1. Operating water pressure: 200 psi (Factory set)
2. Maximum water source temperature 140° F
3. Operating ambient temperature: 40-120° F
4. Electrical supply
a. 120 / 208 Single Phase
b. 1.5 hp pump 20 amps @ 120 volts or 10 amps @ 208
volts
5. Operate valves with 24VAC, 24VDC, 120VAC
a. < .5 amp
6. Water supply
a. 3/4” ID flooded inlet required
b. Inlet pressure: 2-60 psi
7. Air supply
a. Pneumatics inlet pressure: 60 - 100 psi
b. 3/8” feed per system
c. 20 CFM @ 100 psi supply
d. 3 x 3/8” poly lines to applicator
8. Chemical Solution lines
a. 3x ½” poly lines to applicator
9. Space requirements:
a. AQUA-LAB Systems
i. Valve panel – 1.5’ wide x 2’ tall
ii. Pump – 1’ wide x 4’ tall
b. Pump assemblies – pumps need to be within 6’ of the
system to allow connection with the connection hose
included (longer connections can be run with customer
supplied hose)
c. Electrical enclosures
i. Motor starter – 12” wide x 6” tall
Tsunami Applicator Specifications
1. Operating water pressure: 100 psi max
2. Pneumatics inlet pressure: 0-60 psi
3. Maximum water source temperature 140° F
4. Operating ambient temperature: 40-120° F
5. Water supply
a. 4 – 10 GPM
b. Inlet pressure: 10-65 psi
6. Air supply
a. 3x 3/8” feed per system
b. ~20 CFM
7. Solution lines
a. 3x ½” poly lines to applicator
8. Mounting requirements
a. This Applicator requires an arch to mount to
b. Must be level
c. Recommend roughly centering on car or slightly favoring the driver’s side
1. At the ChemFlex injector, remove the chemical feed line from the injector hose barb.
2. Attach the tubing of the vacuum gauge to the ChemFlex hose barb
3. With the pump(s) on, manually activate the chemical that is to be tested at the main car wash control cabinet. An
injector that is working properly will have a reading greater than or equal to (≥) 20 in Hg
4. If injector is not functioning:
a) Check metering tip for clogs (can be tested with no metering tip to ensure injector is performing)
b) Try smaller injector (this will produce less flow and thus less backpressure)
5. Repeat steps 2-4 for each chemical lane that a vacuum reading is needed for.
6. Once testing is complete, turn off the AQUA-LAB pump from the main car wash control cabinet.
There is a variation of performance in the injectors that comes from slight variations in the dimensions of the parts and in assembly that are
unavoidable. It is common to see the resultant vacuum range from 20 in Hg all the way up to 28. There is also variation in the through hole size on the
meter tips from Dema (within their manufacturing tolerances). Using the same tip color from site to site is a good starting point. However with the
potential for variation from part to part it is reasonable to still need to do some adjustments from there.
Check all in-line filters and inlet plumbing for restrictions. Check
valves and backflow preventers
Inadequate water supply
Check pressure on inlet side of pump to be sure positive
pressure is maintained
Undersized pipingReplace with larger piping
Leak on the Inlet sideMake sure connections are tight
Worn or defective pump partsReplace worn parts or entire pump, Clean parts if required
Blown fuse or circuit breakerReplaced fuses or close circuit breaker
Defective Motor Starter
contactor
Replaced motor starter contactor
Incorrect Motor VoltageVoltage must be within 10% of motor rated voltage
Defective motor Replace motor
Pump components damagedReplace worn part or entire pump
Not PrimedReprime pump
Pump not secured firmlySecure properly
Restricted InletClean or correct restriction
Water regulator fluttering
Try to adjust regulator down and then back up or replace
regulator
Cavitation (Sounds like marbles
in pump)
Increase inlet size
Worn mechanical sealReplace shaft seal
Worn o-ring sealsReplace
Injector is not
drawing chemical -
Passes Vacuum
Pressure check
Clogged chemical feed
Check chemical hose, foot valve, metering tip and hose barb
for debris or clogs
Too much back pressure on
injector
Perform back pressure check test outlined in Section IV if this
manual. If the result is higher than 66 psi then, use larger
tubing, or use a smaller flow injector
Clogged injector check valve
Blow compressed air through the chemical hose barb on the
injector to remove debris
Clogged injector nozzleRemove injector and blow out any debris with compressed air
Defective Injector
If Vacuum check fails but Back Pressure is less than 66 psi,
replace the injector
Debris in regulatorRemove regulator and clean out debris
Defective Check ValveReplace check valve if broken
Defective RegulatorReplace Regulator
Defective PumpReplace Pump
Incorrect Injector Flow Rate
Selection
Replace with desired injector size
System pressure too lowEnsure system pressure is set at 200 psi
Foam Generator PluggedEnsure cleaned and clear
Downstream pluming restrictive
Perform Back pressure test outlined in Section IV of this
manual, if over 66 psi, increase tube size and reduce elbows,
turns or other restrictive plumbing
Valve malfunctionEnsure valve is receiving at least 60 psi
Valve may be assembled incorrectly
Disassemble valve and clean out debris (See valve
replacement instructions)
Clogged InjectorRemove injector and blow out debris with compressed air
No water supplyCheck that the system has a supply of water
Valve stuck open –
Staying open when
signal is off
Debris in valve seatRemove and clean valve (See valve replacement instructions)
Starter Tripping
Incorrect Overload Setting
Verify overload setting versus settings found in specs page of
this document
Hydra-Flex Inc warrants its equipment to be free from defect in material or workmanship under proper
normal proper use for a period of one (1) year beginning the date of purchase.
The Hydra-Flex Inc’s liability shall be limited to repair or replacement of parts found to be defective
within the warranty period and following Hydra-Flex Inc’s inspection. Hydra-Flex Inc shall have the
option requiring the return of defective material to establish the purchaser’s claim. In the event of
repair or replacement this limited warranty is non-cumulative. Neither labor nor transportation charges
are included in this warranty.
This warranty is based upon the proper care and maintenance of the warranted equipment. Warranty
does not apply if the merchandise is altered or modified in any way. Warranty does not apply to any
equipment which has been subject to misuse, inappropriate use of tools, including exposure to harsh
chemicals, neglect, lack of maintenance, freezing, fluid hammer, accident, third party damage, fluid
impurities such as sand or minerals, acts of God or acts of war. Nor does it apply to any equipment
which has been repaired or altered by anyone not so authorized by Hydra-Flex Inc. All equipment
must be properly installed in accordance with specified plumbing, electrical, and mechanical
requirements. The warranty does not apply to normal wear and tear or routine maintenance
components as described in the equipment manual.
Except as expressly stated herein, Hydra-Flex Inc shall not be liable for damages of any kind in
connection with the purchase, maintenance, or use of this equipment including loss of profits and all
claims for consequential damages. This limited warranty is in lieu of all other warranties expressed or
implied. Hydra-Flex Inc neither assumes nor authorizes any person to assume for it any other
obligation or liability in connection herewith. This warranty is neither assignable nor transferable.
Transportation damage claims are to be submitted to the carrier of the damaged material.