the PVD100 display:
Page 14
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
14 Issue 1.00 4 Apr, 2007
CALIBRT
YES
CONNECTD
When a wireless connection is made, the STATUS indicator
will show a solid blue color continuously – and the display will read “CONNECTD” for a few seconds before automatically jumping to the Main Display.
When making a connection with another Bluetooth
®
device for the first time, first examine
and prepare the required enabling features of the external device for Bluetooth
®
connection.
Enable the PVD100 Bluetooth
®
function and allow the external device to ‘discover’ the
PVD100. When the external device makes ‘discovery’, the user must select the PVD100
within the device’s menu setup and enable the connection. The external device should
respond with a ‘connected’ message. The PVD100 can now be used in a normal manner. All serial data sent from the PVD100 will be transmitted via Bluetooth
®
. The connection can be
disengaged by selection in the ‘-BT COM’ menu. The operator must monitor the integrity of the connection from the external device. If the connection becomes broken, the PVD100 will revert back to Bluetooth® DISCOVER mode. This message will alert the operator of a disconnection between the PVD100 and the external device. Connection will have to be
rediscovered by the external device. Any active measurement occurring during a broken communications link will terminate and the measurement will need to be retaken. The
PVD100 will not automatically power down when Bluetooth
®
is enabled and connected. If
Bluetooth
®
is On (or ‘connected’) when the PVD100 is manually powered Off, the PVD100 will go directly into ‘DISCOVER’ mode when the PVD100 is turned back On. If the PVD100 is in a ‘DISCOVER’ mode, the PVD100 will remain in this mode for 5 minutes, then automatically turn Off if the PVD100 does not become ‘connected’. However, when the
PVD100 is turned back On, the PVD100 will automatically go into ‘DISCOVER’ mode.
4.3.9 Calibrate
A calibration is performed automatically when the PVD100 is turned on and when a
measurement is actioned through the comms interface. Performing a calibration within the
menu can be used to check the input electronics of the PVD100 for proper operation when
not connected to a meter. It is also useful to check the connectivity to the current meter. If a
detectable fault is found, the COUNT indicator will turn On and stay On and the STATUS indicator will turn On and Off to indicate a connection short or some other electronics fault.
When the fault is cleared, the indicators will reflect their fault-free operating condition.
Finally, it may be necessary to recalibrate the PVD100 in areas where the water has a high
conductivity. Use this function when the meter and wading rod or meter mounting are completely in the water. This will insure a proper calibration and adjustment of the PVD100 to the water conductivity prior to making a measurement.
To perform a manual calibration, use the FUNCTION button to select this menu, then press SELECT to toggle to YES and perform the calibration. (This takes less than 1 second)
Page 15
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
15 Issue 1.00 4 Apr, 2007
COMTEST
YES
SER PORT
TEST
4.3.10 Comms Test
The ‘Serial Port Test’ function, called ‘COMTEST’ in the menu, provides a way to test the
setup of the serial communications interface between the PVD100 and an external device connected (cabled or wireless) to the PVD100.
Set the external device’s communication port as follows :
Baud Rate : 19200
Data bits : 8
Stop bits : 1
Parity : None
Flow control : None
Use the FUNCTION button to select the COMTEST menu, then press SELECT to toggle the arrow to YES, then press ON/OFF to start the test.
When the test is invoked, the display will read “SER PORT TEST” and the single ASCII character ‘1’ is transmitted through the serial interface. If the connected device’s serial communications interface is properly configured, the character will be received. The character is sent at a rate of about two times a second. Depressing and holding the ‘SELECT’ button for more than two seconds can send an alternate ASCII character string. This alternate character string represents the
data string format that would normally be sent through the serial interface when making a current meter measurement. To deselect this alternate character string, depress and hold the ‘SELECT’ button again until the single character ‘1’
appears. The screen capture in the figure to the left, shows both character strings sent through the serial interface while this test is being performed. The string sent through the serial interface can be
viewed when using the Microsoft program HyperTerminal. When the test
is invoked, the ‘COUNT’ indicator flashes and the Signal Driver O/C output operates (to simulate a contact closure) and the ‘STATUS’ indicator shows an
alternating Red/Green flash.
To end the test, simply depress and hold the ‘ON/OFF’ switch until the ‘STATUS’ indicator
stops flashing, then release the ‘ON/OFF’ switch. HyperTerminal is an ‘Accessories’ program in the Microsoft
®
operating system found on most laptops and computers. Operating
selections, measurement data, and special PVD100 operating functions can all be selected and controlled through the serial interface using HyperTerminal or any other comms terminal
application.
Page 16
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
16 Issue 1.00 4 Apr, 2007
DWELL?
YES
CAT- - -
- - - - - -
4.3.11 Dwell Measurement
Using the Function button scroll to the “Dwell ?” menu and press Select to move the arrow to Yes, and then press the ON/OFF button to enter the dwell function. Refer to
Appendix B Measuring Meter Dwell for a further
explanation of this function.
The opening display for the special operating function called ‘DWELL’ is shown. This function measures the dwell angle (ratio of meter contact time to the total time for a single
rotation) for a meter contact and is useful for making adjustments to a cat whisker type meter in the field. For a cat whisker contact head, the optimum adjustment is 60º or a single meter cup width (US Geological Survey Price AA or Pygmy meter). A variation to an adjustment of +/- 10% is acceptable. For this function, (3) measurements are taken from
a rotating meter. The results of the measurement are displayed in whole degrees of dwell. The measure of dwell angle for the magnetic head contact does not apply but can be performed. To perform the measurement, the following steps are used :
1. Before enabling this function from the menu, select the type of contact head, i.e magnetic head or cat whisker head contact. This is done from a different menu.
2. After selecting the type of meter contact head, when the ‘DWELL’ measurement function is enabled, the opening display presentation will show the type of meter
contact selected to make the measurement. CAT is shown when a cat whisker head is chosen and MAG is shown when a magnetic head contact is chosen.
3. From the ‘DWELL’ measurement function display as shown above, free spin the meter at a rate of about 2 rotations per second.
