WARRANTY
ONE YEAR LIMITED WARRANTY
The PH-200, manufactured by HM Digital, Inc. ("the Company") is warranted to the purchaser
against defective materials and workmanship for one (1) year from the date of purchase.
What is covered: Repair parts and labor, or replacement at the Company's option.
Transportation charges for repaired or new product to be returned to the purchaser.
What is not covered: Transportation charges for the defective product to be sent to the
Company. Any consequential damages, incidental damages, or incidental expenses, including
damages to property. This includes damages from abuse or improper maintenance such as
tampering, wear and tear, water damage, or any other physical damage. The PH-200 is
watertight and completely submersible. Please ensure that the battery compartment and probe
gasket ring are firmly tightened before submersing in water. The warranty does not cover water
damage to the PH-200 due to parts not securely closed. Products with any evidence of such
damage will not be repaired or replaced.
How to obtain warranty performance: Attach to the product your name, address, description
of problem, phone number, and proof of date of purchase, package and return to:
HM Digital, Inc.
ATTN: Returns
5819 Uplander Way
Culver City, CA 90230
U.S.A.
Implied Warranties: Any implied warranties, including implied warranties of merchantability and
fitness for a particular purpose, are limited in duration to five years from date of purchase. Some
states do not allow limitations on how long an implied warranty lasts, so the above limitation may
not apply to you. To the extent any provision of this warranty is prohibited by federal and state
law and cannot be preempted, it shall not be applicable. This warranty gives you specific legal
rights, and you may also have other rights, which vary from state to state.
NOTE: Warranties are product-specific. Third-party products and products deemed by HM
Digital as "accessories" are not covered under warranty. Third-party products include, but are
not limited to, batteries and fittings. Accessories include, but are not limited to, precipitator rods,
fuses, lanyards and product cases.
This manual and all information contained herein is copyright HM Digital, Inc.
USER’S GUIDE
pH / TEMP METER
PH-200
PH-200 1/08
CONTENTS
Overview……………………………………………………………. 1
Contact Information………………………………………………...1
Specifications………………………………………………………. 1
Housing and Screen Description………………………………… 2
Taking Measurements…………………………………………….. 3
Calibration………………………………………………………….. 4
Care, Maintenance & Techniques………………………...…….. 5
Warranty……………………………………………………………. 7
OVERVIEW
Thank you for purchasing HM Digital’s PH-200 Meter. The PH-200 is an advanced, microprocessor-based handheld, watertight pH meter with superior accuracy. It can be used for
testing the acidity or alkaline levels of almost any liquid, in a wide variety of applications,
provided the liquid contains less than 50% alcohol and a conductivity of at least 10 µS (5 ppm in
TDS). (It will not work in pure, distilled or de-ionized water.) It also measures temperature in
both Celsius and Fahrenheit. The meter incorporates automatic temperature compensation
(ATC) to accommodate for discrepancies in temperature that would naturally occur. The meter
also features digital calibration for easy and precise calibration that will yield better results.
CONTACT INFORMATION
If you have any problems or questions regarding your meter, please contact HM Digital, Inc.
HM Digital, Inc.
5819 Uplander Way
info@hmdigital.com
www.hmdigital.com
Culver City, CA 90230 1-800-383-2777
SPECIFICATIONS
pH Range: 0.00 – 14.0 pH
Temperature Range: 0-80 °C; 32-176 °F
Resolution:
pH: 0.01 pH
Temperature: 0.1 °C/F
Accuracy: +/- 0.02 pH
Temperature Compensation: Automatic (ATC) to 25 °C
Calibration: Auto-calibration to 4, 7 or 10 pH with fine tune digital calibration by push button.
Auto Shut-Off: After 5 minutes
Probe: Glass sensor and glass reference tube (single junction)
Safe Operating Temperature: 0-80 °C; 32-176 °F (measuring in extreme temperatures is not
recommended)
Minimum Conductivity: For accurate and stable readings, the liquid to be tested must have a
conductivity of at least 10 µS (approx. 5 ppm in TDS).
Display: LCD panel
Housing: Waterproof (submersible); floats
Power source: 3 x 1.5V button cell batteries (included), model 357A
Dimensions: 18.5 x 3.4 x 3.4 cm (7.3 x 1.3 x 1.3 inches)
Weight: 95 grams (3.35 ounces)
1
Changing the batteries:
When the meter displays a flashing battery symbol, your batteries
are getting weak and should be replaced soon. To change the
batteries:
1. Twist open the battery compartment on the top of the meter.
2. Remove the three batteries.
3. Insert new batteries in the direction as depicted inside the
compartment. The meter uses 357A batteries.
4. Close the battery compartment. Make sure it is tightly closed to
retain waterproofness.
NOTE: Do not reverse the polarity of the batteries. This will short circuit the meter.
Cleaning and sensor care:
1. Never touch the sensor!
2. To clean the unit, use a soft rag or towel. Wipe with water and a mild soap or rubbing alcohol.
3. To clean the sensor, rinse in distilled water or a pH 7 buffer solution. If cleaning with a pH 7
buffer solution, lightly dab (touch) the sensor on a soft tissue. Never rub the sensor with a tissue
or anything else, as this may scratch the glass.
4. If it is found that the glass sensor is scratched, it needs to be replaced.
5. If the sensor is dried up (or the gel within it), you can attempt to re-moisten the sensor by
keeping it in distilled water or a pH 7 solution for at least two hours. If the meter still does not
function properly or accuracy is not at optimum performance, the sensor may be irreparably
damaged and must be replaced.
6. For best results, clean the sensor after each use, especially if used in very low or very high pH
liquids, or liquids other than water.
7. PROBE STORAGE: Storing the probe in a 4 pH KCl buffer solution is preferred (the meter is
shipped with this solution in the cap). It is okay to rinse the probe in distilled water, but never
store the probe in distilled water. Storing the probe in a higher pH buffer solution will not
damage it.
Sensor Replacement:
If your sensor has been damaged, you can purchase a new one without having to purchase a
new meter. To replace the sensor:
1. Remove the sensor gasket ring by twisting it counter-clockwise.
2. Gently pull the sensor off the unit.
3. Gently insert the new sensor into the unit. Be sure to align the grooves properly. Never force
the sensor into the unit!
4. Make sure the rubber ring is properly positioned on the sensor.
5. Screw the gasket ring back onto the unit by twisting it clockwise. Tighten.
Sponge Care:
1. Never add distilled or tap water to the sponge/cotton ball, as mildew or mold may form.
2. Rewet with HM Digital’s bottling solution or a professionally made KCl solution with
bactericide.
6
PH-200 1/08