Hydro 38741GB, 38751GB Quick User Manual

Page 1
AccuDose Series Proportioner
Models 38741GB & 38751GB
with E-Gap Eductors
Package Should Contain:
1. Proportioner unit.
2. Supply tubing 6.4m total.
3. Foot valve assemblies & weights (5).
4. Discharge tubing for each eductor.
5. Metering tip kits.
6. Mounting anchor kit.
7. Hose hook for 14 LPM eductor.
8. Instruction sheet.
Installation and Operation:
1. Remove cabinet cover. Drill holes for the three wall anchors with an 8mm drill bit, using the cabinet back as a template for correct spacing of the mounting screws. Install mounting anchors, and then screws in top two anchors. Slide key holes in cabinet back over screw heads, tighten screws, then install bottom screw. Do not mount more than 1.8 metres above the bottom of the concentrate container, nor below the highest concentrate level (never mount your concentrate higher than the proportioner).
2. Select metering tips (up to 4) for the selector valve. (see next two sections) Push each tip firmly into a separate hose barb extending from the selector valve. A tip with no hole (clear plastic) can be used to block any valve port not being used. (This may be used for dispensing water only).
3. Attach the chemical suction tube assembly to the selector valve, sliding the open end of each piece of tubing over one barb on the valve. Make sure suction tubes are on the barb far enough to prevent air from leaking into tube.
4. Slip other end of supply tube through an opening in either side of the cabinet and slide a ceramic weight over the tube, then push a footvalve into the end of the tube. (Repeat for all eductors.)
5. Place foot valve ends of supply tubes into concentrate containers. REMEMBER TO CHECK FOOT VALVE STRAINERS PERIODICALLY FOR CLOGGING: CLEAN IF NECESSARY.
6. A short discharge tube is used with the 4 LPM eductor; minimum tube length is 20cm for correct operation. Longer tubes (1.2 m) are used with a 14 LPM eductor. Do not remove the flooding rings from inside the tubes. Slide end of tube with flooding ring over eductor discharge outlet. (Repeat for all eductors.) Hooks may be installed on longertubes to allow discharge tube to hang from dispenser when not in use.
7. Replace cabinet cover. Push the sides in, behind the latch holes, to snap the cover in place. The two screws provided may be installed in the holes in the cabinet sides to prevent easy removal of cover.
8. Connect water supply hose of at least 13mm ID to water inlet swivel. (Minimum 1.76 Bar pressure, with water running, is required for correct operation.) Connect opposite end of hose to water supply. Turn water supply on.
9. Push button to start flow of desired water/concentrate solution, and hold until supply tube is primed (filled). Then push the button whenever dispensing is desired, and release button to stop flow of solution. If you wishto be able to lock the button in the “on” position: clip or bend the two tabs behind the lower front portion of the button (see diagram). This allows the button to be pushed fully in and so lock in the “on” position. To unlock, PULL THE BUTTON OUT.
Clip or bend these tabs to depress button into locked position.
Page 2
Metering Tip Selection:
The final concentration of the dispensed solution is related to both the size of the metering tip opening and the viscosity of the liquid being siphoned. For water-thin products, the chart at right can be used as a guideline. If product is noticeably thicker than water, consult the Measurement of Concentration Procedure below to achieve your desired water-to-product ratio. Because dilution can vary with water temperature and pressure, actual dilution achieved can only be ascertained by using the Measurement of Concentration Procedure. The clear, undrilled tip is provided to permit drilling to size not listed should you need a dilution ratio that falls between standard tip sizes.
NOTE: A 4 LPM eductor is grey; a 14 LPM eductor is yellow. Refer to parts diagram if unfamiliar with names of system components.
Dilution Ratio (X:1) where X = Amount of Mixed Solution — Amount of Concentrate Drawn
Dilution Ratio, then, equals X parts water to one part concentrate (X:1). If the test does not yield the desired ratio, choose a different tip and repeat the test. Alternative methods to this test are 1) pH (using litmus paper), and 2) titration. Contact your concentrate supplier for further information on these alternative methods and the materials required to
perform them.
AT 40 PSI FOR WATER-THIN PRODUCTS (1.0 CP)
No Tip .187 (3/16) 3:1 6.5:1 Grey .128 (30) 3:1 6.5:1 Black .098 (40) 3:1 7:1 Beige .070 (50) 4:1 10:1 Red .052 (55) 5:1 16:1 White .043 (57) 7:1 20:1 Blue .040 (60) 8:1 24:1 Tan .035 (65) 10:1 30:1 Green .028 (70) 16:1 45:1 Orange .025 (72) 20:1 56:1 Brown .023 (74) 24:1 64:1 Yellow .020 (76) 32:1 90:1 Aqua .018 (77) 38:1 128:1 Purple .014 (79) 64:1 180:1 Pink .010 (87) 128:1 350:1
Amount of Concentrate Drawn
APPROXIMATE DILUTIONS
Orifice Std. Drill Size Number)
Ratio (per Eductor Flow)
1 GPM 3.5 GPMTip Color
AccuDose Select Parts Diagram List:
4
6
3
2
1
5
Key DescriptionPart No.