4. Depress the ‘SELECT’ key.
5. As each measurement is taken, a whole number display of dwell angle (in degrees) will be shown. If no contact closures are detected within a 4 second period, the display will read “NO DWELL”. Conversely, if the contact closures are too long, the display will read “EX DWELL” indicating exceeded dwell.
6. To repeat the measurement, simply spin the meter and depress the ‘SELECT’ key.
7. During the measuring of dwell angle, the ‘STATUS’ indicator will flash ‘red’ at
each detected contact closure to show a cat whisker head is selected and the
display shows CAT. If the STATUS indicator is flashing green during the
measurement, the meter selection is Magnetic Head and the display will show
MAG. Again, select the type of meter contact head before selecting this function
to measure the dwell angle of the meter.
8. To exit from this function, simply depress the ‘ON/OFF’ button.
Page 17
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
17 Issue 1.00 4 Apr, 2007
SPIN 270
FINL 243
SPIN ?
YES
SPIN 0
TEST 0
4.3.12 Spin Test
Using the Function button scroll to the “Spin ?” menu and press SELECT to move the arrow to Yes, and then press the ON/OFF button to enter the spin function.
This opening display for the special function called ‘SPIN TEST’ is shown. A ‘spin test’ is an elapsed time measurement from the time a meter is spun to when rotation of the meter stops. The operator controls the start and stop time for a total elapsed time measurement. The test measures total elapsed
time and total number of meter counts detected. Start and stop time is controlled by
depressing the ‘Select’ button. Perform the following steps to ‘spin test’ a meter:
1. Before selecting the ‘SPIN TEST’ function, select the type of meter contact, i.e. cat whisker head contact or magnetic head contact.
2. From the opening display presentation as shown above, depress the ‘SELECT’ button.
3. Spin the meter. When the PVD100 detects the first contact closure (which is not
counted) the elapsed time counts begins. Each following contact closure detected is counted and shown on the top line of the display. The bottom line of the display
shows the continuing elapsed time in seconds. When the meter stops rotating, depress the ‘SELECT’ button again. The display presentation shown in the figure on the left gives the final summary of the total number of contact closures detected and total elapsed time in seconds. Maximum number of elapsed time is 999 seconds and the maximum total number of meter counts that is displayed is 999.
4. To exit from the ‘SPIN TEST’ function at any time, depress the ON/OFF button.
The PVD100 will exit from the function; perform a calibration, and then return to
the Main display.
***NOTE*** : While the Spin Test is in progress, serial data is automatically transmitted on the serial port. See Appendix A for an explanation of how to collect and display the results of
a Spin Test. The Spin Test can also be invoked using the serial interface command ‘N’.
Refer to Section 4.4.4 Operating the Special Functions Using Serial Communications and Appendix A for a detailed description of how to perform a Spin Test using the serial
communications port.
Page 18
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
18 Issue 1.00 4 Apr, 2007
BATTERY
-OK-
BATTERY
*LOW*
4.3.13 Battery Condition
For battery conditions, the PVD100 continuously monitors the power level of the batteries. In the event a low battery condition is detected, the PVD100 turns On both the red and green
LED on the STATUS indicator to inform the operator. The menu function showing the
battery condition is also messaged to show a low battery condition. If the PVD100 is turned Off with a low battery condition, upon turning On the PVD100, only the red LED on the STATUS indicator is turned On (at a lower indication level) to inform the operator that the PVD100 will not turn On until the batteries are changed. This prevents poor operation from a low battery condition. Whenever the PVD100 is turned On, the battery is checked and the
condition of the battery is displayed for the operator. The operator must acknowledge the
message before any other operation of the PVD100 can be performed. A low battery condition is set at approximately 2.8 V dc (or 1.4Vdc/battery). The PVD100 will work at
lower battery voltages but is not recommended since batteries tend to deplete very quickly at lower voltages.
Use the FUNCTION button to select this menu and view the battery condition.
If the battery voltage falls below 2.8V then it will be indicated as low.
Page 19
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
19 Issue 1.00 4 Apr, 2007
A=S/N
1000-00
1:2.2048
+0.0178
2:2.1234
+0.0456
4.3.14 Meter Selection
The meter selection menu is used to select the meter that will be used in the measurement and for the final computation of velocity. There are two meters that may be selected and they are
designated by the PVD100 as ‘A’ and ‘B’. The ratings for the two meters must have already been entered using the special serial communications program designated ‘User Serial Interface Program’ (refer to Section 4.4.2 for a description of this interface program). Each
meter has a serial number and a rating for the meter, which can have up to (3) equations and (3) ranges of rotations/second (designated as ‘n’). It should be noted that during a measurement, only the first term (slope) of any rating selected within the menus is used in the calculation of velocity. This is to give the operator a close approximation of velocity while the measurement is being performed and is intended only for real time display purposes. At the end of the measurement, the full calculation is performed. If a meter has more than (1)
rating, then the proper range (n) value (rotations/second) is computed to select the proper
rating and perform the final velocity calculation. There is no LCD menu for displaying the range values in the format they are entered via the serial interface.
Use the Select button to view each meter’s serial number and rating(s). To select a meter for the final velocity calculation, use the Select button to bring the serial number into view, then press ON/OFF to return to the main display or press FUNCTION to step to the next menu item. To select a specific rating to use for calculating velocity only while the measurement is taking place, use the select button to view the specific rating, and then press
ON/OFF to return to the main display or press FUNCTION to step to the next menu item.
When the serial number is viewed, the PVD100 automatically selects the first rating equation
(whether a meter has one or more ratings) for the real time display of velocity during a measurement. After a measurement and the final calculation of velocity, the user can select a different rating in the menu to see a different velocity calculation on the same measurement.
This display shows the first rating for the meter. The number of the rating is shown as the first digit on the top row of the display followed by a colon (:). Only numeric values are shown to verify the selection for the first and second terms for the meter. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item. Pressing the Select button moves the
display to the next rating of the meter or to the serial number of the next meter.
This display shows the second rating for the meter. Only the numeric values are shown to verify the selection for the first and second terms. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item. Press the Select button to go to the next rating or to the serial number of the next meter.
Page 20
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
20 Issue 1.00 4 Apr, 2007
V 2.5
3:2.3456
-0.1234
1:2.1234
-0.0456
B=S/N
2000-00
This display shows the third rating for the meter. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item. If the ON/OFF or FUNCTION button is pressed, this rating will be used to compute velocity only during a measurement using only the first term of the rating. For the real time display of velocity during a measurement, only the first term of the rating is used (i.e. 2.3456).
This display shows the serial number for meter ‘B’. Press
ON/OFF to return to the main display or press FUNCTION to step to the next menu item. In this display setting, the computation of velocity for the real time display will use the first rating of the meter even if the meter has more than one rating. Press the Select button to show the values of the first rating of the meter.
This display shows the first rating for the meter. Only the numeric values are shown to verify the selection for the first and second term for the meter. Press ON/OFF to return to the main display or press FUNCTION to step to the next menu item. If the Select button is pressed, the display will show the next rating or show the serial number of the next meter.
4.3.15 Software Version
Use the FUNCTION button to select this menu and view the
PVD100 Software Version.
Page 21
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
21 Issue 1.00 4 Apr, 2007
4.4 Serial Communications Interface
The PVD100 can be connected to a hand-held or notebook computer or any device that can transmit and receive serial communications using RS-232C or Bluetooth
®
using an
application such as HyperTerm. Commands can be sent to the PVD100 to perform special
functions or select operating modes. The following describes the serial data requirements, commands, and data formats that are used for the serial communications interface :
Baud Rate : 19,200 RS-232C Format : 8 data bits, no parity, 1 stop bit Flow Control : None
The following list shows the commands and functions that can be performed through the serial communications interface.
Command Function Performed
S Start a measurement T Terminate the measurement I Abort the measurement R Resend the last data string U Buzzer ON Z Buzzer OFF H Speed Select, set to Normal L Speed Select, set to Slow M Select Magnetic Head meter processing C Select Cat Whisker Head meter processing E Enter User Serial Interface Program (Meter ratings) P Start a measurement without performing a calibration Q Start a measurement with no time limit (continuous measure) V Software Version
NOTE: Commands sent to the PVD100 are case sensitive – and must be upper case only.
When the commands ‘S’, ‘T’, and ‘I’ are received by the PVD100 from an externally
connected device, the PVD100 sends an acknowledge character ‘A’ back to the device. For the ‘S’ command, the PVD100 sends acknowledge character ‘A’ after the PVD100 performs and completes the calibration. Note that invalid or unrecognized
commands will result in a ‘?’ being transmitted through the serial port.
Page 22
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
22 Issue 1.00 4 Apr, 2007
S Start a measurement: Begins a measurement at the first contact closure detected.
Number of contact closures and elapsed time counts are accumulated. At the first
contact closure, the PVD100 begins sending the accumulated contact closures and
elapsed time counts through the serial port every second. The first data string transmitted through the serial port begins the start of the measurement interval with the meter contact closure count tally and elapsed time tally of ‘0’. Note that when
this command is received by the PVD100, it automatically performs a calibration
then monitors the input for meter contact closures. Following is the output data format sent to the externally connected device at the 1-second output rate:
dnn,xxxx_
d = preamble for indicating display data
nn = number of meter contact closures, transmitted in HEX ASCII
(hexadecimal notation)(i.e. 0-9,A-F) , = delimiter xxxx = counts of elapsed time measured as 0.003333 seconds/count,
transmitted in HEX ASCII (i.e. 0-9,A-F) while in Normal mode
and 0.03333 seconds/count while in Slow mode. _ = space
This data format that is transmitted to the externally connected device can be used
to display elapsed time and velocity while the measurement is being performed.
At the end of the measurement interval or if the ‘T’ command is received, the
PVD100 sends the last tally of accumulated counts and time in the following
format:
fnn,xxxx or enn,xxxx
f = final data, no detected faults
e = final data, error detected during measurement;
restart the measurement
The measuring interval is based on the user selected measurement time. When the
measuring interval has been reached, (or the T command is received) the PVD100,
upon the next meter contact closure, will send the final data for that measurement. If there is no contact closure detected when the measurement interval has been
reached, the PVD100 will continue to send data and the user must send the ‘I’
command to abort the function. Note that the meter contact number, nn, is limited to a total of 255 counts (hexadecimal notation ‘FF’). If more than 255 counts are detected throughout the measurement interval, the count will rollover to ‘00’ and continue to accumulate counts. No fault condition is set if this occurs. This allows for continuous counting in excess of 255 counts occurring during the measuring interval. External processing software can easily detect this condition and process total meter counts when the measuring interval is completed. The same condition occurs for the elapsed time tally. This permits an elapsed time count of 218 seconds for a measuring interval before the count rolls over to ‘0000’.
Page 23
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
23 Issue 1.00 4 Apr, 2007
T – Terminate the measurement: PVD100 stops the measurement at the next contact
closure and, at the next contact closure (which is counted), sends the last tally of accumulated counts and elapsed time as previously described.
I – Abort the measurement: The PVD100 aborts the measurement. No final
measurement data is transmitted.
P Start a measurement without doing a calibration: This command is used to begin
a measurement without performing a calibration. The command should only be used for a high velocity measurement that does not allow a calibration to be
performed due to the rapid rotation of the meter cups.
Q Continuous measurement: This command is used to continuously output a 1-
second data display without a time limit. The ‘T’ or ‘I’ command must be used to
terminate the operation.
V – The software version of the PVD100 is sent in the following manner: vX.X Example: v1.0
4.4.1 Data Conversion Example
The data format of the PVD100 is transmitted in HEX ASCII (base 16 numbers using the
ASCII code which is displayable on any terminal) to provide efficiency in the number of data characters sent through the serial interface. Following is an example of the interpretation of
the data. The example is given with the PVD100 in NORMAL Speed mode. Example: d0C,0AF6
Accumulated meter counts = 0C16 = 12
10
Elapsed time = 0AF616 = 2806
10
= (2806 counts x 0.003333 sec/count) = 9.35 secs
The data format says a total of 12 meter counts have been accumulated in 9.35
seconds.