1 238100 Strainer washer 2 10082835 Swivel collar (moulded) 3 10082816 Swivel stem (moulded)
10075950 O-ring (two per nipple) 4 10075925 Pipe plug 5 10080710 Button, grey (standard)
10080711 Button, sky blue
10080712 Button, red
10080713 Button, green
10080714 Button, light grey
10080715 Button, yellow 6 10080894 Cabinet 7 10020700 Selector valve grommet
7
10020900 Back up ring for grommet
Page 3
AccuDose Select Parts Diagram:
12a 12b
9
a
b
c d
15
e
g
f
h
16
10 11
Suction tube assembly (materials for 4 provided) attach open ends to barbs on valve.
13a 13b
14
Key DescriptionPart No.
9 10075980 Valve parts kit
a. diaphragm, b. armature,
c. spring, d. valve bonnet 10 10079010 Spring 11 10079000 Magnet 12 290GB 4 LPM eductor assembly (kit)
291GB 14 LPM eductor assembly (kit)
13 a 10088822 4 LPM discharge tube assembly
b 10074815 14 LPM discharge tube assembly
14 10080730 Hose hook, dark grey (standard)
10080731 Hose hook, sky blue 10080732 Hose hook, red 10080733 Hose hook, green 10080734 Hose hook, light grey
10080735 Hose hook, yellow * Hose hooks are for 14 LPM discharge tubes 15 10080957 Selector replacement kit:
e. support clip f. Suction stub,
g. O-ring, h. selector valve assembly 16 690014 Metering tip (kit) 17 500870 Tubing, 6mm x 2.1m 18 509900 Weight 19 10076301 Footvalve, viton (EPDM also
available order 10076302)
NOT SHOWN:
641750 Screws (cabinet sides)
17
18
19
Page 4
Troubleshooting Chart:
Problem
1. No discharge a. No water a. Open water supply
2. No concentrate draw a. Clogged foot valve a. Clean or replace
3. Excess concentrate draw a. Metering tip not in place a. Press correct tip firmly into barb on
Cause Solution
b. Magnetic valve not functioning b. Install valve parts kit c. Excessive water pressure c. Install regulator if water pressure
exceeds 4.3 Bar (flowing) d. Eductor clogged d. Clean* or replace e. Clogged water inlet strainer e. Disconnect inlet water line and clean
strainer
b. Scale build-up in metering tip b. Clean (descale)* or replace
or eductor
c. Low water pressure c. Minimum 1.76 Bar (with water running)
required to operate unit correctly d. Discharge tube and/or flooding d. Push tube firmly onto eductor discharge
ring not in place hose barb, or replace tube if it does not
have a flooding ring. e. Concentrate container empty e. Replace with full container f. Inlet suction stub not screwed f. Tighten, but do not overtighten
into eductor tightly
g. Clogged water inlet strainer g. Disconnect inlet water line and clean
strainer h. Air leak in chemical pick-up tube h. Put clamp on tube or replace tube if
brittle i. Selector out of position i. Ensure selector is in desired position
selector valve b. Chemical above eductor b. Place concentrate below the eductor
4. Failure of unit to turn off a. Water valve parts dirty or a. Clean* or replace with valve parts kit
5. Excess foaming in discharge a. Air leak in pick-up tube a. Put clamp on tube or replace tube if
* In hard water areas, scale may form inside the discharge end of the eductor, as well as in other areas of the unit which are exposed to water. This scale may be removed by soaking the eductor in a descaling solution (deliming solution). To remove an eductor located in the cabinet, firmly grasp water valve and unthread eductor. Replace in same manner. Alternatively, a scaled eductor can be cleaned (or kept from scaling) by drawing the descaling solution through the unit. Operate the unit with the suction tube in the descaling solution. Operate the unit until solution is drawn consistently, then flush the unit by drawing clear water through it for a minute. Replace concentrate container and put suction tube into concentrate.
defective b. Magnet does not fully return b. Make sure magnet moves freely. c. Push button stuck c. Remove button and clean cabinet/
button to remove any dirt lodged in slide recess
brittle
Hydro Systems Europe Ltd, Unit 3 The Sterling Centre, Eastern Road, Bracknell, Berks RG12 2PW UK Tel: +44 (0)1344 48 88 80 Fax: +44 (0)1344 48 88 79 www. hydrosystemseurope.com
10089611
Loading...