Shown below is a simple Excel program that easily converts hexadecimal numbers to decimal (base 10) numbers. The column ‘Decimal Conversion’ shows the Excel formula.
Hexadecimal
Decimal
Number Conversion
1F =hex2dec(A3)
0 1 2 3
A B
Page 24
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
24 Issue 1.00 4 Apr, 2007
Data Format Notes
The data format presented by the PVD100 during a measurement provides for displaying a
computed velocity by an external device at a 1-second rate. The accumulating contact closure counts and time interval counts can be used in a rating equation to display velocity. At the end of the measuring interval, a final total of contact closures and interval time is then used to compute the measured velocity. For the Normal mode, the 1-second output data format displays the accumulating time interval counts based on a 300 Hz sample rate. The
following example below shows how it would be presented using the ‘HyperTerminal’
program: NORMAL mode, 300 Hz sample rate, counts accumulating at a 1 Hz rate.
For Slow Speed mode operation, the display of the accumulating time interval counts differs in that the value is based on a 30 Hz sample rate. The following example below shows how it would be presented in the Slow Speed mode:
30 Hz sample rate, counts accumulating at a 0.25 Hz rate (1 count every 4 seconds)
Page 25
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
25 Issue 1.00 4 Apr, 2007
4.4.2 The User Serial Interface Program
The User Serial Interface Program allows the user to input the rating(s) for a current meter. The user connects the PVD100 to a laptop, desktop, PocketPC, or other type of field
computing device having a terminal emulator software program (such as HyperTerm) for communication. Once connected, the user then enters on the remote terminal device the
upper case letter ‘E’. Once entered, the user selects between meter ‘A’ or ‘B’ and then enters
the following information about the current meter:
a) Serial Number, b) Number of equations (or ratings), c) First term (slope) and signed second term (intercept) for the rating, d) A rotations/second range n (if a current meter has more than one equation for its
rating).
The following procedure can be used to enter all information necessary for the PVD100 to compute velocity for a measurement. To exit anytime from the User Serial Interface Program, use the ‘Esc’ key.
Step 1: Connect the PVD100 using the serial communications cable or Bluetooth®
connection. If a Bluetooth® connection is desired, first enable the Bluetooth® connection between the PVD100 and the remote terminal device before entering the User Serial Interface Program.
Step 2: To enter the User Serial Interface Program (USIP), enter ‘E’ on the keyboard.
The USIP responds with the following message:
A=S/N 1000-00 B=S/N 2000-00
A, B or S? _
The Serial Number for each meter is displayed followed by a selection request
from the user. Selecting ‘A’ means choosing meter ‘A’ or meter #1. ‘B’ designates meter #2. ‘S’ is an entry that will provide a summary of the rating(s)
and ranges for each meter and gives the user a way to verify entries.
Step 3: The user must enter ‘A’, ‘B’, or ‘S’.
The following instructions will assume Meter ‘A’ has been selected.
The USIP responds with the Serial Number for Meter ‘A’ (or meter #1). For this
example, the USIP responds with the following default Serial Number:
A: S/N 10000-00 The letter ‘A’ designated above specifies the meter number and is not alterable by
Page 26
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
26 Issue 1.00 4 Apr, 2007
the user. This is what the PVD100 uses to distinguish between the two meters.
The user enters any characters and numbers for the Serial Number. Up to seven alphanumeric, numeric, and special characters all can be used to give a serial number for the meter. Three control keys are used for position and data entry control:
- The ‘Space bar’ key inserts a ‘space’ into the character position.
- The ‘Backspace’ key moves the cursor position backwards without changing
the character.
- The ‘Enter’ or ‘Carriage Return’, <CR> key accepts the entry and moves to
the next data entry element.
Step 4: Enter the Serial Number for the meter. When the user has completed the entry,
press the <CR> key to enter the Serial Number. The USIP responds with the
following message:
NUMBER OF EQUATIONS? 3
Each meter (‘A’ or ‘B’) can have up to three equations for the full rating of the meter. The number 3 shown above indicates what was previously selected for the number of equations for the meter rating. The user enters ‘1’, ‘2’, or ‘3’ for the
number of equations for the rating. If the meter has been calibrated, the calibration certificate will show the number of equations that describe the full
rating for the meter. Refer to Appendix D for an example of a calibration
certificate supplied with a meter.
Step 5: The user enters ‘1’, ‘2’, or ‘3’ for the number of equations for the rating. For this
example, the number ‘3’ is entered. The USIP responds with the following Range 1 value, n, in meters/second or feet/second;
1: n < 0.50
This is a default range and allows the user to enter the proper range as shown on
the fan calibration certificate provided by HyQuest Solutions Pty Ltd. when the calibration of the meter was performed. There is no need to enter feet/second or meters/second since units are selected in a selection menu.
Step 6: Enter the Range 1 value as shown on the calibration certificate, then press <CR>
on the keyboard (i.e.’Enter’ key). Any numbers up to 9.99 can be used for the
range. The PVD100 will only allow the user to enter numbers. Three control
keys are used for position and data entry control:
- The ‘Space bar’ key moves the cursor position forward without changing the
number.
- The ‘Backspace’ key moves the cursor position backwards without changing
the number.
- The ‘Enter’ or ‘Carriage Return’, <CR> key accepts the entry and moves to
the next data entry element.
Page 27
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
27 Issue 1.00 4 Apr, 2007
Once entered, the USIP responds with the following Range 2 value:
2: 0.50 < n < 3.75
This is a default range and allows the user to enter the proper range shown on the
provided fan calibration certificate. Range 2 for a three equation fan rating is listed
as the second range given for the second equation listed. Refer to Appendix D for
an example. Again, any numbers up to 9.99 can be used for the range and only
numbers can be entered. After entering the Range 2 value, press the <CR> key on
the keyboard. The USIP responds with the final Range 3 value which the user is not required to enter:
3: n > 3.75
Step 7: After entry of the Range values, the user presses the <CR> key and the USIP
responds with the first rating equation for the meter:
1: 2.2048[n]+0.0178
The first number shown on the line is the rating equation number. Up to three
equations can be shown depending on the user’s response to the ‘Number of Equations…’ entry. For data entry position and control, the following keys can
be used:
- The ‘Space bar’ key moves the cursor position forward without changing the
number.
- The ‘Backspace’ key moves the cursor position backwards without changing
the number.
- The ‘Enter’ or ‘Carriage Return’, <CR> key accepts the entry and moves to
the next data entry element.
When a number is entered, the cursor automatically positions to the next number
entry position. For the sign, the USIP will only recognize the ‘+’ or ‘-‘ keys. When the entry is completed, press the <CR> key to go to the next rating
equation. For any rating equation, the maximum allowable number for the first term (slope) is 6.5535. The maximum allowable number for the second term
(intercept) is 0.9999. At the completion of the entry, pressing the <CR> accepts
the complete entry and displays the next equation (if there is more than one),
otherwise the USIP moves to the beginning (Step 2 above) to choose meter ‘B’ or review and make entry changes to meter ‘A’ or to request a Summary.
Note: Incorrect or unreasonable entries can give unexpected results
for velocity calculations. Be sure to provide the proper entries for these ratings.
Page 28
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
28 Issue 1.00 4 Apr, 2007
Step 8: After all entries have been completed, the user presses <CR> to display the Serial
Numbers of both meters and make a selection. The USIP responds with the
following message (as is shown in Step 2 above):
A=S/N 1000-00 B=S/N 2000-00
A, B or S? _
For this display (as in Step 2 above) ‘S’ is an entry that will provide a summary of
the rating(s) and ranges for each meter.
Entering ‘S’ will show the following results (numbers shown are arbitrary and are
for example):
----------------------------
A=S/N 1000-00 3 Ratings
Range 1: n<1.23
2.3456[n]+0.2345
Range 2: 1.23<n<4.56
1.2345[n]-0.1234
Range 3: n>4.56
2.3456[n]-0.2345
---------------------------- B=S/N 2000-00 1 Rating
0.1234[n]+0.0123
----------------------------
The summary listing shows meter ‘A’ having three equations to describe the rating while meter ‘B’ shows only one equation for the rating. When the summary has been displayed, the user enters any key entry including the ‘space bar’ key, <CR>, or ‘ESC’. Entering <CR> or ‘ESC’ causes the program to end. Entering any
character or the ‘space bar’ will move to display the Serial Numbers of both
meters and make a selection as shown in Step 2 above.
Page 29
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
29 Issue 1.00 4 Apr, 2007
4.4.3 Quick-Step Procedure for the User Serial Interface Program
(Numbers are shown for example)
Step 1: Press E’ on the keyboard.
USIP response: A=S/N 1000-00
B=S/N 2000-00
A, B or S? _
Step 2: Enter A, B, or S. (see Step 7 for entry of ‘S’)
USIP response: A=S/N 1000-00 if the user enters ‘A’
or
B=S/N 2000-00 if the user enters ‘B’
Step 3: Edit the Serial Number, then enter <CR> (‘Carriage Return’).
USIP response: NUMBER OF EQUATIONS? 3
Step 4: Enter 1, 2, or 3.
USIP response: 1: 2.2048[n]+0.0178 if the user enters 1 for Number of Equations
or
1: n < 0.50 if the user enters 2 or 3 for Number of
Equations
Enter the first range value for n (rotations/second).
USIP response: 2: n > 0.50 if the user enters 2 for Number of Equations then
1: 2.2048[n]+0.0178
or
2: 0.50 < n < 3.75 if the user enters 3 for Number of Equations
Page 30
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
30 Issue 1.00 4 Apr, 2007
Enter the second range value for n (rotations/second).
USIP response: 3: n > 3.75 if the user enters 3 for Number of Equations then
USIP response: 1: 2.2048[n]+0.0178
Step 5: Edit the equation, then press <CR>.
USIP response: 2: 2.2048[n]+0.0178 if the user enters 2 or 3 for Number
of Equations
or 3: 0.9604[n]+0.0312 if the user enters 3 for Number of
Equations
Step 6: Edit the final equation, then press <CR>.
USIP response: A=S/N 1000-00
B=S/N 2000-00
A, B or S? _
Step 7: Enter S’. (Values are arbitrary and for example)
USIP response: ----------------------------
A=S/N 1000-00 3 Ratings
Range 1: n<1.23
2.3456[n]+0.2345
Range 2: 1.23<n<4.56
1.2345[n]-0.1234
Range 3: n>4.56
2.3456[n]-0.2345
---------------------------- B=S/N 2000-00 1 Rating
0.1234[n]+0.0123
----------------------------
Page 31
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
31 Issue 1.00 4 Apr, 2007
Step 8: Press <CR> or ‘Esc’ key to exit the program or any key entry or ‘space bar’ key
to repeat Step 1.
USIP response: A (acknowledge) and program ends.
or A=S/N 1000-00
B=S/N 2000-00
A, B or S? _
4.4.4 Operating the Special Functions Using Serial Communications
The special function ‘SPIN Test’ can be operated using the serial communications interface.
A laptop, desktop PC, or PocketPC using a terminal emulating package or Microsoft®’s HyperTerminal can be used to invoke the function. No menu selections in the PVD100 are needed. The selection of the meter type and turning the buzzer On and Off can all be done
using the serial communications interface. These functions can also operate using
Bluetooth
®
.
The following list shows the commands and functions that can performed through the serial communications interface.
Command Function Performed
N Spin Test S Start Spin Measurement (subcommand in ‘Spin Test’) A Stop Spin Measurement (subcommand in ‘Spin Test’) U Buzzer ON Z Buzzer OFF M Magnetic Head Meter C Cat Whisker Head Meter
I Abort function
When the ‘N’ command is received by the PVD100 from an externally connected device, the PVD100 echoes acknowledge character ‘N’ back to the device.
Refer to Appendix A for details in using the Spin Test and capturing a ‘spin test file’.
Page 32
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
32 Issue 1.00 4 Apr, 2007
4.5 Changing the Batteries
NOTE : For optimum performance, the
batteries should always be replaced with two good quality AA alkaline batteries !!
Simply clip open the two side panels as shown and remove the 4 screws, then flip the top cover forward.
Remove the old batteries and discard thoughtfully.
Insert the new batteries taking care of the battery orientation – which is stamped on the bottom of the battery enclosure.
After replacing with new batteries, if the Power ON/OFF button is depressed and the unit
appears not to turn On, do the following:
- Depress and release all three buttons at the same time. This is the same as
performing a hardware Reset. All settings will be restored and the unit will operate
normally. Be sure not to hold down the FUNCTION button too long as this causes the PVD100 to power up with Factory Defaults. If this occurs, simply remove the
batteries while the unit is On, then replace the batteries. All previous settings should be restored.
Please Note: Installing the batteries backwards will not damage the PVD100.
4.6 Maintenance and Repair Procedure
There is no specific maintenance for the PVD100. As with any electronic device, care must
be taken to insure minimal exposure to moisture, dirt, or other contaminants. The unit has waterproof connectors and an “O” ring seal between the two case halves, which prevents the ingress of water and is rated to IP65.
There are no user serviceable parts inside the PVD100. If the unit does not function properly,
the unit must be returned to HyQuest Solutions for component repair.
The PVD100 does not require any calibration or adjustments. The PVD100 is self-
calibrating and will perform this operation whenever the unit is turned on and whenever
CALIBRT is selected via the menu, or whenever a start measurement command (‘S’) is
received through the serial communications interface.
Page 33
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
33 Issue 1.00 4 Apr, 2007
5. Specifications
Bluetooth
®
Class 2 / Output power 2.5mW (4dBm)
Compliant Bluetooth
®
Spec V1.1
Up to 10m (30ft) range (operating at 2.4GHz)
Communications RS232 Port (Tx, Rx) @ 19200 baud 8/N/1
Connections 2 x Banana plug sockets (Meter input)
1 x DB9 female socket (RS232 + filtered meter o/p)
Indicators 8 character x 2 line LCD
1 x Orange LED
1 x Red/Green/Blue LED
Power Source 2 x Internal AA alkaline 1.5V batteries Low battery indication 2.8V
Dimensions 120mm x 80mm x 22mm (L x W x D)
Weight 218 grams (with batteries)
Environmental -5C to + 50C, Non Condensing, IP65
(Note: 60º = 1 Meter cup width for 6 cups)
Water Velocity Flow (absolute maximum)
(1)
(Normal Mode ) Price AA meter - Mag 13.7m/s (45 ft/s) @ 99o to 204o dwell
- CatW 6.7m/s (22 ft/s) @ 10o to 62o dwell Pygmy meter - CatW 2.8m/s (9.4ft/s) @ 10o to 62o dwell
(Slow speed mode) Price AA meter - Mag 0.76m/s (2.5ft/s) @ 54o to 265o dwell
- CatW 0.61m/s (2.0ft/s) @ 17
o
to 70o dwell
(2-cup width) 0.33m/s (1.1ft/s) @ 120
o
dwell Pygmy meter - CatW 0.27m/s (0.9ft/s) @ 17o to 70o dwell (2-cup width) 0.15m/s (0.5ft/s) @ 120o dwell
Water Velocity Flow (absolute minimum) – before setting fault detection
(1)
(Normal Mode ) Price AA meter - Mag 0.04m/s (0.14ft/s) @ 15o to 210o dwell
- CatW 0.03m/s (0.1 ft/s) @ 10o to 90o dwell Pygmy meter - CatW 0.018m/s (0.06ft/s)@ 10o to 90o dwell
(Slow speed mode) Price AA meter - Mag 0.012m/s(0.04ft/s)@ 10o to 340o dwell
- CatW 0.021m/s(0.07ft/s)@ 10
o
to 170o dwell
Pygmy meter - CatW 0.012m/s(0.04ft/s)@ 10
o
to 90o dwell
Maximum Conductivity > 50,000 µSiemens (max velocity ranges must be derated 20%)
Page 34
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
34 Issue 1.00 4 Apr, 2007
Specifications… (continued)
Contact time to Fault Detection (absolute maximum)
(Normal Mode) Magnetic Head Contact 11 Seconds Cat Whisker Head Contact 7 Seconds
(Slow speed Mode) Magnetic Head Contact 30 Seconds Cat Whisker Head Contact 20 Seconds
Minimum dwell angle for 10o Cat Whisker Head
Signal Output driver pulse Normal mode : 100mS, active high to low pulse (FET open drain O/P) Slow mode : 1 Sec, active high to low pulse (Fixed width)
(1)
NOTES: a) Measurements and Computations based on the following Rating Table
Equations.
Price AA Meter: V=2.2048 +0.0178 where: V= Velocity
T T = Seconds/rotation
Pygmy Meter: V=0.9604 +0.0312 T
b) Measurements made with 0 microSiemens conductivity, open air, ambient
room temperature.
User Serial Interface Program Example: 2.3456[n]+0.1234 where: n=number of rotations/second
First Term: 2.3456 Second Term: 0.1234
First Term Entry: Maximum Allowable: 6.5535 Second Term Entry: Maximum Allowable: 0.9999 Range Values: Maximum Allowable: 9.99
Note: Incorrect or unreasonable entries can give unexpected results.
Page 35
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
35 Issue 1.00 4 Apr, 2007
Appendix A Capturing a Spin Test
The spin test function allows the user to perform a simple spin test of the meter. The
PVD100 outputs a character string representing the meter count and elapse time. Whenever the ‘spin test’ measurement begins, the PVD100 will output through the serial data port a
character string at every contact closure. This character string is of the following format:
nxxx,tttt[cr][lf] n = preamble for indicating contact closure data
xxx = number of meter contact closures, transmitted in decimal (i.e. 001, 002, 003, etc) , = delimiter
tttt = counts of elapse time measured as 0.00666 second counts (150 Hz), transmitted in HEX ASCII (i.e. 0-9,A-F)
[cr] = carriage return (ASCII ‘0D’)
[lf] = line feed (ASCII ‘0A’)
At the end of the spin test, the PVD100 sends a final data string that displays the total number
of counts and total time all shown in decimal notation. This format makes it easy to copy the final total number of counts and total elapse time information onto a form. Total time is given in seconds and displays the time to a tenth of a second. The format is shown below.
dxxx,ttt.t[cr][lf]
d = preamble for indicating spin measurement data in decimal
xxx = number of meter contact closures, transmitted in decimal (i.e. 001, 002, 003, etc)
, = delimiter
ttt.t = total seconds of elapse time displayed to a tenth of a second
[cr] = carriage return (ASCII ‘0D’)
[lf] = line feed (ASCII ‘0A’)
Shown below is an example of both formats, as they would be displayed using the
HyperTerminal program.
N
Perform ‘Spin Test’
n000,0000
First contact closure
n001,0013
n002,0027
n003,003A
.
.
.
n162,427A
n163,4607 Final contact closure
n163,4736 Data string when “Stop Spin” executed
d163,121.4 Final meter counts and time (in seconds)
A PVD100 initialization completed
Page 36
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
36 Issue 1.00 4 Apr, 2007
The following information details some specifics about each format, specifically the limitations and alternate characters that may be presented.
1) The maximum time that can be displayed, during the measurement, is ‘FFFF16’. If the measurement continues beyond the maximum data form available, the count ‘rolls over’ and begins from ‘0’. When this occurs, the delimiter character ‘,’ is replaced by the ‘>’ character to indicate the rollover condition. When the data is transferred to a spreadsheet, the character, ‘>’, can be easily detected and used to add ‘65535’ (= FFFF16) to the elapse time tally. No further change in the delimiter character occurs if the accumulating time continues to further rollover.
2) The maximum time that can be displayed as given by the final data string is 436.4 seconds as displayed (FFFF16 x 0.00666 seconds). This equates to 7 minutes and 16.4 seconds. If the accumulating time continues during the measurement (as discussed above), the final count tally will reflect a calculated value rolled over and beginning from ‘0’. If this occurs, the delimiter character ‘,’ is replaced by the ‘>’ character to indicate a rollover. To determine the total accumulated time from the roll over condition, simply add 436.4 seconds to the total number displayed.
3) The maximum number of measured meter contact closures (clicks) is 999. If the accumulating contact closures continues beyond the maximum data form, the count ‘rolls over’ and the most significant digit position number is replaced by the colon (‘:’) character. If the count continues, other characters will replace the most significant digit position number. Meter counts should not exceed 999, however a detection scheme is provided if such an occurrence happens.
The “Spin Test” has three subcommands as follows:
S – Start spin measurement.
A – Stop spin measurement, transmit final results, then quit.
I – Abort or quit the “Spin Test” function. No final data is transmitted.
The “Spin Test” is performed in the following manner (through the serial interface) :
1. Before entering the “Spin Test” function, select the type of meter, i.e. Cat Whisker Head, or Magnetic Head Meter. Also enable or disable the buzzer.
2. Press N” on the computer keyboard to enter the “Spin Test”. The COUNT LED will become illuminated. The OPERATING STATUS indicator will illuminate green if the meter is selected for a Magnetic Head and red if Cat Whisker Head is selected. The PVD100 echoes with ‘N’.
3. Depress the letter “S” on the computer keyboard or press the SELECT button on the PVD100. The OPERATING STATUS indicator will turn off.
4. When the meter is spun, the OPERATING STATUS indicator will blink when
a contact closure is detected.
5. The PVD100 will output through the serial port a character string (as described
Page 37
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
37 Issue 1.00 4 Apr, 2007
previously) every time there is a meter contact closure. At the first contact
closure, the PVD100 outputs the first contact closure as the start of the
measurement. The number of meter contact closures begins with “zero” and the counts of elapsed time begins with “zero.” Contact closures and counts of elapsed time will begin to increment following every contact closure until the
A” command is sent or the SELECT button is pressed. The PVD100 then sends the final data string (in decimal) followed by the PVD100 reinitializing. The character ‘A’ is sent when the PVD100 completes a calibration cycle, as it
initializes.
Sending the ‘I’ command anytime before or during the test aborts (or quits) the “Spin
Test” function. The PVD100 sends the character ‘A’ to acknowledge exiting. When the PVD100 exits, the PVD100 reinitializes with a calibration. When the calibration has been completed, the PVD100 sends the character ‘A’ again, and then goes to normal
operation mode.
Capturing a “Spin Test” file
“Spin test” transmitted data can be saved by using the options in the HyperTerminal
program. HyperTerminal is a program that can be found on most computers in the ‘Accessories’ folder listed in the Programs list on the desktop. Within HyperTerminal and before spinning the meter, select “Capture Text” found in the Windows label “Transfer”
tab. When the “Capture Text” tab is clicked, Windows will ask for a file name. The file
name extension will be “.txt”. Select a file name, then click “Start” to save the screen
results. After completing the spin test, select “Capture Text” again under the “Transfer” tab. Click “Stop”. The file will be saved under the file name given. The file can be
opened using Windows “NotePad” or “WordPad” and can be easily imported into Excel
for analysis.
To better align the data being displayed, the PVD100 will respond to the Carriage Return
key,
C
R
. Entering a
C
R
, the PVD100 will send a
C
R
followed by a Line Feed,
L
F
. This will cause the display cursor to return to the left most position on the display
and move down the page by one line position.
Use the following steps to import a text file into Excel. This procedure uses Microsoft
Excel 2000.
1. Open up Excel for Windows.
2. Under the ‘Data’ tab, select ‘Get External Data’.
3. Within ‘Get External Data’, select ‘Import Text File’.
4. Select the location of the text file and click ‘Import’.
5. The ‘Text Import Wizard’ pops up with the ‘Delimited’ radio button on. Also, a portion of the actual text file is shown at the bottom of the window.
6. Click up the ‘Start import at row:’ tab to start at row 2.
7. Click ‘Next’ at the bottom to go to Step 2 of the Wizard.
Page 38
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
38 Issue 1.00 4 Apr, 2007
8. In the ‘Delimiters’ window,
a) deselect ‘Tab’, b) select ‘Comma’, c) select ‘Other’, then type ‘n’ in the box next to ‘Other’.
The text file preview should show a vertical line separating the meter counts
column data and the time data column.
9. Click ‘Next’, this will bring you to Step 3 of the Wizard.
10. In Step 3 window, click ‘Do not import column (skip)’ in the ‘Column data format’ window. This will show ‘Skip’ in the column 1 heading.
11. Click on column 3 (where the Time data is shown in HEX-ASCII) and change in the ‘Column data format’ window from ‘General’ to ‘Text’.
12. In the ‘Data Preview’ window, the columns should show the following selections on the column heading:
a) ‘Skip’ for column 1 b) ‘General’ for column 2—Current meter counts c) ‘Text’ for column 3—Time
13. Click ‘Finish’, and then click ‘OK’ to import data into the existing worksheet.
14. Save the workbook under a file name.
15. Shown below is an example of an imported ‘Spin Test’ file.
Page 39
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
39 Issue 1.00 4 Apr, 2007
Shown in the figure below is an example of a graph showing the response of a current
meter created from data in a “Spin Test” file.
Spin Test for Meter 1234567
0
10
20
30
40
50
60
70
80
90
100
110
120
130
1 6 11 16 21 26 31 36 41 46 51 56 61 66 71 76 81 86 91 96 101 106 111 116 121 126 131 136 141 146 151 156 161 166
(Total Time) minus (Start-to-End Time)
Total Spin Time = 119.3 seconds
Total Number of Meter Counts = 163
Page 40
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
40 Issue 1.00 4 Apr, 2007
CAT 62 57 60
Appendix B Measuring Meter Dwell
The meter dwell measurement function performs a measure of the contact time in relation to a single rotation of the meter. For the cat whisker contact, it is the measure of the cat whisker contact time on the cam lobe. The resulting measurement is called “dwell”. Dwell is expressed in terms of degrees and for a cat whisker contact; the contact time should be adjusted to be the width of a single cup. This adjustment results in the dwell angle of 60º for a meter with 6 cups and represents optimum performance. The magnetic contact meter (Mag Head) requires no adjustment and therefore measuring the meter dwell angle does not apply
but can be performed. When a meter dwell measurement is performed, the PVD100 takes
three measurements and displays the result on the LCD after each measurement, as shown.
The “Dwell Angle” can be computed by using the following simple formula:
Dwell Angle = (Contact time) x 360 (Total Time)
The contact time and total time shown above in the formula are expressed in terms of the
number of samples measured at a sample rate of 1200 Hz as it is measured by the PVD100.
A high sample rate is used to reflect a more accurate measure of the contact time and total time measured. The repeatability and variability in measuring the dwell is directly related to mechanical factors such as the condition of the pivot bearings, and the cat whisker wire contact surface and cam lobe. Wide variations in readings reflect conditions that may require further inspection of the meter and its mechanical parts. Some variation in readings is acceptable and represents proper adjustment; however there should be no more than a 10% difference between all the readings. Optimal setting is near 60º (a single meter cup width). High conductivity of >25,000 micro Siemens can affect the operating performance of the
PVD100 and should a current meter be required to operate in high conductivity water, a
lower dwell angle setting should be used. A setting of not less than 20º for higher velocities (>0.65 m/sec or 2 ft/sec) is suggested.
Page 41
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
41 Issue 1.00 4 Apr, 2007
Appendix C Serial Port Echo Command
The PVD100 has a special serial port character echo command that can be used to echo
characters or short strings appearing through the serial port. The command can be used to
‘see’ what characters are being transmitted to the PVD100. This command can be useful to
software application developers. To invoke the command, simply send the single ASCII character ‘~’. On a laptop keyboard, this character is located on the numbers key row and is the left most character key in the row. The character is a shift key character and the ‘shift’ key must be used to send the character. This echo character command acts like a switch
which turns On character echo or turns Off character echo. When ‘character echo’ is turned On, any character received through the serial interface is echo’d back through the port and any character sent will not be interpreted as a function or operating command to the PVD100.
Shown below is a simple pictorial of the Serial Port function of the PVD100. Using
Bluetooth
®
and an additional serial port device connected to the hardware port of the
PVD100, can assist developers in their applications and ‘seeing’ what commands are sent to the PVD100.
PVD100
PVD100
Page 42
HyQuest Solutions Pty Ltd
Point Velocity Display - Model PVD100
© Copyright
42 Issue 1.00 4 Apr, 2007
Appendix D Example Calibration Certificate
HYQUEST SOLUTIONS PTY. LTD., AUSTRALIA
------------------------------------------
CERTIFICATE OF CALIBRATION Date: 27/4/06
-------------------------­Current Meter Model: xxxxxx Serial No. 06-999 Fan No. 1 Serial No. 06-123 Diameter: 125mm Pitch: 0.25m Type of Support: Method of Calibration: Average Value Equation
n < 0.42 V = 0.2190 n +0.0153 m/s
0.42 < n < 3.73 V = 0.2459 n +0.0041 m/s n > 3.73 V = 0.2508 n -0.0142 m/s
Starting Velocity = 0.025 m/s Maximum Velocity = 5.000 m/s
Note: 'n' denotes the number of revolutions of the propeller per second 'V' is the water velocity in meters per second.
Date: 27/4/06 INSPECTED:__________________
_____________________________________________________________ | The above meter has been calibrated in accordance with | | AS 3778.6.3-1992. This standard is identical with and has | | been reproduced from ISO 3455:2007. | | The clause below has been printed with the permission of | | Standards Australia from AS 3778.6.1-1992. This standard is | | identical with and has been reproduced from ISO 2537:2007. | | Clause 7.6 Recalibration. Meters shall be recalibrated | | whenever their performance is suspect. In practice for | | individually rated meters, recalibration is sometimes | | carried out at yearly intervals or after 300 hours of use, | | whichever is the shorter. | |_____________________________________________________________| D127.08.00
Range Values First Term
(Slope)
Second Term
(Intercept)